Copyright Notice8
Manual Part Number8
Edition8
Published by8
Warranty8
Technology Licenses8
U.S. Government Rights9
Waste Electrical and Electronic Equipment (WEEE)9
Declarations of Conformity10
Safety Information10
Safety and Regulatory Information11
Safety Considerations11
BEFORE APPLYING POWER11
GROUND THE INSTRUMENT11
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE OR WET ENVIRONMENTS11
DO NOT OPERATE DAMAGED OR DEFECTIVE INSTRUMENTS12
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT12
BE CAUTIOUS WHEN DEALING WITH OUTPUTS ABOVE 60 VDC12
USE THE POWER CORD PROVIDED12
USE THE DEVICE AS SPECIFIED12
CLEAN WITH SLIGHTLY DAMPENED CLOTH13
DO NOT BLOCK VENTILATION HOLES13
OBSERVE ALL DEVICE MARKINGS BEFORE CONNECTING TO DEVICE13
TURN DEVICE OFF BEFORE CONNECTING TO OUTPUT TERMINALS13
ENSURE COVER IS SECURED IN PLACE13
Safety Symbols14
Regulatory Markings15
South Korean Class A EMC declaration:15
Safety and EMC Requirements16
1Getting Started18
Introduction19
Front Panel20
Rear Panel21
Display22
To Set up the Instrument22
Options and Fuse Information24
Ensure the correct AC input voltage and fuse settings24
Use the proper switch settings24
Use the correct fuse25
Programming Ranges26
Table of Programming and Readback Resolutions26
Extending the Voltage Range and Current Range27
Series connections27
Parallel connections27
Front-panel Operation28
Configure the LAN interface28
Set voltage and current29
Constant voltage and constant current29
Specify 2- or 4-wire measurement29
Configure Overcurrent Protection (OCP) and Overvoltage Protection (OVP)30
Clear an OCP or OVP event30
Clear an overtemperature protection (OTP) event31
Lock and unlock the front panel31
Save or recall the instrument's state32
To save or recall the instrument state32
Configure the Power-on State32
Read error codes33
Remote Control34
Keysight IOLibraries Suite34
E36100 Series Web interface34
Technical Connection Details37
2SCPI Programming39
Introduction to the SCPI Language40
Command format used in this manual41
Command separators42
Using the MIN and MAX parameters42
Querying parameter settings43
SCPI command terminators43
IEEE-488.2 common commands43
SCPI parameter types44
OUTPut[:STATe] ON |1 | OFF |0OUTPut[:STATe]?62
OUTPut:PROTection:CLEar62
OUTPut:PON:STATe
RST|RCL0|RCL1|RCL2|RCL3|RCL4|RCL5|RCL6|RCL7|RCL8|RCL9|OUTPut:PON:STATe?62
Voltage programming and readback accuracy78
CV load and line regulation79
CV load regulation79
CV line regulation80
Transient response verification80
Output noise verification83
Constant Current (CC) verification85
Current programming and readback accuracy85
CC load regulation86
CCline regulation87
No part of this manual may be reproduced in any form or by any means (including electronic storage
and retrieval or translation into a foreign language) without prior agreement and written consent from
Keysight Technologies as governed by United States and international copyright laws.
Manual Part Number
E36100-90001
Edition
Edition 2, November 11, 2016
Published by
Keysight Technologies
900 S. Taft Ave.
Loveland, CO 80537
USA
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT
PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN,
INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE
OF THIS DOCUMENT OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE
USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL
IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be
used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”)
2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”)
227.7202, the U.S. government acquires commercial computer software under the same terms by
which the software is customarily provided to the public. Accordingly, Keysight provides the Software to
U.S. government customers under its standard commercial license, which is embodied in its End User
License Agreement (EULA), a copy of which can be found at http://www.keysight.com/find/sweula. The
license set forth in the EULA represents the exclusive authority by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require
or permit, among other things, that Keysight: (1) Furnish technical information related to commercial
computer software or commercial computer software documentation that is not customarily provided
to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose
commercial computer software or commercial computer software documentation. No additional
government requirements beyond those set forth in the EULA shall apply, except to the extent that
those terms, rights, or licenses are explicitly required from all providers of commercial computer
software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the
EULA. Keysight shallbe under no obligation to update, revise or otherwise modify the Software. With
respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR
27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Waste Electrical and Electronic Equipment (WEEE)
This product complies with the WEEE Directive) marketing requirement. The affixed product label (see
below) indicates that you must not discard this electrical/electronic product in domestic household
waste.
Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is
classified as “Monitoring and Control instrumentation” product. Do not dispose in domestic household
waste.
To return unwanted products, contact your local Keysight office, or see
about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.
Declarations of Conformity
Declarations of Conformity for this product and for other Keysight products may be downloaded from
the Web. Go to http://regulations.corporate.keysight.com/DoC/search.htm and click on “Declarations
of Conformity.” You can then search by product number to find the latest Declaration of Conformity.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice
until the indicated conditions are fully understood and met.
Safety and Regulatory Information
Safety Considerations
The following general safety precautions must be observed during all phases of operation, service, and
repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in
this manual violates safety standards of design, manufacture, and intended use of the instrument.
Keysight Technologies assumes no liability for the customer's failure to comply with these requirements.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage and that the correct fuse is
installed. Use the label on the bottom of the instrument to configure the power.
See Options and Fuse Information for additional details.
GROUND THE INSTRUMENT
This product is a Safety Class I instrument (provided with a protective earth terminal). To
minimize shock hazard, the instrument chassis and cabinet must be connected to an
electrical ground. The instrument must be connected to the AC power supply mains
through a three-conductor power cable, with the third wire firmly connected to an
electrical ground (safety ground) at the power outlet. Any interruption of the protective
(grounding) conductor or disconnection of the protective earth terminal will cause a
potential shock hazard that could result in personal injury. If the instrument is to be
energized via an external autotransformer for voltage reduction, be certain that the
autotransformer common terminal is connected to the neutral (earthed pole) of the AC
power lines (supply mains).
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE OR WET ENVIRONMENTS
Do not operate the device around flammable gases or fumes, vapor, or wet environments.
DO NOT OPERATE DAMAGED OR DEFECTIVE INSTRUMENTS
Instruments that appear damaged or defective should be made inoperative and secured
against unintended operation until they can be repaired by qualified service personnel.
DO NOT SUBSTITUTE PARTS OR MODIFY INSTRUMENT
Because of the danger of introducing additional hazards, do not installsubstitute parts or
perform any unauthorized modification to the instrument. Return the instrument to a
Keysight Technologies Sales and Service Office for service and repair to ensure that safety
features are maintained.
To contact Keysight for sales and technical support, refer to the support links on the
following Keysight website:www.keysight.com/find/assist (worldwide contact information
for repair and service).
BE CAUTIOUS WHEN DEALING WITH OUTPUTS ABOVE 60 VDC
Always be cautious when dealing with output voltage above 60 VDC. Such voltage poses
electrical shock hazard.
USE THE POWER CORD PROVIDED
Use the device with the power cord provided with the shipment.
USE THE DEVICE AS SPECIFIED
If the device is used in a manner not specified by manufacturer, the device protection may
be impaired.
CLEAN WITH SLIGHTLY DAMPENED CLOTH
Clean the outside of the instrument with a soft, lint-free, slightly dampened cloth. Do not
use detergent, volatile liquids, or chemical solvents.
DO NOT BLOCK VENTILATION HOLES
Do not block any of the ventilation holes of the device.
OBSERVE ALL DEVICE MARKINGS BEFORE CONNECTING TO DEVICE
Observe all markings on the device before connecting any wiring to the device.
TURN DEVICE OFF BEFORE CONNECTING TO OUTPUT TERMINALS
Turn off the device power before connecting to the output terminals.
ENSURE COVER IS SECURED IN PLACE
Do not operate the device with the cover removed or loosened.
Safety Symbols
SymbolMeaning
Caution, refer to accompanying documentation.
Protective conductor terminal.
Frame or chassis terminal.
In position of bi-stable push control.
Out position of bi-stable push control.
Alternating current.
Direct current.
Plus, positive polarity.
Minus, negative polarity.
The WARNING sign denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not
correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign
until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the like, which, if not
correctly performed or adhered to, could result in damage to or destruction of part or all of the product. Do
not proceed beyond CAUTION sign until the indicated conditions are fully understood and met.
The NOTE sign denotes important information. It calls attention to a procedure, practice, condition or the like,
which is essential to highlight.
Regulatory Markings
SymbolMeaning
The RCM Mark is a Compliance Mark according to the ACMA Labeling Requirement.
The CE marking is the legal required labeling for several EU Directives of the European Union. This CE
marking shows that the product complies with all the relevant European Legal Directives.
ISM 1-AThis product is an Industrial Scientific and Medical Group 1 Class A product. (CISPR11, Clause 4)
ICES/NMB-001This is a marking of a product in compliance with the Canadian Interference-Causing Equipment
Standard (ICES-001).
This symbol indicates the time period during which no hazardous or toxic substance elements are
expected to leak or deteriorate during normal use. Forty years is the expected useful life of the product.
This KC mark shows that the product complies with the relevant Korean compulsory certifications.
South Korean Class A EMC declaration:
This equipment is Class A suitable for professional use and is for use in electromagnetic environments
outside of the home.
Safety and EMC Requirements
This power supply is designed to comply with the following safety and EMC (Electromagnetic
Compatibility) requirements:
– Low Voltage Directive (2006/95/EC)
– EMC Directive (2004/108/EC)
– IEC61010-1:2010/EN61010-1:2010
– IEC61326-1:2012/EN61326-1:2013
1Getting Started
Introduction
Options and Fuse Information
Programming Ranges
Extending the Voltage Range and Current Range
Front-panel Operation
Remote Control
Introduction
The Keysight E36100 Series is a series of DC bench and system power supplies with the following
features and characteristics:
– Single output up to 100 V or 5 A
– Small size: 2U, 1/4 rack
– Remote sense capability
– Intuitive front panel
– High visibility organic light emitting diode (OLED)display
– High performance functionality: accuracy, transient response, and rise/fall time
– LAN (LXICore) and USB interfaces for remote programming with SCPI
Front Panel
ItemDescription
ATough carrying handle
BInformation-packed, high-contrast OLED display; easily viewable even from sharp angles
CRotary knob for quick and easy configuration
DFast voltage/current setting and front-panel electronic calibration
EMenu key opens intuitive user interface
FFront-panel lock prevents accidental changes during tests
GOutput enable/disable switch to protect your DUT quickly
HDual-position power switch
ISense terminals
JOutput terminals
KEarth ground reference point
Rear Panel
The rear panel includes the power input, standard LAN and USB ports, and a security slot. Always use
the power cord that arrived with the instrument or one with equivalent ratings.
Display
The power supply's front panel features a high-visibility OLED display.
The upper left corner of the display shows either CV or CC to indicate that the instrument is in constant
voltage or constant current mode. Below that is either 2W or 4W, to indicate whether 2-wire (normal)
or 4-wire (remote sensing) measurement is in use. The right side of the display shows the output voltage
and current (when the output is enabled), and the voltage and current settings are shown near the
bottom of the screen.
The very bottom of the screen can show arrows, each of which represents a different annunciator.
LabelMeaning
ERR An error has occurred.
OCP An overcurrent protection event has occurred.
OVP An overvoltage protection event has occurred.
RMT The instrument is being programmed remotely.
LCK The front panel is locked.
To Set up the Instrument
Place the instrument's feet on a flat, smooth horizontal surface. Connect output and sense leads to the
front panel, being careful not to short the leads together. Attach the power cable to the rear panel, then
plug it into main power. Connect LAN or USB cables as desired, and you may also secure the instrument
with a security lock cable.
Before disconnecting cables and cords from the instrument, turn the instrument off using the frontpanel power switch and disconnect from the supply source by unplugging the detachable power cord.
Options and Fuse Information
Be certain to order the proper instrument option for the mains power that will be used for
the instrument.
Options 0EM, 0E3, and 0E9 determine which power-line voltage is selected at the factory. The default is
configured for 230 VAC ± 10%, 47-63 Hz input voltage (option 0E3).
OptionDescription
0EM115 VAC ± 10%, 47-63Hz input voltage
0E3230 VAC ± 10%, 47-63Hz input voltage
0E9100 VAC ± 10%, 47-63Hz input voltage
Ensure the correct AC input voltage and fuse settings
Use the proper switch settings
To change the input AC voltage selector on the power supply, use the two AC selector switches on the
underside of the power supply as shown.
For example, to select 230 V, move the switch on the left to the left and the switch on the right to the
right, as illustrated in the diagram below the switches. To select 115 V, move both switches to the
position on the right, and to select 100 V, move the two switches toward each other.
Use the correct fuse
You must use a time delay 2-A fuse for 100- and 115-VAC inputs, and a time delay 1-A fuse for 230-VAC
inputs:
Part Number DescriptionManufacturer Application
2110-1639Fuse 1A 250V Time-delay Littelfuse230-V line voltage
2110-1640Fuse 2A 250V Time-delay100- and 115-V line voltage
To configure the correct fuse, follow the three steps shown below:
Pull the fuse holder out of the power supply.
Insert the proper fuse into the fuse holder.
Re-insert the fuse holder into the power supply.
Programming Ranges
The following table shows the maximum voltage, current, and power that can be programmed for each
model. The DEFault voltage is always 0 V.
Model Max Voltage (V) Max Current (A) Max Power(W)
E36102A65.030
E36103A202.040
E36104A351.035
E36105A600.636
E36106A1000.440
Note that the actual programmable values are 3% above these nominal values, but instrument
specifications are only guaranteed up to the values shown in this table.
Table of Programming and Readback Resolutions
Model Voltage Programming Voltage Readback Current Programming Current ReadbackSmall Current Read-
back
E36102A.001 V.001 V.001 A.001 A.001 mA
E36103A.001 V.001 V.001 A.001 A.001 mA
E36104A.001 V.001 V.001 A.001 A.001 mA
E36105A.001 V.01 V.001 A.0001 A.001 mA
E36106A.001 V.01 V.001 A.0001 A.001 mA
Extending the Voltage Range and Current Range
The power supply may be able to provide voltages and currents greater than its rated maximum outputs
if the power- line voltage is at or above its nominal value.
Operation can be extended up to 3% over the rated output without damage to the power supply, but
performance cannot be guaranteed to meet specifications in this region. If the power- line voltage is
maintained in the upper end of the input voltage range, the power supply will probably operate within its
specifications. The power supply is more likely to stay within specifications if only one of the voltage or
current outputs is exceeded.
Series connections
You can connect up to four E36100 Series power supplies up to the output isolation rating of any one
supply to obtain a higher voltage than a single power supply can produce. Series-connected power
supplies can be operated with one load across both power supplies or with a separate load for each
power supply. The power supply has a reverse polarity diode connected across the output terminals so
that if operated in series with other power supplies, damage will not occur if the load is short-circuited
or if one power supply is turned on separately from its series partners.
When the series connection is used, the output voltage is the sum of the voltages of the individual
power supplies. The current is the current of any one power supply. Each of the individual power
supplies must be adjusted in order to obtain the total output voltage.
Parallel connections
You can connect up to four E36100 Series power supplies in parallelto obtain a total output current
greater than a single power supply can produce. The total output current is the sum of the output
currents of the individual power supplies. The output of each power supply can be set separately. The
output voltage controls of one power supply should be set to the desired output voltage; the other
power supply should be set for a slightly higher output voltage. The supply with the higher output
voltage setting will deliver its constant current output, and drop its output voltage until it equals the
output of the other supply, and the other supply will remain in the constant voltage operation and only
deliver that fraction of its rated output current which is necessary to fulfill the total load demand.
Front-panel Operation
This section describes how to perform the following operations from the instrument's front panel.
Configure the LAN interface
Set voltage and current
Specify 2- or 4-wire measurement
Configure Overcurrent Protection (OCP) and Overvoltage Protection (OVP)
Clear an OCP or OVP event
Clear an overtemperature protection (OTP) event
Lock and unlock the front panel
Save or recall the instrument's state
Configure the Power-on State
Read error codes
Configure the LAN interface
You should generally be able to connect to the LAN by simply connecting the LAN cable and turning the
instrument on. However, you can use the procedure below if you want to specify connections other
than the default values or if you need to see what a given value is. This section provides procedures for
viewing and changing LANsettings.
1. Press [Menu].
2. Turn the knob to I/O Config and press the knob.
3. Turn the knob to LAN and press the knob.
4.
Turn the knob to the desired IP Address to a value of the form xxx.xxx.xxx.xxx, where each xxx represents a base 10 number from 0 to 255. Press [Voltage/Current] to move between numbers, then
press the knob to finish.
5.
The instrument will briefly display CHANGE SAVED to indicate success.
6.
Turn the knob to the desired Subnet address using the same method with which you set the IP
Address. Then press the knob to select.
7.
The instrument will briefly display CHANGE SAVED to indicate success.
8.
Turn the knob to the desired Gateway address. Then press the knob to select.
9.
The instrument will briefly display CHANGE SAVED to indicate success.
10. Press [Menu] to exit the menu system.
Set voltage and current
If you are in a menu, you must exit the menu before setting the voltage and current.
1. Press [Voltage / Current].
2.
Turn the knob to the desired voltage value, pushing the knob to move between digits.
3. Press [Voltage / Current].
4.
Turn the knob to the desired current value, pushing the knob to move between digits.
5. Press [Voltage / Current].
6. If the displaycurrently shows OFF, press [Output On / Off] to enable the output. Press this key again
to turn the output off.
Constant voltage and constant current
If the output load resistance exceeds the voltage setting divided by the current setting, the instrument
will operate in constant voltage mode. The current willequal the voltage divided by the load resistance.
If the output load resistance is less than the voltage setting divided by the current setting, the
instrument will operate in constant current mode. The voltage will equal the current multiplied by the
load resistance.
Specify 2- or 4-wire measurement
You have the option to use the remote sense terminals on the front panel to measure the voltage at the
device under test (DUT). To specify whether you want to use this option:
1. Press [Menu].
2. Turn the knob to Sense Setting and press the knob.
3. Turn the knob to Internal or External and press the knob.
The internal setting sets a relaywithin the power supply to connect the output and sensing connector.
This means that only two wires are used; and remote sensing is disabled. The external setting sets a relay
within the power supply to disconnect the output and remote sensing inputs. This means that four wires
are used; and remote sensing is enabled.
Configure Overcurrent Protection (OCP) and Overvoltage Protection (OVP)
1. Press [Menu].
2. Turn the knob to OCP Set or OVP Set and press the knob.
3.
Press the knob again to choose OCP Delay or OVP Level.
4.
Turn the knob to the desired OCP or OVPlimit value. Then press the knob.
5. Turn the knob to OCP On, OCP Off, OVPOn, or OVP Off and press the knob to enable or disable
OCP or OVP.
6. Turn the knob to Exit Menu and press the knob.
If OCP or OVP is enabled, the instrument will display a triangle above the OCP or OVP
annunciator.
Clear an OCP or OVP event
The instrument prominently displays an OCPTRIPPED or OVPTRIPPEDmessage to indicate an
overcurrent or overvoltage event. The triangle above OCP or OVP flashes continuouslyuntil the event is
cleared.
The output will be automatically turned OFF when OCP/OVP occurs.
1.
To begin, correct the condition that caused the OCP or OVP event. There are three ways to do this:
a.
Turn off OCP or OVP as described above.
b.
Set the current or voltage level to be less than the OCP or OVP level. You may be able to do this
by adjusting the voltage setting, the current setting, or the resistance of the load.
c.
Set the OCP or OVP limit value to be above the current or voltage level.
2. Press [Menu].
3. Turn the knob to OCP Set or OVP Set and press the knob.
4.
Turn the knob to the desired OCP or OVP limit value (if it needs to be changed). Then press the knob.
5. Turn the knob to OCP Clear or OVP Clear and press the knob.
Clear an overtemperature protection (OTP) event
The instrument prominently displays an OTPTRIPPED message to indicate an OTP event. In addition,
the triangle above ERR will flash continuously until the error event is cleared.
The output will be automatically turned OFF when OTP occurs.
The overtemperature event willbe logged in the Error list, and you must clear the OTP event before the
output can be turned ON again.
1.
To begin, correct the condition that caused the OTP event and allow the instrument to cool.
2. Press [Menu].
3. Turn the knob to Error and press the knob. The instrument will briefly display a message like 3
ERRORS (the number will vary) to indicate how many errors are in the queue.
4.
Turn the knob to cycle through the error codes, recording them as you go. Press the knob to view
the error description. Press [Menu] to exit and clear the error list.
An OTP event will occur again after it is cleared if the instrument remains overheated.
Lock and unlock the front panel
If you are in a menu, you must exit the menu before locking the front panel.
1. Press [Lock / Unlock] to lock the front panel. This produces an LCK annunciator triangle, as shown
below.
2.
If you press a key after the front panel is locked, the display willread HOLDKEY. That message indicates that you must press and hold the [Lock / Unlock] key to unlock the front panel and clear the
LCK triangle.
Save or recall the instrument's state
The instrument allows you to save and recallthe instrument state in one of 10 memory locations, labeled
0through9. This allows you to quickly configure the instrument for commonly used applications.
An instrument state includes the instrument's output enable status, voltage and current settings, and
OCP and OCV settings. It does not include error codes or LAN configuration information.
To save or recall the instrument state
1. Press [Menu].
2. Turn the knob to Store (or Recall) and press the knob.
3. Turn the knob to an option from Store 0 to Store 9 (or Recall 0 to Recall 9)and press the knob to
store or recall the present state of the instrument in the specified memory location.
4.
The instrument will briefly display DONE.
Configure the Power-on State
The instrument allows you to recall the power-on state as either the default (*RST) state or the state
stored in one of 10 memory locations (0through9). To configure the power-on state:
1. Press [Menu].
2. Turn the knob to Power-on and press the knob.
3.
Turn the knob to Default, or a state from Recall0 to Recall9 and press the knob to make your selection.
4.
The instrument will briefly display CHANGED SAVED if the recall state is different from the previous
state, and NOCHANGE if the same state was selected.
Read error codes
You only need to do this if there is a triangle over
shown below. Be sure to follow the procedure below carefully, and record each error code in
order. Once you exit this menu, there is no way to retrieve the error codes again.
in the lower left corner of the screen, as
ERR
1. Press [Menu].
2. Turn the knob to Error and press the knob. The instrument will briefly display a message like 3
ERRORS (the number will vary) to indicate how many errors are in the queue.
3.
Turn the knob to cycle through the error codes, recording them as you go.
Remote Control
You can control the instrument remotely in two ways. To program the instrument via SCPI, use Keysight
IO Libraries. To control the instrument via a simulated front panel, use the instrument's Web interface.
Keysight IOLibraries Suite
Keysight IOLibraries Suite is a collection free instrument control software that automatically discovers
instruments and allows you to control instruments over the LAN, USB, GPIB, RS-232, and other
interfaces. For more information, or to download IO Libraries, go to www.keysight.com/find/iosuite.
E36100 Series Web interface
You can monitor control the instrument from a Web browser by using the instrument's Web interface.
To connect, simply enter the instrument's IP address or hostname in your browser's address bar and hit
Enter.
If you see an error indicating 400: Bad Request, that is related to an issue with "cookies" in your
Web browser. To avoid this issue, either start the Web interface by using the IPaddress (not host
name) in the address bar), or clear cookies from your browser immediately before starting the
Web interface.
Check the checkbox below the picture of the instrument to enable an indicator on the instrument's
front panel. This is helpful if you have severalE36100 Series instruments and you wish to identify the one
to which you are connected.
The Configure LANtab on the top allows you to change the instrument's LAN parameters; exercise
caution when doing so, as you may interrupt your ability to communicate with the instrument.
When you click the Control Instrument tab, the instrument willask you for a password (default is
keysight, all lower case), and then it will open a new page, shown below.
This interface allows you to use the instrument just as you would from the front panel. Note the curved
arrow keys that allow you to "turn" the knob. You can click the knob to "push" it, just as you would push
any of the other keys on the front panel.
Read warning
Be sure to read and understand the warning at the top of the Control Instrument page.
Technical Connection Details
In most cases, you can easily connect to the instrument with the IO Libraries Suite or Web interface. In
certain circumstances, it may be helpful to know the following information.
The vendor ID: 0x2A8D, the product ID is 0x0902, and the instrument serial number is MY55160003.
The product ID varies by model: 0x0602 (E36102A), 0x0702 (E36103A), 0x0802 (E36104A), 0x0902 (E36105A),
0x0A02 (E36106A).
2SCPI Programming
Introduction to the SCPI Language
Error Messages
SCPI Status Registers
APPLy Subsystem
CALibration Subsystem
CURRent Subsystem
DISPlay Subsystem
IEEE-488 Subsystem
MEASure Subsystem
OUTPut Subsystem
STATus Subsystem
SYSTem Subsystem
Triggering Commands
VOLTage Subsystem
Introduction to the SCPI Language
Standard Commands for Programmable Instruments (SCPI) is an ASCII- based instrument command
language designed for test and measurement instruments. Refer to Simplified Programming Overview
for basic techniques for programming the power supply over the remote interface.
SCPI commands are based on a hierarchical structure, also known as a tree system. In this system,
associated commands are grouped together under a common node or root, thus forming subsystems.
A portion of the SOURce subsystem is shown below to illustrate the tree system.
[SOURce:]
CURRent {<current>|MIN|MAX|UP|DOWN} CURRent? [MIN|MAX]
CURRent:
TRIGgered {<current>|MIN|MAX} TRIGgered? {MIN|MAX}
VOLTage {<voltage>|MIN|MAX|UP|DOWN} VOLTage? [MIN|MAX]
VOLTage:
SOURce is the root keyword of the command, CURRent and VOLTage are second-level keywords, and
TRIGgered is the third- levelkeyword. A colon (:) separates a command keyword from a lower- level
keyword.
Command format used in this manual
The format used to show commands in this manual is shown below:
CURRent {<current>|MINimum|MAXimum|UP|DOWN}
The command syntax shows most commands (and some parameters) as a mixture of upper- case and
lower- case letters. The upper- case letters indicate the abbreviated spelling for the command. For
shorter program lines, send the abbreviated form. For better program readability, send the long form.
For example, in the above syntax statement, CURR and CURRENT are both acceptable forms. You can
use upper- case or lower- case letters. Therefore, CURRENT, curr, and Curr are all acceptable. Other
forms, such as CUR and CURREN, willgenerate an error.
Braces ({}) enclose the parameter choices for a given command string. The braces are not sent with the
command string.
A vertical bar (|) separates multiple parameter choices for a given command string.
Triangle brackets (<>) indicate that you must specify a value for the enclosed parameter. For example,
the above syntax statement shows the current parameter enclosed in triangle brackets. The brackets are
not sent with the command string. You must specify a value for the parameter (such as CURR 0.1).
Some parameters are enclosed in square brackets ([]). The brackets indicate that the parameter is
optional and can be omitted. The brackets are not sent with the command string. If you do not specify a
value for an optional parameter, the power supply chooses a default value.
A colon (:) separates a command keyword from a lower-level keyword. You must insert a blank space to
separate a parameter from a command keyword. If a command requires more than one parameter, you
must separate adjacent parameters using a comma as shown below:
SOURce:CURRent:TRIGgered
APPLy 3.5,1.5
Command separators
A colon (:) separates a command keyword from a lower- level keyword as shown below:
SOURce:CURRent:TRIGgered
A semicolon (;) is used to separate two commands within the same subsystem, and can also minimize
typing. For example, sending the following command string,
SOUR:VOLT MIN;CURR MAX
is the same as sending the following two commands:
SOUR:VOLT MIN
SOUR:CURR MAX
Use a colon and a semicolon to link commands from different subsystems. For example, in the following
command string, an error is generated if you do not use the colon and semicolon:
DISP:TEXT:CLE;:SOUR:CURR MIN
Using the MIN and MAX parameters
You can substitute MINimum or MAXimum in place of a parameter for many commands. For example,
consider the following command:
CURRent {<current>|MIN|MAX}
Instead of selecting a specific current, you can substitute MINimum to set the current to its minimum
value or MAXimum to set the current to its maximum value.
Querying parameter settings
You can query the value of most parameters by adding a question mark (?) to the command. For
example, the following command sets the output current to 5 A:
CURR 5
You can query the value by executing:
CURR?
You can also query the maximum or minimum value allowed with the present function as follows:
CURR? MAX CURR? MIN
If you send two query commands without reading the response from the first, and then attempt to
read the second response, you may receive some data from the first response followed by the complete second response. To avoid this, do not send a query command without reading the response.
When you cannot avoid this situation, send a device clear before sending the second query command.
SCPI command terminators
A command string sent to the power supply must terminate with a <new line> character. The IEEE- 488
EOI (end-or-identify) message is interpreted as a <new line> character and can be used to terminate a
command string in place of a <new line> character. A <carriage return> followed by a <new line> is also
accepted. Command string termination will always reset the current SCPI command path to the root
level. The <new line> character has the ASCII decimal code of 10.
IEEE-488.2 common commands
The IEEE- 488.2 standard defines a set of common commands that perform functions like reset, selftest, and status operations. Common commands always begin with an asterisk (*), are four to five
characters in length, and may include one or more parameters. The command keyword is separated
from the first parameter by a blank space. Use a semicolon (;) to separate multiple commands as shown
below:
*RST; *CLS; *ESE 32; *OPC?
SCPI parameter types
The SCPI language defines several different data formats to be used in program messages and response
messages.
Numeric parameters
Commands that require numeric parameters will accept all commonly used decimal representations of
numbers including optional signs, decimal points, and scientific notation. Special values for numeric
parameters like MINimum, MAXimum, and DEFault are also accepted.
You can also send engineering unit suffixes (V, A, or SEC) with numeric parameters. If only specific
numeric values are accepted, the power supply will automatically round the input numeric parameters.
The following command uses a numeric parameter:
CURR {<current>|MIN|MAX|UP|DOWN}
Discrete parameters
Discrete parameters are used to program settings that have a limited number of values such as BUS and
IMM. Query responses willalways return the short form in allupper- case letters. The following
command uses discrete parameters:
TRIG:SOUR {BUS|IMM}
Boolean parameters
Boolean parameters represent a single binary condition that is either true or false. For a false condition,
the power supply will accept OFF or 0. For a true condition, the power supply will accept ON or 1. When
you query a boolean setting, the power supply will always return 0 or 1. The following command uses a
boolean parameter:
DISP {OFF|ON}
String parameters
String parameters can contain virtually any set of ASCII characters. A string must begin and end with
matching quotes; either with a single quote or with a double quote. You can include the quote delimiter
as part of the string by typing it twice without any characters in between. The following command uses a
string parameter:
DISP:TEXT <quoted string>
Error Messages
The instrument returns error messages in accordance with the SCPI standard.
— Up to 20 errors can be stored in the instrument's error queue, and the triangle annunciator above
ERR turns on when one or more errors are in the error queue.
— Error retrieval is first-in-first-out (FIFO), and errors are cleared as you read them. When you have read
all errors from the error queue, the ERR annunciator turns off.
— If more than 20 errors have occurred, the last error stored in the queue (the most recent error) is
replaced with -350,"Queue overflow". No additional errors are stored until you remove errors from the
queue. If no errors have occurred when you read the error queue, the instrument responds with
+0,"No error".
— Send SYSTem:ERRor? to read the most recent error. Each error is in the format: -104,"Data type
error".
—
To read the error queue from the front panel, press [Menu], then turn the knob to Error and press the
knob to select it. Then rotate the knob to view the error codes. Be sure to record the error messages,
because they will be cleared when you exit the error list.
— The error queue is cleared by power cycles and *CLS., but not *RST.
Error Codes
The instrument's error codes are listed below:
-440,"Query UNTERMINATED after indefinite response"
-430,"Query DEADLOCKED"
-420,"Query UNTERMINATED"
-410,"Query INTERRUPTED"
-400,"Query error"
-363,"Input buffer overrun"
-350,"Queue overflow"
-330,"Self-test failed"
-310,"System error"
-277,"Macro redefinition not allowed"
-276,"Macro recursion error"
-273,"Illegal macro label"
-272,"Macro execution error"
-270,"Macro error"
-241,"Hardware missing"
-230,"Data corrupt or stale"
-225,"Out of memory"
-224,"Illegal parameter value"
-223,"Too much data"
-222,"Data out of range"
-221,"Settings conflict"
-213,"Init ignored"
-211,"Trigger ignored"
-200,"Execution error"
-183,"Invalid inside macro definition"
-181,"Invalid outside macro definition"
-178,"Expression data not allowed"
-171,"Invalid expression"
-170,"Expression error"
-168,"Block data not allowed"
-161,"Invalid block data"
-158,"String data not allowed"
-151,"Invalid string data"
-150,"String data error"
-148,"Character data not allowed"
-141,"Invalid character data"
-138,"Suffix not allowed"
-134,"Suffix too long"
-131,"Invalid suffix"
-128,"Numeric data not allowed"
-124,"Too many digits"
-123,"Exponent too large"
-121,"Invalid character in number"
-114,"Header suffix out of range"
-113,"Undefined header"
-112,"Program mnemonic too long"
-109,"Missing parameter"
-108,"Parameter not allowed"
-105,"GET not allowed"
-104,"Data type error"
-103,"Invalid separator"
-102,"Syntax error"
-101,"Invalid character"
-100,"Command error"
514,"LAN config error"
561,"Analog board - failed to save to EEPROM"
564,"Analog board - failed to load from EEPROM"
565,"Analog board - over temperature"
566,"Analog board - command timed out"
601,"Front panel does not respond"
609,"System ADC test failed"
610,"I/O board not plugged in"
611,"Unsupported I/O board"
612,"Analog board does not respond"
613,"Analog bias output ±15V test failed"
614,"EEPROM test failed"
615,"EEPROM save failed"
616,"Model no mismatched"
630,"Fan test failed"
631,"System DAC test failed"
701,"Cal security disabled by jumper"
702,"Invalid state. Cal secured"
703,"Invalid secure code"
704,"Secure code too long"
708,"Cal output disabled"
717,"Cal OVP or OCP status enabled"
721,"Failed to calibrate voltage DAC"
722,"Failed to calibrate voltage ADC"
723,"Failed to calibrate OVP"
724,"Failed to calibrate current DAC"
725,"Failed to calibrate current ADC"
726,"Failed to calibrate OCP"
727,"Invalid Calibration sequence"
728,"Calibration failed"
SCPI Status Registers
All SCPI instruments implement status registers in the same way. The status system records various
instrument conditions in three register groups: the Status Byte register, the Standard Event register,
and the Questionable Status register groups. The Status Byte register records high-level summary
information reported in the other register groups.
What is an event register?
An event register is a read-only register that reports defined conditions within the instrument. Bits in an
event register are latched. Once an event bit is set, subsequent state changes are ignored. Bits in an
event register are automatically cleared by a query of that register (such as *ESR? or STAT:QUES:EVEN?)
or by sending the *CLS (clear status) command. A reset (*RST) or device clear willnot clear bits in event
registers. Querying an event register returns a decimal value corresponding to the binary-weighted sum
of all bits set in the register.
What is an enable register?
An enable register defines which bits in the corresponding event register are logically ORed together to
form a single summary bit. Enable registers are both readable and writable. Querying an enable register
will not clear it. The *CLS (clear status) command does not clear enable registers but it does clear the
bits in the event registers. To enable bits in an enable register, you must write a decimal value which
corresponds to the binary- weighted sum of the bits you wish to enable in the register.
Standard Event Status Enable Register
Bit Value NameDescription
01OPCOperation complete
12(unused) (reserved for future use)
24QYEQuery error
38DDEDevice dependent error
416EXEExecution error
532CMECommand error
664(unused) (reserved for future use)
7128PONPower-on
Operation Status Register
Bit Value NameDescription
01CALThe output is computing new calibration constants
1-4 2-16 (unused) (reserved for future use)
532WTGThe output is waiting for a trigger.
6-7 64-128 (unused) (reserved for future use)
8256CVThe output is in constant voltage mode.
9512 (unused) (reserved for future use)
10 1024CCThe output is in constant current mode.
Questionable Register
Bit Value NameDescription
01OVOutput disabled by overvoltage protection
12OCOutput disabled by overcurrent protection
2-3 4-8(unused) (reserved for future use)
416OTOutput disabled by overtemperature protection
Specifies and changes the output voltage and current in one command.
– The DEFault, MINimum, and MAXimum values for each mode are shown in Programming Ranges.
– The query returns a quoted string with the voltage setting before the current setting.
– Both values returned by the query have five decimal points: "25.00000,0.75000"
CALibration Subsystem
For a detailed discussion on the calibration procedures, see Calibration Adjustment Procedures.
CALibration:COUNt?
Returns the number of times the instrument has saved calibration data as a signed whole number, for
example +21. Your instrument was calibrated at the factory; read and record the initial count when you
receive the instrument.
CALibration:CURRent[:DATA][:HIGH] <value>
Enters an output current value (in amps) that you obtained by reading a DMM. You must first select a
calibration level (CAL:CURR:LEV) for the value being entered. This command can only be used when
calibration is unsecured and output is ON.
CALibration:CURRent[:DATA]:LOW <value>
Enters an output current value (in amps) that you obtained by reading a DMM. This is similar to
CALibration:CURRent[:DATA][:HIGH], but it is only used for small current calibration. This command can
only be used when calibration is unsecured and output is ON.
CALibration:CURRent:LEVel[:HIGH] MINimum|MAXimum
Initiates calibration for a specific current level. The MINimumcalibration must be performed before the
MAXimum, and you must unsecure this instrument to use this command.
CALibration:CURRent:LEVel:LOW MINimum|MAXimum
Initiates small current calibration for a specific current level. The MINimumcalibration must be
performed before the MAXimum, and you must unsecure this instrument to use this command.
Unsecures or secures the power supply with the calibration security code Of up to nine digits. The query
returns 1 (ON - unsecured for calibration) or 0 (OFF - secured against calibration).
– The query only returns the state; it does not return the passcode.
– The security code is nonvolatile, and does is not changed by power cycling or *RST.
– The front panel limits the user to entering codes of up to six digits.
– The default passcodes are 36102, 36103, 36104, 36105, and 36106.
– Example: Unlock the instrument to enable calibration: CAL:STATe 1,36102
CALibration:STRing "<string>"
CALibration:STRing?
Record or read up to 40 characters of instrument calibration information.
– You can store any information related to calibration, such as date or contact information.
– The power supply must be unsecured before sending a calibration message.
– You can read the string regardless of whether the supply is secured.
CALibration:VOLTage[:DATA] <value>
Enters an output voltage value (in volts) that you obtained by reading a DMM.
– You must first select a calibration level (CAL:CURR:LEV) for the value being entered.
– This command can only be used when calibration is unsecured and output is ON.
CALibration:VOLTage:LEVel MINimum|MAXimum
Initiates calibration for a specific voltage level. You must do the MINimum calibration before the
MAXimum, and this command can only be used when calibration is unsecured.
CURRent Subsystem
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <current> | MINimum |
MAXimum |UP | DOWN
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? [MINimum | MAXimum]
Programs the immediate current level of the instrument's output.See Programming Ranges for details.
This command changes the output immediately. The UP and DOWN parameters increase or decrease
the immediate current by the amount specified by CURRent:STEP. A setting that exceeds the maximum
or minimum rated current will cause error -222 (Data out of range).
Sets the time (in milliseconds) that the overcurrent protection is temporarily disabled after a current
level change. The query returns a number of the form +#.########E+##.
Enables or disables overcurrent protection, which causes the instrument to go into a protected state
when the power supply status is in constant current mode. The query returns 1 (ON)or 0 (OFF).
[SOURce:]CURRent:PROTection:TRIPped?
Indicates whether an overcurrent protection occurred (1) or not (0). This is reset to 0 by
CURRent:PROTection:CLEar.
Displays a message of up to 12 characters on the front panel. Additional characters are truncated.
DISPlay[:WINDow][:STATe] ON | 1 | OFF | 0
DISPlay[:WINDow][:STATe]?
Turns the display off or on. When the display is turned off, outputs are not sent to the display and all
annunciators except ERROR are disabled. The display state is automatically turned on when you return
to the local mode. Press and hold [Lock/Unlock] for a few seconds to return to the local mode.
IEEE-488 Subsystem
*CLS
Clears all event registers, the Status Byte register, and the error queue.
*ESE <enable value>
*ESE?
Enables bits in the Standard Event Enable register. The selected bits are then reported to the Status
Byte. The query returns a decimal value corresponding to the binary-weighted sum of all bits in the
register. All of the enabled events of the Standard Event Status Event Register are logically ORed to
cause the Event Summary Bit (ESB) of the Status Byte Register to be set.
Standard Event Status Enable Register
Bit Value Name Description
01OPC Operation complete
120(reserved for future use)
24QYE Query error
38DDE Device-dependent error
416EXE Execution error
532CME Command error
6640(reserved for future use)
7128PON Power-on has occurred
*ESR?
Returns a decimal value corresponding to the binary-weighted sum of all bits in the Standard Event
register and clears the register to 0. The bit configuration is the same as the Standard Event Status
Enable register (see *ESE).
*IDN?
Returns the power supply’s identification string in the form shown below.
The four comma-separated fields are the manufacturer name, instrument model number, instrument
serial number, and revision code.
*OPC
*OPC?
Sets the “Operation Complete” bit (bit 0) of the Standard Event register after the command is executed.
The query returns 1 to the output buffer after the command is executed.
*OPT?
Returns the option number of the instrument.
*PSC 0|1
*PSC?
(Power- on status clear) Enables (1)or disables (0) the clearing of the Status Byte and the Standard
Event register enable masks when power is turned on.
*RST
Resets the instrument to its default state, but does not clear any of the status registers or the error
queue. It also does not affect any interface error conditions.
Parameter Default Value
OCP delay50 ms
OVP levelMaximum (varies by model)
Voltage0
CurrentMaximum (varies by model)
OCP stateOff
OVP stateOff
Output state Off
Relay sense Internal
*RCL <state>
*SAV <state>
Saves (*SAV) the current instrument state or recalls (*RCL) a previously saved state file. All state storage
locations (0through9) are nonvolatile.
– The parameters stored in a state file are the same parameters affected by *RST.
– States are not affected by the *RSTcommand.
– Saving a state overwrites the previous state (if any) stored in that location.
– When shipped from the factory, storage locations 0through9 are empty.
*SRE <enable value>
*SRE?
Enables bits in the Status Byte enable register. The query returns the decimal value corresponding to the
binary-weighted sum of allbits set in the register.
*STB?
Queries the Status Byte summary register. This is similar to a serial poll but it is processed like any other
instrument command. The *STB? command returns the same result as a serial poll but the “Request
Service” bit (bit 6) is not cleared if a serial poll has occurred.
*TRG
Generates an event trigger to the trigger system when the trigger system has a BUS (software) trigger as
its trigger source (TRIG:SOUR BUS). If the trigger system is not initiated, the *TRG command is simply
ignored.
*TST?
Performs a complete self-test of the power supply and returns “0” (pass) or a non-zero value (fail). If the
self-test fails, an error message is also generated with additional information on why the test failed.
The bits added together to produce a non-zero value in the case of a failing self-test are shown below.
Bit Value NameDescription
01EEPROMstatus
12Voltage monitor for +15 V
24Model number detection failure
38(unused) (reserved for future use)
416(unused) (reserved for future use)
532ADC failure
664DAC failure
7128 (unused) (reserved for future use)
*WAI
Waits for all pending operations to complete before executing any additional remote interface
commands. This command is used only in the triggered mode to wait for a pending delayed trigger.
MEASure Subsystem
MEASure:CURRent[:DC]?
Returns the sensed DCoutput current in amps in the format 1.23456789E+00.
MEASure[:VOLTage][:DC]?
Returns the sensed DCoutput voltage in volts in the format 1.23456789E+00. Use the
[SOURce:]VOLTage:SENSe[:SOURce] command to specify whether the voltage uses internal or external
(remote) sensing.
OUTPut Subsystem
OUTPut[:STATe] ON |1 | OFF |0
OUTPut[:STATe]?
Enables or disables the instrument's output.The query returns 0 (OFF) or 1 (ON). At *RST, the output
state is off.
OUTPut:PROTection:CLEar
Clears the latch that disables the output due to an overvoltage or overcurrent condition. You must clear
the conditions that cause the fault before executing this command. You can then restore the output to
the state that existed before the fault condition occurred.
Specifies whether the instrument's power-on state is the *RST state (default) or the state stored in one
of ten memory locations.
STATus Subsystem
STATus:OPERation:[EVENT]?
Returns a decimal value corresponding to the binary-weighted sum of the bits in the event register of
the Operation Status group. This read-only register stores (latches) all events that are passed by the
Operation NTR and/or PTR filter.
– Reading the Operation Status Event register clears it.
– *RST has no effect on this register.
STATus:OPERation:CONDition?
Returns a decimal value corresponding to the binary-weighted sum of the bits in the condition register
of the Operation Status group. This read-only register holds the live (unlatched) operational status of
the instrument.
– Reading the Operation Status Condition register does not clear it.
– *RST clears this register, other than those bits where the condition still exists after *RST.
STATus:OPERation:ENABle <value>
Sets the value of the enable register for the Operation Status group. The enable register is a mask for
enabling specific bits from the Operation Event register to set the OPER (operation summary) bit of the
Status Byte register.
– STATus:PRESet clears all bits in the enable register.
– *CLS does not clear the enable register, but does clear the event register .
STATus:PRESet
Sets the all registers to their power-on values.
STATus:QUEStionable:CONDition?
Returns a decimal value corresponding to the binary-weighted sum of the bits in the Questionable
Status condition register.
– This register is a read-only register, which holds the instrument's live (unlatched) operational status.
– Reading the Questionable Status Condition register does not clear it.
– *RST clears this register, other than those bits where the condition still exists after *RST
Enables bits in the Questionable Status enable register, which is a mask for enabling specific bits from
the Operation Event register to set the QUES (questionable summary) bit of the Status Byte register. The
selected bits are then reported to the Status Byte.
– STATus:PRESet clears all bits in the enable register.
– *CLS does not clear the enable register, but does clear the event register .
STATus:QUEStionable[:EVENt]?
Returns a decimal value corresponding to the binary-weighted sum of all bits in the Questionable Status
event register.
– These bits are latched.
– Reading the event register clears it, but *RST has no effect on this register.
SYSTem Subsystem
SYSTem:COMMunicate:RLSTate LOCal |REMote | RWLock
SYSTem:COMMunicate:RLSTate?
Places the instrument in remote or local mode. The LOCal parameter is the same as SYSTem:LOCal, the
REMote parameter is the same as SYSTem:REMote, and the RWLock parameter is the same as
SYSTemRWLock. The query returns LOC, REM, or RWL.
SYSTem:ERRor[:NEXT]?
Queries the power supply's error queue. When the front-panel ERR annunciator turns on, one or more
errors have been detected. Up to 20 errors can be stored in the error queue. See Error Messages.
— Errors are retrieved in first-in-first-out (FIFO) order. The ERR annunciator turns off after the last error
is read.
— If more than 20 errors have occurred, the last error stored in the queue (the most recent error) is
replaced with -350, “Queue overflow”. No additional errors are stored until you remove errors from
the queue. If no errors have occurred when you read the error queue, the power supply responds with
+0, “No error”.
— The error queue is cleared when power has been off or after a *CLS (clear status) command has been
executed. The *RST (reset) command does not clear the error queue.
SYSTem:LOCal
Places the power supply in the local mode. All keys on the front panel are fully functional.
SYSTem:REMote
Places the power supply into remote mode for remote operation. When the unit is being controlled
remotely, the power supply will go to remote mode automatically. You can switch the instrument to
local mode using SYSTem:LOCal or by pressing any key on the front panel.
SYSTem:RWLock
Places the power supply in the remote mode. This command is the same as SYSTem:REMote, except
that all front-panel keys, excluding the [Lock/Unlock] key are disabled. You can unlock the front panel
by holding the [Lock/Unlock] key. This switches the instrument to local mode at the same time. The
front panel willbe unlocked when the instrument is switched to remote or local mode.
SYSTem:SECurity:IMMediate
Clears and sanitizes all user memory (states, LAN information, and so on), usuallyto prepare the
instrument for removal from a secure area. Instrument identification data (instrument firmware, model
number, serial number, MAC address and calibration data) is not erased. This procedure is not
recommended for routine use because of the possibility of unintended data loss.
SYSTem:VERSion?
Returns the SCPI version on which the instrument's command set is based, which is 2005.0.
Triggering Commands
ABORt
Clears any pending delayed trigger and returns the trigger system to idle. If INIT:CONT is enabled, the
trigger system is continuously initiated.
INITiate[:IMMediate]
Cause the trigger system to initiate. When the trigger system is initiated, an event on the specified
trigger source causes the corresponding trigger action on the power supply output. This command
completes one full trigger cycle when the trigger source is IMMediate or initiates the trigger subsystem
when the trigger source is BUS.
INITiate:CONTinuous ON | 1 | OFF | 0
INITiate:CONTinuous?
Enables or disables continuous initiation of the trigger system. If this is disabled, an INIT initiates the
trigger system for only one trigger action. If this is enabled, the trigger system is continuously initiated
and INIT is redundant. The query returns 1 (ON) or 0 (OFF).
*TRG
Generates an event trigger to the trigger system when the trigger system has a BUS (software) trigger as
its trigger source (TRIG:SOUR BUS). If the trigger system is not initiated, the *TRG command is simply
ignored.
TRIGger[:SEQuence]:DELay <seconds> MINimum | MAXimum
TRIGger[:SEQuence]:DELay? [MINimum | MAXimum]
Sets the time delay (in seconds, from 0 to 32.767)from the detection of a bus trigger to the start of any
corresponding trigger action on the power supply output. The default is 0, and the query returns a
number of the form +#.########E+##.
TRIGger[:SEQuence]:SOURce BUS | IMMediate
TRIGger[:SEQuence]:SOURce?
Specifies the source from which the instrument will accept a trigger, either a bus (software) trigger or an
internal immediate trigger. At *RST, the bus trigger source is selected. To ensure synchronization when
the bus source is selected, send the *WAI (wait) command. You can also use the *OPC? (operation
complete) query or the *OPC command to indicate when the operation has completed. The query
returns BUS (wait for trigger command) or IMM (continuous triggering.
Enables or disables overvoltage protection, which causes the instrument to go into a protected state
when the power supply goes above the protection level specified by the VOLTage:PROTection[:LEVel].
The query returns 1 (ON)or 0 (OFF).
[SOURce:]VOLTage:PROTection:TRIPped?
Indicates whether an overvoltage protection occurred (1) or not (0). This is reset to 0 by
VOLTage:PROTection:CLEar.
[SOURce:]VOLTage:PROTection[:LEVel] <voltage>|MINimum | MAXimum
[SOURce:]VOLTage:PROTection[:LEVel]? [MINimum|MAXimum]
Sets the level at which overvoltage protection trips, in volts. The query returns a number of the form
+#.########E+##.
Specifies whether the instrument uses remote or local sensing. The query returns 0 (INT)or 1 (EXT).
The internal setting sets a relaywithin the power supply to connect the output and sensing connector.
This means that only two wires are used; and remote sensing is disabled. The external setting sets a relay
within the power supply to disconnect the output and remote sensing inputs. This means that four wires
are used; and remote sensing is enabled.. The Internal setting displays 2w in the upper left corner of the
display, and the External setting shows 4w in the upper left corner.
3Service and Support
Service and Repair
Performance Verification
Test Record Forms
Calibration Adjustment Procedures
Specifications and Typical Characteristics
Service and Repair
Types of Service Available
If your instrument fails during the warranty period, Keysight Technologies will repair or replace it under
the terms of your warranty. After your warranty expires, Keysight offers repair services at competitive
prices. You also have the option to purchase a service contract that extends the coverage after the
standard warranty expires.
Obtaining Repair Service (Worldwide)
To obtain service for your instrument, contact your nearest Keysight Technologies Service Center. They
will arrange to have your unit repaired or replaced, and can provide warranty or repair–cost information
where applicable. Ask the Keysight Technologies Service Center for shipping instructions, including
what components to ship. Keysight recommends that you retain the original shipping carton for return
shipments.
Repackaging for Shipment
Ensure the following to ship the unit to Keysight for service or repair:
– Attach a tag to the unit identifying the owner and indicating the required service or repair. Include the
model number and full serial number.
– Place the unit in its original container with appropriate packaging material.
– Secure the container with strong tape or metal bands.
– If the original shipping container is unavailable, use a container that will ensure at least 10 cm (4 in.) of
compressible packaging material around the entire instrument. Use static-free packaging materials.
Keysight suggests that you always insure your shipments.
Cleaning and Handling
Cleaning
To prevent electrical shock, disconnect the instrument from AC mains power and disconnect all test
leads before cleaning. Clean the outside of the instrument using a soft, lint-free, cloth slightly
dampened with water.
– Do not use detergent or solvents.
– Do not attempt to clean internally.
If required, contact a Keysight Technologies Sales and Service office to arrange for proper cleaning to
ensure that safety features and performance are maintained.
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling.
Component damage can occur at electrostatic discharge voltages as low as 50 V.
The following guidelines will help prevent ESD damage during service operations:
– Disassemble instruments only in a static-free work area.
– Use a conductive work area to reduce static charges.
– Use a conductive wrist strap to reduce static charge accumulation.
– Minimize handling.
– Keep replacement parts in original static-free packaging.
– Remove all plastic, foam, vinyl, paper, and other static-generating materials from the immediate work
area.
Troubleshooting
Before troubleshooting or repairing the instrument, make sure the failure is in the instrument rather
than any external connections. Also make sure that the instrument was accurately calibrated within the
last year (see Calibration Adjustment Procedures > Calibration Interval for details).
Perform the following verifications if the unit is inoperative:
– Verify that the ac power cord is connected to the power supply.
– Verify that the front-panel power switch is depressed.
– Verify that the power-line fuse is installed:
– Use the 2.0 AT, 250 V fuse for 100 or 115 Vac operation.
– Use the 1.0 AT, 250 V fuse for 230 Vac operation.
– Verify the power-line voltage setting.
– See Getting Started > Options and Fuse Information for details.
Self-Test Procedures
A power-on self-test occurs automaticallywhen you turn on the power supply. This limited test assures
you that the power supply is operational.
The complete self-test of the power supply is enabled by pressing the Recall key or any front panel keys
except the Error key and the power-line switch simultaneously . Then continue to press the Recall key
for 5 seconds. It will take another 2 seconds for the complete self-test to end.
You can also perform a complete self-test from the remote interface see SCPI Programming for details.
– If the self-test is successful, "PASS" is displayed on the front panel.
– If the self-test fails, "FAIL" is displayed and the ERROR annunciator turns on. Record the error code
and message and contact Keysight support if necessary.
– If the self-test is successful, this indicates a high chance that the power supply is operational.
To Replace the Power-Line Fuse
The power-line fuse is located within the power supply's fuse-holder assembly on the rear panel see
Getting Started > Options and Fuse Information for details. For 100 or 115 Vac operation, you must use
a 2.5 AT slow-blow fuse. For 230 Vac operation, you must use a 2 AT slow-blow fuse.
User Replaceable Parts
You can find the instrument support part list at Keysight's Test & Measurement Parts Catalog
http://www.keysight.com/find/parts.
Performance Verification
Performance verification ensures that the instrument performs within the specifications stated in the
data sheet. Some specifications are included in this document and were accurate at the time of
publication, but the latest data sheet from the Keysight website should be referred to see if there are any
changes since this document was published.
Recommended test equipment
The test equipments recommended for the performance verification and adjustment procedures are
listed below. If the exact instrument is not available, use the accuracy requirements shown to select
substitute calibration standards.
TypeSpecificationRecommended Model
Digital MultimeterReadout: 6 1/2 digits Basic DC Accuracy: 0.0035%Keysight 34401A or equivalent
Current Monitor15 A (0.1 Ω), TC = 4 ppm/°CGuildline 9230A-15R
Electronic Load150 V, 5 A minimum, with transient capability and a slew
rate of 833 kA/s or better.
LAN / USB ControllerPC with Keysight Connection Expert
RMS VoltmeterTrue RMS; Bandwidth: 20 MHz Sensitivity: 100 μVRhode and Schwartz Model URE3
Differential AmplifierBandwidth: 20 MHzLeCroy 1855A or equivalent
Terminations1 – 50 Ω BNC termination
2 – 50 Ω, 1/8 W termination resistors
Variable voltage transformer
or AC source
Calibration resistors for small
current
Adjustable to highest rated input voltage range. Power:
500 VA
300 Ω (E36102A) 1/8 W minimum
2.5 kΩ (E36103A) 1/8 W minimum
9 kΩ (E36104A) 1/8 W minimum
20 kΩ (E36105A) 1/8 W minimum
51 kΩ (E36106A) 1/8 W minimum
Keysight N3300A mainframe, with
N330xA modules
loaded
Keysight Infiniium/6054A or
equivalent
or equivalent
Keysight 6813B or equivalent
Test considerations
– Ensure that the calibration ambient temperature is stable and between 20 °C and 30 °C.
– Ensure ambient relative humidity is less than 80%.
– Allow a 1-hour warm-up period before verification or calibration.
– Keep cables as short as possible, consistent with the impedance requirements.
The tests should be performed by qualified personnel. During performance verification tests, hazardous voltages may be present at the outputs of the power supply.
Measurement techniques
Voltmeter
To ensure that the values read by the voltmeter during both the verification procedure and the
calibration procedure are not affected by the instantaneous measurement of the AC peaks of the output
current ripple, make severalDC measurements and average them.
Current-monitoring resistor
The 4-terminal current shunt is used to eliminate output current measurement error caused by voltage
drops in the load leads and connections. It has special current-monitoring terminals inside the load
connection terminals. Connect the voltmeter directly to these current-monitoring terminals.
Electronic load
Many of the test procedures require the use of a variable load capable of dissipating the required power.
If a variable resistor is used, switches should be used to connect, disconnect, or short the load resistor.
For most tests, an electronic load can be used. The electronic load is considerably easier to use than
load resistors, but it may not be fast enough to test transient recovery time and may be too noisy for the
noise (PARD) tests.
Fixed load resistors may be used in place of a variable load, with minor changes to the test procedures.
Also, if computer controlled test setups are used, the relatively slow (compared to computers and
system voltmeters) settling times and slew rates of the power system may have to be taken into account.
"Wait" statements can be used in the test program if the test system is faster than the power system.
Setup for most tests
This setup is used for most tests and it requires the DMM, electronic load, and power supply being
verified. Some wire is also required for connection between instruments. A LAN or USB cable is needed
for readback data. The DMM measures the power supply output, and the electronic load draws current
from the power supply. The accuracy of the current monitoring resistor must be 0.01% or better, which
should include any self-heating effects.
Constant Voltage (CV) verification
Voltage programming and readback accuracy
These tests verify that the voltage programming and the LAN or USB readback functions are within
specifications. Note that the readback values over the remote interface should be identical to those
displayed on the front panel, but with maximum resolution.
1.
Turn off the power supply using the AC line switch.
2.
Connect a DMM between the (+) and (–) terminals of the output.
3.
If you are using a computer to control the power supply, connect a LAN or USB cable from the
power supply to the PC.
4.
Turn on the power supply using the AC line switch.
Voltage programming accuracy
StepFront PanelSCPI
5.Set the instrument settings as described in the test record form under the Voltage
Programming and Readback
6.Enable the output by pressing the Output ON key.OUTP ON
. See
Test Record Forms
for details.
VOLT 6; CURR 5 (for example for
the E36102A, 6 V, 5 A output)
7. The output status should be CV and the output current should be close to zero.
8.
Record the voltage measured by the DMM and verify whether it is within the limits calculated.
Voltage readback accuracy
StepFront PanelSCPI
9.MEAS:VOLT? (for example for the
E36102A, 6 V, 5 A output)
10.
Record the voltage returned by the SCPI command query via Keysight Connection Expert, and verify
whether it is within the limits calculated.
CV load and line regulation
These tests verify that the voltage variation due to load or line variation are within specifications.
CV load regulation
This test measures the change in output voltage resulting from a change in output current from full load
to no load.
1.
Turn off the power supply using the AC line switch.
2.
Connect the power supply output with a DMM and an electronic load.
3.
If you are using a PC to control the power supply, connect a LAN or USB cable from the power supply to the PC.
4.
Turn on the power supply using the AC line switch.
5.
Set the power supply settings as described in the test record form under “CV Load Regulation”. See
the Test Record Forms under “CV Load Regulation” for details. Enable the output.
6.
Operate the electronic load in constant current mode and set its current to the value in the test
record form under “CV Load Regulation”. See the Test Record Forms under Test Description of “CV
Load Regulation” for details. Check that the front panel CV annunciator of the power supply remains
lit. If it turns to CC or UNREG, adjust the load so that the output current drops slightly until the CV
annunciator lights up. Record the output voltage reading on the DMM as V
7.
Operate the electronic load in open mode (input off). Record the output voltage reading on the
DMM immediately as V
8.
Take the difference between the DMM readings in steps 6 and 7 that is the CV load regulation (V
– V
noload
cification limits.
). The difference of the readings during the immediate change should be within the spe-
noload
.
load
.
load
CV line regulation
This test measures the change in output voltage that results from a change in AC line voltage from the
minimum to maximum value within the line voltage specifications.
1.
Turn off the power supply using the AC line switch.
2.
Connect the power supply output with a DMM and an electronic load.
3.
Connect a variable AC Source or Variac to the AC input, set to an appropriate line voltage for the
power supply configuration.
4.
If you are using a PC to control the power supply, connect a LAN or USB cable from the power supply to the PC.
5.
Turn on the power supply using the AC line switch.
6.
Set the power supply settings as described in the test record form under “CV Line Regulation”. See
the Test Record Forms under “CV Line Regulation” for details. Enable the output.
7.
Operate the electronic load in constant current mode and set its current to the value in the test
record form under “CV Line Regulation”. See the Test Record Forms under Test Description of “CV
Line Regulation” for details. Check that the front panel CV annunciator of the power supply remains
lit. If it turns to CC or UNREG, adjust the load so that the output current drops slightly until the CV
annunciator lights up.
8.
Adjust the AC power source to low line voltage limit (90 VAC for nominal 100 VAC, 104 VAC for nominal 115 VAC, 194 VAC for nominal 215 VAC, or 207 VAC for nominal 230 VAC). Record the output
reading on the DMM as V
9.
Adjust the AC power source to high line voltage (110 VAC for nominal 100 VAC, 127 VAC for nominal
115 VAC, 237 VAC for nominal 215 VAC or 253 VAC for nominal 230 VAC). Record the voltage reading on the DMM immediately as V
10.
Take the difference between the DMM readings in steps 8 and 9 that is the CV line regulation
(V
lowline
the limit calculated from the specification.
– V
highline
lowline
). The difference of the readings during the immediate change should be within
.
highline
.
Transient response verification
This test measures the time for the output voltage to recover to within the specified value following a
50% change in the load current.
1.
Turn off the power supply using the AC line switch.
2.
Connect an oscilloscope and electronic load between the (+) and (–) terminals of the output to be
tested as shown below.
3.
If you are using a PC to control the power supply, connect a LAN or USB cable from the power supply to the PC.
4.
Turn on the power supply using the AC line switch.
5.
Set the instrument settings as described in the test record form under “Transient Response”. See
the Test Record Forms under “Transient Response” for details. Enable the output.
6.
Operate the electronic load in constant current mode and set its current to the value in the test
record form under “Transient Response”. See the Test Record Forms under Test Description of “Tran-
sient Response” for details. Set the transient levelto ½ the maximum current. Set the transient duty
cycle to 50% and transient frequency to 1 kHz. Check that the front panel CV annunciator of the
power supply remains lit. If it turns to CC or UNREG, adjust the maximum current load so that the
output current drops slightly until the CV annunciator lights up.
7.
Adjust the oscilloscope to displaytransients as shown below. Note that the pulse width (t2-t1) of the
transient at the voltage settling band, for example 15 mV for the E36102 from the base line is no
more than 50 ms.
The oscilloscope cursors X1 and X2 represent t1 and t2.
8.
Transient response specification is met when the voltage recovers within 50 μs.
Output noise verification
Periodic and random output deviations superimpose a residual AC voltage on the DC output. This
residual voltage is specified as the rms or peak-to-peak noise in and is specified in the product data
sheet.
1.
Turn off the power supply using the AC line switch.
2.
Connect a load resistor or electronic load, differential amplifier, and an oscilloscope (AC coupled) to
the output as shown below.
3.
Use an appropriate load resistor (see load resistor value Recommended test equipment list) to keep
the power system at the instrument setting specified in the test record form under “CV Ripple and
Noise”. See the Test Record Forms under “CV Ripple and Noise” for details.
4.
As shown in the figure, use two BNC cables to connect the differential amplifier to the (+) and (−) output terminals. Each cable should be terminated by a 50 Ω resistor. The shields of the two BNC cables
should be connected together. Connect the differential amplifier output to the oscilloscope with a
50 Ω termination at the oscilloscope input.
5.
Set the differential amplifier to multiply by ten, divide by one, and 1 MΩ input resistance. Set the differential amplifier's positive and negative inputs to AC coupling. Set the oscilloscope’s time base to 5
ms/div, and the vertical scale to 10 mV/div. Turn the bandwidth limit on (usually 20 or 30 MHz), and
set the sampling mode to peak detect.
6.
Program the power supply to the settings indicated in the in the test record form for the appropriate
model under “CV Ripple and Noise” see the Test Record Forms under Test Description of “CV Ripple
and Noise” for details and enable the output. Let the oscilloscope run for a few seconds to generate
enough measurement points. On the Keysight Infiniium oscilloscope, the maximum peak-to-peak
voltage measurement is indicated at the bottom of the screen on the right side. Divide this value by
10 to get the CV peak-to-peak noise measurement. The result should not exceed the peak-to-peak
upper limits for instrument's “CV Ripple and Noise, peak-to-peak.” See the Test Record Forms
under “CV Ripple and Noise” for details.
7.
Disconnect the oscilloscope and connect an rms voltmeter in its place. Do not disconnect the 50 Ω
termination. Divide the reading of the rms voltmeter by 10. The result should not exceed the rms limits in the test record form for the appropriate model under “CV Ripple and Noise, rms”. See the Test
Record Forms under “CV Ripple and Noise” for details.
Constant Current (CC) verification
Current programming and readback accuracy
These tests verify that the current programming and the LAN or USB readback measurement functions
are within specifications. Note that the readback values over the remote interface should be identical to
those displayed on the front panel, but with maximum resolution.
1.
Turn off the power supply using the AC line switch.
2.
Connect the current shunt directly across the output terminals. Connect the DMM directly across
the current shunt.
3.
If you are using a PC to control the power supply, connect a LAN or USB cable from the power supply to the PC.
4.
Turn on the power supply using the AC line switch.
Current programming accuracy
StepFront PanelSCPI
5.Set the instrument settings as described in the test record form under Current Programming and Readback
6.Enable the output by pressing the Output ON keyOUTP ON
. See
Test Record Forms
for details.
VOLT 6; CURR 5 (for example for
the E36102A, 6 V, 5 A output)
7. The output status should be CC and the output voltage should be close to zero.
8.
Divide the voltage drop (DMM reading) across the current shunt by the shunt resistance to convert
to amps.
9.
Record the current measured by the DMM and verify whether it is within the limits calculated.
Current readback accuracy
StepFront PanelSCPI
10.MEAS:CURR? (for example for the
E36102A, 6 V, 5 A output)
11.
Record the current returned by the SCPI command query via Keysight Connection Expert, and verify
whether it is within the limits calculated.
CC load regulation
This test measures the change in output current resulting from a change in output voltage from full
scale to short circuit.
1.
Turn off the power supply using the AC line switch.
2.
Connect the power supply output with a DMM, an electronic load and a current shunt as shown
below.
3.
Turn on the power supply using the AC line switch.
4.
Program the output voltage and output current as described in the test record form under CC load
regulation.
5.
Enable the output by sending the command OUTP ON or by pressing the Output ON key.
6.
Operate the electronic load in constant voltage mode and set its voltage to the power supply output
value as described in the test record form under “CC Load Regulation”. See the Test Record Forms
under Test Description of “CC Load Regulation” for details. Check that the front panel CC annunciator of the power supply remains lit. If it turns to CV or UNREG, adjust the load so that the output
current drops slightly until the CC annunciator lights up. Record the current reading (I
ing the voltage reading on the DMM by the resistance of the current monitoring resistor.
7.
Operate the electronic load in short (input short) mode. Record the current reading, (I
dividing the voltage reading on the DMM by the resistance of the current monitoring resistor.
load
short
), by divid-
), by
8.
Take the difference between the current readings in steps 6 and 7 is the load regulation current
(I
– I
load
specification limits.
). The difference of the readings during the immediate change should be within the
short
CCline regulation
1.
Turn off the power supply using the AC line switch.
2.
Connect the power supply output with a DMM, an electronic load and a current shunt. See Recom-
mended test equipment for details.
3.
Connect the AC power cord of the power supply to the AC power source.
4.
Turn on the power supply using the AC line switch.
5.
Program the output voltage and output current as described in the test record form under CC line
regulation.
6.
Operate the electronic load in constant voltage mode and set its voltage to the power supply output
value as described the test record form under “CC Line Regulation”. See the Test Record Forms
under Test Description of “CC Line Regulation” for details. Check that the front panel CC annunciator of the power supply remains lit. If it turns to CV or UNREG, adjust the load so that the output
current drops slightly until the CC annunciator lights up.
7.
Adjust the AC power source to low line voltage limit (90 VAC for nominal 100 VAC, 104 VAC for nominal 115 VAC, 194 VAC for nominal 215 VAC, or 207 VAC for nominal 230 VAC). Record the output
current reading (I
rent monitoring resistor.
8.
Adjust the AC power source to high line voltage (110 VAC for nominal 100 VAC, 127 VAC for nominal
115 VAC, 237 VAC for nominal 215 VAC or 253 VAC for nominal 230 VAC). Record the current reading (I
highline
the current monitoring resistor.
9.
Take the difference between the DMM readings in steps 7 and 8 is the CC line regulation (I
I
highline
cification limits.
). The difference of the readings during the immediate change should be within the spe-
lowline
) again immediately by dividing the voltage reading on the DMM by the resistance of
) by dividing the voltage reading on the DMM by the resistance of the cur-
lowline
–
Test Record Forms
Keysight E36102A
Report number: ______________________
Date: _______________________________
DescriptionLower limitResultUpper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
– Maximum Voltage Ouput (V
Voltage Readback
max
)
– Zero Voltage measured over interface
– Maximum Voltage measured over interface
CV Load Regulation (V
CV Line Regulation (V
CV Ripple and Noise
load
lowline
- V
)-2.6 mV______________2.6 mV
noload
- V
)-1.6 mV______________1.6 mV
highline
– peak-to-peak
– rms
Transient Response—______________50 μs
Constant Current Tests
Current Programming
– Zero Current Output (I0)
– Maximum Current Ouput (I
max
)
-0.003 V______________0.003 V
5.994 V______________6.006 V
V0- 0.003 V______________V0+ 0.003 V
V
- 0.006 V______________V
max
—______________10 mV
—______________10 μV
-0.004 A______________0.004 A
4.9925 A______________5.0075 A
max
+ 0.006 V
Current Readback
– Zero Current measured over interface
– Maximum Current measured over interface
CC Load Regulation (I
CC Line Regulation (I
load
lowline
- I
)-1250 μA______________1250 μA
short
- I
)-1250 μA______________1250 μA
highline
I0- 0.004 A______________I0+ 0.004 A
I
- 0.0065 A______________I
max
max
+ 0.0065 A
Test DescriptionInstrument settings
Voltage Programming & Readback, Zero Voltage (V0)0 V5 A
Voltage Programming & Readback, Maximum Voltage (V
)6 V5 A
max
Test DescriptionInstrument settings
CV Load Regulation, Line Regulation, Ripple and Noise6 V5 A
Transient Response6 V2.5 Ato 5 A
Current Programming & Readback, Zero Current (I0)0 A6 V
Current Programming & Readback, Maximum Current (I
CC Load Regulation, Line Regulation, Ripple and Noise5 A6 V
)5 A6 V
max
Keysight E36103A
Report number: ______________________
Date: _______________________________
DescriptionLower limitResultUpper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
– Maximum Voltage Ouput (V
Voltage Readback
max
)
– Zero Voltage measured over interface
– Maximum Voltage measured over interface
CV Load Regulation (V
CV Line Regulation (V
CV Ripple and Noise
load
lowline
- V
)-5 mV______________5 mV
noload
- V
)-4 mV______________4 mV
highline
-0.007 V______________0.007 V
19.983 V______________20.017 V
V0- 0.005 V______________V0+ 0.005 V
V
- 0.015 V______________V
max
max
+ 0.015 V
– peak-to-peak
– rms
Transient Response—______________50 μs
Constant Current Tests
Current Programming
– Zero Current Output (I0)
– Maximum Current Ouput (I
Current Readback
max
)
– Zero Current measured over interface
– Maximum Current measured over interface
CC Load Regulation (I
CC Line Regulation (I
load
lowline
- I
)-500 μA______________500 μA
short
- I
)-500 μA______________500 μA
highline
—______________30 mV
—______________2 mV
-0.001 A______________0.001 A
1.998 A______________2.002 A
I0- 0.001 A______________I0+ 0.001 A
I
- 0.002 A______________I
max
max
+ 0.002 A
Test DescriptionInstrument settings
Voltage Programming & Readback, Zero Voltage (V0)0 V2 A
Voltage Programming & Readback, Maximum Voltage (V
CV Load Regulation, Line Regulation, Ripple and Noise20 V2 A
Transient Response20 V1 Ato 2 A
Current Programming & Readback, Zero Current (I0)0 A20 V
Current Programming & Readback, Maximum Current (I
CC Load Regulation, Line Regulation, Ripple and Noise2 A20 V
)20 V2 A
max
)2 A20 V
max
Keysight E36104A
Report number: ______________________
Date: _______________________________
DescriptionLower limitResultUpper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
– Maximum Voltage Ouput (V
max
)
-0.012 V______________0.012 V
34.9705 V______________35.0295 V
Voltage Readback
– Zero Voltage measured over interface
– Maximum Voltage measured over interface
CV Load Regulation (V
CV Line Regulation (V
CV Ripple and Noise
load
lowline
- V
)-9.5 mV______________9.5 mV
noload
- V
)-7.5 mV______________7.5 mV
highline
– peak-to-peak
– rms
Transient Response—______________50 μs
Constant Current Tests
Current Programming
– Zero Current Output (I0)
– Maximum Current Ouput (I
Current Readback
max
)
V0- 0.008 V______________V0+ 0.008 V
V
- 0.0255 V______________V
max
—______________60 mV
—______________4 mV
-0.0006 A______________0.0006 A
0.9989 A______________1.0011 A
max
+ 0.0255 V
DescriptionLower limitResultUpper limit
– Zero Current measured over interface
– Maximum Current measured over interface
CC Load Regulation (I
CC Line Regulation (I
load
lowline
- I
)-250 μA______________250 μA
short
- I
)-250 μA______________250 μA
highline
I0- 0.0005 A______________I0+ 0.0005 A
I
- 0.001 A______________I
max
max
+ 0.001 A
Test DescriptionInstrument settings
Voltage Programming & Readback, Zero Voltage (V0)0 V1 A
Voltage Programming & Readback, Maximum Voltage (V
CV Load Regulation, Line Regulation, Ripple and Noise35 V1 A
Transient Response35 V0.5 Ato 1 A
Current Programming & Readback, Zero Current (I0)0 A35 V
Current Programming & Readback, Maximum Current (I
CC Load Regulation, Line Regulation, Ripple and Noise1 A35 V
)35 V1 A
max
)1 A35 V
max
Keysight E36105A
Report number: ______________________
Date: _______________________________
DescriptionLower limitResultUpper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
– Maximum Voltage Ouput (V
Voltage Readback
max
)
– Zero Voltage measured over interface
– Maximum Voltage measured over interface
CV Load Regulation (V
CV Line Regulation (V
CV Ripple and Noise
load
lowline
- V
)-16 mV______________16 mV
noload
- V
)-13 mV______________13 mV
highline
– peak-to-peak
– rms
-0.02 V______________0.02 V
59.95 V______________60.05 V
V0- 0.012 V______________V0+ 0.012 V
V
- 0.042 V______________V
max
—______________100 mV
—______________5 mV
max
+ 0.042 V
Transient Response—______________50 μs
DescriptionLower limitResultUpper limit
Constant Current Tests
Current Programming
– Zero Current Output (I0)
– Maximum Current Ouput (I
Current Readback
max
)
– Zero Current measured over interface
– Maximum Current measured over interface
CC Load Regulation (I
CC Line Regulation (I
load
lowline
- I
)-150 μA______________150 μA
short
- I
)-150 μA______________150 μA
highline
-0.0004 A______________0.0004 A
0.5993 A______________0.6007 A
I0- 0.0003 A______________I0+ 0.0003 A
I
- 0.0006 A______________I
max
max
+ 0.0006 A
Test DescriptionInstrument settings
Voltage Programming & Readback, Zero Voltage (V0)0 V0.6 A
Voltage Programming & Readback, Maximum Voltage (V
CV Load Regulation, Line Regulation, Ripple and Noise60 V0.6 A
Transient Response60 V0.3 Ato 0.6 A
Current Programming & Readback, Zero Current (I0)0 A60 V
Current Programming & Readback, Maximum Current (I
CC Load Regulation, Line Regulation, Ripple and Noise0.6 A60 V
)60 V0.6 A
max
)0.6 A60 V
max
Keysight E36106A
Report number: ______________________
Date: _______________________________
DescriptionLower limitResultUpper limit
Constant Voltage Tests
Voltage Programming
– Zero Voltage Output (V0)
– Maximum Voltage Ouput (V
Voltage Readback
max
)
– Zero Voltage measured over interface
– Maximum Voltage measured over interface
CV Load Regulation (V
load
- V
)-30 mV______________30 mV
noload
-0.04 V______________0.04 V
99.91 V______________100.09 V
V0- 0.02V______________V0+ 0.02 V
V
- 0.07V______________V
max
max
+ 0.07 V
DescriptionLower limitResultUpper limit
CV Line Regulation (V
CV Ripple and Noise
– peak-to-peak
– rms
Transient Response—______________50 μs
Constant Current Tests
Current Programming
– Zero Current Output (I0)
– Maximum Current Ouput (I
Current Readback
– Zero Current measured over interface
– Maximum Current measured over interface
CC Load Regulation (I
CC Line Regulation (I
lowline
lowline
load
- I
- V
max
short
- I
highline
)-22 mV______________22 mV
highline
—______________150 mV
—______________15 mV
-0.0003 A______________0.0003 A
)
)-100 μA______________100 μA
)-100 μA______________100 μA
0.3995 A______________0.4005 A
I0- 0.0002 A______________I0+ 0.0002 A
I
- 0.0004 A______________I
max
max
+ 0.0004 A
Test DescriptionInstrument settings
Voltage Programming & Readback, Zero Voltage (V0)0 V0.4 A
Voltage Programming & Readback, Maximum Voltage (V
CV Load Regulation, Line Regulation, Ripple and Noise100 V0.4 A
Transient Response100 V0.2 Ato 0.4 A
Current Programming & Readback, Zero Current (I0)0 A100 V
Current Programming & Readback, Maximum Current (I
CC Load Regulation, Line Regulation, Ripple and Noise0.4 A100 V
)100 V0.4 A
max
)0.4 A100 V
max
Calibration Adjustment Procedures
This chapter includes calibration adjustment procedures for Keysight E36100 series power supply.
Instructions are applicable for performing the procedures from either the front panel or a controller over
the LAN or USB.
Perform the verification tests before calibrating your instrument. If the instrument passes the verification tests, the unit is operating within its calibration limits and does not need to be re-calibrated.
Closed–case electronic calibration
The instrument uses closed-case electronic calibration; no internal mechanical adjustments are
required. The instrument calculates correction factors based on reference signals that you apply and
stores the correction factors in non-volatile memory. This data is not changed by cycling power, *RST,
or SYSTem:PRESet.
Calibration interval
The recommended calibration interval for Keysight E36100 series power supply is one year.
Calibration adjustment process
The following general procedure is recommended to complete a full calibration adjustment.
1.
Adhere to the test considerations. See Performance Verification > Test considerations for details.
2.
Perform the performance verification tests to characterize the instrument. See Performance Veri-
fication for details.
3.
Unsecure the instrument for calibration. See Calibration security for details.
4.
Perform the calibration procedures. See Calibration procedure for details.
5.
Secure the instrument against the calibration. See Calibration security for details.
6.
Take note of the security code and calibration count in the instrument's maintenance records.
7.
Perform the performance verification tests to verify the calibration.
Calibration security
The instrument has a calibration security code to prevent accidental or unauthorized calibration. When
you receive your power supply, it is secured by a default security code, as shown below. The security
code is not changed by a power cycle or *RST.
ModelDefault Code
E36102A36102
E36103A36103
E36104A36104
E36105A36105
E36106A36106
You can enter a security code of up to 6 digits on the front panel or up to 9 digits from the remote
interface.
To reset a forgotten security code to its factory default, you can turn on the instrument after
shorting CAL SECURE jumper J3 inside the instrument as shown below.
Calibration count
The instrument counts the number of times it has saved calibration data. Your instrument was
calibrated at the factory; when you receive your instrument, read and record the initial count. You can
only read the calibration count by sending the CAL:COUNt? query, and the calibration count is not
change by a power cycle or *RST.
If Auto Save is enabled, the count increments when you exit the calibration state. To avoid double
counting, do not manually save the count with Auto Save enabled.
Calibration message
You can use the CALibration:STRing command to store a message of up to 40 characters in calibration
memory. For example, you could store the last calibration date, the calibration due date, or contact
information for the person responsible for calibration. The calibration message is not affected by a
power cycle or *RST.
You can only store the calibration message when the instrument is unsecured, but you can execute the
CALibration:STRing? query regardless of whether the instrument is secured. A new calibration message
overwrites the previous message, and messages over 40 characters are truncated.
Saving calibration data
You must always save new calibration data before cycling instrument power or leaving the calibration
state with the Auto Save feature off. To save calibration data, send CAL:SAVE or save the calibration data
from the front panel.
Calibration auto save
The instrument includes a calibration Auto Save feature. This feature automatically saves the calibration
data to non-volatile memory and increments the calibration count when you exit the calibration state.
To enable or disable the CAL auto Save feature, send CAL:ASAV ON or CAL:ASAV OFF. To query the CAL
auto Save state, send CAL:ASAV?
Calibration procedure
Enter the calibration state
To begin the calibration procedure, you must enter the calibration state.
StepFront PanelSCPI
1Press and hold [Calibrate] while powering ON the unit.
2Press [Secure], then turn the knob to the security code, pressing the knob to move
between digits. Press [Secure] when done.
CAL:STATON,<code>
Calibrate voltage
Let the unit sit with output ON for one minute, then connect the DMM voltage input to the power
supply.
StepFront PanelSCPI
1Turn the knob to ‘Cal Voltage’ and press the knob.CAL:VOLT:LEV MIN
2Measure the output voltage (low point) with the DMM.
3Enter the measured value, and press the knob.CAL:VOLT <reading>
CAL:VOLT:LEV MAX
4Measure the output voltage (high point) with the DMM.
5Enter the measured value, and press the knob.CAL:VOLT <reading>
6Read DONE or FAIL on the display.(wait 10 seconds)
SYST:ERR?
Calibrate current
Let the unit sit with output ON for one minute, then connect a current monitoring resistor across the
output terminals to be calibrated and then connect a DMM across the terminals of the monitoring
resistor.
StepFront PanelSCPI
1Turn the knob to Cal Current and press the knob.CAL:CURR:LEV MIN
2Measure the low output current with the DMM (DMM reading / shunt resistance).
3Enter the measured value, and press the knob.CAL:CURR <reading>
CAL:CURR:LEV MAX
4Measure the high output current with the DMM (DMM reading / shunt resistance).
5Enter the measured value, and press the knob.CAL:CURR <reading>
6Read DONE or FAIL on the display.
Calibrate small current
Allow the unit to sit with output ON for one minute before continuing.
StepFront PanelSCPI
1Turn the knob to Cal Current and press the knob.CAL:CURR:LEV:LOW MIN
2Wait approximately 40 seconds for the ‘Cal in Prog’ message to leave the display.
StepFront PanelSCPI
3When the display shows second point calibration, connect the DMM current input
with the appropriate calibration resistor (see
Recommended test equipment
ply.
4Measure output current with the DMM.CAL:CURR:LEV:LOW MAX
5Enter the measured value, and press the knob.CAL:CURR:LOW <rdg>
6Read DONE or FAIL on the display.(wait 25 seconds)
Performance Verification >
) for your instrument model to the power sup-
SYST:ERR?
Save the calibration data
To save calibration data, go to the ‘CAL Save’ menu to save the calibration data or enable the ‘Auto Save’
feature. With Auto Save, calibration data willbe saved when the user exits the calibration menu.
After completing the Voltage, Current and Small Current calibrations, save the calibration data before
exiting the calibration state, or simply exit the calibration state if Auto Save is on.
To save the CAL data: CAL:SAVE
To enable the CAL Auto Save: CAL:ASAVE ON
To exit CAL State: CAL:STAT OFF, 36102
Specifications and Typical Characteristics
Product specifications and descriptions in this document are subject to change without
notice.
Performance Specifications
The following items are the performance specifications of the E36100 Series power supplies.