Keysight Technologies 85056D, 85563A, 85564A User's And Service Manual

User's and Service Guide
Keysight Technologies
85056D 2.4 mm -f- and -m­85563A 2.4 mm -f- (3-in-1)
85564A 2.4 mm -m- (3-in-1) Economy Calibration Kits

Notices

No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Trademark Acknowledgments
Manual Part Number
85056-90021
Edition
Edition 2, September 2016
Printed in USA/Malaysia
Published by: Keysight Technologies
1400 Fountaingrove Parkway Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at
http://www.keysight.com/find/sweula
The license set forth in the EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose commercial computer software or commercial computer software
documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS 227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents

1. General Information
Calibration Kit Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Calibration Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Incoming Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Recording the Device Serial Numbers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Precision Slotless Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Clarifying the Terminology of a Connector Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
When to Calibrate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
cifications
2. Spe
Environmental Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Temperature—What to Watch Out For . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Pin Depth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Electrical Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Contents
3. Use,
Maintenance, and Care of the Devices
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Inspecting Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Look for Obvious Defects and Damage First . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect the Mating Plane Surfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Inspect Female Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Gaging Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Connector Gage Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
When to Gage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
How to Make a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
How to Separate a Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
4. Performance
Verification
How Keysight Verifies the Devices in Your Kit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
How Often to Recertify . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Where to Send a Kit for Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
85056D/ 85563A/85564A
i i i
Contents
5. Troubleshooting
Troubleshooting Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Where to Look for More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Returning a Kit or Device to Keysight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Contacting Keysight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
6. Replaceable Parts
A. Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web . . . . . . . . . . . . . . . A-1
i v
85056D/85563A/85564A
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

1 General Information

Calibration Kit Overview

The Keysight 85056D 2.4 mm standard calibration kit and 85563A/85664A 3-in-1 2.4 mm calibration kits are used to calibrate Keysight network analyzers up to 50 GHz for measurements of components with 2.4 mm connectors.

Kit Contents

Figure 1-1 85056D 2.4 mm calibration kit
The 85056D calibration kit includes the following items:
— offset opens and shorts
— broadband terminations
—2.4 mm adapters
— 5/16 in, 90 N-cm (8 in-lb) torque wrench
— 7 mm open-end wrench
1- 1
1-
General Information Calibration Kit Overview
Figure 1-2 85563A 3-in-1 (female) and 85564A 3-in-1 (male) calibration kits
The 85563A 3-in-1 calibration kit includes the following items:
— offset open and short (female)
— broadband termination (female)
The 85564A 3-in-1 calibration kit includes the following items:
— offset open and short (male)
— broadband termination (male)
85563A and 85564A are mostly subsets of the 85056D calibration kit. 85563A and 85564A contain a single sex connector (i.e., female only or male only). Also, while adapters are standard with the 85056D, they are optional on the 85563A and 85564A.
The open and shorts are identical in the kits. For example, the open female standard in the 85056D is the same open that is included in the 85563A. The load standards are different parts that have the same electrical performance.
Refer to Chapter 6, “Replaceable Parts,” for a complete list of kit contents and their associated part numbers.
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good broadband performance and to ensure stable, repeatable connections.
The adapters are designed so that their nominal electrical lengths are the same, allowing them to be used in calibration procedures for non-insertable devices.
Broadband Loads
The broadband loads are metrology-grade, 50Ω terminations that have been optimized for performance up to 50 GHz. The rugged internal structure provides for highly repeatable connections. A distributed resistive element on sapphire provides excellent stability and return loss.
1-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
General Information Calibration Kit Overview
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current state-of-the-art in precision machining.
The offset short’s inner conductors have a one-piece construction, common with the shorting plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong, low-dielectric-constant plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly, thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are designed so that the difference in phase of their reflection coefficients is approximately 180 degrees at all frequencies.

Calibration Definitions

The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
— resident within the analyzer
— entered from the front panel
Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at http://na.support.keysight.com/pna/caldefs/stddefs.html
Installation of the Calibration Definitions
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be manually entered. Refer to your network analyzer user’s guide for instructions.

Equipment Required but Not Supplied

Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are not supplied with the calibration kit but are required to ensure successful operation of the kit. Refer to Table 6-2 on page 2 for ordering information.
Keysight 85056D and 85563A/85564A User’s and Service Guide 1-3
1-
General Information Incoming Inspection

Incoming Inspection

Verify that the shipment is complete by referring to Table 6-1.
Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case and its contents are not damaged.
If the case or any device appears damaged, or if the shipment is incomplete, contact Keysight Technologies. See “Contacting Keysight” on page 5-5. Keysight will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company.
When you send the kit or device to Keysight, include a service tag (found near the end of this manual) with the following information:
— your company name and address
— the name of a technical contact person within your company, and the
person's complete phone number
— the model number and serial number of the kit
— the part number and serial number of the device
— the type of service required
—a detailed description of the problem
1-4 Keysight 85056D and 85563A/85564A User’s and Service Guide
General Information Recording the Device Serial Numbers

Recording the Device Serial Numbers

In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are labeled onto the body of each device). Record these serial numbers in
Table 1-1. Recording the serial numbers will prevent confusing the devices in
this kit with similar devices from other kits.
The adapters included in the kit are for measurement convenience only and are not serialized.
85563A and 85564A parts are serialized at the calibration kit level only.
Table 1-1 Serial Number Record for the 85056D
Device Serial Number
Calibration kit
–m– open
–f– open
–m– short
–f– short
–m– broadband load
–f– broadband load
–m– to –m– 2.4 mm adapter
–m– to -f- 2.4 mm adapter
–f– to –f– 2.4 mm adapter
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
Keysight 85056D and 85563A/85564A User’s and Service Guide 1-5
1-
General Information Precision Slotless Connectors

Precision Slotless Connectors

The female connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to national measurement standards through their well-defined mechanical dimensions.
Conventional female center conductors are slotted. When mated, the female center conductor is flared by the male pin. Because physical dimensions determine connector impedance, electrical characteristics of the female connector (and connection pair) are dependent upon the mechanical dimensions of the male pin. While connectors are used in pairs, their male and female halves are always specified separately as part of a standard, instrument, or device under test. Because of these facts, making precision measurements with the conventional slotted connector is very difficult, and establishing a direct traceability path to primary dimensional standards is nearly impossible.
The precision slotless connector was developed to eliminate these problems. All PSCs are female. A PSC incorporates a center conductor with a solid cylindrical shell that defines the outside diameter of the female center pin. Its outside diameter and, therefore, the impedance in its region does not change. The inner part provides an internal contact that flexes to accept the allowed range of male pin diameters.
The calibration of a network analyzer having a conventional slotted female connector on the test port remains valid only when the device under test and all calibration standards have identical male pin diameters. For this reason PSC test port adapters are supplied in most calibration kits.
Precision slotless connectors have the following characteristics:
— There is no loss of traceable calibration on test ports when the male pin
diameter of the connector on the device under test is different from the male pin diameter of the calibration standard.
— The female PSC and its mating male connector can be measured and
specified separately as part of the device either is attached to.
— All female connectors can have a known, stable impedance based only on
the diameters of their inner and outer conductors.
— Female calibration standards can be fully specified. Their specifications and
traceability are unaffected by the diameter of the male mating pin.
— A fully traceable performance verification is made using a precision 50 ohm
airline having a PSC.
— Measurement repeatability is enhanced due to non-changing connector
characteristics with various pin diameters.
With PSCs on test ports and standards, the percentage of accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
1-6 Keysight 85056D and 85563A/85564A User’s and Service Guide
General Information Clarifying the Terminology of a Connector Interface

Clarifying the Terminology of a Connector Interface

In this document and in the prompts of the PNA calibration wizard, the sex of cable connectors and adapters is referred to in terms of the center conductor. For example, a connector or device designated as 2.4 mm –f– has a 2.4 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector gage has a female center conductor on the gage so that it can measure male devices.

Preventive Maintenance

The best techniques for maintaining the integrity of the devices in the kit include:
— routine visual inspection
—cleaning
—proper gaging
— proper connection techniques
All of these techniques are described in Chapter 3. Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can damage any connector mated to it. Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-2 on page 3), or from bad connection techniques, can also damage these devices.

When to Calibrate

A network analyzer calibration remains valid as long as the changes in the systematic error are insignificant. This means that changes to the uncorrected leakages (directivity and isolation), mismatches (source match and load match), and frequency response of the system are small (<10%) relative to accuracy specifications.
Change in the environment (especially temperature) between calibration and measurement is the major cause in calibration accuracy degradation. The major effect is a change in the physical length of external and internal cables. Other important causes are dirty and damaged test port connectors and calibration standards. If the connectors become dirty or damaged, measurement repeatability and accuracy is affected. Fortunately, it is relatively easy to evaluate the general validity of the calibration. To test repeatability, remeasure one of the calibration standards. If you can not obtain repeatable measurements from your calibration standards, maintenance needs to be performed on the test port connectors, cables and calibration standards. Also,
Keysight 85056D and 85563A/85564A User’s and Service Guide 1-7
1-
General Information When to Calibrate
maintain at least one sample of the device under test or some known device as your reference device. A verification kit may be used for this purpose. After calibration, measure the reference device and note its responses. Periodically remeasure the device and note any changes in its corrected response which can be attributed to the test system. With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again.
1-8 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

2 Specifications

Environmental Requirements

Table 2-1 Environmental Requirements
Parameter Limits
Temperature
Operating
Storage 40 °C to +75 °C
Error-corrected range
Relative humidity Type tested, 0% to 95% at 40 °C, non-condensing
a
b
a. The temperature range over which the calibration standards maintain conformance to their specifications. b. The allowable network analyzer ambient temperature drift during measurement calibration and during
measurements when the network analyzer error correction is turned on. Also, the range over which the network analyzer maintains its specified performance while correction is turned on.
+20 °C to +26 °C
± 1 °C of measurement calibration temperature

Temperature—What to Watch Out For

Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a critical factor in performance. During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in
Table 2-1.
IMPORTANT! Avoid unnecessary handling of the devices during calibration because your fingers are a heat source.
2- 1
2-
Specifications Mechanical Characteristics

Mechanical Characteristics

Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Keysight Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing. This ensures that the device connectors do not exhibit any center conductor protrusion or improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-7 explains how to use gages to determine if
the kit devices have maintained their mechanical integrity. Refer to Table 2-2
on page 3 for typical and observed pin depth limits.

Pin Depth

Pin depth is the distance the center conductor mating plane differs from being flush with the outer conductor mating plane. See Figure 2-1. The pin depth of a connector can be in one of two states: either protruding or recessed.
Protrusion is the condition in which the center conductor extends beyond the outer conductor mating plane. This condition will indicate a positive value on the connector gage.
Recession is the condition in which the center conductor is set back from the outer conductor mating plane. This condition will indicate a negative value on the connector gage.
Figure 2-1 Connector Pin Depth
2-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
Specifications Mechanical Characteristics
The pin depth value of each calibration device in the kit is not specified, but is an important mechanical parameter. The electrical performance of the device depends, to some extent, on its pin depth. The electrical specifications for each device in the kit take into account the effect of pin depth on the device’s performance. Table 2-2 lists the typical pin depths and measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for a visual representation of proper pin depth (slightly recessed).
Table 2-2 Pin Depth Limits
Device Typical Pin Depth
Opens 0 to -0.0127 mm
0 to -0.00050 in
Shorts 0 to -0.0127 mm
0 to -0.00050 in
Fixed loads -0.0025 to -0.0203 mm
-0.00010 to -0.00080 in
Adapters 0 to -0.0381 mm
0 to -0.00150 in
a. Approximately +2 sigma to -2 sigma of gage uncertainty based on studies done at the factory according to
recommended procedures.
b. Observed pin depth limits are the range of observation limits seen on the gage reading due to measurement
uncertainty. The depth could still be within specifications.
Measurement Uncertainty
+0.0030 to -0.0030 mm +0.00012 to -0.00012 in.
+0.0015 to -0.0015 mm +0.00006 to -0.00006 in
+0.0030 to -0.0030 mm +0.00012 to -0.00012 in
+0.0030 to -0.0030 mm +0.00012 to -0.00012 in
a
Observed Pin Depth Limits
+0.0030 to -0.0157 mm +0.00012 to -0.00062 in
+0.0015 to -0.0142 mm +0.00006 to -0.00056 in
+0.0005 to -0.0234 mm +0.00002 to -0.00092 in
+0.0030 to -0.0411 mm +0.00012 to -0.00162 in
b
Keysight 85056D and 85563A/85564A User’s and Service Guide 2-3
2-
Specifications Electrical Specifications

Electrical Specifications

The electrical specifications in Tab le 2-3 apply to the devices in your calibration kit when connected with a Keysight precision interface.
Table 2-3 Electrical Specifications for 85056D/85563A/85564A 2.4 mm Devices
Device Specification Frequency (GHz)
Broadband loads Return loss ≥ 42 dΒ (ρ ≤ 0.00794) dc to ≤ 4
(male and female) Return loss 34 dΒ (ρ ≤ 0.01995) > 4 to 20
Return loss ≥ 30 dB (ρ ≤ 0.03162) > 20 to ≤ 26.5
Return loss ≥ 26 dB (ρ ≤ 0.05012) > 26.5 to ≤ 50
Adapters Return loss ≥ 32 dΒ (ρ ≤ 0.02512) dc to ≤ 4
Return loss 30 dΒ (ρ ≤ 0.03162) > 4 to 26.5
Return loss ≥ 25 dB (ρ ≤ 0.05623) > 26.5 to ≤ 40
Return loss ≥ 20 dB (ρ ≤ 0.10000) > 40 to ≤ 50
Offset opens
a
± 0.5° deviation from nominal
dc to ≤ 2
(male and female) ± 1.25° deviation from nominal > 2 to 20
± 1.75° deviation from nominal > 20 to 40
± 2.25° deviation from nominal > 40 to 50
Offset shorts
a
± 0.50° deviation from nominal
dc to ≤ 2
(male and female) ± 1.25° deviation from nominal > 2 to 20
± 1.5° deviation from nominal > 20 to 40
± 2.0° deviation from nominal > 40 to 50
a. The specifications for the opens and shorts are given an allowed deviation from the nominal model as
defined in the standard definitions (see “Class Assignments and Standard Definitions Values are
Available on the Web” on page 1).

Certification

Keysight Technologies certifies that this product met its published specifications at the time of shipment from the factory. Keysight further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities of other International Standards Organization members. See “How Keysight Verifies
the Devices in Your Kit” on page 4-1 for more information.
2-4 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

3 Use, Maintenance, and Care of the Devices

Electrostatic Discharge

Protection against electrostatic discharge (ESD) is essential while connecting, inspecting, or cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage. Devices such as calibration components and devices under test (DUT), can also carry an electrostatic charge. To prevent damage to the test set, components, and devices:
Always wear a grounded wrist strap having a 1 ΜΩ resistor in series with it
when handling components and devices or when making connections to
the test set. — Always use a grounded antistatic mat in front of your test equipment. — Always wear a heel strap when working in an area with a conductive floor.
If you are uncertain about the conductivity of your floor, wear a heel strap. — Always ground yourself before you clean, inspect, or make a connection to
a static-sensitive device or test port. You can, for example, grasp the
grounded outer shell of the test port or cable connector briefly. — Always ground the center conductor of a test cable before making a
connection to the analyzer test port or other static-sensitive device. This
can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Figure 3-1 shows a typical ESD protection setup using a grounded mat and
wrist strap. Refer to Table 6-2 on page 6-2 for information on ordering supplies for ESD protection.
3- 1
3-
Use, Maintenance, and Care of the Devices Electrostatic Discharge
Figure 3-1 ESD Protection Setup
3-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Inspecting Connectors

Inspecting Connectors

Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal particles from the connector threads may fall into the connector when it is disconnected. One connection made with a dirty or damaged connector can damage both connectors beyond repair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be misleading. Defects and damage that cannot be seen without magnification generally have no effect on electrical or mechanical performance. Magnification is of great use in analyzing the nature and cause of damage and in cleaning connectors, but it is not required for inspection.

Look for Obvious Defects and Damage First

Examine the connectors first for obvious defects and damage: badly worn plating on the connector interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets, the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or are not connected properly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life if their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the network analyzer test set may have many connections each day, and are therefore more subject to wear. It is recommended that an adapter be used as a test port saver to minimize the wear on the test set’s test port connectors.

Inspect the Mating Plane Surfaces

Flat contact between the connectors at all points on their mating plane surfaces is required for a good connection. See Figure 2-1 on page 2-2. Look especially for deep scratches or dents, and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow circular marks distributed more or less uniformly over the mating plane surface. Other small defects and cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-3
3-
Use, Maintenance, and Care of the Devices Inspecting Connectors
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration.

Inspect Female Connectors

Inspect the contact fingers in the female center conductor carefully. These can be bent or broken, and damage to them is not always easy to see. A connector with damaged contact fingers will not make good electrical contact and must be replaced.
This is particularly important when mating nonprecision to precision devices.
The female 2.4 mm connectors in this calibration kit are metrology-grade, precision slotless connectors (PSC). Precision slotless connectors are used to improve accuracy. With PSCs on test ports and standards, the accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times.
Conventional female center conductors are slotted and, when mated, are flared by the male pin. Because physical dimensions determine connector impedance, this change in physical dimension affects electrical performance, making it very difficult to perform precision measurements with conventional slotted female connectors.
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, this center conductor has an internal contact that flexes to accept the male pin.
3-4 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Cleaning Connectors

Cleaning Connectors

Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
1. Use Compressed Air or Nitrogen
Always use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. Clean air cannot damage a connector or leave particles or residues behind.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause electrostatic effects when directed into a connector. These electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this chapter for additional information.
2. Clean the Connector Threads
Keep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to Table 6-2 on page 6-2 for part numbers for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning
swab.
b. Clean the connector threads.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-5
3-
Use, Maintenance, and Care of the Devices Cleaning Connectors
c. Let the alcohol evaporate, then blow the threads dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning
swab.
b. Clean the center and outer conductor mating plane surfaces. Refer
to Figure 2-1 on page 2-2. When cleaning a female connector, avoid snagging the swab on the center conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle
stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
4. Inspect
Inspect the connector to make sure that no particles or residue remain. Refer to “Inspecting Connectors” on page 3-3.
3-6 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Gaging Connectors

Gaging Connectors

The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. See Table 6-2 on page 6-2 for part number information. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices. Do not use the gages for precise pin depth measurements.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices. This is partially due to the repeatability uncertainties that are associated with the measurement. Only the factory—through special gaging processes and electrical testing— can accurately verify the mechanical characteristics of the devices.
With proper technique, the gages are useful in detecting gross pin depth errors on device connectors. To achieve maximum accuracy, random errors must be reduced by taking the average of at least three measurements having different gage orientations on the connector. Even the resultant average can be in error by as much as ± 0.0001 inch due to systematic (biasing) errors usually resulting from worn gages and gage masters. The information in Table 2-2 on
page 2-3 assumes new gages and gage masters. Therefore, these systematic
errors were not included in the uncertainty analysis. As the gages undergo more use, the systematic errors can become more significant in the accuracy of the measurement.
The measurement uncertainties in Table 2-2 on page 2-3 are primarily a function of the assembly materials and design, and the unique interaction each device type has with the gage. Therefore, these uncertainties can vary among the different devices. For example, note the difference between the uncertainties of the opens and shorts in Table 2-2 on page 2-3.
The observed pin depth limits in Table 2-2 on page 2-3 add these uncertainties to the typical factory pin depth values to provide practical limits that can be referenced when using the gages. See “Pin Depth” on page 2-2. Refer to “Kit
Contents” on page 1-1 for more information on the design of the calibration
devices in the kit.
When measuring pin depth, the measured value (resultant average of three or more measurements) contains measurement uncertainty and is not necessarily the true value. Always compare the measured value with the observed pin depth limits (which account for measurement uncertainties) in Table 2-2 on page 2-3 to evaluate the condition of device connectors.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-7
3-
Use, Maintenance, and Care of the Devices Gaging Connectors

When to Gage Connectors

Gage a connector at the following times:
— Prior to using a device for the first time: record the pin depth measurement
so that it can be compared with future readings. (It will serve as a good troubleshooting tool when you suspect damage may have occurred to the device.)
— If either visual inspection or electrical performance suggests that the
connector interface may be out of typical range (due to wear or damage, for example).
— If a calibration device is used by someone else or on another system or
piece of equipment.
— Initially after every 100 connections, and after that as often as experience
indicates.
3-8 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Gaging Connectors

Gaging Procedures

Gaging 2.4 mm Connectors
Always hold a connector gage by the gage barrel, below the dial indicator. This gives the best stability, and improves measurement accuracy. (Cradling the gage in your hand or holding it by the dial applies stress to the gage plunger mechanism through the dial indicator housing.)
1. Select the proper gage for your connector. Refer to Table 6-2 on page 6-2 for gage part numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to
“Inspecting Connectors” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage
or the device, connect the gage to the gage master by interconnecting the male and female connectors. Connect the nut finger tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating,
use the torque wrench included in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-13 for additional information.
c. As you watch the gage pointer, gently tap the barrel of the gage to
settle the reading.
The gage pointer should line up exactly with the zero mark on the gage. If not, adjust the zero set knob until the gage pointer lines up exactly with the zero mark.
d. Remove the gage master.
4. Gage the device connector (refer to Figure 3-2):
a. While holding the gage by the barrel, and without turning the gage
or the device, connect the gage to the device by interconnecting the male and female connectors. Connect the nut finger-tight. Do not overtighten.
b. Using an open-end wrench to keep the device body from rotating,
use the torque wrench included in the kit to tighten the connecting nut to the specified torque. Refer to “Final Connection Using a
Torque Wrench” on page 3-13 for additional information.
c. Gently tap the barrel of the gage with your finger to settle the gage
read ing.
d. Read the gage indicator dial. Read only the black ± signs; not the
red ± signs.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-9
3-
Use, Maintenance, and Care of the Devices Gaging Connectors
For maximum accuracy, measure the connector a minimum of three times and take an average of the readings. After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in
Table 2-2 on page 2-3.
3-10 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Gaging Connectors
Figure 3-2 Gaging 2.4 mm Connectors
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-11
3-
Use, Maintenance, and Care of the Devices Making Connections

Making Connections

Good connections require a skilled operator. The most common cause of measurement error is bad connections. The following procedures illustrate how to make good connections.

How to Make a Connection

Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table mat. Refer to “Electrostatic Discharge” on
page 3-1 for ESD precautions.
2. Visually inspect the connectors. Refer to “Inspecting Connectors” on
page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on
page 3-5.
4. Use a connector gage to verify that all center conductors are within the observed pin depth values in Table 2-2 on page 2-3. Refer to “Gaging
Connectors” on page 3-7.
5. Carefully align the connectors. The male connector center pin must slip concentrically into the contact finger of the female connector.
6. Push the connectors straight together and tighten the connector nut finger-tight.
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform, light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the connection from long or heavy devices or cables.
3-12 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Making Connections
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Table 3-1 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Refer to Table 6-1 on
page 6-1 for replacement part number and ordering information.
Table 3-1 Torque Wrench Information
Connector Type Torque Setting Torque Tolerance
2.4 mm 90 N-cm (8 in-lb) ±9.0 N-cm (±0.8 in-lb)
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible to do this by hand if one of the connectors is fixed (as on a test port). However, it is recommended that you use an open-end wrench to keep the body of the device from turning.
1. Position both wrenches within 90 degrees of each other before applying force. See Figure 3-3. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together.
Figure 3-3 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-4.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-13
3-
Use, Maintenance, and Care of the Devices Handling and Storage
Figure 3-4 Using the Torque Wrench
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque.
4. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its internal pivot point. See Figure 3-4. Do not tighten the connection further.
You don’t have to fully break the handle of the torque wrench to reach the specified torque; doing so can cause the handle to kick back and loosen the connection. Any give at all in the handle is sufficient torque.

How to Separate a Connection

To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and connections.
Do not turn the device body. Only turn the connector nut. Damage to the center conductor can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connecting nut.
3. Complete the separation by hand, turning only the connecting nut.
4. Pull the connectors straight apart without twisting, rocking, or bending
either of the connectors.

Handling and Storage

— Install the protective end caps and store the calibration devices in the
foam-lined storage case when not in use.
3-14 Keysight 85056D and 85563A/85564A User’s and Service Guide
Use, Maintenance, and Care of the Devices Handling and Storage
— Never store connectors loose in a box, or in a desk or bench drawer. This is
the most common cause of connector damage during storage.
— Keep connectors clean.
— Do not touch mating plane surfaces. Natural skin oils and microscopic
particles of dirt are easily transferred to a connector interface and are very difficult to remove.
— Do not set connectors contact-end down on a hard surface. The plating and
the mating plane surfaces can be damaged if the interface comes in contact with any hard surface.
Keysight 85056D and 85563A/85564A User’s and Service Guide 3-15
3-
Use, Maintenance, and Care of the Devices Handling and Storage
3-16 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

4 Performance Verification

The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for recertification. The equipment required to verify the specifications of the devices in the kit has been specially manufactured and is not commercially available.

How Keysight Verifies the Devices in Your Kit

Keysight verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts that are directly traced to the National Institute of Standards and Technology (NIST). The airline and short characteristics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurements are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the calibration devices, Keysight includes the test measurement uncertainty as a guardband to guarantee each device meets the published specification. For recertifications (after sale), no guardband is used and the measured data is compared directly with the specification to determine the pass or fail status. The measurement uncertainty for each device is, however, recorded in the calibration report that accompanies recertified kits.
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Keysight Technologies.
4- 1
4-
Performance Verification Recertification

Recertification

The following will be provided with a recertified kit:
— a new calibration sticker affixed to the case
— a certificate of calibration
— a calibration report for each device in the kit listing measured values,
specifications, and uncertainties
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report.
Keysight Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Keysight Technologies. See “Contacting
Keysight” on page 5-5.

How Often to Recertify

The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a different recertification interval that reflects the usage and wear of the kit.
The recertification interval should begin on the date the kit is first used after the recertification date.

Where to Send a Kit for Recertification

Contact Keysight Technologies for information on where to send your kit for recertification. See “Contacting Keysight” on page 5-5. Refer to “Returning a
Kit or Device to Keysight” on page 5-4 for details on sending your kit.
4-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

5 Troubleshooting

Troubleshooting Process

If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1.
5- 1
5-
Troubleshooting Troubleshooting Process
Figure 5-1 Troubleshooting Flowchart
5-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
Troubleshooting Where to Look for More Information

Where to Look for More Information

This manual contains limited information about network analyzer system operation. For detailed information on using an ENA or PNA series network analyzer, refer to the appropriate user guide or Help file.
— To view the ENA or PNA Help, press the Help key on the front panel of the
network analyzer.
— To view Help or a user guide online, use the following steps:
1.Go to www.keysight.com.
2.Enter your analyzer model number (Ex: N5242A) in the Search box and click Search.
3.Under the heading Manuals, click on the title/hyperlink for the document PDF you want to view.
If you need additional information, see “Contacting Keysight” on page 5.
Keysight 85056D and 85563A/85564A User’s and Service Guide 5-3
5-
Troubleshooting Returning a Kit or Device to Keysight

Returning a Kit or Device to Keysight

If your kit or device requires service, contact Keysight Technologies for information on where to send it. See “Contacting Keysight” on page 5. Include a service tag (located near the end of this manual) on which you provide the following information:
— your company name and address
— a technical contact person within your company, and the person's complete
phone number
— the model number and serial number of the kit
— the part number and serial number of each device
— the type of service required
—a detailed description of the problem and how the device was being used
when the problem occurred (such as calibration or measurement)
5-4 Keysight 85056D and 85563A/85564A User’s and Service Guide
Troubleshooting Contacting Keysight

Contacting Keysight

Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at:
www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
In any correspondence or telephone conversation, refer to the Keysight product by its model number and full serial number. With this information, the Keysight representative can determine whether your product is still within its warranty period.
Keysight 85056D and 85563A/85564A User’s and Service Guide 5-5
5-
Troubleshooting Contacting Keysight
5-6 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

6 Replaceable Parts

Keysight does not offer replaceable parts on the 85563A or 85564A.
Table 6-1 lists the replacement part numbers for items included in the 85056D
calibration kit and Figure 6-1 illustrates each of these items.
Table 6-1 lists the replacement part numbers for items included in the 85056D
calibration kit and Figure 6-1 illustrates each of these items.
Table 6-2 lists the replacement part numbers for items not included in the
calibration kit that are either required or recommended for successful operation of the kit.
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Keysight Technologies. See
“Contacting Keysight” on page 5.
Table 6-1 Replaceable Parts for the 85056D Calibration Kit
Descriptiona
Calibration Devices (2.4 mm)
Broadband load -m- 1 00901-60013
Broadband load -f- 1 00901-60011
Offset open -m- 1 85056-60022
Offset open -f- 1 85056-60023
Offset short -m- 1 85056-60020
Offset short -f- 1 85056-60021
Adapters (2.4 mm)
Qty Per Kit
Keysight Part Number
00901-60003
00901-60004
b
b
-m- to -m- 1 85056-60005
-f- to -f- 1 85056-60006
6-1
6-
Replaceable Parts
Table 6-1 Replaceable Parts for the 85056D Calibration Kit
Descriptiona
Qty Per
Keysight Part Number
Kit
-m- to -f- 1 85056-60007
Calibration Kit Storage Case
Box (without foam pads) 1 5180-7862
Foam pad (for lid) 1 5181-5544
Foam pad (for lower case) 1 85052-80026
Was 85056-80023
Wrenches
5/16 in, 90 N-cm (8 in-lb) torque wrench 1 8710-1765
7 mm open-end wrench 1 8710-1761
Miscellaneous Items
User’s and service guide
a. Refer to “Class Assignments and Standard Definitions Values are Available on the Web” on
page 1.
b. Part numbers in italic typeface are for production assemblies and cannot be ordered by customers. Both the
replacement part number and the production part number are interchangeable and have identical performance.
c. Refer to
www.keysight.com.
c
1 85056-90021
Table 6-2 Items Not Included in the Calibration Kit
Description Qty
Per Kit
Connector Gages (2.4 mm)
Male gage set (includes gage master) 1 11752-60108
Female gage set (includes gage master) 1 11752-60107
Open-End Wrench
5/16 in open-end wrench 1 8720-0015
ESD Protection Devices
Heel strap 9300-1308
Grounding wrist strap 1 9300-1367
5 ft grounding cord for wrist strap 1 9300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire 1 9300-0797
Keysight Part Number
6-2 Keysight 85056D and 85563A/85564A User’s and Service Guide
Replaceable Parts
Table 6-2 Items Not Included in the Calibration Kit
Description Qty
Per Kit
Connector Cleaning Supplies
Anhydrous isopropyl alcohol (>92% pure)
Foam-tipped cleaning swabs 100 9301-1243
a. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally.
a
-- --
Keysight Part Number
Keysight 85056D and 85563A/85564A User’s and Service Guide 6-3
6-
Replaceable Parts
Figure 6-1 Replaceable Parts for the 85056D Calibration Kit
6-4 Keysight 85056D and 85563A/85564A User’s and Service Guide
Keysight Technologies 85056D and 85563A/85564A
2.4 mm Economy Calibration Kits
User’s and Service Guide

A: Standard Definitions

Class Assignments and Standard Definitions Values are Available on the Web

Class assignments and standard definitions may change as more accurate model and calibration methods are developed. You can download the most recent class assignments and standard definitions from Keysight’s Calibration Kit Definitions Web page at
http://na.support.keysight.com/pna/caldefs/stddefs.html.
For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying Calibration Standards and Kits for Keysight Vector Network Analyzers.” This application note covers calibration standard definitions, calibration kit content and its structure requirements for Keysight vector network analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an existing calibration kit definition file. To download a free copy, go to www.keysight.com and enter literature number 5989-4840EN in the Search window.
A-1
-
A­Class Assignments and Standard Definitions Values are Available on the Web
A-2 Keysight 85056D and 85563A/85564A User’s and Service Guide

Index

Index
A
adapters
part numbers
1-2
6-1
alcohol
isopropyl
as cleaning solvent precautions for use of
3-5
3-5
altitude
specifications
2-1
antistatic mat
part number
6-2
B
box
pads
part numbers
broadband loads
part numbers specifications
6-2
1-2
6-1
2-4
C
cal kit
performance
verifying
4-1
calibration
bad
5-1
certificate of
4-2
constants, See calibration definitions definitions
entering
permanently stored frequency report temperature when to perform
1-3
1-3
1-7
4-2
2-1
1-7
calibration kit
contents
drawing of Keysight Application Note modifying definition files overview
1-1, 6-4
6-4
A-1
A-1
1-1
performance
how Keysight verifies
verifying
4-1
4-1
case
part number
certificate of calibration
6-2
4-2
certification
of device specifications
2-4
characteristics
mechanical
2-2
class assignments
downloading from Keysight Web site cleaning connectors cleaning supplies
part numbers cleaning swabs
part number
6-3
3-5
1-3
6-3
3-5
A-1
compressed air
for cleaning connections
ESD protection
3-13
final
preliminary
separating
using torque wrench
3-5
3-12
3-12
3-12
3-14
3-13
connector
cleaning
damage
defects
female
3-5
3-3
3-3
3-4
gage
accuracy handling part numbers use of zeroing
gaging
to determine pin depth when to do
mating plane surfaces
cleaning precision slotless (female) slotless terminology
3-7
3-7, 3-9
6-2
3-7
3-7, 3-9
3-7, 3-9
3-7
3-8
3-6
3-6
1-6
3-4
1-7
threads
cleaning
inspecting visual inspection wear
affect on electrical performance
3-5
3-3
3-3
3-3
3-3
constants, calibration, See calibration definitions contacting Keysight Technologies
5-4
contents
85056D/85563A/85564A Index
1
Index
calibration kit 6-4 drawing of
6-4
incomplete
what to do
1-4
cord
grounding
part number
6-2
D
damage
caused by electrostatic discharge device inspecting for to connectors what to do
damaged connectors
3-3
3-3 3-3
1-4
3-3
3-1
data
recertification
defective connectors
4-2
3-3
defects
connector
3-3
definitions
calibration
deviation from nominal phase
1-3
2-4
device
connecting damage disconnecting handling maintenance
3-12
3-3
3-14
3-14
1-7
performance
verifying
specifications
certification of
traceability storage temperature visual inspection
disconnections
4-1
2-4
2-4
4-1, 4-2
3-14
2-1
3-3
3-14
downloading class assignments & std definitions from the Web
A-1
equipment required ESD
3-1
1-3
antistatic mat
part number
6-2
heel strap
part number precautions protection
setup
3-1
supplies
part numbers
6-2
3-5
3-1
3-1
6-2
wrist strap
part number
6-2
F
female connectors
inspection of
3-4
3-4
frequency
specifications
2-4
frequency of calibration
G
gage
connector
handling
part numbers zeroing
1-3
3-9
6-2
3-9
gaging
connectors
when to do
3-7, 3-9
3-8
to determine pin depth
grounding cord
part number
6-2
H
handling
3-14
heel strap
part number
6-2
how often to calibrate humidity
specifications
2-1
1-7
3-7
1-7
E
electrical specifications
2-4
electrostatic discharge, See ESD environmental
regulations requirements specifications
2
3-5
2-1
2-1
I
incoming inspection
1-4
information, troubleshooting inspection
damage defects female connectors
3-3
3-3
3-4
85056D/85563A/85564A Index
5-3
Index
incoming 1-4 mating plane surfaces visual
3-3
3-3
isopropyl alcohol
as cleaning solvent precautions for use of
3-5
3-5
K
Keysight Technologies
application note contacting
A-1
5-4, 5-5
kit
contents
drawing of
overview
1-1, 6-4
6-4
1-1
L
limits
pin depth
2-3
loads
broadband
part numbers
1-2
6-1
M
maintenance
of devices preventive
making connections
ESD protection precautions
3-1
1-7
1-7
3-12
3-12
3-12
mat
antistatic
part number
6-2
conductive table
part number
6-2
mating plane surfaces
cleaning connector inspection of
mechanical characteristics
affect on electrical performance verifying
modifying calibration kit definition files
3-6
3-6
3-3
2-2
2-2
3-7
A-1
N
National Institute of Standards and Technology (NIST)
2-4, 4-1
nitrogen
for cleaning
3-5
numbers
replaceable parts serial
1-5
recording
1-5
6-1
O
observed limits
pin depth
2-3
offset opens
part numbers
6-1
offset shorts
part numbers
offsets
1-3
open-end wrench
6-1
3-14
5/16 in
part number
6-2
7-mm
part number
opens
1-3
part numbers specifications
6-2
6-1
2-4
ordering
parts
6-1
P
pads
box
part numbers
part numbers
of items in kit of items not in kit
6-2
6-1
6-1
6-2
parts
included in kit not included in kit ordering replaceable
6-1
6-1, 6-2
6-1
6-1
performance verification
5-1
fail
permanently stored calibration definitions pin depth
affect on electrical performance gaging to determine observed limits protrusion recession typical values
2-2
2-3
3-7
2-3, 3-7
2-2
2-2
2-3
precision adapters
part numbers
precision slotless connectors
6-1
1-6
1-3
85056D/85563A/85564A Index
3
Index
preventive maintenance 1-7 protrusion
pin depth
2-2
R
recertification
how to order interval what's included where it’s done
4-2
4-2
4-2
4-2
recession
pin depth
2-2
regulations
environmental
replaceable parts
drawing of
report, calibration
3-5
6-1, 6-4
6-4
4-2
requirements
environmental
2-1
return
kit or device to Keysight
5-4
return loss
specifications
2-4
S
separating connections serial numbers
devices recording
service
5-4
service tag
1-5
1-5
1-5
1-4, 4-2, 5-4
3-14
setup
ESD protection
1-3
shorts
part numbers specifications
specifications
3-1
6-1
2-4
2-1
altitude
operating storage
certification of
2-1
2-1
2-4
deviation from nominal phase device electrical environmental frequency humidity
return loss
2-4
storage
2-4
2-1
2-4 2-1 2-1
2-4
2-4
temperature torque wrench traceability
2-1
3-13
4-1, 4-2
standard definitions
downloading from Keysight Web site
A-1
standards
international
2-4
National Institute of Standards and Technology
(NIST)
2-4, 4-1
static
discharge electricity
storage
3-1
3-1
3-14
storage case
part number
6-2
strap
heel
part number
6-2
wrist
part number
6-2
supplies
cleaning
part number
1-3
6-3
swabs
cleaning
part number
3-5
6-3
T
tag
service
1-4, 4-2, 5-4
temperature
affect on electrical performance calibration cautions about changes in device error-corrected measurement specifications
operating storage
2-1
2-1
2-1
2-1
2-1 2-1 2-1
2-1
2-1
verification and measurement terminology, connector test data
4-2
1-7
2-1
2-1
threads
connector
cleaning inspecting
3-5
3-3
tools
4
85056D/85563A/85564A Index
part numbers 6-2
torque wrench
part number specifications
1-3
6-2
3-13
traceability
of device specifications
troubleshooting
5-1
4-1, 4-2
U
user’s and service guide
part number
6-2
V
verification
temperature
visual inspection
2-1
3-3
W
wear
connector
3-3
affect on electrical performance
when to calibrate
1-7
wrench
7 mm open-end
part number
open-end
part number proper positioning of torque
1-3, 3-13, 3-14
part number
precautions for use of
proper use of
6-2
1-3, 3-13, 3-14
6-2
3-13
6-2
3-14
3-14
wrist strap
part number
6-2
Index
3-3
Z
zeroing
connector gage
85056D/85563A/85564A Index
3-9
5
Index
6
85056D/85563A/85564A Index
This information is subject to change without notice. © Keysight Technologies 1996-2016
Edition 1, September 2016
www.keysight.com
Loading...