No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
85059-90005
Edition
Edition 1, August 2017
Printed in USA/Malaysia
Published by:
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
U.S. Government Rights
The Software is “commercial
computer software,” as defined
by Federal Acquisition Regulation
(“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and
Department of Defense FAR
Supplement (“DFARS”) 227.7202,
the U.S. government acquires
commercial computer software
under the same terms by which
the software is customarily
provided to the public.
Accordingly, Keysight provides
the Software to U.S. government
customers under its standard
commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of
which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA
represents the exclusive authority
by which the U.S. government
may use, modify, distribute, or
disclose the Software. The EULA
and the license set forth therein,
does not require or permit,
among other things, that
Keysight: (1) Furnish technical
information related to
commercial computer software
or commercial computer
software documentation that is
not customarily provided to the
public; or (2) Relinquish to, or
otherwise provide, the
government rights in excess of
these rights customarily provided
to the public to use, modify,
reproduce, release, perform,
display, or disclose commercial
computer software or
commercial computer software
documentation. No additional
government requirements
beyond those set forth in the
EULA shall apply, except to the
extent that those terms, rights, or
licenses are explicitly required
from all providers of commercial
computer software pursuant to
the FAR and the DFARS and are
set forth specifically in writing
elsewhere in the EULA. Keysight
shall be under no obligation to
update, revise or otherwise
modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant
to FAR 12.211 and 27.404.2 and
DFARS 227.7102, the U.S.
government acquires no greater
than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5
(c), as applicable in any technical
data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Class Assignments and Standard Definitions Values are Available on the Web. . . . . . . . . . 7-1
Contents
Keysight 85059V Operation and Service Guide 3
Contents
4 Keysight 85059V Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
1General Information
Chapter One at-a-Glance
About this Manual
This manual describes the 85059B calibration kit and provides replacement
part numbers, specifications, and procedures for using, maintaining and
troubleshooting this kit.
Chapter One at-a-Glance
Section TitleSummary of Content
“Calibration Kit Overview” on page 1-3Describes the 85059B calibration kit contents, how to identify shorts,
and compatible network analyzers.
“Offset-Short Calibration Using the
85059B (Precision) Calibration Kit and the
PNA’s Guided Power Calibration” on
page 1-7
“Calibration Residual Error Specifications”
on page 1-8
“Serial Numbers” on page 1-9How to verify your calibration kit’s serial number in a table to avoid
“Incoming Inspection” on page 1-10Verifying your calibration kit’s contents and what to do, if there is an
“Clarifying Connector Gender” on
page 1-12
“Clarifying the Terminology of a
Provides links for performing a Guided Power Calibration and using
the 85059B calibration kit.
Describes the data-based residual error specifications. Refer to
Table 1-1 on page 1-8.
possible confusion with other calibration kits.
problem.
A discussion of how to identify the male and female connectors.
A discussion of some of the connector terminology.
Connector Interface” on page 1-13
“Preventive Maintenance” on
How to avoid damaging the devices in your calibration kit.
page 1-13
“When to Calibrate” on page 1-14
How often should you re-calibrate your network analyzer.
1
1-
General Information
Chapter One at-a-Glance
Section TitleSummary of Content
“Regulatory and Environmental
Information” on page 1-15
Description of regulatory symbols.
1-2 Keysight 85059B Operation and Service Guide
General Information
Calibration Kit Overview
Calibration Kit Overview
The 85059B 1.0 mm precision calibration kit is used to calibrate your PNA
series network analyzer system. This kit is used for the measurement of
components with 50
120 GHz.
Because it is physically impossible to construct a slotless version of the 1.0 mm
female contact, the female devices in this kit use slotted contacts. The slotted
female contact does not have the same electrical characteristics as a solid
conductor, and therefore, the male and female devices in this kit have different
models.
Although the male and female devices are designed to have the same
mechanical length, their electrical delays are different. This reflects the
differences in the connector interface compression. When the male and female
devices are torqued together the male side compresses more than the female
side.
The 1.0 mm connector utilizes an air dielectric interface for the highest
accuracy and repeatability. The coupling diameter and thread size were chosen
to maximize strength, increase durability and provide highly repeatable
connections. The connectors are designed so that the outer conductors
engage before the center conductors.
Ω, 1.0 mm connectors with a frequency range of DC to
To obtain the best performance possible, the manufacturing tolerances of the
connectors are tighter than the standard 1.0 mm specifications per the IEEE
287 precision connector standard.
Kit Contents
Use the Contents List in the shipping container to verify the completeness of
your shipment. Although this list is the most accurate, you can also use the
illustration in Table 6-1 on page 6-2 to verify the items in your shipment. If
your shipment is not complete, contact Keysight Technologies - refer to
“Contacting Keysight” on page 5-4.
Compatible Network Analyzers
The devices in this kit and their data are compatible with the PNA series
network analyzers.
Data-Based Model for Defining the Calibration Standards
The 85059 family of calibration kits uses the Data-Based models to define the
calibration standards.
Keysight 85059B Operation and Service Guide 1-3
1-
General Information
Calibration Kit Overview
Data-Based Model
The data-based model (85059B) of each calibration kit provides higher
accuracy for describing calibration standards than a polynomial model. The
data-based nominal S-Parameter model for each calibration standard is
generated by using a theoretical derivation or a high accuracy S-Parameter
measurement. These nominal S-parameter models are stored as part of the
data-based model. Errors that can be the result of polynomial model
approximations do not exist.
The data-based models may only be used with the SmartCal (Guided)
Calibration method of the Cal Wizard. Using the data-based models with the
Unguided Calibration method of the Cal Wizard is not allowed.
The data-based models may NOT be edited with the Advanced Modify Cal Kit
dialog box. Attempts to programmatically edit the definitions of the
data-based-model calibration standards will result in reported errors (i.e. SCPI
Execution Error).
In the PNA’s Modify Calibration Class Assignments dialog box
Math” feature is turned ON by default when the 85059B data-based model is
selected from a cal kit menu. In order to achieve a more accurate calibration,
Expanded Math uses the data at any given frequency from ALL measured
standards instead of using only the three standards specified in the class
definitions.
85059B (Precision) Kit Contents
The 85059B (Precision) calibration kit contains the following:
— one pair (male and female) of an offset open
— one pair (male and female) of a lowband load
— four pairs (male and female) of offset shorts
—three adapters
— form—tracking recalibration date due
— calibration certificate
— wrench, open end for the 6 mm flats on some of the components
— wrench–torque, 4-in-lb. 6 mm open end, gold handle
1
, the “Expanded
— wrench–torque, 4-in-lb. 14 mm open end, gold handle
— a 10x magnifying glass
— case—plastic storage
1. Pathway to the Modify Calibration Class Assignments dialog box: Calibration >
Advanced Modify Cal Kit > Edit Kit > Class Assignments: Edit.
1-4 Keysight 85059B Operation and Service Guide
General Information
Calibration Kit Overview
Refer to Table 6-1 on page 6-2 and Figure 6-1 on page 6-3 for a complete list
of kit contents and their associated part numbers.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the
current state–of–the–art in precision machining. The offset short's inner
conductors have a one–piece construction, common with the shorting plane.
This construction provides for extremely repeatable connections. The offset
opens have inner conductors that are supported by a strong, low dielectric
plastic to provide repeatability and reliability. Both the opens and shorts are
constructed so that the pin depth can be controlled very tightly, thereby
minimizing phase errors. The length of the offset opens are designed so that
the difference in phase of their reflection coefficients is approximately 180° at
all frequencies, with respect to offset short 2.
How to Identify Devices
The shorts, opens, and loads in the kit are identified by the number of rings (or
bands) around the body of the device.
— lowband loads have no rings or bands
— 1.3 mm offset short has 1 ring or band and is identified as Short 1
— 2.45 mm offset short has 2 rings or bands and is identified as Short 2
— 3.326 mm offset short has 3 rings or bands and is identified as Short 3
— 4.039 mm offset short has 4 rings or bands and is identified as Short 4
— offset open has 3 rings and is designed to be 180 degrees out of phase with
Short 2. In addition, the offset open has a rounded edge to help distinguish
it from the shorts (see Figure 1-1).
Figure 1-1 OPEN Calibration Devices: 85059B
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances
to provide good broadband performance and to ensure stable, repeatable
connections.
Keysight 85059B Operation and Service Guide 1-5
1-
General Information
Calibration Kit Overview
Lowband Loads
The loads have been optimized for broadband performance up to 50 GHz.
The best operating region of the load is from DC to 50 GHz. Performance
degrades quickly above 50 GHz.
Performing a Calibration
Using a Network Analyzer
To find information about performing a calibration with your network analyzer,
refer to your analyzer’s User’s Guide or Help system. To find an online copy on
the Keysight web site:
1. Go to www.keysight.com
2. Enter the analyzer’s model in the search box.
3. Press Enter
4. On the web page that opens, underneath the title of your Network Analyzer, click Support
5. Click the Documentation Library icon
6. Click the link for the document you want to view
Calibration Definitions
The calibration kit must be selected prior to performing a calibration. In
addition, the calibration definitions for the devices in the kit must be installed
in the analyzer as part of a firmware upgrade (≥12.80). Refer to
http://na.support.keysight.com/pna/.
Class assignments and standard definitions may change as more accurate
model and calibration methods are developed. For reference, you can
download and view the most recent class assignments and standard definitions
from Keysight’s Calibration Kit Definitions Web page at
Refer to online PNA Help for instructions on selecting the calibration kit,
installing (if necessary) the calibration definitions, and performing a
calibration.
Equipment Required but Not Supplied
Various connector cleaning supplies and electrostatic discharge safety
supplies are not provided in this kit. (Refer to Chapter 6, “Replacement Parts”,
on page 6- 1 for ordering information.)
1-6 Keysight 85059B Operation and Service Guide
General Information
Offset-Short Calibration Using the 85059B (Precision) Calibration Kit and the PNA’s
Guided Power Calibration
Gage sets (not provided with the 85059B), adapters, ESD protection devices,
and va
but are required to ensure successful operation of the calibration kit. Refer to
Table 6-1 on page 6-2 for ordering information.
rious connector cleaning supplies are not included in the calibration kit
Offset-Short Calibration Using the 85059B (Precision) Calibration
Kit and the PNA’s Guided Power Calibration
IMPORTANT! The 85059B calibration kit must be used with the PNA’s
Guided Calibration. PNA’s guided calibration selects the proper standards
for connection based on the selected frequency range. Refer to “Using a
Network Analyzer” on page 1-6.
To download a copy of the 85059B calibration kit definitions (.xkt file), refer
to Keysight’s Calibration Kit Definitions Web page at
For best measurement results, the isolation standard should be the
equivalent impedance of the DUT.
Keysight 85059B Operation and Service Guide 1-7
1-
General Information
Calibration Residual Error Specifications
Calibration Residual Error Specifications
Please download our free Vector Network Analyzer Uncertainty Calculator from
http://www.keysight.com/find/na_calculator.
Table 1-1 85059B (Precision) Residual System Calibration Error
Specifications
Frequency
Range (GHz)
Data-Based Model (Residuals)
DirectivitySource
Match
DC to 10–29–280.02
10 to 15–31–310.09
15 to 20–35–350.09
20 to 30 –37–360.05
30 to 40 –37–350.05
40 to 50 –33–320.07
50 to 60 –31–300.08
60 to 70 –31–300.15
70 to 80 –28–270.15
80 to 100–28–270.11
100 to 115 –27–270.13
115 to 120 –27–270.13
1
Reflection
Tracking
1. Refer to “Data-Based Model for Defining the
Calibration Standards” on page 1-3.
1-8 Keysight 85059B Operation and Service Guide
General Information
Serial Numbers
Serial Numbers
Serial Number Prefix
A serial number label is attached to the calibration kit. A typical kit serial
number label is shown in Figure 1-2. The prefix is made up of six characters.
The first two characters show the country, the next two digits represent the
year, and the last two digits designate the week of manufacturing.
Serial Number Suffix
The last four digits are the suffix numbers. The suffix numbers are unique to
each calibration kit.
Figure 1-2 Typical Kit Serial Number Label
Device Serial Numbers
In addition to the kit serial number, the devices in this kit are individually
serialized (serial numbers are printed on an attached label, or scribed onto the
body of each device). Record these serial numbers in Figure 1-6 on page 1-16.
This can help you avoid confusing the devices in this kit with similar devices
from other kits.
Keysight 85059B Operation and Service Guide 1-9
1-
General Information
Incoming Inspection
Recording the Device Serial Numbers
Table 1-2 Serial Number Record for 85059B (Precision)
Verify that the shipment is complete by referring to Table on page 6-2 or
Figure 6-1 on page 6-3.
Check for damage. The foam-lined storage case provides protection during
shipping.
If the case or any device appears damaged, or if the shipment is incomplete,
contact Keysight. See “Contacting Keysight” on page 5-4. Keysight will
arrange for repair or replacement of incomplete or damaged shipments
without waiting for a settlement from the transportation company.
When you send the kit or device to Keysight, include the following information:
—your company name and address
— the name of a technical contact person within your company, and the
person's complete phone number
— the model number and serial number of the kit
1-10 Keysight 85059B Operation and Service Guide
General Information
Incoming Inspection
— the part number and serial number of the device
— the type of service required
—a detailed description of the problem
Keysight 85059B Operation and Service Guide 1-11
1-
General Information
Clarifying Connector Gender
Clarifying Connector Gender
In this manual, connectors are referred to in terms of their device gender
unless otherwise stated. For example, a male open has a male connector.
Figure 1-3 Clarifying Connector Gender
Figure 1-4 Male and Female Connectors
1-12 Keysight 85059B Operation and Service Guide
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of
cable connectors and adapters is referred to in terms of the center conductor.
For example, a connector or device designated as 1.0 mm –f– has a 1.0 mm
female center conductor.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit
include:
— routine visual inspection
—cleaning
— proper pin depth
— proper connection techniques
All of the above are described in Chapter 4, “Maintenance and Support.”.
Failure to detect and remove dirt or metallic particles on a mating plane
surface can degrade repeatability and accuracy and can damage any
connector mated to it. Improper connections, resulting from pin depth values
being out of the typical limits (see Table 2-4), or from bad connections, can
also damage these devices.
Keysight 85059B Operation and Service Guide 1-13
1-
General Information
When to Calibrate
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the
systematic error are insignificant. This means that changes to the uncorrected
leakages (directivity and isolation), mismatches (source match and load
match), and frequency response of the system are small (<10%) relative to
accuracy specifications.
Change in the environment (especially temperature) between calibration and
measurement is the major cause in calibration accuracy degradation. The
major effect is a change in the physical length of external and internal cables.
Other important causes are dirty and damaged test port connectors and
calibration standards. If the connectors become dirty or damaged,
measurement repeatability and accuracy is affected. Fortunately, it is relatively
easy to evaluate the general validity of the calibration. To test repeatability,
remeasure one of the calibration standards. If you can not obtain repeatable
measurements from your calibration standards, maintenance needs to be
performed on the test port connectors, cables and calibration standards. Also,
maintain at least one sample of the device under test or some known device as
your reference device. A verification kit may be used for this purpose. After
calibration, measure the reference device and note its responses. Periodically
remeasure the device and note any changes in its corrected response which
can be attributed to the test system. With experience you will be able to see
changes in the reference responses that indicate a need to perform the
measurement calibration again.
1-14 Keysight 85059B Operation and Service Guide
General Information
Regulatory and Environmental Information
Regulatory and Environmental Information
This section contains information that is required by various government
regulatory agencies.
The CE mark is a registered trademark of the European Community. (If accompanied by a year, it is
when the design was proven.)
ccr.keysight@key
sight.com
The Keysight email address is required by EU directives applicable to our product.
China RoHS regulations include requirements related to packaging, and require compliance to China
standard GB18455-2001.
This symbol indicates compliance with China RoHS regulations for paper/fiberboard packaging.
Keysight 85059B Operation and Service Guide 1-15
1-
General Information
Regulatory and Environmental Information
1-16 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
2Specifications
Information in This Chapter
Chapter Two at-a-Glance
Section TitleSummary of Content
“Environmental Requirements” on
page 2
“Mechanical Characteristics” on
page 3
“Electrical Specifications” on page 4A discussion of the electrical specifications of the calibration devices.
Describes the operating and storage environmental conditions for your
calibration kit.
A discussion of the mechanical characteristics of the calibration devices
including: temperature.
1
2-
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Operating temperature
Error-corrected temperature range
Storage temperature−40 °C to +75 °C (−40 °F to +167 °F)
a. The temperature range over which the calibration standards maintain conformance to their specifica-
tions.
b. The allowable network analyzer ambient temperature drift during measurement calibration and during
measurements when the network analyzer error correction is turned on. Also, the range over which the
network analyzer maintains its specified performance while correction is turned on.
a
b
+20 °C to +26 °C (+68 °F to +79 °F)
±1 °C of measurement calibration temperature
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the
operating temperature is a critical factor in performance. During a
measurement calibration, the temperature of the calibration devices must be
stable and within the range specified in Table 2-1.
IMPORTANT! Avoid unnecessary handling of the devices during calibration
because your fingers are a heat source.
Performance verification and measurements of devices under test need not be
performed within the operating temperature range of the calibration devices,
but they must be within the error-corrected temperature of the network
analyzer (±1 °C of the measurement calibration temperature). For example, if
the calibration is performed at +20 °C, the error-corrected temperature range
is +19 °C to +21 °C. It is then appropriate to perform measurements and
performance verifications at +19 °C, which is outside the operating
temperature range of the calibration devices, since only the actual calibration
must be performed within the operating temperature range.
2-2 Keysight 85059B Operation and Service Guide
Specifications
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth
are not performance specifications. They are, however, important
supplemental characteristics related to electrical performance. Keysight
Technologies verifies the mechanical characteristics of the devices in this kit
with special gaging processes and electrical testing. This ensures that the
device connectors do not exhibit any improper pin depth when the kit leaves
the factory.
Keysight 85059B Operation and Service Guide 2-3
2-
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications in Table 2-2 apply to the devices in your
calibration kit when connected with a Keysight precision interface.
Table 2-2 85059B (Precision) Electrical Specifications for 1.0 mm
Keysight Technologies certifies that this product met its published
specifications at the time of shipment from the factory. Keysight further
certifies that its calibration measurements are traceable to the United States
National Institute of Standards and Technology (NIST) to the extent allowed by
the institute’s calibration facility, and to the calibration facilities of other
International Standards Organization members. See “How Keysight Verifies
the Devices in This Kit” on page 4-2 for more information.
2-6 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
3Use, Maintenance, and Care of the Devices
Information in this Chapter
Chapter Three at-a-Glance
Section TitleSummary of Content
“Electrostatic Discharge” on page 2How to protect your instruments and devices against electrostatic
discharge.
“Visual Inspection” on page 4A discussion of visually inspecting your verification kit’s contents.
“Cleaning Connectors” on page 8How to clean your connectors.
“Pin Depth” on page 10How to verify the pin depth of you male and female devices.
“Making Connections” on page 12Discussion of making connections. Includes a discussion of proper
torquing procedures and properly separate connections.
“Using a Torque Wrench” on
page 15
“Disconnection Procedure” on
page 19
“Handling and Storage” on page 19What the system verification does.
How to use a torque wrench for consistent measurements and
avoiding damaging a device.
Discussion of how to disconnect your devices and how to avoid
damaging your devices while torquing.
How to perform the verification test.
How to interpret the results.
1
3-
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting,
inspecting, or cleaning connectors attached to a static-sensitive circuit (such
as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive
internal circuit elements when discharged. Static discharges too small to be
felt can cause permanent damage. Devices such as calibration components
and devices under test (DUTs), can also carry an electrostatic charge. To
prevent damage to the test set, components, and devices:
— always wear a grounded wrist strap having a 1 MΩ resistor in series with it
when handling components and devices or when making connections to
the test set.
— always use a grounded, conductive table mat while making connections.
— always wear a heel strap when working in an area with a conductive floor. If
you are uncertain about the conductivity of your floor, wear a heel strap.
— always ground yourself before you clean, inspect, or make a connection to
a static-sensitive device or test port. You can, for example, grasp the
grounded outer shell of the test port or cable connector briefly.
— always ground the center conductor of a test cable before making a
connection to the analyzer test port or other static-sensitive device. This
can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to
short the center conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the
cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 6, “Replacement Parts.” for part numbers and instructions for
ordering ESD protection devices.
3-2 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Figure 3-1 ESD Protection Setup
Keysight 85059B Operation and Service Guide 3-3
3-
Use, Maintenance, and Care of the Devices
Visual Inspection
Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a
connection is made. Metal particles from the connector threads may fall into
the connector when it is disconnected. One connection made with a dirty or
damaged connector can damage both connectors beyond repair.
Devices with damaged connectors should be immediately discarded or
clearly marked and set aside for repair. A damaged device will in turn
damage any good connector to which it is attached. Determine the cause
of the damage before connecting a new, undamaged connector in the
same configuration.
Magnification is helpful when inspecting connectors, but it is not required and
may actually be misleading. Defects and damage that cannot be seen without
magnification generally have no effect on electrical or mechanical
performance. Magnification is of great use in analyzing the nature and cause of
damage and in cleaning connectors, but it is not required for inspection.
Look for Obvious Defects and Damage First
Examine the connectors first for obvious defects and damage:
—Plating
—Bare metal showing
— Burrs or blisters
—Deformed threads
— Center conductors
—Bent
—Broken
—Misaligned
— Concentricity
Connector nuts should move smoothly and be free of:
—Burrs
—Loose metal particles
— Rough spots
Any connector that has obvious defects should be discarded or sent for repair refer to “Contacting Keysight” on page 5-4.
3-4 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Visual Inspection
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The
more use a connector gets, the faster it wears and degrades. The wear is
greatly accelerated when connectors are not kept clean, or are connected
incorrectly.
Connector wear eventually degrades performance of the device. Calibration
devices should have a long life if their use is on the order of a few times per
week. Replace devices with worn connectors.
The test port connectors on the network analyzer test set may have many
connections each day, and are therefore more subject to wear. It is
recommended that an adapter be used as a test port saver to minimize the
wear on the test set’s test port connectors.
Connector Contacts
See Figure 3-2 on page 3-5 for visual guidelines when evaluating the contact
integrity of a connector.
Notice the location of the cross hairs in relationship to the center of the
figures.
Figure 3-2 Contact Integrity
Concentricity
Figure 3-3 and Figure 3-4 show the concentricity of both the male and female
1.0 mm connectors. Inspect the connectors with a minimum magnification of
10X.
Keysight 85059B Operation and Service Guide 3-5
3-
Use, Maintenance, and Care of the Devices
Visual Inspection
Figure 3-3 Concentricity of a Female Connector
Figure 3-4 Concentricity of a Male Connector
Inspect the Mating Plane Surfaces
Flat contact between the connectors at all points on their mating plane
surfaces is required for a good connection. See Figure 3-6 on page 3-10.
Look especially for deep scratches or dents, and for dirt and metal particles on
the connector mating plane surfaces. Also, look for “dings” on the mating
plane surfaces of the center and outer conductors, and look for signs of
damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light
scratches or shallow circular marks distributed more or less uniformly over the
mating plane surface. Other small defects and cosmetic imperfections are also
normal. None of these affect electrical or mechanical performance.
Clean and inspect the connector again, if it shows:
— Deep scratches or dents
— Particles clinging to the mating plane surfaces
— Uneven wear
3-6 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Visual Inspection
Devices with damaged connectors should be discarded or sent for repair.
Determine the cause of damage before connecting a new, undamaged
connector in the same configuration. Magnification is of great use in analyzing
the nature and cause of damaged connectors.
Inspect Female Connectors
Pay special attention to the contact fingers in the female center conductor.
These can be bent or broken, and damage to them is not always easy to see. A
connector with damaged contact fingers will negatively affect electrical
performance and must be replaced.
Inspection is particularly important when mating nonprecision to precision
devices.
Supplies and Equipment Needed
The supplies and equipment needed to perform the cleaning procedure, and
their Keysight Technologies part numbers are listed in Table 7-1 on page 7-2
and page 7-3.
Keysight 85059B Operation and Service Guide 3-7
3-
Use, Maintenance, and Care of the Devices
Cleaning Connectors
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave
coaxial connections.
1. Use Compressed Air or Nitrogen
Clean connectors are essential for ensuring the integrity of RF and
microwave coaxial connections.
Always use protective eyewear when using compressed air or
nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector
mating plane surfaces.
You can use any source of clean, dry, low-pressure compressed air or
nitrogen that has an effective oil-vapor filter and liquid condensation trap
placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air
pressure to less than 414 kPa (60 psi) to control the velocity of the air
stream. High-velocity streams of compressed air can cause electrostatic
effects when directed into a connector. These electrostatic effects can
damage the device. Refer to “Electrostatic Discharge” on page 3-2 earlier
in this chapter for additional information.
2. Clean the Connector Threads
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a
tightly closed container. It is extremely flammable. In case of fire,
use alcohol foam, dry chemical, or carbon dioxide; water may be
ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact
with eyes, skin, and clothing. It causes skin irritation, may cause eye
damage, and is harmful if swallowed or inhaled. It may be harmful if
absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with
water.
Dispose of isopropyl alcohol in accordance with all applicable
federal, state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to
remove any dirt or stubborn contaminants on a connector that cannot be
removed with compressed air or nitrogen. Refer to Table 6-2 on page 6-4
for a part numbers for cleaning swabs.
3-8 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Cleaning Connectors
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning
swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always
completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning
swab.
b. Gently clean the center and outer conductor mating plane surfaces.
Refer to Figure 3-6 on page 3-10. When cleaning a female
connector, avoid snagging the swab on the center conductor
contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always
completely dry a connector before you reassemble or use it.
4. Reinspect
Inspect the connector again to make sure that no particles or residue are
present.
Figure 3-5 Cleaning Illustration
Keysight 85059B Operation and Service Guide 3-9
3-
Use, Maintenance, and Care of the Devices
Pin Depth
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being
flush with the outer conductor mating plane. See Figure 3-6 The pin depth of a
connector can be in one of two states: either protruding or recessed.
— Protrusion is the condition in which the center conductor extends beyond
the outer conductor mating plane. This condition will indicate a positive
value on the connector gage.
At no time should the pin depth of the 1.0 mm connector be protruding.
— Recession is the condition in which the center conductor is set back from
the outer conductor mating plane. This condition will indicate a negative
value on the connector gage.
Figure 3-6 Connector Pin Depth
The pin depth value of each calibration device in the kit is not specified, but is
an important mechanical parameter. The electrical performance of the device
depends, to some extent, on its pin depth.
Keysight verifies the pin depth characteristics of the connectors in this kit with
special gaging processes and electrical testing. This ensures that the device
connectors do not exhibit any center conductor protrusion and have proper pin
depth when the kit leaves the factory.
3-10 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Pin Depth
The electrical specifications for each device in the kit take into account the
effect of pin depth on the device’s performance. Table 3-1 on page 3-11 lists
the typical pin depths for the devices in the kit. If the pin depth of a device does
not measure within the typical pin depth limits, it may be an indication that the
device fails to meet electrical specifications. Refer to Figure 3-6 on page 3-10
for a visual representation of proper pin depth (slightly recessed).
Table 3-1 Pin Depth Limit
DeviceTypical Pin Depth
Opens
Shorts
Fixed Loads
Adapters
−0.003 to −0.011 mm
−0.00011 to −0.00043 in
−0.003 to −0.007 mm
−0.00011 to −0.00027 in
−0.0020 to −0.020 mm
−0.000079 to −0.00078 in
−0.004 to −0.024 mm
−0.00016 to −0.00094 in
Keysight 85059B Operation and Service Guide 3-11
3-
Use, Maintenance, and Care of the Devices
Making Connections
Making Connections
Good connections require a skilled operator. Instrument sensitivity and coaxial
connector mechanical tolerances are such that slight errors in operator
technique can have a significant effect on measurements and measurement
uncertainties.
The most common cause of measurement error is poor connections.
How to Make a Connection
Connection Procedure
1. Ground yourself and all devices. Wear a grounded wrist strap and work on
grounded, conductive table mat. Refer to “Electrostatic Discharge”
a
ge 3-2 for ESD precautions.
pa
on
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-4.
3. If necessary, clean the connectors. Refer to “Cleaning Connecto
ge 3-8.
pa
4. Carefully align the connectors. The male connector center pin must
c
oncentrically into the contact finger of the female connector.
5. Push the connectors straight together. Do not twist or screw
t
ogether. As the center conductors mate, there is usually a slig
r
esistance.
rs” on
slip
them
ht
Do not twist one connector into the other (like inserting a light bulb). This
happens when you turn the device body, rather than the connector nut.
Majordamage to the center conductor and the outer conductor can occur
if the device body is twisted.
6. Initial tightening can be done by hand, or with a 6 mm open-end wrench.
Tighten until “snug” or where the connectors are first making contact. The
eliminary connection is tight enough when the mating plane surfac
pr
e uniform, light contact. Do not overtighten this connection.
mak
es
At this point, all you want is for the outer conductors to make gentle
contact on both mating surfaces. Use very light finger pressure (no more than 2
inch–pounds of torque).
7. Relieve any side pressure on the connection from long or heavy devices,
cables. This assures consistent torque (refer to “Using a Torque Wrench”
on page 3-15).
3-12 Keysight 85059B Operation and Service Guide
or
Use, Maintenance, and Care of the Devices
Making Connections
Figure 3-7 Alignment
Connector Misalignment
Forced misalignment could damage the female center conductor.
Figure 3-8 Misalignment
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces,
always support the devices and connections.
Turn the connector nut, not the device body. Major damage to the center
conductor can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
Keysight 85059B Operation and Service Guide 3-13
3-
Use, Maintenance, and Care of the Devices
Making Connections
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending
either of the connectors.
3-14 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Using a Torque Wrench
Using a Torque Wrench
Use a torque wrench to make a final connection. Table 3-2 provides
information about the torque wrench recommended for use with this
calibration kit. Refer to Table 6-1 on page 6-2 for part number and ordering
information.
Table 3-2 Torque Wrench Information
Connector TypeTorque SettingTorque Tolerance
1.0 mm45 N-cm (4 in-lb)±5.4 N-cm (±0.5 in-lb)
Using a torque wrench guarantees that the connection is not too tight,
preventing possible connector damage. It also guarantees that all connections
are equally tight each time. Table 3-9 on page 3-16 shows you where to hold
the torque wrench for optimum performance.
Donot pre-tighten the connector nut so much that there is no rotation of
the nut with the torque wrench. Static friction must not be present during
torquing.
1. Use the torque wrench supplied with your kit to make the final
connections.
2. Rotate only the connector nut when you tighten the connector.
In all situations, use an open–end wrench to keep the body of the device from
turning. Position both wrenches within 90 degrees of each other before
applying force (see Table 3-10 on page 3-16). Wrenches opposing each other
(180 degrees apart) will cause a lifting action. This lifting action can misalign,
and stress the connections of the devices involved. This is especially true when
several devices are connected together.
Keysight 85059B Operation and Service Guide 3-15
3-
Use, Maintenance, and Care of the Devices
Using a Torque Wrench
Figure 3-9 Correct Wrench Position
Narrow separation of the wrenches produces a small residual lateral force on
the structure of connected devices.
Figure 3-10 Incorrect Torque Wrench
Wide separation of the wrenches produces a larger residual lateral force on the
structure of connected devices. This can degrade connector repeatability.
3. Hold the torque wrench lightly, at the end of the handle only (beyond the
groove). See Figure 3-11 on page 3-17.
3-16 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Using a Torque Wrench
Figure 3-11 Where to Hold the Torque Wrench
4. Apply downward force perpendicular to the wrench handle. See Figure
3-12. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down
along its length rather than pivoting it, otherwise you apply an unknown
amount of torque.
Figure 3-12 Using the Torque Wrench
5. Tighten the connection just to the torque wrench "break" point. The
wrench handle gives way at its internal pivot point. See Figure 3-12. Do not tighten the connection further.
Keysight 85059B Operation and Service Guide 3-17
3-
Use, Maintenance, and Care of the Devices
Using a Torque Wrench
You don’t have to fully break the handle of the torque wrench to reach the
specified torque; doing so can cause the handle to kick back and loosen
the connection. Any give at all in the handle is sufficient torque.
Do not pivot the wrench handle on your thumb or other fingers, otherwise
you apply an unknown amount of torque to the connection when the
wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating
plane. If you do, you apply more than the recommended torque.
3-18 Keysight 85059B Operation and Service Guide
Use, Maintenance, and Care of the Devices
Disconnection Procedure
Disconnection Procedure
To avoid lateral (bending) force on the connector mating plane surfaces,
always support the devices and connections.
1. Use an open–end wrench to prevent the device body from turning.
2. Use another wrench to loosen the connector nut.
3. Complete the disconnection by hand, turning only the connector nut.
Do not twist one connector out of the other, (like removing a light bulb).
Turn the connector nut, not the device body. Major damage to the center
conductor and the outer conductor can occur if the device body is twisted.
4. Pull the connectors straight apart without twisting or bending.
Handling and Storage
— Install the protective end caps and store the calibration devices in the
foam-lined storage case when not in use.
— Never store connectors loose in a box, desk, or bench drawer. This is the
most common cause of connector damage during storage.
— Keep connectors clean.
— Do not touch mating plane surfaces. Natural skin oils and microscopic
particles of dirt are easily transferred to a connector interface and are very
difficult to remove.
— Do not set connectors contact-end down on a hard surface. The plating and
the mating plane surfaces can be damaged if the interface comes in
contact with any hard surface.
Keysight 85059B Operation and Service Guide 3-19
3-
Use, Maintenance, and Care of the Devices
Handling and Storage
3-20 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
4Performance Verification
Information in This Chapter
Chapter Four at-a-Glance
Section TitleSummary of Content
“Introduction” on page 1Overview of how to recertify the 85059B kit.
“How Keysight Verifies the Devices in
This Kit” on page 2
“Recertification” on page 2Discusses what is provided with the recertification and
“Where to Send a Kit for
Recertification” on page 3
Introduction
A discussion on the steps Keysight uses to verify a
calibration kit.
how often to recertify the calibration kit.
Where to send your kit for certification.
The performance of your calibration kit can only be verified by returning the kit
to Keysight Technologies for recertification. The equipment required to verify
the specifications of the devices in the kit has been specially manufactured and
is not commercially available.
1
4-
Performance Verification
How Keysight Verifies the Devices in This Kit
How Keysight Verifies the Devices in This Kit
Keysight verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with
precision airlines and shorts that are directly traced to NIST (National
Institute of Standards and Technology). The airline and short
characteristics are developed from mechanical measurements. The
mechanical measurements and material properties are carefully modeled
to give very accurate electrical representation. The mechanical
measurements are then traced to NIST through various plug and ring
gages and other mechanical measurements
2. Each calibration device is electrically tested on this system. For the initial
(before sale) testing of the calibration devices, Keysight includes the test
measurement uncertainty as a guardband to guarantee each device meets
the published specification. For recertifications (after sale), no guardband
is used and the measured data is compared directly with the specification
to determine the pass or fail status. The measurement uncertainty for each
device is, however, recorded in the calibration report that accompanies
recertified kits.
Recertification
These two steps establish a traceable link to NIST for Keysight to the extent
allowed by the institute’s calibration facility. The specifications data provided
for the devices in this kit is traceable to NIST through Keysight Technologies.
The following will be provided with a recertified kit:
— a new calibration sticker affixed to the case
— a certificate of calibration
— a calibration report for each device in the kit listing measured values,
specifications, and uncertainties
A list of NIST traceable numbers may be purchased upon request to be
included in the calibration report.
Keysight Technologies offers a Standard calibration for the recertification of
this kit. For more information, contact Keysight Technologies - refer to
“Contacting Keysight” on page 5-4.
4-2 Keysight 85059B Operation and Service Guide
Performance Verification
Where to Send a Kit for Recertification
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The
actual need for recertification depends on the use of the kit. After reviewing the
results of the initial recertification, you may establish a different recertification
interval that reflects the usage and wear of the kit.
In some cases, the first time a kit is used after being recertified occurs
some time after the actual recertification date. The recertification interval
should begin on the date the kit is first used after the recertification date.
Where to Send a Kit for Recertification
Contact Keysight Technologies for information on where to send your kit for
recertification. See “Contacting Keysight” on page 5-4.
Keysight 85059B Operation and Service Guide 4-3
4-
Performance Verification
Where to Send a Kit for Recertification
4-4 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
5Troubleshooting
Information in This Chapter
Chapter Five at-a-Glance
Section TitleSummary of Content
“Troubleshooting Process” on
page 5-2
“Where to Look for More Information”
on page 5-3
“Returning a Kit or Device to Keysight”
on page 5-3
“Contacting Keysight” on page 5-4How to contact Keysight.
How to troubleshoot a calibration problem. Includes a flow
troubleshooting diagram.
Where to find more information on network analyzer system
operation.
Where to return a kit to Keysight and the required
information.
1
5-
Troubleshooting
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass
performance verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5-2 Keysight 85059B Operation and Service Guide
Troubleshooting
Where to Look for More Information
Where to Look for More Information
This manual contains limited information about network analyzer system
operation. For detailed information on using an ENA or PNA series network
analyzer, refer to the appropriate user guide or Help file.
— To view the ENA or PNA Help, press the Help key on the front panel of the
network analyzer.
— To view Help or a user guide online, use the following steps:
1. Go to www.keysight.com/find/assist.
2. Enter your analyzer model number (Ex: N5242B) in the search box
and press return.
3. Under the heading of your network analyzer, click on the Support tab
and then the Document Library icon.
4. Click on the document PDF you want to view.
If you need additional information, see “Contacting Keysight” on page 5-4.
Returning a Kit or Device to Keysight
If your kit or device requires service, contact Keysight Technologies for
information on where to send it and proved the following:
—your company name and address
— a technical contact person within your company, and the person's complete
phone number
— the model number and serial number of the kit
— the part number and serial number of each device
— the type of service required
—a detailed description of the problem and how the device was being used
when the problem occurred (such as calibration or measurement)
— see also, “Contacting Keysight” on page 5-4
Keysight 85059B Operation and Service Guide 5-3
5-
Troubleshooting
Contacting Keysight
Contacting Keysight
Assistance with test and measurements needs and information on finding a
local Keysight office are available on the Web at:
www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field
engineer.
In any correspondence or telephone conversation, refer to the Keysight
product by its model number and full serial number. With this information,
the Keysight representative can determine whether your product is still
within its warranty period.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from
the Keysight web site:
—Go to www.keysight.com.
— Enter the document’s part number (located on the title page) in the Search
box.
— Click Search
— Open the PDF and print the document.
5-4 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
6Replacement Parts
Information in This Chapter
Chapter Six at-a-Glance
Section TitleSummary of Content
“Introduction” on page 1A discussion of the chapter overview:
— Table of Replaceable Parts (See Table 6-1 on page 6-2)
— Table of orderable parts that are not included in this kit (See
Table 6-2 on page 6-4)
Introduction
The following replaceable parts table lists the replacement part numbers for
the 85059B calibration kit contents. To order a listed part, note the description,
part number, and the quantity desired and contact Keysight Technologies refer to “Contacting Keysight” on page 5-4.
1
6-
Replacement Parts
Introduction
Table 6-1 Replaceable Parts for the 85059B (Precision) Calibration Kit
Description Qty Per KitKeysight Part Number
Calibration Devices (1.0 mm)
a,b
1.3 mm offset short 1 –m–185059-60027
2.45 mm offset short 2 –m–185059-60029
3.326 mm offset short 3 –m–185059-60031
4.039 mm offset short 4 –m–185059-60033
1.3 mm offset short 1 –f–185059-60028
2.45 mm offset short 2 –f–185059-60030
3.326 mm offset short 3 –f–185059-60032
4.039 mm offset short 4 –f–185059-60034
Offset open –m–185059-60053
Offset open –f–185059-60054
Lowband load –m–185059-60019
Lowband load –f–185059-60020
Adapters (1.0 mm)
a
Adapter m–m185059-60044
Adapter f–f185059-60045
Adapter m–f185059-60046
Calibration Kit Storage Case
Box with customized foam inserts11540-2330
Wrenches
6 mm open–end wrench18710-2156
6 mm, 45 N–cm (4 in–lb) torque wrench18710-2812
14 mm, 45 N–cm (4 in–lb) torque wrench18710-2813
Miscellaneous Items
Operation and Service Guide1
85059-90005
10x Magnifying Glass11000-1114
Protective end capsas required1401-0202
a. See “Clarifying the Terminology of a Connector Interface” on page 1-13.
b. Refer to “How to Identify Devices” on page 1-5.
c
6-2 Keysight 85059B Operation and Service Guide
Replacement Parts
6 mm6 mm
14 mm
Introduction
c. Refer to “Printing Copies of Documentation from the Web” on page 5-4.
Figure 6-1 Replaceable Parts for the 85059B (Precision) Calibration Kit (not all items are
shown below. For a complete list, refer to See “Replaceable Parts for the
85059B (Precision) Calibration Kit” on page 2.)
Keysight 85059B Operation and Service Guide 6-3
6-
Replacement Parts
Introduction
Table 6-2 Items Not Included in the Calibration Kits
Description QtyKeysight Part Number
a
Adapters
V-band waveguide to 1.0 mm –f– coax 1V281C
V-band waveguide to 1.0 mm –m– coax 1V281D
W-band waveguide to 1.0 mm –f– coax 1W281C
W-band waveguide to 1.0 mm –m– coax 1W281D
1 mm -m- to 1.85 mm -m-111921E
1 mm -f- to 1.85 mm -f-111921F
1 mm -m- to 1.85 mm -f-111921G
(50Ω)
1 mm -f- to 1.85 mm -m-111921H
1.0 mm
1.0 mm
b
–f– to 1.0 mmb –f– (Ruggedized)
b
–f– to 1.0 mmb –m– (Ruggedized)
1 Y1900B
1 Y1900C
ESD Protective Devices
Grounding wrist strap19300-1367
5 ft grounding cord for wrist strap19300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire19300-0797
ESD heel strap19300-1308
Connector Cleaning Supplies
Anhydrous isopropyl alcohol (>92% pure)
c
----
Foam tipped cleaning swabs1009301-1243
a. See “Clarifying the Terminology of a Connector Interface” on page 1-13.
b. Ruggedized connector.
c. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally.
6-4 Keysight 85059B Operation and Service Guide
Keysight 1.0 mm Precision Calibration Kit
85059B
Operation and Service Guide
7Standard Definitions
Information in This Chapter
Chapter Seven at-a-Glance
Section TitleSummary of Content
Class Assignments and Standard
Definitions Values are Available on
the Web
Discussion on standard definitions and where to find more information on the
web.
Class Assignments and Standard Definitions Values are Available on
the Web
Class assignments and standard definitions may change as more accurate
model and calibration methods are developed. You can download the most
recent class assignments and standard definitions from Keysight’s Calibration
Kit Definitions Web page at
For a detailed discussion of calibration kits, refer to the Keysight Application
Note, “Specifying Calibration Standards and Kits for Keysight Vector Network
Analyzers.” This application note covers calibration standard definitions,
calibration kit content and its structure requirements for Keysight vector
network analyzers. It also provides some examples of how to set up a new
calibration kit and how to modify an existing calibration kit definition file. To
download a free copy, go to www.keysight.com and enter literature number
5989-4840EN in the Search window.
1
7-
Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web