duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Manual Part Number
E36200-90009
Edition
Edition 2, September 17, 2019
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, perform, display, or disclose commercial
computer software or commercial computer software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight E36200 Series Service Guide
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of danger (refer to accompanying text for the specific Warning or
Caution information)
Caution, risk of electric shock
Protective earth (ground) terminal
Earth (ground) terminal
Frame or chassis (ground) terminal
Standby supply, the instrument is not completely disconnected from AC mains
when the switch is off
Direct current (DC)
Alternating current (AC)
Plus, positive polarity
Minus, negative polarity
Keysight E36200 Series Service Guide3
Safety Considerations
WARNING
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
– BEFORE APPLYING POWER
Verify that the correct fuse is installed. See “Fuse Information” in the
E36200 Series User’s Guide for additional details. Ensure the mains
supply voltage fluctuation do not exceed ±10% of the nominal supply
voltage.
– GROUND THE INSTRUMENT
This product is a Safety Class I instrument (provided with a protective
earth terminal). To minimize shock hazard, the instrument chassis and
cabinet must be connected to an electrical ground. The instrument must
be connected to the AC power supply mains through a three-conductor
power cable, with the third wire firmly connected to an electrical ground
(safety ground) at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal
will cause a potential shock hazard that could result in personal injury. If
the instrument is to be energized via an external autotransformer for
voltage reduction, be certain that the autotransformer common terminal
is connected to the neutral (earthed pole) of the AC power lines (supply
mains).
– DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE OR WET
ENVIRONMENTS
Do not operate the device around flammable gases or fumes, vapor, or
wet environments.
4Keysight E36200 Series Service Guide
WARNING
– DO NOT OPERATE DAMAGED OR DEFECTIVE INSTRUMENTS
Instruments that appear damaged or defective should be made
inoperative and secured against unintended operation until they can be
repaired by qualified service personnel.
– DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not install
substitute parts or perform any unauthorized modification to the
instrument. Return the instrument to a Keysight Technologies Sales and
Service Office for service and repair to ensure that safety features are
maintained. To contact Keysight for sales and technical support, refer to
the support links on the following Keysight website: www.keysight.com/
find/assist (worldwide contact information for repair and service).
– USE THE POWER CORD PROVIDED
Use the device with the power cord provided with the shipment.
– USE THE DEVICE AS SPECIFIED
If the device is used in a manner not specified by manufacturer, the
device protection may be impaired.
– DO NOT BLOCK VENTILATION HOLES
Keysight E36200 Series Service Guide5
Do not block any of the ventilation holes of the device.
– OBSERVE ALL DEVICE MARKINGS BEFORE CONNECTING TO DEVICE
Observe all markings on the device before connecting any wiring to the
device.
– TURN DEVICE OFF BEFORE CONNECTING TO OUTPUT TERMINALS
Turn off the device power before connecting to the output terminals.
– ENSURE COVER IS SECURED IN PLACE
Do not operate the device with the cover removed or loosened.
WARNING
– TURN DEVICE OFF AND REMOVE ALL CONNECTIONS BEFORE
CAUTION
NOTE
INSTALLING THE GPIB INTERFACE
Turn off the power and remove all connections, including the power cord,
from the instrument prior installation of the GPIB interface.
– ENSURE PROPER AWG CABLE IS USED
Use a cable with the correct voltage and AWG rating based on the
intended setup when operating the E36200A Series autoranging DC
power supplies.
– DO NOT TOUCH CABLES DURING OPERATION
Do not touch the cable while the instrument output is operational to
prevent electric shock hazard and burn hazard.
CLEAN WITH SLIGHTLY DAMPENED CLOTH
Clean the outside of the instrument with a soft, lint-free, slightly dampened
cloth. Do not use detergent, volatile liquids, or chemical solvents.
CONNECT USB CABLE WITH FERRITE CORE
Connect a USB cable with a ferrite core to the rear panel USB port of the
instrument for better performance.
6Keysight E36200 Series Service Guide
Environmental Conditions
The E36200 Series is designed for indoor use and in an area with low
condensation. The table below shows the general environmental requirements for
this instrument.
Environmental conditionRequirement
Temperature
Humidity
AltitudeUp to 2000 m
Pollution degree2
Regulatory Information
Operating condition
– 0 °C to 40 °C
Storage condition
– –20 °C to 70 °C
Operating condition
– Up to 80% RH at 40 °C (non-condensing)
Storage condition
– Up to 90% RH at 65 °C (non-condensing)
The E36200 Series complies with the following safety and Electromagnetic
Compatibility (EMC) compliances:
– Low Voltage Directive 2014/35/EU
– EMC Directive 2014/30/EU
Keysight E36200 Series Service Guide7
Regulatory Markings
The RCM mark is a registered trademark of the Australian Communications and
Media Authority.
The CE mark is a registered trademark of the European Community. This CE
mark shows that the product complies with all the relevant European Legal
Directives.
ICES/NMB-001 indicates that this ISM device complies with the Canadian
ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada.
ISM GRP.1 Class A indicates that this is an Industrial Scientific and Medical
Group 1 Class A product.
This symbol is a South Korean Class A EMC Declaration. This is a Class A
instrument suitable for professional use and in electromagnetic environment
outside of the home.
The CSA mark is a registered trademark of the Canadian
Standards Association.
This symbol indicates the time period during which no hazardous or toxic
substance elements are expected to leak or deteriorate during normal use. Forty
years is the expected useful life of the product.
This instrument complies with the WEEE Directive marking requirement. This
affixed product label indicates that you must not discard this electrical or
electronic product in domestic household waste.
8Keysight E36200 Series Service Guide
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive marking requirement. This
affixed product label indicates that you must not discard this electrical or
electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
– www.keysight.com/find/e36200
(product-specific information and support, software and
documentation updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
Keysight E36200 Series Autoranging DC Power Supplies
Service Guide
1Service and Maintenance
Specifications and Characteristics 14
General Information 14
Troubleshooting 16
Self-Test Procedures 16
Replacing the Fuse 16
User Replaceable Parts 17
Battery Replacement 17
Disassembly 20
This chapter provides the specifications and service information on cleaning,
troubleshooting, repair, and replaceable parts of the E36200 Series autoranging
DC power supplies. This chapter also explains how to assemble and disassemble
the E36200 Series autoranging DC power supplies.
13
1Service and Maintenance
NOTE
Specifications and Characteristics
For the specifications and characteristics of the E36200 Series autoranging DC
power supplies, refer to its datasheet at http://literature.cdn.keysight.com/
litweb/pd f/5992-3747EN.pdf.
General Information
Types of service available
If your instrument fails during the warranty period, Keysight Technologies will
repair or replace it under the terms of your warranty. After your warranty expires,
Keysight offers repair services at competitive prices. You also have the option to
purchase a service contract that extends the coverage after the standard warranty
expires.
Obtaining repair service (worldwide)
To obtain service for your instrument, contact your nearest Keysight Technologies
Service Center. They will arrange to have your unit repaired or replaced, and can
provide warranty or repair–cost information where applicable. Ask the Keysight
Technologies Service Center for shipping instructions, including what
components to ship. Keysight recommends that you retain the original shipping
carton for return shipments.
Repackaging for shipment
Ensure the following to ship the unit to Keysight for service or repair:
– Attach a tag to the unit identifying the owner and indicating the required
service or repair. Include the model number and full serial number.
– Place the unit in its original container with appropriate packaging material.
– Secure the container with strong tape or metal bands.
14Keysight E36200 Series Service Guide
– If the original shipping container is unavailable, use a container that will ensure
at least 10 cm (4 in.) of compressible packaging material around the entire
instrument. Use static-free packaging materials.
Keysight suggests that you always insure your shipments.
Cleaning and handling
Cleaning
To prevent electrical shock, disconnect the instrument from AC mains power and
disconnect all test leads before cleaning. Clean the outside of the instrument
using a soft, lint-free, cloth slightly dampened with water.
– Do not use detergent or solvents.
– Do not attempt to clean internally.
If required, contact a Keysight Technologies Sales and Service office to arrange
for proper cleaning to ensure that safety features and performance are
maintained.
Electrostatic Discharge (ESD) precautions
Service and Maintenance1
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic discharge
voltages as low as 50 V.
The following guidelines will help prevent ESD damage during service operations:
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastic, foam, vinyl, paper, and other static-generating materials
from the immediate work area.
Keysight E36200 Series Service Guide15
1Service and Maintenance
Troubleshooting
Before troubleshooting or repairing the instrument, make sure the failure is in the
instrument rather than any external connections. Also make sure that the
instrument was accurately calibrated within the last year (see “Calibration
Adjustment Procedures” on page 82 and “Calibration interval” on page 82 for
details).
Perform the following verifications if the unit is inoperative:
– Verify that the ac power cord is connected to the power supply.
– Verify that the front-panel power switch is depressed.
Self-Test Procedures
A power-on self-test occurs automatically when you turn on the power supply.
This limited test assures you that the power supply is operational.
Refer to the E36200 Series User’s Guide for more information on how to run a
complete self-test of the power supply (Chapter 2, “General Operating Information” > Utilities Menu - Test/Setup > Self-Test).
Replacing the Fuse
The fuse is located within the power supply's fuse-holder assembly on the rear
panel. Refer to the E36200 Series User’s Guide for more information on how to
replace a faulty or blown fuse (Chapter 1, “Product Information” > Fuse Information).
16Keysight E36200 Series Service Guide
User Replaceable Parts
WARNING
You can find the instrument support part list from the Keysight's Test &
Measurement Parts Catalog at http://www.keysight.com/find/parts.
Battery Replacement
SHOCKHAZARD
Only qualified, service-trained personnel who are aware of the hazards
involved should remove instrument covers. Always d isconnect the power
cable and any external circuits before removing the instrument cover.Some
circuits are active and have power for a short time even when the power
switch is turned off.
The internal battery powers the real-time clock. The primary function of the clock
it to provide a time stamp for the internal file system. If the battery fails, the clock
and time stamp function will not be available. No other instrument functions are
affected.
Under normal use at room temperature, the lithium battery has a life expectancy
between seven and ten years. Note that battery life will be reduced if the
instrument is stored for a prolonged period at temperatures above 40 degrees
Celsius.
Service and Maintenance1
Keysight E36200 Series Service Guide17
The battery type is CR 2032 (3V).
1Service and Maintenance
NOTE
Replacing the battery
You will need a flat-bladed screwdriverto perform this procedure.
1 Remove the instrument top cover.
2 The battery is located at the front PC board.
3 Use a flat-bladed screwdriver and carefully pry up on the side of the battery.
4 Install the new battery.
Make sure that the positive side (+) is facing up. Place the battery under the
small spring clips closest to the ribbon cable connector, then push down on
the opposite end of the battery to seat the battery (see red arrow below). The
top of the small spring clips should be visible after the battery is seated (see
red circle below).
5 Replace the top cover when finished.
18Keysight E36200 Series Service Guide
Service and Maintenance1
NOTE
6 Reset the date and time.
Refer to the E36200 Series User’s Guide for more information on how to reset
the date and time. (Chapter 2, “General Operating Information” > Utilities
Menu - Test/Setup > User Settings > Date/Time)
Properly dispose of the old battery in accordance with local laws and
regulations.
Keysight E36200 Series Service Guide19
1Service and Maintenance
WARNING
WARNING
Disassembly
Tools required
ItemsTorque value
PZ1 driver9.0 in.lbs
T10 Torx driver9.0 in.lbs
T8 Torx driver9.0 in.lbs
1/4" nut socket9.0 in.lbs
1/2" nut socket3.5 in.lbs
Removing/Installing the optional GPIB interface
TURN OFF POWER AND REMOVE ALL CONNECTIONS BEFOREPROCEED
Turn off the power and remove all connections, including the power cord,
from the instrument prior installation/removal of the GPIB interface.
RETAIN GPIB COVER PLATE
After installing the GPIB interface, retain the cover plate for use in the event
that you ever remove the GPIB interface. The instrument should never be
connected to power or inputs on the output terminals without either the
GPIB interface or the cover plate securely covering the rear-panel opening.
Refer to the E36200 Series User’s Guide for more information on how to remove or
install the optional GPIB interface (Chapter 1, “Product Information” >
Installation > Installing the optional GPIB interface).
20Keysight E36200 Series Service Guide
Replacing the front panel binding post
WARNING
NOTE
Only qualified, service-trained personnel should remove and replace the
front panel binding post.
The GPIB interface must be removed before you proceed to remove and replace
the front panel binding post.
Disassembly procedures are model-specific. Refer to the procedures listed below
for your respective E36200 Series model.
– “For E36231A and E36232A models only” on page 22
– “For E36233A models only” on page 32
– “For E36234A models only” on page 42
Service and Maintenance1
Keysight E36200 Series Service Guide21
1Service and Maintenance
For E36231A and E36232A models only
StepInstructionsVisuals
1Remove the side trims from the front panel.
2Remove four screws from the front panel using a
T10 driver (two screws on each side).
22Keysight E36200 Series Service Guide
StepInstructionsVisuals
3Remove the top panel. Slide the top panel
towards the rear and lift it up.
4Remove the IO board thermal pad.
Service and Maintenance1
5Remove four Pozi screws from the front panel
using a PZ1 driver (two screws on each side).
Keysight E36200 Series Service Guide23
1Service and Maintenance
StepInstructionsVisuals
6Gently pull out the front frame from the chassis.
7Detach the mainboard connector from the IO
board.
24Keysight E36200 Series Service Guide
StepInstructionsVisuals
8Detach the power PSU connector from the IO
board.
9Remove four screws from the IO board using a
T10 driver. Remove the IO board from the
chassis.
Service and Maintenance1
Keysight E36200 Series Service Guide25
1Service and Maintenance
StepInstructionsVisuals
NOTE: Perform step 10 and step 12 only if you have installed the optional GPIB interface (see “Removing/Installing the optional
GPIB interface” on page 20 for more details).
10Remove one screw from the GPIB module plate
using a T10 driver.
11Detach the GPIB module connector from the IO
board.
12Remove the GPIB module from the chassis.
26Keysight E36200 Series Service Guide
StepInstructionsVisuals
13Detach the AC inlet connector from the PSU
module.
14Detach the power PSU connector from the
mainboard.
Service and Maintenance1
15Remove one screw from the rear chassis using a
T10 bit.
Keysight E36200 Series Service Guide27
1Service and Maintenance
StepInstructionsVisuals
16Remove two screws from the bracket using a T10
bit. Remove the bracket from the chassis.
17Remove seven screws from the PCA mainboard
using a T10 bit.
28Keysight E36200 Series Service Guide
StepInstructionsVisuals
18Lift and remove the PCA mainboard from the
chassis.
Take caution not to bend any of the connector pins on the
IO board.
Service and Maintenance1
Ensure that the two thermal pad are present on the PCA
mainboard before proceeding with the next step.
Keysight E36200 Series Service Guide29
1Service and Maintenance
StepInstructionsVisuals
19Remove two screws from the IO board using a
T10 bit.
20Remove three screws from the IO board using a
T10 bit. Remove the IO board from the chassis.
30Keysight E36200 Series Service Guide
StepInstructionsVisuals
21Unsolder the binding post assembly from the IO
board. Replace the existing binding post
assembly with a new binding post assembly.
22Perform the steps above in reverse order to reassamble back the instrument.
Service and Maintenance1
Keysight E36200 Series Service Guide31
1Service and Maintenance
For E36233A models only
StepInstructionsVisuals
1Remove the side trims from the front panel.
2Remove four screws from the front panel using a
T10 driver (two screws on each side).
3Remove the top panel. Slide the top panel
towards the rear and lift it up.
32Keysight E36200 Series Service Guide
StepInstructionsVisuals
4Remove the IO board thermal pad.
5Remove four Pozi screws from the front panel
using a PZ1 driver (two screws on each side).
Service and Maintenance1
6Detach the mainboard connectors from the IO
board.
Keysight E36200 Series Service Guide33
1Service and Maintenance
StepInstructionsVisuals
7Gently pull out the front frame from the chassis.
8Detach the power PSU connector from the IO
board.
34Keysight E36200 Series Service Guide
Service and Maintenance1
StepInstructionsVisuals
9Remove four screws from the IO board using a
T10 driver. Remove the IO board from the
chassis.
NOTE: Perform step 10 and step 12 only if you have installed the optional GPIB interface (see “Removing/Installing the optional
GPIB interface” on page 20 for more details).
10Remove one screw from the GPIB module plate
using a T10 driver.
Keysight E36200 Series Service Guide35
1Service and Maintenance
StepInstructionsVisuals
11Detach the GPIB module connector from the IO
board.
12Remove the GPIB module from the chassis.
36Keysight E36200 Series Service Guide
StepInstructionsVisuals
13Remove two screws from the insulator using a
T10 bit. Remove the insulator.
14Detach the AC inlet connector from the PSU
module.
Service and Maintenance1
15Detach the power PSU connectors from the
mainboard.
Keysight E36200 Series Service Guide37
1Service and Maintenance
StepInstructionsVisuals
16Remove one screw from the rear chassis using a
T10 bit.
17Remove two screws from the bracket using a T10
bit. Remove the bracket from the chassis.
38Keysight E36200 Series Service Guide
StepInstructionsVisuals
1
2
18Remove 14 screws from the PCA mainboard
using a T10 bit.
19Lift and remove the the two PCA mainboards
from the chassis.
Service and Maintenance1
Keysight E36200 Series Service Guide39
1Service and Maintenance
StepInstructionsVisuals
Take caution not to bend any of the connector pins on the
IO board.
Ensure that the two thermal pad are present on each of
the PCA mainboards before proceeding with the next
step.
20Remove two screws from the IO board using a
T10 bit.
40Keysight E36200 Series Service Guide
StepInstructionsVisuals
21Remove three screws from the IO board using a
T10 bit. Remove the IO board from the chassis.
22Unsolder the binding post assembly from the IO
board. Remove the four screws with a T-10 bit.
Replace the existing binding post assembly with
a new binding post assembly.
Service and Maintenance1
23Perform the steps above in reverse order to reassamble back the instrument.
Keysight E36200 Series Service Guide41
1Service and Maintenance
For E36234A models only
StepInstructionsVisuals
1Remove the side trims from the front panel.
2Remove four screws from the front panel using a
T10 driver (two screws on each side).
3Remove the top panel. Slide the top panel
towards the rear and lift it up.
42Keysight E36200 Series Service Guide
StepInstructionsVisuals
4Remove the IO board thermal pad.
5Remove four Pozi screws from the front panel
using a PZ1 driver (two screws on each side).
Service and Maintenance1
6Detach the mainboard connectors from the IO
board.
Keysight E36200 Series Service Guide43
1Service and Maintenance
StepInstructionsVisuals
7Gently pull out the front frame from the chassis.
8Detach the power PSU connector from the IO
board.
44Keysight E36200 Series Service Guide
Service and Maintenance1
StepInstructionsVisuals
9Remove four screws from the IO board using a
T10 driver. Remove the IO board from the
chassis.
NOTE: Perform step 10 and step 12 only if you have installed the optional GPIB interface (see “Removing/Installing the optional
GPIB interface” on page 20 for more details).
10Remove one screw from the GPIB module plate
using a T10 driver.
Keysight E36200 Series Service Guide45
1Service and Maintenance
StepInstructionsVisuals
11Detach the GPIB module connector from the IO
board.
12Remove the GPIB module from the chassis.
46Keysight E36200 Series Service Guide
StepInstructionsVisuals
13Remove two screws from the insulator using a
T10 bit. Remove the insulator.
14Detach the AC inlet connector from the PSU
module.
Service and Maintenance1
15Detach the power PSU connectors from the
mainboard.
Keysight E36200 Series Service Guide47
1Service and Maintenance
StepInstructionsVisuals
16Remove one screw from the rear chassis using a
T10 bit.
17Remove two screws from the bracket using a T10
bit. Remove the bracket from the chassis.
48Keysight E36200 Series Service Guide
StepInstructionsVisuals
1
2
18Remove 14 screws from the PCA mainboard
using a T10 bit.
19Lift and remove the the two PCA mainboards
from the chassis.
Service and Maintenance1
Keysight E36200 Series Service Guide49
1Service and Maintenance
StepInstructionsVisuals
Take caution not to bend any of the connector pins on the
IO board.
Ensure that the two thermal pad are present on each of
the PCA mainboards before proceeding with the next
step.
20Remove two screws from the IO board using a
T10 bit.
50Keysight E36200 Series Service Guide
StepInstructionsVisuals
21Remove three screws from the IO board using a
T10 bit. Remove the IO board from the chassis.
Service and Maintenance1
22Unsolder the binding post assembly from the IO
board. Replace the existing binding post
assembly with a new binding post assembly.
23Perform the steps above in reverse order to reassamble back the instrument.
Keysight E36200 Series Service Guide51
1Service and Maintenance
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
52Keysight E36200 Series Service Guide
Keysight E36200 Series Autoranging DC Power Supplies
Service Guide
2Verification and
Adjustments
Performance Verification 54
Test Record Forms 74
Calibration Adjustment Procedures 82
This chapter contains the performance verification procedures, which verifies that
the E36200 Series is operating within its published specifications. This chapter
also provides information on the adjustments performed after the instrument fails
any of the performance verification tests.
53
2Verification and Adjustments
Performance Verification
Performance verification ensures that the instrument performs within the
specifications stated in its data sheet (http://literature.cdn.keysight.com/litweb/
pdf/5992-3747EN.pd f.)
Recommended test equipment
The test equipments recommended for the performance verification and
adjustment procedures are listed below. If the exact instrument is not available,
use the accuracy requirements shown to select substitute calibration standards.
Type Specification Recommended model
Digital multimeter
Current monitor 50 A (0.05 Ω), TC = 4 ppm/°C Guildline 9230A-50
Electronic load
Fixed load
Oscilloscope
RMS voltmeter
Differential amplifier Bandwidth: 20 MHz LeCroy 1855A or equivalent
Readout: 6 ½ digits
Basic DC accuracy: 0.0035%
150 V, 20 A minimum, with transient capability
and a slew rate of 833 kA/s or better
Sensitivity: 1 mV
Bandwid th limit: 20 MHz
Probe: 1:1 with RF tip
True RMS
Bandwid th: 10 MHz
Sensitivity: 100 μV
Keysight 34465A or equivalent
Keysight N3300A mainframe, with
N330xA modules
DSR-1000W40RJ-F-V1
DSR-1000W10RJ-F-V1
DSR-2000W2R5J-F-V1
DSR-2000WR6J-F-V1
Keysight Infiniium/6054A or
equivalent
Keysight 3458A or equivalent
54Keysight E36200 Series Service Guide
Verification and Adjustments2
CAUTION
NOTE
NOTE
Type Specification Recommended model
Terminations
Variable voltage
transformer or AC
source
Resistors for low
range current
1 – 50 Ω BNC termination
2 – 50 Ω, ≥ 1/8 W resistor
Adjustable to highest rated input voltage range
Power: 500 VA
E36231/E36233A output: 300 Ω high power resistor, 5 WARCOL HS100 300R F
E36232A/E36234A output: 1.2 kΩ power resistor, 5 WARCOL HS75 1K2 F
N/A
Keysight 6813B or equivalent
Test considerations
– Ensure that the calibration ambient temperature is stable and between 20 °C
and 30 °C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
– Performance verification and calibration procedure must be performed
through rear panel output.
The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the
power supply.
Refer to the E36200 Series Programming Guide for more information on the SCPI
commands used in this chapter.
For E36233A and E36234A models, perform the verification test described below
for Output 1 and then repeat the test again for Output 2.
Keysight E36200 Series Service Guide55
2Verification and Adjustments
Measurement techniques
Voltmeter
To ensure that the values read by the voltmeter during both the verification
procedure and the calibration procedure are not affected by the instantaneous
measurement of the AC peaks of the output current ripple, make several DC
measurements and average them.
Current-monitoring resistor
The 4-terminal current shunt is used to eliminate output current measurement
error caused by voltage drops in the load leads and connections. It has special
current-monitoring terminals inside the load connection terminals. Connect the
voltmeter directly to these current-monitoring terminals.
Electronic load
Many of the test procedures require the use of a variable load capable of
dissipating the required power. If a variable resistor is used, switches should be
used to connect, disconnect, or short the load resistor. For most tests, an
electronic load can be used. The electronic load is considerably easier to use than
load resistors, but it may not be fast enough to test transient recovery time and
may be too noisy for the noise (PARD) tests.
Fixed load resistors may be used in place of a variable load, with minor changes to
the test procedures. Also, if computer controlled test setups are used, the
relatively slow (compared to computers and system voltmeters) settling times and
slew rates of the power system may have to be taken into account. “Wait”
statements can be used in the test program if the test system is faster than the
power system.
56Keysight E36200 Series Service Guide
Verification and Adjustments2
Setup for most tests
This setup is used for most tests and it requires the DMM, electronic load, and
power supply being verified. Some wire is also required for connection between
instruments. A LAN or USB cable is needed for readback data. The DMM
measures the power supply output, and the electronic load draws current from
the power supply. The accuracy of the current monitoring resistor must be 0.01%
or better, which should include any self-heating effects.
This setup is used for most tests and it requires the DMM, electronic load, and
power supply being verified. Some wire is also required for connection between
instruments. A LAN or USB cable is needed for readback data. The DMM
measures the power supply output, and the electronic load draws current from
the power supply. The accuracy of the current monitoring resistor must be 0.01%
or better, which should include any self-heating effects.
Keysight E36200 Series Service Guide57
2Verification and Adjustments
Constant Voltage (CV) verification
Voltage programming and readback accuracy
These tests verify that the voltage programming and the LAN or USB readback
functions are within specifications. Note that the readback values over the remote
interface should be identical to those displayed on the front panel, but with
maximum resolution.
1 Turn off the power supply using the AC line switch.
2 Connect a DMM between the (+) and (–) terminals of the output.
3 If you are using a PC to control the power supply, connect a LAN, GPIB, or USB
cable from the power supply to the PC.
4 Turn on the power supply using the AC line switch.
Voltage programming accuracy
Step Front panelSCPI
5Set the instrument settings as
described in Table 2-1.
6Enable the output by pressing the
On key for the selected output.
The output status should be CV and the output current should be close to zero.
7 Record the voltage measured by the DMM in the “Test Record Forms” on
page 74 and verify whether it is within the limits calculated.
Voltage readback accuracy
Step Front panelSCPI
8MEAS:VOLT?, (@<channel>)
VOLT 30, (@<channel>; CURR
20, (@<channel>)
This is an example for the E36231A,
20 V, 20 A output.
OUTP ON, (@<channel>)
This is an example for the E36231A,
30 V, 20 A output.
58Keysight E36200 Series Service Guide
Verification and Adjustments2
9 Record the voltage returned by the SCPI command query via the Keysight
Connection Expert in the “Test Record Forms” on page 74 , and verify whether
it is within the limits calculated.
Table 2-1Voltage programming and readback accuracy instrument settings
This test measures the change in output voltage that results from a change in AC
line voltage from the minimum to maximum value within the line voltage
specifications.
1 Turn off the power supply using the AC line switch.
2 Connect the power supply output with a DMM and an electronic load.
3 Connect a variable AC Source or Variac to the AC input, and set to an
appropriate line voltage for the power supply configuration.
4 If you are using a PC to control the power supply, connect a LAN, GPIB, or USB
cable from the power supply to the PC.
60Keysight E36200 Series Service Guide
Verification and Adjustments2
5 Turn on the power supply using the AC line switch.
6 Set the power supply settings as described in Ta ble 2- 3.
7 Enable the output.
8 Operate the electronic load in constant current mode and set its current to the
value described in Table 2-3.
Check that the front panel CV annunciator of the power supply remains lit. If it
turns to CC or UNREG, adjust the load so that the output current drops slightly
until the CV annunciator lights up.
9 Adjust the AC power source to the low line voltage limit (see Table 2-3).
Record the output reading on the DMM as V
lowline
.
10 Adjust the AC power source to the high line voltage limit (see Table 2-3).
Record the output reading on the DMM as V
highline
.
11 Take the difference between the DMM readings in step 9 and step 10 that is
the CV line regulation (V
lowline
– V
). Record the calculated value in the
highline
“Test Record Forms” on page 74. The difference of the readings during the
immediate change should be within the limit calculated from the specification.
This test measures the time for the output voltage to recover to within the
specified value following a 50% change in the load current.
1 Turn off the power supply using the AC line switch.
2 Connect an oscilloscope and electronic load between the (+) and (–) terminals
of the output to be tested as shown below.
3 If you are using a PC to control the power supply, connect a LAN, GPIB, or USB
cable from the power supply to the PC.
4 Turn on the power supply using the AC line switch.
5 Set the instrument settings as described in Ta ble 2-4.
6 Enable the output.
62Keysight E36200 Series Service Guide
Verification and Adjustments2
NOTE
7 Operate the electronic load in constant current mode and set its current to the
value described in Table 2-4.
Set the transient level to ½ the maximum current. Set the transient duty cycle
to 50% and transient frequency to 1 kHz. Check that the front panel CV
annunciator of the power supply remains lit. If it turns to CC or UNREG, adjust
the maximum current load so that the output current drops slightly until the
CV annunciator lights up.
8 Adjust the oscilloscope to display transients as shown below. Note that the
pulse width (t2 – t1) of the transient at the voltage settling band, for example
15 mV for the E36234A Output 1 from the base line is no more than 50 µs.
– The oscilloscope cursors X1 and X2 represent t1 and t2.
– The oscilloscope green trace and yellow trace represent output current and
output voltage trace.
Record the measured value in the “Test Record Forms” on page 74. The
transient response specification is met when the voltage recovers within
50 µs.
Keysight E36200 Series Service Guide63
2Verification and Adjustments
Tab le 2-4Transient response verification instrument setting
DUT SettingE-LoadTransient Response
ModelVoltage (V)Current (A)Current (A)T-LevelLower l imitUpper limit
E36231A /
E36233A
E36232A /
E36234A
30MAX6.6673.333-50 μs
10MAX2010-50 μs
60MAX3.3331.67-50 μs
20MAX105-50 μs
Output noise verification
Periodic and random output deviations superimpose a residual AC voltage on the
DC output. This residual voltage is specified as the rms or peak-to-peak noise in
and is specified in the product data sheet.
1 Turn off the power supply using the AC line switch.
2 Connect a fixed load, differential amplifier, and an oscilloscope (AC coupled) to
the output as shown below.
3 Turn on the power supply using the AC line switch.
64Keysight E36200 Series Service Guide
Verification and Adjustments2
4 Use an appropriate load resistor (see the fixed load value in the Recommended
test equipment list) to keep the power system at the instrument setting
specified in Table 2-6.
5 As shown in the figure, use two BNC cables to connect the differential
amplifier to the (+) and (−) output terminals. Each cable should be terminated
by a 50 Ω resistor. The shields of the two BNC cables should be connected
together. Connect the differential amplifier output to the oscilloscope with a
50 Ω termination at the oscilloscope input.
6 Set the differential amplifier to multiply by ten, divide by one, and 1 MΩ input
resistance. Set the differential amplifier's positive and negative inputs to AC
coupling. Set the oscilloscope’s time base to 5 ms/div, and the vertical scale to
10 mV/div. Turn the bandwidth limit on (usually 20 or 30 MHz), and set the
sampling mode to peak detect.
7 Program the power supply to the settings indicated in the in the test record
form for the appropriate model under Tab le 2-6 and enable the output.
Let the oscilloscope run for a few seconds to generate enough measurement
points. On the Keysight Infiniium oscilloscope, the maximum peak-to-peak
voltage measurement is indicated at the bottom of the screen on the right
side. Divide this value by 10 to get the CV peak-to-peak noise measurement.
The result should not exceed the peak-to-peak upper limits for instrument's
“CV ripple and noise, peak-to-peak” value. See the “Test Record Forms” on
page 74 for details.
8 Disconnect the oscilloscope and connect an rms voltmeter in its place. Do not
disconnect the 50 Ω termination. Divide the reading of the rms voltmeter by
10. The result should not exceed the rms limits in the test record form for the
appropriate model under “CV ripple and noise, rms”. See the “Test Record
Forms” on page 74 for details.
Keysight E36200 Series Service Guide65
2Verification and Adjustments
Tab le 2-5CV ripple and noise, peak-to-peak instrument settings
Model
Voltage (V)Current (A)Value (W)Lower limitUpper limit
E36231A /
E36233A
E36232A /
E36234A
DUT settingsFixed loadRMS noise
30MAX4.5-4.5 mV
10MAX0.5-4.5 mV
60MAX18-3.5 mV
20MAX2-3.5 mV
Tab le 2-6CV ripple and noise, rms instrument settings
Model
Voltage (V)Current (A)Value (W)Lower limitUpper limit
E36231A /
E36233A
E36232A /
E36234A
DUT settingsFixed loadPeak-to-peak noise
30MAX4.5-0.35 mV
10MAX0.5-0.35 mV
60MAX18-0.35 mV
20MAX2-0.35 mV
66Keysight E36200 Series Service Guide
Low range current verification
Low range current readback accuracy
These tests verify that the LAN, GPIB, or USB readback measurement functions
are within specifications. Note that the readback values over the remote interface
should be identical to those displayed on the front panel, but with maximum
resolution.
1 Turn off the power supply using the AC line switch.
2 Connect a DMM between the (+) and (–) terminals of the output.
3 Connect appropriate resistor in series with the (+) terminals of the power
supply and (+) terminals of the DMM as shown below.
Verification and Adjustments2
4 If you are using a PC to control the power supply, connect a LAN, GPIB, or USB
cable from the power supply to the PC.
5 Turn on the power supply using the AC line switch.
Keysight E36200 Series Service Guide67
2Verification and Adjustments
Low range current readback accuracy
Step Front panelSCPI
6Set the instrument settings as
described in Table 2-7.
VOLT 60, (@<channel>); CURR
0.05, (@<channel>)
This is an example for the E36234A,
60 V, 50 mA output.
7Enable the output by pressing the
OUTP ON, (@<channel>)
On key for the selected output.
8Record the current measured by the
DMM and current returned by the
SCPI MEAS:CURR?, (@channel)
command query via the Keysight
MEAS:CURR?, (@<channel>)
This is an example for the E36234A,
60 V, 50 mA output.
Connection Expert in the “Test
Record Forms” on page 74, and
verify whether it is within the limits
calculated.
Tab le 2-7Low range current readback accuracy instrument settings
Model
Voltage (V)Current (A)Lower limitUpper limit
E36231A/E36233A300.1–410 μA410 μA
DUT settingsRead back accuracy
E36232A/E36234A600.05–205 μA–205 μA
68Keysight E36200 Series Service Guide
Constant Current (CC) verification
Current programming and readback accuracy
These tests verify that the current programming and the LAN, GPIB, or USB
readback measurement functions are within specifications. Note that the
readback values over the remote interface should be identical to those displayed
on the front panel, but with maximum resolution.
1 Turn off the power supply using the AC line switch.
2 Connect the current shunt directly across the output terminals. Connect the
DMM directly across the current shunt.
3 If you are using a PC to control the power supply, connect a LAN, GPIB or USB
cable from the power supply to the PC.
4 Turn on the power supply using the AC line switch.
Current programming accuracy
Step Front panelSCPI
Verification and Adjustments2
5Set the instrument settings as
described in Tab le 2-8.
6Enable the output by pressing the
On key for the selected output.
7 The output status should be CC and the output voltage should be close to
zero.
8 Divide the voltage drop (DMM reading) across the current shunt by the shunt
resistance to convert to amps.
9 Record the current measured by the DMM and verify whether it is within the
limits calculated.
Keysight E36200 Series Service Guide69
VOLT MAX, (@<channel>); CURR
10, (@<channel>)
This is an example for the E36234A,
MAX V, 5 A output.
OUTP ON, (@<channel>)
2Verification and Adjustments
Current readback accuracy
Step Front panelSCPI
10MEAS:CURR?, (@<channel>)
This is an example for the E36234A,
MAX V, 5 A output.
11 Record the current returned by the SCPI command query via the Keysight
Connection Expert in the “Test Record Forms” on page 74, and verify whether
it is within the limits calculated.
Tab le 2-8Current programming and readback accuracy instrument settings
Model
E36231A /
E36233A
E36232A /
E36234A
DUT settingsProgramming accuracyReadback accuracy
Voltage (V)Current (A)Lower limitUpper limitLower limitUpper limit
MAX0–0.006 A0.006 A–0.006 A0.006 A
MAX2019.974 A20.026 A–0.026 A0.026 A
MAX0–0.003 A0.003 A–0.003 A0.003 A
MAX109.987 A10.013 A–0.013 A0.013 A
70Keysight E36200 Series Service Guide
Verification and Adjustments2
CC load regulation
This test measures the change in output current resulting from a change in output
voltage from full scale to short circuit.
1 Turn off the power supply using the AC line switch.
2 Connect the power supply output with a DMM, an electronic load, and a
current shunt as shown below.
3 Program the output voltage and output current as described in the test record
form under CC load regulation.
4 Turn on the power supply using the AC line switch.
5 Enable the output by sending the command OUTP ON or by pressing the
output [On] key.
6 Operate the electronic load in constant voltage mode and set its voltage to the
power supply output value as described in Table 2-9.
Check that the front panel CC annunciator of the power supply remains lit. If it
turns to CV or UNREG, adjust the load so that the output current drops slightly
until the CC annunciator lights up. Record the current reading (I
load
), by
dividing the voltage reading on the DMM by the resistance of the current
monitoring resistor.
Keysight E36200 Series Service Guide71
2Verification and Adjustments
7 Operate the electronic load in short (input short) mode. Record the current
reading, (I
), by dividing the voltage reading on the DMM by the resistance
short
of the current monitoring resistor.
8 Take the difference between the current readings in step 6 and step 7 is the
load regulation current (I
load
– I
). Record the calculated value in the “Test
short
Record Forms” on page 74. The difference of the readings during the
immediate change should be within the specification limits.
Maximum current measured over interfaceMax V, 10 A
CC load regulation (I
CC line regulation (I
load
lowline
- I
- I
short
highline
)
)
Max V, 3.333 A–0.5833 mA____________0.5833 mA
Max V, 10 A–1.250 mA1.250 mA
Max V, 3.333 A–0.5833 mA____________0.5833 mA
Max V, 10 A–1.250 mA____________1.250 mA
Low range current measurement
Maximum low range current measured over interface60 V, 0.05 A
I
0
I
max
I
low max
– 0.003 A
– 0.013 A
– 205 μA
____________
____________
____________
I0 + 0.003 A
I
+ 0.013 A
max
I
low max
+ 205 μA
Keysight E36200 Series Service Guide77
2Verification and Adjustments
Table 2-13Test record form - Keysight E36233A
E36233AReport number _______________Date _______________
DescriptionDUT settingLower limit
Constant voltage tests
Voltage programming
Zero voltage output (V
Maximum voltage output (V
)
0
)
max
0 V, 20 A–0.004 V____________ ____________0.004 V
30 V 20 A29.987 V____________ ____________30.013 V
Voltage readback
Zero voltage measured over interface0 V, 20 A
V0 – 0.004 V
____________ ____________
Result
Output 1Output 2
Upper limit
V0 + 0.004 V
Maximum voltage measured over
interface
CV load regulation (V
CV line regulation (V
load
lowline
- V
noload
- V
CV ripple and noise
peak-to-peak
rms
Transient response
)
highline
V
30 V 20 A
– 0.013 V
max
____________ ____________
30 V, Max A–5 mV____________ ____________5 mV
10 V, Max A–3 mV____________ ____________3 mV
30 V, Max A–5 mV____________ ____________5 mV
)
10 V, Max A–3 mV____________ ____________3 mV
30 V, Max A-____________ ____________4.5 mV
10 V, Max A-____________ ____________4.5 mV
30 V, Max A-____________ ____________0.35 mV
10 V, Max A-____________ ____________0.35 mV
30 V, Max A-____________ ____________50 μs
10 V, Max A-____________ ____________50 μs
V
max
+ 0.013 V
78Keysight E36200 Series Service Guide
Verification and Adjustments2
Table 2-13Test record form - Keysight E36233A (continued)
E36233AReport number _______________Date _______________
DescriptionDUT settingLower limit
Constant current tests
Current programming
Zero current output (I
Maximum current ouput (I
)
0
)
max
Max V, 0 A–0.0060 A____________ ____________0.0060 A
Max V, 20 A19.974 A____________ ____________20.026 A
Current readback
Zero current measured over interfaceMax V, 0 A
I0 – 0.0060 A
____________ ____________
Result
Output 1Output 2
Upper limit
I0 + 0.0060 A
Maximum current measured over
interface
- I
- I
short
highline
)
)
CC load regulation (I
CC line regulation (I
load
lowline
Low range current measurement
Maximum low range current measured
over interface
Max V, 20 A
max
– 0.026 A
____________ ____________
I
+ 0.026 A
max
I
Max V, 6.667 A–0.9167 mA____________ ____________0.9167 mA
Max V, 20 A–2.250 mA2.250 mA
Max V, 6.667 A–0.9167 mA____________ ____________0.9167 mA
Max V, 20 A–2.250 mA____________ ____________2.250 mA
30 V, 0.1 A
I
low max
– 410 μA
____________ ____________
I
low max
+ 410 μA
Keysight E36200 Series Service Guide79
2Verification and Adjustments
Table 2-14Test record form - Keysight E36234A
E36234AReport number _______________Date _______________
DescriptionDUT settingLower limit
Constant voltage tests
Voltage programming
Zero voltage output (V
Maximum voltage output (V
)
0
)
max
0 V, 10 A–0.008 V____________ ____________0.008 V
60 V 10 A59.974 V____________ ____________60.026 V
Voltage readback
Zero voltage measured over interface0 V, 10 A
V0 – 0.008 V
____________ ____________
Result
Output 1 Output 2
Upper limit
V0 + 0.008 V
Maximum voltage measured over
interface
CV load regulation (V
CV line regulation (V
load
lowline
- V
noload
- V
CV ripple and noise
peak-to-peak
rms
Transient response
)
highline
V
60 V 10 A
– 0.026 V
max
____________ ____________
60 V, Max A–8 mV____________ ____________8 mV
20 V, Max A–4 mV____________ ____________4 mV
60 V, Max A–8 mV____________ ____________8 mV
)
20 V, Max A–4 mV____________ ____________4 mV
60 V, Max A-____________ ____________3.5 mV
20 V, Max A-____________ ____________3.5 mV
60 V, Max A-____________ ____________0.35 mV
20 V, Max A-____________ ____________0.35 mV
60 V, Max A-____________ ____________50 μs
20 V, Max A-____________ ____________50 μs
V
max
+ 0.026 V
80Keysight E36200 Series Service Guide
Verification and Adjustments2
Table 2-14Test record form - Keysight E36234A (continued)
E36234AReport number _______________Date _______________
DescriptionDUT settingLower limit
Constant current tests
Current programming
Zero current output (I
Maximum current ouput (I
)
0
)
max
Max V, 0 A–0.003 A____________ ____________0.003 A
Max V, 10 A9.987 A____________ ____________10.013 A
Current readback
Zero current measured over interfaceMax V, 0 A
I0 – 0.003 A
____________ ____________
Result
Output 1 Output 2
Upper limit
I0 + 0.003 A
Maximum current measured over
interface
- I
- I
short
highline
)
)
CC load regulation (I
CC line regulation (I
load
lowline
Low range current measurement
Maximum low range current measured
over interface
Max V, 10 A
max
– 0.013 A
____________ ____________
I
+ 0.013 A
max
I
Max V, 3.333 A–0.5833 mA____________ ____________0.5833 mA
Max V, 10 A–1.250 mA1.250 mA
Max V, 3.333 A–0.5833 mA____________ ____________0.5833 mA
Max V, 10 A–1.250 mA____________ ____________1.250 mA
60 V, 0.05 A
I
low max
– 205 μA
____________ ____________
I
low max
+ 205 μA
Keysight E36200 Series Service Guide81
2Verification and Adjustments
Calibration Adjustment Procedures
This chapter includes calibration adjustment procedures for the E36200 Series
autoranging DC power supplies. Instructions are applicable for performing the
procedures from either the front panel or a controller over the LAN, USB, or GPIB.
Perform the verification tests (see “Performance Verification” on page 54) before
calibrating your instrument. If the instrument passes the verification tests, the unit
is operating within its calibration limits and does not need to be re-calibrated.
Closed-case electronic calibration
The instrument uses closed-case electronic calibration; no internal mechanical
adjustments are required. The instrument calculates the correction factors based
on reference signals that you apply and stores the correction factors in the
instrument’s non-volatile memory. This data is not changed by cycling power,
*RST, or SYSTem:PRESet.
Calibration interval
The recommended calibration interval for the E36200 Series autoranging DC
power supplies is one year.
Calibration adjustment process
The following general procedure is recommended to complete a full calibration
adjustment.
1 Adhere to the test considerations. See “Test considerations” on page 55 for
details.
2 Perform the performance verification tests to characterize the instrument. See
“Performance Verification” on page 54 for details.
3 Unsecure the instrument for calibration. See “Calibration security” on page 83
for details.
4 Perform the calibration procedures. See “Calibration procedure” on page 86
for details.
5 Secure the instrument against the calibration. See “Calibration security” on
page 83 for details.
82Keysight E36200 Series Service Guide
6 Take note of the security code and calibration count in the instrument's
maintenance records.
7 Perform the performance verification tests again to verify the calibration.
Calibration security
The instrument has a calibration passcode to prevent accidental or unauthorized
calibration. When you receive your power supply, it is secured by a default
passcode. The default passcode is 0. The security code cannot be changed by a
power cycle or *RST.
You can enter a passcode of up to 9 digits.
You can change the passcode from both front panel and remote interface.
To change a forgotten passcode to a new passcode, you can turn on the
instrument after shorting CAL SECURE jumper J10 on the front panel board as
shown below and send CAL:SEC:CODE <code> to change the passcode.
Calibration count
The instrument counts the number of times it has saved calibration data. Your
instrument was calibrated at the factory; when you receive your instrument, read
and record the initial count. You can only read the calibration count by sending
the CALibration:COUNt? query, and the calibration count is not change by a
power cycle or *RST.
If Auto Save is enabled, the count increments when you exit the calibration state.
To avoid double counting, do not manually save the count with Auto Save
enabled.
84Keysight E36200 Series Service Guide
Calibration message
You can use the CALibration:STRing command to store a message of up to 40
characters in calibration memory. For example, you could store the last calibration
date, the calibration due date, or contact information for the person responsible
for calibration. The calibration message is not affected by a power cycle or *RST.
You can only store the calibration message when the instrument is unsecured, but
you can execute the CALibration:STRing? query regardless of whether the
instrument is secured. A new calibration message overwrites the previous
message, and messages over 40 characters are truncated.
Saving calibration data
You must always save new calibration data before cycling instrument power or
leaving the calibration state with the Auto Save feature off. To save calibration
data, send CALibration:SAVE or save the calibration data from the front panel.
Calibration auto save
The instrument includes a calibration Auto Save feature. This feature
automatically saves the calibration data to non-volatile memory and increments
the calibration count when you exit the calibration state.
To enable or disable the calibration auto save feature, send CALibration:ASAVe
ON or CALibration:ASAVe OFF. To query the calibration auto save state, send
CALibration:ASAV?.
Verification and Adjustments2
Keysight E36200 Series Service Guide85
2Verification and Adjustments
Calibration procedure
Enter the calibration state
To begin the calibration procedure, you must enter the calibration state.
StepFront panelSCPI
Press Utilities > Test / Setup > Calibration.
1
Enter the default passcode (default passcode is 0). Press Login to
enter the calibration page.
2To change a forgotten passcode to a new passcode, you can turn on
the instrument after shorting the CAL SECURE jumper J10 on the
front panel board and sending the SCPI command shown on the
right to change the passcode.
Calibrate voltage
Allow the unit to sit with its output turned ON for one minute, then connect the
DMM voltage input to the power supply.
CAL:SEC:STAT 0, <code>
CAL:SEC:CODE <new code>
StepFront panelSCPI
Press Perform Calibration. For E36233A and E36234A models,
1
select the channel for calibration.
Press Cal Volt.
2Measure the output voltage (low point) with the DMM.
3Enter the measured value, and press Next.
4Measure the output voltage (high point) with the DMM.
5Enter the measured value, and press Next.
6Read DONE or FAIL on the display.(wait 30 seconds)
Alllow the unit to sit with its output turned ON for one minute, then connect a
current monitoring resistor across the output terminals to be calibrated and then
connect a DMM across the terminals of the monitoring resistor.
StepFront panelSCPI
Press Perform Calibration. For E36233A and E36234A models,
1
select the channel for calibration.
Press Cal Curr.
2Measure the output current (low point) with the DMM.
3Enter the measured value, and press Next.
4Measure the output current (high point) with the DMM.
5Enter the measured value, and press Next.
6Read DONE or FAIL on the display.(wait 30 seconds)
Allow the unit to sit with its output turned ON for one minute before continuing.
StepFront panelSCPI
Press Perform Calibration. For E36233A and E36234A models,
1
select the channel for calibration.
Press Cal Low Curr.
2Wait approximately 50 seconds.
Connect the calibration resistor with the correct value as stated in the instructions shown on the display.
Wait until a stable reading is obtained before you measure the output low range current (low point) with the DMM.
3Enter the measured value, and press Next.
4Wait until a stable reading is obtained before you measure the output current (high point) with the DMM.
6Read DONE or FAIL on the display.(wait 30 seconds)
7Press Cal Save.
CAL:CURR <reading>, (@<channel>)
SYST:ERR?
CAL:SAVE
Save the calibration data
To save calibration data, go to the CAL Save menu to save the calibration data or
enable the Auto Save feature. With Auto Save, the calibration data will be saved
when you exit the calibration menu.
After completing the Voltage, Current, and Low Range Current calibrations, save
the calibration data before exiting the calibration state, or simply exit the
calibration state if Auto Save is turned On.
–To save the CAL data: CAL:SAVE
– To enable the CAL Auto Save: CAL:ASAVE ON
– To exit the CAL State: CAL:STAT 1 <code>
88Keysight E36200 Series Service Guide
This information is subject to change
without notice. Always refer to the
Keysight website for the latest revision.