No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
E5072-90000
Edition
Edition 4, May 2019
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight
products may be downloaded from
the Web. Go to
http://www.keysight.com/go/confor
mity. You can then search by product
number to find the latest Declaration
of Conformity.
perform, display, or disclose
commercial computer software or
commercial computer software
documentation. No additional
government requirements beyond
those set forth in the EULA shall
apply, except to the extent that those
terms, rights, or licenses are
explicitly required from all providers
of commercial computer software
pursuant to the FAR and the DFARS
and are set forth specifically in
writing elsewhere in the EULA.
Keysight shall be under no obligation
to update, revise or otherwise modify
the Software. With respect to any
technical data as defined by FAR
2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the
U.S. government acquires no greater
than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data.
Printed in Malaysia
Published by:
Keysight Technologies International
Japan G.K,
1-3-3 Higashikawasaki-cho
Chuo-ku
Kobe-shi, Hyogo, Japan
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
U.S. Government Rights
The Software is “commercial
computer software,” as defined by
Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202, the
U.S. government acquires
commercial computer software
under the same terms by which the
software is customarily provided to
the public. Accordingly, Keysight
provides the Software to U.S.
government customers under its
standard commercial license, which
is embodied in its End User License
Agreement (EULA), a copy of which
can be found at
http://www.keysight.com/find/sweu
la The license set forth in the EULA
represents the exclusive authority by
which the U.S. government may use,
modify, distribute, or disclose the
Software. The EULA and the license
set forth therein, does not require or
permit, among other things, that
Keysight: (1) Furnish technical
information related to commercial
computer software or commercial
computer software documentation
that is not customarily provided to
the public; or (2) Relinquish to, or
otherwise provide, the government
rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release,
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Page 3
Caution
Keysight E5072A Network Analyzer
Installation Guide
Do not exceed the operating input power, voltage, and current level and signal
type appropriate for the instrument being used, refer to your instrument’s
Function Reference.
Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in
your instrument. ESD damage is most likely to occur as the test fixtures are
being connected or disconnected. Protect them from ESD damage by wearing
a grounding strap that provides a high resistance path to ground. Alternatively,
ground yourself to discharge any static charge built-up by touching the outer
shell of any grounded instrument chassis before touching the test port
connectors.
Safety Summary
When you notice any of the unusual conditions listed below, immediately
terminate operation and disconnect the power cable.
Contact your local Keysight Technologies sales representative or authorized
service company for repair of the instrument. If you continue to operate
without repairing the instrument, there is a potential fire or shock hazard to the
operator.
•Instrument operates abnormally.
•Instrument emits abnormal noise, smell, smoke or a spark-like light during
•Instrument generates high temperature or electrical shock during
•Power cable, plug, or receptacle on instrument is damaged.
•Foreign substance or liquid has fallen into the instrument.
operation.
operation.
3
Page 4
-
Caution
Manufacturer’s Declaration
Manufacturer’s Declaration
Herstellerbescheinigung
GERA- USCHEMISSION
LpA < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer’s Declaration
ACOUSTIC NOISE EMISSION
LpA < 70 dB
operator position
normal operation
per ISO 7779
Regulatory Compliance Information
This product complies with the essential requirements of the following
applicable European Directives, and carries the CE marking accordingly:
•The Low Voltage Directive 2006/95/EC
•The EMC Directive 2004/108/EEC
To obtain Declaration of Conformity, please contact your local Keysight
Technologies sales office, agent or distributor.
4 Keysight E5072A Network Analyzer
Page 5
Caution
Safety Notice Supplement
Safety Notice Supplement
•This equipment complies with EN/IEC61010-1:2001.
•This equipment is of MEASUREMENT CATEGORY I (CAT I). Do not use for
CAT II, III, or IV.
•Do not connect the measuring terminals to mains.
•This equipment is a POLLUTION DEGREE 2, INDOOR USE product.
•This equipment is tested in stand-alone condition and in combination with
the accessories supplied by Keysight Technologies against the requirement
of the standards described in the Declaration of Conformity. If it is used as a
system component, compliance of related regulations and safety
requirements are to be confirmed by the builder of the system.
Keysight E5072A Network Analyzer 5
Page 6
-
Caution
General Safety Precautions
General Safety Precautions
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific WARNINGS elsewhere in this manual may impair
the protection provided by the equipment. Such noncompliance would also
violate safety standards of design, manufacture, and intended use of the
instrument. Keysight Technologies assumes no liability for the customer’s
failure to comply with these precautions.
The E5072A complies with INSTALLATION CATEGORY II as well as
POLLUTION DEGREE 2 in IEC61010-1. The E5072A is an INDOOR USE
product.
The LEDs in the E5072A are Class 1 in accordance with IEC60825-1,
CLASS 1 LED PRODUCT.
•Ground the Instrument
To avoid electric shock, the instrument chassis and cabinet must be
grounded with the supplied power cable’s grounding prong.
•Do NOT Operate in an Explosive Atmosphere
Do not operate the instrument in the presence of inflammable gases or
fumes. Operation of any electrical instrument in such an environment
clearly constitutes a safety hazard.
•Keep Away from Live Circuits
Operators must not remove instrument covers. Component replacement
and internal adjustments must be made by qualified maintenance
personnel. Do not replace components with the power cable connected.
Under certain conditions, dangerous voltage levels may remain even after
the power cable has been disconnected. To avoid injuries, always
disconnect the power and discharge circuits before touching them.
•DO NOT Service or Adjust the Instrument Alone
Do not attempt internal service or adjustment unless another person,
capable of rendering first aid and resuscitation, is present.
•Do NOT Substitute Parts or Modify the Instrument
To avoid the danger of introducing additional hazards, do not install
substitute parts or perform unauthorized modifications to the instrument.
Return the instrument to a Keysight Technologies Sales and Service Office
for service and repair to ensure that safety features are maintained in
operational condition.
6 Keysight E5072A Network Analyzer
Page 7
Safety Symbols
Caution
Safety Symbols
•Dangerous Procedure Warnings
Warnings, such as the example below, precede potentially dangerous
procedures throughout this manual. Instructions contained in the warnings
must be followed.
Dangerous voltage levels, capable of causing death, are present in this
instrument. Use extreme caution when handling, testing, and adjusting
this instrument.
General definitions of safety symbols used on the instrument or in manuals are
listed below.
Instruction Manual symbol: the product is marked with this
symbol when it is necessary for the user to refer to the instrument
manual.
Certification
Keysight Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Keysight Technologies
further certifies that its calibration measurements are traceable to the United
States National Institute of Standards and Technology, to the extent allowed
by the Institution’s calibration facility or by the calibration facilities of other
International Standards Organization members.
Alternating current.
Direct current.
On (Supply).
Off (Supply).
A chassis terminal; a connection to the instrument’s chassis,
which includes all exposed metal structure.
Standby.
Keysight E5072A Network Analyzer 7
Page 8
-
Caution
Safety Symbols
Documentation Warranty
The material contained in this document is provided “as is,” and is subject to
being changed, without notice, in future editions. Further, to the maximum
extent permitted by applicable law, Keysight disclaims all warranties, either
express or implied with regard to this manual and any information contained
herein, including but not limited to the implied warranties of merchantability
and fitness for a particular purpose. Keysight shall not be liable for errors or for
incidental or consequential damages in connection with the furnishing, use, or
performance of this document or any information contained herein. Should
Keysight and the user have a separate written agreement with warranty terms
covering the material in this document that conflict with these terms, the
warranty terms in the separate agreement will control.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive remedies.
Keysight Technologies shall not be liable for any direct, indirect, special,
incidental, or consequential damages, whether based on contract, tort, or any
other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements
are available for Keysight Technologies products.
For any assistance, contact your nearest Keysight Technologies Sales and
Service Office. Addresses are provided at the back of this manual.
8 Keysight E5072A Network Analyzer
Page 9
Caution
Manuals for E5072A
Manuals for E5072A
Keysight provides the following three manuals for E5072A. The latest version of
all documentations can be downloaded from
http://www.keysight.com/manual/e5072a.
Installation Guide
The installation guide (this manual) provides start up setup information when
you use the E5072A for the first time and troubleshooting information when the
Windows cannot be boot up. See this manual first when you use the E5072A
for the first time.
Online Help
The online help provides the information about the quick start, measurement
operation, programming, built-in VBA, I/O interface This is pre-installed in the
E5072A. Press [Help] hard key on the front panel to open. Quick Start helps in
understanding the E5072A operation quickly.
The latest version of online help is available at:
http://www.keysight.com/manuals/e5072a.
The online help has context sensitive help, which is a great feature of the
E5072A help. It allows you to get information about the selected softkey by
pressing the Help key in the E5072A or by pressing F1 in a keyboard attached
to the E5072A or by clicking the help button in a dialog box. With context
sensitive help, users can receive information quickly about the area the user is
working in the firmware of the E5072A. It provides information relevant to the
task that needs to be accomplished and reduces the time to search relevant
information required to complete a task.
Service Guide
The service manual provides information about the parts, troubleshooting,
performance test, adjustment and service menu.
This chapter provides information on how to set up the Keysight E5072A
microwave downconverter.
Checking the Shipment
After you receive the analyzer, carry out checks during unpacking according to
the following procedure.
When unpacking the analyzer, if the exterior of the analyzer (such as the
cover, front/rear panel, LCD screen, power switch, and port connectors)
appear to be damaged during transport, do not turn on the power switch.
Otherwise, you may get an electric shock.
Step 1. Check that the packing box or shock- absorbing material used to package the
analyzer has not been damaged.
If the packing box or shock-absorbing material has been damaged, leave
the packing box and shock-absorbing material as is until other inspection
items are checked as follows.
Step 2. Check the packaged items supplied with the analyzer for any damage or
defects.
Step 3. By referring to the furnished contents list, check that all packaged items
supplied with the analyzer have been received as per the specified options.
Step 4. After checking, if any of the following applies, contact your nearest Keysight
Technologies sales and service office.
1. The packing box or shock- absorbing material used to package the analyzer
has been damaged or the shock- absorbing material has traces where
extreme force has been applied.
2. A packaged item supplied with the analyzer has mechanical damage or
defects.
13
Page 14
1-
Installation
Environmental Requirements
3. A packaged item supplied with the analyzer is missing.
4. A fault has been detected in the subsequent operation check of the
analyzer.
If an abnormality is detected in step 1, contact the company that transported
the analyzer as well as your nearest Keysight Technologies sales and service
office. For inspection by the transport company, save the packing box, shockabsorbing material, and packaged items as you received them.
Environmental Requirements
Set up the E5072A where the following environmental requirements are met.
Operating Environments
Ensure that the operating environment meets the following requirements.
Temp erat ure5 ºC to 40 ºC
Temperature range at the
error-correction
Humidity20% to 80% at wet bulb temperature < +29 °C
Altitude0 to 2,000 m (0 to 6,561 feet)
Vibration0.21 G maximum, 5 Hz to 500 Hz
Above environmental requirements are NOT for the specifications and
measurement accuracy of the analyzer, but for the operating environment
of the analyzer.
Ventilation Requirement
To ensure safety requirements, the specifications and measurement accuracy
of the analyzer, you must keep the environmental temperature within the
specified range by providing appropriate cooling clearance around the analyzer
or, for the rack mount type, by forcefully air- cooling inside the rack housing.
For more information on environmental temperature to satisfy the
specifications and measurement accuracy of the analyzer, see the specification
in the E5072A Online Help.
23 ºC ± 5 ºC (< 1 ºC deviation from the temperature
when performing the error-correction)
(non-condensation)
When the environmental temperature around the analyzer is kept within the
temperature range of the operating environment specification (See Section
“Operating Environments” on page 14), the analyzer conforms to the
14 Keysight E5072A Network Analyzer
Page 15
Installation
Environmental Requirements
requirements of the safety standard. Furthermore, under that temperature
requirement, the analyzer still conforms to the requirements of the safety
standard even when placing the analyzer with cooling clearance as follows:
Requirements
Back≤ 180mm
Sides≤ 60 mm (both right and left)
Figure 1-1 Ventilation space at the installation
Place the E5072A in a proper position as show in Figure 1-1.
Keysight E5072A Network Analyzer 15
Page 16
1-
Installation
Environmental Requirements
Protection Against Electrostatic Discharge (ESD)
Set up a static-safe work-station to protect the electronic components from
the damage by the electrostatic discharge (ESD) as shown in Figure 1-2.
Figure 1-2 Example of the static-safe work station
Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of
Power Cable in Case of Emergency
As described in “Disconnection from Supply Source" on page 31, the power
supply is disconnected by removing the power cable’s connector plug from
either the AC outlet or the E5072A unit. When installing the E5072A, ensure
that there is sufficient free space around the unit to permit quick disconnection
of the plug (from AC outlet or E5072A unit) in case of emergency.
16 Keysight E5072A Network Analyzer
Page 17
Installation
Installing Front Handles/Rack Mounting Flanges
Installing Front Handles/Rack Mounting Flanges
The E5072A can be installed on a workbench or in a rack. This section
describes how to install the front handles (Option 1CN) used for moving or
transporting the instrument, and how to install the analyzer in an equipment
rack as part of a measurement system (Option 1CM: without the handles,
Option 1CP: with the handles).
Installation
Installing Front Handles/Rack Mounting Flanges
Figure 1-3 Installing front handle/rack-mount kits
How to Install the Handle Kit (Option 1CN)
The handle kit is used for transport and relocation of the E5072A. While
referring to Figure 1-3 install the handle kit by following these steps.
Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame
of the E5072A front panel.
Step 2. Use the screws provided screws to mount the front handles (2) on each side
of the E5072A front panel frame.
Step 3. Attach the modified trim strip (3) provided to each front handle in order to
cover the front panel locking screws.
If the installed front handle becomes damaged, replace it with a new one
immediately. A damaged handle can break while moving or lifting the
instrument and cause personal injury or damage to the instrument.
18 Keysight E5072A Network Analyzer
Page 19
Installation
Installing Front Handles/Rack Mounting Flanges
How to Install the Rack-mount Kit (Option 1CM)
The rack- mount kit includes two flanges (locking side plates) for mounting the
E5072A on a rack (482.6 mm/19 inches) conforming to the EIA Standard. While
referring to Figure 1-3, install the rack- mount kit by following these steps.
Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame
of the E5072A front panel.
Step 2. Use the screws provided to mount a rack- mounting flange (4) on each side of
the E5072A front panel frame.
Step 3. Remove the four bottom feet of the E5072A (lift the bar marked TAB on the
inner side of the foot and slide the foot toward the bar).
Step 4. Mount the E5072A on the rack.
How to Install the Rack-mount and Handle Kit (Option 1CP)
The rack- mount and handle kit includes both the rack- mounting flanges
(locking side plates) and front handles. While referring to Figure 1-3, install the
rack- mount kit by following these steps.
Step 1. Remove the adhesive- backed trim strip (1) from each side of the outer frame
of the E5072A front panel.
Step 2. Use the screws provided to mount a front handle (2) and rack- mounting
flange (4) on each side of the E5072A front panel frame.
Use both a front handle and a rack-mounting flange in the same time
certainly. Do not attempt to install flanges or handles separately with
hardware provided. Serious electrical damage will occur to the instrument.
Step 3. Remove the four bottom feet of the E5072A (lift the bar marked TAB on the
inner side of the foot and slide the foot toward the bar).
Step 4. Mount the E5072A on the rack.
Keysight E5072A Network Analyzer 19
Page 20
1-
Installation
Connecting the Accessories
Connecting the Accessories
You need to connect some accessories before use.
Connecting the BNC Adaptor (Option 1E5)
When the E5072A is equipped with Option 1E5 (high- stability frequency
standard), connect the BNC adaptor from the Ref Oven terminal to the Ref In
terminal on the rear panel as shown in Figure 1-4 The BNC adaptor is included
in Option 1E5. When Option 1E5 is installed, the frequency accuracy and
stability of the E5072A improves further.
Figure 1-4 Connecting the BNC adaptor
Connecting Mouse and Keyboard
The E5072A allows you to connect mouse and/or keyboard through USB. USB
mouse and keyboard can be connected with the USB ports on the front or rear
panels.
Initial registration of the E5072A requires the mouse and keyboard before
turning on the power.
20 Keysight E5072A Network Analyzer
Page 21
Power Supply
Verification of the Power Supply
Installation
Power Supply
Before turning on the E5072A power, check the following.
Confirm that the power supplied to the E5072A meets the following
requirements:
Requirements
Voltage
90 to 132 VAC or 198 to 264 VAC
1
Frequency47 to 63 Hz
Maximum Power consumption350 VA
1. Switched automatically by the E5072A in conformity to the voltage used.
Verification and Connection of Power Cable
The three- wire power cable attached to the E5072A has one wire serving as a
ground. Using this power cable allows the E5072A to be grounded, thereby
protecting you against electrical shock from the power outlet.
Step 1. Confirm that the power cable is not damaged.
NEVER use a power cable showing any sign of damage. Faulty cables can
cause electrical shock.
Step 2. Use the supplied cable to connect between the power cable receptacle
(Figure 5 on page 30) on the rear panel of the E5072A and a three- wire power
outlet with the grounding prong firmly connected in the ground slot.
Use the supplied power cable with grounding wire to securely ground the
E5072A.
Power cord information is available on
https://www.keysight.com/find/powercords.
Keysight E5072A Network Analyzer 21
Page 22
1-
Installation
Starting the E5072A
Starting the E5072A
This section describes how to turn on/off the E5072A power and how to cut off
the power supply in an emergency.
Turning the Power ON and OFF
The standby switch can turn/off the E5072A. The color on the button shows
the status, as shown below:
Indicator ColorDescription
GreenNormal power on status
OrangeStandby status
RedIllegal power on status
Perform the following steps to turn the power ON or OFF.
Turning the Power ON
Step 1. Confirm if the Line Switch on the rear panel is on. The switch should always
be turned on.
Step 2. Press the standby switch once and it changes to green color.
This operation turns ON the power, and the E5072A starts the self- test
Step 3. Confirm that the self- test indicates normal operation.
Normal operation is confirmed by the self- test if no error message appears.
Turning the Power OFF
Step 1. To turn off the power of the E5072A, first, press this standby switch or send a
shutdown command from the external controller to activate the shutdown
process (the process of software and hardware necessary to turn off the
power supply). This puts the E5072A into the standby state and the button
changes to orange color.
22 Keysight E5072A Network Analyzer
Page 23
Installation
Starting the E5072A
Step 2. Next, if necessary, turn off power supply to the power cable receptacle (Line
Switch) on the rear panel.
Under normal use, never directly interrupt the power supply to the power cable receptacle on the rear panel when the power supply is on. Always
keep the Line Switch ON and never turn it.
If you directly interrupt the power supply to the power cable receptacle
when the power supply is on, or turn off the Line Switch, the shutdown
process will not work. This could damage the software and hardware of
the E5072A and lead to device failure.
Turning on the power supply after a faulty shutdown may cause the
system to start up in a condition called “safe mode.” If this occurs, first
shut down the system to set it to the standby state and then turn on the
power supply again to start up the system in normal mode.
Figure 1-5 Line switch (Always ON) and power cable receptacle
Disconnection from Supply Source
The power supply of the E5072A is cut off by disconnecting the plug of the
power cable (on either AC outlet side or E5072A side). When it is necessary to
disconnect the power supply in order to avoid shock hazards, etc., pull out the
power cable plug from either the AC outlet side or the E5072A side.
Keysight E5072A Network Analyzer 23
Page 24
1-
Installation
Starting the E5072A
To allow this operation to be performed smoothly, be sure to follow the
guidelines in “Ensuring Adequate Free Space around Analyzer for
Immediate Disconnection of Power Cable in Case of Emergency” on
page 16.
When turning the power OFF under normal circumstances, always follow
the methods described in “Turning the Power OFF” on page 22.
24 Keysight E5072A Network Analyzer
Page 25
Installation
Initial Registration of E5072A
Initial Registration of E5072A
When you start up the E5072A for the first time, you need to perform the initial
registration of the Windows OS operating system of the E5072A.
You cannot use the front panel keys during the initial registration of the
E5072A therefore connect the mouse and the keyboard before turning on
the power.
If you perform the following procedure incorrectly, a message will prompt
to give you the option to return to the previous registration screen and
perform the registration. In such case, follow the instruction to return to
the previous registration screen.
Failure to comply with the illustrated procedures may cause serious
damage to the E5072A.
Windows Embedded Standard 7
Step 1. Turns on the E5072A.
Step 2. Wait for several minutes for initial registration procedure. (Windows may
reboot a few times.) The license agreement dialog box as shown in Figure?
appears. Click “Accept” if you agree it.
Figure 1-6 License Agreement
Step 3. After a while, the E5072A measurement application starts automatically.
Keysight E5072A Network Analyzer 25
Page 26
1-
Installation
Initial Registration of E5072A
Windows XP
Step 1. Turn on the E5072A. Do not touch any key on the front panel during boot up.
Step 2. The screen as shown in the Figure 1-7 appears. Select Windows XP
Professional and press [Enter].
Figure 1-7 System start up screen (1/2)
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 3. The screen as shown in the Figure 1-8 appears. Press [Enter].
Figure 1-8 System start up screen (2/2)
26 Keysight E5072A Network Analyzer
Page 27
Installation
Initial Registration of E5072A
Step 4. The windows start up screen as shown in Figure 1-9 appears.
Figure 1-9 Windows start up screen
Step 5. In the License agreement dialog box, select the I accept this agreement box
and click the Next > button. Figure 1-10 shows the license agreement dialog
box.
Figure 1-10 License agreement dialog box
Step 6. In the next dialog box, set the date and time for your Windows. Then, click
the Next> button (Figure 1-11).
Keysight E5072A Network Analyzer 27
Page 28
1-
Installation
Initial Registration of E5072A
Figure 1-11 Date and Time Setting dialog box
Step 7. After a while, the E5072A restarts automatically.
Calibration of the Touch Screen
When E5072A measurement screen appears, you have to calibrate the touch
screen. Follow the procedure described below to calibrate the touch screen.
Step 1. Press .
Step 2. Press Service Menu.
Step 3. Press Test Menu.
Step 4. Press Adjust Touch Screen.
Step 5. Touch the symbol on the upper left of the screen with your finger. The
symbols also appears on the upper right, lower left, and lower right of the
screen. Touch all the symbols in that order with your finger.
Touching all the four locations described above with your finger automatically
concludes the touch screen calibration.
If you do not start the touch screen calibration within several seconds, the
previous measurement screen re-appears automatically.
28 Keysight E5072A Network Analyzer
Page 29
Keysight E5072A Network Analyzer
Installation Guide
2Troubleshooting
This chapter describes the troubleshooting process during start up and the
procedure of the operating system (OS) recovery when the Windows OS has
been damaged.
Troubleshooting during Startup
When you encounter problems during start up, see Table 2-1. System recovery
saves most of problems.
Table 2-1 Troubleshooting during startup
SymptomSolution
Turning on (|) the standby switch does not
start up the system.
The system starts up, but it automatically
shuts down immediately.
The system starts up, but it enters the
service mode (The status bar at the lower
right of the screen displays SVC in red).
The measurement screen appears after
start up, but the power-on test fails with
Error Message 241 prompting on a red
background in the instrument’s
message/warning area at the lower left of
the screen.
Confirm that the power cable is properly
plugged in.
Confirm that the line switch at the rear
panel is turned on.
Execute the system recovery.
29
Page 30
2-
Troubleshooting
System Recovery
System Recovery
By executing system recovery, you can return the E5072A’s system (the
Windows operating system and the firmware) to the factory state (at the time
1
of purchase
Strictly follow the steps described below. If you do any operation other
than the following steps, the system may not be recovered.
Notes on executing the factory recovery function
Executing the factory recovery function causes the following conditions:
•In addition to the Windows operating system and the firmware, the
following settings of the E5072A are returned to the factory state.
—Network setting
).
— GPIB setting
—Printer setting
•The driver for the supported printer which is installed after purchase is
deleted.
•You need to execute initial registration again.
Files that you have created using the save function (files in the D drive) are not
affected. However, Keysight recommends backing them up before executing
system recovery for precautionary purposes. For more information on backup,
refer to “Backing Up the Data” as described in E5072A Online Help.
1. The default setting of the hard disk is recovered during the system recovery. For the
hard disk that had failed earlier and had been replaced (after purchase), its default
setting may not be the same as the factory state.
30 Keysight E5072A Network Analyzer
Page 31
Troubleshooting
System Recovery
Procedure to execute the factory recovery
This section describes how to return the contents of the C drive to the factory
state.
You need a keyboard for this operation.
Step 1. Shut down the E5072A.
Step 2. Disconnect all of the USB device from the USB ports.
Step 3. Connect the keyboard to the E5072A.
Step 4. Press the standby switch of the E5072A to turn it on.
Step 5. When the screen as shown in the figure below appears, select Keysight
Recovery System and press [Enter].
Figure 2-1 Keysight Recovery System selection screen
After several seconds, the next screen appears automatically even if you
do not press any key.
Step 6. Type 2 to select Recover Factory Backup Image option and press [Enter].
Keysight E5072A Network Analyzer 31
Page 32
2-
Troubleshooting
System Recovery
Figure 2-2 System recovery selection screen
Step 7. A confirmation message appears. Click OK to continue the process.
Figure 2-3 System recovery confirmation screen
Step 8. The progress of the system configuration is displayed on the screen. The
recovery takes a few minutes depending on the amount of data.
Figure 2-4 System configuration progress screen
32 Keysight E5072A Network Analyzer
Page 33
Troubleshooting
System Recovery
Step 9. Once the recovery process is completed, completion message prompts. Click