duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Manual Part Number
E3634-90010
Edition
Edition 9, March 26, 2019
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, mod ify, reprod uce, release, perform, display, or disclose commercial
computer software or commercial computer software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight E3633A and E3634A Service Guide
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Earth (ground) terminalFrame or chassis (ground) terminal
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
Alternating current (AC)Protective earth (ground) terminal
Out position of a bi-stable push controlIn position of a bi-stable push control
Positive binding postNegative binding post
Direct current (DC)
Keysight E3633A and E3634A Service Guide3
Regulatory Markings
This symbol is a South Korean Class A
EMC Declaration. This is a Class A
instrument suitable for professional
use and in electromagnetic
environment outside of the home.
This symbol indicates the time period
during which no hazardous or toxic
substance elements are expected to
leak or deteriorate during normal use.
Forty years is the expected useful life
of the product.
The CE mark is a registered trademark
of the European Community. This CE
mark shows that the product complies
with all the relevant European Legal
Directives.
This instrument complies with the
WEEE Directive (2002/96/EC) marking
requirement. This affixed product label
indicates that you must not discard
this electrical or electronic product in
domestic household waste.
The CSA mark is a registered
trademark of the Canadian
Standards Association.
ICES/NMB-001 indicates that this ISM
device complies with the
Canadian ICES-001.
Cet appareil ISM est conforme a la
norme NMB-001 du Canada.
The RCM mark is a registered
trademark of the Australian
Communications and Media Authority.
4Keysight E3633A and E3634A Service Guide
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
– www.keysight.com/find/powersupply
(product-specific information and support, software and
documentation updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight E3633A and E3634A DC Power Supplies
Service Guide
1Specifications
Performance Specifications 18
Supplemental Characteristics 20
The performance specifications are listed in the following pages. Specifications
are warranted in the temperature range of 0 to 40°C with a resistive load.
Supplemental characteristics, which are not warranted but are descriptions of
performance determined either by design or testing. The Service Guide contains
procedures for verifying the performance specifications.
17
1Specifications
Performance Specifications
Tab le 1-1Performance specifications
ParameterKeysight E3633AKeysight E3634A
Output Ratings
°C- 40 °C)
(@ 0
Programming Accuracy
12 months (@ 25 °C ± 5 °C),
±(% of output + offset)
Readback Accuracy
12 months (over GPIB and
RS-232 or front panel with
respect to actual output @ 25 °C
± 5 °C), ±(% of output + offset)
Ripple and Noise
(with outputs ungrounded, or
with either output terminal
grounded, 20 Hz to 20 MHz)
Load Regulation,
±(% of output + offset)
Line Regulation,
±(% of output + offset)
Programming ResolutionVoltage1 mV3 mV
Readback ResolutionVoltage0.5mV1.5 mV
Front Panel ResolutionVoltage1 mV
[a]
[a][b]
Low Range0 to +8 V/ 0 to 20 A0 to 25 V/ 0 to 7 A
High Range0 to +20V/ 0 to 10 A 0 to+50 V/ 0 to 4 A
Voltage0.05% + 10 mV
Current0.2% + 10 mA
Voltage0.05% + 5 mV
Current0.15% + 5 mA
Normal mode voltage<0.35 mV rms and
3 mV p-p
Normal mode current<2 mA rms
Common mode current <1.5 uA rms
Voltage<0.01% + 2 mV
Current
Voltage<0.01% + 2 mV
Current<0.01% + 250 μA
Current1 mA0.5 mA
Current1 mA0.5 mA
Current1 mA (< 10A), 10mA (. 10A)
<0.01% + 250 μA
<0.5 mV rms and
3 mV p-p
[a] Accuracy specifications are after an 1-hour warm-up with no load and calibration at 25 °C.
[b] This specification may degrade when the unit is subjected to an RF field >= 3V/meter.
18Keysight E3633A and E3634A Service Guide
Transient response time
Less than 50 μsec for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Command processing time
Average time for output voltage to begin to change after receipt of digital data
when the power supply is connected directly to the GPIB or RS-232 is less than
100 msec
OVP and OCP accuracy, ±(% of output + offset)
OVP0.5% + 0.5 V
OCP0.5% + 0.5 A
Activation time
Specifications1
Average time for output to start to drop after OVP and OCP condition occurs.
OVP<1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
OCP<10 msec
Keysight E3633A and E3634A Service Guide19
1Specifications
Supplemental Characteristics
Tab le 1-2Supplemental Characteristics
ParameterKeysight E3633AKeysight E3634A
Output Programming Range
(maximum programmable values)
Voltage Programming Speed:
Maximum time required for output
voltage to settle within 1% of its total
excursion (for resistive load). Excludes
command processing time.
Remote sensing capability
Low Range
High Range
OVP1 V to 22 V1 V to 55 V
OCP0 A to 22 A0 A to 7.5 A
Up95 msec45 msec80 msec100 msec
Down30 msec450 msec30 msec450 msec
0 to +8.24 V/
0 to 20.6 A
0 to +20.6 V/
0 to 10.3 A
Full LoadNo LoadFull LoadNo Load
0 to +25.75 V/
0 to 7.21 A
0 to +51.5V/
0 to 4.12 A
Voltage dropUp to 0.7 V per each lead
Load regulation
Load VoltageSubtract voltage drop in load leads from specified output voltage rating.
Add 5 mV to spec for each 1-volt change in the + output lead due to load
current changes
Temperature coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage0.01% + 3 mV
Current0.02% + 3 mA
20Keysight E3633A and E3634A Service Guide
Stability, ±(% of output + offset)
Following a 30-minute warm-up, with the output in the ON state according to the
operating mode (CC with load or CV), and with a change in the output over 8
hours under constant load, line, and ambient temperature.
Voltage0.02% + 1mV
Current 0.1% + 1mA
Output voltage overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V
if the output control is set to less than 1 V. If the output control is set to 1 V or
higher, there is no overshoot.
Programming language
SCPI (Standard Commands for Programmable Instruments)
Specifications1
State Storage Memory
Three (3) user-configurable stored states
Recommended calibration interval
1 year
Output Terminal Isolation (maximum, from chassis ground)
±60 Vdc when connecting shorting conductors without insulation to the
(+) output to the (+) sense and the (-) output and the (-) sense terminals.
±240 Vdc when connecting insulated shorting conductors to the (+) output
to the (+) sense and the (-) output and the (-) sense terminals.
Keysight E3633A and E3634A Service Guide21
1Specifications
AC input ratings (selectable via rear panel selector
Maximum input power
700 VA with full load
Cooling
Fan cooled.
Operating temperature
0 to 40 °C for full rated output. At higher temperatures, the output current is
derated linearly to 50% at 55 °C maximum temperature.
std115 Vac ± 10%, 47 to 63 Hz
opt 0E3230 Vac ± 10%, 47 to 63 Hz
opt 0E9100 Vac ± 10%, 47 to 63 Hz
Storage temperature
-20 to 70 °C for storage environment.
Environmental Conditions
Designed for indoor use in an installation category II, pollution degree 2
environment. Designed to operate at a maximum relative humidity of 95 % and at
altitudes of up to 2000 meters.
Dimensions
[1] See below for detailed information
22Keysight E3633A and E3634A Service Guide
[1]
213 mmW x 133 mmH x 348 mmD (8.4 x 5.2 x 13.7 in)
Weight
Net9.5 kg (21 lb)
Shipping12 kg (26 lb)
Weight (MY53xx6xxx)
Net11.4 kg (E3633A)
10.9 kg (E3634A)
Shipping13.9 kg (E3633A)
13.4 kg (E3634A)
Specifications1
Figure 1-1Dimensions of Keysight E3633A and E3634A power supplies
Keysight E3633A and E3634A Service Guide23
1Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies
Service Guide
2Quick Start
Quick StartQuick Start 26
To prepare the Power Supply for Use 28
To Check the Rated Voltages of the Power Supply 31
To Check the Rated Currents of the Power Supply 32
To Use the Power Supply in Constant Voltage Mode 34
To Use the Power Supply in Constant Current Mode 37
To store and Recall the Instrument State 39
To Program Overvoltage Protection 41
To Program Overcurrent Protection 45
To Rack Mount the Power Supply 49
25
2Quick Start
Quick Start
One of the first things you will want to do with your power supply is to become
acquainted with its front panel. Written procedures in this chapter prepare the
power supply for use and familiarize you with most front-panel operations.
– The power supply is shipped from the factory configured in the front-panel
– The power supply has two output ranges. This feature allows more voltage at a
– When you press key (the Limit annunciator blinks), the display of the
operation mode. At power-on, the power supply is automatically set to operate
in the front-panel operation mode. When in this mode, the front-panel keys
can be used. When the power supply is in remote operation mode, you can
return to front-panel operation mode at any time by pressing the (Local)
key if you did not previously send the front-panel lockout command. A change
between front-panel and remote operation modes will not result in a change in
the output parameters
lower current or more current at a lower voltage. The desired output range is
selected from the front panel or over the remote interfaces. The 8V or 20V for
the E3633A and 25V or 50V for the E3634A annunciator indicates the
presently selected range.
power supply goes to the limit mode and the present limit values will be
displayed. In this mode, you can also observe the change of the limit values
when adjusting the knob. If you press the key again or let the display
time-out after several seconds, the power supply will return the display to the
meter mode (the Limit annunciator turns off). In this mode, the actual output
voltage and current will be displayed.
– The output of the power supply can be enabled or disabled from the front
panel by pressing key. When the output is off, the OFF annunciator turns
on and the output is disabled.
26Keysight E3633A and E3634A Service Guide
Quick Start2
– The display provides the present operating status of the power supply with
annunciators and also informs the user of error codes. For example, the power
[1]
supply is operating in CV mode in the 8V/20A
or 25V/7A
controlled from the front panel, then the CV and 8V
[1]
[2]
or 25V
range and
[2]
annunciators
will turn on. If, however, the power supply is remotely controlled, the Rmt
annunciator will also turn on, and when the power supply is being addressed
over HPIB interface, the Adrs annunciator will turn on.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide27
2Quick Start
To prepare the Power Supply for Use
The following steps help you verify that the power supply is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your power supply. If
anything is missing, contact your nearest Keysight Technologies Sales Office.
– One power cord for your location.
– This User’s Guide.
– One Service Guide.
– Certificate of Calibration.
2 Verify that the correct power-line voltage setting is selected and that the
correct power-line fuse is installed.
The line voltage is set to 100, 115 or 230 Vac from the factory according to the
input power option selected when you ordered the power supply. Change the
voltage setting if it is not correct for your location (see the next page for
detailed information). For 100 or 115 Vac operation, the correct fuse is 6.3 AT
(Keysight part number 2110-1030) and for 230 Vac operation, the correct fuse
is 3.15 AT (Keysight part number 2110-1031).
3 Connect the power cord and turn on the power supply.
A power-on self-test occurs automatically when you turn on the power supply.
The front-panel display will light up while the power supply performs its
power-on self- test. After performing the self-test, the power supply will go
into the power-on / reset state; the output is disabled (the OFF annunciator
turns on); the 8V/20A
[1]
or 25V/7A
[2]
range is selected (the 8V
[1]
or 25V
[2]
annunciator turns on); the knob is selected for voltage control.
Notice that the OVP and OCP annunciators also turn on.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
28Keysight E3633A and E3634A Service Guide
Quick Start2
4 Enable the outputs.
Press key to enable the outputs. The OFF annunciator turns off and the
8V
[1]
or 25V
[2]
, OVP, OCP, and CV annunciators are lit. The blinking digit can be
adjusted by turning the knob. Notice that the display is in the meter mode.
“Meter mode” means that the display shows the actual output voltage and
current.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide29
2Quick Start
NOTE
1 Remove the power cord. Remove the fuse-holder assembly
with a flat-blade screwdriver from the rear panel.
3 Rotate the power-line vol tage selector until the correct
voltage appears.
2 Install the correct line fuse. Remove the power-line
voltage selector from the power-line module.
4 Replace the power-line voltage selector and the
fuse-holder assembly in the rear panel.
Install the correct fuse and verify that the correct line voltage appears in the
window.
30Keysight E3633A and E3634A Service Guide
To Check the Rated Voltages of the Power Supply
NOTE
The following procedures check to ensure that the power supply produces its
rated voltage output with no load and properly responds to operation from the
front panel.
For each step, use the keys shown on the left margins.
1Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
2 Enable the outputs.
The OFF annunciator turns off and the 8V
annunciators are lit. The blinking digit can be adjusted by turning the knob.
Notice that the display is in the meter mode. “Meter mode” means that the
display shows the actual output voltage and current.
[1]
or 25V/7A
[1]
or 25V
[2]
Quick Start2
[2]
range is selected
OVP, OCP, and CV
3 Check that the front-panel voltmeter properly responds to knob control for
the 8V/20A
[1]
or 25V/7A
[2]
range.
Turn the knob clockwise or counter clockwise to check that the voltmeter
responds to knob control and the ammeter indicates nearly zero.
4 Ensure that the voltage can be adjusted from zero to the full rated value.
Adjust the knob until the voltmeter indicates 0 volts and then adjust the knob
until the voltmeter indicates “8.0 volts”
[1]
or “25.0 volts”
[2]
You can use the resolution selection keys to move the flashing digit to the right
or left when setting the voltage.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide31
2Quick Start
To Check the Rated Currents of the Power Supply
The following steps check basic current functions with a short across the power
supply’s output.
1 Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
[1]
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
2 Connect a short across (+) and (-) output terminals with an insulated test
lead.
3 Enable the output
The OFF annunciator turns off and the 8V
annunciators turn on. The CV or CC annunciator is lit depending on the
resistance of the test lead. The flashing digit can be adjusted by turning the
knob. Notice that the display is in the meter mode. “Meter mode” means that
the display shows the actual output voltage and current.
or 25V/7A
[1]
or 25V
[2]
[2]
range is selected
, OVP, and OCP
4 Adjust the voltage limit to 1.0 volt
Set the display to the limit mode (the Limit annunciator will be flashing).
Adjust the voltage limit to 1.0 volt to assure CC operation. The CC annunciator
will light.
5 Check that the front-panel ammeter properly responds to knob control.
Set the knob to the current control, and turn the knob clockwise or counter
clockwise when the display is in the meter mode (the Limit annunciator is off).
Check that the ammeter responds to knob control and the voltmeter indicates
nearly zero (actually, the voltmeter will show the voltage drop caused by the
test lead)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
32Keysight E3633A and E3634A Service Guide
Quick Start2
NOTE
6 Ensure that the current can be adjusted from zero to the full rated value.
Adjust the knob until the ammeter indicates 0 amps and then until the
ammeter indicates 20.0 amps
[1]
or 7.0 amps
[2]
.
If an error has been detected during the output checkout procedures, the
ERROR annunciator will turn on. See “Error Messages’’ for more information,
starting on page 123 in chapter 5 of the User’s Guide.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide33
2Quick Start
NOTE
To Use the Power Supply in Constant Voltage Mode
To setup the power supply for constant voltage (CV) operation, proceed as
follows.
For each step, use the keys shown on the left margin.
1 Connect a load to the output terminals
With the power-off, connect a load to the (+) and (-) output terminals.
2 Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
control. Press
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or
[2]
key to operate the power supply in the 20V/
10A* or 50V/4A** range before proceeding to the next step. The 20V
[2]
annunciator turns on.
50V
3 Set the display to the limit mode
[1]
or 25V/7A
[2]
range is selected
[1]
or
Notice that the Limit annunciator blinks, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the power supply.
In constant voltage mode, the voltage values between the meter and limit
modes are the same, but the current values are not. Moreover, if the display is in
the meter mode, you cannot see the change of current limit value when
adjusting the knob. We recommend that you should set the display to “limit”
mode to see the change of current limit value in the constant voltage mode
whenever adjusting the knob.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
34Keysight E3633A and E3634A Service Guide
Quick Start2
4 Adjust the knob for the desired current limit.
Check that the Limit annunciator still blinks. Set the knob for current control.
The second digit of the ammeter will be blinking. The blinking digit can be
changed using the resolution selection keys and the blinking digit can be
adjusted by turning the knob. Adjust the knob to the desired current limit.
5 Adjust the knob for the desired output voltage.
Check that the Limit annunciator still blinks. Set the knob for voltage control.
The second digit of the voltmeter will be blinking. Change the blinking digit
using the resolution selection keys and adjust the knob to the desired output
voltage.
6 Return to the meter mode.
Press key or let the display time-out after several seconds to return to
the meter mode. Notice that the Limit annunciator turns off and the displays
shows “OUTPUT OFF” message.
7 Enable the output.
The OFF annunciator turns off and the 8V
[1]
(or 25V
[2]
) or 20V
[1]
(or 50V
OVP, OCP and CV annunciators are lit. Notice that the display is in the meter
mode. In the meter mode, the display shows the actual output voltage and
current.
[2]
),
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify that
the CV annunciator is lit. If the CC annunciator is lit, choose a higher current
limit.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide35
2Quick Start
NOTE
NOTE
During actual CV operation, if a load change causes the current limit to be
exceeded, the power supply will automatically crossover to the constant current
mode at the preset current limit and the output voltage will drop
proportionately.
You can use the resolution selection keys to move the flashing digit to the right
or left when setting the current.
36Keysight E3633A and E3634A Service Guide
To Use the Power Supply in Constant Current Mode
NOTE
To set up the power supply for constant current (CC) operation, proceed as
follows.
1 Connect a load to the output terminals.
With power-off, connect a load to the (+) and (-) output terminals.
2Turn on the power supply
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
[2]
annunciator turns on); and the knob is selected for voltage
control.
[1]
Press
[2]
50V
or
annunciator turns on.
[2]
key before proceeding to the next step. The 20V
3 Set the display to the limit mode.
[1]
or 25V/7A
Quick Start2
[2]
range is selected
[1]
or
Notice that the Limit annunciator blinks, indicating that the display is in the
limit mode. When the display is in the limit mode, you can see the voltage and
current limit values of the selected supply.
In constant current mode, the current values between the meter mode and limit
mode are the same, but the voltage values are not. Moreover, if the d isplay is in
the meter mode, you cannot see the change of voltage limit value when
adjusting the knob. We recommend that you should set the display to “limit”
mode to see the change of voltage limit value in the constant current mode
whenever adjusting the knob.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide37
2Quick Start
NOTE
4 Adjust the knob for the desired voltage limit.
5 Adjust the knob for the desired output current.
6 Return to the meter mode.
7 Enable the output.
Check that the Limit annunciator still blinks and the second digit of voltmeter
blinks to indicate the knob is selected for voltage control.The blinking digit can
be changed using the resolution keys and the blinking digit can be adjusted by
turning the knob. Adjust the knob for the desired voltage limit.
Check that the Limit annunciator still blinks. Set the knob for current control.
The second digit of the ammeter will be blinking. Change the blinking digit
using the resolution selection keys and adjust the knob to the desired output
current.
Press key or let the display time-out after several seconds to return to
the meter mode. Notice that the Limit annunciator turns off and the display
shows “OUTPUT OFF” message.
The OFF annunciator turns off and the 8V
[1]
(or 25V
[2]
) or 20V
[1]
(or 50V
[2]
),
OVP, OCP and CC annunciators are lit. Notice that the display is in the meter
mode. In the meter mode, the display shows the actual output voltage and
current.
8 Verify that the power supply is in the constant current mode.
If you operate the power supply in the constant current (CC) mode, verify that
the CC annunciator is lit. If the CV annunciator is lit, choose a higher voltage
limit.
During actual CC operation, if a load change causes the voltage limit to be
exceeded, the power supply will automatically crossover to constant voltage
mode at the preset voltage limit and the output current will drop
proportionately.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
38Keysight E3633A and E3634A Service Guide
To store and Recall the Instrument State
You can store up to three different operating states in non-volatile memory. This
also enables you to recall the entire instrument configuration with just a few key
presses from the front panel.
The memory locations are supplied with the reset states from the factory for
front-panel operation. Refer to the description of *RST command, starting on
page 84 in chapter 4 for more information. The following steps show you how to
store and recall an operating state.
1 Set up the power supply for the desired operating state.
The storage feature “remembers” output range selection, the limit value
settings of voltage and current, output on/off state, OVP and OCT on/off state
and OVP and OCP trip levels.
2 Turn on the storage mode.
Three memory locations (numbered 1, 2 and 3) are available to store the
operating states. The operating states are stored in non-volatile memory and
are remembered when being recalled.
Quick Start2
STORE 1
This message appears on the display for approximately 3 seconds.
3 Store the operating state in memory location “3”.
Turn the knob to the right to specify the memory location 3.
STORE 3
To cancel the store operation, let the display time-out after about 3 seconds
or press any other function key except the ion key except the key. The
power supply returns to the normal operating mode and to the function
pressed. Save the operating state.
Keysight E3633A and E3634A Service Guide39
2Quick Start
4 Save the operating state.
5 Turn on the recall mode.
6 Recall the stored operating state.
The operating state is now stored. To recall the stored state, go to the
following steps.
DONE
Memory location “1” will be displayed in the recall mode.
RECALL 1
This message appears on the display for approximately 3 seconds.
Turn the knob to the right to change the displayed storage location to 3.
RECALL 3
If this setting is not followed within 3 seconds with key stroke, the power
supply returns to normal operating mode and will not recall the instrument
state 3 from memory.
7 Restore the operating state.
The power supply should now be configured in the same state as when you
stored the state on the previous steps.
DONE
This message appears on the display for approximately 1 second.
40Keysight E3633A and E3634A Service Guide
To Program Overvoltage Protection
Overvoltage protection guards the load against output voltages that reach a
specified value greater than the programmed protection level. It is accomplished
by shorting the output via an internal SCR when the trip level is set to equal or
greater than 3 volts, or by programming the output to 1 volt when the trip level is
set to less than 3 volts.
The following steps show how to set the OVP trip level, how to check OVP
operation, and how to clear overvoltage condition.
Setting the OVP level and enable the OVP circuit
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
control.
2 Enable the output.
The OFF annunciator turns off and the display will go to the meter mode.
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or 25V/7A
Quick Start2
[2]
range is selected
3 Enter the OVP menu and set the trip level.
LEVEL 22.0V (E3633A)
LEVEL 55.0V (E3634A)
You will see the above message on the display when you enter the OVP menu.
Adjust the control knob for the desired OVP trip level.
Note that you cannot set the trip levels to lower than 1.0 volt.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide41
2Quick Start
4 Enable the OVP circuit.
5Exit the OVP menu.
Checking OVP operation
To check OVP operation, raise the output voltage to near the trip point. Then very
gradually increase the output by turning the knob until the OVP circuit trips. This
will cause the power supply output to drop to near zero, the OVP annunciator to
blink, and the CC annunciator to turn on. The “OVP TRIPPED” message also
appears on the display.
OVP ON
You will see the above message after pressing key.
CHANGED
The “CHANGED” message is highlighted for a second to show that the new
OVP trip level is now in effect. If the OVP settings are not changed, “NO
CHANGE” will be displayed. The power supply will exit the OVP menu and the
display will return to the meter mode. Check that the OVP annunciator turns
on.
Clearing the overvoltage condition
When the OVP condition occurs (the “OVP TRIPPED” message is shown on the
display), the OVP annunciator blinks. When it was caused by an external voltage
source such as a battery, disconnect it first. Clear the overvoltage condition by
adjusting output voltage level or by adjusting OVP trip level.
The following steps show how to clear the overvoltage condition and get back to
normal mode operation. In the following steps, the display will go back to “OVP
TRIPPED” if you let the display time out after about several seconds.
42Keysight E3633A and E3634A Service Guide
Quick Start2
– Adjust the output voltage level
1 Lower the output voltage level.
Lower the output voltage level below the OVP trip point after pressing
key. The OVP and Limit annunciators are blinking.
2 Move to the clear mode.
OVP CLEAR
Press key twice to move to the OVP CLEAR mode. The “OVP ON”
message appears on the display. Turn the knob to the right until the above
message appears on the display.
3 Clear the overvoltage cond ition and exit this menu.
Now, when you press key again, the “DONE” message is displayed for
a second and the OVP annunciator will not blink any more. The output will
return to meter mode.
Keysight E3633A and E3634A Service Guide43
2Quick Start
– Adjust OVP trip level
1Raise the OVP trip level.
Press key and turn the knob to raise the OVP trip level.
2 Move to the OVP CLEAR mode.
OVP CLEAR
Press key to move the OVP clear mode. The “OVP ON” message
appears on the display. Turn the knob to the right until the above message
appears on the display.
3 Clear the overvoltage condition and exit this menu.
Now, when you press key again, the “DONE’’ message is displayed for
a second and the OVP annunciator will not blink any more. The output will
return to the meter mode.
44Keysight E3633A and E3634A Service Guide
To Program Overcurrent Protection
Overcurrent protection guards the load against output currents that reach a
specified value greater than the programmed protection level. It is accomplished
by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how to
check OCP operation and how to clear overcurrent condition.
Setting the OCP level and enable the OCP circuit
1Turn on the power supply.
The power supply will go into the power-on / reset state; the output is disabled
(the OFF annunciator turns on); the 8V/20A
(the 8V
[1]
or 25V
control.
2 Enable the output.
The OFF annunciator turns off and the display will go to the meter mode.
[2]
annunciator turns on); and the knob is selected for voltage
[1]
or 25V/7A
Quick Start2
[2]
range is selected
3 Enter the OCP menu and set the trip level.
LEVEL 22.0 A (E3633A)
LEVEL 7.5 A (E3634A)
You will see the above message on the display when you enter the OCP menu.
Adjust the knob for the desired OCP trip level.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide45
2Quick Start
4Enable the OCP circuit
5 Exit the OCP menu.
Checking OCP operation
To check OCP operation, raise the output current to near the trip point. Then very
gradually increase the output by turning the knob until the OCP circuit trips. This
will cause the power supply’s output current to drop to zero and the OCP
annunciator to blink. The “OCP TRIPPED” message also appears on the display.
OCP ON
You will see the above message after pressing the key.
CHANGED
The “CHANGED” message is displayed for a second to show that the new OCP
trip level is now in effect. If the OCP settings are not changed, “NO CHANGE”
will be displayed. The power supply will exit the OCP menu and the display will
return to the meter mode. Check that the OCP annunciator turns on.
Clearing the overcurrent condition
When the OCP condition occurs (the “OCP TRIPPED” message is shown on the
display), the OCP annunciator blinks. When it was caused by external voltage
sources such as a battery, disconnect it first. Clear the overcurrent condition by
adjusting output current level or by adjusting OCP trip level.
The following steps show how to clear the overcurrent condition and get back to
normal mode operation. In the following steps, the display will go back to “OCP
TRIPPED” if you let the display time out after about several seconds.
46Keysight E3633A and E3634A Service Guide
Quick Start2
– Adjust output current level
1 Lower the output current level.
Press key and set the knob for current control by pressing key
then lower the output current level below the OCP trip point.
2 Move to the clear mode.
OCP CLEAR
Press key twice to move to the OCP CLEAR mode. The “OCP ON”
message appears on the display. Turn the knob to the right until the above
message appears on the display
3 Clear the overcurrent condition and exit this menu.
Now, when you press key again, the “DONE” message is displayed for
a second and the OCP annunciator will not blink any more. The output will
return to meter mode. The knob is selected for current control.
Notice that the power supply is operated in the constant current (CC)
mode.
Keysight E3633A and E3634A Service Guide47
2Quick Start
– Adjust OCP trip level
1 Raise the OCP trip level.
Press key and turn the knob to raise the OCP trip level.
2 Move to the OCP CLEAR mode.
OCP CLEAR
Press the key to move to the OCP CLEAR mode. The “OCP ON” message
appears on the display. Turn the knob to the right until the above message
appears on the display.
3 Clear the overcurrent condition and exit this menu.
Now, when you press again, the “DONE” message is displayed for a
second and the OCP annunciator will not blink any more. The output will
return the meter mode.
48Keysight E3633A and E3634A Service Guide
To Rack Mount the Power Supply
The power supply can be mounted in a standard 19-inch rack cabinet using one of
three optional kits available. A rack-mounting kit for a single instrument is
available as Option 1CM (P/N 5063-9243). Installation instructions and hardware
are included with each rack-mounting kit. Any Keysight System II instrument of
the same size can be rack-mounted beside the Keysight E3633A or E3634A DC
power supply.
Remove the front and rear bumpers before rack-mounting the power supply.
Quick Start2
To remove the rubber bumper, pull the bumper off from the top as there are
protrusions on the sides and bottom of the cover.
To rack mount a single instrument, order adapter kit 5063-9243.
Keysight E3633A and E3634A Service Guide49
2Quick Start
To rack mount two instruments of the same depth side-by-side, order lock-link
kit 5061-9694 and flange kit 5063-9214.
To install two instruments in a sliding support shel f, order support shel f
5063-9256, and slide kit 1494-0015.
50Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies
Service Guide
3Calibration Procedures
Keysight Technologies Calibration Services 52
Calibration Interval 52
Automating Calibration Procedures 53
Test Considerations 53
Recommended Test Equipment 54
Performance Verification Tests 55
Measurement Techniques 56
Constant Voltage (CV) Verifications 59
Constant Current (CC) Verifications 65
Common Mode Current Noise 70
Performance Test Record for Keysight E3633A and E3634A 71
Calibration Security Code 73
Calibration Count 77
Calibration Message 78
General Calibration/ Adjustment Procedure 79
Aborting a Calibration in Progress 85
Calibration Record for Keysight E3633A/E3634A 86
Error Messages 87
An Example Program of Excel 97 for Calibration 90
This chapter contains procedures for verification of the power supply’s
performance and calibration (adjustment).
The performance verification tests for constant voltage (CV) and constant current
(CC) operations use the power supply’s specifications listed in “Specifications” on
page 17.
51
3Calibration Procedures
NOTE
If you calibrate the power supply over the remote interface, you must send the
*RST
command to the power supply or turn the power supply off and on again
after performing a calibration to ensure proper power supply operation.
Closed-Case Electronic Calibration
The power supply features closed-case electronic calibration since no internal
mechanical adjustments are required for normal calibration. The power supply
calculates correction factors based upon the input reference value you enter. The
new correction factors are stored in non-volatile memory until the next calibration
adjustment is performed. (Non-volatile memory does not change when power has
been off or after a remote interface reset.)
Keysight Technologies Calibration Services
When your power supply is due for calibration, contact your local Keysight
Technologies Service Center for a low-cost calibration. The Keysight E3633A and
E3634A power supplies are supported on calibration processes which allow
Keysight Technologies to provide this service at competitive prices.
Calibration Interval
The power supply should be calibrated on a regular interval determined by the
accuracy requirements of your application. A 1-year interval is adequate for most
applications. Keysight Technologies does not recommend extending calibration
intervals beyond 1 year for any application. Keysight Technologies recommends
that complete re-adjustment should always be performed at the calibration
interval. This will increase your confidence that the Keysight E3633A and E3634A
will remain within specification for the next calibration interval. This criteria for
re-adjustment provides the best long-term stability.
52Keysight E3633A and E3634A Service Guide
Automating Calibration Procedures
CAUTION
You can automate the complete verification procedures outlined in this chapter if
you have access to programmable test equipment. You can program the
instrument configurations specified for each test over the remote interface. You
can then enter readback verification data into a test program and compare the
results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote
operation is similar to the local front-panel procedure. You can use a computer to
perform the adjustment by first selecting the required setup. The calibration value
is sent to the power supply and then the calibration is initiated over the remote
interface. The power supply must be unsecured prior to initiating the calibration
procedure. An example program of Excel 97 for calibration over the GPIB interface
is listed at the end of this chapter.
For further details on programming the power supply, see chapters 3 and 4 in the
Keysight E3633A and E3634A User’s Guide.
Test Considerations
Calibration Procedures3
To ensure proper instrument operation, verify that you have selected the correct
power-line voltage prior to attempting any test procedure in this chapter. See
page 30in Chapter 2 for more information.
Ensure that all connections of terminals (both front panel and rear panel) are
removed while the power supply internal self-test is being performed.
For optimum performance verification, all test procedures should comply with the
following recommendations:
– Assure that the calibration ambient temperature is stable and between 20°C
and 30°C.
– Assure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Use short cables to connect test set-ups.
The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the
power supply.
Keysight E3633A and E3634A Service Guide53
3Calibration Procedures
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available, use
the accuracy requirements shown to select substitute calibration standards. If you
use equipment other than that recommended in Table 3-1, you must recalculate
the measurement uncertainties for the actual equipment used.
Tab le 3-1Recommended test equipment
InstrumentRequirementsRecommended
Model
GPIB controllerFull GPIB or RS-232 capabilities Keysight 82341C
interface card
Oscilloscope100 MHz with 20 MHz
bandwidth
RMS voltmeter20 Hz to 20 MHzMeasure rms ripple & noise
[c] It is recommended to use a current monitoring resistor after calibration to find the accurate resistance.
54Keysight E3633A and E3634A Service Guide
Performance Verification Tests
The performance verification tests use the power supply’s specifications listed in
“Specifications” on page 17.
You can perform two different levels of performance verification tests:
– Self-Test
A series of internal verification tests that provide high confidence that the
power supply is operational.
– Performance Verification Tests
These tests are used to verify that the power supply is operating as specified in
the performance specifications.
Self-test
A power-on self-test occurs automatically when you turn on the power supply.
This limited test assures you that the power supply is operational.
The complete self-test is enabled by holding down key (actually any front
panel keys except key) as you turn on the power supply and hold down the
key until you hear a long beep. The self-test will begin when you release the key
following the beep. The complete self-test takes approximately two seconds to
execute.
You can also perform a self-test from the remote interface (see chapter 3 in the Keysight E3633A and E3634A User’s Guide).
– If the self-test is successful, “PASS” is displayed on the front panel.
Calibration Procedures3
– If the self-test fails, “FAIL” is displayed and the ERROR annunciator turns on. If
repair is required, see chapter 5, “Service”, for further details.
– If self-test passes, you have a high confidence that the power supply is
operational.
Performance verification tests
These tests should be used to verify the power supply’s specifications following
repairs to specific circuits. The following sections explain all verification
procedures in detail. All of the performance test limits are shown in each test.
Keysight E3633A and E3634A Service Guide55
3Calibration Procedures
Measurement Techniques
Setup for most tests
Most tests are performed at the front terminals as shown in the following figure.
Measure the dc voltage directly at the (+) and (–) terminals on the front panel.
Figure 3-1Performance verification test setup
Electronic load
Many of the test procedures require the use of a variable load resistor capable of
dissipating the required power. Using a variable load resistor requires that
switches should be used to connect, disconnect, and short the load resistor. An
electronic load, if available, can be used in place of a variable load resistor and
switches. The electronic load is considerably easier to use than load resistors. It
eliminates the need for connecting resistors or rheostats in parallel to handle
power, it is much more stable than carbon-pile load, and it makes easy work of
switching between load conditions as is required for the load regulation and load
response tests. Substitution of the electronic load requires minor changes to the
test procedures in this chapter.
56Keysight E3633A and E3634A Service Guide
General measurement techniques
To achieve best results when measuring load regulation, peak to peak voltage,
and transient response time of the power supply, measuring devices must be
connected through the hole in the neck of the binding post at (A) while the load
resistor is plugged into the front of the output terminals at (B). A measurement
made across the load includes the impedance of the leads to the load. The
impedance of the load leads can easily be several orders of the magnitude greater
than the power supply impedance and thus invalidate the measurement. To avoid
mutual coupling effects, each measuring device must be connected directly to the
output terminals by separate pairs of leads.
To eliminate output current measurement error caused by the voltage drops in the
leads and connections, connect the current monitoring resistor between the (-)
output terminal and the load as a four-terminal device. Connect the
current-monitoring leads inside the load-lead connections directly at the
monitoring points on the resistor element (see R
Keysight E3633A and E3634A Service Guide57
in Figure 3-1).
M
3Calibration Procedures
Programming
Most performance tests can be performed from the front panel. However, an GPIB
or RS-232 controller is required to perform the voltage and current programming
accuracy and readback accuracy tests.
The test procedures are written assuming that you know how to program the
power supply either from the front panel or from an GPIB or RS-232 controller.
Complete instructions on front panel and remote programming are given in the
Keysight E3633A and E3634A User’s Guide.
58Keysight E3633A and E3634A Service Guide
Constant Voltage (CV) Verifications
Constant voltage test setup
If more than one meter or if a meter and an oscilloscope are used, connect each to
the (+) and (–) terminals by a separate pair of leads to avoid mutual coupling
effects. Use coaxial cable or shielded 2-wire cable to avoid noise pick-up on the
test leads.
Voltage programming and readback accuracy
This test verifies that the voltage programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the remote
interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test to
avoid round off errors.
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(–) terminals of the output to be tested as shown in Figure 3-1
2 Turn on the power supply. Select the 20V/10A
enable the output by sending the commands:
Calibration Procedures3
[1]
or 50V/4A
[2]
range and
VOLT:RANG P20V (E3633A)
VOLT:RANG P50V (E3634A)
OUTP ON
3 Program the output voltage to zero volt and current to full rated value
(10.0 A)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or (4.0 A)
VOLT 0
CURR 10 (E3633A)
CURR 4 (E3634A)
[2]
by sending the commands:
Keysight E3633A and E3634A Service Guide59
3Calibration Procedures
4 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (0 V ± 10 mV). Also, note that the CV, Adrs, Limit,
and Rmt annunciators are on.
5 Readback the output voltage over the remote interface by sending the
command:
MEAS:VOLT?
6 Record the value displayed on the controller. This value should be within the
limit of (DVM ±5 mV).
7 Program the output voltage to full rated value (20.0 V)
sending the command:
VOLT 20.0 (E3633A)
VOLT 50.0 (E3634A)
8 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (20 V ± 20 mV)
9 Readback the output voltage over the remote interface by sending the
command:
[1]
or (50.0 V)
[1]
or (50 V ± 35 mV)
[2]
[2]
by
MEAS:VOLT?
10 Record the value displayed on the controller. This value should be within the
limit of (DVM ± 15 mV)
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or (DVM ± 30 mV)
[2]
.
60Keysight E3633A and E3634A Service Guide
CV load effect (load regulation)
This test measures the immediate change in the output voltage resulting from a
change in the output current from full to no load.
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output as shown in Figure 3-1.
2 Turn on the power supply. Select the 20V/10A
output, and set the display to the limit mode. When the display is in the limit
mode, program the output current to the maximum programmable value and
the voltage to the full rated value (20.0 V)
3 Operate the electronic load in constant current mode and set its current to
[1]
(10.0 A)
or (4.0 A)
not lit, adjust the load so that the output current drops slightly until the CV
annunciator lights. Record the output voltage reading on the digital voltmeter.
4 Operate the electronic load in open mode (input off). Record the output
voltage reading on the digital voltmeter immediately. The difference between
the digital voltmeter readings in steps (3) and (4) is the CV load regulation. The
difference of the readings during the immediate change should be within the
limit of (4 mV)
[2]
[1]
or (7 mV)
. Check that the front panel CV annunciator remains lit. If
[2]
[1]
or 50V/4A
[1]
or (50.0 V)
Calibration Procedures3
[2]
range, enable the
[2]
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide61
3Calibration Procedures
CV source effect (line regulation)
This test measures the immediate change in output voltage that results from a
change in ac line voltage from the minimum value (10% below the nominal input
voltage) to maximum value (10% above the nominal input voltage).
1 Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in Figure 3-1.
2 Connect the ac power line through a variable voltage transformer.
3 Turn on the power supply. Select the 20V/10A
the output, and set the display to the limit mode. When the display is in the
limit mode, program the current to the maximum programmable value and the
voltage to full rated value (20.0 V)
4 Operate the electronic load in constant current mode and set its current to
(10.0 A)
adjust the load so that the output current drops slightly until the CV
annunciator lights.
5 Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the
output reading on the digital voltmeter.
6 Adjust the transformer to high line voltage (127 Vac for nominal 115 Vac, 110
Vac for nominal 100 Vac, or 253 Vac for nominal 230 Vac). Record the voltage
reading on the digital voltmeter immediately. The difference between the
digital voltmeter readings in steps (5) and (6) is the CV line regulation. The
difference of the readings during the immediate change should be within the
limit of (4 mV)
[1]
or (4.0 A)
[2]
[1]
or (7 mV)
[1]
or (50.0 V)
[1]
or 50V/4A
[2]
.
[2]
range, enable
. Check that the CV annunciator remains lit. If not lit,
[2]
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
62Keysight E3633A and E3634A Service Guide
CV PARD (ripple and noise)
Periodic and random deviations (PARD) in the output (ripple and noise) combine
to produce a residual ac voltage superimposed on the dc output voltage. CV
PARD is specified as the rms or peak-to-peak output voltage in the frequency
range from 20 Hz to 20 Mhz.
VRMS measurement techniques
When measuring Vrms ripple and noise, the monitoring device should be plugged
into the front of the terminals at (B) in Figure 3-2. Use the coaxial cable and BNC
female to banana plug adapter to connect the monitor device to the power
supply. To reduce the measurement error caused by common mode noise, it is
recommended that you use a common mode choke between the cable and the
BNC adapter. The common mode choke is constructed by inserting coaxial cable
into Ferrite Toroidal core. The load resistor should be connected to the terminal at
(A) in Figure 3-2. Twisted leads between the load resistor and the power supply
helps reduce noise pickup for these measurements.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 to an oscilloscope (ac coupled) between (+) and (-) terminals. Set
the oscilloscope to AC mode and bandwidth limit to 20 MHz. Connect a
resistive load (2.0
2 Turn on the power supply. Select the 20V/10A
the output, and set the display to the limit mode. When the display is in the
limit mode, program the current to the full rated value (10.0 A)
and the voltage to the full rated value (20.0 V)
3 Check that the front panel CV annunciator remains lit. If not lit, adjust the load
down slightly.
[1]
Ω
)
or (12.5 Ω)
Calibration Procedures3
[2]
to the terminal at (A) in Figure 3-2.
[1]
or 50V/4A
[1]
or (50.0 V)
[2]
range, enable
[1]
or (4.0 A)
[2]
.
[2]
4 Note that the waveform on the oscilloscope does not exceed the peak-to-peak
limit of 3 mV.
5 Disconnect the oscilloscope and connect the ac rms voltmeter in its place
according to the VRMS measurement techniques above. The rms voltage
reading does not exceed the rms limit of 0.35 mV
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
[1]
or 0.5 mV
[2]
.
Keysight E3633A and E3634A Service Guide63
3Calibration Procedures
Load transient response time
This test measures the time for the output voltage to recover to within 15 mV of
nominal output voltage following a load change from full load to half load, or half
load to full load.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with an oscilloscope. Operate the electronic load in constant
current mode.
2 Turn on the power supply. Select 20V/10A
output, and set the display to the limit mode. When the display is in the limit
mode, program the current to the full rated value (10.0 A)
the voltage to the full rated value (20.0 V)
3 Set the electronic load to transient operation mode between one half of the
output’s full rated value and the output’s full rated value at a 1 kHz rate with
50% duty cycle.
4 Set the oscilloscope for ac coupling, internal sync, and lock on either the
positive or negative load transient.
5 Adjust the oscilloscope to display transients as shown in Figure 3-3. Note that
the pulse width (t2 - t1) of the transients at 15 mV from the base line is no
more than 50 μsec for the output.
[1]
or 50V/4A
[1]
or (50.0 V)
[2]
range, enable the
[2]
[1]
.
or (4.0 A)
[2]
and
Figure 3-3Transient response time
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
64Keysight E3633A and E3634A Service Guide
Constant Current (CC) Verifications
Constant current test setup
Follow the general setup instructions in the Measurement Techniques and the
specific instructions will be given in the following paragraphs.
Current programming and readback accuracy
This test verifies that the current programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the remote
interface should be identical to those displayed on the front panel. The accuracy
of the current monitoring resistor must be 0.01% or better.
You should program the power supply over the remote interface for this test to
avoid round off errors.
1 Turn off the power supply and connect a 0.01 Ω current monitoring resistor
(R
) across the output to be tested and a digital voltmeter across the current
M
monitoring resistor (R
2 Turn on the power supply. Select the 8V/20A
the output by sending the commands:
).
M
Calibration Procedures3
[1]
or 25V/7A
[2]
range and enable
VOLT:RANG P8V (E3633A)
VOLT:RANG P25V(E3634A)
OUTP ON
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide65
3Calibration Procedures
3 Program the output voltage to full rated voltage (8.0 V)
[1]
or (25.0 V)
[2]
and
output current to zero amps by sending the commands:
VOLT 8 (E3633A)
VOLT 25 (E3634A)
CURR 0
4 Divide the voltage drop (DVM reading) across the current monitoring resistor
(R
) by its resistance to convert to amps and record this value (IO). This value
M
should be within the limit of (0 A ± 10 mA). Also, note that the CC, Adrs, Limit,
and Rmt annunciators are on.
5 Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
6 Record the value displayed on the controller. This value should be within the
limit of (I
7 Program the output current to the full rated value (20.0 A)
± 5 mA).
O
[1]
or (7.0 A)
[2]
by
sending the command:
CURR 20.0 (E3633A)
CURR 7.0 (E3634A)
8 Divide the voltage drop (DVM reading) across the current monitoring resistor
(RM) by its resistance to convert to amps and record this value (I
[1]
This value should be within the limit of (20 A ± 50 mA)
or (7A ± 24 mA)
).
O
[2]
.
9 Readback the output current over the remote interface by sending the
command:
MEAS:CURR?
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
66Keysight E3633A and E3634A Service Guide
10 Record the value displayed on the controller. This value should be within the
limit of (I
± 35 mA)
O
[1]
or (IO ± 15.5 mA)
CC load effect (load regulation)
This test measures the immediate change in output current resulting from a
change in the load from full rated output voltage to short circuit.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with the digital voltmeter connected across the 0.01 Ω current
monitoring resistor (R
2 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit
mode, program the output voltage to the maximum programmable value and
the output current to the full rated value (20.0 A)
3 Set the voltage of the electronic load to (6.0 V)
constant voltage mode since a voltage drop occurs on the load wires when
[1]
(20.0 A)
or (7.0 A)
is on. If it is not, adjust the load so that the output voltage drops slightly.
Record the current reading by dividing the voltage reading on the digital
voltmeter by the resistance of the current monitoring resistor.
4 Operate the electronic load in short (input short) mode. Record the current
reading immediately by dividing the voltage reading on the digital voltmeter
by the resistance of the current monitoring resistor. The difference between
the current readings in step (3) and (4) is the load regulation current. The
difference of the readings during the immediate change should be within the
limit of (2.25 mA)
).
M
[2]
flows on the load wires. Check that the CC annunciator
[1]
or (0.95 mA)
[2]
Calibration Procedures3
[2]
.
[1]
or 25V/7A
[1]
[1]
or (23.0 V)
[2]
or (7.0 A)
range, enable the
[2]
.
[2]
to operate it in
.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide67
3Calibration Procedures
CC source effect (line regulation)
This test measures the immediate change in output current that results from a
change in ac line voltage from the minimum value (10% below the nominal input
voltage) to the maximum value (10% above nominal voltage).
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1 with the digital voltmeter connected across the current monitoring
resistor (R
2 Connect the ac power line through a variable voltage transformer.
3 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit
mode, program the voltage to the full rated value (8.0 V)
the current to the full rated value (20.0 A)
4 Operate the electronic load in constant voltage mode and set its voltage to
(8.0 V)
adjust the load so that the output voltage drops slightly until the CC
annunciator lights.
5 Adjust the transformer to low line voltage limit (104 Vac for nominal 115 Vac,
90 Vac for nominal 100 Vac, or 207 Vac for nominal 230 Vac). Record the
output current reading by dividing the voltage reading on the digital voltmeter
by the resistance of the current monitoring resistor.
6 Adjust the transformer to 10% above the nominal line voltage (127 Vac for a
115 Vac nominal input, 110 Vac for a 100 Vac nominal input or 253 Vac for a
230 Vac nominal input). Record the current reading again immediately by
dividing the voltage reading on the digital voltmeter by the resistance of the
current monitoring resistor. The difference between the current readings in
step (5) and (6) is the load regulation current. The difference of the readings
during the immediate change should be within the limit of (2.25 mA)
[2]
mA)
).
M
[1]
or (25.0 V)
.
[1]
or 25V/7A
[1]
or (7.0 A)
[2]
. Check that the CC annunciator remains lit. If not lit,
[2]
range, enable the
[1]
or (25.0 V)
[2]
.
[1]
[2]
and
or (0.95
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
68Keysight E3633A and E3634A Service Guide
CC PARD (ripple and noise)
Periodic and random deviations (PARD) in the output (ripple and noise) combine
to produce a residual ac current, as well as an ac voltage superimposed on the dc
output. CC PARD is specified as the rms output current in a frequency range 20 Hz
to 20 MHz with the power supply in constant current operation.
1 Turn off the power supply and connect the output to be tested as shown in
Figure 3-1with a load resistor (0.4 Ω)
Connect a rms voltmeter across the load resistor. Use only a resistive load for
this test.
2 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit
mode, program the current to full rated value (20.0 A)
voltage to the full rated value (8.0 V)
3 The output current should be at the full rated rating with the CC annunciator
on. If not lit, adjust the load so that the output voltage drops slightly until the
CC annunciator lights.
4 Divide the reading on the rms voltmeter by the load resistance to obtain rms
current. The readings should be within the limit of 2 mA.
[1]
or (3.5 Ω)
[1]
or (25.0 V)
Calibration Procedures3
[2]
across output terminals.
[1]
or 25V/7A
[2]
[1]
or (7.0 A)
.
[2]
range, enable the
[2]
and the
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide69
3Calibration Procedures
Common Mode Current Noise
The common mode current is that ac current component which exists between
the output or output lines and chassis ground. Common mode noise can be a
problem for very sensitive circuitry that is referenced to earth ground. When a
circuit is referenced to earth ground, a low level line-related ac current will flow
from the output terminals to earth ground. Any impedance to earth ground will
create a voltage drop equal to the output current flow multiplied by the
impedance.
1 Turn off the power supply and connect a 100 kΩ resistor (R
capacitor in parallel between the (-) terminal and chassis ground.
2 Connect a digital voltmeter across R
3 Turn on the power supply. Select the 8V/20A
output, and set the display to the limit mode. When the display is in the limit
mode, program the output to the full rated value (8.0 V and 20.0 A)
V and 7.0 A)
4 Record the voltage across R
resistance (DVM reading/100 kΩ). Note that the current is less than 1.5 μA.
[2]
.
) and a 2200 pF
S
.
S
[1]
or 25V/7A
and convert it to current by dividing by the
S
[2]
range, enable the
[1]
or (25.0
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
70Keysight E3633A and E3634A Service Guide
Calibration Procedures3
Performance Test Record for Keysight E3633A and E3634A
CV Programming Accuracy
@ Full Scale (DVM reading)
CV Readback Accuracy @ Full Scale
CV Load Effect (Load Regulation)
CV Source Effect (Line Regulation)
CV PARD (Normal mode)
Load Transient Response TimeBoth
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
Both+0.0100 V-0.0100 V
(E3633A)+20.0200 V19.9800 V
(E3634A)+50.0350 V49.9650 V
(E3633A)DVM + 0.0150 VDVM - 0.0150 V
(E3634A)DVM + 0.0300 VDVM - 0.0300 V
Actual
Result
Upper LimitLower Limit
Maximum change: < (4 mV)
Maximum change: < (4 mV)
< 3 mV p-p, (350
Specifications
[a]
μ
V)
or (500 μV)
< 50 μsec
[a]
or (7 mV)
[a]
or (7 mV)
[b]
[b]
[b]
rms
Keysight E3633A and E3634A Service Guide71
3Calibration Procedures
CC performance test records
Test DescriptionModels
CC Programming Accuracy @ 0
amps (IO)
Both+0.0100 A-0.0100 A
CC Readback Accuracy @ 0 amps Both
CC Programming Accuracy @ Full
Scale (IO)
CV Readback Accuracy @ Full
Scale
(E3633A)20.0500 A19.9500 A
(E3634A)7.0230 A6.9770 A
(E3633A)
(E3634A)
CC Load Effect (Load Regulation)
CC Source Effect (Line Regulation)
Actual
Result
Upper LimitLower Limit
I
O
IO + 0.0350 AIO - 0.0350 A
I
O
Maximum change: < (2.25 mA)
Maximum change: < (2.25 mA)
Specifications
+ 0.0050 AIO - 0.0050 A
+ 0.0155 AIO - 0.0155 A
CC PARD (Normal mode)Both< 2 mA rms
CC PARD (Common mode)Both
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
< 1.5
μ
A rms
[a]
or (0.95 mA)
[a]
or (0.95 mA)
[b]
[b]
72Keysight E3633A and E3634A Service Guide
Calibration Security Code
NOTE
This feature allows you to enter a security code (electronic key) to prevent
accidental or unauthorized calibrations of the power supply. When you first
receive your power supply, it is secured. Before you can calibrate the power
supply, you must unsecure it by entering the correct security code. A procedure to
unsecure the power supply is given on the following page.
– The security code is set to “HP003633”
supply is shipped from the factory. The security code is stored in non-volatile
memory, and does not change when power has been off or after a remote
interface reset.
– To secure the power supply from the remote interface, the security code may
contain up to 12 alphanumeric characters as shown below. The first character
must be a letter, but the remaining characters can be letters or numbers. You
do not have to use all 12 characters but the first character must always be a
letter.
A _ _ _ _ _ _ _ _ _ _ _ (12 characters)
– To secure the power supply from the remote interface so that it can be
unsecured from the front panel, use the eight-character format shown below.
The first two characters must be “H P” and the remaining characters must be
numbers. Only the last six characters are recognized from the front panel, but
all eight characters are required. To unsecure the power supply from the front
panel, omit the “H P” and enter the remaining numbers as shown on the
following pages.
Calibration Procedures3
[1]
or “HP003634”
[2]
when the power
H P _ _ _ _ _ _ (8 characters)
If you forget your security code, you can disable the security feature by adding a
jumper inside the power supply, and then entering a new code.
See“To unsecure the power supply without the security code” on page 76
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
Keysight E3633A and E3634A Service Guide73
3Calibration Procedures
To unsecure the power supply for calibration
The power supply can use a calibration security code to prevent unauthorized or
accidental calibration. This procedure shows you how to unsecure the power
supply for calibration from the front panel.
1 Turn on the front-panel calibration mode.
SECURED
Turn on the calibration mode by holding down (Calibrate) key as you turn
on the power supply and hold down the key until you hear a long beep.
If the power supply is secured, you will see the above message from the front
panel for approximately one second. The “CAL MODE” message is then
displayed on the front panel.
2 Move to the security code by pressing (Secure) key.
000000 CODE
3 Enter the security code using the knob and resolution selection keys.
003633 CODE (E3633A)
003634 CODE (E3634A)
The security code is set to “HP003633”
[1]
or “HP003634”
[2]
when the power
supply is shipped from the factory. The security code is stored in non-volatile
memory and does not change when the power has been off or after a remote
interface reset.
– To enter the security code from the front panel, enter only the last six
digits.
– To enter the security code from the remote interface, you may enter up to
12 characters.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
74Keysight E3633A and E3634A Service Guide
Calibration Procedures3
NOTE
Use the resolution selection keys to move left or right between digits. Use the
knob to change the digits. Notice that the security code may be different if the
security code has been changed from the default setting.
4 Unsecure the power supply.
UNSECURED
The power supply is unsecured when you press (Secure) key. You will see
the above message from the front panel for one second. The “CAL MODE”
message is displayed on the front panel after above message.
5 Turn off the calibration mode.
Turn off the power supply to exit the calibration mode.
To re-secure the power supply (following calibration), perform this procedure
again.
Keysight E3633A and E3634A Service Guide75
3Calibration Procedures
To unsecure the power supply without the security code
To unsecure the power supply without the correct security code (when you forget
the security code), follow the steps below. See Electrostatic Discharge (ESD)
Precautions in chapter 5 before beginning this procedure.
1 Disconnect the power cord and all load connections from the power supply.
2 Remove the instrument cover.
3 Connect the power cord and turn on the calibration mode by holding down
(Calibrate) key as you turn on the power supply and hold down the key
until you hear a long beep. Be careful not to touch the power line connections.
4 Apply a short between the two exposed metal pads on JP5
U13). The JP5 is outlined with a circle on the“E3633-60002/E3634-60002
Component Locator for the Main Board” on page 125.
5 While maintaining the short, move to the security code and enter any unsecure
code in the calibration mode. The power supply is now unsecured.
6 Remove the short at JP5
7 Turn off and reassemble the power supply.
[2]
. (An error occurs if not removed.)
[1]
(located near
Now you can enter a new security code. Be sure you take note of the new
security code.
[1] For serial MY53xx6xxx, apply a short between the two exposed metal pads on JP102 (located near
U103). The JP102 is outlined with a circle on the component locator drawing on page 128
[2] For serial MY53xx6xxx, remove the short at JP102.
76Keysight E3633A and E3634A Service Guide
Calibration Count
The calibration count feature provides an independent “serialization” of your
calibrations. You can determine the number of times that your power supply has
been calibrated. By monitoring the calibration count, you can determine whether
an unauthorized calibration has been performed. Since the value increments by
one for each calibration parameter (see Table 3-2 on the next page), a complete
calibration increases the value by 5 counts.
– The calibration count is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset. Your power supply
was calibrated before it left the factory. When you receive the power supply,
read the calibration count to determine its value.
– The calibration count increments up to a maximum of 32,767 after which it
wraps around to 0. No way is provided to program or reset the calibration
count.
Calibration Procedures3
Keysight E3633A and E3634A Service Guide77
3Calibration Procedures
Calibration Message
You can use the calibration message feature to record calibration information
about your power supply. For example, you can store such information as the last
calibration date, the next calibration due date, the power supply’s serial number,
or even the name and phone number of the person to contact for a new
calibration.
You can record and read information in the calibration message from the remote
interface only.
– The calibration message may contain up to 40 characters.
– The calibration message is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset.
– Remote Interface Operation
CAL:SEC:STAT {OFF|ON},<code>Secure or unsecure the power supply
CAL:SEC:CODE <new code>Change the security code
CAL:STR <quoted string>Store the calibration message
CAL:STR?Query the calibration information
See “Calibration Overview”, starting on page 68 in chapter 3 of the User’s Guide
for more details.
78Keysight E3633A and E3634A Service Guide
General Calibration/ Adjustment Procedure
NOTE
NOTE
The calibration procedures from the front panel are described in this section. For
voltage calibration, disconnect all loads from the power supply and connect a
DVM across the output terminals. For current calibration, disconnect all loads
from the power supply, connect an appropriate current monitoring resistor 0.01
across the output terminals, and connect a DVM across the terminals of the
monitoring resistor.
Note that the power supply should be calibrated after 1-hour warm-up with no
load connected.
The following table shows calibration parameters and points which should be
used to calibrate the output voltage and current.
Table 3-2Parameters for calibration
Calibration ParameterVoltage/ CurrentCalibration Point Mnemonic
CAL SETUP 1Voltage
Calibration Procedures3
Ω
V LO
V MI
V HI
CAL SETUP 2OVPNone
I LO
CAL SETUP 3Current
CAL SETUP 4OCPNone
You can terminate any CAL SETUP without changing its calibration constants by
turning off power.
Perform the voltage calibration prior to the OVP calibration and the current
calibration prior to the OCP calibration.
Keysight E3633A and E3634A Service Guide79
I MI
I HI
3Calibration Procedures
Front panel voltage and current calibration
1 Unsecure the power supply.
To calibrate the voltage and current, you must unsecure the power supply
according to the “To unsecure the power supply for calibration” on page 74.
2 Disconnect all loads from the power supply and connect a DVM across
output terminals.
3 Turn on calibration mode.
CAL MODE
Turn on the calibration mode by holding down (Calibrate) key as you turn
on the power supply and hold down the key until you hear a long beep. Make
sure that the power supply is in “CV” mode. If the power supply is not in “CV”
mode, an error occurs.
Voltage and OVP calibration
4 Move down a level to the voltage calibration mode.
CAL SETUP 1
The display shows the above message to indicate that the power supply is
ready for voltage calibration.
5 Calibrate DAC and select the low voltage calibration point.
30 LEFT
The “START BITCAL” message is displayed for about 3 seconds to indicate that
the power supply is ready for DAC calibration. Then it counts down numbers
from 30 to 0. Then, the display shows the low voltage calibration point.
V LO 0.5000V
80Keysight E3633A and E3634A Service Guide
Calibration Procedures3
6 Read the DVM and change the low voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V
using the knob and resolution selection keys.
V LO 0.4500 V
7 Pressing (Calibrate) key saves the change and selects the middle
voltage calibration point. (“V MI 25.000 V” - E3634A model)
V MI 10.000 V (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message
appears for one second. If the entered number is not correct, an “MIN VALUE”
or “MAX VALUE” message appears for one second and the display shows the
low voltage calibration point again. The display now shows the middle voltage
calibration point.
8 Read the DVM and change the middle voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reads 10.001 V, adjust the voltage to 10.001 V using
the knob and arrow keys.
V MI 10.001 V
9 Pressing (Calibrate) key saves the changes and selects the high
voltage calibration point. (“V HI 49.500 V” - E3634A model)
V HI 19.500 V (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message
appears for one second. If the entered number is not correct, an “MIN VALUE”
or “MAX VALUE” message appears for one second and the display shows the
middle voltage calibration point again. The display now shows the high
voltage calibration point.
Keysight E3633A and E3634A Service Guide81
3Calibration Procedures
NOTE
10 Read the DVM and change the high voltage value on the d isplay to match
the measured voltage.
For example, if the DVM reads 19.495 V, adjust the voltage to 19.495 V using
the knob and arrow keys.
V HI 19.495 V
11 Pressing (Calibrate) key saves the new voltage calibration constants,
and goes to the OVP calibration mode.
CAL SETUP 2
A “CALIBRATING” message appears for one second to indicate that the
voltage calibration is progressing and new voltage calibration constants of
“SETUP 1” are stored. Then, the display shows above message to indicate that
the power supply is ready for the OVP calibration.
If the calibration fails, a “DAC CAL FAIL” or “ADC CAL FAIL” message appears
for one second and the display shows the “CAL SETUP 1” for voltage
calibration again.
Current and OCP calibration
12 Pressing (Calibrate) key saves the new calibration constants for OVP
circuit and goes to the current calibration mode.
CAL SETUP 3
A “CALIBRATING” message appears for about several seconds to indicate that
the OVP calibration is progressing and new calibration constants of “SETUP 2”
are stored. Then, the display shows the above message to indicate that the
power supply is ready for the current calibration.
If the calibration fails, a “OVP CAL FAIL” message appears for one second and
the display shows the “CAL SETUP 2” for OVP calibration again.
Connect an appropriate shunt 0.01 Ω across the output terminals, and connect a
digital voltmeter across the shunt resistor for the current calibration.
82Keysight E3633A and E3634A Service Guide
Calibration Procedures3
13 Select the low current calibration point. (‘‘I LO 0.2000 A’’ - E3634A model)
I LO 0.5000 A (E3633A)
The display shows the low current calibration point.
14 Read the DVM and change the low current value on the display to match
the computed current (DVM reading ÷ by shunt resistance).
For example, if the computed value is 0.4990 A, adjust the current to 0.4990 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
I LO 0.4990 A
15 Pressing the “Calibrate” key saves the change and selects the middle
current calibration point. (“I MI 3.5000 A’’ - E3634A model)
I MI 10.000 A (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message
appears for one second. If the entered number is not correct, an “MIN VALUE”
or “MAX VALUE” message appears for one second and the display shows the
low current calibration point again. The display now shows the middle current
calibration point.
16 Read the DVM and change the middle current value on the display to match
the computed current (DVM reading ÷ by shunt resistance).
For example, if the computed value is 9.999 A, adjust the current to 9.999 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
I MI 09.999 A
Keysight E3633A and E3634A Service Guide83
3Calibration Procedures
17 Pressing the “Calibrate” key saves the change and selects the high current
calibration point. (“I HI 6.9000 A” - E3634A model)
I HI 19.500 A (E3633A)
If the entered number is within an acceptable range, an “ENTERED” message
appears for one second. If the entered number is not correct, an “MIN VALUE”
or “MAX VALUE” message appears for one second and the display shows the
middle current calibration point again. The display now shows the high current
calibration point.
18 Read the DVM and change the high current value on the d isplay to match
the computed current (DVM reading ¸ by shunt resistance).
For example, if the computed value is 19.499 A, adjust the current to 19.499 A
using the knob and arrow keys.
Notice that you should wait for the DVM reading to stabilize for accurate
calibration.
I HI 19.499 A
19 Pressing the ‘‘Calibrate’’ key saves the new calibration constants for the
output current and goes to the OCP calibration mode.
CAL SETUP 4
A “CALIBRATING” message appears for one second to indicate that the current
calibration is progressing and new calibration constants of “SETUP 3” are
stored. Then, the display shows the above message to indicate that the power
supply is ready for the OCP calibration.
If the calibration fails, an “DAC CAL FAIL” or “ADC CAL FAIL” message appears
for one second and the display shows the “CAL SETUP 3” for current
calibration again.
84Keysight E3633A and E3634A Service Guide
20 Pressing the “Calibrate” key saves the new OCP calibration constants and
return to the calibration mode.
CAL MODE
A “CALIBRATING” message appears for several seconds to indicate that the
OCP calibration is progressing and new OCP calibration constants of “SETUP
4” are stored. Then the display will return to the calibration mode.
If the calibration fails, a “OCP CAL FAIL” message appears for one second and
the display shows the “CAL SETUP 4” for OCP calibration again.
Turn off the power supply to exit the calibration mode.
Aborting a Calibration in Progress
Sometimes it may be necessary to abort a calibration after the procedure has
already been initiated. You can abort a calibration at any time by turning the
power supply off from the front panel. When performing a calibration from the
remote interface, you can abort a calibration by issuing a remote interface device
clear message or by pressing the front-panel “Local” key.
Calibration Procedures3
Keysight E3633A and E3634A Service Guide85
3Calibration Procedures
Calibration Record for Keysight E3633A/E3634A
StepCalibration Description
Measurement
Mode (DVM)
Supply Being
Adjusted
1Unsecure the power supply (see page 62).
2Turn on “CAL MODE” (hold down “Calibrate” and “Power” keys as you turn on the
power supply until you hear a long beep).
3Move down menu to “CAL SETUP 1” (press “Calibrate” key).Voltage
Calibration
4Calibrate DAC and select the low point for voltage calibration; “START BITCAL”
appears for about 3 seconds and the display counts down numbers from 30 to 0.
Then, “V LO 0.5000 V” appears on the display (press “Calibrate” key and wait
VDAC and low
voltage point
calibration
about 30 seconds; then change the display to match the DVM reading).
5
“V MI 10.000 V”
[a]
or “V MI 25.000 V”
(press “Calibrate” key; then change the display to match the DVM reading).
6
“V HI 19.500 V”
[a]
or “V HI 49.500 V”
(press “Calibrate” key; then change the display to match the DVM reading).
[b]
appears on the display
[b]
appears on the display
VMiddle voltage
point calibration
VHigh vol tage point
calibration
7“CAL SETUP 2” now appears on the display (press “Calibrate” key).VOVP calibration
8“CAL SETUP 3” now appears on the display
(press “Calibrate” key; then connect 0.01
9
“I LO 0.5000 A”
[a]
or “I LO 0.2000 A”
Ω
resistor across the output terminals).
[b]
appears on the display
(press “Calibrate” key; then change the display to match the computed current
ALow current point
Current
calibration
calibration
through the resistor).
10
“I MI 10.000 A”
[a]
or “I MI 3.5000 A”
(press “Calibrate” key; then change the display to match the computed current
[b]
appears on the display
AMiddle current
point calibration
through the resistor).
11
“I HI 19.500 A”
[a]
or “I HI 6.9000 A”
(press “Calibrate” key; then change the display to match the computed current
[b]
appears on the display
AHigh current point
calibration
through the resistor).
12 “CAL SETUP 4” now appears on the display (press “Calibrate” key).AOCP calibration
13 Press “Calibrate” key, then press “Power” switch.Exit CAL MODE
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
86Keysight E3633A and E3634A Service Guide
Error Messages
The following tables are abbreviated lists of error messages for the E3633A and
E3634A. The errors listed are the most likely errors to be encountered during
calibration and adjustment. A more complete list of error messages and
descriptions is contained in chapter 5 of the E3633A and E3634A User’s Guide.
System error messages
ErrorError Messages
-330Self-test failed
-350Queue overflow
501Isolator UART framing error
502Isolator UART overrun error
503
511RS-232 framing error
512RS-232 overrun error
513RS-232 parity error
514Command allowed only with RS-232
521Input buffer overflow
522Output buffer overflow
550Command not allowed in local
Calibration Procedures3
[a]
SPI data error
[a] This error message is only applicable for serial MY53xx6xxx.
Keysight E3633A and E3634A Service Guide87
3Calibration Procedures
Self-test error messages
ErrorError Messages
601Front panel does not respond
602RAM read/write failed
603A/D sync stuck
604A/D slope convergence failed
605Cannot calibrate rundown gain
606Rundown gain out of range
607Rundown too noisy
608Serial configuration readback failed
[a]
609
624Unable to sense line frequency
625I/O processor does not respond
626I/O processor failed self-test
630Fan test failed
631System DAC test failed
632Hardware test failed
System ADC test failed
[a] This error message is only applicable for serial MY53xx6xxx.
88Keysight E3633A and E3634A Service Guide
Calibration error messages
ErrorError Messages
701Cal security disabled by jumper
702Cal secured
703Invalid secure code
704Secure code too long
705Cal aborted
708Cal output disabled
712Bad DAC cal data
713Bad readback cal data
714Bad OVP cal data
715Bad OCP cal data
716Bad DAC DNL error correction data
717Cal OVP or OCP status enabled
718Gain out of range for gain error correction
740Cal checksum failed, secure state
741Cal checksum failed, string data
742Cal checksum failed, store/recall data in location 0
743Cal checksum failed, store/recall data in location 1
744Cal checksum failed, store/recall data in location 2
745Cal checksum failed, store/recall data in location 3
746Cal checksum failed, DAC cal constants
747Cal checksum failed, readback cal constants
748Cal checksum failed, GPIB address
749Cal checksum failed, internal data
750Cal checksum failed, DAC DNL error correction data
Calibration Procedures3
Keysight E3633A and E3634A Service Guide89
3Calibration Procedures
An Example Program of Excel 97 for Calibration
This section contains an Excel Macros (Visual Basic® for Applications) program
for calibration over the GPIB interface. This program makes software adjustments
to the E3633A power supply using a current shunt and a digital multimeter which
is connected to the controller. In this program a 0.001 ohm current shunt is used.
Be sure to change the value of the variable “shunt” to the value of the current
shunt used and the GPIB address for the power supply and the digital voltmeter.
'**********************************************************************
' This program was written on a PC with Excel Macros (Visual Basic® for
' Applications) for Windows 95 or Windows NT 4.0. It will make software
' adjustments to the E3633A Power Supply on the GPIB bus using a
' Keysight 34401A Digital Multimeter and a current shunt. In the program
' a 0.001 ohm current shunt is used to measure current. Be sure to
' change the value of the variable ‘shunt’ to the value of the current
' shunt used.
'**********************************************************************
Global id_power As Long
Global id_DMM As Long
Global power As Long
Global DMM As Long
Global Const VI_SUCCESS = &H0&
Global Const VoltageMin = 0
Global Const VoltageMid = 1
Global Const VoltageMax = 2
Global Const CurrentMin = 3
Global Const CurrentMid = 4
Global Const CurrentMax = 5
Sub Calibration_Click()Max
Dim shunt As Single 'Current shunt value in Ohms
Dim UserAnswer 'User response
Range("B4:B6").ClearContents
If OpenPort = False Then
ClosePort
Exit Sub
End If
InitializeDevice
EnableOVPandOCP False
If CheckSecurity = False Then
ClosePort
Exit Sub
End If
Continued on next page
90Keysight E3633A and E3634A Service Guide
Calibration Procedures3
If DacErrorCorrection = False Then
ClosePort
Exit Sub
End If
shunt = 0.001 'Change the current shunt value
UserAnswer = MsgBox(“Voltage & OVP calibration. Connect the output to the
DMM”, vbYesNo + vbQuestion, “E3633A Calibration”)
If UserAnswer = vbNo Then
ClosePort
Exit Sub
End If
Range("B4").Select
ActiveCell.Value = "Begin Voltage Calibration"
StartCalibration VoltageMin, True, shunt 'Set output to minimum voltage cal
ActiveCell.Value = "End Minimum Voltage Calibration"
StartCalibration VoltageMid, True, shunt 'Set output to middle voltage cal
ActiveCell.Value = "End Middle Voltage Calibration"
StartCalibration VoltageMax, True, shunt 'Set output to maximum voltage cal
ActiveCell.Value = "End Maximum Voltage Calibration"
Message = SendSCPI(power, "Syst:Err?")
If InStr(Message, "0") Then
ActiveCell.Value = "Voltage Calibration Complete"
Else
ActiveCell.Value = Message
ClosePort
Exit Sub
End If
OVPandOCPCalibration True
UserAnswer = MsgBox("Connect the Current shunt to the DMM input to
measure a current. Connect the output to the shunt.",
vbYesNo + vbQuestion, "E3633A Calibration")
If UserAnswer = vbNo Then
ClosePort
Exit Sub
End If
ActiveCell.Value = "Begin Current Calibration"
StartCalibration CurrentMin, False, shunt 'Set output to minimum current
cal
ActiveCell.Value = "End Minimum Current Calibration"
StartCalibration CurrentMid, False, shunt 'Set output to middle current cal
ActiveCell.Value = "End Middle Current Calibration"
StartCalibration CurrentMax, False, shunt 'Set output to maximum current
cal
ActiveCell.Value = "End Maximum Current Calibration"
Continued on next page
Keysight E3633A and E3634A Service Guide91
3Calibration Procedures
OVPandOCPCalibration False
Message = SendSCPI(power, "Syst:Err?")
If InStr(Message, "0") Then
Else
End If
EnableOVPandOCP True
SaveDate
ActiveCell.Value = "Calibration Complete"
Message = SendSCPI(power, "*RST")
ClosePort
End Sub
Private Function OpenPort() As Boolean 'Open communications on GPIB
Dim Power_Address As String
Dim DMM_Address As String
Dim Error As Long
Power_Address = "5" 'Select power supply GPIB address between 0 to 30
DMM_Address = "22" 'Select DMM GPIB address between 0 to 30
Error = viOpenDefaultRM(id_power) 'Open the power supply VISA session
If OpenPort = CheckError(Error, "Open ID Error to Power Supply") Then
End If
Error = viOpenDefaultRM(id_DMM) 'Open the DMM VISA session
If OpenPort = CheckError(Error, "Open ID Error to Digital Multi Meter")
Then
End If
Error = viOpen(id_power, "GPIB0::" & Power_Address & "::INSTR", 0, 1000,
power)
If OpenPort = CheckError(Error, "Unable Open to Power Supply") Then
End If
Error = viOpen(id_DMM, "GPIB0::" & DMM_Address & "::INSTR", 0, 1000, DMM)
If OpenPort = CheckError(Error, "Unable Open to Digital Multi Meter") Then
End If
OpenPort = True
End Function
ActiveCell.Value = "Current Calibration Complete"
ActiveCell.Value = Message
ClosePort
Exit Sub
Exit Function
Exit Function
Exit Function
Exit Function
Private Function InitializeDevice()
SendSCPI power, "*Cls"
SendSCPI DMM, "*Rst" 'Set power-on condition for DMM
SendSCPI power, "*Rst" 'Set power-on condition for power supply
End Function
Continued on next page
92Keysight E3633A and E3634A Service Guide
Calibration Procedures3
Private Function CheckError(Error As Long, Message As String) As Boolean
Dim commandString As String
Dim ReturnString As String
Dim crlfpos As Integer
Dim ReadBuffer As String * 512
Dim actual As Long
Dim Error As Long
commandString = command & Chr$(10)
Error = viWrite(device, ByVal commandString, Len(commandString), actual)
If InStr(commandString, "?") Then
Error = viRead(device, ByVal ReadBuffer, 512, actual)
ReturnString = ReadBuffer
crlfpos = InStr(ReturnString, Chr$(0))
If crlfpos Then
ReturnString = Left(ReturnString, crlfpos - 2)
End If
SendSCPI = ReturnString
End If
Dim Finish As Single
Finish = Timer + delay_time
Do
Loop Until Finish <= Timer
Continued on next page
Keysight E3633A and E3634A Service Guide93
3Calibration Procedures
Private Function CheckSecurity() As Boolean
Dim Message As String
Dim SecurityCode As String
Message = SendSCPI(power, "Cal:Str?")
Range("B5").Select
ActiveCell.Value = Message
Message = SendSCPI(power, "Cal:Count?")
Range("B6").Select
ActiveCell.Value = Message
Range("B4").Select
SecurityCode = "HP003633"
SendSCPI power, "Cal:Sec:Stat Off," & SecurityCode
Message = SendSCPI(power, "Cal:Sec:Stat?")
If InStr(Message, "1") Then
Else
Else
End If
End If
End Function
Private Function DacErrorCorrection() As Boolean
Dim Message As String
SendSCPI power, "Output On" 'Turn on the power supply output
SendSCPI power, "Cal:Dac:Error"
For I = 1 To 27
Next I
SendSCPI power, "Output Off" 'Turn off the power supply output
Message = SendSCPI(power, "Syst:Err?")
If InStr(Message, "0") Then
Else
End If
End Function
ActiveCell.Value = "Unable to Unsecure the Power Supply"
CheckSecurity = False
Message = SendSCPI(power, "Syst:Err?")
If InStr(Message, "0") Then
Dim Message As String
'calibration string enclosed in quotes
Message = "Cal:Str " & Chr(34) & Date & Chr(34)
SendSCPI power, Message
Message = SendSCPI(power, "Cal:Str?")
Range("B5").Select
ActiveCell.Value = Message
Message = SendSCPI(power, "Cal:Count?")
Range("B6").Select
ActiveCell.Value = Message
Range("B4").Select
End Function
End of program
96Keysight E3633A and E3634A Service Guide
Calibration Procedures3
Declaration for Windows 95/NT 4.0
Declare Function viOpenDefaultRM Lib "visa32.dll" (instrumentHandle As Long) As
Long
Declare Function viOpen Lib "visa32.dll" (ByVal instrumentHandle As Long, _
ByVal viDesc As String, ByVal mode As Long, ByVal timeout As Long, _
vi As Long) As Long
Declare Function viClose Lib "visa32.dll" (ByVal vi As Long) As Long
Declare Function viWrite Lib "visa32.dll" (ByVal vi As Long, ByVal Buffer As
String, ByVal count As Long, retCount As Long) As Long
Declare Function viRead Lib "visa32.dll" (ByVal vi As Long, ByVal Buffer As
String, _ByVal count As Long, retCount As Long) As Long
Keysight E3633A and E3634A Service Guide97
3Calibration Procedures
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
98Keysight E3633A and E3634A Service Guide
Keysight E3633A and E3634A DC Power Supplies
Service Guide
4Theory of Operation
Block Diagram Overview 100
AC Input and Bias Supplies 102
Floating Logic 103
D-to-A Converter 105
A-to-D Converter 106
Power Mesh and Control 107
Earth-Referenced Logic 108
Front Panel 109
This chapter provides block diagram level descriptions of the power supply. The
descriptions provide a basic understanding of circuit operation and are intended
as an aid in troubleshooting. It is assumed in the following discussions that you
are familiar with the operating and programming instructions presented in the
E3633A and E3634A User’s Guide. Subjects covered include the following:
99
4Theory of Operation
Block Diagram Overview
This discussion pertains to the block diagram on the next page. The power
supply’s circuitry is divided into two major blocks: the floating circuitry and the
earth referenced circuitry. All power mesh and control circuits, display circuit, and
digital circuits are contained in the floating circuitry. This circuitry also contains
the power supply’s main controller. The earth referenced circuitry provides the
interface between the user and the power supply.
The floating circuitry can be viewed in four pieces; the DAC system, the digital
logic section (floating logic), the power mesh and control section, and the front
panel (display and keyboard) section.
The floating logic receives digital signals from the earth-referenced logic and the
DAC converts them to analog signals which are sent to the power control circuits
in order to program the power supply’s output voltage and current. The power
supply can also be commanded to send measurement and status data back to the
remote interface controller and/or the VFD (vacuum fluorescent display) display
on the front panel. The data is processed and sent back via the floating logic and
earth-referenced logic.
The power mesh and control circuits contains voltage and current control circuits
which allows the power supply to operate in either the constant voltage (CV) or
constant current (CC) mode. The control circuits compare the power supply’s
output voltage or current with the programmed value and generates a control
signal which varies the conduction of the series pass transistor to raise or lower
the output as required.
The front panel circuits consist of VFD control, display high voltage drivers, and
keyboard scanning. Communication between the front panel and floating logic
circuits is accomplished through a 4-wire bi-directional serial interface.
The earth referenced circuitry uses a controller configured as a slave to the main
controller. This controller establishes external I/O communication with the main
controller through a bi-directional, optically isolated, serial communications link.
The earth referenced controller controls low-level GPIB (IEEE-488) and RS-232
interface operation.
Separate reference and bias supplies are provided for the floating and ground
reference circuitry. The front panel operates from the floating circuitry with its
logic common different from the main controller logic common.
100Keysight E3633A and E3634A Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.