This document contains proprietary information that is protected by copyright. All rights
are reserved.
No part of this document may
reproduced in (including electronic storage and retrieval or
translation into a foreign language) without prior agreement
and written consent from Keysight Technologies Deutschland
GmbH as governed by United
States and international copywright laws.
The material in this document is
subject to change without notice.
Keysight Technologies makes no
warranty of any kind with regard
to this printed material, including, but not limited to, the
implied warranties of merchantability and fitness for a particular purpose.
Keysight Technologies shall not
be liable for errors contained
herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this material.
Warranty
This Keysight Technologies
instrument product is warranted
against defects in material and
workmanship for a period of one
year from date of shipment.
During the warranty period,
Keysight will, at its option,
either repair or replace products
that prove to be defective.
For warranty service or repair,
this product must be returned to
a service facility designated by
Keysight. Buyer shall prepay
shipping charges to Keysight
and Keysight shall pay shipping
charges to return the product to
Buyer. However, Buyer shall
pay all shipping charges, duties,
and taxes for products returned
to Keysight from another country. Keysight warrants that its
software and firmware designated by Keysight for use with
an instrument will execute its
programming instructions when
properly installed on that instrument. Keysight does not warrant that the operation of the
instrument, software, or firmware will be uninterrupted or
error free.
Limitation of Warranty
The foregoing warranty shall not
apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse,
operation outside of the environmental specifications for the
product, or improper site preparation or maintenance.
No other warranty is expressed
or implied. Keysight Technologies specifically disclaims the
implied warranties of Merchantability and Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein
are Buyer’s sole and exclusive
remedies. Keysight Technologies shall not be liable for any
direct, indirect, special, incidental, or consequential damages
whether based on contract, tort,
or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Keysight Technologies products. For any assistance contact
your nearest Keysight Technologies Sales and Service Office.
Certification
Keysight Technologies certifies
that this product met its published specifications at the time
of shipment from the factory.
Keysight Technologies further
certifies that its calibration measurements are traceable to the
United States National Institute
of Standards and Technology,
NIST to the extent allowed by
the Institutes’s calibration facility, and to the calibration facilities of other International
Standards Organization members.
ISO 9001 Certification
Produced to ISO 9001 international quality system standard as
part of our objective of continually increasing customer satisfaction through improved
process control.
Safety Notices
CAUTION
Caution denotes a hazard. It
calls attention to a procedure
which, if not correctly performed or adhered to, could
result in damage to or destruction of the product. Do not proceed beyond a caution sign until
the indicated conditions are fully
understood and met.
WARNING
Warning denotes a hazard. It
calls attention to a procedure
which, if not correctly performed or adhered to, could
result in injury or loss of life. Do
not proceed beyond a warning
sign until the indicated conditions are fully understood and
met.
The instruction manual
symbol. The product is
marked with this warning symbol when it is
necessary for the user to
refer to the instructions
in the manual.
The laser radiation symbol. This warning symbol is marked on
products which have a
laser output.
The AC symbol is used
to indicate the required
nature of the line module input power.
| The ON symbols are
used to mark the positions of the instrument
power line switch.
m The OFF symbols
are used to mark the
positions of the instrument power line switch.
The CE mark is a registered trademark of the
European Community.
The CSA mark is a registered trademark of the
Canadian Standards
Association.
The C-Tick mark is a
registered trademark of
the Australian Spectrum Management
Agency.
This text denotes the
instrument is an Industrial Scientific and Medical Group 1 Class A
product.
Typographical Conventions.
The following conventions are
used in this book:
Key type for keys or text located
on the keyboard or instrument.
Softkeytype for key names that
are displayed on the instrument’s
screen.
Display type for words or characters displayed on the computer’s screen or instrument’s
display.
Usertype for words or characters
that you type or enter.
Emphasis type for words or
characters that emphasize some
point or that are used as place
holders for text that you type.
Edition 3:
86120-90B03: November 2014
Edition 2:
86120-90B03: July 2004
Edition 1:
2
The Keysight 86120B—At a Glance
The Keysight 86120B—At a Glance
The Keysight 86120B Multi-Wavelength Meter measures the wavelength and optical power of laser light in the 700-1650 nm wavelength range. Because the
Keysight 86120B simultaneously measures multiple laser lines, you can characterize wavelength-division-multiplexed (WDM) systems and the multiple lines of
Fabry-Perot lasers.
NOTE
The front-panel OPTICAL INPUT connector uses a single-mode input fiber.
What’s new with the Keysight 86120B
This book directly applies to Keysight 86120B instruments with firmware version
number 2.0. When first turned on, the instrument briefly displays the firmware version. These instruments have the added capability of measuring broadband devices
and chirped lasers. Refer to “Connect the fiber-optic cable to the front-panel OPTICAL INPUT connector.” on page 2-40.
Characterize laser lines easily
With the Keysight 86120B you can quickly and easily measure any of the following
parameters:
• Wavelengths and powers
• Average wavelength
• Total optical power
• Laser line separation
• Laser drift (wavelength and power)
• Signal-to-noise ratios
• Coherence length
3
The Keysight 86120B—At a Glance
In addition to these measurements, a “power bar” is displayed that shows power
changes like a traditional analog meter. You can see the power bar shown in the following figure of the Keysight 86120B’s display.
CAUTIONThe input circuitry of the Keysight 86120B can be damaged when total input
power levels exceed +18 dBm. To prevent input damage, this specified level
must not be exceeded.
Print measurement results
You can get hardcopy results of your measurements by connecting a printer to the
rear-panel PARALLEL PRINTER PORT connector.
Program the instrument for automatic measurements
The Keysight 86120B offers an extensive set of GPIB programming commands.
These commands allow you to perform automated measurements on manufacturing
production lines and remote sites. Chapter 4, “Programming” and Chapter 5, “Com-
mon Commands” provide all the information you’ll need to know in order to pro-
gram the Keysight 86120B.
Display wavelengths as if measured in vacuum or standard air
Although all measurements are made in air, displayed results are corrected for air
dispersion to accurately show wavelength values in vacuum or in “standard air.” To
ensure accurate wavelength measurements, make sure that you enter the elevation
from which you will be making measurements as described in Chapter 1, “Getting
Started”.
4
The Keysight 86120B—At a Glance
Measurement accuracy—it’s up to you!
Fiber-optic connectors are easily damaged when connected to dirty or damaged
cables and accessories. The Keysight
86120B’s front-panel INPUT connector is
no exception. When you use improper cleaning and handling techniques, you
risk expensive instrument repairs, damaged cables, and compromised measure
ments.
Before you connect any fiber-optic cable to the Keysight 86120B, refer to
“Cleaning Connections for Accurate Measurements” on page 23.
-
5
General Safety Considerations
General Safety Considerations
This product has been designed and tested in accordance with IEC 61010-1 and has
been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and
to maintain the product in a safe condition.
There is no output laser aperture
The Keysight 86120B does not have an output laser aperture. However, light
less than
Operator maintenance or precautions are not necessary to maintain safety. No
controls, adjustments, or performance of procedures result in hazardous radiation exposure.
1 nW escapes out of the front-panel OPTIC
AL INPUT connector.
6
General Safety Considerations
Laser Safety Information
The laser sources contained in the product specified by this user guide are classified
according to IEC 60825-1 (2007).
The laser sources comply with 21 CFR 1040.10 except for deviations pursuant to
Laser Notice No. 50 dated 2007-June-24.
Table 1-1. Safety Information for Wavelength Meter
86120B
Laser typeHeNe Laser
Wavelength (±0.1 nm)632.8 nm
Max. CW output power
Beam waist diameter9 µm
Numerical aperture0.1
1
< 1 nW
Laser Class according to IEC 608251 (2007)
Max. permissible CW output power0.39 mW
1
Max. CW output power is defined as the highest possible optical power that
the laser source can produce at its output connector
Class 1
7
General Safety Considerations
WARNINGIf this instrument is not used as specified, the protection provided
by the equipment could be impaired. This instrument must be used
in a normal condition (in which all means for protection are intact)
only.
WARNINGNo operator serviceable parts inside. Refer servicing to qualified
personnel. To prevent electrical shock, do not remove covers.
WARNINGTo prevent electrical shock, disconnect the Keysight 86120B from
mains before cleaning. Use a dry cloth or one slightly dampened
with water to clean the external case parts. Do not attempt to clean
internally.
WARNINGThis is a Safety Class 1 product (provided with protective earth).
The mains plug shall only be inserted in a socket outlet provided
with a protective earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
WARNINGFor continued protection against fire hazard, replace line fuse only
with same type and ratings, (Fuse type F 6.3 A/250V, IEC 60127
type 5x20mm). The use of other fuses or materials is prohibited.
CAUTIONThis product complies with Overvoltage Category II and Pollution Degree 2.
8
General Safety Considerations
CAUTIONVENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be restricted. The
ambient temperature (outside the cabinet) must be less than the maximum
operating temperature of the product by 4
⋅C for every 100 watts dissipated in
the cabinet. If the total power dissipated in the cabinet is greater than
800 watts, then forced convection must be used.
CAUTIONAlways use the three-prong ac power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
CAUTIONDo not connect ac power until you have verified the line voltage is correct as
described in
“Line Power Requirements” on page 1-16. Damage to the equipment
could result.
CAUTIONThis instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
9
General Safety Considerations
10
Contents
The Keysight 86120B—At a Glance 3
General Safety Considerations 6
1 Getting Started
Step 1. Inspect the Shipment 15
Step 2. Connect the Line-Power Cable 16
Step 3. Connect a Printer 17
Step 4. Turn on the Keysight 86120B 18
Step 5. Enter Your Elevation 20
Step 6. Select Medium for Wavelength Values 21
Step 7. Turn Off Wavelength Limiting 22
Cleaning Connections for Accurate Measurements 23
Returning the Instrument for Service 34
2 Using the Multi-Wavelength Meter
Displaying Wavelength and Power 39
Changing the Units and Measurement Rate 49
Defining Laser-Line Peaks 52
Measuring Laser Separation 56
Measuring Modulated Lasers 60
Measuring Total Power Greater than 10 dBm 62
Calibrating Measurements 63
Printing Measurement Results 65
Addressing and Initializing the Instrument 83
Making Measurements 85
Monitoring the Instrument 96
Reviewing SCPI Syntax Rules 103
Example Programs 108
Lists of Commands 123
Test 1. Absolute Wavelength Accuracy 221
Test 2. Sensitivity 222
Test 3. Polarization Dependence 223
Test 4. Optical Input Return Loss 224
Test 5. Amplitude Accuracy and Linearity 226
7 Specifications and Regulatory Information
Definition of Terms 233
Specifications 235
General Safety Information 239
Compliance with Canadian EMC Requirements 239
Notice for Germany: 239
Declaration of Conformity 240
Product Overview 241
8 Reference
Instrument Preset Conditions 244
Menu Maps 246
Error Messages 253
Connector interfaces (order separately) 259
Power Cords 260
For Assistance and Support 262
12
1
Step 1. Inspect the Shipment 15
Step 2. Connect the Line-Power Cable 16
Step 3. Connect a Printer 17
Step 4. Turn on the Keysight 86120B 18
Step 5. Enter Your Elevation 20
Step 6. Select Medium for Wavelength Values 21
Step 7. Turn Off Wavelength Limiting 22
Cleaning Connections for Accurate Measurements 23
Returning the Instrument for Service 34
Getting Started
Getting Started
Getting Started
Getting Started
The instructions in this chapter show you how to install your Keysight 86120B. You
should be able to finish these procedures in about ten to twenty minutes. After
you’ve completed this chapter, continue with Chapter 2, “Using the Multi-Wave-
length Meter”. Refer to Chapter 7, “Specifications and Regulatory Information” for
information on operating conditions such as temperature.
CAUTIONInstall the instrument so that the ON/OFF switch is readily identifiable and is
easily reached by the operator. The ON/OFF switch or the detachable power
cord is the instrument disconnecting device. It disconnects the mains circuits
from the mains supply before other parts of the instrument. Alternately, an
externally installed switch or circuit breaker (which is really identifiable and
is easily reached by the operator) may be used as a disconnecting device.
CAUTIONInstall the instrument according to the enclosure protection provided. This
instrument does not protect against the ingress of water. This instrument
protects against finger access to hazardous parts within the enclosure.
14
Getting Started
Step 1. Inspect the Shipment
Step 1. Inspect the Shipment
1 Verify that all system components ordered have arrived by comparing the shipping
forms to the original purchase order. Inspect all shipping containers.
If your shipment is damaged or incomplete, save the packing materials and notify
both the shipping carrier and the nearest Keysight Technologies sales and service
office. Keysight Technologies will arrange for repair or replacement of damaged
or incomplete shipments without waiting for a settlement from the transportation
company. Notify the Keysight Technologies customer engineer of any problems.
2 Make sure that the serial number and options listed on the instrument’s rear-panel
label match the serial number and options listed on the shipping document.
15
Getting Started
Step 2. Connect the Line-Power Cable
Step 2. Connect the Line-Power Cable
WARNINGThis is a Safety Class 1 Product (provided with protective earth).
The mains plug shall only be inserted in a socket outlet provided
with a protective earth contact. Any interruption of the protective
conductor inside or outside of the instrument is likely to make the
instrument dangerous. Intentional interruption is prohibited.
CAUTIONAlways use the three-prong AC power cord supplied with this instrument.
Failure to ensure adequate earth grounding by not using this cord may cause
instrument damage.
CAUTIONDo not connect ac power until you have verified the line voltage is correct as
described in the following paragraphs. Damage to the equipment could result.
CAUTIONThis instrument has autoranging line voltage input. Be sure the supply
voltage is within the specified range.
1 Verify that the line power meets the requirements shown in the following table.
Line Power Requirements
Voltagemax. 100 / 115 / 230 / 240 V~
Frequency50 / 60 Hz
2 Connect the line-power cord to the instrument’s rear-panel connector.
3 Connect the other end of the line-power cord to the power receptacle.
16
Getting Started
Step 3. Connect a Printer
Various power cables are available to connect the Keysight 86120B to ac power outlets unique to specific geographic areas. The cable appropriate for the area to which
the Keysight 86120B is originally shipped is included with the unit. The cable
shipped with the instrument also has a right-angle connector so that the
Keysight 86120B can be used while sitting on its rear feet. You can order additional
ac power cables for use in different geographic areas. Refer to “Connector interfaces
(order separately)” on page 259.
Step 3. Connect a Printer
The Keysight 86120B can print hardcopies of measurement results on a printer. The
output is ASCII text. If you don’t have a printer, continue with .
• Using a standard parallel printer cable, connect the printer to the Keysight 86120B’s
rear-panel PARALLEL PRINTER PORT connector.
17
Getting Started
Step 4. Turn on the Keysight 86120B
Step 4. Turn on the Keysight 86120B
CAUTIONThe front panel LINE switch disconnects the mains circuits from the mains
supply after the EMC filters and before other parts of the instrument.
1 Press the front-panel LINE key. After approximately 20 seconds, the display should
look similar to the following figure:
The front-panel LINE switch disconnects the mains circuits from the mains supply
after the EMC filters and before other parts of the instrument.
2 If the Keysight 86120B fails to turn on properly, consider the following possibilities:
• Is the line fuse good?
• Does the line socket have power?
• Is it plugged into the proper ac power source?
If the instrument still fails, return it to Keysight Technologies for repair. Refer to
“Returning the Instrument for Service” on page 34.
18
Getting Started
Step 4. Turn on the Keysight 86120B
Instrument firmware version
When the instrument is first turned on, the display briefly shows the instrument’s firmware version number. In the unlikely event that you have a problem
with the Keysight
86120B, you may need to indicate this number when commu-
nicating with Keysight Technologies.
There is no output laser aperture
The Keysight 86120B does not have an output laser aperture. However, light
less than 1
nw escapes out of the front-panel OPTICAL INPUT connector.
Operator maintenance or precautions are not necessary to maintain safety. No
controls, adjustments, or performance of procedures result in hazardous radia
-
tion exposure.
Measurement accuracy—it’s up to you!
Fiber-optic connectors are easily damaged when connected to dirty or damaged
cables and accessories. The Keysight
86120B’s front-panel INPUT connector is
no exception. When you use improper cleaning and handling techniques, you
risk expensive instrument repairs, damaged cables, and compromised measure
ments.
Before you connect any fiber-optic cable to the Keysight 86120B, refer to
“Cleaning Connections for Accurate Measurements” on page 23.
-
19
Getting Started
m
ft
3.281
-------------
=
Step 5. Enter Your Elevation
Step 5. Enter Your Elevation
In order for your Keysight 86120B to accurately measure wavelengths and meet its
published specifications, you must enter the elevation where you will be performing
your measurements.
1 Press the Setup key.
2 Press the MORE softkey.
3 Press the CAL softkey.
4 Press ELEV.
5 Use the and softkeys to enter the elevation in meters. Entries jump in 500
meter steps from 0 m to 5000 m.
The elevation value selected with the softkeys must be within 250 meters of the actual elevation.
6 Press RETURN to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by
using the following equation:
20
Getting Started
Step 6. Select Medium for Wavelength Values
Step 6. Select Medium for Wavelength Values
Because wavelength varies with the material that the light passes through, the
Keysight 86120B offers wavelength measurements in two mediums: vacuum and
standard air.
1 Press the Setup key.
2 Press the MORE softkey.
3 Press the CAL softkey.
4 Make the following selection:
• Press VACUUM for wavelength readings in a vacuum.
• Press STD AIR for wavelength readings in standard air.
5 Press RETURN to complete the entry.
Definition of standard air
Standard air is defined to have the following characteristics:
After the Preset key is pressed, the input wavelength range is limited to measuring
lasers between 1200 nm and 1650 nm. You can easily expand the input range to the
full 700 nm to 1650 nm range with the following steps:
1 Press the Preset key.
2 Press the Setup key.
3 Press the WL LIM softkey.
4 Press LIM OFF to remove the limits on wavelength range.
All responses in the full 700 nm to 1650 nm range are now displayed.
22
Getting Started
Cleaning Connections for Accurate Measurements
Cleaning Connections for Accurate Measurements
Today, advances in measurement capabilities make connectors and connection techniques more important than ever. Damage to the connectors on calibration and verification devices, test ports, cables, and other devices can degrade measurement
accuracy and damage instruments. Replacing a damaged connector can cost thousands of dollars, not to mention lost time! This expense can be avoided by observing
the simple precautions presented in this book. This book also contains a brief list of
tips for caring for electrical connectors.
Choosing the Right Connector
A critical but often overlooked factor in making a good lightwave measurement is
the selection of the fiber-optic connector. The differences in connector types are
mainly in the mechanical assembly that holds the ferrule in position against another
identical ferrule. Connectors also vary in the polish, curve, and concentricity of the
core within the cladding. Mating one style of cable to another requires an adapter.
Keysight Technologies offers adapters for most instruments to allow testing with
many different cables. Figure 1-1 on page 24 shows the basic components of a typical connectors.
The system tolerance for reflection and insertion loss must be known when selecting
a connector from the wide variety of currently available connectors. Some items to
consider when selecting a connector are:
• How much insertion loss can be allowed?
• Will the connector need to make multiple connections? Some connectors are better
than others, and some are very poor for making repeated connections.
• What is the reflection tolerance? Can the system take reflection degradation?
• Is an instrument-grade connector with a precision core alignment required?
• Is repeatability tolerance for reflection and loss important? Do your specifications
23
Getting Started
Cleaning Connections for Accurate Measurements
take repeatability uncertainty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be required?
For example, many DFB lasers cannot operate with reflections from connectors. Of
ten as much as 90 dB isolation is needed.
-
Figure 1-1. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most popular
connector for fiber-optic applications. While not the highest performing connector,
it represents a good compromise between performance, reliability, and cost. If properly maintained and cleaned, this connector can withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most connectors, including the FC/PC style, can deliver. These instruments cannot tolerate
connectors with the large non-concentricities of the fiber common with ceramic
style ferrules. When tighter alignment is required, Keysight
Technologies instruments typically use a connector such as the Diamond HMS-10,
which has concentric tolerances within a few tenths of a micron. Keysight Technologies then uses a special universal adapter, which allows other cable types to mate
with this precision connector. See Figure 1-2.
24
Getting Started
Cleaning Connections for Accurate Measurements
Figure 1-2. Universal adapters to Diamond HMS-10.
The HMS-10 encases the fiber within a soft nickel silver (Cu/Ni/Zn) center which is
surrounded by a tough tungsten carbide casing, as shown in Figure 1-3.
Figure 1-3. Cross-section of the Diamond HMS-10 connector.
The nickel silver allows an active centering process that permits the glass fiber to be
moved to the desired position. This process first stakes the soft nickel silver to fix
the fiber in a near-center location, then uses a post-active staking to shift the fiber
into the desired position within 0.2 μm. This process, plus the keyed axis, allows
very precise core-to-core alignments. This connector is found on most Keysight
Technologies lightwave instruments. The soft core, while allowing precise centering, is also the chief liability of the connector. The soft material is easily damaged.
Care must be taken to minimize excessive scratching and wear. While minor wear is
not a problem if the glass face is not affected, scratches or grit can cause the glass
fiber to move out of alignment. Also, if unkeyed connectors are used, the nickel sil-
25
Getting Started
Cleaning Connections for Accurate Measurements
ver can be pushed onto the glass surface. Scratches, fiber movement, or glass contamination will cause loss of signal and increased reflections, resulting in poor
return loss.
Inspecting Connectors
Because fiber-optic connectors are susceptible to damage that is not immediately
obvious to the naked eye, poor measurements result without the user being aware.
Microscopic examination and return loss measurements are the best way to ensure
good measurements. Good cleaning practices can help ensure that optimum connector performance is maintained. With glass-to-glass interfaces, any degradation of a
ferrule or the end of the fiber, any stray particles, or finger oil can have a significant
effect on connector performance. Where many repeat connections are required, use
of a connector saver or patch cable is recommended.
Figure 1-4 shows the end of a clean fiber-optic cable. The dark circle in the center of
the micrograph is the fiber’s 125 μm core and cladding which carries the light. The
surrounding area is the soft nickel-silver ferrule. Figure 1-5 shows a dirty fiber end
from neglect or perhaps improper cleaning. Material is smeared and ground into the
end of the fiber causing light scattering and poor reflection. Not only is the precision
polish lost, but this action can grind off the glass face and destroy the connector.
Figure 1-6 shows physical damage to the glass fiber end caused by either repeated
connections made without removing loose particles or using improper cleaning
tools. When severe, the damage of one connector end can be transferred to another
good connector endface that comes in contact with the damaged one. Periodic
checks of fiber ends, and replacing connecting cables after many connections is a
wise practice. The cure for these problems is disciplined connector care as described
in the following list and in “Cleaning Connectors” on page 30.
26
Getting Started
Cleaning Connections for Accurate Measurements
Use the following guidelines to achieve the best possible performance when making
measurements on a fiber-optic system:
• Never use metal or sharp objects to clean a connector and never scrape the connector.
• Avoid matching gel and oils.
Figure 1-4. Clean, problem-free fiber end and ferrule.
Figure 1-5. Dirty fiber end and ferrule from poor cleaning.
27
Getting Started
Cleaning Connections for Accurate Measurements
Figure 1-6. Damage from improper cleaning.
While these often work well on first insertion, they are great dirt magnets. The oil or
gel grabs and holds grit that is then ground into the end of the fiber. Also, some early
gels were designed for use with the FC, non-contacting connectors, using small
glass spheres. When used with contacting connectors, these glass balls can scratch
and pit the fiber. If an index matching gel or oil must be used, apply it to a freshly
cleaned connector, make the measurement, and then immediately clean it off. Never
use a gel for longer-term connections and never use it to improve a damaged connector. The gel can mask the extent of damage and continued use of a damaged fiber
can transfer damage to the instrument.
• When inserting a fiber-optic cable into a connector, gently insert it in as straight a line
as possible. Tipping and inserting at an angle can scrape material off the inside of the
connector or even break the inside sleeve of connectors made with ceramic material.
• When inserting a fiber-optic connector into a connector, make sure that the fiber end
does not touch the outside of the mating connector or adapter.
• Avoid over tightening connections.
Unlike common electrical connections, tighter is not better. The purpose of the connector is to bring two fiber ends together. Once they touch, tightening only causes a
greater force to be applied to the delicate fibers. With connectors that have a convex
fiber end, the end can be pushed off-axis resulting in misalignment and excessive
return loss. Many measurements are actually improved by backing off the connector
pressure. Also, if a piece of grit does happen to get by the cleaning procedure, the
tighter connection is more likely to damage the glass. Tighten the connectors just
until the two fibers touch.
• Keep connectors covered when not in use.
• Use fusion splices on the more permanent critical nodes. Choose the best connector
possible. Replace connecting cables regularly. Frequently measure the return loss of
28
Getting Started
Cleaning Connections for Accurate Measurements
the connector to check for degradation, and clean every connector, every time.
All connectors should be treated like the high-quality lens of a good camera. The
weak link in instrument and system reliability is often the inappropriate use and care
of the connector. Because current connectors are so easy to use, there tends to be
reduced vigilance in connector care and cleaning. It takes only one missed cleaning
for a piece of grit to permanently damage the glass and ruin the connector.
Measuring insertion loss and return loss
Consistent measurements with your lightwave equipment are a good indication that
you have good connections. Since return loss and insertion loss are key factors in
determining optical connector performance they can be used to determine connector
degradation. A smooth, polished fiber end should produce a good return-loss measurement. The quality of the polish establishes the difference between the “PC”
(physical contact) and the “Super PC” connectors. Most connectors today are physical contact which make glass-to-glass connections, therefore it is critical that the
area around the glass core be clean and free of scratches. Although the major area of
a connector, excluding the glass, may show scratches and wear, if the glass has
maintained its polished smoothness, the connector can still provide a good low level
return loss connection.
If you test your cables and accessories for insertion loss and return loss upon receipt,
and retain the measured data for comparison, you will be able to tell in the future if
any degradation has occurred. Typical values are less than 0.5 dB of loss, and sometimes as little as 0.1 dB of loss with high performance connectors. Return loss is a
measure of reflection: the less reflection the better (the larger the return loss, the
smaller the reflection). The best physically contacting connectors have return losses
better than 50 dB, although 30 to 40 dB is more common.
29
Getting Started
Cleaning Connections for Accurate Measurements
Visual inspection of fiber ends
Visual inspection of fiber ends can be helpful. Contamination or imperfections on
the cable end face can be detected as well as cracks or chips in the fiber itself. Use a
microscope (100X to 200X magnification) to inspect the entire end face for contamination, raised metal, or dents in the metal as well as any other imperfections.
Inspect the fiber for cracks and chips. Visible imperfections not touching the fiber
core may not affect performance (unless the imperfections keep the fibers from contacting).
WARNINGAlways remove both ends of fiber-optic cables from any
instrument, system, or device before visually inspecting the fiber
ends. Disable all optical sources before disconnecting fiber-optic
cables. Failure to do so may result in permanent injury to your
eyes.
Cleaning Connectors
The procedures in this section provide the proper steps for cleaning fiber-optic
cables and Keysight Technologies universal adapters. The initial cleaning, using the
alcohol as a solvent, gently removes any grit and oil. If a caked-on layer of material
is still present, (this can happen if the beryllium-copper sides of the ferrule retainer
get scraped and deposited on the end of the fiber during insertion of the cable), a
second cleaning should be performed. It is not uncommon for a cable or connector
to require more than one cleaning.
CAUTIONKeysight Technologies strongly recommends that index matching
compounds not be applied to their instruments and accessories. Some
compounds, such as gels, may be difficult to remove and can contain
30
Getting Started
Cleaning Connections for Accurate Measurements
damaging particulates. If you think the use of such compounds is necessary,
refer to the compound manufacturer for information on application and
cleaning procedures.
Table 1-2. Cleaning Accessories
Item Keysight Technologies Part Number
Pure isopropyl alcohol—
Cotton swabs8520-0023
Small foam swabs9300-1223
Compressed dust remover (non-residue)8500-5262
Table 1-3. Dust Caps Provided with Lightwave Instruments
Item Keysight Technologies Part Number
Laser shutter cap08145-64521
FC/PC dust cap08154-44102
Biconic dust cap08154-44105
DIN dust cap5040-9364
HMS10/dust cap5040-9361
ST dust cap5040-9366
To clean a non-lensed connector
CAUTIONDo not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1 Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
31
Getting Started
Cleaning Connections for Accurate Measurements
Cotton swabs can be used as long as no cotton fibers remain on the fiber end after
cleaning.
2 Clean the ferrules and other parts of the connector while avoiding the end of the fiber.
3 Apply isopropyl alcohol to a new clean lint-free cotton swab or lens paper.
4 Clean the fiber end with the swab or lens paper.
Do not scrub during this initial cleaning because grit can be caught in the swab and
become a gouging element.
5 Immediately dry the fiber end with a clean, dry, lint-free cotton swab or lens paper.
6 Blow across the connector end face from a distance of 6 to 8 inches using filtered,
dry, compressed air. Aim the compressed air at a shallow angle to the fiber end face.
Nitrogen gas or compressed dust remover can also be used.
CAUTIONDo not shake, tip, or invert compressed air canisters, because this releases
particles in the can into the air. Refer to instructions provided on the
compressed air canister.
7 As soon as the connector is dry, connect or cover it for later use.
If the performance, after the initial cleaning, seems poor try cleaning the connector
again. Often a second cleaning will restore proper performance. The second cleaning should be more arduous with a scrubbing action.
To clean an adapter
The fiber-optic input and output connectors on many Keysight Technologies instruments employ a universal adapter such as those shown in the following picture.
These adapters allow you to connect the instrument to different types of fiber-optic
cables.
Figure 1-7. Universal adapters.
1 Apply isopropyl alcohol to a clean foam swab.
Cotton swabs can be used as long as no cotton fibers remain after cleaning. The foam
32
Getting Started
Cleaning Connections for Accurate Measurements
swabs listed in this section’s introduction are small enough to fit into adapters.
Although foam swabs can leave filmy deposits, these deposits are very thin, and the
risk of other contamination buildup on the inside of adapters greatly outweighs the
risk of contamination by foam swabs.
2 Clean the adapter with the foam swab.
3 Dry the inside of the adapter with a clean, dry, foam swab.
4 Blow through the adapter using filtered, dry, compressed air.
Nitrogen gas or compressed dust remover can also be used. Do not shake, tip, or invert compressed air canisters, because this releases particles in the can into the air.
Refer to instructions provided on the compressed air canister.
33
Getting Started
Returning the Instrument for Service
Returning the Instrument for Service
The instructions in this section show you how to properly return the instrument for
repair or calibration. Always call the Keysight Technologies Instrument Support
Center first to initiate service before returning your instrument to a service office.
This ensures that the repair (or calibration) can be properly tracked and that your
instrument will be returned to you as quickly as possible. Call this number regardless of where you are located. Refer to “For Assistance and Support” on page 262
for a list of service offices.
If the instrument is still under warranty or is covered by a Keysight Technologies
maintenance contract, it will be repaired under the terms of the warranty or contract
(the warranty is at the front of this manual). If the instrument is no longer under warranty or is not covered by a Keysight Technologies maintenance plan, Keysight
Technologies will notify you of the cost of the repair after examining the unit.
When an instrument is returned to a Keysight Technologies service office for servicing, it must be adequately packaged and have a complete description of the failure
symptoms attached. When describing the failure, please be as specific as possible
about the nature of the problem. Include copies of additional failure information
(such as the instrument failure settings, data related to instrument failure, and error
messages) along with the instrument being returned.
Preparing the instrument for shipping
1 Write a complete description of the failure and attach it to the instrument. Include any
specific performance details related to the problem. The following information
34
Returning the Instrument for Service
should be returned with the instrument.
• Type of service required.
• Date instrument was returned for repair.
• Description of the problem:
• Whether problem is constant or intermittent.
• Whether instrument is temperature-sensitive.
• Whether instrument is vibration-sensitive.
• Instrument settings required to reproduce the problem.
• Performance data.
• Company name and return address.
• Name and phone number of technical contact person.
• Model number of returned instrument.
• Full serial number of returned instrument.
• List of any accessories returned with instrument.
2 Cover all front or rear-panel connectors that were originally covered when you first
received the instrument.
CAUTIONCover electrical connectors to protect sensitive components from
electrostatic damage. Cover optical connectors to protect them from damage
due to physical contact or dust.
Getting Started
CAUTIONInstrument damage can result from using packaging materials other than the
original materials. Never use styrene pellets as packaging material. They do
not adequately cushion the instrument or prevent it from shifting in the
carton. They may also cause instrument damage by generating static
electricity.
3 Pack the instrument in the original shipping containers. Original materials are
available through any Keysight Technologies office. Or, use the following
guidelines:
• Wrap the instrument in antistatic plastic to reduce the possibility of damage
caused by electrostatic discharge.
• For instruments weighing less than 54 kg (120 lb), use a double-walled, corrugat-
35
Getting Started
Returning the Instrument for Service
ed cardboard carton of 159 kg (350 lb) test strength.
• The carton must be large enough to allow approximately 7 cm (3 inches) on all
sides of the instrument for packing material, and strong enough to accommodate
the weight of the instrument.
• Surround the equipment with approximately 7 cm (3 inches) of packing material,
to protect the instrument and prevent it from moving in the carton. If packing foam
is not available, the best alternative is S.D-240 Air Cap™ from Sealed Air Corpo
ration (Commerce, California 90001). Air Cap looks like a plastic sheet filled with
air bubbles. Use the pink (antistatic) Air Cap™ to reduce static electricity. Wrap
ping the instrument several times in this material will protect the instrument and
prevent it from moving in the carton.
4 Seal the carton with strong nylon adhesive tape.
5 Mark the carton “FRAGILE, HANDLE WITH CARE”.
6 Retain copies of all shipping papers.
-
-
36
2
Displaying Wavelength and Power 39
Changing the Units and Measurement Rate 49
Defining Laser-Line Peaks 52
Measuring Laser Separation 56
Measuring Modulated Lasers 60
Measuring Total Power Greater than 10 dBm 62
Calibrating Measurements 63
Printing Measurement Results 65
Using the Multi-Wavelength Meter
Using the Multi-Wavelength Meter
Using the Multi-Wavelength Meter
Using the Multi-Wavelength Meter
In this chapter, you’ll learn how to make a variety of fast, accurate measurements.
As you perform these measurements, keep in mind the following points:
• 700 nm to 1650 nm maximum input wavelength range
The range is normally limited from 1200 nm to 1650 nm. To use the full range, refer
to “Measuring lasers between 700 nm and 1200 nm” on page 44.
• +10 dBm maximum total displayed input power
• Laser linewidths assumed to be less than 10 GHz
• If you change the elevation where you will be using your Agilent 86120B, refer to
“Calibrating Measurements” on page 63.
• Press the green Preset key to return the Agilent 86120B to its default state.
CAUTIONDo not exceed +18 dBm source power. The Agilent 86120B’s input circuitry
can be damaged when total input power exceeds 18 dBm. You can measure
power levels that are greater by adding attenuation and entering a power
offset as described in “To measure total power exceeding 10 dBm” on page 62.
38
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Displaying Wavelength and Power
This section gives you step-by-step instructions for measuring peak wavelength,
average wavelength, peak power, and total input power. There are three display
modes:
• Peak wavelength
• List-by-wavelength or power
• Average wavelength and total power
If the measured amplitudes are low, clean the front-panel OPTICAL INPUT connector.
This section includes:
Peak WL mode 40
List by WL or power modes 42
Total power and average wavelength 43
Measuring lasers between 700 nm and 1200 nm 44
Limiting the wavelength range 44
Measuring broadband devices and chirped lasers 46
Graphical display of optical power spectrum 47
Instrument states 48
Power bar 48
39
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Peak WL mode
When Peak WL is pressed, the display shows the largest amplitude line in the spectrum. This is the peak wavelength mode. The word PEAK is shown on the screen. If
multiple laser lines are present at the input, the number of lines located will be
shown along the right side of the screen.
Display after “Peak WL” key pressed
In addition to the digital readouts, there is a power bar. It provides a convenient analog “meter movement” for tuning laser power.
Although the Peak WL mode shows one signal at a time, softkeys are provided that
allow you to scroll through and display all the measured laser lines. You can scroll
through the list according to the wavelengths or powers measured. The signals are
displayed in order from shortest to longest wavelengths. The Agilent 86120B can
measure up to 100 laser lines simultaneously.
To display peak wavelength and power
1 Connect the fiber-optic cable to the front-panel OPTICAL INPUT connector.
40
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
2 To display the peak wavelength and power, do one of the following:
• Press the green Preset key.
• Press Peak WL.
3 To move the cursor to view other signals, press:
• PREV WL to select next (previous) shorter wavelength.
• NEXT WL to select next longer wavelength.
• PEAK to signal with greatest power.
• PREV PK to select next lower power signal.
• NEXT PK to select next higher power signal.
41
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
List by WL or power modes
In the list-by-wavelength or list-by-power modes, the measurements of five laser
lines can be displayed at any one time. In list by wavelength mode, the signals are
displayed in order from shortest to longest wavelengths. The Agilent 86120B can
measure up to 100 laser lines simultaneously. Use the and softkeys to move
the cursor through the list of signals; the list can contain up to 100 entries. Press
the SELECT key, and the display changes to peak wavelength mode with the signal
at the cursor displayed.
Annotation in the upper right corner of the display indicates whether the signals are
ordered according to wavelength (BY WL) or power (BY PWR). The cursor shows
the currently selected laser line. As you scroll through the responses, the current
position of the selection cursor is shown along the screen’s right side.
Display after “List by WL” key pressed
Also notice that power bars graphically show the relative power levels between laser
lines.
To display multiple laser lines
1 Connect the fiber-optic cable to the front-panel OPTICAL INPUT connector.
2 Press the green Preset key.
3 Press List by WL to display the laser lines from the shortest wavelength to the longest
wavelength.
4 Press List by Power to display the laser lines in order of decreasing amplitudes.
42
Using the Multi-Wavelength Meter
λ
avg
Piλ
i
i1=
n
P
i
i1=
n
--------------------
=
P
total
P
i
i1=
n
=
Displaying Wavelength and Power
Total power and average wavelength
In the third available display mode, the Agilent 86120B displays the average wavelength as shown in the following figure. The displayed power level is the total input
power to the instrument. It is the sum of the powers of each laser line; it is not a
measure of the average power level of the laser lines.
The following equation shows how individual wavelengths of laser lines are
summed together to obtain the average wavelength value:
where,
n is the number of laser lines included in the average.
Pi is the peak power of an individual laser line. Power units are in Watts (linear).
The following equation shows how individual powers of laser lines are summed
together to obtain the total power value:
where,
n is the number of laser lines included in the measurement.
Pi is the peak power of an individual laser line. Power units are in Watts (linear).
43
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
To display average wavelength and total power
• Press the Avg WL key.
Measuring lasers between 700 nm and 1200 nm
After the Preset key is pressed, the input wavelength range is limited to measuring
lasers between 1200 nm and 1650 nm. This prevents the accidental display of spurious signals that may not exist. You can easily expand the input range to the full 700
nm to 1650 nm range, however you should learn how to identify spurious signals.
Spurious signals below 1200 nm may be displayed whenever low-power laser lines
(power levels near the Agilent 86120B’s specified sensitivity) are present at the
input. For example, a low-power laser line at 1550 nm has a second harmonic line at
775 nm. If this second harmonic is above the peak threshold level relative to the fundamental line, it is considered a peak. Its displayed power level may be greater than
that of the fundamental because the amplitude correction at 775 nm is much greater
(by about 15 dB) than that at 1550 nm (the interferometer is less sensitive at 775
nm).
You can also avoid displaying this second harmonic line by reducing the peak
threshold below its preset value. Because the peak threshold level is used to determine which signals are to be displayed before amplitude corrections are applied, the
harmonic will be eliminated. Refer to “Defining Laser-Line Peaks” on page 52.
To use the full wavelength range
1 Press the Setup key.
2 Press the WL LIM softkey.
3 Press LIM OFF to remove the limits on wavelength range. All responses in the full
700 nm to 1650 nm range are now displayed.
Limiting the wavelength range
The wavelength range of measurement can be limited with the wavelength limit
function. Both start and stop wavelengths can be chosen. The units of wavelength
start and stop are the same as the currently selected wavelength units. If wavelength
44
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
units are later changed, the start and stop wavelength units will change accordingly.
Note that a start wavelength limit in nm will become a stop wavelength limit if THz
-1
or cm
is chosen. See “To change the units of measure” on page 49.
The wavelength limit can be useful when laser modulation causes spurious wavelengths to be displayed. Reducing the wavelength range to the region of interest
minimizes the number of spurious wavelengths displayed. Also, the graphical display uses these start and stop wavelength values to plot the power spectrum, whether
the wavelength limit function is on or off.
Preset will turn the wavelength limit on and will set the start wavelength to 1200 nm
and the stop wavelength to 1650 nm.
To limit the wavelength range
1 Press the Setup key.
2 Press the WL LIM softkey.
3 Press the LIM ON softkey if it is not already highlighted.
4 Press the STARTWL softkey to adjust the start wavelength value.
5 Press the STOP WL softkey to adjust the stop wavelength value.
45
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Measuring broadband devices and chirped lasers
When first turned on (or the green Preset key is pressed), the Agilent 86120B is configured to measure narrowband devices such as DFB lasers and modes of FP lasers.
If you plan to measure broadband devices such as LEDs, optical filters, and chirped
lasers, use the Setup menu first to reconfigure the instrument. When broadband
devices are selected, the display shows the BROAD annotation on the screen.
The measurement algorithm for broadband devices determines the wavelength
based upon the center-of-mass of the power spectrum. The
is used to determine the value of the integration limits. Care must be taken to ensure
that the integration limits are above any noise. This is especially true when measuring devices with sloping noise floors, like an EDFA amplifier. For more information
on peak excursion, refer to “Defining Laser-Line Peaks” on page 52.
Instrument specifications apply when the Agilent 86120B is configured to measure
narrowband devices. Specifications do not apply when the instrument is configured
to measure broadband devices.
This feature applies to Agilent 86120B instruments with firmware version number
2.0. When first turned on, the instrument briefly displays the firmware version.
Instruments with a firmware version number less than 2.0 do not have this fea
ture.
peak excursion function
-
To measure broadband devices
1 Press the Setup key.
2 Press MORE twice, and then the DEVICE softkey.
3 Press the BROAD softkey.
To return to measuring narrowband devices, press NARROW.
46
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Graphical display of optical power spectrum
A graphical display of optical power versus wavelength is shown from the start
wavelength value to the stop wavelength value. The start wavelength value is shown
in the upper left corner of the graphical display, and the stop wavelength value is
shown in the upper right corner of the graphical display. The power scale is a fixed
dB scale, with +10 dBm at the display top and –53 dBm at the display bottom. The
power scale is not affected by the Power Offset value. In most cases, the noise floor
will be visible if the total input power is greater than about –5 dBm.
The Agilent 86120B graphical display.
The Peak Threshold value is displayed as a dotted line. All peaks above this dotted
line are displayed in the List by Wavelength and List by Power modes. All peaks
below this line are not displayed. Adjust the Peak Threshold value with the Setup
key, and the THRSHLD softkey.
The wavelength limit start and stop wavelength values are used for the graphical
display even if the wavelength limit function is off.
The graphical display cannot be printed.
To see the graphical display
1 Press the List by WL or List by Power key.
2 Press the GRAPH softkey.
3 To exit the graphical display, press any softkey.
47
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Instrument states
Four different instrument states can be saved and recalled at a later time. The actual
instrument conditions that are saved are identical to those saved from the previous
state after power is turned on. These conditions are shown in Table 8-23 on
page 8-244. If drift measurements or an application (such as signal-to-noise) is on
when an instrument state is saved, it is off when that state is recalled.
To save an instrument state
1 Press the Setup key.
2 Press the SAV/RCL softkey.
3 Press the SAVE softkey.
4 Press one of the four SAVE softkeys to save the instrument state.
To recall a state
1 Press the Setup key.
2 Press the SAV/RCL softkey.
3 Press the RECALL softkey.
4 Press one of the four RCL softkeys to recall an instrument state.
Power bar
To control the power bar
1 Press the Setup key.
2 Press MORE twice, and then PWR BAR.
3 Press BAR ON to display the power bar, and press BAR OFF to hide the power bar
display.
48
Using the Multi-Wavelength Meter
Changing the Units and Measurement Rate
Changing the Units and Measurement Rate
This section includes step-by-step instructions for changing the units and measurement rate.
This section includes:
Displayed units 49
Measurement rate 50
Continuous or single measurements 51
Displayed units
As described below, it’s easy to change the wavelength and amplitude units. You can
choose between the following units:
Table 2-4. Available Units
WavelengthPower
nmdBm
–
1
cm
THz
To change the units of measure
1 Press Setup.
2 Press the MORE softkey.
3 Press the UNITS softkey.
mW
μW
49
Using the Multi-Wavelength Meter
Changing the Units and Measurement Rate
4 Press WL and select one of the following units. Then, press RETURN to complete
your selection:
• NM for nanometers
• THZ for terahertz
• CM –1 for wave number
5 Press POWER and select one of the following units:
• DBM for decibels relative to a milliwatt
• MW for milliwatts
• UW for microwatts
Measurement rate
Under normal operation, the Agilent 86120B makes a measurement and displays the
results about once every second. It is in this normal update mode that maximum
accuracy and wavelength resolution are achieved. However, should a faster update
be desired, for example when real-time feedback is required to tune a laser to its designated channel, the Agilent 86120B can be set to update approximately three times
per second. This reduces both wavelength resolution and accuracy but can be beneficial in some applications.
When FA S T update is selected, one less digit of resolution is displayed. Also, if multiple wavelengths are present, these individual responses, with the reduced resolution, may no longer be recognized.
To change the measurement speed
1 Press the Setup key.
2 Press the MORE softkey.
3 Press the UPDATE softkey.
4 Select either NORMAL or FA S T .
50
Using the Multi-Wavelength Meter
Changing the Units and Measurement Rate
Continuous or single measurements
The Agilent 86120B continuously measures the input spectrum at the front-panel
OPTICAL INPUT connector. Whenever measurements are being acquired, an asterisk (*) is displayed in the display’s upper right corner. When you switch between
normal and fast update modes the rate that the asterisk blinks changes.
You can specify that the instrument perform a measurement only when the frontpanel Single key is pressed. This is the single-acquisition measurement mode, and it
is useful for capturing and preserving data. After capturing the data, you can display
it using many of the procedures included in this chapter. You can return to continuous measurement mode at any time by pressing the Cont key.
To select single measurement acquisition
• Press the Single key.
51
Using the Multi-Wavelength Meter
8 dBm–2 dBm 10 dB –=()
Defining Laser-Line Peaks
Defining Laser-Line Peaks
The Agilent 86120B uses two rules to identify valid laser-line peaks. Understanding
these rules is essential to getting the most from your measurements. For example,
these rules allow you to “hide” AM modulation sidebands or locate laser lines with
small amplitudes.
In order to identify a laser line, the laser-line must meet both of the following rules:
• Power must be greater than the power established by the peak threshold limit
• Power must rise and then fall by at least the peak excursion value
In addition, the input wavelength range can be limited as described in this section.
Peak threshold limit The peak threshold limit is set by subtracting the peak threshold value from the
power of the largest laser line. So, if the largest laser line is 2 dBm and the peak
threshold value is 10 dB, the peak threshold limit is –8 dBm
. You can set the peak threshold value between 0 to 40
dB.
The peak threshold’s default value is 10 dB. This ensures that any modulated signals
being measured are not confused with their AM sidebands. For unmodulated lasers,
or Fabry-Perot lasers, it may be desirable to increase this threshold to look for
responses that are more than 10 dB from the peak.
Peak threshold can be used to suppress spurious signals. For example, a laser that is
amplitude modulated in the audio frequency range can cause spurious wavelengths
to be displayed below and above the correct wavelength. The power of these spurious wavelengths is below that of the correct wavelength. These spurious signals can
be eliminated by decreasing Peak threshold from its Preset value.
Peak excursionThe peak excursion defines the rise and fall in amplitude that must take place in
order for a laser line to be recognized. The rise and fall can be out of the noise, or in
the case of two closely spaced signals, out of the filter skirts of the adjacent signal.
The peak excursion’s default value is 15 dB. Any laser line that rises by 15 dB and
then falls by 15 dB passes the rule. You can set the peak excursion value from 1 to
30 dB.
52
Using the Multi-Wavelength Meter
Defining Laser-Line Peaks
Examples of valid
and invalid signals
In the following figure, three laser lines are identified: responses 1, 3, and 4.
Response ¡ is not identified because it is below the peak threshold. The portion of
each signal that is within the peak excursion limits is shown in bold lines.
Because of the peak excursion rule, responses 4 and 5 are identified as one laser
line—the minimum point between 4 and 5 does not drop to the peak excursion limit.
This response has the highest power shown which is peak 4.
Whenever the peak threshold limit or peak excursion value is changed, the new limits are applied to the current displayed measurements even if the instrument is in the
Single measurement mode.
The following figure shows the same laser lines as the previous figure, but the peakexcursion value has been changed from 15 to 3 dB. Four laser lines are now identified with responses Ð and ƒ identified as two distinct laser lines.
53
Using the Multi-Wavelength Meter
Defining Laser-Line Peaks
Limiting the input
wavelength range
The Agilent 86120B’s preset condition limits the wavelength measurement range
from 1200 nm to 1650 nm. You can expand the wavelength range to cover the entire
700 nm to 1650 nm range. Although wavelength range limiting reduces the number
of laser lines found, its main purpose is to eliminate the identification of second harmonic distortion products as described in the following sidebar.
Distortion caused by low-power laser lines
Low-power laser lines (power level near the Agilent 86120B’s specified sensitivity) may be accompanied by second harmonic (or other) distortion. For
example, a low-power laser line at 1550 nm has a second harmonic line at 775
nm. If this second harmonic is above the peak threshold level relative to the
fundamental line, it is considered a peak. Its displayed power level may be
greater than that of the fundamental because the amplitude correction at 775
nm is much greater (by about 15 dB) than that at 1550 nm (the interferometer is
less sensitive at 775 nm).
To avoid displaying this second harmonic line, limit the input wavelength range
from 1200 nm to 1650 nm.
Or, reduce the peak threshold below its preset value. Because the peak threshold level is used to determine which signals are to be displayed before amplitude corrections are applied, the harmonic will be eliminated.
To define laser-line peaks
1 Press the Setup key.
54
Using the Multi-Wavelength Meter
Defining Laser-Line Peaks
2 Press the THRSHLD softkey.
3 Press PX EXC, and enter the peak excursion value. Use the softkey to select the
digit that requires editing. Use the and softkeys to change the value.
The peak excursion value can range from 1 to 30 dB. The default value is 15 dB.
4 Press RETURN.
5 Press PK THLD and then enter the peak threshold value.
The peak threshold value can range from 0 to 40 dB. Setting this value to 0 dB ensures that only the peak wavelength is identified. The default value is 10 dB.
Pressing the green PRESET key changes the peak excursion and peak threshold values to their default settings. It also turns wavelength range limiting on. Turning the
Agilent 86120B’s power off and then on does not change these settings.
If too many lines are identified
If the following message is displayed, too many laser lines have been identified:
E15 MAX NUMBER OF SIGNALS FOUND
The maximum number of laser lines that the instrument can measure is 100. If
this message appears, decrease the peak threshold value, increase the peak
excursion value, or decrease the wavelength range of operation with the WL
LIM ....START WL and STOP WL functions.
55
Using the Multi-Wavelength Meter
Measuring Laser Separation
Measuring Laser Separation
It is often important to measure the wavelength and power separation between multiple laser lines. This is especially true in wavelength-division-multiplexed (WDM)
systems where channel spacing must be adhered to. The Agilent 86120B can display
the wavelength and amplitude of any laser line relative to another. In fact, the following types of relative measurements can be made compared to the reference:
• Relative wavelength, absolute power
• Relative power, absolute wavelength
• Relative wavelength and power
This section includes:
Channel separation 57
Measuring flatness 58
56
Using the Multi-Wavelength Meter
Measuring Laser Separation
Channel separation
Suppose that you want to measure separation on a system having the spectrum
shown in the following figure.
The Agilent 86120B displays separation on this spectrum as shown in the following
figure. Notice that the 1541.747 nm laser line is selected as the reference. It is shown
in absolute units. The wavelengths and powers of the remaining responses are
shown relative to this reference. For example, the first response is 2.596 nm below
the reference.
To determine channel spacing, simply read the relative wavelength measurement of
the laser lines immediately preceding and following the reference. Use the , ,
and SELECT softkeys to change the reference laser line and read the channel spacing between each channel.
To measure channel separation
1 Press the front-panel Preset key.
57
Using the Multi-Wavelength Meter
Measuring Laser Separation
2 Press List by WL.
3 Press the Delta On key.
Use the Off key to turn off the measurement.
4 Select the type of separation to observe:
• ΔWL displays channel separation.
• ΔWL / ΔPWR displays both channel separation and differences in power.
5 Use the and softkeys to select the reference laser line.
6 Press SELECT.
Press SELECT at any time to select a new reference. Press RESET at any time to turn
off the delta calculation.
Measuring flatness
You can use relative power measurements to measure flatness (pre-emphasis) in a
WDM system. Simply select one carrier as the reference and measure the remaining
carriers relative to the reference level. The power differences represent the system
flatness. Press RESET to turn off the delta calculations so that all responses are
shown in absolute wavelength and powers.
58
Using the Multi-Wavelength Meter
Measuring Laser Separation
To measure flatness
1 Press the front-panel Preset key.
2 Press List by Power.
This lists the input signals by power with the largest response listed first.
3 Press the Delta On key.
4 Select Δ PWR.
5 Use the and softkeys to select the first laser line.
6 Press SELECT.
7 Since the largest power signal is the reference, the relative power measurements for
the other responses shows system flatness.
59
Using the Multi-Wavelength Meter
spacing6
10–
×10Fλ
2
=
Measuring Modulated Lasers
Measuring Modulated Lasers
Lasers modulated at
low frequencies
A laser that is amplitude modulated at low frequencies (for example, modulated in
the audio frequency range) can cause spurious wavelengths to be displayed below
and above the correct wavelength. The power of these spurious wavelengths is
below that of the correct wavelength. These spurious signals can be eliminated by
decreasing the peak threshold. Refer to “Defining Laser-Line Peaks” on page 52.
Even when the laser is amplitude modulated, the correct wavelength and power is
displayed.
The spurious wavelengths caused by low frequency amplitude modulation will be
located above and below the correct wavelength by the following wavelength spacing:
where F is the modulation frequency in Hz, and λ is the correct wavelength in nm.
For example, an amplitude modulation of 10 kHz on a 1550 nm laser will produce
spurious wavelengths spaced by 15 nm from the correct wavelength, and the spurious wavelengths will be at 1535 and 1565 nm.
Low frequency (10 kHz) AM modulation graph showing rounded sideband spurs.
60
Using the Multi-Wavelength Meter
Measuring Modulated Lasers
The graphical display is useful for locating these spurious wavelengths. Their amplitude will be below that of the correct wavelength and they will be broad, rounded
peaks compared to the sharp peak of the correct wavelength. Use the Peak Threshold function to place the dotted line above the spurious peaks so they will not be displayed in the List by WL or List by Power table.
Lasers modulated at
high frequencies
A laser modulated at high frequency (in the RF or microwave range) can also cause
spurious wavelengths to be displayed, especially when the modulation is of a repetitive nature such as that of PRBS or SONET digital formats. In general, no spurious
wavelengths will be displayed using preset instrument conditions. The preset condition includes peak excursion, peak threshold, and wavelength range limiting. However, increasing peak threshold can cause spurious wavelengths to be displayed. To
control the wavelength range, refer to “” on page 44.
Even when the laser being tested is modulated with repetitive formats, the carrier’s
correct wavelength and power is displayed; the wavelength and power of the spurious sidebands are incorrect.
The graphical display is useful to see the effects of high frequency modulation.
Without modulation, the noise floor is typically 45 dB below the laser power. In
general, high frequency modulation will raise the noise floor to about 25 dB below
the laser power. The noise floor is typically flat, or white. The actual level of the
noise floor depends on the type of data format and the data rate.
Directly modulated
lasers
PRBS modulation graph showing raised noise floor.
Directly modulated lasers exhibit a linewidth that is broadband. To measure directly
modulated lasers, refer to “Measuring broadband devices and chirped lasers” on
page 46.
61
Using the Multi-Wavelength Meter
Measuring Total Power Greater than 10 dBm
Measuring Total Power Greater than 10 dBm
The maximum total power that can be measured by the Agilent 86120B is 10 dBm.
However, with the addition of an external attenuator, more power can be applied.
This may be necessary at the transmit end of a wavelength-division-multiplexed
system where large signal levels are present. By entering an amplitude offset equal
to the amount of attenuation at the instrument’s input, accurate amplitude measurements are shown on the display. Additional amplification can also be accounted for.
To measure total power exceeding 10 dBm
CAUTIONThe maximum total input power that can be applied to the Agilent 86120B
before damage occurs is 18 dBm. The maximum total input power that can be
measured is 10 dBm.
1 Connect an optical attenuator between the front-panel OPTICAL INPUT connector
and the fiber-optic cable.
The attenuator must reduce the total input power to the Agilent 86120B so that it is
below +10 dBm.
2 Press Setup, MORE, CAL, and then PWR OFS.
Notice that the PWR OFS annotation appears on the screen to indicate an offset is
applied.
3 Use the softkey to select the digit that requires editing.
4 Use the and softkeys to change the value.
Power offset values are added to the display power readings. For example, if you
placed a 10 dB attenuator on the front-panel connector, enter a power offset value of
+10 dB. Negative values can also be entered if you connect an amplifier instead of
an attenuator.
62
Using the Multi-Wavelength Meter
Calibrating Measurements
Calibrating Measurements
The wavelength of light changes depending on the material that the light is passing
through. To display meaningful wavelength measurements, the Agilent 86120B performs two steps:
1 Measures the wavelength in air.
2 Converts the wavelength to show values in either a vacuum or “standard air”.
For example, a laser line with a wavelength of 1550.000 nm in a vacuum would
have a wavelength in standard air of 1549.577 nm.
Because all measurements made inside the Agilent 86120B are performed in air, the
density of air, due to elevation, affects the wavelength results. You must calibrate
the Agilent 86120B by entering the elevation. Elevations from 0 to 5000 meters can
be entered. The elevation correction is immediately applied to the current measurement even if the instrument is in the single measurement acquisition mode.
Annotation on the display shows the current calibration elevation in meters and
whether the wavelength measurements are shown for a vacuum (VAC) or standard
air (STD AIR).
If you select frequency instead of wavelength measurements, switching between
vacuum and standard air will not affect the measurement results. This is because the
frequency of an optical signal does not change in different mediums—only the
wavelength changes.
Definition of standard air
Standard air is defined to have the following characteristics:
5 Use the and softkeys to enter the elevation in meters. Entries jump in 500
meter steps from 0 m to 5000 m.
In order for the Agilent 86120B to meet its published specifications, the elevation
value selected with the softkeys must be within 250 meters of the actual elevation.
6 Press RETURN to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by
using the following equation:
To select the medium for light
1 Press the Setup key.
2 Press the MORE softkey.
3 Press the CAL softkey, and make the following selection:
• Press VAC UUM for wavelengths in a vacuum.
• Press STD AIR for wavelengths in standard air.
4 Press RETURN to complete the entry.
64
Using the Multi-Wavelength Meter
Printing Measurement Results
Printing Measurement Results
Measurement results can be sent directly to a printer. Simply connect a compatible
printer to the rear-panel PARALLEL PRINTER PORT connector. The output is
ASCII text. An example of a compatible printer is Hewlett-Packard
series printer. Be sure to use a parallel printer cable to connect the printer.
The printer output is not a copy of the display. Rather, it is a listing of all signals
present at the input (up to 100). The measurement values printed depend on the settings of the instrument when the Print key is pressed.
The following is an example of a typical printout:
Agilent 86120B SER US36151025
Firmware Ver. 2.000
List By Wavelength
8 Lines
Power Offset 0.0 dB
Vacuum
Elevation 0 Meters
Update Normal
Peak Excursion 15 dB
Peak Threshold 10 dB
Device Narrow
®
’s L ase rJet1
Input
WavelengthPower
---------------------
1280.384nm-16.97dBm
1281.473-13.14
1282.569-13.92
1283.651-13.34
1284.752-11.69
1285.840 -8.11
1286.944-10.38
1288.034-14.65
To create a hardcopy
1 Connect the printer to the Agilent 86120B’s rear-panel PARALLEL PRINTER
PORT connector.
2 Press Print.
1. Hewlett-Packard and LaserJet are registered trademarks of Hewlett-Packard Company.
65
You can use the ABORT and CONT softkey to stop and restart a print job that is in
progress.
In this chapter, you’ll learn how to make a variety of fast, accurate measurements
using the measurement tools accessed by pressing the Appl’s key.
68
Measurements Applications
Measuring Signal-to-Noise Ratios
Measuring Signal-to-Noise Ratios
Signal-to-noise measurements provide a direct indication of system performance.
Signal-to-noise measurements are especially important in WDM systems because
there is a direct relation between signal-to-noise and bit error rate. The
Keysight 86120B displays signal-to-noise measurements in the third column. For
example, the selected signal in the following figure has a signal-to-noise ratio of
30.0 dB.
Signal-to-noise display.
During a signal-to-noise measurement, the absolute power of the carrier, in dBm, is
compared to the absolute power of the noise at the carrier wavelength. See the following figure. The noise power at the carrier must be determined by interpolation
because the carrier, in most cases, can not or should not be turned off.
You can select one of two methods used to determine the wavelength where the
noise is measured: automatic interpolation or a user-entered wavelength. In the figure above, notice that “S/N AUTO” is displayed to indicate that automatic interpolation is selected.
69
Measurements Applications
Measuring Signal-to-Noise Ratios
Location of noise measurements
Automatic
interpolation
When the signal-to-noise “auto” function is selected, the Keysight 86120B first
determines the proximity of any adjacent signal. If the next closest signal is ≤200
GHz (approximately 1.6 nm at 1550 nm) away from the signal of interest, then the
noise power is measured half way between the two channels and an equal distance
to the other side of the signal of interest. See points P
and Pn2 in the following fig-
n1
ure.
If the closest signal is more than 200 GHz from the signal of interest, or if there is no
other signals present, then the noise power is measured at 100 GHz on either side of
the signal of interest. The two measured noise power levels are then averaged to
estimate the noise power level at the signal wavelength. The noise power measurements use linear interpolation to estimate the noise power level at the signal of interest’s wavelength.
Automatic interpolation
70
Measurements Applications
Measuring Signal-to-Noise Ratios
User-entered
wavelength
When the signal-to-noise “user” function is selected, the Keysight 86120B uses only
one wavelength to measure the noise power for all signals. This wavelength is set by
the user and all signals are compared to the noise level at this wavelength to determine their corresponding signal-to-noise ratios.
Noise bandwidthWhen measuring noise power, the Keysight 86120B must account for the noise
bandwidth used during the measurement. Because noise bandwidth varies with measurement bandwidth (a wide bandwidth allows more noise to the Keysight 86120B’s
detector than a narrow bandwidth), the Keysight 86120B normalizes all noise power
measurements to a bandwidth of 0.1 nm. The annotation 0.1 nm is displayed to
show that the noise bandwidth is being normalized to a 0.1 nm bandwidth.
Repetitive data formats
The Keysight 86120B signal-to-noise application works best when the laser
being tested is not modulated, or modulated with non-repetitive data formats.
With repetitive data formats, such as PRBS data and SONET formats, there is
significant low-frequency amplitude modulation of the laser. This modulation
raises the noise floor of the Keysight
measured can be limited to about 15
86120B significantly. The signal-to-noise
dB while measuring lasers modulated by
repetitive data formats. For improved performance when the laser is modulated
with repetitive data formats, use the Signal-to-Noise with Averaging applica
-
tion.
To measure signal-to-noise
1 Press the front-panel Preset key.
2 Press List by WL or List by Power.
3 Press Appl’s and then S/N.
71
Measurements Applications
Measuring Signal-to-Noise Ratios
4 To select the wavelength reference for measuring the noise, do the following steps:
a Press WL REF, and
•press AUTO to let the instrument interpolate the wavelength,
or
•press USER to select the last wavelength manually entered.
b If you chose USER, you can specify the wavelength by pressing USER WL. Use
the
softkey to select the digit that requires editing. Use the and softkeys
to change the value.
c Press RETURN.
5 While the signal-to-noise measurements are displayed, you can press PEAK anytime
to select the signal with the highest power.
72
Measurements Applications
Measuring Signal-to-Noise Ratios with Averaging
Measuring Signal-to-Noise Ratios with Averaging
When the lasers being measured are modulated, especially with repetitive data formats such as SONET or PRBS, the noise floor is raised. Averaging reduces the noise
floor and allows an improvement of greater than 10 dB in a signal-to-noise measurement. In general, averaging will decrease the noise floor caused by modulation until
the true optical noise level is reached. The displayed signal-to-noise will improve
with each average until the true optical noise level is reached, and then the displayed
signal-to-noise will remain approximately constant. If, however, the true signal-tonoise is below the instrument sensitivity of approximately 40 dB (in a 0.1 nm noise
bandwidth), it will not be measured.
Averaging can also improve the accuracy of measuring signal-to-noise of unmodulated lasers.
Signal-to-noise with averaging display.
Averaging is performed on the noise, not on the wavelength or power of the laser
signals.
The signal-to-noise with averaging measurement uses the automatic interpolation
method to determine the wavelengths where the noise is measured. Refer to “Mea-
suring Signal-to-Noise Ratios” on page 69 for a description of automatic interpola-
tion. There is no user-entered wavelength selection in signal-to-noise with
averaging.
During a signal-to-noise with averaging measurement, the display indicates S/N A xx, where A indicates averaging and xx is the number of averages taken so far.
The maximum number of averages is 900, the minimum number of averages is 10,
and the default (Preset) value is 100 averages. A measurement with 100 averages
takes about 2 minutes to complete. When the measurement is complete, the instrument switches to single measurement mode. Then, pressing the Cont key will start a
73
Measurements Applications
Measuring Signal-to-Noise Ratios with Averaging
completely new measurement. During a measurement and before the number of
averages has been reached, pressing the Single key will stop the measurement.
Then, pressing the Cont key will continue with the current measurement.
While making a signal-to-noise with averaging measurement, the number of averages can be changed. As long as the new number of averages is greater than the
number of averages taken so far, the measurement continues. If the new number of
averages selected is less than the number of averages taken so far, the measurement
stops and the instrument switches to single measurement mode. Then, pressing the
Cont key will start a completely new measurement.
Noise bandwidth
affects measurement
When measuring noise power, the Keysight 86120B must account for the noise
bandwidth used during the measurement. Because noise bandwidth varies with measurement bandwidth (a wide bandwidth allows more noise to the Keysight 86120B’s
detector than a narrow bandwidth), the Keysight 86120B normalizes all noise power
measurements to a bandwidth of 0.1 nm. The annotation 0.1 nm is displayed to
show that the noise bandwidth is being normalized to a 0.1 nm bandwidth.
To measure signal-to-noise with averaging
1 Press the front panel Preset key.
2 Press List by WL or List by Power.
3 Press Appl's and then S/N AVG.
4 To change the number of averages, press NUM AVG. The default (Preset) value is
100.
5 To stop the measurement at the current number of averages shown, press the Single
key. Then press the Cont key to continue the present measurement.
6 When the measurement is complete, the instrument will switch to the single
measurement mode and stop.
7 To make a new measurement, press the Cont key.
8 To exit, press the EXIT softkey, then press the Cont key for continuous measurement.
74
Measurements Applications
Measuring Laser Drift
Measuring Laser Drift
In this section, you’ll learn how the Keysight 86120B can be used to monitor drift
(changes to a laser’s wavelength and amplitude over time). Drift is measured simultaneously for every laser line that is identified at the input. The Keysight 86120B
keeps track of each laser line’s initial, current, minimum, and maximum values and
displays their differences relative to itself. This allows the Keysight 86120B to be
used for laser transmitter evaluation, burn-in, or development. In addition, you can
monitor system performance over time, temperature, or other condition.
The following display shows power and wavelength drift measured on five laser
lines. The DRIFT annotation, item ¿, tells you that drift measurements are being
performed. The current relative drift values for wavelength and power are shown in
items ¡ and ¬ respectively. Item Ð indicates the absolute reference values for the
laser line indicated by the cursor . The reference values are measured before the
measurement starts.
You can restart the drift measurement at any time by pressing the RESET softkey.
All minimum and maximum values are reset to the reference values, and the
Keysight 86120B begins to monitor drift from the current laser line values. Move
the cursor up and down the listing to see the reference wavelength and power of
each laser line.
75
Measurements Applications
Measuring Laser Drift
If measurement updating stops or the values become blanked
If, in the middle of a measurement, the number of laser lines present changes,
the measurement stops until the original number of lines returns. You’ll notice
that a CLEAR softkey appears and one of the following message is displayed:
E46 NUM LINES < NUM REFS
E47 NUM LINES > NUM REFS
To view the data measured before the conditions changed, press CLEAR and
then MAX-MIN. Notice that the measurement acquisition is changed from con
tinuous to single.
To restart testing, press CLEAR, the CONT key, and then RESET to use the
new number of lines as the reference. Pressing CONT restarts continuous mea
surement acquisition. Or, you can restore the original number of lines on the
input so that the drift measurement can continue.
To measure drift
1 Press the front-panel Preset key.
-
-
2 Press Peak WL, List by WL, or List by Power to select the display style for observing
drift.
3 Press Appl’s and then DRIFT.
Pressing DRIFT sets the current laser-line values as the reference from which to compare all drift.
4 Press MAX-MIN for the desired type of drift measurement as described in the
following paragraphs:
Display shows the current values of laser lines relative to the
wavelength and power values measured when the test was
begun or the RESET softkey was pressed.
Display shows absolute maximum values since the drift
measurement was started. This measurement gives the longest
wavelength and greatest power measured. The laser line of
interest may have since drifted to a lesser value. Note that the
maximum wavelength and maximum power may not have
occurred simultaneously.
Display shows absolute minimum values since the drift
measurement was started. This measurement gives the shortest
76
Measurements Applications
Measuring Laser Drift
wavelength and smallest power measured. The laser line of
interest may have since drifted to a greater value. Note that the
minimum wavelength and minimum power may not have
occurred simultaneously.
Display shows the total drift from the reference since the drift
measurement was started. Values represent the minimum
wavelength and power drift values subtracted from the
maximum drift values.
5 In the List by WL and List by Power displays, use the and softkeys to view
the reference values (wavelength and power values of each laser line before the test
was started).
During the measurement, you can change the display mode to Peak WL, List by WL,
List by Power, or Avg WL. When List by WL or List by Power is selected, the signal
list is sorted by reference values and not by the current, maximum, or minimum val
ues.
To restart the drift measurements, press RESET. This resets the reference values.
-
77
Measurements Applications
Measuring Coherence Length
Measuring Coherence Length
Coherence length is a measure of the distance over which a laser’s light retains the
phase relationships of its spectrum. The Keysight 86120B measures coherence
length of Fabry-Perot semiconductor diode lasers. The Keysight 86120B cannot
measure coherence length of light emitting diodes (LEDs) or distributed feedback
(DFB) lasers.
When you select coherence length measurements, the Keysight 86120B displays the
following four values:
• Coherence length (Lc)
• Round trip optical length of diode laser cavity (2nLd)
• Alpha factor
• Beta factor
Coherence length in the region of 1 mm to 200 mm can be measured. The following
figure shows a coherence length measurement.
To measure coherence length
1 Press the front-panel Preset key.
2 Press Appl’s and then COH LEN.
78
Measurements Applications
decay curvee
OPD
L
c
------------
–
=
1e⁄
Measuring Coherence Length
Coherence length
(L
)
c
The interferogram of the laser being tested is sampled and the envelope of the interferogram is found. This envelope has peaks (regions of high fringe visibility) at zero
optical path delay and at delays equal to multiples of the laser cavity round-trip optical length. This is shown in the following figure of the interferogram envelope:
The amplitudes of the peaks decreases exponentially from the largest peak at zero
path delay. The exponential decay constant is defined as the coherence length, Lc.
The curve that connects the tops of the envelope peaks is given by the following
equation:
OPD is the optical path delay and Lc is the coherence length. Thus, at an optical
path delay equal to the coherence length, the envelope peaks are down to of
their value at zero path delay peak. All envelope peaks found are used to determine
the exponential decay constant (coherence length) using a least squares fit.
Round trip optical
length of diode laser
The average optical path delay spacing of the envelope peaks is measured. This is
equal to the diode laser cavity round trip optical length, 2nLd.
cavity (2nLd)
Alpha factorThe alpha factor is defined as the height of the first envelope peak away from zero
path delay relative to the height of the envelope peak at zero path delay. The alpha
factor is always between 0 and 1.
79
Measurements Applications
Alpha factor
V
A
V
o
------
=
Beta factor
V
B
V
o
------
=
Measuring Coherence Length
The smaller the alpha factor, the shorter the coherence length.
Beta factorThe beta factor is defined as the height of the fringe visibility envelope midway
between the zero optical path delay peak and the next peak relative to the height of
the envelope peak at zero path delay. The beta factor is always between 0 and 1.
The smaller the beta factor, the more longitudinal modes (wavelengths) the laser
has.
80
4
Addressing and Initializing the Instrument 83
To change the GPIB address 83
Making Measurements 85
Commands are grouped in subsystems 87
Measurement instructions give quick results 89
The format of returned data 95
Monitoring the Instrument 96
Status registers 96
Queues 101
Reviewing SCPI Syntax Rules 103
Example Programs 108
Example 1. Measure a DFB laser 110
Example 2. Measure WDM channels 112
Example 3. Measure WDM channel drift 114
Example 4. Measure WDM channel separation 117
Example 5. Measure SN ratio of WDM channels 119
Example 6. Increase a source’s wavelength accuracy 121
Lists of Commands 123
Programming
Programming
Programming
Programming
This chapter explains how to program the Keysight 86120B. The programming syntax conforms to the IEEE 488.2 Standard Digital Interface for Programmable Instrumentation and to the Standard Commands for Programmable Instruments (SCPI).
Where to begin…
The programming examples for individual commands in this manual are written in
®1
BASIC 6.0 for an HP 9000 Series 200/300 Controller.
HP
For more detailed information regarding the GPIB, the IEEE 488.2 standard, or the
SCPI standard, refer to the following books:
Hewlett-Packard Company. Tutorial Description of Hewlett-Packard Interface Bus,
1987.
Hewlett-Packard Company. SCPI—Standard Commands for Programmable Instru-ments, 1995.
International Institute of Electrical and Electronics Engineers. IEEE Standard 488.1-1987, IEEE Standard Digital Interface for Programmable Instrumentation. New
York, NY, 1987.
International Institute of Electrical and Electronics Engineers. IEEE Standard 488.2-
1987, IEEE Standard Codes, Formats, Protocols and Common commands For Use
with ANSI/IEEE Std 488.1-1987. New York, NY, 1987.
Types of commands
The Keysight 86120B responds to three types of commands:
• Common commands
• Measurement instructions
• Subsystem commands
All of these commands are documented in Chapter 5, “Programming Commands”.
1. HP is a registered trademark of Hewlett-Packard Company.
82
Programming
Addressing and Initializing the Instrument
Addressing and Initializing the Instrument
The Keysight 86120B’s GPIB address is configured at the factory to a value of 20.
You must set the output and input functions of your programming language to send
the commands to this address.
To change the GPIB address
1 Press the Setup key.
2 Press MORE twice, then GPIB.
3 Use the and softkeys to change the GPIB address.
4 Press RETURN.
Remote mode and front-panel lockout
Whenever the instrument is controlled by a computer, the Remote message is displayed on the instrument’s screen and the softkey menu is blanked except for the
LOCAL softkey. This softkey can be pressed by the user to restore front panel control of the instrument.
You can specify a local lockout mode that prevents the LOCAL softkey from being
displayed. If the instrument is in local lockout mode, all the softkeys may be
blanked. For example, if the instrument is first placed in local lockout mode and
then placed in remote mode, no softkeys are displayed.
Consult the documentation for your programming environment to determine which
commands are used to put an instrument in the remote and local lockout modes.
These are not Keysight 86120B commands; they control GPIB control lines and do
not send any characters to the Keysight 86120B.
Initialize the instrument at start of every program
It is good practice to initialize the instrument at the start of every program. This
ensures that the bus and all appropriate interfaces are in a known state. HP BASIC
provides a CLEAR command which clears the interface buffer and also resets the
instrument’s parser. (The parser is the program that reads the instructions that you
send.) Whenever the instrument is under remote programming control, it should be
83
Programming
Addressing and Initializing the Instrument
in the single measurement acquisition mode. This is automatically accomplished
when the *RST common command is used. The *RST command initializes the
instrument to a preset state:
CLEAR 720
OUTPUT 720;”*RST”
Notice in the example above, that the commands are sent to an instrument address of
720. This indicates address 20 on an interface with select code 7. Pressing the green
Preset key does not change the GPIB address.
Set single acquisition mode
An advantage of using the *RST command is that it sets the Keysight 86120B into
the single measurement acquisition mode. Because the READ and MEASure data
queries expect this mode, their proper operation is ensured.
Establish the wavelength range
At the start of each program, be sure to establish the input wavelength range using
the Keysight 86120B’s :CALCulate2:WLIMit command. Setting this command to
off enables the full wavelength range of the instrument. If you are measuring signals
over a narrow wavelength range, use this command to ensure that spurious second
harmonic peaks are not identified. Refer to “WLIMit[:STATe]” on page 166,
“WLIMit:STARt[:WAVelength]” on page 169, and “WLIMit:STOP[:WAVelength]”
on page 172. Refer also to “To limit the wavelength range” on page 45.
84
Programming
Making Measurements
Making Measurements
Making measurements remotely involves changing the Keysight 86120B’s settings,
performing a measurement, and then returning the data to the computer. The simplified block diagram of the Keysight 86120B shown here lists some of the available
programming commands. Each command is placed next to the instrument section it
configures or queries data from.
Notice that there are two buffers from which data can be queried: an uncorrected
data buffer and a corrected data buffer. With each scan of the input wavelength
range, the analog-to-digital converter loads 65,536 data values into the uncorrected
data buffer. This is considered to be one “measurement”. A fast-update measurement mode is available for quicker measurement acquisition. But, because only
8,192 data values are collected in fast-update measurement mode, the ability to
resolve closely spaced signals is reduced.
After collecting the uncorrected data, the Keysight 86120B searches the data for the
first 100 peak responses. (Searching starts at 1700 nm and progresses towards 700
nm for WLIMit OFF. If WLIMit is on, searching starts at WLIMit:STARt to
85
Programming
Making Measurements
WLIMit:STOP). These peak values are then placed into the corrected data buffer.
Each peak value consists of an amplitude and wavelength measurement. Amplitude
and wavelength correction factors are applied to this data.
For a listing of the programming commands (including a cross reference to frontpanel keys), refer to the following tables:
Table 4-11, “Programming Commands,” on page 4-123
Table 4-12, “Keys Versus Commands,” on page 4-128
86
Programming
Making Measurements
Commands are grouped in subsystems
The Keysight 86120B commands are grouped in the following subsystems. You’ll
find a description of each command in Chapter 5, “Programming Commands”.
CALCulate2 Queries corrected peak data and sets wavelength limits.
CALCulate3 Performs delta, drift, and signal-to-noise measurements.
DISPlay Applies markers and displays power bars.
HCOPy Prints measurement results.
SENSe Sets elevation-correction values, selects readings for air or
Perform frequency, wavelength, wavenumber, and coherence
length measurements.
vacuum, and enters amplitude offsets. Configures instrument
for measuring broadband devices and chirped lasers. Queries
time-domain values of the input data.
STATus Queries instrument status registers.
SYSTem Presets Keysight 86120B and queries error messages.
TRIGger Stops current measurement. Acquires new measurement data.
Also used to select single or continuous acquisition of
measurement data.
UNIT Sets the amplitude units to watts or dBm.
Table 2-5 on page 4-88 shows the kinds of measurements that the Keysight 86120B
can perform and the associated programming commands used to return that data. In
some cases, there is more than one method that can be used to obtain the desired
data. Refer to Chapter 5, “Programming Commands” for the correct syntax for these
commands.
87
Programming
Making Measurements
Table 2-5. Commands for Capturing Data
Desired
Measurement
Wavelength (nm)CONFigure, FETCh, READ, and
Frequency (THz)CONFigure, FETCh, READ, and
Wavenumber (m–1)
Coherence Length (m)CONFigure, FETCh, READ, and
Power (W, dBm)CONFigure, FETCh, READ, and
Average Wavelength,
Wavenumber, or Frequency
Total Power (W, dBm)CALCulate2:PWAVerage:STATeCALCulate2:DATA?
Signal-to-Noise Ratio AverageCALCulate3:ASNR:STATeCALCulate3:DATA?
Time-Domain DataCALCulate1:TRANsform:FREQuency:P
Corrected Frequency Domain
Data
Uncorrected Frequency
Domain Data
Command to Configure Measurement
(partial listing)
MEASure
MEASure
CONFigure, FETCh, READ, and
MEASure
MEASure
MEASure
CALCulate2:PWAVerage:STATeCALCulate2:DATA?
OINts
CALCulate1:TRANsform:FREQuency:P
OINts
CALCulate1:TRANsform:FREQuency:P
OINts
Command to Query Data
MEASure:ARRay:POWer:WAVelen
gth?
MEASure:ARRay:POWer:FREQuen
cy?
MEASure:ARRay:POWer:WNUMb
er?
FETCh, READ, or MEASure
MEASure:ARRay:POWer?
SENSe:DATA?
CALCulate2:DATA?
CALCulate1:DATA?
88
Measurement instructions give quick results
The easiest way to measure wavelength, frequency, power, or coherence length is to
use the MEASure command. The MEASure command is one of four measurement
instructions: MEASure, READ, FETCh, and CONFigure. The syntax for measurement instructions is documented in “Measurement Instructions” on page 146.
Each measurement instruction has an argument that controls the measurement
update rate. This is equivalent to using the NORMAL and FA S T softkeys.
:MEASure command
MEASure configures the Keysight 86120B, captures new data, and queries the data
all in one step. For example, to measure the longest wavelength, send the following
command:
:MEASure:SCALar:POWer:WAVelength? MAX
Table 2-6. The Different Forms of MEASure
Programming
Making Measurements
Desired
Measurement Data
Power (W, dBm):MEASure:ARRay:POWer?List by Power
Frequency (Hz):MEASure:ARRay:POWer:FREQuency?List by WL (frequency)
Wavelength (m)MEASure:ARRay:POWer:WAVelength?List by WL
Specifying SCALar places the display in the single wavelength format and returns a
single value to the computer. Specifying ARRay places the display in the List by
Power or List by WL modes; an array of data is returned to the computer.
89
Programming
Making Measurements
A common programming error is to send the :MEASure command when the instrument is in the continuous measurement acquisition mode. Because :MEASure contains an :INIT:IMM command, which expects the single measurement acquisition
mode, an error is generated, and the INIT command is ignored.
:READ command
The READ command works like the MEASure command except that it does not
configure the instrument’s settings. You can use the CONFigure command to configure the instrument for a particular measurement without returning any data.
The MEASure and READ commands are identical to combining the following commands:
Command Equivalent Commands
:MEASure :ABORt;:CONFigure;:READ
:READ :ABORt;:INITiate:IMMediate;:FETCh
A common programming error is to send the :READ command when the instrument
is in the continuous measurement acquisition mode. Because :READ contains an
:INIT:IMM command, which expects the single measurement acquisition mode, an
error is generated, and the INIT command is ignored.
:FETCh command
The FETCh command returns data from previously performed measurements; it
does not initiate the collection of new data. Because FETCh does not configure the
instrument or acquire new input data, you can use FETCh repeatedly on the same set
of acquired data. For example, use two FETCh commands to return wavelength and
then power values for the same measurement. This is shown in the following program fragment:
OUTPUT 720;”:INIT:CONT OFF;”
OUTPUT 720;”:CONF:ARR:POW MAX”
OUTPUT 720;”:INIT:IMM”
OUTPUT 720;”:FETC:ARR:POW?”
ENTER 720:powers$
OUTPUT 720;”:FETC:ARR:POW:WAV?”
ENTER 720:wavelengths$
In the example above, the data in the power and wavelength arrays are returned in
the same order so that powers can be matched to wavelengths.
Also, because new data is not collected, FETCh is especially useful when characterizing transient data.
90
Programming
Making Measurements
FETCh does not reconfigure the display. For example, if the display is in the Peak
WL mode, sending :FETCh:ARRay does not configure the display to the List by
WL even though an array of data is returned to the computer.
A common programming error occurs when the :FETCh command is used after an
*RST command. This generates error number –230, “Data corrupt or stale”. In this
instance, you must send :INIT:IMM after the *RST command and before :FETCh
command to capture a new array of measurement data.
:CONFigure command
The CONFigure command changes measurement settings without taking a measurement. The instrument is placed in the List by WL, List by Ampl, Peak WL display,
or in the coherence length application.
CONFigure can be queried. The query returns the last configuration setup by the
CONFigure command. The instrument returns a string which is the last instrument
function sent by a CONFigure command or MEASure query. The returned string is
in the short command form. Use caution when using this query, because if any
instrument settings were changed since the last CONFigure command or MEASure
query these changes may not be included in the returned string.
For example, if the last CONFigure command was
:CONFigure:SCALar:POWer:WAVelength 1300NM, MAX
a CONFigure? query would return a string that is similar to the following line:
“POW:WAV 1.300000e-6,0.01”
The 1300NM and resolution values track the actual instrument settings and input
signals. Notice that the quotation marks are part of the returned string.
Return single or multiple measurement values
You can specify whether FETCh, READ, or MEASure returns a single value (SCALar) or multiple values (ARRay). The following example specifies SCALar data
which returns a single value.
:MEASure:SCALar:POWer:WAVelength? MAX
91
Programming
Making Measurements
ARRay and the SCPI standard
According to the SCPI command reference, ARRay command causes an instrument to take multiple measurements. (A <size> parameter indicates the number
of measurements to take.) However, the Keysight
86120B’s ARRay command
refers to the measurements performed for one measurement sweep; this results
in an array of measured signals. Because the <size> parameter does not apply,
any <size> parameter sent will be ignored by the instrument. No syntax error
will be generated if a <size> parameter is sent.
Always force the Keysight 86120B to wait for non-sequential commands
The Keysight 86120B normally processes its remote programming commands
sequentially. The instrument waits until the actions specified by a particular command are completely finished before reading and executing the next command.
However, there are a few non-sequential commands where this is not true. Nonsequential commands do not finish executing before the next command is inter-
preted.
The following is a list of the Keysight 86120B’s non-sequential commands:
The benefit of non-sequential commands is that, in some situations, they can reduce
the overall execution times of programs. For example, you can set the peak excursion, peak threshold, and elevation and use a *WAI command at the end to save
92
Programming
Making Measurements
time. However, non-sequential commands can also be a source of annoying errors.
Always use the *OPC query or *WAI command with the non-sequential commands
to ensure that your programs execute properly.
For example, suppose that you wanted to set the elevation correction value and then
send an :INIT:IMM command. The following programming fragment results in an
error –213 “Init ignored”. This occurs because the :ELEVation
command causes
the recalculation of the data which is like sending the :INIT:IMM command. When
the actual :INIT:IMM is sent, the error occurs because the command is already in
progress.
The :CALCulate3:DRIFt:PRESet command turns off the minimum, maximum, difference, and reference states but leaves the drift state on.
Attempting to turn more than one state on at a time results in a “–221 Settings Con-flict” error.
The *RST and SYSTem:PRESet commands turn all calculations off.
CALCulate3:PRESet turns off any CALCulate3 calculations.
94
Programming
Making Measurements
The format of returned data
Measurements are returned as strings
All measurement values are returned from the Keysight 86120B as ASCII strings.
When an array is returned, the individual values are separated by the comma character.
Determine the number of data points
When a FETCh, READ, or MEASure command is used (with ARRay specified), the
first returned value indicates the total number of measurement values returned in the
query.
If you use the:CALCulate1:DATA?, :CALCulate2:DATA?, or :CALCulate3:DATA?
queries to query data, send the :POINts? query first to determine the number of values returned in the string. The string does not contain a first value which specifies
the string length. This is shown in the following example:
OUTPUT 720;”:CALCulate1:POINts?”
ENTER 720;Length
OUTPUT 720;”:CALCulate1:DATA?”
ENTER 720;Result$
Data can be corrected for elevation and vacuum
Normally, the Keysight 86120B provides measurement values calculated for conditions in air at sea level. Use the :SENSe:CORRection:ELEVation command to compensate for air dispersion. Altitudes up to 5000 meters can be entered. Use the
:SENSe:CORRection:MEDium command to switch to readings in a vacuum.
Amplitude units
The default amplitude units are dBm. If you need measurements in watts, use the
:UNIT:POWer command. When the Keysight 86120B is turned on, the amplitude
units are automatically set to the units used before the instrument was last turned off.
95
Programming
Monitoring the Instrument
Monitoring the Instrument
Almost every program that you write will need to monitor the Keysight 86120B for
its operating status. This includes querying execution or command errors and determining whether or not measurements have been completed. Several status registers
and queues are provided to accomplish these tasks.
In this section, you’ll learn how to enable and read these registers. In addition to the
information in this section, you should review the commands documented in “Com-
mon Commands” on page 133 and “STATus Subsystem” on page 206.
Status registers
The Keysight 86120B provides four registers which you can query to monitor the
instrument’s condition. These registers allow you to determine the following items:
• Status of an operation
• Availability of the measured data
• Reliability of the measured data
All three registers are shown in the figure on the following page and have the following uses:
Register Definition
Status Byte Monitors the status of the other three registers.
Standard Event Status This is the standard IEEE 488.2 register. Contains
bits which indicate the status of the other two regis
ters.
OPERation Status Contains bits that report on the instrument’s normal
operation.
QUEStionable Status Contains bits that report on the condition of the sig-
nal.
96
-
Programming
Monitoring the Instrument
Status Byte register
The Status Byte Register contains summary bits that monitor activity in the other
status registers and queues. The Status Byte Register’s bits are set and cleared by the
presence and absence of a summary bit from other registers or queues. Notice in the
following figure that the bits in the Standard Event Status, OPERation status, and
QUEStionable status registers are “or’d” to control a bit in the Status Byte Register.
If a bit in the Status Byte Register goes high, you can query the value of the source
register to determine the cause.
The Status Byte Register can be read using either the *STB? common command or
the GPIB serial poll command. Both commands return the decimal-weighted sum of
all set bits in the register. The difference between the two methods is that the serial
poll command reads bit 6 as the Request Service (RQS) bit and clears the bit which
clears the SRQ interrupt. The *STB? command reads bit 6 as the Master Summary
Status (MSS) and does not clear the bit or have any effect on the SRQ interrupt. The
value returned is the total bit weights of all of the bits that are set at the present time.
OPERation Status and QUEStionable Status registers
You can query the value of the OPERation Status and QUEStionable Status registers
using commands in the STATus subsystem.
The STATus subsystem also has transition filter software which give you the ability
to select the logic transitions which set bits in the OPERation Status and QUEStionable Status registers. For example, you can define the POWer bit of the QUEStionable Status register to report an event when the condition transitions from false to
true. This is a positive transition. You can also specify a negative transition where
the bit is set when the condition transitions from true to false.
97
Programming
Monitoring the Instrument
98
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.