Keysight Technologies E36200 Series, E36231A, E36232A, E36233A, E36234A Service Manual

Keysight E36200 Series Autoranging DC Power Supplies
Service Guide
Notices
CAUTION
WARNING
Copyright Notice
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
E36200-90009
Edition
Edition 2, September 17, 2019
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Technology Licenses
The hardware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight E36200 Series Service Guide

Safety Symbols

The following symbols on the instrument and in the documentation indicate precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of danger (refer to accompanying text for the specific Warning or Caution information)
Caution, risk of electric shock
Protective earth (ground) terminal
Earth (ground) terminal
Frame or chassis (ground) terminal
Standby supply, the instrument is not completely disconnected from AC mains when the switch is off
Direct current (DC)
Alternating current (AC)
Plus, positive polarity
Minus, negative polarity
Keysight E36200 Series Service Guide 3

Safety Considerations

WARNING
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
– BEFORE APPLYING POWER
Verify that the correct fuse is installed. See “Fuse Information” in the E36200 Series User’s Guide for additional details. Ensure the mains supply voltage fluctuation do not exceed ±10% of the nominal supply voltage.
– GROUND THE INSTRUMENT
This product is a Safety Class I instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The instrument must be connected to the AC power supply mains through a three-conductor power cable, with the third wire firmly connected to an electrical ground (safety ground) at the power outlet. Any interruption of the protective (grounding) conductor or disconnection of the protective earth terminal will cause a potential shock hazard that could result in personal injury. If the instrument is to be energized via an external autotransformer for voltage reduction, be certain that the autotransformer common terminal is connected to the neutral (earthed pole) of the AC power lines (supply mains).
– DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE OR WET
ENVIRONMENTS
Do not operate the device around flammable gases or fumes, vapor, or wet environments.
4 Keysight E36200 Series Service Guide
WARNING
– DO NOT OPERATE DAMAGED OR DEFECTIVE INSTRUMENTS
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel.
– DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained. To contact Keysight for sales and technical support, refer to the support links on the following Keysight website: www.keysight.com/
find/assist (worldwide contact information for repair and service).
– USE THE POWER CORD PROVIDED
Use the device with the power cord provided with the shipment.
– USE THE DEVICE AS SPECIFIED
If the device is used in a manner not specified by manufacturer, the device protection may be impaired.
– DO NOT BLOCK VENTILATION HOLES
Keysight E36200 Series Service Guide 5
Do not block any of the ventilation holes of the device.
– OBSERVE ALL DEVICE MARKINGS BEFORE CONNECTING TO DEVICE
Observe all markings on the device before connecting any wiring to the device.
– TURN DEVICE OFF BEFORE CONNECTING TO OUTPUT TERMINALS
Turn off the device power before connecting to the output terminals.
– ENSURE COVER IS SECURED IN PLACE
Do not operate the device with the cover removed or loosened.
WARNING
– TURN DEVICE OFF AND REMOVE ALL CONNECTIONS BEFORE
CAUTION
NOTE
INSTALLING THE GPIB INTERFACE
Turn off the power and remove all connections, including the power cord, from the instrument prior installation of the GPIB interface.
– ENSURE PROPER AWG CABLE IS USED
Use a cable with the correct voltage and AWG rating based on the intended setup when operating the E36200A Series autoranging DC power supplies.
– DO NOT TOUCH CABLES DURING OPERATION
Do not touch the cable while the instrument output is operational to prevent electric shock hazard and burn hazard.
CLEAN WITH SLIGHTLY DAMPENED CLOTH
Clean the outside of the instrument with a soft, lint-free, slightly dampened cloth. Do not use detergent, volatile liquids, or chemical solvents.
CONNECT USB CABLE WITH FERRITE CORE
Connect a USB cable with a ferrite core to the rear panel USB port of the instrument for better performance.
6 Keysight E36200 Series Service Guide

Environmental Conditions

The E36200 Series is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument.
Environmental condition Requirement
Temperature
Humidity
Altitude Up to 2000 m
Pollution degree 2

Regulatory Information

Operating condition – 0 °C to 40 °C Storage condition – –20 °C to 70 °C
Operating condition – Up to 80% RH at 40 °C (non-condensing) Storage condition – Up to 90% RH at 65 °C (non-condensing)
The E36200 Series complies with the following safety and Electromagnetic Compatibility (EMC) compliances:
– Low Voltage Directive 2014/35/EU
– EMC Directive 2014/30/EU
Keysight E36200 Series Service Guide 7

Regulatory Markings

The RCM mark is a registered trademark of the Australian Communications and Media Authority.
The CE mark is a registered trademark of the European Community. This CE mark shows that the product complies with all the relevant European Legal Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this is an Industrial Scientific and Medical Group 1 Class A product.
This symbol is a South Korean Class A EMC Declaration. This is a Class A instrument suitable for professional use and in electromagnetic environment outside of the home.
The CSA mark is a registered trademark of the Canadian Standards Association.
This symbol indicates the time period during which no hazardous or toxic substance elements are expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
8 Keysight E36200 Series Service Guide

Waste Electrical and Electronic Equipment (WEEE) Directive

This instrument complies with the WEEE Directive marking requirement. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.

Product category:

With reference to the equipment types in the WEEE directive Annex 1, this instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.

Sales and Technical Support

To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites:
www.keysight.com/find/e36200
(product-specific information and support, software and documentation updates)
www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight E36200 Series Service Guide 9
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
10 Keysight E36200 Series Service Guide

Table of Contents

Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Regulatory Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Regulatory Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Waste Electrical and Electronic Equipment (WEEE) Directive . . . . . . . .9
Product category: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Sales and Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
1 Service and Maintenance
Specifications and Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Types of service available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Obtaining repair service (worldwide) . . . . . . . . . . . . . . . . . . . . . . . . .14
Repackaging for shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
Cleaning and handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Self-Test Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Replacing the Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
User Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Replacing the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Removing/Installing the optional GPIB interface . . . . . . . . . . . . . . .20
Replacing the front panel binding post . . . . . . . . . . . . . . . . . . . . . . .21
2 Verification and Adjustments
Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Recommended test equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
Keysight E36200 Series Service Guide 11
Test considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Measurement techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Constant Voltage (CV) verification . . . . . . . . . . . . . . . . . . . . . . . . . . 58
CV load and line regulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Low range current verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Constant Current (CC) verification . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Test Record Forms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Calibration Adjustment Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Closed-case electronic calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Calibration interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Calibration adjustment process . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Calibration security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
Calibration count . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Calibration message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Saving calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Calibration auto save . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
Save the calibration data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
12 Keysight E36200 Series Service Guide
Keysight E36200 Series Autoranging DC Power Supplies Service Guide

1 Service and Maintenance

Specifications and Characteristics 14 General Information 14 Troubleshooting 16 Self-Test Procedures 16 Replacing the Fuse 16 User Replaceable Parts 17 Battery Replacement 17 Disassembly 20
This chapter provides the specifications and service information on cleaning, troubleshooting, repair, and replaceable parts of the E36200 Series autoranging DC power supplies. This chapter also explains how to assemble and disassemble the E36200 Series autoranging DC power supplies.
13
1 Service and Maintenance
NOTE

Specifications and Characteristics

For the specifications and characteristics of the E36200 Series autoranging DC power supplies, refer to its datasheet at http://literature.cdn.keysight.com/
litweb/pd f/5992-3747EN.pdf.

General Information

Types of service available

If your instrument fails during the warranty period, Keysight Technologies will repair or replace it under the terms of your warranty. After your warranty expires, Keysight offers repair services at competitive prices. You also have the option to purchase a service contract that extends the coverage after the standard warranty expires.

Obtaining repair service (worldwide)

To obtain service for your instrument, contact your nearest Keysight Technologies Service Center. They will arrange to have your unit repaired or replaced, and can provide warranty or repair–cost information where applicable. Ask the Keysight Technologies Service Center for shipping instructions, including what components to ship. Keysight recommends that you retain the original shipping carton for return shipments.

Repackaging for shipment

Ensure the following to ship the unit to Keysight for service or repair:
– Attach a tag to the unit identifying the owner and indicating the required
service or repair. Include the model number and full serial number.
– Place the unit in its original container with appropriate packaging material.
– Secure the container with strong tape or metal bands.
14 Keysight E36200 Series Service Guide
– If the original shipping container is unavailable, use a container that will ensure
at least 10 cm (4 in.) of compressible packaging material around the entire instrument. Use static-free packaging materials.
Keysight suggests that you always insure your shipments.

Cleaning and handling

Cleaning
To prevent electrical shock, disconnect the instrument from AC mains power and disconnect all test leads before cleaning. Clean the outside of the instrument using a soft, lint-free, cloth slightly dampened with water.
– Do not use detergent or solvents.
– Do not attempt to clean internally.
If required, contact a Keysight Technologies Sales and Service office to arrange for proper cleaning to ensure that safety features and performance are maintained.
Electrostatic Discharge (ESD) precautions
Service and Maintenance 1
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 V.
The following guidelines will help prevent ESD damage during service operations:
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastic, foam, vinyl, paper, and other static-generating materials
from the immediate work area.
Keysight E36200 Series Service Guide 15
1 Service and Maintenance

Troubleshooting

Before troubleshooting or repairing the instrument, make sure the failure is in the instrument rather than any external connections. Also make sure that the instrument was accurately calibrated within the last year (see “Calibration
Adjustment Procedures” on page 82 and “Calibration interval” on page 82 for
details).
Perform the following verifications if the unit is inoperative:
– Verify that the ac power cord is connected to the power supply.
– Verify that the front-panel power switch is depressed.

Self-Test Procedures

A power-on self-test occurs automatically when you turn on the power supply. This limited test assures you that the power supply is operational.
Refer to the E36200 Series User’s Guide for more information on how to run a complete self-test of the power supply (Chapter 2, “General Operating Information” > Utilities Menu - Test/Setup > Self-Test).

Replacing the Fuse

The fuse is located within the power supply's fuse-holder assembly on the rear panel. Refer to the E36200 Series User’s Guide for more information on how to replace a faulty or blown fuse (Chapter 1, “Product Information” > Fuse Information).
16 Keysight E36200 Series Service Guide

User Replaceable Parts

WARNING
You can find the instrument support part list from the Keysight's Test & Measurement Parts Catalog at http://www.keysight.com/find/parts.

Battery Replacement

SHOCKHAZARD
Only qualified, service-trained personnel who are aware of the hazards involved should remove instrument covers. Always d isconnect the power cable and any external circuits before removing the instrument cover.Some circuits are active and have power for a short time even when the power switch is turned off.
The internal battery powers the real-time clock. The primary function of the clock it to provide a time stamp for the internal file system. If the battery fails, the clock and time stamp function will not be available. No other instrument functions are affected.
Under normal use at room temperature, the lithium battery has a life expectancy between seven and ten years. Note that battery life will be reduced if the instrument is stored for a prolonged period at temperatures above 40 degrees Celsius.
Service and Maintenance 1
Keysight E36200 Series Service Guide 17
The battery type is CR 2032 (3V).
1 Service and Maintenance
NOTE

Replacing the battery

You will need a flat-bladed screwdriverto perform this procedure.
1 Remove the instrument top cover.
2 The battery is located at the front PC board.
3 Use a flat-bladed screwdriver and carefully pry up on the side of the battery.
4 Install the new battery.
Make sure that the positive side (+) is facing up. Place the battery under the
small spring clips closest to the ribbon cable connector, then push down on
the opposite end of the battery to seat the battery (see red arrow below). The
top of the small spring clips should be visible after the battery is seated (see
red circle below).
5 Replace the top cover when finished.
18 Keysight E36200 Series Service Guide
Service and Maintenance 1
NOTE
6 Reset the date and time.
Refer to the E36200 Series User’s Guide for more information on how to reset the date and time. (Chapter 2, “General Operating Information” > Utilities
Menu - Test/Setup > User Settings > Date/Time)
Properly dispose of the old battery in accordance with local laws and regulations.
Keysight E36200 Series Service Guide 19
1 Service and Maintenance
WARNING
WARNING

Disassembly

Tools required

Items Torque value
PZ1 driver 9.0 in.lbs
T10 Torx driver 9.0 in.lbs
T8 Torx driver 9.0 in.lbs
1/4" nut socket 9.0 in.lbs
1/2" nut socket 3.5 in.lbs

Removing/Installing the optional GPIB interface

TURN OFF POWER AND REMOVE ALL CONNECTIONS BEFOREPROCEED
Turn off the power and remove all connections, including the power cord, from the instrument prior installation/removal of the GPIB interface.
RETAIN GPIB COVER PLATE
After installing the GPIB interface, retain the cover plate for use in the event that you ever remove the GPIB interface. The instrument should never be connected to power or inputs on the output terminals without either the GPIB interface or the cover plate securely covering the rear-panel opening.
Refer to the E36200 Series User’s Guide for more information on how to remove or install the optional GPIB interface (Chapter 1, “Product Information” > Installation > Installing the optional GPIB interface).
20 Keysight E36200 Series Service Guide

Replacing the front panel binding post

WARNING
NOTE
Only qualified, service-trained personnel should remove and replace the front panel binding post.
The GPIB interface must be removed before you proceed to remove and replace the front panel binding post.
Disassembly procedures are model-specific. Refer to the procedures listed below for your respective E36200 Series model.
“For E36231A and E36232A models only” on page 22
“For E36233A models only” on page 32
“For E36234A models only” on page 42
Service and Maintenance 1
Keysight E36200 Series Service Guide 21
1 Service and Maintenance
For E36231A and E36232A models only
Step Instructions Visuals
1 Remove the side trims from the front panel.
2 Remove four screws from the front panel using a
T10 driver (two screws on each side).
22 Keysight E36200 Series Service Guide
Step Instructions Visuals
3 Remove the top panel. Slide the top panel
towards the rear and lift it up.
4 Remove the IO board thermal pad.
Service and Maintenance 1
5 Remove four Pozi screws from the front panel
using a PZ1 driver (two screws on each side).
Keysight E36200 Series Service Guide 23
1 Service and Maintenance
Step Instructions Visuals
6 Gently pull out the front frame from the chassis.
7 Detach the mainboard connector from the IO
board.
24 Keysight E36200 Series Service Guide
Step Instructions Visuals
8 Detach the power PSU connector from the IO
board.
9 Remove four screws from the IO board using a
T10 driver. Remove the IO board from the chassis.
Service and Maintenance 1
Keysight E36200 Series Service Guide 25
1 Service and Maintenance
Step Instructions Visuals
NOTE: Perform step 10 and step 12 only if you have installed the optional GPIB interface (see “Removing/Installing the optional
GPIB interface” on page 20 for more details).
10 Remove one screw from the GPIB module plate
using a T10 driver.
11 Detach the GPIB module connector from the IO
board.
12 Remove the GPIB module from the chassis.
26 Keysight E36200 Series Service Guide
Step Instructions Visuals
13 Detach the AC inlet connector from the PSU
module.
14 Detach the power PSU connector from the
mainboard.
Service and Maintenance 1
15 Remove one screw from the rear chassis using a
T10 bit.
Keysight E36200 Series Service Guide 27