Keysight Technologies E3633A, E3634A Service Manual

Keysight E3633A and E3634A DC Power Supplies
Service Guide
Notices
CAUTION
WARNING
Copyright Notice
© Keysight Technologies 1998 - 2019 No part of this manual may be repro-
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
E3634-90010
Edition
Edition 9, March 26, 2019
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, mod ify, reprod uce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight E3633A and E3634A Service Guide

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Earth (ground) terminal Frame or chassis (ground) terminal
Caution, risk of danger (refer to this manual for specific Warning or Caution information)
Alternating current (AC) Protective earth (ground) terminal
Out position of a bi-stable push control In position of a bi-stable push control
Positive binding post Negative binding post
Direct current (DC)
Keysight E3633A and E3634A Service Guide 3

Regulatory Markings

This symbol is a South Korean Class A EMC Declaration. This is a Class A instrument suitable for professional use and in electromagnetic environment outside of the home.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
The CSA mark is a registered trademark of the Canadian Standards Association.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
The RCM mark is a registered trademark of the Australian Communications and Media Authority.
4 Keysight E3633A and E3634A Service Guide

Waste Electrical and Electronic Equipment (WEEE) Directive

This instrument complies with the WEEE Directive (2002/96/EC) marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/powersupply
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight E3633A and E3634A Service Guide 5
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
6 Keysight E3633A and E3634A Service Guide

Table of Contents

Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . .5
Product category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
1Specifications
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Transient response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Command processing time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
OVP and OCP accuracy, ±(% of output + offset) . . . . . . . . . . . . . . . .17
Activation time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Remote sensing capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Temperature coefficient, ±(% of output + offset) . . . . . . . . . . . . . . .18
Stability, ±(% of output + offset) . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Output voltage overshoot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Programming language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
State Storage Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
Recommended calibration interval . . . . . . . . . . . . . . . . . . . . . . . . . .19
Output Terminal Isolation (maximum, from chassis ground) . . . . . .19
AC input ratings (selectable via rear panel selector . . . . . . . . . . . . .20
Maximum input power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Operating temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Weight (MY53xx6xxx) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Keysight E3633A and E3634A Service Guide 7
2Quick Start
Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
To prepare the Power Supply for Use . . . . . . . . . . . . . . . . . . . . . . . . . . 26
To Check the Rated Voltages of the Power Supply . . . . . . . . . . . . . . . . 29
To Check the Rated Currents of the Power Supply . . . . . . . . . . . . . . . . 30
To Use the Power Supply in Constant Voltage Mode . . . . . . . . . . . . . . 32
To Use the Power Supply in Constant Current Mode . . . . . . . . . . . . . . 35
To store and Recall the Instrument State . . . . . . . . . . . . . . . . . . . . . . . 37
To Program Overvoltage Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
To Program Overcurrent Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
To Rack Mount the Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
3 Calibration Procedures
Keysight Technologies Calibration Services . . . . . . . . . . . . . . . . . . . . . 50
Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Automating Calibration Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Test Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Performance Verification Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Performance verification tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Measurement Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Setup for most tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Electronic load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
General measurement techniques . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Current-Monitoring Resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
8 Keysight E3633A and E3634A Service Guide
Constant Voltage (CV) Verifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Constant voltage test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
Voltage programming and readback accuracy . . . . . . . . . . . . . . . . .57
CV load effect (load regulation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
CV source effect (line regulation) . . . . . . . . . . . . . . . . . . . . . . . . . . .60
CV PARD (ripple and noise) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Load transient response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
Constant Current (CC) Verifications . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Constant current test setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
Current programming and readback accuracy . . . . . . . . . . . . . . . . .63
CC load effect (load regulation) . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
CC source effect (line regulation) . . . . . . . . . . . . . . . . . . . . . . . . . . .66
CC PARD (ripple and noise) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
Common Mode Current Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
Performance Test Record for Keysight E3633A and E3634A . . . . . . . .69
CV performance test record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
CC performance test records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Calibration Security Code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
To unsecure the power supply for calibration . . . . . . . . . . . . . . . . . .72
To unsecure the power supply without the security code . . . . . . . .74
Calibration Count . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Calibration Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76
General Calibration/ Adjustment Procedure . . . . . . . . . . . . . . . . . . . . .77
Front panel voltage and current calibration . . . . . . . . . . . . . . . . . . .78
Aborting a Calibration in Progress . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
Calibration Record for Keysight E3633A/E3634A . . . . . . . . . . . . . . . . .84
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
An Example Program of Excel 97 for Calibration . . . . . . . . . . . . . . . . . .88
Keysight E3633A and E3634A Service Guide 9
4Theory of Operation
Block Diagram Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
AC Input and Bias Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
Floating Logic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
D-to-A Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
A-to-D Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
Power Mesh and Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Earth-Referenced Logic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
5Service
Operating Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Is the power supply inoperative? . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Does the power supply fail self-test? . . . . . . . . . . . . . . . . . . . . . . . 110
Types of Service Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Standard repair service (worldwide) . . . . . . . . . . . . . . . . . . . . . . . . 111
Express Exchange (U.S.A. only) . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Repacking for Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 112
Electrostatic Discharge (ESD) Precautions . . . . . . . . . . . . . . . . . . . . . 113
Surface Mount Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
To Replace the Power-Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
To Disconnect the Output Using an External Relay . . . . . . . . . . . . . . 114
Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
Troubleshooting Hints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Unit is inoperative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Unit reports errors 740 to 750 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Unit fails self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Bias supplies problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116
Self-Test Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Power-on self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Complete self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
10 Keysight E3633A and E3634A Service Guide
6 Component Drawings
E3633-60002/E3634-60002 Component Locator for
the Main Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
E3633-60003 Component Locator for the Front Panel . . . . . . . . . . . . 124
E3633-60019/E3634-60019 Component Locator for
the main board (serial MY53xx6xxx) . . . . . . . . . . . . . . . . . . . . . . . .125
Component Locator for the front panel (serial MY53xx6xxx) . . . . . . . .126
Keysight E3633A and E3634A Service Guide 11
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
12 Keysight E3633A and E3634A Service Guide

List of Figures

Figure 1-1 Dimensions of Keysight E3633A and E3634A
Figure 3-1 Performance verification test setup . . . . . . . . . . . . . . .54
Figure 3-2 Front panel terminal connections (side view) . . . . . . .55
Figure 3-3 Transient response time . . . . . . . . . . . . . . . . . . . . . . . .62
power supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Keysight E3633A and E3634A Service Guide 13
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
14 Keysight E3633A and E3634A Service Guide

List of Tables

Table 1-1 Performance specifications . . . . . . . . . . . . . . . . . . . . . 16
Table 1-2 Supplemental Characteristics . . . . . . . . . . . . . . . . . . . 18
Table 3-1 Recommended test equipment . . . . . . . . . . . . . . . . . .52
Table 3-2 Parameters for calibration . . . . . . . . . . . . . . . . . . . . . .77
Table 5-1 Bias supplies voltages . . . . . . . . . . . . . . . . . . . . . . . . 116
Table 5-2 Bias supplies voltages . . . . . . . . . . . . . . . . . . . . . . . . 116
Keysight E3633A and E3634A Service Guide 15
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
16 Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies Service Guide

1 Specifications

Performance Specifications 18
Supplemental Characteristics 20
The performance specifications are listed in the following pages. Specifications are warranted in the temperature range of 0 to 40°C with a resistive load. Supplemental characteristics, which are not warranted but are descriptions of performance determined either by design or testing. The Service Guide contains procedures for verifying the performance specifications.
17
1 Specifications

Performance Specifications

Tab le 1-1 Performance specifications
Parameter Keysight E3633A Keysight E3634A
Output Ratings
°C- 40 °C)
(@ 0
Programming Accuracy
12 months (@ 25 °C ± 5 °C), ±(% of output + offset)
Readback Accuracy
12 months (over GPIB and RS-232 or front panel with respect to actual output @ 25 °C ± 5 °C), ±(% of output + offset)
Ripple and Noise
(with outputs ungrounded, or with either output terminal grounded, 20 Hz to 20 MHz)
Load Regulation, ±(% of output + offset)
Line Regulation, ±(% of output + offset)
Programming Resolution Voltage 1 mV 3 mV
Readback Resolution Voltage 0.5mV 1.5 mV
Front Panel Resolution Voltage 1 mV
[a]
[a][b]
Low Range 0 to +8 V/ 0 to 20 A 0 to 25 V/ 0 to 7 A
High Range 0 to +20V/ 0 to 10 A 0 to+50 V/ 0 to 4 A
Voltage 0.05% + 10 mV
Current 0.2% + 10 mA
Voltage 0.05% + 5 mV
Current 0.15% + 5 mA
Normal mode voltage <0.35 mV rms and
3 mV p-p
Normal mode current <2 mA rms
Common mode current <1.5 uA rms
Voltage <0.01% + 2 mV
Current
Voltage <0.01% + 2 mV
Current <0.01% + 250 μA
Current 1 mA 0.5 mA
Current 1 mA 0.5 mA
Current 1 mA (< 10A), 10mA (. 10A)
<0.01% + 250 μA
<0.5 mV rms and 3 mV p-p
[a] Accuracy specifications are after an 1-hour warm-up with no load and calibration at 25 °C.
[b] This specification may degrade when the unit is subjected to an RF field >= 3V/meter.
18 Keysight E3633A and E3634A Service Guide

Transient response time

Less than 50 μsec for output to recover to within 15 mV following a change in output current from full load to half load or vice versa

Command processing time

Average time for output voltage to begin to change after receipt of digital data when the power supply is connected directly to the GPIB or RS-232 is less than 100 msec

OVP and OCP accuracy, ±(% of output + offset)

OVP 0.5% + 0.5 V
OCP 0.5% + 0.5 A

Activation time

Specifications 1
Average time for output to start to drop after OVP and OCP condition occurs.
OVP <1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
OCP <10 msec
Keysight E3633A and E3634A Service Guide 19
1 Specifications

Supplemental Characteristics

Tab le 1-2 Supplemental Characteristics
Parameter Keysight E3633A Keysight E3634A
Output Programming Range
(maximum programmable values)
Voltage Programming Speed:
Maximum time required for output voltage to settle within 1% of its total excursion (for resistive load). Excludes
command processing time.

Remote sensing capability

Low Range
High Range
OVP 1 V to 22 V 1 V to 55 V
OCP 0 A to 22 A 0 A to 7.5 A
Up 95 msec 45 msec 80 msec 100 msec
Down 30 msec 450 msec 30 msec 450 msec
0 to +8.24 V/ 0 to 20.6 A
0 to +20.6 V/ 0 to 10.3 A
Full Load No Load Full Load No Load
0 to +25.75 V/ 0 to 7.21 A
0 to +51.5V/ 0 to 4.12 A
Voltage drop Up to 0.7 V per each lead
Load regulation
Load Voltage Subtract voltage drop in load leads from specified output voltage rating.
Add 5 mV to spec for each 1-volt change in the + output lead due to load current changes

Temperature coefficient, ±(% of output + offset)

Maximum change in output/readback per °C after a 30-minute warm-up
Voltage 0.01% + 3 mV
Current 0.02% + 3 mA
20 Keysight E3633A and E3634A Service Guide

Stability, ±(% of output + offset)

Following a 30-minute warm-up, with the output in the ON state according to the operating mode (CC with load or CV), and with a change in the output over 8 hours under constant load, line, and ambient temperature.
Voltage 0.02% + 1mV
Current 0.1% + 1mA

Output voltage overshoot

During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if the output control is set to less than 1 V. If the output control is set to 1 V or higher, there is no overshoot.

Programming language

SCPI (Standard Commands for Programmable Instruments)
Specifications 1

State Storage Memory

Three (3) user-configurable stored states

Recommended calibration interval

1 year

Output Terminal Isolation (maximum, from chassis ground)

±60 Vdc when connecting shorting conductors without insulation to the (+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output to the (+) sense and the (-) output and the (-) sense terminals.
Keysight E3633A and E3634A Service Guide 21
1 Specifications

AC input ratings (selectable via rear panel selector

Maximum input power

700 VA with full load

Cooling

Fan cooled.

Operating temperature

0 to 40 °C for full rated output. At higher temperatures, the output current is derated linearly to 50% at 55 °C maximum temperature.
std 115 Vac ± 10%, 47 to 63 Hz
opt 0E3 230 Vac ± 10%, 47 to 63 Hz
opt 0E9 100 Vac ± 10%, 47 to 63 Hz

Storage temperature

-20 to 70 °C for storage environment.

Environmental Conditions

Designed for indoor use in an installation category II, pollution degree 2 environment. Designed to operate at a maximum relative humidity of 95 % and at altitudes of up to 2000 meters.

Dimensions

[1] See below for detailed information
22 Keysight E3633A and E3634A Service Guide
[1]
213 mmW x 133 mmH x 348 mmD (8.4 x 5.2 x 13.7 in)

Weight

Net 9.5 kg (21 lb)
Shipping 12 kg (26 lb)

Weight (MY53xx6xxx)

Net 11.4 kg (E3633A)
10.9 kg (E3634A)
Shipping 13.9 kg (E3633A)
13.4 kg (E3634A)
Specifications 1
Figure 1-1 Dimensions of Keysight E3633A and E3634A power supplies
Keysight E3633A and E3634A Service Guide 23
1 Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24 Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies Service Guide

2 Quick Start

Quick StartQuick Start 26
To prepare the Power Supply for Use 28
To Check the Rated Voltages of the Power Supply 31
To Check the Rated Currents of the Power Supply 32
To Use the Power Supply in Constant Voltage Mode 34
To Use the Power Supply in Constant Current Mode 37
To store and Recall the Instrument State 39
To Program Overvoltage Protection 41
To Program Overcurrent Protection 45
To Rack Mount the Power Supply 49
25
2Quick Start

Quick Start

One of the first things you will want to do with your power supply is to become acquainted with its front panel. Written procedures in this chapter prepare the power supply for use and familiarize you with most front-panel operations.
– The power supply is shipped from the factory configured in the front-panel
– The power supply has two output ranges. This feature allows more voltage at a
– When you press key (the Limit annunciator blinks), the display of the
operation mode. At power-on, the power supply is automatically set to operate in the front-panel operation mode. When in this mode, the front-panel keys can be used. When the power supply is in remote operation mode, you can return to front-panel operation mode at any time by pressing the (Local) key if you did not previously send the front-panel lockout command. A change between front-panel and remote operation modes will not result in a change in the output parameters
lower current or more current at a lower voltage. The desired output range is selected from the front panel or over the remote interfaces. The 8V or 20V for the E3633A and 25V or 50V for the E3634A annunciator indicates the presently selected range.
power supply goes to the limit mode and the present limit values will be displayed. In this mode, you can also observe the change of the limit values
when adjusting the knob. If you press the key again or let the display time-out after several seconds, the power supply will return the display to the meter mode (the Limit annunciator turns off). In this mode, the actual output voltage and current will be displayed.
– The output of the power supply can be enabled or disabled from the front
panel by pressing key. When the output is off, the OFF annunciator turns on and the output is disabled.
26 Keysight E3633A and E3634A Service Guide
Quick Start 2
– The display provides the present operating status of the power supply with
annunciators and also informs the user of error codes. For example, the power
[1]
supply is operating in CV mode in the 8V/20A
or 25V/7A
controlled from the front panel, then the CV and 8V
[1]
[2]
or 25V
range and
[2]
annunciators
will turn on. If, however, the power supply is remotely controlled, the Rmt
annunciator will also turn on, and when the power supply is being addressed
over HPIB interface, the Adrs annunciator will turn on.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 27
2Quick Start

To prepare the Power Supply for Use

The following steps help you verify that the power supply is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your power supply. If anything is missing, contact your nearest Keysight Technologies Sales Office.
– One power cord for your location.
– This User’s Guide.
– One Service Guide.
– Certificate of Calibration.
2 Verify that the correct power-line voltage setting is selected and that the
correct power-line fuse is installed.
The line voltage is set to 100, 115 or 230 Vac from the factory according to the input power option selected when you ordered the power supply. Change the voltage setting if it is not correct for your location (see the next page for detailed information). For 100 or 115 Vac operation, the correct fuse is 6.3 AT (Keysight part number 2110-1030) and for 230 Vac operation, the correct fuse is 3.15 AT (Keysight part number 2110-1031).
3 Connect the power cord and turn on the power supply.
A power-on self-test occurs automatically when you turn on the power supply. The front-panel display will light up while the power supply performs its power-on self- test. After performing the self-test, the power supply will go into the power-on / reset state; the output is disabled (the OFF annunciator
turns on); the 8V/20A
[1]
or 25V/7A
[2]
range is selected (the 8V
[1]
or 25V
[2]
annunciator turns on); the knob is selected for voltage control.
Notice that the OVP and OCP annunciators also turn on.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
28 Keysight E3633A and E3634A Service Guide
Quick Start 2
4 Enable the outputs.
Press key to enable the outputs. The OFF annunciator turns off and the
8V
[1]
or 25V
[2]
, OVP, OCP, and CV annunciators are lit. The blinking digit can be
adjusted by turning the knob. Notice that the display is in the meter mode.
“Meter mode” means that the display shows the actual output voltage and
current.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 29
2Quick Start
NOTE
1 Remove the power cord. Remove the fuse-holder assembly with a flat-blade screwdriver from the rear panel.
3 Rotate the power-line vol tage selector until the correct voltage appears.
2 Install the correct line fuse. Remove the power-line voltage selector from the power-line module.
4 Replace the power-line voltage selector and the fuse-holder assembly in the rear panel.
Install the correct fuse and verify that the correct line voltage appears in the window.
30 Keysight E3633A and E3634A Service Guide

To Check the Rated Voltages of the Power Supply

NOTE
The following procedures check to ensure that the power supply produces its rated voltage output with no load and properly responds to operation from the front panel.
For each step, use the keys shown on the left margins.
1Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
2 Enable the outputs.
The OFF annunciator turns off and the 8V
annunciators are lit. The blinking digit can be adjusted by turning the knob.
Notice that the display is in the meter mode. “Meter mode” means that the
display shows the actual output voltage and current.
[1]
or 25V/7A
[1]
or 25V
[2]
Quick Start 2
[2]
range is selected
OVP, OCP, and CV
3 Check that the front-panel voltmeter properly responds to knob control for
the 8V/20A
[1]
or 25V/7A
[2]
range.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero.
4 Ensure that the voltage can be adjusted from zero to the full rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob
until the voltmeter indicates “8.0 volts”
[1]
or “25.0 volts”
[2]
You can use the resolution selection keys to move the flashing digit to the right or left when setting the voltage.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 31
2Quick Start

To Check the Rated Currents of the Power Supply

The following steps check basic current functions with a short across the power supply’s output.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
[1]
(the OFF annunciator turns on); the 8V/20A (the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
2 Connect a short across (+) and (-) output terminals with an insulated test
lead.
3 Enable the output
The OFF annunciator turns off and the 8V annunciators turn on. The CV or CC annunciator is lit depending on the resistance of the test lead. The flashing digit can be adjusted by turning the knob. Notice that the display is in the meter mode. “Meter mode” means that the display shows the actual output voltage and current.
or 25V/7A
[1]
or 25V
[2]
[2]
range is selected
, OVP, and OCP
4 Adjust the voltage limit to 1.0 volt
Set the display to the limit mode (the Limit annunciator will be flashing). Adjust the voltage limit to 1.0 volt to assure CC operation. The CC annunciator will light.
5 Check that the front-panel ammeter properly responds to knob control.
Set the knob to the current control, and turn the knob clockwise or counter clockwise when the display is in the meter mode (the Limit annunciator is off). Check that the ammeter responds to knob control and the voltmeter indicates nearly zero (actually, the voltmeter will show the voltage drop caused by the test lead)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
32 Keysight E3633A and E3634A Service Guide
Quick Start 2
NOTE
6 Ensure that the current can be adjusted from zero to the full rated value.
Adjust the knob until the ammeter indicates 0 amps and then until the
ammeter indicates 20.0 amps
[1]
or 7.0 amps
[2]
.
If an error has been detected during the output checkout procedures, the ERROR annunciator will turn on. See “Error Messages’’ for more information,
starting on page 123 in chapter 5 of the User’s Guide.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 33
2Quick Start
NOTE

To Use the Power Supply in Constant Voltage Mode

To setup the power supply for constant voltage (CV) operation, proceed as follows.
For each step, use the keys shown on the left margin.
1 Connect a load to the output terminals
With the power-off, connect a load to the (+) and (-) output terminals.
2 Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled (the OFF annunciator turns on); the 8V/20A (the 8V
[1]
or 25V
control. Press
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or
[2]
key to operate the power supply in the 20V/
10A* or 50V/4A** range before proceeding to the next step. The 20V
[2]
annunciator turns on.
50V
3 Set the display to the limit mode
[1]
or 25V/7A
[2]
range is selected
[1]
or
Notice that the Limit annunciator blinks, indicating that the display is in the limit mode. When the display is in the limit mode, you can see the voltage and current limit values of the power supply.
In constant voltage mode, the voltage values between the meter and limit modes are the same, but the current values are not. Moreover, if the display is in the meter mode, you cannot see the change of current limit value when adjusting the knob. We recommend that you should set the display to “limit” mode to see the change of current limit value in the constant voltage mode whenever adjusting the knob.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
34 Keysight E3633A and E3634A Service Guide
Quick Start 2
4 Adjust the knob for the desired current limit.
Check that the Limit annunciator still blinks. Set the knob for current control.
The second digit of the ammeter will be blinking. The blinking digit can be
changed using the resolution selection keys and the blinking digit can be
adjusted by turning the knob. Adjust the knob to the desired current limit.
5 Adjust the knob for the desired output voltage.
Check that the Limit annunciator still blinks. Set the knob for voltage control.
The second digit of the voltmeter will be blinking. Change the blinking digit
using the resolution selection keys and adjust the knob to the desired output
voltage.
6 Return to the meter mode.
Press key or let the display time-out after several seconds to return to
the meter mode. Notice that the Limit annunciator turns off and the displays
shows “OUTPUT OFF” message.
7 Enable the output.
The OFF annunciator turns off and the 8V
[1]
(or 25V
[2]
) or 20V
[1]
(or 50V
OVP, OCP and CV annunciators are lit. Notice that the display is in the meter
mode. In the meter mode, the display shows the actual output voltage and
current.
[2]
),
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify that
the CV annunciator is lit. If the CC annunciator is lit, choose a higher current
limit.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 35
2Quick Start
NOTE
NOTE
During actual CV operation, if a load change causes the current limit to be exceeded, the power supply will automatically crossover to the constant current mode at the preset current limit and the output voltage will drop proportionately.
You can use the resolution selection keys to move the flashing digit to the right or left when setting the current.
36 Keysight E3633A and E3634A Service Guide

To Use the Power Supply in Constant Current Mode

NOTE
To set up the power supply for constant current (CC) operation, proceed as follows.
1 Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
[1]
Press
[2]
50V
or
annunciator turns on.
[2]
key before proceeding to the next step. The 20V
3 Set the display to the limit mode.
[1]
or 25V/7A
Quick Start 2
[2]
range is selected
[1]
or
Notice that the Limit annunciator blinks, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the selected supply.
In constant current mode, the current values between the meter mode and limit mode are the same, but the voltage values are not. Moreover, if the d isplay is in the meter mode, you cannot see the change of voltage limit value when adjusting the knob. We recommend that you should set the display to “limit” mode to see the change of voltage limit value in the constant current mode whenever adjusting the knob.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 37
2Quick Start
NOTE
4 Adjust the knob for the desired voltage limit.
5 Adjust the knob for the desired output current.
6 Return to the meter mode.
7 Enable the output.
Check that the Limit annunciator still blinks and the second digit of voltmeter blinks to indicate the knob is selected for voltage control.The blinking digit can be changed using the resolution keys and the blinking digit can be adjusted by turning the knob. Adjust the knob for the desired voltage limit.
Check that the Limit annunciator still blinks. Set the knob for current control. The second digit of the ammeter will be blinking. Change the blinking digit using the resolution selection keys and adjust the knob to the desired output current.
Press key or let the display time-out after several seconds to return to the meter mode. Notice that the Limit annunciator turns off and the display shows “OUTPUT OFF” message.
The OFF annunciator turns off and the 8V
[1]
(or 25V
[2]
) or 20V
[1]
(or 50V
[2]
),
OVP, OCP and CC annunciators are lit. Notice that the display is in the meter mode. In the meter mode, the display shows the actual output voltage and current.
8 Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that the CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage limit.
During actual CC operation, if a load change causes the voltage limit to be exceeded, the power supply will automatically crossover to constant voltage mode at the preset voltage limit and the output current will drop proportionately.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
38 Keysight E3633A and E3634A Service Guide

To store and Recall the Instrument State

You can store up to three different operating states in non-volatile memory. This also enables you to recall the entire instrument configuration with just a few key presses from the front panel.
The memory locations are supplied with the reset states from the factory for front-panel operation. Refer to the description of *RST command, starting on
page 84 in chapter 4 for more information. The following steps show you how to
store and recall an operating state.
1 Set up the power supply for the desired operating state.
The storage feature “remembers” output range selection, the limit value
settings of voltage and current, output on/off state, OVP and OCT on/off state
and OVP and OCP trip levels.
2 Turn on the storage mode.
Three memory locations (numbered 1, 2 and 3) are available to store the
operating states. The operating states are stored in non-volatile memory and
are remembered when being recalled.
Quick Start 2
STORE 1
This message appears on the display for approximately 3 seconds.
3 Store the operating state in memory location “3”.
Turn the knob to the right to specify the memory location 3.
STORE 3
To cancel the store operation, let the display time-out after about 3 seconds
or press any other function key except the ion key except the key. The
power supply returns to the normal operating mode and to the function
pressed. Save the operating state.
Keysight E3633A and E3634A Service Guide 39
2Quick Start
4 Save the operating state.
5 Turn on the recall mode.
6 Recall the stored operating state.
The operating state is now stored. To recall the stored state, go to the following steps.
DONE
Memory location “1” will be displayed in the recall mode.
RECALL 1
This message appears on the display for approximately 3 seconds.
Turn the knob to the right to change the displayed storage location to 3.
RECALL 3
If this setting is not followed within 3 seconds with key stroke, the power supply returns to normal operating mode and will not recall the instrument state 3 from memory.
7 Restore the operating state.
The power supply should now be configured in the same state as when you stored the state on the previous steps.
DONE
This message appears on the display for approximately 1 second.
40 Keysight E3633A and E3634A Service Guide

To Program Overvoltage Protection

Overvoltage protection guards the load against output voltages that reach a specified value greater than the programmed protection level. It is accomplished by shorting the output via an internal SCR when the trip level is set to equal or greater than 3 volts, or by programming the output to 1 volt when the trip level is set to less than 3 volts.
The following steps show how to set the OVP trip level, how to check OVP operation, and how to clear overvoltage condition.
Setting the OVP level and enable the OVP circuit
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
control.
2 Enable the output.
The OFF annunciator turns off and the display will go to the meter mode.
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or 25V/7A
Quick Start 2
[2]
range is selected
3 Enter the OVP menu and set the trip level.
LEVEL 22.0V (E3633A)
LEVEL 55.0V (E3634A)
You will see the above message on the display when you enter the OVP menu.
Adjust the control knob for the desired OVP trip level.
Note that you cannot set the trip levels to lower than 1.0 volt.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 41
2Quick Start
4 Enable the OVP circuit.
5Exit the OVP menu.
Checking OVP operation
To check OVP operation, raise the output voltage to near the trip point. Then very gradually increase the output by turning the knob until the OVP circuit trips. This will cause the power supply output to drop to near zero, the OVP annunciator to blink, and the CC annunciator to turn on. The “OVP TRIPPED” message also appears on the display.
OVP ON
You will see the above message after pressing key.
CHANGED
The “CHANGED” message is highlighted for a second to show that the new OVP trip level is now in effect. If the OVP settings are not changed, “NO CHANGE” will be displayed. The power supply will exit the OVP menu and the display will return to the meter mode. Check that the OVP annunciator turns on.
Clearing the overvoltage condition
When the OVP condition occurs (the “OVP TRIPPED” message is shown on the display), the OVP annunciator blinks. When it was caused by an external voltage source such as a battery, disconnect it first. Clear the overvoltage condition by adjusting output voltage level or by adjusting OVP trip level.
The following steps show how to clear the overvoltage condition and get back to normal mode operation. In the following steps, the display will go back to “OVP TRIPPED” if you let the display time out after about several seconds.
42 Keysight E3633A and E3634A Service Guide
Quick Start 2
– Adjust the output voltage level
1 Lower the output voltage level.
Lower the output voltage level below the OVP trip point after pressing key. The OVP and Limit annunciators are blinking.
2 Move to the clear mode.
OVP CLEAR
Press key twice to move to the OVP CLEAR mode. The “OVP ON” message appears on the display. Turn the knob to the right until the above message appears on the display.
3 Clear the overvoltage cond ition and exit this menu.
Now, when you press key again, the “DONE” message is displayed for a second and the OVP annunciator will not blink any more. The output will return to meter mode.
Keysight E3633A and E3634A Service Guide 43
2Quick Start
– Adjust OVP trip level
1Raise the OVP trip level.
Press key and turn the knob to raise the OVP trip level.
2 Move to the OVP CLEAR mode.
OVP CLEAR
Press key to move the OVP clear mode. The “OVP ON” message appears on the display. Turn the knob to the right until the above message appears on the display.
3 Clear the overvoltage condition and exit this menu.
Now, when you press key again, the “DONE’’ message is displayed for a second and the OVP annunciator will not blink any more. The output will return to the meter mode.
44 Keysight E3633A and E3634A Service Guide

To Program Overcurrent Protection

Overcurrent protection guards the load against output currents that reach a specified value greater than the programmed protection level. It is accomplished by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how to check OCP operation and how to clear overcurrent condition.
Setting the OCP level and enable the OCP circuit
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
control.
2 Enable the output.
The OFF annunciator turns off and the display will go to the meter mode.
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or 25V/7A
Quick Start 2
[2]
range is selected
3 Enter the OCP menu and set the trip level.
LEVEL 22.0 A (E3633A)
LEVEL 7.5 A (E3634A)
You will see the above message on the display when you enter the OCP menu.
Adjust the knob for the desired OCP trip level.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 45
2Quick Start
4Enable the OCP circuit
5 Exit the OCP menu.
Checking OCP operation
To check OCP operation, raise the output current to near the trip point. Then very gradually increase the output by turning the knob until the OCP circuit trips. This will cause the power supply’s output current to drop to zero and the OCP annunciator to blink. The “OCP TRIPPED” message also appears on the display.
OCP ON
You will see the above message after pressing the key.
CHANGED
The “CHANGED” message is displayed for a second to show that the new OCP trip level is now in effect. If the OCP settings are not changed, “NO CHANGE” will be displayed. The power supply will exit the OCP menu and the display will return to the meter mode. Check that the OCP annunciator turns on.
Clearing the overcurrent condition
When the OCP condition occurs (the “OCP TRIPPED” message is shown on the display), the OCP annunciator blinks. When it was caused by external voltage sources such as a battery, disconnect it first. Clear the overcurrent condition by adjusting output current level or by adjusting OCP trip level.
The following steps show how to clear the overcurrent condition and get back to normal mode operation. In the following steps, the display will go back to “OCP TRIPPED” if you let the display time out after about several seconds.
46 Keysight E3633A and E3634A Service Guide
Quick Start 2
– Adjust output current level
1 Lower the output current level.
Press key and set the knob for current control by pressing key then lower the output current level below the OCP trip point.
2 Move to the clear mode.
OCP CLEAR
Press key twice to move to the OCP CLEAR mode. The “OCP ON” message appears on the display. Turn the knob to the right until the above message appears on the display
3 Clear the overcurrent condition and exit this menu.
Now, when you press key again, the “DONE” message is displayed for a second and the OCP annunciator will not blink any more. The output will return to meter mode. The knob is selected for current control.
Notice that the power supply is operated in the constant current (CC) mode.
Keysight E3633A and E3634A Service Guide 47
2Quick Start
– Adjust OCP trip level
1 Raise the OCP trip level.
Press key and turn the knob to raise the OCP trip level.
2 Move to the OCP CLEAR mode.
OCP CLEAR
Press the key to move to the OCP CLEAR mode. The “OCP ON” message appears on the display. Turn the knob to the right until the above message appears on the display.
3 Clear the overcurrent condition and exit this menu.
Now, when you press again, the “DONE” message is displayed for a second and the OCP annunciator will not blink any more. The output will return the meter mode.
48 Keysight E3633A and E3634A Service Guide

To Rack Mount the Power Supply

The power supply can be mounted in a standard 19-inch rack cabinet using one of three optional kits available. A rack-mounting kit for a single instrument is available as Option 1CM (P/N 5063-9243). Installation instructions and hardware are included with each rack-mounting kit. Any Keysight System II instrument of the same size can be rack-mounted beside the Keysight E3633A or E3634A DC power supply.
Remove the front and rear bumpers before rack-mounting the power supply.
Quick Start 2
To remove the rubber bumper, pull the bumper off from the top as there are protrusions on the sides and bottom of the cover.
To rack mount a single instrument, order adapter kit 5063-9243.
Keysight E3633A and E3634A Service Guide 49
2Quick Start
To rack mount two instruments of the same depth side-by-side, order lock-link kit 5061-9694 and flange kit 5063-9214.
To install two instruments in a sliding support shel f, order support shel f 5063-9256, and slide kit 1494-0015.
50 Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies Service Guide

3 Calibration Procedures

Keysight Technologies Calibration Services 52
Calibration Interval 52
Automating Calibration Procedures 53
Test Considerations 53
Recommended Test Equipment 54
Performance Verification Tests 55
Measurement Techniques 56
Constant Voltage (CV) Verifications 59
Constant Current (CC) Verifications 65
Common Mode Current Noise 70
Performance Test Record for Keysight E3633A and E3634A 71
Calibration Security Code 73
Calibration Count 77
Calibration Message 78
General Calibration/ Adjustment Procedure 79
Aborting a Calibration in Progress 85
Calibration Record for Keysight E3633A/E3634A 86
Error Messages 87
An Example Program of Excel 97 for Calibration 90
This chapter contains procedures for verification of the power supply’s performance and calibration (adjustment).
The performance verification tests for constant voltage (CV) and constant current (CC) operations use the power supply’s specifications listed in “Specifications” on page 17.
51
3 Calibration Procedures
NOTE
If you calibrate the power supply over the remote interface, you must send the
*RST
command to the power supply or turn the power supply off and on again
after performing a calibration to ensure proper power supply operation.
Closed-Case Electronic Calibration
The power supply features closed-case electronic calibration since no internal mechanical adjustments are required for normal calibration. The power supply calculates correction factors based upon the input reference value you enter. The new correction factors are stored in non-volatile memory until the next calibration adjustment is performed. (Non-volatile memory does not change when power has been off or after a remote interface reset.)

Keysight Technologies Calibration Services

When your power supply is due for calibration, contact your local Keysight Technologies Service Center for a low-cost calibration. The Keysight E3633A and E3634A power supplies are supported on calibration processes which allow Keysight Technologies to provide this service at competitive prices.

Calibration Interval

The power supply should be calibrated on a regular interval determined by the accuracy requirements of your application. A 1-year interval is adequate for most applications. Keysight Technologies does not recommend extending calibration intervals beyond 1 year for any application. Keysight Technologies recommends that complete re-adjustment should always be performed at the calibration interval. This will increase your confidence that the Keysight E3633A and E3634A will remain within specification for the next calibration interval. This criteria for re-adjustment provides the best long-term stability.
52 Keysight E3633A and E3634A Service Guide

Automating Calibration Procedures

CAUTION
You can automate the complete verification procedures outlined in this chapter if you have access to programmable test equipment. You can program the instrument configurations specified for each test over the remote interface. You can then enter readback verification data into a test program and compare the results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote operation is similar to the local front-panel procedure. You can use a computer to perform the adjustment by first selecting the required setup. The calibration value is sent to the power supply and then the calibration is initiated over the remote interface. The power supply must be unsecured prior to initiating the calibration procedure. An example program of Excel 97 for calibration over the GPIB interface is listed at the end of this chapter.
For further details on programming the power supply, see chapters 3 and 4 in the Keysight E3633A and E3634A User’s Guide.

Test Considerations

Calibration Procedures 3
To ensure proper instrument operation, verify that you have selected the correct power-line voltage prior to attempting any test procedure in this chapter. See
page 30 in Chapter 2 for more information.
Ensure that all connections of terminals (both front panel and rear panel) are removed while the power supply internal self-test is being performed.
For optimum performance verification, all test procedures should comply with the following recommendations:
– Assure that the calibration ambient temperature is stable and between 20°C
and 30°C.
– Assure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Use short cables to connect test set-ups.
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages may be present at the outputs of the power supply.
Keysight E3633A and E3634A Service Guide 53
3 Calibration Procedures

Recommended Test Equipment

The test equipment recommended for the performance verification and adjustment procedures is listed below. If the exact instrument is not available, use the accuracy requirements shown to select substitute calibration standards. If you use equipment other than that recommended in Table 3-1, you must recalculate the measurement uncertainties for the actual equipment used.
Tab le 3-1 Recommended test equipment
Instrument Requirements Recommended
Model
GPIB controller Full GPIB or RS-232 capabilities Keysight 82341C
interface card
Oscilloscope 100 MHz with 20 MHz
bandwidth
RMS voltmeter 20 Hz to 20 MHz Measure rms ripple & noise
Coaxial cable Keysight 10502A Measure rms ripple & noise
BNC female to banana plug adapter
Digital voltmeter Resolution: 0.1 mV
Accuracy: 0.01%
Electronic load Voltage range: 60 Vdc
Current range: 60 Adc Open and short switches Transient on/off
Resistive loads (RL)
Current monitoring resistor (shunt)
(0.4 , 300 W/2.0 , 300 W) (3.5 , 300 W/12.5 , 300 W)
(0.01 , 0.1%)
[c]
Keysight 54602B Display transient response
Keysight 1251-2277
Keysight 34401A Measure dc voltages
Keysight 60502B Measure load and line
[a]
[b]
ISOTEK Co. Model: A-H
Programming and readback accuracy
and ripple & noise waveform
Measure rms ripple & noise
regulations and transient response time.
Measure rms ripple & noise
Constant current test setup
Use
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
[c] It is recommended to use a current monitoring resistor after calibration to find the accurate resistance.
54 Keysight E3633A and E3634A Service Guide

Performance Verification Tests

The performance verification tests use the power supply’s specifications listed in
“Specifications” on page 17.
You can perform two different levels of performance verification tests:
Self-Test
A series of internal verification tests that provide high confidence that the
power supply is operational.
Performance Verification Tests
These tests are used to verify that the power supply is operating as specified in
the performance specifications.

Self-test

A power-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the power supply is operational.
The complete self-test is enabled by holding down key (actually any front panel keys except key) as you turn on the power supply and hold down the
key until you hear a long beep. The self-test will begin when you release the key following the beep. The complete self-test takes approximately two seconds to execute.
You can also perform a self-test from the remote interface (see chapter 3 in the Keysight E3633A and E3634A User’s Guide).
– If the self-test is successful, “PASS” is displayed on the front panel.
Calibration Procedures 3
– If the self-test fails, “FAIL” is displayed and the ERROR annunciator turns on. If
repair is required, see chapter 5, “Service”, for further details.
– If self-test passes, you have a high confidence that the power supply is
operational.

Performance verification tests

These tests should be used to verify the power supply’s specifications following repairs to specific circuits. The following sections explain all verification procedures in detail. All of the performance test limits are shown in each test.
Keysight E3633A and E3634A Service Guide 55
3 Calibration Procedures

Measurement Techniques

Setup for most tests

Most tests are performed at the front terminals as shown in the following figure. Measure the dc voltage directly at the (+) and (–) terminals on the front panel.
Figure 3-1 Performance verification test setup

Electronic load

Many of the test procedures require the use of a variable load resistor capable of dissipating the required power. Using a variable load resistor requires that switches should be used to connect, disconnect, and short the load resistor. An electronic load, if available, can be used in place of a variable load resistor and switches. The electronic load is considerably easier to use than load resistors. It eliminates the need for connecting resistors or rheostats in parallel to handle power, it is much more stable than carbon-pile load, and it makes easy work of switching between load conditions as is required for the load regulation and load response tests. Substitution of the electronic load requires minor changes to the test procedures in this chapter.
56 Keysight E3633A and E3634A Service Guide

General measurement techniques

To achieve best results when measuring load regulation, peak to peak voltage, and transient response time of the power supply, measuring devices must be connected through the hole in the neck of the binding post at (A) while the load resistor is plugged into the front of the output terminals at (B). A measurement made across the load includes the impedance of the leads to the load. The impedance of the load leads can easily be several orders of the magnitude greater than the power supply impedance and thus invalidate the measurement. To avoid mutual coupling effects, each measuring device must be connected directly to the output terminals by separate pairs of leads.
Calibration Procedures 3
Figure 3-2 Front panel terminal connections (side view)

Current-Monitoring Resistor

To eliminate output current measurement error caused by the voltage drops in the leads and connections, connect the current monitoring resistor between the (-) output terminal and the load as a four-terminal device. Connect the current-monitoring leads inside the load-lead connections directly at the monitoring points on the resistor element (see R
Keysight E3633A and E3634A Service Guide 57
in Figure 3-1).
M
3 Calibration Procedures

Programming

Most performance tests can be performed from the front panel. However, an GPIB or RS-232 controller is required to perform the voltage and current programming accuracy and readback accuracy tests.
The test procedures are written assuming that you know how to program the power supply either from the front panel or from an GPIB or RS-232 controller. Complete instructions on front panel and remote programming are given in the
Keysight E3633A and E3634A User’s Guide.
58 Keysight E3633A and E3634A Service Guide

Constant Voltage (CV) Verifications

Constant voltage test setup

If more than one meter or if a meter and an oscilloscope are used, connect each to the (+) and (–) terminals by a separate pair of leads to avoid mutual coupling effects. Use coaxial cable or shielded 2-wire cable to avoid noise pick-up on the test leads.

Voltage programming and readback accuracy

This test verifies that the voltage programming and GPIB or RS-232 readback functions are within specifications. Note that the readback values over the remote interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test to avoid round off errors.
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(–) terminals of the output to be tested as shown in Figure 3-1
2 Turn on the power supply. Select the 20V/10A
enable the output by sending the commands:
Calibration Procedures 3
[1]
or 50V/4A
[2]
range and
VOLT:RANG P20V (E3633A)
VOLT:RANG P50V (E3634A)
OUTP ON
3 Program the output voltage to zero volt and current to full rated value
(10.0 A)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or (4.0 A)
VOLT 0
CURR 10 (E3633A)
CURR 4 (E3634A)
[2]
by sending the commands:
Keysight E3633A and E3634A Service Guide 59
3 Calibration Procedures
4 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (0 V ± 10 mV). Also, note that the CV, Adrs, Limit, and Rmt annunciators are on.
5 Readback the output voltage over the remote interface by sending the
command:
MEAS:VOLT?
6 Record the value displayed on the controller. This value should be within the
limit of (DVM ±5 mV).
7 Program the output voltage to full rated value (20.0 V)
sending the command:
VOLT 20.0 (E3633A)
VOLT 50.0 (E3634A)
8 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (20 V ± 20 mV)
9 Readback the output voltage over the remote interface by sending the
command:
[1]
or (50.0 V)
[1]
or (50 V ± 35 mV)
[2]
[2]
by
MEAS:VOLT?
10 Record the value displayed on the controller. This value should be within the
limit of (DVM ± 15 mV)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or (DVM ± 30 mV)
[2]
.
60 Keysight E3633A and E3634A Service Guide

CV load effect (load regulation)

This test measures the immediate change in the output voltage resulting from a change in the output current from full to no load.
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output as shown in Figure 3-1.
2 Turn on the power supply. Select the 20V/10A
output, and set the display to the limit mode. When the display is in the limit
mode, program the output current to the maximum programmable value and
the voltage to the full rated value (20.0 V)
3 Operate the electronic load in constant current mode and set its current to
[1]
(10.0 A)
or (4.0 A)
not lit, adjust the load so that the output current drops slightly until the CV
annunciator lights. Record the output voltage reading on the digital voltmeter.
4 Operate the electronic load in open mode (input off). Record the output
voltage reading on the digital voltmeter immediately. The difference between
the digital voltmeter readings in steps (3) and (4) is the CV load regulation. The
difference of the readings during the immediate change should be within the
limit of (4 mV)
[2]
[1]
or (7 mV)
. Check that the front panel CV annunciator remains lit. If
[2]
[1]
or 50V/4A
[1]
or (50.0 V)
Calibration Procedures 3
[2]
range, enable the
[2]
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 61
3 Calibration Procedures

CV source effect (line regulation)

This test measures the immediate change in output voltage that results from a change in ac line voltage from the minimum value (10% below the nominal input voltage) to maximum value (10% above the nominal input voltage).
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in Figure 3-1.
2 Connect the ac power line through a variable voltage transformer.
3 Turn on the power supply. Select the 20V/10A
the output, and set the display to the limit mode. When the display is in the limit mode, program the current to the maximum programmable value and the
voltage to full rated value (20.0 V)
4 Operate the electronic load in constant current mode and set its current to
(10.0 A) adjust the load so that the output current drops slightly until the CV annunciator lights.
5 Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the output reading on the digital voltmeter.
6 Adjust the transformer to high line voltage (127 Vac for nominal 115 Vac, 110
Vac for nominal 100 Vac, or 253 Vac for nominal 230 Vac). Record the voltage reading on the digital voltmeter immediately. The difference between the digital voltmeter readings in steps (5) and (6) is the CV line regulation. The difference of the readings during the immediate change should be within the
limit of (4 mV)
[1]
or (4.0 A)
[2]
[1]
or (7 mV)
[1]
or (50.0 V)
[1]
or 50V/4A
[2]
.
[2]
range, enable
. Check that the CV annunciator remains lit. If not lit,
[2]
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
62 Keysight E3633A and E3634A Service Guide

CV PARD (ripple and noise)

Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac voltage superimposed on the dc output voltage. CV PARD is specified as the rms or peak-to-peak output voltage in the frequency range from 20 Hz to 20 Mhz.
VRMS measurement techniques
When measuring Vrms ripple and noise, the monitoring device should be plugged into the front of the terminals at (B) in Figure 3-2. Use the coaxial cable and BNC female to banana plug adapter to connect the monitor device to the power supply. To reduce the measurement error caused by common mode noise, it is recommended that you use a common mode choke between the cable and the BNC adapter. The common mode choke is constructed by inserting coaxial cable into Ferrite Toroidal core. The load resistor should be connected to the terminal at (A) in Figure 3-2. Twisted leads between the load resistor and the power supply helps reduce noise pickup for these measurements.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 to an oscilloscope (ac coupled) between (+) and (-) terminals. Set
the oscilloscope to AC mode and bandwidth limit to 20 MHz. Connect a
resistive load (2.0
2 Turn on the power supply. Select the 20V/10A
the output, and set the display to the limit mode. When the display is in the
limit mode, program the current to the full rated value (10.0 A)
and the voltage to the full rated value (20.0 V)
3 Check that the front panel CV annunciator remains lit. If not lit, adjust the load
down slightly.
[1]
)
or (12.5 Ω)
Calibration Procedures 3
[2]
to the terminal at (A) in Figure 3-2.
[1]
or 50V/4A
[1]
or (50.0 V)
[2]
range, enable
[1]
or (4.0 A)
[2]
.
[2]
4 Note that the waveform on the oscilloscope does not exceed the peak-to-peak
limit of 3 mV.
5 Disconnect the oscilloscope and connect the ac rms voltmeter in its place
according to the VRMS measurement techniques above. The rms voltage
reading does not exceed the rms limit of 0.35 mV
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or 0.5 mV
[2]
.
Keysight E3633A and E3634A Service Guide 63
3 Calibration Procedures

Load transient response time

This test measures the time for the output voltage to recover to within 15 mV of nominal output voltage following a load change from full load to half load, or half load to full load.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with an oscilloscope. Operate the electronic load in constant
current mode.
2 Turn on the power supply. Select 20V/10A
output, and set the display to the limit mode. When the display is in the limit mode, program the current to the full rated value (10.0 A) the voltage to the full rated value (20.0 V)
3 Set the electronic load to transient operation mode between one half of the
output’s full rated value and the output’s full rated value at a 1 kHz rate with 50% duty cycle.
4 Set the oscilloscope for ac coupling, internal sync, and lock on either the
positive or negative load transient.
5 Adjust the oscilloscope to display transients as shown in Figure 3-3. Note that
the pulse width (t2 - t1) of the transients at 15 mV from the base line is no more than 50 μsec for the output.
[1]
or 50V/4A
[1]
or (50.0 V)
[2]
range, enable the
[2]
[1]
.
or (4.0 A)
[2]
and
Figure 3-3 Transient response time
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
64 Keysight E3633A and E3634A Service Guide

Constant Current (CC) Verifications

Constant current test setup

Follow the general setup instructions in the Measurement Techniques and the specific instructions will be given in the following paragraphs.

Current programming and readback accuracy

This test verifies that the current programming and GPIB or RS-232 readback functions are within specifications. Note that the readback values over the remote interface should be identical to those displayed on the front panel. The accuracy of the current monitoring resistor must be 0.01% or better.
You should program the power supply over the remote interface for this test to avoid round off errors.
1 Turn off the power supply and connect a 0.01 Ω current monitoring resistor
(R
) across the output to be tested and a digital voltmeter across the current
M
monitoring resistor (R
2 Turn on the power supply. Select the 8V/20A
the output by sending the commands:
).
M
Calibration Procedures 3
[1]
or 25V/7A
[2]
range and enable
VOLT:RANG P8V (E3633A)
VOLT:RANG P25V (E3634A)
OUTP ON
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 65
3 Calibration Procedures
3 Program the output voltage to full rated voltage (8.0 V)
[1]
or (25.0 V)
[2]
and
output current to zero amps by sending the commands:
VOLT 8 (E3633A)
VOLT 25 (E3634A)
CURR 0
4 Divide the voltage drop (DVM reading) across the current monitoring resistor
(R
) by its resistance to convert to amps and record this value (IO). This value
M
should be within the limit of (0 A ± 10 mA). Also, note that the CC, Adrs, Limit, and Rmt annunciators are on.
5 Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
6 Record the value displayed on the controller. This value should be within the
limit of (I
7 Program the output current to the full rated value (20.0 A)
± 5 mA).
O
[1]
or (7.0 A)
[2]
by
sending the command:
CURR 20.0 (E3633A)
CURR 7.0 (E3634A)
8 Divide the voltage drop (DVM reading) across the current monitoring resistor
(RM) by its resistance to convert to amps and record this value (I
[1]
This value should be within the limit of (20 A ± 50 mA)
or (7A ± 24 mA)
).
O
[2]
.
9 Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
66 Keysight E3633A and E3634A Service Guide
10 Record the value displayed on the controller. This value should be within the
limit of (I
± 35 mA)
O
[1]
or (IO ± 15.5 mA)

CC load effect (load regulation)

This test measures the immediate change in output current resulting from a change in the load from full rated output voltage to short circuit.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with the digital voltmeter connected across the 0.01 current
monitoring resistor (R
2 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit
mode, program the output voltage to the maximum programmable value and
the output current to the full rated value (20.0 A)
3 Set the voltage of the electronic load to (6.0 V)
constant voltage mode since a voltage drop occurs on the load wires when
[1]
(20.0 A)
or (7.0 A) is on. If it is not, adjust the load so that the output voltage drops slightly. Record the current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.
4 Operate the electronic load in short (input short) mode. Record the current
reading immediately by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (3) and (4) is the load regulation current. The difference of the readings during the immediate change should be within the
limit of (2.25 mA)
).
M
[2]
flows on the load wires. Check that the CC annunciator
[1]
or (0.95 mA)
[2]
Calibration Procedures 3
[2]
.
[1]
or 25V/7A
[1]
[1]
or (23.0 V)
[2]
or (7.0 A)
range, enable the
[2]
.
[2]
to operate it in
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 67
3 Calibration Procedures

CC source effect (line regulation)

This test measures the immediate change in output current that results from a change in ac line voltage from the minimum value (10% below the nominal input voltage) to the maximum value (10% above nominal voltage).
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with the digital voltmeter connected across the current monitoring
resistor (R
2 Connect the ac power line through a variable voltage transformer.
3 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit mode, program the voltage to the full rated value (8.0 V) the current to the full rated value (20.0 A)
4 Operate the electronic load in constant voltage mode and set its voltage to
(8.0 V) adjust the load so that the output voltage drops slightly until the CC annunciator lights.
5 Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the output current reading by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor.
6 Adjust the transformer to 10% above the nominal line voltage (127 Vac for a
115 Vac nominal input, 110 Vac for a 100 Vac nominal input or 253 Vac for a 230 Vac nominal input). Record the current reading again immediately by dividing the voltage reading on the digital voltmeter by the resistance of the current monitoring resistor. The difference between the current readings in step (5) and (6) is the load regulation current. The difference of the readings
during the immediate change should be within the limit of (2.25 mA)
[2]
mA)
).
M
[1]
or (25.0 V)
.
[1]
or 25V/7A
[1]
or (7.0 A)
[2]
. Check that the CC annunciator remains lit. If not lit,
[2]
range, enable the
[1]
or (25.0 V)
[2]
.
[1]
[2]
and
or (0.95
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
68 Keysight E3633A and E3634A Service Guide

CC PARD (ripple and noise)

Periodic and random deviations (PARD) in the output (ripple and noise) combine to produce a residual ac current, as well as an ac voltage superimposed on the dc output. CC PARD is specified as the rms output current in a frequency range 20 Hz to 20 MHz with the power supply in constant current operation.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1with a load resistor (0.4 Ω)
Connect a rms voltmeter across the load resistor. Use only a resistive load for this test.
2 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit mode, program the current to full rated value (20.0 A) voltage to the full rated value (8.0 V)
3 The output current should be at the full rated rating with the CC annunciator
on. If not lit, adjust the load so that the output voltage drops slightly until the
CC annunciator lights.
4 Divide the reading on the rms voltmeter by the load resistance to obtain rms
current. The readings should be within the limit of 2 mA.
[1]
or (3.5 Ω)
[1]
or (25.0 V)
Calibration Procedures 3
[2]
across output terminals.
[1]
or 25V/7A
[2]
[1]
or (7.0 A)
.
[2]
range, enable the
[2]
and the
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 69
3 Calibration Procedures

Common Mode Current Noise

The common mode current is that ac current component which exists between the output or output lines and chassis ground. Common mode noise can be a problem for very sensitive circuitry that is referenced to earth ground. When a circuit is referenced to earth ground, a low level line-related ac current will flow from the output terminals to earth ground. Any impedance to earth ground will create a voltage drop equal to the output current flow multiplied by the impedance.
1 Turn off the power supply and connect a 100 kΩ resistor (R
capacitor in parallel between the (-) terminal and chassis ground.
2 Connect a digital voltmeter across R
3 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit mode, program the output to the full rated value (8.0 V and 20.0 A) V and 7.0 A)
4 Record the voltage across R
resistance (DVM reading/100 k). Note that the current is less than 1.5 μA.
[2]
.
) and a 2200 pF
S
.
S
[1]
or 25V/7A
and convert it to current by dividing by the
S
[2]
range, enable the
[1]
or (25.0
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
70 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3

Performance Test Record for Keysight E3633A and E3634A

CV performance test record

Test Description Models
CV Programming Accuracy @ 0 volts (DVM reading)
CV Readback Accuracy @ 0 volts Both DVM + 0.0050 V DVM - 0.0050 V
CV Programming Accuracy @ Full Scale (DVM reading)
CV Readback Accuracy @ Full Scale
CV Load Effect (Load Regulation)
CV Source Effect (Line Regulation)
CV PARD (Normal mode)
Load Transient Response Time Both
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
Both +0.0100 V -0.0100 V
(E3633A) +20.0200 V 19.9800 V
(E3634A) +50.0350 V 49.9650 V
(E3633A) DVM + 0.0150 V DVM - 0.0150 V
(E3634A) DVM + 0.0300 V DVM - 0.0300 V
Actual Result
Upper Limit Lower Limit
Maximum change: < (4 mV)
Maximum change: < (4 mV)
< 3 mV p-p, (350
Specifications
[a]
μ
V)
or (500 μV)
< 50 μsec
[a]
or (7 mV)
[a]
or (7 mV)
[b]
[b]
[b]
rms
Keysight E3633A and E3634A Service Guide 71
3 Calibration Procedures

CC performance test records

Test Description Models
CC Programming Accuracy @ 0 amps (IO)
Both +0.0100 A -0.0100 A
CC Readback Accuracy @ 0 amps Both
CC Programming Accuracy @ Full Scale (IO)
CV Readback Accuracy @ Full Scale
(E3633A) 20.0500 A 19.9500 A
(E3634A) 7.0230 A 6.9770 A
(E3633A)
(E3634A)
CC Load Effect (Load Regulation)
CC Source Effect (Line Regulation)
Actual
Result
Upper Limit Lower Limit
I
O
IO + 0.0350 A IO - 0.0350 A
I
O
Maximum change: < (2.25 mA)
Maximum change: < (2.25 mA)
Specifications
+ 0.0050 A IO - 0.0050 A
+ 0.0155 A IO - 0.0155 A
CC PARD (Normal mode) Both < 2 mA rms
CC PARD (Common mode) Both
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
< 1.5
μ
A rms
[a]
or (0.95 mA)
[a]
or (0.95 mA)
[b]
[b]
72 Keysight E3633A and E3634A Service Guide

Calibration Security Code

NOTE
This feature allows you to enter a security code (electronic key) to prevent accidental or unauthorized calibrations of the power supply. When you first receive your power supply, it is secured. Before you can calibrate the power supply, you must unsecure it by entering the correct security code. A procedure to unsecure the power supply is given on the following page.
– The security code is set to “HP003633”
supply is shipped from the factory. The security code is stored in non-volatile memory, and does not change when power has been off or after a remote interface reset.
– To secure the power supply from the remote interface, the security code may
contain up to 12 alphanumeric characters as shown below. The first character must be a letter, but the remaining characters can be letters or numbers. You do not have to use all 12 characters but the first character must always be a letter.
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
– To secure the power supply from the remote interface so that it can be
unsecured from the front panel, use the eight-character format shown below. The first two characters must be “H P” and the remaining characters must be numbers. Only the last six characters are recognized from the front panel, but all eight characters are required. To unsecure the power supply from the front
panel, omit the “H P” and enter the remaining numbers as shown on the following pages.
Calibration Procedures 3
[1]
or “HP003634”
[2]
when the power
H P _ _ _ _ _ _ (8 characters)
If you forget your security code, you can disable the security feature by adding a jumper inside the power supply, and then entering a new code.
See“To unsecure the power supply without the security code” on page 76
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide 73
3 Calibration Procedures

To unsecure the power supply for calibration

The power supply can use a calibration security code to prevent unauthorized or accidental calibration. This procedure shows you how to unsecure the power supply for calibration from the front panel.
1 Turn on the front-panel calibration mode.
SECURED
Turn on the calibration mode by holding down (Calibrate) key as you turn on the power supply and hold down the key until you hear a long beep.
If the power supply is secured, you will see the above message from the front panel for approximately one second. The “CAL MODE” message is then displayed on the front panel.
2 Move to the security code by pressing (Secure) key.
000000 CODE
3 Enter the security code using the knob and resolution selection keys.
003633 CODE (E3633A)
003634 CODE (E3634A)
The security code is set to “HP003633”
[1]
or “HP003634”
[2]
when the power supply is shipped from the factory. The security code is stored in non-volatile memory and does not change when the power has been off or after a remote interface reset.
– To enter the security code from the front panel, enter only the last six
digits.
– To enter the security code from the remote interface, you may enter up to
12 characters.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
74 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
NOTE
Use the resolution selection keys to move left or right between digits. Use the knob to change the digits. Notice that the security code may be different if the security code has been changed from the default setting.
4 Unsecure the power supply.
UNSECURED
The power supply is unsecured when you press (Secure) key. You will see the above message from the front panel for one second. The “CAL MODE” message is displayed on the front panel after above message.
5 Turn off the calibration mode.
Turn off the power supply to exit the calibration mode.
To re-secure the power supply (following calibration), perform this procedure again.
Keysight E3633A and E3634A Service Guide 75
3 Calibration Procedures

To unsecure the power supply without the security code

To unsecure the power supply without the correct security code (when you forget the security code), follow the steps below. See Electrostatic Discharge (ESD)
Precautions in chapter 5 before beginning this procedure.
1 Disconnect the power cord and all load connections from the power supply.
2 Remove the instrument cover.
3 Connect the power cord and turn on the calibration mode by holding down
(Calibrate) key as you turn on the power supply and hold down the key
until you hear a long beep. Be careful not to touch the power line connections.
4 Apply a short between the two exposed metal pads on JP5
U13). The JP5 is outlined with a circle on the“E3633-60002/E3634-60002
Component Locator for the Main Board” on page 125.
5 While maintaining the short, move to the security code and enter any unsecure
code in the calibration mode. The power supply is now unsecured.
6 Remove the short at JP5
7 Turn off and reassemble the power supply.
[2]
. (An error occurs if not removed.)
[1]
(located near
Now you can enter a new security code. Be sure you take note of the new security code.
[1] For serial MY53xx6xxx, apply a short between the two exposed metal pads on JP102 (located near
U103). The JP102 is outlined with a circle on the component locator drawing on page 128
[2] For serial MY53xx6xxx, remove the short at JP102.
76 Keysight E3633A and E3634A Service Guide

Calibration Count

The calibration count feature provides an independent “serialization” of your calibrations. You can determine the number of times that your power supply has been calibrated. By monitoring the calibration count, you can determine whether an unauthorized calibration has been performed. Since the value increments by one for each calibration parameter (see Table 3-2 on the next page), a complete calibration increases the value by 5 counts.
– The calibration count is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset. Your power supply was calibrated before it left the factory. When you receive the power supply, read the calibration count to determine its value.
– The calibration count increments up to a maximum of 32,767 after which it
wraps around to 0. No way is provided to program or reset the calibration count.
Calibration Procedures 3
Keysight E3633A and E3634A Service Guide 77
3 Calibration Procedures

Calibration Message

You can use the calibration message feature to record calibration information about your power supply. For example, you can store such information as the last calibration date, the next calibration due date, the power supply’s serial number, or even the name and phone number of the person to contact for a new calibration.
You can record and read information in the calibration message from the remote interface only.
– The calibration message may contain up to 40 characters.
– The calibration message is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset.
– Remote Interface Operation
CAL:SEC:STAT {OFF|ON},<code> Secure or unsecure the power supply
CAL:SEC:CODE <new code> Change the security code
CAL:STR <quoted string> Store the calibration message
CAL:STR? Query the calibration information
See “Calibration Overview”, starting on page 68 in chapter 3 of the User’s Guide for more details.
78 Keysight E3633A and E3634A Service Guide

General Calibration/ Adjustment Procedure

NOTE
NOTE
The calibration procedures from the front panel are described in this section. For voltage calibration, disconnect all loads from the power supply and connect a DVM across the output terminals. For current calibration, disconnect all loads
from the power supply, connect an appropriate current monitoring resistor 0.01 across the output terminals, and connect a DVM across the terminals of the monitoring resistor.
Note that the power supply should be calibrated after 1-hour warm-up with no load connected.
The following table shows calibration parameters and points which should be used to calibrate the output voltage and current.
Table 3-2 Parameters for calibration
Calibration Parameter Voltage/ Current Calibration Point Mnemonic
CAL SETUP 1 Voltage
Calibration Procedures 3
V LO
V MI
V HI
CAL SETUP 2 OVP None
I LO
CAL SETUP 3 Current
CAL SETUP 4 OCP None
You can terminate any CAL SETUP without changing its calibration constants by turning off power.
Perform the voltage calibration prior to the OVP calibration and the current calibration prior to the OCP calibration.
Keysight E3633A and E3634A Service Guide 79
I MI
I HI
3 Calibration Procedures

Front panel voltage and current calibration

1 Unsecure the power supply.
To calibrate the voltage and current, you must unsecure the power supply according to the “To unsecure the power supply for calibration” on page 74.
2 Disconnect all loads from the power supply and connect a DVM across
output terminals.
3 Turn on calibration mode.
CAL MODE
Turn on the calibration mode by holding down (Calibrate) key as you turn on the power supply and hold down the key until you hear a long beep. Make sure that the power supply is in “CV” mode. If the power supply is not in “CV” mode, an error occurs.
Voltage and OVP calibration
4 Move down a level to the voltage calibration mode.
CAL SETUP 1
The display shows the above message to indicate that the power supply is ready for voltage calibration.
5 Calibrate DAC and select the low voltage calibration point.
30 LEFT
The “START BITCAL” message is displayed for about 3 seconds to indicate that the power supply is ready for DAC calibration. Then it counts down numbers from 30 to 0. Then, the display shows the low voltage calibration point.
V LO 0.5000V
80 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
6 Read the DVM and change the low voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V using the knob and resolution selection keys.
V LO 0.4500 V
7 Pressing (Calibrate) key saves the change and selects the middle
voltage calibration point. (“V MI 25.000 V” - E3634A model)
V MI 10.000 V (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message appears for one second. If the entered number is not correct, an “MIN VALUE” or “MAX VALUE” message appears for one second and the display shows the low voltage calibration point again. The display now shows the middle voltage calibration point.
8 Read the DVM and change the middle voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reads 10.001 V, adjust the voltage to 10.001 V using the knob and arrow keys.
V MI 10.001 V
9 Pressing (Calibrate) key saves the changes and selects the high
voltage calibration point. (“V HI 49.500 V” - E3634A model)
V HI 19.500 V (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message appears for one second. If the entered number is not correct, an “MIN VALUE” or “MAX VALUE” message appears for one second and the display shows the middle voltage calibration point again. The display now shows the high voltage calibration point.
Keysight E3633A and E3634A Service Guide 81
3 Calibration Procedures
NOTE
10 Read the DVM and change the high voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reads 19.495 V, adjust the voltage to 19.495 V using the knob and arrow keys.
V HI 19.495 V
11 Pressing (Calibrate) key saves the new voltage calibration constants,
and goes to the OVP calibration mode.
CAL SETUP 2
A “CALIBRATING” message appears for one second to indicate that the voltage calibration is progressing and new voltage calibration constants of “SETUP 1” are stored. Then, the display shows above message to indicate that the power supply is ready for the OVP calibration.
If the calibration fails, a “DAC CAL FAIL” or “ADC CAL FAIL” message appears for one second and the display shows the “CAL SETUP 1” for voltage calibration again.
Current and OCP calibration
12 Pressing (Calibrate) key saves the new calibration constants for OVP
circuit and goes to the current calibration mode.
CAL SETUP 3
A “CALIBRATING” message appears for about several seconds to indicate that the OVP calibration is progressing and new calibration constants of “SETUP 2” are stored. Then, the display shows the above message to indicate that the power supply is ready for the current calibration.
If the calibration fails, a “OVP CAL FAIL” message appears for one second and the display shows the “CAL SETUP 2” for OVP calibration again.
Connect an appropriate shunt 0.01 across the output terminals, and connect a digital voltmeter across the shunt resistor for the current calibration.
82 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
13 Select the low current calibration point. (‘‘I LO 0.2000 A’’ - E3634A model)
I LO 0.5000 A (E3633A)
The display shows the low current calibration point.
14 Read the DVM and change the low current value on the display to match
the computed current (DVM reading ÷ by shunt resistance).
For example, if the computed value is 0.4990 A, adjust the current to 0.4990 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
I LO 0.4990 A
15 Pressing the “Calibrate” key saves the change and selects the middle
current calibration point. (“I MI 3.5000 A’’ - E3634A model)
I MI 10.000 A (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message appears for one second. If the entered number is not correct, an “MIN VALUE” or “MAX VALUE” message appears for one second and the display shows the low current calibration point again. The display now shows the middle current calibration point.
16 Read the DVM and change the middle current value on the display to match
the computed current (DVM reading ÷ by shunt resistance).
For example, if the computed value is 9.999 A, adjust the current to 9.999 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
I MI 09.999 A
Keysight E3633A and E3634A Service Guide 83
3 Calibration Procedures
17 Pressing the “Calibrate” key saves the change and selects the high current
calibration point. (“I HI 6.9000 A” - E3634A model)
I HI 19.500 A (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message appears for one second. If the entered number is not correct, an “MIN VALUE” or “MAX VALUE” message appears for one second and the display shows the middle current calibration point again. The display now shows the high current calibration point.
18 Read the DVM and change the high current value on the d isplay to match
the computed current (DVM reading ¸ by shunt resistance).
For example, if the computed value is 19.499 A, adjust the current to 19.499 A using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate calibration.
I HI 19.499 A
19 Pressing the ‘‘Calibrate’’ key saves the new calibration constants for the
output current and goes to the OCP calibration mode.
CAL SETUP 4
A “CALIBRATING” message appears for one second to indicate that the current calibration is progressing and new calibration constants of “SETUP 3” are stored. Then, the display shows the above message to indicate that the power supply is ready for the OCP calibration.
If the calibration fails, an “DAC CAL FAIL” or “ADC CAL FAIL” message appears for one second and the display shows the “CAL SETUP 3” for current calibration again.
84 Keysight E3633A and E3634A Service Guide
20 Pressing the “Calibrate” key saves the new OCP calibration constants and
return to the calibration mode.
CAL MODE
A “CALIBRATING” message appears for several seconds to indicate that the OCP calibration is progressing and new OCP calibration constants of “SETUP 4” are stored. Then the display will return to the calibration mode.
If the calibration fails, a “OCP CAL FAIL” message appears for one second and the display shows the “CAL SETUP 4” for OCP calibration again.
Turn off the power supply to exit the calibration mode.

Aborting a Calibration in Progress

Sometimes it may be necessary to abort a calibration after the procedure has already been initiated. You can abort a calibration at any time by turning the power supply off from the front panel. When performing a calibration from the remote interface, you can abort a calibration by issuing a remote interface device clear message or by pressing the front-panel “Local” key.
Calibration Procedures 3
Keysight E3633A and E3634A Service Guide 85
3 Calibration Procedures

Calibration Record for Keysight E3633A/E3634A

Step Calibration Description
Measurement
Mode (DVM)
Supply Being
Adjusted
1 Unsecure the power supply (see page 62).
2 Turn on “CAL MODE” (hold down “Calibrate” and “Power” keys as you turn on the
power supply until you hear a long beep).
3 Move down menu to “CAL SETUP 1” (press “Calibrate” key). Voltage
Calibration
4 Calibrate DAC and select the low point for voltage calibration; “START BITCAL”
appears for about 3 seconds and the display counts down numbers from 30 to 0. Then, “V LO 0.5000 V” appears on the display (press “Calibrate” key and wait
VDAC and low
voltage point calibration
about 30 seconds; then change the display to match the DVM reading).
5
“V MI 10.000 V”
[a]
or “V MI 25.000 V”
(press “Calibrate” key; then change the display to match the DVM reading).
6
“V HI 19.500 V”
[a]
or “V HI 49.500 V”
(press “Calibrate” key; then change the display to match the DVM reading).
[b]
appears on the display
[b]
appears on the display
VMiddle voltage
point calibration
V High vol tage point
calibration
7 “CAL SETUP 2” now appears on the display (press “Calibrate” key). V OVP calibration
8 “CAL SETUP 3” now appears on the display
(press “Calibrate” key; then connect 0.01
9
“I LO 0.5000 A”
[a]
or “I LO 0.2000 A”
resistor across the output terminals).
[b]
appears on the display
(press “Calibrate” key; then change the display to match the computed current
A Low current point
Current calibration
calibration
through the resistor).
10
“I MI 10.000 A”
[a]
or “I MI 3.5000 A”
(press “Calibrate” key; then change the display to match the computed current
[b]
appears on the display
A Middle current
point calibration
through the resistor).
11
“I HI 19.500 A”
[a]
or “I HI 6.9000 A”
(press “Calibrate” key; then change the display to match the computed current
[b]
appears on the display
A High current point
calibration
through the resistor).
12 “CAL SETUP 4” now appears on the display (press “Calibrate” key). A OCP calibration
13 Press “Calibrate” key, then press “Power” switch. Exit CAL MODE
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
86 Keysight E3633A and E3634A Service Guide

Error Messages

The following tables are abbreviated lists of error messages for the E3633A and E3634A. The errors listed are the most likely errors to be encountered during calibration and adjustment. A more complete list of error messages and descriptions is contained in chapter 5 of the E3633A and E3634A User’s Guide.
System error messages
Error Error Messages
-330 Self-test failed
-350 Queue overflow
501 Isolator UART framing error
502 Isolator UART overrun error
503
511 RS-232 framing error
512 RS-232 overrun error
513 RS-232 parity error
514 Command allowed only with RS-232
521 Input buffer overflow
522 Output buffer overflow
550 Command not allowed in local
Calibration Procedures 3
[a]
SPI data error
[a] This error message is only applicable for serial MY53xx6xxx.
Keysight E3633A and E3634A Service Guide 87
3 Calibration Procedures
Self-test error messages
Error Error Messages
601 Front panel does not respond
602 RAM read/write failed
603 A/D sync stuck
604 A/D slope convergence failed
605 Cannot calibrate rundown gain
606 Rundown gain out of range
607 Rundown too noisy
608 Serial configuration readback failed
[a]
609
624 Unable to sense line frequency
625 I/O processor does not respond
626 I/O processor failed self-test
630 Fan test failed
631 System DAC test failed
632 Hardware test failed
System ADC test failed
[a] This error message is only applicable for serial MY53xx6xxx.
88 Keysight E3633A and E3634A Service Guide
Calibration error messages
Error Error Messages
701 Cal security disabled by jumper
702 Cal secured
703 Invalid secure code
704 Secure code too long
705 Cal aborted
708 Cal output disabled
712 Bad DAC cal data
713 Bad readback cal data
714 Bad OVP cal data
715 Bad OCP cal data
716 Bad DAC DNL error correction data
717 Cal OVP or OCP status enabled
718 Gain out of range for gain error correction
740 Cal checksum failed, secure state
741 Cal checksum failed, string data
742 Cal checksum failed, store/recall data in location 0
743 Cal checksum failed, store/recall data in location 1
744 Cal checksum failed, store/recall data in location 2
745 Cal checksum failed, store/recall data in location 3
746 Cal checksum failed, DAC cal constants
747 Cal checksum failed, readback cal constants
748 Cal checksum failed, GPIB address
749 Cal checksum failed, internal data
750 Cal checksum failed, DAC DNL error correction data
Calibration Procedures 3
Keysight E3633A and E3634A Service Guide 89
3 Calibration Procedures

An Example Program of Excel 97 for Calibration

This section contains an Excel Macros (Visual Basic® for Applications) program for calibration over the GPIB interface. This program makes software adjustments to the E3633A power supply using a current shunt and a digital multimeter which is connected to the controller. In this program a 0.001 ohm current shunt is used. Be sure to change the value of the variable “shunt” to the value of the current shunt used and the GPIB address for the power supply and the digital voltmeter.
'********************************************************************** ' This program was written on a PC with Excel Macros (Visual Basic® for ' Applications) for Windows 95 or Windows NT 4.0. It will make software ' adjustments to the E3633A Power Supply on the GPIB bus using a ' Keysight 34401A Digital Multimeter and a current shunt. In the program ' a 0.001 ohm current shunt is used to measure current. Be sure to ' change the value of the variable ‘shunt’ to the value of the current ' shunt used. '**********************************************************************
Global id_power As Long Global id_DMM As Long Global power As Long Global DMM As Long Global Const VI_SUCCESS = &H0& Global Const VoltageMin = 0 Global Const VoltageMid = 1 Global Const VoltageMax = 2 Global Const CurrentMin = 3 Global Const CurrentMid = 4 Global Const CurrentMax = 5
Sub Calibration_Click()Max
Dim shunt As Single 'Current shunt value in Ohms Dim UserAnswer 'User response Range("B4:B6").ClearContents
If OpenPort = False Then
ClosePort
Exit Sub End If InitializeDevice EnableOVPandOCP False If CheckSecurity = False Then
ClosePort
Exit Sub End If
Continued on next page
90 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
If DacErrorCorrection = False Then
ClosePort Exit Sub
End If
shunt = 0.001 'Change the current shunt value
UserAnswer = MsgBox(“Voltage & OVP calibration. Connect the output to the
DMM”, vbYesNo + vbQuestion, “E3633A Calibration”)
If UserAnswer = vbNo Then
ClosePort Exit Sub
End If
Range("B4").Select ActiveCell.Value = "Begin Voltage Calibration" StartCalibration VoltageMin, True, shunt 'Set output to minimum voltage cal ActiveCell.Value = "End Minimum Voltage Calibration" StartCalibration VoltageMid, True, shunt 'Set output to middle voltage cal ActiveCell.Value = "End Middle Voltage Calibration" StartCalibration VoltageMax, True, shunt 'Set output to maximum voltage cal ActiveCell.Value = "End Maximum Voltage Calibration" Message = SendSCPI(power, "Syst:Err?") If InStr(Message, "0") Then
ActiveCell.Value = "Voltage Calibration Complete"
Else
ActiveCell.Value = Message ClosePort Exit Sub
End If
OVPandOCPCalibration True
UserAnswer = MsgBox("Connect the Current shunt to the DMM input to measure a current. Connect the output to the shunt.", vbYesNo + vbQuestion, "E3633A Calibration")
If UserAnswer = vbNo Then
ClosePort Exit Sub
End If
ActiveCell.Value = "Begin Current Calibration" StartCalibration CurrentMin, False, shunt 'Set output to minimum current
cal
ActiveCell.Value = "End Minimum Current Calibration" StartCalibration CurrentMid, False, shunt 'Set output to middle current cal ActiveCell.Value = "End Middle Current Calibration" StartCalibration CurrentMax, False, shunt 'Set output to maximum current
cal
ActiveCell.Value = "End Maximum Current Calibration"
Continued on next page
Keysight E3633A and E3634A Service Guide 91
3 Calibration Procedures
OVPandOCPCalibration False Message = SendSCPI(power, "Syst:Err?") If InStr(Message, "0") Then
Else
End If EnableOVPandOCP True SaveDate ActiveCell.Value = "Calibration Complete" Message = SendSCPI(power, "*RST") ClosePort
End Sub
Private Function OpenPort() As Boolean 'Open communications on GPIB
Dim Power_Address As String Dim DMM_Address As String Dim Error As Long
Power_Address = "5" 'Select power supply GPIB address between 0 to 30 DMM_Address = "22" 'Select DMM GPIB address between 0 to 30 Error = viOpenDefaultRM(id_power) 'Open the power supply VISA session If OpenPort = CheckError(Error, "Open ID Error to Power Supply") Then
End If Error = viOpenDefaultRM(id_DMM) 'Open the DMM VISA session If OpenPort = CheckError(Error, "Open ID Error to Digital Multi Meter")
Then
End If Error = viOpen(id_power, "GPIB0::" & Power_Address & "::INSTR", 0, 1000,
power)
If OpenPort = CheckError(Error, "Unable Open to Power Supply") Then
End If Error = viOpen(id_DMM, "GPIB0::" & DMM_Address & "::INSTR", 0, 1000, DMM) If OpenPort = CheckError(Error, "Unable Open to Digital Multi Meter") Then
End If OpenPort = True
End Function
ActiveCell.Value = "Current Calibration Complete"
ActiveCell.Value = Message
ClosePort
Exit Sub
Exit Function
Exit Function
Exit Function
Exit Function
Private Function InitializeDevice()
SendSCPI power, "*Cls" SendSCPI DMM, "*Rst" 'Set power-on condition for DMM SendSCPI power, "*Rst" 'Set power-on condition for power supply
End Function
Continued on next page
92 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
Private Function CheckError(Error As Long, Message As String) As Boolean
If Error < VI_SUCCESS Then
ActiveCell.Value = Message CheckError = False 'Exit Calibration
Else
CheckError = True
End Function
Private Function ClosePort()
End Function
Private Function EnableOVPandOCP(bEnable As Boolean)
End Function
'This routine send a SCPI command string to the GPIB port. 'If the command contains a question mark, the response is read, and returned
Private Function SendSCPI(device As Long, command As String) As String
End Function
Private Function delay(delay_time As Single) 'This routine is used to create delays
End Function
End If
Dim Error As Long Error = viClose(power) Error = viClose(id_power) Error = viClose(DMM) Error = viClose(id_DMM)
If bEnable Then
SendSCPI power, "Volt:Prot:Stat On" SendSCPI power, "Curr:Prot:Stat On"
Else
SendSCPI power, "Volt:Prot:Stat Off" SendSCPI power, "Curr:Prot:Stat Off"
End If
Dim commandString As String Dim ReturnString As String Dim crlfpos As Integer Dim ReadBuffer As String * 512 Dim actual As Long Dim Error As Long commandString = command & Chr$(10)
Error = viWrite(device, ByVal commandString, Len(commandString), actual) If InStr(commandString, "?") Then
Error = viRead(device, ByVal ReadBuffer, 512, actual) ReturnString = ReadBuffer crlfpos = InStr(ReturnString, Chr$(0)) If crlfpos Then
ReturnString = Left(ReturnString, crlfpos - 2) End If SendSCPI = ReturnString
End If
Dim Finish As Single Finish = Timer + delay_time Do Loop Until Finish <= Timer
Continued on next page
Keysight E3633A and E3634A Service Guide 93
3 Calibration Procedures
Private Function CheckSecurity() As Boolean
Dim Message As String Dim SecurityCode As String Message = SendSCPI(power, "Cal:Str?") Range("B5").Select ActiveCell.Value = Message Message = SendSCPI(power, "Cal:Count?") Range("B6").Select ActiveCell.Value = Message Range("B4").Select SecurityCode = "HP003633" SendSCPI power, "Cal:Sec:Stat Off," & SecurityCode Message = SendSCPI(power, "Cal:Sec:Stat?")
If InStr(Message, "1") Then
Else
Else
End If
End If
End Function
Private Function DacErrorCorrection() As Boolean
Dim Message As String SendSCPI power, "Output On" 'Turn on the power supply output SendSCPI power, "Cal:Dac:Error" For I = 1 To 27
Next I SendSCPI power, "Output Off" 'Turn off the power supply output Message = SendSCPI(power, "Syst:Err?") If InStr(Message, "0") Then
Else
End If
End Function
ActiveCell.Value = "Unable to Unsecure the Power Supply" CheckSecurity = False
Message = SendSCPI(power, "Syst:Err?") If InStr(Message, "0") Then
CheckSecurity = True
ActiveCell.Value = Message CheckSecurity = False
delay 1 ActiveCell.Value = "Waitting for " & Str$(27 - I) & "secs"
ActiveCell.Value = "DAC DNL Error Correction completed for power supply" DacErrorCorrection = True
ActiveCell.Value = Message DacErrorCorrection = False
Continued on next page
94 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
Private Function StartCalibration(mode As Integer, bVolt As Boolean, shunt As Single)
Dim DMMdata As Single
SendSCPI power, "Output On" 'Turn on the power supply outpu Select Case mode
Case VoltageMin
SendSCPI power, "Cal:Volt:Level Min" 'Set output to minimum cal
value Case VoltageMid
SendSCPI power, "Cal:Volt:Level Mid" 'Set output to middle cal
value Case VoltageMax
SendSCPI power, "Cal:Volt:Level Max" 'Set output to maximum cal
value Case CurrentMin
SendSCPI power, "Cal:Curr:Level Min" 'Set output to minimum cal
value Case CurrentMid
SendSCPI power, "Cal:Curr:Level Mid" 'Set output to middle cal
value Case CurrentMax
SendSCPI power, "Cal:Curr:Level Max" 'Set output to maximum cal
End Select
delay 4 DMMdata = SendSCPI(DMM, "Meas:Volt:DC?") 'measure output
If bVolt Then
Else
End If SendSCPI power, "Output Off" 'Turn off the power supply output
End Function
Private Function OVPandOCPCalibration(bVolt As Boolean)
SendSCPI power, "Output On" 'Turn on the power supply output If bVolt Then
Else
End If delay 9 SendSCPI power, "Output Off" 'Turn off the power supply output
End Function
value
'Send the measured voltage value to the power supply
SendSCPI power, "Cal:Volt:Data " & Str(DMMdata)
'Send the measured current value to the power supply SendSCPI power, "Cal:Curr:Data " & Str(DMMdata / shunt)
ActiveCell.Value = " Begin OVP Calibration" SendSCPI power, "Cal:Volt:Prot" 'Perform OVP circuit calibration
ActiveCell.Value = " Begin OCP Calibration" SendSCPI power, "Cal:Curr:Prot" 'Perform OCP circuit calibration
Continued on the next page
Keysight E3633A and E3634A Service Guide 95
3 Calibration Procedures
Private Function SaveDate()
Dim Message As String 'calibration string enclosed in quotes Message = "Cal:Str " & Chr(34) & Date & Chr(34) SendSCPI power, Message Message = SendSCPI(power, "Cal:Str?") Range("B5").Select ActiveCell.Value = Message Message = SendSCPI(power, "Cal:Count?") Range("B6").Select ActiveCell.Value = Message
Range("B4").Select
End Function
End of program
96 Keysight E3633A and E3634A Service Guide
Calibration Procedures 3
Declaration for Windows 95/NT 4.0
Declare Function viOpenDefaultRM Lib "visa32.dll" (instrumentHandle As Long) As Long
Declare Function viOpen Lib "visa32.dll" (ByVal instrumentHandle As Long, _
ByVal viDesc As String, ByVal mode As Long, ByVal timeout As Long, _ vi As Long) As Long
Declare Function viClose Lib "visa32.dll" (ByVal vi As Long) As Long
Declare Function viWrite Lib "visa32.dll" (ByVal vi As Long, ByVal Buffer As String, ByVal count As Long, retCount As Long) As Long
Declare Function viRead Lib "visa32.dll" (ByVal vi As Long, ByVal Buffer As String, _ByVal count As Long, retCount As Long) As Long
Keysight E3633A and E3634A Service Guide 97
3 Calibration Procedures
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
98 Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies Service Guide

4 Theory of Operation

Block Diagram Overview 100 AC Input and Bias Supplies 102 Floating Logic 103 D-to-A Converter 105 A-to-D Converter 106 Power Mesh and Control 107 Earth-Referenced Logic 108 Front Panel 109
This chapter provides block diagram level descriptions of the power supply. The descriptions provide a basic understanding of circuit operation and are intended as an aid in troubleshooting. It is assumed in the following discussions that you are familiar with the operating and programming instructions presented in the E3633A and E3634A User’s Guide. Subjects covered include the following:
99
4 Theory of Operation

Block Diagram Overview

This discussion pertains to the block diagram on the next page. The power supply’s circuitry is divided into two major blocks: the floating circuitry and the earth referenced circuitry. All power mesh and control circuits, display circuit, and digital circuits are contained in the floating circuitry. This circuitry also contains the power supply’s main controller. The earth referenced circuitry provides the interface between the user and the power supply.
The floating circuitry can be viewed in four pieces; the DAC system, the digital logic section (floating logic), the power mesh and control section, and the front panel (display and keyboard) section.
The floating logic receives digital signals from the earth-referenced logic and the DAC converts them to analog signals which are sent to the power control circuits in order to program the power supply’s output voltage and current. The power supply can also be commanded to send measurement and status data back to the remote interface controller and/or the VFD (vacuum fluorescent display) display on the front panel. The data is processed and sent back via the floating logic and earth-referenced logic.
The power mesh and control circuits contains voltage and current control circuits which allows the power supply to operate in either the constant voltage (CV) or constant current (CC) mode. The control circuits compare the power supply’s output voltage or current with the programmed value and generates a control signal which varies the conduction of the series pass transistor to raise or lower the output as required.
The front panel circuits consist of VFD control, display high voltage drivers, and keyboard scanning. Communication between the front panel and floating logic circuits is accomplished through a 4-wire bi-directional serial interface.
The earth referenced circuitry uses a controller configured as a slave to the main controller. This controller establishes external I/O communication with the main controller through a bi-directional, optically isolated, serial communications link. The earth referenced controller controls low-level GPIB (IEEE-488) and RS-232 interface operation.
Separate reference and bias supplies are provided for the floating and ground reference circuitry. The front panel operates from the floating circuitry with its logic common different from the main controller logic common.
100 Keysight E3633A and E3634A Service Guide
Loading...