Keysight Technologies B2900 Series User Manual

Keysight Technologies B2900 Series Precision Source/Measure Unit
User’s Guide
Notices
Copyright Notice
© Keysight Technologies 2011-2016
No part of this manual may be reproduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technolo­gies as governed by United States and inter­national copyright laws.
Manual Part Number
B2910-90010
Edition
Edition 1, March 2011 Edition 2, July 2011 Edition 3, November 2014 Edition 4, December 2015 Edition 5, December 2016
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies International Japan G.K. 9-1, Takakura-cho, Hachioji-shi, Tokyo 192-0033 Japan
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer soft­ware,” as defined by Federal Acquisition Reg­ulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan-
dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://
www.keysight.com/find/sweula. The license
set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish techni­cal information related to commercial com­puter software or commercial computer software documentation that is not customar­ily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose com­mercial computer software or commercial computer software documentation. No addi­tional government requirements bey ond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of com­mercial computer software pursuant to the FAR and the DFARS and are set forth specifi­cally in writing elsewhere in the EULA. Key­sight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and
27.404.2 and DFARS 227.7102, the U.S. gov­ernment acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCU­MENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXI­MUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRAN­TIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFOR­MATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WAR­RANTIES OF MERCHANTABILITY AND FIT­NESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR
ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMA­TION CONTAINED HEREIN. SHOULD KEY­SIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEP­ARATE AGREEMENT SHALL CONTROL.
Open Software License
A portion of the software in this product is licensed under terms of the General Public License Version 2 (“GPLv2”). The text of the license and source code can be found at:
www.keysight.com/find/GPLV2
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/go/
conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
Latest Information
To get the latest firmware/software/electronic manuals/specifications/support information, go to www.keysight.com and type in the prod­uct number in the Search field at the top of the page.
About Customer Feedback
We love hearing from you. Please take a few moments and let us know how we can improve this manual. Click here to open the
Keysight B2900 manual feedback form.
We respect your privacy. Be assured that Key­sight will never sell or rent your information. Nor will Keysight share this information with other companies without your expressed con­sent. We make a commitment to you that we will respect and protect your privacy. Please see the details of this commitment in our Pri­vacy Statement. Click here to open the state-
ment.
COMPLIANCE WITH GERMAN NOISE REQUIREMENTS
This is to declare that this product is in conformance with the German Regulation on Noise Declaration for Machines (L
ärmangabe nach der Maschinenlärminformation-Verordnung -3.GSGV Deutschland).
Herstellerbescheinigung
GERÄUSCHEMISSION
Lpa < 70 dB
am Arbeitsplatz
normaler Betrieb
nach DIN 45635 T. 19
Manufacturer’s Declaration
ACOUSTIC NOISE EMISSION
Lpa < 70dB
operator position
normal operation
per ISO 7779
South Korean Class A EMC declaration
This equipment is Class A suitable for professional use and is for use in electromagnetic environments outside of the home.
Safety Summary
The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual may impair the protections provided by the instrument. In addition, it violates safety standards of design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for customer’s failure to comply with these requirements.
Product manuals may be provided on CD-ROM or in printed form. Printed manuals are an option for many products. Manuals may also be available on the Web. Go to
www.keysight.com and type the product model number in the Search field at the top of
the page.
NOTE Do not use this instrument in any manner not specified by the manufacturer. The
protective features of this instrument may be impaired if it is used in a manner not specified in the operation instructions.
This instrument is an INDOOR USE product.
This instrument complies with INSTALLATION CATEGORY II for mains input and INSTALLATION CATEGORY I for measurement input terminals, and POLLUTION DEGREE 2 defined in IEC 61010-1.
If an instrument is marked CAT I (IEC Measurement Category I), or it is not marked with a measurement category, its measurement terminals must not be connected to line-voltage mains.
Safety of any system incorporating the equipment is the responsibility of the assembler of the system.
WARNING Hazardous voltage of up to the instrument’s maximum voltage may appear at High Force,
Guard, and High Sense terminals if Interlock terminal is closed. Open the Interlock terminal when the High Force, Guard, and High Sense terminals are accessible. Voltage applied to the terminals will be limited up to ±42
Do not work the interlock function intentionally in order to bring the output voltage to the safe level. While the high voltage indicator is lit, the dangerous voltage by the output voltage or the residual charge appears on the measurement terminal.
V.
DANGEROUS PROCEDURE WARNINGS
Warnings, such as WARNING on the previous page, shall be complied. Procedures throughout in this manual prevent you from potentially hazard. Their instructions contained in the warnings must be followed.
BEFORE APPLYING POWER
Verify that all safety precautions are taken. Make all connections to the instrument before applying power. Note the instrument's external markings described under “Safety Symbols”.
GROUND THE INSTRUMENT
This is Safety Class I instrument. To minimize shock hazard, the instrument chassis and cabinet must be connected to an electrical ground. The power terminal and the power cable must meet International Electrotechnical Commission (IEC) safety standards.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard.
DO NOT REMOVE COVERS
No operator serviceable parts inside. Refer servicing to qualified personnel. To prevent electrical shock do not remove covers.
IN CASE OF DAMAGE
Instruments that appear damaged or defective should be made inoperative and secured against unintended operation until they can be repaired by qualified service personnel. Return the instrument to a Keysight Technologies sales or service office for services and repair to ensure that safety features are maintained.
USE ONLY THE SPECIFIC ACCESSORIES
Specific accessories satisfy the requirements for specific characteristics for using the instrument. Use the specific accessories, cables, adapters, and so on for safety reasons.
Safety Symbols
>
^
CAT I
The general definitions of safety symbols used on equipment or in manuals are listed below.
Direct current.
Alternating current.
Earth ground terminal.
Protective conductor terminal. For protection against electrical shock in case of a fault. Used with field wiring terminals to indicate the terminal which must be connected to ground before operating equipment.
Frame or chassis terminal. A connection to the frame (chassis) of the equipment which normally includes all exposed metal structures.
Grounded terminal which indicates the earth potential.
On supply.
Off supply.
Standby supply. The equipment will be marked with this symbol is not completely disconnected from AC mains when power switch is in the standby position.
In position of a bi-stable push switch.
Out position of a bi-stable push switch.
Hazardous voltage and potential for electrical shock. Do not touch terminals that have this symbol when the equipment is on.
Hot surface. Avoid contact. Surfaces are hot and may cause personal injury if touched.
Caution, refer to accompanying documentation. The equipment will be marked with this symbol when it is necessary for the user to refer to the instruction manual.
Read operator's manual. To indicate that the operator's manual or card should be read before continuing the operation.
IEC Measurement Category I
The CE mark shows that the product complies with all applicable European Directives.
The CSA mark is a registered trademark of the Canadian Standards Association.
The RCM mark is a registered trademark of the Australian Communications Authority. This signifies compliance with the Australian EMC Framework Regulations under the terms of the Radio communications Act.
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 du Canada.
This is the symbol for an Industrial, Scientific and Medical, Group 1 Class A product. (CISPR 11)
Korea’s safety and EMC mark
China RoHS - Environmentally Green Product Label
China RoHS - Product with Toxic Substance 40 yr EPUP
The Chinese mark for paper-based packaging materials; Paperboard and Corrugated Fiberboard
Plastic Material Coding Identification
WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure,
practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice,
or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
Power Supply and Measurement Safety
•Power Supply Safety
This instrument can output high currents and voltages. Make sure that the load or device under test can safely handle the output current and voltage. Also, make sure that the connection leads can safely withstand the expected currents and are insulated for the expected voltages.
The instrument outputs may be connected so as to float relative to earth ground. Isolation or floating voltage ratings are indicated on the instrument, near the output terminal or the Chassis ground terminal. There is the danger of electric shock by touching the floated measurement terminals. Keep in mind it to protect yourself. And it is a reason of using the recommended accessories.
Voltage/Current Measurement Safety
Multimeters and other instruments capable of measuring high voltages and currents are subject to specific safety concerns because of the circuits to which they may be connected. To safely use these instruments, you need to understand the markings on the instrument near the input terminals, which include the Protection Limits and the IEC Measurement Category.
Protection Limits
Keysight multimeters and other voltage measurement instruments provide protection circuitry to prevent damage to the instrument and to protect against the danger of electric shock, provided the Protection Limits are not exceeded. To ensure safe operation of the instrument, do not exceed the Protection Limits shown on the input terminals.
Source/Measure Terminals
High Sense, for 4-wire connection
High Force
Guard
Chassis ground
Low Force
Low Sense, for 4-wire connection
Source/measure unit, SMU, can simultaneously perform DC voltage or current output and measurement. Typical SMU has the Force, Sense, and Guard terminals as shown below. Normally the Force, Sense, and Guard are the same potential. Voltage marked around the terminals indicates the Protection Limits.
Force and Sense must be connected to a terminal of a device under test (DUT) for the 4-wire connection (Kelvin connection) which is effective for high current measurement and low resistance measurement. For the 2-wire connection to ease the connections, connect Force only. Do not connect Sense. It must be opened.
Guard should be connected to the guard shield which covers the DUT high side wiring for reducing leakage current caused by the wire. Or else, the Guard must be opened.
Chassis ground (green terminal) should be connected to the ground shield which covers the DUT including the guard shield to minimize the affect of noise. Or else, the chassis ground should be opened.
The following image is the source and measurement terminals of Keysight B2900. For the B2900, the High Force, High Sense, and Guard are the same potential. And the Low Force and Low Sense are the same potential.
High Voltage Shock Hazard
Keysight B2900 can force dangerous voltages (±210 V) at the High Force, Guard, and High Sense terminals. To prevent electric shock hazard, the following safety precautions must be observed during the use of Keysight B2900.
Use a three-conductor AC power cable to appliance coupler (inlet) and the instrument to an electric ground (safety ground).
Prepare shielding box which covers interface to a device under test and equipped with interlock circuit that opens when the door is opened.
Before performing measurement, connect the interlock circuit to the Interlock terminal of this instrument.
Confirm periodically that the interlock function works normally.
Before touching the connections of the High Force, Guard, and High Sense terminals, turn the instrument off and discharge any capacitors of the measurement path. If you do not turn the instrument off, complete “all” of the following items, regardless of any instrument’s settings.
Terminate source output by pressing the On/Off switch, confirm that the On/Off
switch turns off.
Confirm that the HV (high voltage) status indicator is not lit.
Open the shielding box access door (open the Interlock terminal).
Discharge any capacitors if the capacitance is connected to this instrument.
Warn workers in the vicinity of the instrument about hazardous conditions.
Gefahr durch Hochspannung
Von den Geräten Keysight B2900 können Spannungen an den Anschlüssen “High Force”, “Guard” und “High Sense” von bis zu 210 V ausgehen. Um elektrischem Schlag vorzubeugen, ist bei der Benützung der Geräte Keysight B2900 folgendes zu beachten.
Verwenden Sie ein dreiphasiges AC-Stromkabel für die Gerätsteckvorrichtung (Eingang) und schließen Sie das Instrument an eine Erdung an (Sicherheitserdung).
Bereiten Sie das Abschirmungsgehäuse vor, dass die Oberfläche eines zu testenden Geräts abdeckt und mit einem Verriegelungsstromkreis ausgestattet ist, der bei geöffneter Tür unterbrochen wird.
Vor der Messung verbinden Sie den Verriegelungsstromkreis mit dem Interlock-Anschluss dieses Instruments.
Prüfen Sie in regelmäßigen Abständen, dass die Verriegelungsfunktion ordnungsgemäß funktioniert.
Bevor Sie die Verbindungen zu den Anschlüssen “High Force”, “Guard” und “High Sense” berühren, schalten Sie das Instrument aus und entladen alle Kondensatoren des Messwegs. Wenn Sie das Instrument nicht ausschalten, führen Sie, unabhängig von den Instrumenteinstellungen, alle folgenden Schritte durch.
Beenden Sie die Messung, indem Sie auf die Taste “On/Off” drücken. Stellen Sie
sicher, dass die Statusanzeige “On/Off” nicht leuchtet.
Stellen Sie sicher, dass die Anzeige “HV” nicht leuchtet.
Öffnen Sie die Tür des Abschirmungsgehäuses (öffnen des
Interlock-Anschlusses).
Entladen Sie alle Kondensatoren, wenn die Kapazität mit das Instrument
verbunden ist.
Warnen Sie Mitarbeiter in der Umgebung des Instruments vor den Gefahren.
Danger de choc dû à une haute tension
Une tension dangereuse (max. ± pour; 210 Vdc) émanant du dispositif Keysight B2900 peut être sortie aux bornes High Force, Guard et High Sense, d de détection. Les précautions suivantes doivent être obserées contre commotion élec trique accidentelle.
Utilisez un câble d’alimentation CA à trois conducteurs vers le coupleur secteur (entrée) et branchez l’instrument sur une mise électrique à la terre (prise de terre de sécurité).
Préparez le boîtier de protection qui couvre l’interface avec le dispositif à tester et équipez-le d’un circuit de sécurité qui s’ouvre lors de l’ouverture d’une porte.
Avant de procéder aux mesures, connectez le circuit de sécurité à la borne Interlock de l’instrument.
Vérifiez régulièrement le bon fonctionnement de la fonction de sécurité.
Avant de toucher les connexions des bornes High Force, Guard et High Sense, mettez l’instrument hors tension et déchargez tout condensateur du chemin de mesure. Si vous ne mettez pas l’instrument hors tension, effectuez « toutes » les opérations ci-dessous, quels que soient les paramètres de l’instrument.
Terminez les mesures en appuyant sur la touche On/Off ; vérifiez que l’indicateur
d’état On/Off est éteint.
Vérifiez que le témoin HV est éteint.
'appareil de protection ou
-
Ouvrez la trappe d’accès au boîtier de protection (ouvrez la borne Interlock).
Déchargez les éventuels condensateurs si la capacité est connectée à
l’instrument.
Informez les personnes travaillant à proximité de l’instrument des conditions.
高電圧感電注意
Keysight B2900 High ForceGuard、High Sense 端子には、危険電圧が出力されること があります(最大 ±210 Vdc)。感電事故防止のため、必ず以下の事柄を守ってくだ さい。
•3 極電源ケーブルを使用して本器を接地してください。
ドアを開くことによって開放されるインターロック回路を装備し、被測定デバ
イスとのインタフェースを覆うことのできるシールド・ボックスを用意してく ださい。
測定を開始する前にはインターロック回路を本器の Interlock 端子に接続してく ださい。
インターロック機能が正常であることを定期的に確認してください。
•High Force、Guard、High Sense 端子に繋がる接続部に触れる前には、本器の電源
を切断してください。また、測定系のキャパシタを放電してください。電源を 切らない場合は、以下の事項を全て実施してください。
•On/Offスイッチを押して On/Off スイッチが消灯したことを確認してくださ
い。
高電圧警告インジケータ(HV)が消灯していることを確認してください。
シールド・ボックスのドアを開けてください(Interlock 端子を開放してくだ
さい)。
本器にキャパシタが接続されているならば、キャパシタを放電してくださ
い。
周囲のほかの作業者に対しても、高電圧危険に対する注意を徹底してください。
Product Stewardship
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control instrumentation” product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Keysight office or visit the following website for more information.
http://about.keysight.com/en/companyinfo/environment/
•LCD Fluorescent Lamp
Certain products sold by Keysight contain a liquid crystal display (LCD); backlighting for the LCD is provided by a fluorescent lamp which contains mercury, and must be managed, recycled, and/or disposed in accordance with all applicable laws, ordinances and regulations.
For information on how to recycle or dispose of the fluorescent lamp contained in your own product, visit the following website.
http://about.keysight.com/en/quality/env_compliance.shtml
If you live in the U.S., also visit the following websites.
http://www.lamprecycle.org
http://www.eiae.org
If you have additional questions, please visit the following website.
http://www.keysight.com/go/contactus
•Perchlorate Information
Perchlorate Material - special handling may apply. Visit the following website.
http://www.dtsc.ca.gov/hazardouswaste/perchlorate/
Equipment's real-time clock battery or coin cell battery may contain perchlorate and may require special handling when recycled or disposed of in California.
In This Manual
This manual describes the front panel operation, installation, and functions of the Keysight Technologies B2900 Precision Source/Measure Unit, SMU. This manual consists of the following chapters.
1. "Getting Started"
This chapter briefly explains how to use the Keysight B2900 by the front panel operation.
2. "Introduction"
This chapter describes overview, accessories, and options of the Keysight B2900.
3. "Installation"
This chapter explains how to install the Keysight B2900, and how to connect the device under test to a test fixture.
4. "Front Panel Reference"
This chapter provides the reference information of the Keysight B2900 front panel keys and graphical user interface.
5. "Front Panel Operations"
This chapter explains how to use the Keysight B2900 in the local mode.
6. "Function Details"
This chapter explains the several functions and initial settings of the Keysight B2900.
NOTE For the specifications of the B2900, see Data Sheet.
To get the latest Data Sheet, go to www.keysight.com/find/b2901a and click “Technical Support” and “Specifications”.
NOTE The information is subject to change without notice due to the future enhancement.
The actual screen image on the B2900 may be different from the image shown in this manual.
Contents
1. Getting Started
Applying DC Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Performing Spot Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Performing Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Operation Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
Operation Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
2. Introduction
Keysight B2900 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
Source/Measure Unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Measurement Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Limit/Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Output and Measurement Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Operation and Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Front Panel Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Source/Measurement Capabilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Program and Interface Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Software and Drivers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Furnished Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Available Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
3. Installation
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspecting the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Checking the Operation of Keysight B2900 . . . . . . . . . . . . . . . . . . . . . . . 3-4
Checking for Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Installing the Keysight B2900. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Environment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Connecting the Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Setting the Power Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Setting Date and Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Bench Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Rack Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Self-calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Connecting a DUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
2-Wire Connections or 4-Wire Connections. . . . . . . . . . . . . . . . . . . . . . 3-15
Floating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Using Test Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Using the N1295A Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Using the 16442B Test Fixture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Guarding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Performing Low Current Measurements . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Installing the Interlock Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Connecting to the Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
GPIB/USB Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
LAN Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Communicating Over the LAN. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Using the Graphical Web Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Using Telnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Using Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-32
Using Digital I/O. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
Accessory for Digital I/O Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
4. Front Panel Reference
Hard Keys and Rotary Knob. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Display and Assist Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Dual View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Single View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Graph View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-19
Roll View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
Status Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Function Keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
Config key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
Output Filter dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
Output Connection dialog box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
Sweep dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
Ranging dialog box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Wait Control dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-31
Function key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Math Expression dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Composite Limit Test Setup dialog box . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Limit Test Setup d ialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-34
Trace Buffer Setup d ialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Trigger key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Trigger Configuration dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
Result key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
Measure Result dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
Limit Test Resul t dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-41
Trace Statistical Result d ialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
File key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
File Selection dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
Program key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
I/O key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Data Output Format. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-46
Format (Measure) dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
Format (Math/Limit) dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
Format (Trace) dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-47
LAN Configuration dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
DIO Configuration dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-48
DIO Read/Write dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-49
Display key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-50
Display Preference dialog box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-51
System key group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-52
5. Front Panel Operations
Basic Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
To Change the Setting in a Field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
To Change the Settings on a Dialog Box . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
To Use Functions other than Measurement . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
To Set Power Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
To Return to Factory Shipment Condition . . . . . . . . . . . . . . . . . . . . . . . . 5-6
To Apply Initial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
To Set Beeper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
To Set Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
To Perform Self-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
To Perform Self-Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
To Set Operations at Power On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
To Display Error Message. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
To Clear Error Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
To Clear Timestamp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
To Set Automatic Clear of Timestamp . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
To Display Firmware Revision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
To Display License Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
To Set GPIB Address. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
To Set Remote Control Command Set . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
To Set Remote Display Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
To Save/Recall All Setup Information . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
To Enable/Disable Media Transfer Protocol for Easy File Access . . . . . 5-10
To Enable/Disable the Function to Restore the Trigger Mode Settings on
the Local State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
To Enable/Disable Source/Limit Real Time Update by Using Digit Pointer
and Knob. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
To Set Source Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
To Set Source Output Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
To Apply DC Voltage/Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
To Stop Source Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
To Set Limit (Compliance) Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
To Set Output Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
To Set Pulse Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
To Set Sweep Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-15
To Set List Sweep Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-16
To Set Source Output Trigger Parameters. . . . . . . . . . . . . . . . . . . . . . . 5-17
To Set Source Wait Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-17
To Set Output Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
To Set Connection Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
To Set Low Terminal State . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-18
To Enable or Disable High Capacitance Mode . . . . . . . . . . . . . . . . . . . . 5-19
To Enable or Disable Over Voltage/Current Protection . . . . . . . . . . . . . 5-19
To Specify Output-Off Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
To Enable or Disable Automatic Output-On Function . . . . . . . . . . . . . . 5-20
To Enable or Disable Automatic Output-Off Function . . . . . . . . . . . . . . 5-20
To Set Ranging Mode of the Sweep Source . . . . . . . . . . . . . . . . . . . . . . 5-20
To Set Sweep Direction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21
To Set Source Output Value after Sweep . . . . . . . . . . . . . . . . . . . . . . . . 5-21
To Execute Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
To Enable Resistance Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-22
To Set Measurement Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
To Perform Spot Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
To Stop Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
To Set Measurement Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-23
To Set Measurement Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
To Perform Sweep Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
To Set Measurement Trigger Parameters . . . . . . . . . . . . . . . . . . . . . . . . 5-25
To Set Measurement Wait Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
To Set Measurement Auto Range Operation . . . . . . . . . . . . . . . . . . . . . 5-26
To Enable or Disable Resistance Compensation . . . . . . . . . . . . . . . . . . 5-26
To Use Math Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
To Execute Limit Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
To Set Composite Limit Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
To Set Individual Limit Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
To Display Limit Test Result . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
To Use Trace Buffer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
To Set Trace Buffer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
To Display Statistical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
To Use Program Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
To Select a Program . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
To Control Program Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-34
6. Function Details
Limit/Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
To Set Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Ranging Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Setting the Ranging Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Measurement Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Aperture Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Overhead Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
To Control Source/Measure Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Pulse Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
To Control Pulse Output/Measure Timing . . . . . . . . . . . . . . . . . . . . . . . . 6-7
To Set Pulse Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Sweep Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
To Set Sweep Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
List Sweep . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Output Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
Over Voltage/Current Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Output-Off Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Automatic Output-ON/OFF Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
High Capacitance Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-15
Resistance Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Resistance Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Math Expression. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Predefined Math Expressions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Resources Used in the Expressions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-18
Limit Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
Trace Buffer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
Program Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Channel Grouping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Trigger System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Trigger Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
Device Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
Trigger Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
File Access Function (Easy File Access) . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-32
Interlock Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
Over Temperature Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
Initial Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35

1 Getting Started

Getting Started
D: Dual view
1: Single view
This chapter describes basic operations for Keysight B2900. Before learning the details of B2900, let us briefl y co ver the u se of the B2 900 sou rce/meas ure u nit . The operations require only B2900 and the power cord. To get started with the operations, open the measurement terminals.
This chapter consists of the following sections.
“Applying DC Output”
“Performing Spot Measurement”
“Performing Sweep Measurement”
“Operation Tips”
“Operation Su mmary”
NOTE Turning B2900 on
1. Connect the power cord from the AC input connector (receptacle) on B2900’s rear panel to an AC power outlet at your site.
2. Press the stan dby swit ch on the front panel. When the pow er is on, the LED below the switch is turned green.
B2900 boots up and performs a self-test. After B2900 boots normally, the front panel LCD displays the following view. It will be the Single view for the 1-channel model or the Dual view for the 2-channel model.
NOTE Connecting your DUT
If you want to connect your device under test, see “Connecting a DUT” on page
3-14.
1-2 Keysight B2900 User’s Guide, Edition 5

Applying DC Output

B2900 can be used as a DC voltage or current source. The following procedure configures B2900 as a voltage source and applies a voltage of +500 mV.
Step 1. Setting the Source mode ( voltage output)
Getting Started
Applying DC Output
1. Press the
Mode
assist key on a 1-ch model, or the
model. The field pointer appears on the Source mode, and the status changes to EDIT
(green). The status information will show .
2. Press the
VOLTS (V)
assist key. Or use the rotary knob or arrow keys
and to specify the mode, and press the knob to fix the setting.
The status changes to MOVE (blue). The status information will show .
Step 2. Setting the Source value (voltage output value)
Source
1. Press the
assist key on a 1-ch model, or the 2-ch model. Or , use the rota ry knob or arrow keys to move t he field poin ter onto the Source value and press the knob. The field pointer appears on the Source value, and the status changes to EDIT (green).
2. Turn the rotary knob clockwise to increase the value, or counter clockwise to decrease the value. Set the desired value.
Ch1 Mode
Ch1 Source
assist key on a 2-ch
assist key on a
When you press an arrow key, a digit pointer will appear on a digit. Turning the knob will change the value of the digit, or move the decimal point if the pointer is on it.
3. Press the rotary knob to fix the value. The status changes to MOVE (blue).
The numeric/alpha keys and the unit assist keys are also available for setting the Source value in the same way as for setting the Limit value, shown in the next step.
Keysight B2900 User’s Guide, Edition 5 1-3
Getting Started Applying DC Output
This example sets the Source value to +500 mV.
Step 3. Setting the Limit value (compliance current)
1. Press the
Limit
assist key on a 1-ch model, or the model. Or, use the rotary knob or arrow keys to move the field pointer onto the Limit value and press the knob. The field pointer app ears on the Limit value, and the status changes to EDIT (green).
2. Use the rotary knob, arrow keys, or numeric/alpha keys to set the value.
3. Press one of the unit assist keys to set the unit and fix the value. The status changes to MOVE (blue).
This example sets the Limit value to 1 mA.
Step 4. Enabling the channel
Press the Ch 1 switch to enable channel 1. This turns the switch green. Channel 1 starts applying the voltage specified by the Source value.
Step 5. Disabling the channel
Ch1 Limit
assist key on a 2-ch
Press the Ch 1 switch to disable channel 1. This turns off the switch light.
1-4 Keysight B2900 User’s Guide, Edition 5

Performing Spot Measurement

Performing Spot Measurement
B2900 can be used as a DC voltage or current meter. The following procedure performs one point current measurement.
Step 1. Setting the measurement mode (current measurement)
Getting Started
Press the
Measure
assist key on a 1-ch mo del, or the
model. And press the The rotary knob and arrow keys are also available for setting the measurement
mode, in the same way as for setting the Sou rce mode under “Applying DC Output”.
Step 2. Enabling the channel
Press the Ch 1 switch to enable channel 1. This turns the switch green. Channel 1 starts applying the voltage specified by the Source value.
Step 3. Starting measurement
Press the key to start a single (one shot) measurement. Spot measurement is performed once.
Press the key to start a repeat (continuous) measurement. Spot measurement is performed repeatedly.
The measurement results will be displayed in the left half of the LCD. The following example shows the measu rement results for a 1 k resistor co nnected
between the High Force and Low Force terminals of channel 1.
AMPS (I)
assist key.
Ch1 Measure
assist key on a 2-ch
Step 4. Disabling the channel
Press the Ch 1 switch to disable channel 1. This turns off the switch light.
Keysight B2900 User’s Guide, Edition 5 1-5
Getting Started Performing Sweep Measurement

Performing Sweep Measurement

B2900 supports sweep measurements. The following example applies a staircase sweep voltage and measures the current at each step voltage.
Step 1. Press the key to display the Single view. Step 2. Set the Source mode, Limit value, and measurement mode as shown in “Applying
DC Output” and “Performing Spot Measurement”.
This example sets the voltage output, a 10 mA limit value, and both current and voltage measurements.
Step 3. Setting the Sweep parameters
More
1. Press the
2. Press the
assist key to change the assist keys.
Show Sweep
assist key to display the Sweep setup parameters.
3. Press the rotary knob to change the status to EDIT (green).
4. Press the
LINEAR SINGLE
The source shape indicator shows the staircase icon, and the status changes to MOVE (blue).
1-6 Keysight B2900 User’s Guide, Edition 5
assist key to set the linear single sweep mode.
Getting Started
G: Graph View
Performing Sweep Measurement
5. Use the rotary knob, arrow keys, or numeric/alpha keys to set the Start (sweep start value), Stop (sweep stop value), and Step (sweep step value) or Points (number of sweep steps).
This example sets Start to 1 V, Stop to 2 V, and Points to 11.
Step 4. Press the key to display the Graph view. Step 5. Enabling the channel
Press the Ch 1 switch to enable channel 1. This turns the switch green. Channel 1 starts applying the voltage specified by the Source value.
Step 6. Press the key to start a single (one shot) measurement. Sweep measurement
is performed once. And the measurement results will be displayed on the graph.
Step 7. Press the
Auto Scale
The following example shows the measu rement results for a 1 k resistor co nnected between the High Force and Low Force terminals of channel 1.
Step 8. Disabling the channel
Press the Ch 1 switch to disable channel 1. This turns off the switch light.
Keysight B2900 User’s Guide, Edition 5 1-7
assist key to fit the trace in the graph scale.
Getting Started
Graph
G
Single Ch1
1
Roll
R
Dual
D
2
Single Ch2
B2912A
Graph
G
Single
1
B2901A
Graph
G
Single Ch1
1
Dual
D 2
Single Ch2
B2902A
Graph
G
Single
1
Roll
B2911A
R

Operation Tips

Operation Tips
This section introduces key operation and status information on B2900.
“Changing the View Mode”
“Editing the Set up”
“Status Information”
Figure 1-1 Changing the View Mode
For details about the View mode, see “Display and Assist Keys” on page 4-6.
1-8 Keysight B2900 User’s Guide, Edition 5
Figure 1-2 Editing the Setup
1 2
34
MOVE status EDIT status
1
Press to select
MOVE status
2
EDIT status
3 4
Select
Press to fix
Fix
Trigger auto
Trigger active
High voltage
Floating
Remote
Local lockout
Error
MOVE/EDIT
LXI LAN status indicator
View mode
Getting Started
Operation Tips
For more information about the fro nt panel k eys, see “Hard Keys and Rotary Knob”
on page 4-3.
If the field pointer is in the EDIT (green) status on the Source or Limit (Compliance) field, turning the knob changes the setting value of the source channel in real time.
Figure 1-3 Status Information
For details about the status information, see “Status Information” on page 4-24.
Keysight B2900 User’s Guide, Edition 5 1-9
Getting Started Operation Summary

Operation Summary

This section summarizes front panel operations on B2900.
“Basic Operations”
“Channel Setup”
“Source Setup”
“Measurement Setup”
“Display Setup”
“File Operations”
“Miscellaneous Functions”
“Interface Setup”
“System Setup and Operations” For details about front panel operations, see Chapter 4, “Front Panel Reference.”
1-10 Keysight B2900 User’s Guide, Edition 5
Table 1-1 Basic Operations
Task Relevant front panel key
To turn the B2900 on/off Standby switch
Getting Started
Operation Summary
To change the display mode To enable/disable the specified channel Ch 1 To cancel the previous setup operation To return to the upper menu of a function key or softkey To return to the local status from the remote status To start a single (one shot) output/measurement To start a repeat (continuous) measurement
View
key
On/Off
Cancel/Local
Cancel/Local
Cancel/Local
Trigger
key
Auto
key
switch or Ch 2
key key key
On/Off
switch
To move the field pointer Rotary knob or arrow keys To move the digit pointer Rotary knob or arrow keys To switch the EDIT/MOVE status Rotary knob on a setup field To select the setup value Assist keys, rotary knob, or arrow keys To save/recall all setup information of the B2900 Config > Save or Recall function keys
Table 1-2 Channel Setup
Task Relevant front panel key
To enable/disable the specified channel Ch 1
On/Off
switch or Ch 2
On/Off
switch To select the sensing type; 2-wire or 4-wire Config > Source > Connection function keys T o select the low terminal status; grounded or floating Config > Source > Connection function keys To enable/disable high capacitance mode Config > Source > Connectio n function keys To enable/disable over voltage/current protection Config > Source > Connection function keys To enable/disable resistance compensation Config > Measure > R Compen function keys To enable/disable 2-channel synchronous operation Config > Common > Group function keys
Keysight B2900 User’s Guide, Edition 5 1-11
Getting Started Operation Summary
Table 1-3 Source Setup
Task Relevant front panel key
To enable/disable source output Ch 1
switch or Ch 2
On/Off
switch
On/Off
To select source output mode Mode, Ch1 Mode, or Ch2 Mode assist key To set source output value Source, Ch1 Source, or Ch2 Source assist key To set limit/compliance value Limit, Ch1 Limit, or Ch2 Limit assist key To set constant source ranging mode Speed assist key or Hide Sweep/Pulse/Trigger assist key on
Single view To set sweep source Show Sweep assist key on Single view To set list sweep source Edit assist key in the EDIT status for the LIST sweep
Start/Stop/Points field To set sweep source ranging mode Config > Source > Sweep function keys To set sweep direction Config > Source > Sweep function keys To set source output value after sweep Config > Source > Sweep function keys To set pulse source Show Pulse assist key on Single view To set source wait time Config > Common > Wait function keys To set output filter Config > Source > Filter function keys To select output-off status Config > Sour ce > Connection function keys To enable/disable automatic output ON Config > Source > Connection function keys To enable/disable automatic output OFF Config > Source > Connection function keys
Table 1-4 Measurement Setup
Task Relevant front panel key
To enable/disable the specified channel Ch 1
switch or Ch 2
On/Off
switch
On/Off
To select measurement mode Measure, Ch1 Measure, or Ch2 Measure assist key To set measurement speed Speed assist key on Single view
1-12 Keysight B2900 User’s Guide, Edition 5
Getting Started
Operation Summary
Task Relevant front panel key
To set measurement ranging mode Speed assist key or Hide Sweep/Pulse/Trigger assist key on
Single view To define automatic ranging operation Config > Measure > Ranging function keys To set measurement wait time Config > Common > Wait function keys
Table 1-5 Display Setup
Task Relevant front panel key
To change the display mode
View
key To change the color set Display > Color function keys To enable zoom-in Display > Zoom > ON function keys To disable zoom-in Zoom Out assist key for the zoom-in display status To change the data digits Display > Digits function keys T o enable/disable the front panel display
Display > Remote function keys
in the remote condition
Table 1-6 File Operations
Task Relevant front panel key
To save a measurement result data to USB memory File > Save > Measure function keys To save a math result data to USB memory File > Save > Math function keys To save a limit test result data to USB memory File > Save > Limit Test function keys To save a trace buffer data to USB memory File > Save > Trace function keys To save a system setting data to USB memory File > Save > Config function keys To save a graph screen damp to USB memory Dump Screen assist key on Graph or Roll view To load a system setting data from USB memory File > Load > Config function keys To load a list sweep data from USB memory Load assist key in the EDIT status for the LIST
sweep Start/Stop/Points field
Keysight B2900 User’s Guide, Edition 5 1-13
Getting Started Operation Summary
Table 1-7 Miscellaneous Functions
Task Relevant front panel key
T o enable/disable Limit (Compliance) for each polarity
To enable/disable Source/Limit real time update by using digit pointer and knob
To see measurement result Result > Measure function keys To use math expression Function > Math function keys To see math result Result > Measure function keys To set limit test Function > Limit Test function keys To see limit test result Result > Limit Test function keys To set trace buffer Function > Trace function keys To see trace statistical result Result > Trace function keys To select program memory Program > Catalog function keys To control program memory Program > Control function keys To set trigger parameters easily Show Trigger assist key on Single view To set trigger parameters in detail Function > Trigger > Config function keys T o control trigger system Function > Trigger > Initiate/Abort/Immediate function keys
Display > Pref. function keys
Display > Pref. function keys
Table 1-8 Interface Setup
Task Relevant front panel key
To specify measurement data elements I/O > Format > Measure function keys T o specify math and limit test data elements I/O > Format > Math/Limit function keys To specify trace statistical data elements I/O > Format > Trace function keys To select data output format I/O > Format > Data Type function keys To enable/disable byte swap of binary data I/O > Format > Data Swap function keys
1-14 Keysight B2900 User’s Guide, Edition 5
Task Relevant front panel key
To set LAN configuration I/O > LAN > Config function keys To see status of LAN interface I/O > LAN > Status function keys To reset all of LAN connections I/O > LAN > Reset function keys To sets LAN settings to the default settings I/O > LAN > Default function keys
Getting Started
Operation Summary
To enable/disable Media Transfer Protocol for easy file access
To see status of USB interface I/O > USB function keys To set GPIB address I/O > GPIB function keys To see status of GPIB interface I/O > GPIB function keys To set configuration of Digital I/O I/O > DIO > Config function keys To read/write a value set to Digital I/O I/O > DIO > R/W function keys
Table 1-9 System Setup and Operations
Task Relevant front panel key
To set power line frequency System > PLC function keys To return to the factory shipment
condition To apply initial settings System > Reset function keys To perform self-calibration System > Cal/Test > Self-Cal function keys To perform self-test System > Cal/Test > Self-Test function keys
I/O > USB function keys
System > More > Factory function keys
To check errors System > Error > Log function keys To clear error log System > Error > Clear function keys To clear time stamp System > Timestamp > Clear function keys To set automatic clear of time stamp System > Timestamp > Auto CLR function keys To set start-up operation System > More > Start-up function keys
Keysight B2900 User’s Guide, Edition 5 1-15
Getting Started Operation Summary
Task Relevant front panel key
To enable/disable beep and sound System > M ore > Sound function keys To select remote control command set System > More > Language function keys T o enable/disable the function to restore
the trigger mode settings on the local state
To set date and time System > More > Info. > Date/Time function keys To perform firmware update System > More > Info. > Update > Firmware function keys To perform demonstration System > More > Info. > Demo. function keys
Display > Pref. function keys
1-16 Keysight B2900 User’s Guide, Edition 5

2Introduction

Introduction
This chapter describes the basic functions and features of the Keysight B2900, and consists of the following sections.
“Keysight B29 00 Ser ie s”
“Front View”
“Rear View”
“Source/Measure Unit”
“Operation and Fu nctions”
“Software and Drivers”
“Accessories”
“Options”
2-2 Keysight B2900 User’s Guide, Edition 5

Keysight B2900 Series

Keysight B2900 is a series of precision SMU (source/measure unit, SMU). B2900 provides the LCD, front p a nel keys , an d ro tar y kn ob f or applying voltage/current or measuring voltage/current/resistance. B2900 also supports several functions, such as sweep output, pulse output, arbitrary waveform generation, limit test, trace buf fer , math expressions, and graph plot. Hence, B2900 can be used as a DC (constant) voltage/current source, sweep voltage/current source, pulse generator, arbitrary waveform generator, and multimeter.
The product lineup of the Keysight B2900 series includes the following products.
Table 2-1 Keysight B2900 Series
Introduction
Keysight B2900 Series
Effective set and measure value
Model
No.
B2901A 1 1 pA, B2902A 2 B2911A 1 0.01 pA, 0.1 V B2912A 2
B2900 is a LAN eXtended Interface Core compliant instrument.
Number
of SMU
channels
Minimum resolution Maximum
current
Set Measure
1 V
0.1 pA,
0.1 V
DC: ± 3.03 A
Pulse: ± 10.5 A
Maximum
voltage
± 210 V
Keysight B2900 User’s Guide, Edition 5 2-3
Introduction Front View

Front View

This section describes the front view of the Keysight B2900 series.
Standby switch Turns the instrument on or off. When the power is on, the LED below the switch
is turned green.
•Display Displays the source setup, measurement result, status information, etc. For
details, see Chapter 4, “Front Panel Reference.”
•Trigger key Starts a single (one shot) output/measurement or initiates trigger system. If a
repeat measurement is in progress, stops it. The single measurement is performed with the DC bias output, staircase sweep
output, pulsed bias output, or pulsed sweep output set to the channel. One single measurement can contain the maximum of 100000 measurement points.
Once a single measurement starts, the data buffer (maximum 100000 data) is cleared, and the last single measurement result is stored in the buffer. The measurement result is displayed on the Single view, Dual view, Graph view, or Roll view.
NOTE If measurement is not performed properly , check the trigger setting. The trigger type
must be set to AUTO, or the trigger count (Count) must be set properly. See
“Trigger parameters” on page 4-17.
2-4 Keysight B2900 User’s Guide, Edition 5
Introduction
Front View
Auto key Starts a repeat (continuous) measurement. If a repeat measurement is in
progress, stops it. The repeat measurement is performed with the DC bias output of the Source
value. And the measurement result is displayed on the Single view, Dual view, or Roll view. The repeat measurement result is not stored in the buffer.
USB-A connector Used to connect a USB memory. After discon necting the USB memory, wait 10
seconds before connecting it again or new one.
CAUTION Turning the instrument off while the USB memory is being accessed may damage
the device.
Function keys Six function keys are available below the display. They are assigned to the
softkeys, Config, Function, Trigger, Result, File, Program, I/O, System, and More. For details, see Chapter 4, “Front Panel Reference.”
Assist keys Five assist keys are available to the right of the display. They are assigned to
several softkeys, such as Mode, Source, Limit, Measure, and More. The softkey assignment depends on the display mode (Single, Graph, Roll, or Dual). For details, see Chapter 4, “Front Panel Reference.”
•View key Changes the display mode. Pressing the key changes the mode as shown below.
On B2901A Single Graph (return to Single) On B2902A Dual Single for channel 1 Single for channel 2
Graph (return to Dual)
On B2911A Single Graph Roll (return to Single) On B2912A Dual Single for channel 1 Single for channel 2
Graph Roll (return to Dual)
Cancel / Local key Cancels the setup operation if the instrument is in the local status. Returns the
instrument to the local status if it is in the remote status.
Keysight B2900 User’s Guide, Edition 5 2-5
Introduction
Front View
Numeric/alpha keys Used to enter the value of setup parameters such as the source output value, limit
(compliance) value, and message, specified by the field pointer.
Rotary kno b If the field pointer is in the MOVE (blue) status, turning the knob moves the
pointer. Pressing the knob fixes the pointer position and changes the pointer status to EDIT (green).
If the field pointer is in the EDIT (green) status, turning the knob changes the value of the setup parameter specified by the pointer. Pressing the knob f ixes the value and changes the pointer status to MOVE (blue).
Left and right keys If the field pointer is in the MOVE (blue) status, pressing the key moves the
pointer. If the field pointer is in the EDIT (green) status, pressing the key changes the
value of the setup parameter specified by the pointer. If the field pointer is in the EDIT (green) status on a numeric value entry field,
pressing the key changes the pointer to a digit pointer.
On/Off switch(es) Used to enable or disable the SMU channel. Turns the channel off if it is in the
output status even if it is in the remote status. One switch on 1-channel models, and two switches on 2-channel models.
The switch turns green if the channel is enabled. The switch turns red if the channel is in the high voltage state.
Channel 1 source/measure terminals T erm inals for SMU channel 1 . High For ce, Low Force, High Sense, Low Sense,
Guard, and chassis ground. For details, see “Connecting a DUT” on page 3-14
CAUTION Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the B2900.
CAUTION Do not apply current to the chassis ground terminal. Doing so will damage the
B2900.
2-6 Keysight B2900 User’s Guide, Edition 5

Rear View

This section describes the rear view of the Keysight B2900 series.
Channel 2 source/measure terminals Only on 2-channel models. Terminals for SMU channel 2. High Force, Low
Force, High Sense, Low Sense, Guard, and chassis ground. For details, see
“Connecting a DUT” on page 3-14
Introduction
Rear View
CAUTION Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the B2900.
CAUTION Do not apply current to the chassis ground terminal. Doing so will damage the
B2900.
GPIB interface connector Use a Keysight 82357A/B USB/GPIB interface or Keysight 10833A/B/C/D
GPIB cable to connect to an external computer or equipment.
Cooling fan
AC input connector AC power cord is connected to this receptacle.
LAN interface connector Connects to 10/100 Base-T interface. Left LED indicates activity. Right LED
indicates link integrity.
Keysight B2900 User’s Guide, Edition 5 2-7
Introduction
Rear View
USB-B connector Connects to USB interface.
Digital I/O connector D-sub 25 pin female connector for general purpose I/O (GPIO). Can be used as
trigger input/output terminals or as an interface to a handler or the likes. For details, see “Using Digital I/O” on page 3-34
Pins 16 and 24 and pins 17 and 25 are reserved for the interlock function. If the terminals are open, the instrument output is limited to 42 V. Be sure to connect the terminals to the Keysight 16442B test fixture or another DUT interface before performing measurements. If you do not use 16442B, you need to install an interlock circuit. For details on how to install the interlock circuit, see
“Installing the Interlock Circuit” on page 3-24.
WARNING Dangerous voltage, instrument maximum output voltage may appear at the
High Force, High Sense, and Guard terminals if the Interlock terminal is closed.
Une tension dangereuse, une tension de sortie maximale de l'appareil peut apparaître aux bornes High Force, High Sense et Guard si la borne Interlock est fermée.
NOTE Serial Number
Y o u will need the instrument’ s serial number when using the Keysight T echn ologies telephone assistance program. The serial number label is attached to the bottom of the instrument.
2-8 Keysight B2900 User’s Guide, Edition 5

Source/Measure Unit

High Force
Chassis ground
V
A
+
-
+-
High Sense
4-wire connections
Low Sense
Low Force
High Force
Chassis ground
V
A
+
-
+-
Low Force
2-wire connections
The source/measure unit (SMU) can apply voltage or current, and can measure current or voltage.
Figure 2-1 shows a simplified SMU circuit diagrams. The SMU can perform the
following operations.
Applies voltage and measures current
Applies current and measures voltage
Applies voltage and measures voltage
Applies current and measures current
Figure 2-1 Simplified SMU Circuit Diagram
Introduction
Source/Measure Unit
Keysight B2900 User’s Guide, Edition 5 2-9
Introduction Source/Measure Unit

Measurement Parameters

B2900 supports the following measurement parameters.
Current
Voltage
Resistance
•Power
Resistance data is given by Resistance=Vmeas/Imeas. Power data is given by Power=VmeasImeas. In the above formula, Vmeas is the voltage measurement data, and Imeas is the
current measurement data. For using the resistance compensation, see “Resistance Compensation” on page
6-16.

Limit/Compliance

B2900 has a compliance feature that limits the output voltage or current to prevent damage to the device under test. Wh en the SMU applies v oltage, you can sp ecify the current limit/compliance. When the SMU applies current, you can specify the voltage limit/compliance. For details about limit/compliance, see
“Limit/Compliance” on page 6-3.
2-10 Keysight B2900 User’s Guide, Edition 5

Output and Measurement Ranges

6
10.5
10.5
Voltage (V)
Current (A)
6
1.515
-
-
-
1.515
Pulsed output
200
200
-
-
210
210
6
3.03
3.03
Voltage (V)
Current (A)
6
0.105
-
-
-
0.105
DC output
1.515
1.515
-
-
21
21
Duty cycle 2.5 %
Duty cycle 99.9999 %
Pulsed output
~
~
This section describes typical specifications of the Keysight B2900 series.
Maximum voltage and current: see Figure 2-2.
Maximum power: 31.8 W
Output/measurement value and resolution: see Table 2-2 to Table 2-7.
Figure 2-2 Maximum Voltage and Current
Introduction
Source/Measure Unit
Keysight B2900 User’s Guide, Edition 5 2-11
Introduction Source/Measure Unit
Table 2-2 Current Output Range
Maximum voltage
DC
output
Pulsed output
Pulse width t
Range
value
10 nA
Setting resolution
B2901A B2902A
d
DC output current or
B2911A B2912A
pulse peak/base
current
a b
10 fA 0 |I| 10.5 nA 210 V 210 V 50 s  t  t
100 nA 1 pA 100 fA 0 |I| 105 nA
1 A 10 pA 1 pA 0 |I| 1.05 A
10 A 100 pA 10 pA 0 |I| 10.5 A
100 A 1 nA 100 pA 0 |I| 105 A
1 mA 10 nA 1 nA 0 |I| 1.05 mA
10 mA 100 nA 10 nA 0 |I| 10.5 mA
100 mA 1 A 100 nA 0 |I| 105 mA
1 A 10 A1 A0  |I|  105 mA
105 mA < |I| 1.05 A 21 V 21 V
0 |I| 1.05 A
― ―
200 V 50 s t 2.5 ms180 V 50 s t 10 ms
1.5 A 0 |I| 105 mA 210 V 210 V 50 s  t 105 mA < |I| 1.515 A 21 V 21 V
t
c
max
max
0 |I| 1.515 A
0 |I| 1.05 A
― ―
200 V 50 s t 2.5 ms180 V 50 s t 10 ms
3 A 100 A 10 A0  |I|  105 mA 210 V 210 V 50 s  t  t
105 mA < |I| 1.515 A 21 V 21 V
1.515 A < |I| 3.03 A 6 V 6 V
10 A 0 |I|  10.5 A
0 |I| 1.515 A
0 |I| 1.05 A
― ― ―
6 V 50 s t 1 ms200 V 50 s t 2.5 ms180 V 50 s t 10 ms
2-12 Keysight B2900 User’s Guide, Edition 5
max
Source/Measure Unit
a. Table 2-4 shows the limitations when using Channels 1 and 2 for DC output or Pulsed output
with 50 s  t  t
(=99.9999 ks).
max
b. Maximum base current is 500 mA for the pulse with 50 s  t 1 ms, and 50 mA for the pulse
with 50 s  t 2.5 ms or 50 s  t 10 ms.
c. Maximum duty cycle is 99.9999 % for the pulse with 50 s  t  t
, and 2.5 % for the pulse
max
with 50 s  t 1 ms, 50 s  t 2.5 ms, or 50 s  t 10 ms.
d. Available for the B2911A and B2912A. Not available for the B2901A and B2902A.
Table 2-3 Voltage Output Range
Introduction
Pulsed output
a
Pulse width t
Range
value
Setting resolution
B2901A B2902A
B2911A
B2912A
DC output voltage or pulse peak/base
voltage
Maximum current
DC
output
0.2 V 1 V0.1 V0  |V|  0.21 V 3.03 A 3.03 A with 50 s  t  t 2 V 10 V1 V0  |V|  2.1 V
10.5 A with 50 s t 1 ms
20 V 100 V10 V0  |V|  6 V
6 V < |V| 21 V 1.515 A 1.515 A with 50 s  t  t
200 V 1 mV 100 V0  |V|  6 V 3.03 A 3.03 A with 50 s  t  t
10.5 A with 50 s t 1 ms
6 V < |V| 21 V 1.515 A 1.515 A 50 s  t  t
21 V < |V| 210 V 105 mA 105 mA
0 |V| 180 V 0 |V| 200 V
― ―
1.05 A 50 s t 10 ms
1.515 A 50 s t 2.5 ms
a. Table 2-4 shows the limitations when using Channels 1 and 2 for DC output or Pulsed output
with 50 s  t  t
b. Maximum duty cycle is 99.9999 % for the pulse with 50 s  t  t
(=99.9999 ks).
max
, and 2.5 % for the pulse
max
with 50 s  t 1 ms, 50 s  t 2.5 ms, or 50 s  t 10 ms.
b
max
max
max
max
Keysight B2900 User’s Guide, Edition 5 2-13
Introduction Source/Measure Unit
Table 2-4 Limitations for using Channels 1 and 2
Channel 1 voltage V1 Channel 2 voltage V2
0 < |V1| 6 V 0 < |V2| 6 V I1 + I2 4 A
6 V < |V1| 21 V 0 < |V2| 6 V I1 + I2 0.625 2.5 A
a. I1: Channel 1 current, I2: Channel 2 current
Table 2-5 Current Measurement Range
Range value Current measurement value Resolution
a
10 nA 100 nA 0 |I| 106 nA 100 fA
1 A0  |I|  1.06 A1 pA
10 A0  |I|  10.6 A 10 pA
100 A0  |I|  106 A 100 pA
1 mA 0 |I| 1.06 mA 1 nA
Current limit
6 V < |V2| 21 V I1 + I2 1.6 4 A
6 V < |V2| 21 V I1 + I2 2.5 A
0 |I| 10.6 nA 10 fA
a
10 mA 0 |I| 10.6 mA 10 nA
100 mA 0 |I| 106 mA 100 nA
1 A 0 |I| 1.06 A 1 A
1.5 A 0 |I| 1.53 A 3 A 0 |I| 3.06 A 10 A
10 A
b
0 |I| 10.6 A
a. Available for the B2911A and B2912A. Not available for the B2901A
and B2902A.
b. Available for pulse mode. Not available for DC mode.
2-14 Keysight B2900 User’s Guide, Edition 5
Table 2-6 Voltage Measurement Range
Range value Voltage measurement value Resolution
0.2 V 0 |V| 0.212 V 0.1 V 2 V 0 |V| 2.12 V 1 V
20 V 0 |V| 21.2 V 10 V
200 V 0  |V| 212 V 100 V
Introduction
Source/Measure Unit
Table 2-7 Resistance Measurement Range
Range value Resistance measurement value
2 0 < R 2 1  1 A
20 2  < R 20 10  100 mA
200 20 < R 200 100  10 mA
2 k 200 < R 2 k 1 m 1 mA
20 k 2 k < R 20 k 10 m 100 A
200 k 20 k < R 200 k 100 m 10 
2 M 200 k < R 2 M 1 1 A
20 M 2 M < R 20 M 10 100 nA
200 M 20 M < R 200 M 100 10 nA
1
Display
resolution
Test
current
A
1. The res istan ce meas urem ent r ange is effective for the resistance measurements set to the AUTO mode which is selected by the :SENS:RES:MODE command.
Keysight B2900 User’s Guide, Edition 5 2-15
Introduction Operation and Functions

Operation and Functions

Front Panel Interface

4.3” color TFT WQVGA (480272, with LED backlight) provides a graphical user interface (GUI) with the following features.
4-view modes (Single, Dual, Graph, Roll view)
Easy-to-use key operations
Easy access to basic and advanced source/measurement capabilities
Single view
Single view provides basic and advanced settings and display capabilities for the selected channel from the front panel of the instrument. No additional controller or software is required.
Dual view
Dual view provides the basic settings and display capabilities for both channels 1 and 2. Up to 6½ digits can be displayed in the Dual view mode. This mode is available only for the B2902A and B2912A.
Graph view
Graph view displays measurement results on XY graphs (such as I-V and I-t/V-t curves) on up to 2 channels. This is useful for quick evaluation of device characteristics, especially those obtained from sweep measurements.
Roll view
Roll view draws I-t or V-t curves similar to the curves drawn by a strip chart recorder. Up to 1000 acquired data points can be displayed and updated while the measurement is still in progress. Roll view’s continuous measurement capability is especially useful for monitoring low frequency phenomena. Roll view is available only for the B2911A and B2912A.
Easy-to-Use Key Operations
Numeric/alpha keys, assist keys, and rotary knob for easy operation
Numeric/alpha keys enable direct input of alphanumeric values.
Assist key guides operation on the front panel.
2-16 Keysight B2900 User’s Guide, Edition 5
Introduction
Operation and Functions
Easy Access to Source/Measurement Capabilities
Settings:
V/I source mode
Source value
Limit (Compliance) value
Sweep parameters
Pulse parameters
Measure speed
Range setting
Trigger setting
Result Displays:
Numeric results fully up to 6½ digits both in Single and Dual view modes
Characterization in Graph view and Roll view
Data list
Advanced Capabilities:
Limit testing by user definable limits working with component handlers
T race memory for accumu lating measurement (or math) result and collecting their statistics
File Operations (USB memory): Save:
System configuration
Measurement/Math result
Limit test result
•Trace data
Graph screen dump Load:
System configuration
List sweep data
Keysight B2900 User’s Guide, Edition 5 2-17
Introduction Operation and Functions

Source/Measurement Capabilities

Sweep Measurement
Number of sweep steps: 1 to 100000. Or 1 to 2500 for the following B2900.
B2900 series without the license “SWS”
B2900 series with the firmware revision 3.0 or earlier
B2901A with the serial number MY51140470 or below
B2902A with the serial number MY51140683 or below
B2911A with the serial number MY51140213 or below
B2912A with the serial number MY51140498 or below
Sweep mode: linear, logarithmic (log), or list
Sweep direction: single or double
Type: DC or pulse
Minimum programmable value to create list sweep waveform: B2901A/B2902A: 20 s with 1 s resolution B2911A/ B2 912 A : 10 s with 1 s resolution
Digitizing/Sampling Measurement
Digitizing/sampling measurement can be performed by appropriate setting of measure speed and trigger interval.
Minimum trigger interval (maximum speed of measurement): B2901A/B2902A: 20 s (50000 points/s) B2911A/ B2 912 A : 10 s (100000 points/s)
Data Buffers
Data buffers are array of measurement elements. Each elements hold the following values.
Measured voltage
Measured current
Measured resistance
2-18 Keysight B2900 User’s Guide, Edition 5
Introduction
Operation and Functions
Time stamp
Measurement status
Source setting Maximum buffer size: 100000 points/channel

Program and Interface Capabilities

Programming Language
The B2900 supports SCPI (Standard Commands for Programmable Instruments).
Default command set: Supports all of the advanced features of the B2900.
Conventional command set: Supports industry standard conventional SCPI command set for basic compatibility.
Program Memory
Program memory allows you to s t ore lo ng s t ring s o f S CPI co mmand lines once into the built-in volatile memory and then recall those strings multiple times while the program is executing us ing a single SCPI command . By storing the command strings in memory, the time that would have been spent sending those same commands over a communication bus is eliminated. For tests that utilize lots of repeated code (such as subroutines), program memory can dramatically reduce test times.
Maximum number of characters of program name: 32 with alphabets, numbers, hyphens, and underscores
Maximum memory size: 100 KB (2500 lines typical)
LXI
LXI core conformant. The B2900 follows specified LAN protocols and adhere to LXI requirements such as a built-in Web control server and IVI-COM driver.
Ethernet: 10/100Base-T
USB2.0: USB-TMC488 protocol (rear 1)
GPIB: IEEE-488.2 compliant
USB File System
USB 2.0 high-speed mass storage (MSC) class device (front 1)
Keysight B2900 User’s Guide, Edition 5 2-19
Introduction Operation and Functions
Digital I/O Interface
Connector: 25-pin female D
Input/output pins: 14 open drain I/O bits
Absolute maximum input voltage: 5.25 V
Absolute minimum input voltage: -0.25 V
Logic low maximum input voltage: 0.8 V
Logic high minimum input voltage: 2.0 V
Maximum source current: 1 mA, Vout = 0 V
Maximum sink current: 50 mA, Vout = 5 V
5 V power supply pin: Limited to 600 mA, solid state fuse protected
Safety interl ock pin: One active high pin and one active low pin. Activation of both pin enables
output voltage > 42 V.
2-20 Keysight B2900 User’s Guide, Edition 5
Introduction

Software and Drivers

Software and Drivers
Keysight B2900A Quick I/V Measurement Software The B2900 includes PC-based Keysight B2900A Quick I/V Measurement
Software. This powerful software makes it easy to quickly setup and perform I/V measurements and to display the measurement data in a table or graph without the need to perform any programming. The software allows you to control up to four channels over a GPIB or LAN connection, or one B2900 via a USB connection.
Operating System: Windows 7 (64 bit/32 bit), XP SP3 (32 bit) Other requirements: Microsoft .NET framework 4.0 or later and Keysight IO
Libraries 16.0 or later
Graphical Web Interface The Graphical W eb Interface is a web brow ser based instrument control panel. It
enables you to set up and perform a measurement easily and q uickly f rom a web browser using the built-in web server. This allows you to control one B2900 over a LAN connection.
IVI-C or IVI-COM drivers Compatible with W in dows 7 (64 bit/32 b it), XP SP 3 (32 bit ) IO Librari es 16.0 o r
later. Supports Keysight VEE, Microsoft Visual Studio (Visual Basic, Visual C++, Visual C#), National Instruments LabWindows and LabVIEW.
LabVIEW driver (VI) National Instruments LabVIEW 7.0 or later. LabView drivers are available at
NI.COM.
NOTE For the latest system requirements, go to www.keysight.com/find/b2900a and get
the information.
Keysight B2900 User’s Guide, Edition 5 2-21
Introduction Accessories

Accessories

Furnished Accessories

The Keysight B2900 is furnished with the following accessories.
Quick Reference, 1 ea.
Product Reference CD-ROM, 1ea.
Keysight IO Libraries CD-ROM, 1ea.
Keysight EasyEXPERT for PC Software CD, 1ea.
License-to-use for EasyEXPERT
Power cord, 1 ea.
USB cable, 1 ea. The Product Reference C D-RO M st or es Ke ys ight B 29 00A Qui ck I/ V Meas uremen t
Software, sample programs, user manuals, and application notes.

Available Accessories

For the available accessories for the Keysight B2900 series, see the catalog Accessories for the Keysight B2900 Precision Instrument Family.
T o get the latest catalog, go to www.keysight.com/find/b2901a and click “Technical Support” and “Catalogs”.
2-22 Keysight B2900 User’s Guide, Edition 5

Options

Table 2-8 lists the options for the Keysight B2900.
Table 2-8 Options
Option Description
Calibration
UK6 Commercial calibration certificate with test data
Printed Manual (User’s Guide)
AB0 Traditional Chinese AB2 Simplified Chinese
Introduction
Options
A6J ANSI Z540 compliant calibration
0B0 No printed manual (printed manual is not shipped)
ABA English
ABJ Japanese
Rack Mount Kit
1CM Rack mount kit
Keysight B2900 User’s Guide, Edition 5 2-23
Introduction Options
2-24 Keysight B2900 User’s Guide, Edition 5

3 Installation

Installation
This chapter describes how to install Keysight B2900 and accessories and consists of the following sections.
“Inspecting the Shipment”
“Installing the Keysight B2900”
“Maintenance”
“Connecting a DUT”
“Installing the Interlock Circuit”
“Connecting to the Interfaces”
“Communicating Over the LAN”
“Using Digital I/O”
WARNING To avoid electrical shock and instrument damage, turn the instrument off
before connecting or disconnecting measurement cable. Mettez la machine hors tension pour fixer le connecteur ou pour retirer le
connecteur de l’unité centrale.
WARNING There are potentially hazardous voltages (210 V) present at the High Force,
High Sense, and Guard termin als of this instrument. To prevent electrical shock, the following safety precautions must be observed during the use of the instrument.
Use a three-conductor AC power cord to connect the cabinet (if used) and the instrument to an electrical ground (safety ground).
If an interlock circuit is not installed in your test fixture or connection interface, you must install and connect the interlock circuit that opens the interlock terminal when the shielding box access door is opened.
If you change the connection interface, test fixture, prober, and such, connect an interlock cable to the one actually used.
Confirm periodically that the interlock function works normally.
Before touching the connections on the High Force, High Sense, and Guard terminals, turn the instrument off and discharge any capacitors. If you do not turn the instrument off, complete all of the following items, regardless of the instrument settings.
•Press the
3-2 Keysight B2900 User’s Guide, Edition 5
On/Off
switch, and confirm that the switch turns off.
Installation
Confirm that the
Open the fixture cover or the shielding box access door (open interlock).
Discharge any capacitors connected to a channel.
Warn persons working around the instrument about dangerous conditions.
WARNING Une tension dangereuse (max. pour; 210 Vdc) émanant du dispositif
l'instrument peut être sortie aux bornes High Force, High Sense et Guard. Les précautions suivantes doivent être obserées contre commotion électrique accidentelle.
Utilisez un cordon d'alimentation CA à trois connecteurs pour connecter la cabine (si utilisée) et l'instrument à la mise électrique à la terre (sol de sécurité).
Si un circuit de sécurité n'est pas installé dans votre test d'installation ou dans votre interface de connexion, vous devez installer et connecter le circuit de sécurité qui ouvre la borne d'enclenchement lorsque la porte d'accès à la protection de la boîte est ouverte.
Si vous changez l'interface de connexion, un test d'installation, la sonde, ou toute autre élément, connectez un cordon d'enclenchement à celui utilisé actuellement.
Vérifiez régulièrement que la fonction de verrouillage fonctionne normalement.
On/Off
switch does not turn red.
Avant de toucher les connexions des bornes High Force, High Sense et Guard, éteignez l'instrument et déchargez tous les condensateurs. Si vous n'éteignez pas l'appareil, complétez tous les éléments suivants, indépendamment des réglages de l'appareil.
Appuyez sur le bouton On/Off, et confirmez que le signal lumineux est
éteint.
Vérif iez que le bouton On/Off ne devienne pas rouge.
Ouvrez le couvercle d'appareil ou la protection du boîtier de la porte
d'accès (verrouillage ouvert).
Déchargez tous les condensateurs connectés au réseau.
Informez les personnes se trouvant à proximité de l'appareil à propos des conditions dangereuses.
Keysight B2900 User’s Guide, Edition 5 3-3
Installation Inspecting the Shipment

Inspecting the Shipment

Perform the following inspections when the Keysight B2900 and accessories arrive at your site.
1. Before unp acking any component, ins pect all boxes for any signs of damage that might have occurred during shipment, such as:
•dents
scratches
•cuts
water marks If you suspect any damage, contact your nearest Keysight Sales and Support
Office.
2. When you open the boxes that contain the B2900 and accessories, check the components against the contents lists attached to the boxes.
If anything is missing, contact your nearest Keysight Sales and Support Office.
3. Verify the operation of B2900 as described in “Checking the Operation of
Keysight B2900”.
If any problem occurs, contact your nearest Keysight Sales and Support Office.

Checking the Operation of Keysight B2900

1. Make sure that the standby switch is set to off.
2. Connect the power cord from the B2900’s rear panel AC input connector (receptacle) to an AC power outlet at your site.
3. Press the standby switch to turn on the instrument. The initialization screen will appear on the B2900’s front panel display, and the
power-on self-test is automatically executed. If the B2900 is operating normally, the front panel LCD displays as shown in
Figure 3-1 or 3-2.
3-4 Keysight B2900 User’s Guide, Edition 5
Figure 3-1 Display Example for 1-ch Model
Figure 3-2 Display Example for 2-ch Model
Installation
Inspecting the Shipment

Checking for Errors

Errors can be checked as described below.
1. Press the More > Sys tem > Error > Log function keys. This op ens the Error Log dialog box .
2. Check the errors displayed on the dialog box. If no error is detected, “0, No Error” is displayed.
3. Press the OK softkey to close the dialog box.
Keysight B2900 User’s Guide, Edition 5 3-5
Installation

Installing the Keysight B2900

Installing the Keysight B2900
This section describes information you must pay attention to when installing the Keysight B2900.
“Safety Considerations”
“Environment”
“Connecting the Power Cord”
“Setting the Power Line Frequency”
“Setting Date and Time”
“Bench Installation”
“Rack Installation”

Safety Considerations

Refer to the Safety Summary page at the beginning of this guide for general safety information. Before installation or operation, check the B2900 and review this guide for safety warnings and instructions. Safety warnings for specific procedures are located at appropriate places throughout this guide.

Environment

WARNING Do not operate the instrument in dusty environment, or in the presence of
flammable gases, corrosive gases, or fumes. Ne pas utiliser l’appareil dans un endroit poussiéreux, ou en présence de gaz
inflammables, corrosifs ou de fumée.
Environmental conditions for the B2900 are described below. The B2900 is designed for use in indoor facilities.
Temperature range
Operating 0 C to +55 C Storage -30 C to +70 C
Humidity range
Operating 30 % to 80 %RH, non-condensing
3-6 Keysight B2900 User’s Guide, Edition 5
Installation
Installing the Keysight B2900
Storage 10 % to 90 %RH, non-condensing
Altitude
Operating 0 m to 2,000 m (6,561 ft.) Storage 0 m to 4,600 m (15,092 ft.)
Power requireme nt: 100-24 0 V (±10 %), 50 /60 Hz
Maximum volt-amps (VA): 250 VA
•Dimensions Without handle and feet: 88 mm (H) 213 mm (W) 450 mm (D) Bench configuration (with handle and feet): 180 mm (H) 260 mm (W) 480
mm (D)
Fan cools the instrument by drawing air through the sides and exhausting it out the back. The instrument must be installed in a location that allows sufficient space at the sides and back of the instrument for adequate air circulation.

Connecting the Power Cord

WARNING FIRE HAZARD: Use only the power cord supplied with your instrument.
Using other types of power cord may cause overheating of the power cord, resulting in fire.
SHOCK HAZARD: The power cord provides the chassis ground through a third conductor. Be sure to connect to a three-conductor type power outlet with the correct pin grounded.
RISQUE D’INCENDIE : utilisez uniquement le câble d'alimentation fourni avec votre appareil. L’ utilisation d’autres types de câble d’alimentation peut provoquer une surchauffe du câble d’alimentation et provoquer un incendie.
RISQUE DE CHOC ÉLECTRIQUE: le câble d’alimentation fournit la masse du châssis par le biais d’un troisième conducteur. Assurez-vous de connecter la prise d'alimentation de type trois conducteurs av ec la broche correcte mise à la terre.
NOTE The detachable power cord may be used as an emergency disconnecting device.
Removing the power cord will disconnect AC input power to the instrument.
Keysight B2900 User’s Guide, Edition 5 3-7
Installation Installing the Keysight B2900
Option 900
Plug: BS 1363/A, 250 V, 10 A
PN: 8120-4420
Option 904
Plug: NEMA 6-15P, 250 V, 10 A
PN: 8120-3996
Option 918
Plug: JIS C 8303, 125 V, 12 A
PN: 8121-0743
Option 922
Option 901
Plug: AS/N ZS 3112, 250 V, 10 A
PN: 8120-441 9
Option 906
Plug: SEV 1011, 250 V, 10 A
PN: 8120-441 6
Option 919
Plug: Israel SI 32, 250 V, 10 A
PN: 8121-072 4
Option 923
Option 902
Plug: IEC 60277-1, 250 V, 10 A
PN: 8121-1226
Option 912
Plug: SB 107-2 -D1, 250 V, 10 A
PN: 8121-1655
Option 920
Plug: IRAM 2073, 250 V, 10 A
PN: 8121-0725
Option 927
Option 903
Plug: NEMA 5-15P, 125 V, 10 A
PN: 8120-6825
Option 917
Plug: IS 1293 and IS 6538, 250 V, 10 A
PN: 8121-1690
Option 921
Plug: CEI 23-16, 250 V, 10 A
PN: 8121-0722
Option 930
Plug: GB 1002 figure 3, 250 V, 10 A
PN: 8120-8376
Option 931
Plug: CNS 10917-2, 125 V, 10 A
PN: 8121-1635
Plug: SANS 164-1, 250 V, 10 A
PN: 8121-056 4
Option 932
Plug: CS 0017, 250 V, 10 A
PN: 8121-163 8
Plug: NEMA WD-6, 250 V, 10 A
PN: 8120-0674
Plug: NBR 14136, 250 V, 10 A
PN: 8121-1809
3-8 Keysight B2900 User’s Guide, Edition 5
Installation
Installing the Keysight B2900
Connect the power cord to the IEC 60320 connector on the rear of the instrument. If the wrong power cord was shipped with your instrument, contact Keysight Technologies.
The AC input on the back of your instrument is a universal AC input. It accepts nominal line voltages in the range of 100 to 240 VAC.

Setting the Power Line Frequency

The power line frequency must be set properly for the AC power at your site. Press the following function keys to set the frequency to 50 Hz or 60 Hz.
For settin g to 50 Hz: More > System > PLC > 50 Hz
For settin g to 60 Hz: More > System > PLC > 60 Hz

Setting Date and Time

You can set the date and time on the Date and Time dialog box opened by pressing System > Info. > Date/Time function keys.

Bench Installation

Do not block the air intake at the sides and the exhaust at the rear of the B2900. Minimum clearances for bench operation are 5 mm along the sides and 100 mm along the back.
For easier display viewing and measurement terminal access, you can tilt up the front of the instrument by turning the handle. To adjust the handle, grab the handle by the sides and pull outward. Then, rotate the handle to the desired position.
Keysight B2900 User’s Guide, Edition 5 3-9
Installation
1
1
2
3
Installing the Keysight B2900

Rack Installation

NOTE Use the Rack Mount kit (Option 1CM) to rack mount the Keysight B2900.
Keysight B2900 can be mounted in a 19-inch EIA rack cabinet. It is designed to fit in two rack-units (2U) of space.
Remove the front and rear rubber bumpers and the hand le befor e rack mou nting the B2900.
Do not block the air intake at the sides and the exhaust at the rear of the B2900.
Removing the Bumper
Stretch a corner of the rubber bumper and slide it off.
Removing the Handle
1. Grab the handle by the sides and pull outward. This will allow you to rotate the handle.
2. Rotate the handle to a vertical position. Then, position the instrument horizontally.
3. Pull outward and then lift the handle upward.
CAUTION When reattaching the handle, pay attention to its direction. Incorrect attachment
may damage the hardware.
3-10 Keysight B2900 User’s Guide, Edition 5
Installation

Maintenance

Maintenance
Maintenance should be performed periodicall y to keep the Keysigh t B290 0 in good condition. If problems arise, contact your nearest Keysight Sales and Support Office.
“Cleaning”
“Self-test”
“Self-calibration”
“Calibration”

Cleaning

WARNING SHOCK HAZARD: To prevent electric shock, unplug the B2900 before
cleaning. Risques de choc électrique : afin d’éviter tout choc électrique, débranchez le
B2900 avant d’effectuer le nettoyage.
Use a dry cloth or a cloth slightly dampened with water to clean the external case parts. Do not use detergents or chemical solvents. Do not attempt to clean internally.

Self-test

Keysight B2900 provides the self-test function to check the operation. The self-test is automatically performed when the instrument is turned on. It is recommended to perform the self-test for the following condition or purpose. Before performing the self-test, turn the channel output off and disconnect test leads and cables from the terminals.
If a channel is in the lock condition due to over temperature In this condition, the Emer gency dial og box is disp layed on the gray background
screen, however this is not shown in the remote status if the remote display is set to OFF. And the ERR indicator turns on and the On/Off switch is not effective.
Perform the self-test to unlock the channel. The channel can be used soon if the self-test does not report any problem.
If you feel that the instrument may be defective
Keysight B2900 User’s Guide, Edition 5 3-11
Installation Maintenance
For preventive maintenance
To perform self-test
To perform self-calibration
The following procedure shows how to perform the self-test.
1. If it is in the remote state, press the Cancel / Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the channel terminals.
4. Press the following functio n keys. More > System > Cal/Test > Self-Test Confirmation dialog box is opened.
5. Press the OK key. This starts the self-test.

Self-calibration

Keysight B2900 provides the self-calibration function to maintain the measurement performance. If the environmental temperature changes 3 C or more, perform the self-calibration. This is effective for the accurate measurements by minimizing the effect of thermal drift. The self-calibration must be performed after warming-up of 60 minutes. Before performing the self-calibration, turn the channel output off and disconnect test leads and cables from the terminals.
The following procedure shows how to perform the self-calibration.
1. If it is in the remote state, press the Cancel / Local key.
2. Press the On/Off switch and confirm that the switch turns off.
3. Disconnect test leads and cables from the channel terminals.
4. Press the following functio n keys. More > System > Cal/Test > Self-Cal Confirmation dialog box is opened.
5. Press the OK key. This starts the self-calibration.
3-12 Keysight B2900 User’s Guide, Edition 5
Installation
Maintenance

Calibration

Calibration and adjustments must be performed periodically so that the instruments satisfy the specifications, and keep a good condition. It is recommended to perform the calibration once a year at least. For the calibration and adjustments , contact your nearest Keysight Sales and Support Office. Trained service personnel will perform the calibration and adjustments.
Keysight B2900 User’s Guide, Edition 5 3-13
Installation

Connecting a DUT

Connecting a DUT
This section describes how to connect a device under test (DUT) to the Keysight B2900 source/measure terminals.
NOTE Set the channel output off when changing the connections. If not, the DUT may be
damaged. To set the channel output off, press the
LED turns off.
This section includes the following descriptions.
“2-Wire Connections or 4-Wire Connections”
“Floating”
“U sing Test Leads”
“Using the N1295A Test Fixture”
“Using the 16442B Test Fixture”
“Guarding”
“Performing Low Current Measurements”
NOTE Connecting the interlock circuit
B2900 provides an interlock f unction to pr event the user from receiving an electrical shock from high voltages over 42 V. If the Digital I/O interlock terminal is open, B2900 cannot apply a high voltage.
For high voltage measurement, connect the interlock terminal to the interlock circuit as described in “Installing the Interlock Circuit” on page 3-24.
WARNING To avoid touching the end of the extension cable or the terminal area with the
DUT, cover over the conductors with insulator. Also it is important to protect the terminal area by using the grounded shield cover and such.
On/Off
switch and confirm that the switch
To prevent electrical shock and DUT damage, do not connect or disconnect the DUT while the instrument is applying voltage or current.
When you touch the DUT after measurement, devise a countermeasure of residual charge and heat to prevent electrical shock and burn. Use glove and any tool. Al so have enough time for discha rge and radiation.
3-14 Keysight B2900 User’s Guide, Edition 5
Installation
Connecting a DUT
Afin d'éviter de toucher l'extrémité du câble d'allongement ou l'aire de la borne avec l'appareil mis sous tension (AST), couvrez les conducteurs avec l'isolant. En outre, il est important de protéger la zone de la borne en utilisant le couvercle d'écran à la mise à terre, ou tout autre élément.
Afin d’éviter toute décharge électrique et dommage MST, ne branchez ou déconnectez pas la sortie MST alors que la source de sortie est appliquée.
Lorsque vous touchez le MST après la mesure, élaborez une contre-mesure de la charge résiduelle et du chauffage afin d'éviter tout choc électrique et toute brûlure. Utilisez des gants et des outils. Prévoyez également du temps pour la décharge et la radiation.

2-Wire Connections or 4-Wire Connections

When connecting a DUT, you can choose the connection type either 2-wire connections or 4-wire connections.
If you want to simplify the connections, use 2-wire connections by connecting the Force terminals only and opening the Sense terminals. The Force terminals can be used to apply and measure DC voltage or current.
To make 4-wire connections, remote sensing, well known as Kelvin connections, use both Force and Sense terminals. Connecting the Force and Sense lines together at the terminal of the DUT minimizes the measurement error caus ed by the residu al resistance of the test leads or cables. This connection is effective for low resistance measurements and high current measurements.
To specify the connection type (Sensing Type), perform the following procedure.
1. Press the Config > Source > Connection function keys. This opens the Output Connection dialog box.
2. Set the Sensing Type on this dialog box. Choose 2-WIRE for 2-wire connections, or 4-WIRE for 4-wire connections.
3. Press the OK softkey. For 4-wire connections, the following indicator appears on the Single view or the Dual view.
4-wire connection indicator:
Keysight B2900 User’s Guide, Edition 5 3-15
Installation
Connecting a DUT
Figure 3-3 Dual View, 4-Wire Connection Setup for Channel 1

Floating

With the default setting, Low Force terminal is connected to the chassis ground. However, it can be internally disconnected from the ground for making the floating condition. This setup is used for connecting the source channels in series.
To specify the low terminal status, perform the following procedure. Setting is not saved. Turning off the instrument sets the status to the grounded.
1. Press the Config > Source > Connection function keys. This opens the Output Connection dialog box.
2. Set the Low Terminal State on this dialog box. It must be FLOATING for the floating status, or GROUNDED for the grounded status.
3. Press the OK softkey. For the floating status, the following indicator appears on the status information
area. And then the Low Force and Low Sense terminals can be connected to the maximum of 250 V.
Channel 1 floating status indicator: Channel 2 floating status indicator:
3-16 Keysight B2900 User’s Guide, Edition 5
Installation
Connecting a DUT
Figure 3-4 Dual View, Floating Status for Channels 1 and 2
CAUTION Do not apply current to the chassis ground. Doing so will damage the B2900.
WARNING To prevent electrical shock, do not touch any of measurement circuit at any
time while a floating measurement is in progress. Also use accessories that comply with IEC 6101 0-2-03 1. Al l termin als and th e extende d cond uctors must be isolated by using insulation caps, sleeves, etc.
Afin d’éviter toute décharge électrique, ne touchez aucune mesure de circuit à tout moment lorsque la mesure de flotte est en cours. Utiliser également des accessoires qui sont conformes à la norme IEC 61010-2-031. Toutes les bornes et les conducteurs prolongés doivent être isolés en utilisa nt des bouchons d'isolation, des manchons, etc.
Keysight B2900 User’s Guide, Edition 5 3-17
Installation
High Sense
High Force
Low Force
Low Sense
DUT
High Force
Low Force
DUT
2-wire connections
4-wire connections
Connecting a DUT

Using Test Leads

B2900 source/measure terminals use banana jacks. For connecting a DUT, the following test leads are available. Figure 3-5 shows the connections for two-terminal device measurements.
Keysight U8201A Combo Test Lead Kit Couple of test leads, test probes, alligator clips, SMT grabbers, fine-tips test
probes, and banana plugs, CAT III 1000 V, 15 A maximum Two kits are required for 4-wire connections.
Keysight 11059A Kelvin Probe Set Test leads for 4-wire connections, 42 V maximum
Figure 3-5 Connect ing a Two-Terminal Device
3-18 Keysight B2900 User’s Guide, Edition 5
Installation
Bad connection
Good connection (no space)
1
1
2
3
4
3
2
4
± 42V 1.05 A Max
Low Force
High Force
DUT
Connecting a DUT

Using the N1295A Test Fixture

Keysight N1295A is a test fixture which has four triaxial connectors and which supports two 2-wire connections.
Requirements Keysight N1295A test fixture furnished with pin clip wire 4 ea. and pin plug
wire 2 ea.
Keysight N1294A-001 banana to triaxial adapter for 2-wire connections
Triaxial cable, 2 ea. for one 2-wire connection
Connections 1. Attach the banana to triaxial adapter to the B2 900 source/measure ter minals. See
Figure 3-6. If there is some space as “Bad connection”, the contact is not
enough.
2. Connect the triaxial cables between the adapter and N129 5A as shown in Figure
3-7.
3. Connect the DUT between terminals 1 and 2 for the example in Figure 3-7.
Figure 3-6 Attaching Banana to Triaxial Adapter
Figure 3-7 Keysight N1295A Connection Example
NOTE Shielding
N1295A is equipped wit h a cover. T o minimize the e ffect of am bient noise, close the cover when performing measurements.
Keysight B2900 User’s Guide, Edition 5 3-19
Installation Connecting a DUT
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent damage.
Keysight N1294A-001: 250 V maximum, 42 V maximum for connecting N1295A
Keysight N1294A-002: 250 V maximum, 42 V maximum for connecting N1295A
Keysight N1295A: 42 V, 1.05 A maximum

Using the 16442B Test Fixture

Keysight 16442B is a test fixture which has six triaxial connectors for source/measure unit (SMU), the GNDU connector for the ground unit of a parameter/device analyzer, the Intlk connector for the interlock control, and six coaxial connectors for other instruments. See Keysight 16442B User’s Guide for details and accessories of the 16442B.
Requirements Keysight 16442B test fixture furnished with pin clip wire, pin plug wire, etc.
Keysight N1294A-001 banana to triaxial adapter for 2-wire connections
Keysight N1294A-002 banana to triaxial adapter for 4-wire connections
Triaxial cable, 2 ea. for 2-wire connections, and 3 ea. for 4-wire connections
NOTE Shielding
16442B is equipped with a cover. To minimize the ef fect of ambient noise, close the cover when performing measurements.
CAUTION Maximum voltage and current
The test fixture and adapters must be used under the following limitations to prevent damage.
Keysight N1294A-001: 250 V maximum, 200 V maximum for connecting 16442B
Keysight N1294A-002: 250 V maximum, 200 V maximum for connecting 16442B
Keysight 16442B: 200 V, 1 A maximum for SMU input
3-20 Keysight B2900 User’s Guide, Edition 5
Installation
N1294A-001 Adapter for 2-wire connections
N1294A-002 Adapter for 4-wire connections
to SMU 2 Force
to SMU 1 Force
to SMU 1 Force
to SMU 1 Sense
to SMU 3 Sense
(or GNDU)
Low Force/ Low Sense
High Sense
High Force
Low Force
High Force
SMU(±200V= 1A Max)
SenseForce
Guard Guard
1
SenseForce
Guard Guard
2
SenseForce
Guard Guard
3
DUT
SMU(±200V= 1A Max)
SenseForce
Guard Guard
1
SenseForce
Guard Guard
2
SenseForce
Guard Guard
3
2-wire connections
4-wire connections
DUT
(or GNDU Sense)
(or GNDU Force)
Connecting a DUT
Connections 1. Attach the banana to triaxial adapter to the B2 900 source/measure ter minals. See
Figure 3-6. If there is some space as “Bad connection”, the contact is not
enough.
2. Connect the triaxial cables between the adapter and the 16442B. Figure 3-8 shows a connection example.
3. Connect the DUT on the 16442B wiring panel by using the 16442B’s furnished accessories, socket module, blank board, and wire suitable for the DUT.
Figure 3-9 shows the connection example for the connections of Figure 3-8.
For 2-wire connections, connect the SMU 1 Force terminal to a DUT terminal, and connect the SMU 2 Force terminal to the other DUT terminal.
For 4-wire connections, connect the SMU 1 Force and Sense terminals to a DUT terminal. And connect the SMU 3 Sense and Guard terminals to the other DUT terminal if the adapter’s Low Force/Low Sense connector is connected to the SMU 3 Sense connector. The Low Force signal appears on the Guard terminal.
Figure 3-8 Connection Example between N1294A and 16442B
Figure 3-9 Connection Example of 16442B Wiring Panel
Keysight B2900 User’s Guide, Edition 5 3-21
Installation
Connecting a DUT
NOTE Performing high voltage measurement
When the Digital I/O interlock terminal is open, B2900 cannot apply high voltage over 42 V. To perform high voltage measurements, B2900 must be connected to the interlock circuit installed in 16442B.
Prepare the N1294A-011 or 012 interlock cable, and connect it between the B2900 Digital I/O connector and the 16442B Intlk connector. B2900 can apply a high voltage when the test fixture cover is closed.
WARNING Hazardous voltage, instrument maximum output voltage may appear at the
High Force, High Sense, and Guard terminals if the fixture cover is closed. To prevent electrical shock, do not expose these lines.
Une tension dangereuse, une tension de sortie maximale de l'appareil peut apparaître aux bornes High Force, High Sense et Guard si le couvercle de l'équipement est fermé. Afin d’éviter toute décharge électrique, n’exposez pas ces lignes.
3-22 Keysight B2900 User’s Guide, Edition 5

Guarding

A
Buffer
Guard
High Force
DUT
Guarding reduces the leakage current between the instrument and a DUT. This is important when you measure a low current.
Figure 3-10 shows the theory of guarding. The buffer amplifier (1) keeps the
potential of the Guard conductor at the same potential as the High Force conductor, so that the current does not flow between the High Force and Guard conductors. Therefore, the current measured by the instrument is the same as the current at the DUT terminal because no current is leaked.
For the connection examples shown in Figu res 3-7 and 3-8, the guard is extended to the test fixture input connectors with triaxial cables.
Figure 3-10 Guarding
Installation
Connecting a DUT
CAUTION Never connect the Guard terminal to any output, including circuit common, chassis
ground, or any other guard terminal. Doing so will damage the B2900.

Performing Low Current Measurements

With the default setting, 10 nA and 100 nA measurement ranges are not used. For the low current measurements, the measurement range settings must be changed.
To change the measurement range
1. Press the View key to display the Single view. If the Range parameters are not displayed in the lower half of the Single view,
press the More, Hide Sweep, Hide Pulse, or Hide Trigger assist key to display the Range parameters.
2. Change the setting of the Measure Amps field of the Range parameters. Set AUTO for the auto ranging or FIXED for the fixed range. Set 10 nA or 100 nA for the minimum or fixed measurement range.
Keysight B2900 User’s Guide, Edition 5 3-23
Installation
Mechanical switches
Shielding box
LED
Access door
D-sub connector
Interlock circuit
B2900 Digital I/O
15, 19
14
17, 18
16
24, 25
Dsub cable
15, 19
22, 23
16, 17
24
14
Negative
Positive
or
22
R
R: 500 W to 1 kW
25
R

Installing the Interlock Circuit

Installing the Interlock Circuit
The interlock circuit is a simple electric circuit, as shown in Figure 3- 11. The circuit electrically opens when an access door is opened, and closes when the door is closed.
B2900 cannot apply high voltages over 42 V when the Digital I/O interlock terminal is open. To perform high voltage measurements, the B2900 interlock terminal must be connected to the interlock circuit installed in the measurement environment such as the shielding box. The interlock circuit is important and necessary to prevent electrical shock when the user touches the measurement terminals.
WARNING Hazardous voltage, instrument maximum output voltage may appear at the
High Force, High Sense, and Guard terminals if the fixture cover is closed. To prevent electrical shock, do not expose these lines.
Une tension dangereuse, une tension de sortie maximale de l'appareil peut apparaître aux bornes High Force, High Sense et Guard si le couvercle de l'équipement est fermé. Afin d’éviter toute décharge électrique, n’exposez pas ces lignes.
Figure 3-11 Interlock Circuit
Requirements LED (Keysight part number 1450-0641 or equivalent), 1 ea.
Mechanical switch (Keysight N1254A-402 or equivalent), 2 ea.
Resistor, 500 to 1 k (choose lower one to get higher output of LED), 1 ea.
3-24 Keysight B2900 User’s Guide, Edition 5
Installation
Installing the Interlock Circuit
25-pin D-sub connector (for wiring, mounted on the shielding box), 1 ea.
Connection wire
25-pin D-sub cable, 1 ea.
Procedure 1. Mount two mechanical switches onto your shielding box, so that the switches
close when the access door is closed, and open when the door is opened. See
Figure 3-13 for the switch dimensions.
2. Mount an LED onto your shielding box. See Figure 3-12 for the LED dimensions.
The LED is used as a high voltage indicator which is lit when B2900 is in the high voltage output s tatus.
3. Mount a D-sub connector onto your shielding box.
4. Use a wire and sh ort t he fo llo wing pi n s of the D-su b connector. S ee Figu re 3-11 for a connection example.
For the negative logic: Pins 24 and 25 For the positive logic: Pins 16 and 17
5. Use a wire and connect the two switches in series between the following pins of the D-sub connector. See Table 3-1 for the Digital I/O pin assignment.
For the negative logic: Pins 16 and some ground pins The ground pins are pin numbers 15 and 17 to 21. For the positive logic: Pins 24 and some +5 V pins The +5 V pins are pin numbers 22, 23, and 25.
6. Use a wire and connect the L ED between the pin 14 and some groun d pins of the D-sub connector. Then set the pin 14 to HIGH VOLTAG E LAMP. See “DIO
Configuration dialog box” on page 4-48.
7. Connect a resistor between the pin 14 and an unused +5 V pin.
Keysight B2900 User’s Guide, Edition 5 3-25
Installation
6
10
11
5
5.6
5
Units: mm
Anode (+)
Cathode (-)
3.1
35.6
8.1
14
NC
NO
COM
6.4
22.2
27.8
2.8
6.3 15.9 18.8
4.7510.3 4.3
2.8
2.8
3.2
10.3
2.8
3.1
Switch off
Units: mm
15.3
Installing the Interlock Circuit
Figure 3-12 Dimensions of the LED (Keysight part number 1450-0641)
Figure 3-13 Dimensions of the Interlock Switch (Keysight N1254A-402)
3-26 Keysight B2900 User’s Guide, Edition 5
Installation
Connect to GPIB interface card installed in PC.
Connect to GPIB port on instrument.
InstrumentInstrument
Instrument
PC
GPIB cable

Connecting to the Interfaces

Connecting to the Interfaces
CAUTION Electrostatic discharges greater than 1 kV near the interface connectors may cause
the unit to reset and require operator intervention.
B2900 supports GPIB, LAN, and USB interfaces. All three interfaces are live at power-on. Connect your interface cable to the appropriate interface connector. Information on configuring the interfaces can be found later in this section.
The front panel LAN indicator is lit when the LAN port is connected and configured.
B2900 provides Ethernet connection monitoring. With Ethernet connection monitoring, the instrument’s LAN port is continually monitored and automatically reconfigured.

GPIB/USB Interfaces

NOTE For detailed information about GPIB and USB interface connections, refer to
Connectivity Guide installed with the Keysight IO Libraries.
The following steps will help you quickly get started connecting your instrument to the GPIB (General Purpose Interface Bus). The following figure illustrates a typical GPIB interface system.
Keysight B2900 User’s Guide, Edition 5 3-27
Installation
Connect to USB port on PC.
Connect to USB port on instrument.
Instrument
PC
USB cable
Connecting to the Interfaces
1. If you hav e not al read y d one s o, in stal l t he Keys i gh t IO Librari es Sui te fro m t he CD shipped with your product.
2. If you do not have a GPIB interface card installed on your computer, turn off your computer and install the GPIB card.
3. Connect your instrument to the GPIB interface card using a GPIB interface cable.
4. Use the Connection Expert utility from the Keysight IO Libraries Suite to configure the installed GPIB interface card’s parameters.
5. B2900 is shipped with its GPIB address set to 23. To view or change the GPIB address, press the More function key, then the I/O > GPIB softkeys. The GPIB Configuration dialog box appears.
To change the GPIB address, use the rotary knob or the arrow keys, and press OK to set the value.
6. You can now use the Interactive IO within Connection Expert to communicate with your instrument, or program your instrument using various programming environments.
The following steps will help you quickly get started with connecting your USB-enabled instrument to the USB (Universal Serial Bus). The following figure illustrates a typical USB interface system.
1. If you have n ot already done so , install Ke ysigh t IO Libraries Suite from the CD
2. Connect the USB device port located on the back of your instrumen t to the USB
3. With the Connection Expert utility from Keysight IO Libraries Suite running,
3-28 Keysight B2900 User’s Guide, Edition 5
shipped with your product.
port on your computer.
the computer will automatically recognize the instrument. This may take several seconds. When the instrument is recognized, your computer will display the VISA alias, IDN string, and VISA address.
Installation
To network interface card (NIC).
Connect to LAN port on instrument.
Instrument
PC
LAN cable
To site LAN
Connecting to the Interfaces
You can also view the instrument’s VISA address from the fro nt panel. Press the More function key, then the I/O > USB softkeys. The VISA address is shown in the USB Status dialog box.
4. You can now use the Interactive IO within Connection Expert to communicate with your instrument, or program your instrument using various programming environments.

LAN Interface

NOTE For detailed information about LAN interface connections, refer to Connectivity
Guide installed with the Keysight IO Libraries.
The following steps will help you quickly get started with connecting and configuring your instrument on a local area network.
Connecting to a Site LAN
A site LAN is a local area network in which LAN-enabled instruments and computers are connected to the network via routers, hubs, and/or s witches. They are typically large, centrally-managed networks with services such as DHCP and DNS.
1. If you hav e not al read y d one s o, in st all the Ke ysi ght IO Libraries Suite from the CD shipped with your product.
2. Connect the instrument to the site LAN. The factory-shipped instrument LAN settings are configured to automatically obtain an IP address from the network using DHCP (DHCP is enabled). Note that this may take up to one minute. The DHCP server will register the instrument’s hostname with the dynamic DNS server. The hos tname as well as the IP address can then be used to commu nicate with the instrument. The front panel LAN indicator will turn green when the LAN port has been configured successfully, or turn red if the configuration fails.
Keysight B2900 User’s Guide, Edition 5 3-29
Installation Connecting to the Interfaces
NOTE If you need to manually configure any instrument LAN settings, see “LAN
Configuration dialog bo x” on page 4-48 for info rmation about configur ing the LAN
settings from the front panel of the instrument.
3. Use the Connection Expert utility from Keysight IO Libraries Suite to add the B2900 and verify the connection. To add the instrument, you can request Connection Expert to dis cover the inst rument. If t he instrument cannot be fo und, add the instrument using the instrument’s hostname or IP address.
NOTE If this does not work, refer to the troubleshooting section in Connectivity Guide.
4. You can now use the Interactive IO within Connection Expert to communicate with your instrument, or program your instrument using various programming environments. You can also use the Web browser o n your compu ter to connect to the instrument as described under “Using the Graphical Web Interface” on page
3-31.
Viewing the Active LAN Status
T o view the currently active LAN settings, press the More function key, then the I/O > LAN > Status softkeys. The LAN status dialog box appears.
Note that the currently active LAN settings for the IP Address, Subnet Mask, and Default Gateway may be different from the settings specified in the LAN Configuration dialog box, depending on the configuration of the network. If the settings are different, it is because the network has automatically assigned its own settings.
Modifying the LAN Settings
As shipped from the factory, B2900’s pre-configured settings should work in most LAN environments. If you need to manually configure these settings, press the More function key, then the I/O > LAN > Config softkeys. The LAN Configuration dialog box appears.
NOTE B2900 must be rebooted if the host name is changed.
For the LAN setup parameters, see “LAN Configuration dialog box” on page 4-48.
3-30 Keysight B2900 User’s Guide, Edition 5
Installation

Communicating Over the LAN

Communicating Over the LAN

Using the Graphical Web Interface

Your Keysight B2900 Source/Measure Unit has a built-in graphical web interface that lets you control it directly from an Web browser on your computer. Multiple simultaneous connections are allowed, but only from one computer. With additional connections, performance will be reduced. Multiple connections from multiple computers are not allowed.
With the graphical web interface, you can access the front panel control functions including the LAN configuration parameters. This is a convenient way to communicate with the B2900 without using I/O libraries or drivers.
NOTE The built-in graphical web interface only operates over the LAN interface. It
requires Internet Explorer 6+ or Firefox 2+. You will als o need the Java Plug-in. This is included in the Java Runtime Environment. Visit the Java’s website.
The graphical web interface is enabled when shipped.
Keysight B2900 User’s Guide, Edition 5 3-31
Installation Communicating Over the LAN
To launch the graphical web interface:
1. Open a Web browser on your computer.
2. Enter the instrument’s hostname or IP address into the browser’s Address field to launch the graphical web interface. The B2900 home page will appear.
3. Click on the Browser Web Control button in the navigation bar on the left to begin controlling your instrument.
4. For additional help about any page, click “Help with this Page”.
If desired, you can control the access to the graphical web interface using password protection. As shipped fr om the factor y, the password is set to Keysight, keysight, or agilent. T o change the pas sword, click on the View & Modify Con figuration bu tton. Refer to the online help for additional information about setting the password.

Using Telnet

The Telnet utility (as well as sockets), is another way to communicate with B2900 without using I/O libraries or drivers. In all cases, you must first establish a LAN connection from your computer to the B2900 as previously described.
In an MS-DOS Command Prompt box, type “telnet hostname 5024” where hostname is the B2900’s hostname or IP address, and 50 24 is the ins trument’s telnet port. You should get a Telnet session box with a title indicating that you are connected to the B2900. Type the SCPI commands at the prompt.

Using Sockets

NOTE Keysight B2900 allows any combination of up to four simultaneous data socket,
control socket, or telnet connections to be made.
Keysight instruments are standardized on using port 5025 for SCPI socket services. A data socket on this port can be used to send and receive ASCII/SCPI commands, queries, and query responses. All commands must be terminated with a newline for the message to be parsed. All query responses will also be terminated with a newline.
The socket programming interface also allows control socket connections. The control socket can be used by a client to send device clear and to receive service requests. Unlike the data socket, which uses a fixed port number, the port number for a control socket varies and must be obtained by sending the following SCPI query to the data socket: SYSTem:COMMunicate:TCPip:CONTrol?
3-32 Keysight B2900 User’s Guide, Edition 5
Installation
Communicating Over the LAN
Once the port number is obtained, a control socket connection can be opened. As with the data socket, all commands to the control socket must be terminated with a newline, and all query responses returned on the control socket will be terminated with a newline.
T o send a device clear , send the strin g “DCL” to the control s ocket. When the B2900 has finished performing the device clear, it will echo the string “DCL” back to the control socket.
Service requests are enabled for control sockets using the Service Request Enable register. Once service requests have been enabled, the client program listens on the control connection. When SRQ becomes true, the instrument will send the string “SRQ +nn” to the client. The “nn” is the status byte value, which the client can use to determine the source of the service request.
Keysight B2900 User’s Guide, Edition 5 3-33
Installation Using Digital I/O

Using Digital I/O

B2900A has a Digital I/O connector, D-sub 25 pin female, for general purpose input/output (GPIO). It can be used for:
Trigger input
Trigger output
Dig ital signal input/output
Dig ital signal input
Start of Test (SOT) input (for component handlers)
Busy status output (for component handlers)
End of Test (EOT) output (for component handlers)
Interlock control
High voltage status output (shared with the digital signal input/output DIO 14)
The pin assignment of the Digital I/O connector is shown in Table 3-1. Pins DIO 1 to DIO 14 can be assigned to one of the above functions, except for interlock control. T o set the DIO functions, press the More function key, then the I/O > DIO > Config softkeys. For details, see “DIO Configuration dialog box” on page 4-48.
Figure 3-14 shows the input/output circuits internally connected to each pin of the
Digital I/O connector. For the available accessory, see “Accessory for Digital I/O Connector” on page
3-36.
3-34 Keysight B2900 User’s Guide, Edition 5
Figure 3-14 Digital I/O Internal Circuit
Vcc
to Digital I/O
connector
R1
R2
from Output control
Vcc: +5 V R1: 5 kohm R2: 100 ohm
to Input control
Table 3-1 Digital I/O Pin Assignment
Description Pin Number Description
Installation
Using Digital I/O
a
+5 V
Interloc k control
a
+5 V
a
+5 V
b
25 13 DIO 13 (bit 13)
24 12 DIO 12 (bit 12)
23 11 DIO 11 (bit 11)
22 10 DIO 10 (bit 10)
GND 21 9 DIO 9 (bit 9) GND 20 8 DIO 8 (bit 8) GND 19 7 DIO 7 (bit 7) GND 18 6 DIO 6 (bit 6) GND 17 5 DIO 5 (bit 5)
Interlock control
c
16 4 DIO 4 (bit 4)
GND 15 3 DIO 3 (bit 3)
DIO 14 (bit 14) or
High voltage status
Keysight B2900 User’s Guide, Edition 5 3-35
14 2 DIO 2 (bit 2)
1 DIO 1 (bit 1)
Installation
N1294A Opt 031
GPIO-BNC Trigger Adapter
DIO10 DIO 9
Using Digital I/O
a. Current limit: 600 mA (total current to the pins 22, 23, and 25) b. Used for the positive logic. Connected to the pin 25 for the negative
logic.
c. Used for the negative logic. Connected to the pin 17 for the positive
logic.

Accessory for Digital I/O Connector

The following accessory is available for the Digital I/O connector.
Keysight N1294A-031 GPIO-BNC Trigger Adapter This adapter is used to change the Digital I/O connector to the trigger input/
output BNC connectors. The BNC connectors are internally connected to the DIO pins 9 and 10.
This adapter can be connected directly to the Digital I/O connector.
Figure 3-15 N1294A-031 GPIO-BNC Trigger Adapter
NOTE Interlock terminals
If the N1294A-031 adapter is connected to the Digital I/O connector, the Interlock terminals are opened. Then the B2961A/B2962A output voltage is limited to 42 V.
T o use the interlock function, see “Installing the Interlock Circuit” on page 3-24 and extend the Interlock terminals. They can be extended from the adapter internal board through the opening of the adapter top cover.
3-36 Keysight B2900 User’s Guide, Edition 5
Loading...