THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS
SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR
CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
Assistance
Product maintenance agreements and other customer assistance agreements are available for
Keysight products.
For any assistance, contact Keysight Technologies. Refer to “Contacting Keysight” on
page 5-5.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Keysight web
site:
•Go to www.keysight.com.
•Enter the document’s part number (located on the title page) in the Search box.
•Click Search.
•Click on the hyperlink for the document.
•Click the printer icon located in the tool bar.
This manual applies directly to 85054B calibration kits with serial number prefix 3101 and later.
The calibration devices in this kit are individually serialized. Record the device serial numbers
in the table provided in this manual (See “Recording the Device Serial Numbers” on page 5..)
Class Assignments and Standard Definitions Values are Available on the Web . . . . . . . . . . . . . . . . . . . . .A-2
. . . . . . . . . . . 5-5
. . . . 5-3
. . . . . . . 5-4
85054B
1 General Information
General Information
Calibration Kit Overview
Calibration Kit Overview
The Keysight 85054B type-N calibration kit is used to calibrate Keysight network analyzers up
to 18 GHz for measurements of components with 50type-N connectors.
The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM
(thru–reflect–match) calibrations.
This manual describes the 85054B calibration kit and provides replacement part numbers,
specifications, and procedures for using, maintaining, and troubleshooting the kit.
Kit Contents
The 85054B calibration kit contains the following:
•offset opens and shorts, and lowband and sliding load terminations
•four type-N to 7 mm adapters
•two type-N to type-N adapters
•a type-N connector gage set
•a 3/4 inch, 135 N-cm (12 in-lb) torque wrench for use on the type-N connectors
•a spanner wrench
Refer to Table 6-1, Figure 6-1, and Figure 6-2 for a complete list of kit contents and their
associated part numbers.
Offset Opens and Shorts
The offset opens and shorts are built from parts that are machined to the current
state-of-the-art in precision machining.
The offset short’s inner conductors have a one-piece construction, common with the shorting
plane. The construction provides for extremely repeatable connections.
The offset opens have inner conductors that are supported by a strong, low-dielectricconstant
plastic to minimize compensation values.
Both the opens and shorts are constructed so that the pin depth can be controlled very tightly,
thereby minimizing phase errors. The lengths of the offsets in the opens and shorts are
designed so that the difference in phase of their reflection coefficients is approximately 180
degrees at all frequencies.
Lowband Loads
The lowband loads are metrology-grade, 50 ohm terminations which have been optimized for
lowband performance up to 2 GHz. The rugged internal structure provides for highly repeatable
connections. A distributed resistive element on sapphire provides excellent stability and return
loss.
Sliding Loads
The sliding loads in this kit are designed to provide excellent performance from 3 GHz to 18
1- 285054B
General Information
Calibration Kit Overview
GHz. The inner and outer conductors of the airline portion are precision machined to
state-of-the-art tolerances. Although the sliding load has exceptional return loss, its superior
load stability qualifies it as a high-performance device.
The sliding load was designed with the ability to extend the inner conductor for connection
purposes and then pull it back to a preset pin depth. This feature is critical since it minimizes
the possibility of damage during the connection, while maintaining a minimum pin depth to
optimize performance.
Adapters
Like the other devices in the kit, the adapters are built to very tight tolerances to provide good
broadband performance. The adapters utilize a dual-beaded connector structure to ensure
stable, repeatable connections. The beads are designed to minimize return loss and are
separated far enough so that interaction between the beads is minimized.
Calibration Definitions
The calibration kit must be selected and the calibration definitions for the devices in the kit
installed in the network analyzer prior to performing a calibration. Refer to your network
analyzer user’s guide for instructions on selecting the calibration kit and performing a
calibration.
The calibration definitions can be:
•resident within the analyzer
•entered from the front panel
Class assignments and standard definitions may change as more accurate model and
calibration methods are developed. You can download the most recent class assignments and
standard definitions from Keysight’s Calibration Kit Definitions Web page at
The calibration definitions for the kit may be permanently installed in the internal memory or
hard disk of the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer,
they must be manually entered. Refer to your network analyzer user’s guide for instructions.
85054B 1-3
General Information
Equipment Required but Not Supplied
Equipment Required but Not Supplied
Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are
not supplied with the calibration kit but are required to ensure successful operation of the kit.
Refer to Table 6-2 on page 6-3 for ordering information.
Incoming Inspection
Refer to “Kit Contents” on page 1-2 to verify a complete shipment. Use Table 1-1 on page 1-5
to record the serial numbers of all serialized devices in your kit.
Check for damage. The foam-lined storage case provides protection during shipping. If the
case or any device appears damaged, or if the shipment is incomplete, refer to “Contacting
Keysight” on page 5-5. Keysight will arrange for repair or replacement of incomplete or
damaged shipments without waiting for a settlement from the transportation company. See
“Returning a Kit or Device to Keysight” on page 5-4.
1- 485054B
General Information
Serial Numbers
Serial Numbers
A serial number is attached to this calibration kit. The first four digits followed by a letter
comprise the serial number prefix; the last five digits are the suffix, unique to each calibration
kit.
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in the kit are individually serialized (serial
numbers are labeled onto the body of each device). Record these serial numbers in
Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with
similar devices from other kits.
Table 1-1 Serial Number Record for the 85054B
DeviceSerial Number
Calibration kit
Calibration Devices
Lowband load –m–
Lowband load –f–
Open–m–
Open –f–
Short –m–
Short –f–
Sliding load –f–
Sliding load –m–
Gages
Gage Master–m–
Gage Master –f–
Connector gage –m–
Connector gage –f–
Adapters
Type-N –m– to Type-N –m–
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
____________________________
Type-N –f– to Type-N –f–
Type-N –f– to 7 mm
Type-N –m– to 7 mm
85054B 1-5
____________________________
____________________________
____________________________
General Information
Precision Slotless Connectors
Precision Slotless Connectors
The female type-N connectors in this calibration kit are metrology-grade, precision slotless
connectors (PSC). A characteristic of metrology-grade connectors is direct traceability to
national measurement standards through their well-defined mechanical dimensions.
Conventional female center conductors are slotted. When mated, the female center conductor
is flared by the male pin. Because physical dimensions determine connector impedance,
electrical characteristics of the female connector (and connection pair) are dependent upon the
mechanical dimensions of the male pin. While connectors are used in pairs, their male and
female halves are always specified separately as part of a standard, instrument, or device
under test. Because of these facts, making precision measurements with the conventional
slotted connector is very difficult, and establishing a direct traceability path to primary
dimensional standards is nearly impossible.
The precision slotless connector was developed to eliminate these problems. All PSCs are
female. A PSC incorporates a center conductor with a solid cylindrical shell that defines the
outside diameter of the female center pin. Its outside diameter and, therefore, the impedance in
its region does not change. The inner part provides an internal contact that flexes to accept the
allowed range of male pin diameters.
The calibration of a network analyzer having a conventional slotted female connector on the
test port remains valid only when the device under test and all calibration standards have
identical male pin diameters. For this reason PSC test port adapters are supplied in most
calibration kits.
Precision slotless connectors have the following characteristics:
•There is no loss of traceable calibration on test ports when the male pin diameter of the
connector on the device under test is different from the male pin diameter of the calibration
standard.
•The female PSC and its mating male connector can be measured and specified separately
as part of the device either is attached to.
•All female connectors can have a known, stable impedance based only on the diameters of
their inner and outer conductors.
•Female calibration standards can be fully specified. Their specifications and traceability are
unaffected by the diameter of the male mating pin.
•A fully traceable performance verification is made using a precision 50 ohm airline having a
PSC.
•Measurement repeatability is enhanced due to non-changing connector characteristics with
various pin diameters.
With PSCs on test ports and standards, the percentage of accuracy achieved when measuring
at 50 dB return loss levels is comparable to using conventional slotted connectors measuring
devices having only 30 dB return loss. This represents an accuracy improvement of about 10
times.
1- 685054B
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of cable connectors
and adapters is referred to in terms of the center conductor. For example, a connector or
device designated as 1.85 mm –f– has a 1.85 mm female center conductor.
A connector gage is referred to in terms of the connector that it measures. For instance, a male
connector gage has a female connector on the gage so that it can measure male devices.
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in this kit include:
•routine visual inspection
•cleaning
•proper gaging
•proper connection techniques
All of the above are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.”
Failure to detect and remove dirt or metallic particles on a mating plane surface can degrade
repeatability and accuracy and can damage any connector mated to it. Improper connections,
resulting from pin depth values being out of the observed limits (see Table 2-2 on page 2-4), or
from bad connections, can also damage these devices.
85054B 1-7
General Information
When to Calibrate
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the systematic error are
insignificant. This means that changes to the uncorrected leakages (directivity and isolation),
mismatches (source match and load match), and frequency response of the system are small
(<10%) relative to accuracy specifications.
Change in the environment (especially temperature) between calibration and measurement is
the major cause in calibration accuracy degradation. The major effect is a change in the
physical length of external and internal cables. Other important causes are dirty and damaged
test port connectors and calibration standards. If the connectors become dirty or damaged,
measurement repeatability and accuracy is affected. Fortunately, it is relatively easy to
evaluate the general validity of the calibration. To test repeatability, remeasure one of the
calibration standards. If you can not obtain repeatable measurements from your calibration
standards, maintenance needs to be performed on the test port connectors, cables and
calibration standards. Also, maintain at least one sample of the device under test or some
known device as your reference device. A verification kit may be used for this purpose. After
calibration, measure the reference device and note its responses. Periodically remeasure the
device and note any changes in its corrected response which can be attributed to the test
system. With experience you will be able to see changes in the reference responses that
indicate a need to perform the measurement calibration again.
1- 885054B
2 Specifications
Specifications
Environmental Requirements
Environmental Requirements
Table 2-1 Environmental Requirements
Parameter Limits
Temperature
Operating
Storage40 C to +75 C
Error-corrected range
Relative humidityType tested, 0% to 95% at 40 C, non-condensing
a. The temperature range over which the calibration standards maintain conformance to their
b. The allowable network analyzer ambient temperature drift during measurement calibration and
a
b
specifications.
during measurements when the network analyzer error correction is turned on. Also, the range
over which the network analyzer maintains its specified performance while correction is turned
on.
+20 C to +26 C
1 C of measurement calibration temperature
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range specified in Table
2-1.
IMPORTANT Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
2- 285054B
Specifications
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as center conductor protrusion and pin depth are not
performance specifications. They are, however, important supplemental characteristics related
to the electrical performance of devices. Keysight Technologies verifies the mechanical
characteristics of the devices in this kit with special gaging processes and electrical testing.
This ensures that the device connectors do not exhibit any excess center conductor protrusion
or improper pin depth when the kit leaves the factory.
“Gaging Connectors” on page 3-6 explains how to use gages to determine if the kit devices
have maintained their mechanical integrity. (Refer to Table 2-2 on page 2-4 for typical and
observed pin depth limits.)
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with the
outer conductor mating plane. Refer to Figure 2-1. Some coaxial connectors, such as 2.4 mm
and 3.5 mm, are designed to have these planes nearly flush. Type-N connectors, however, are
designed with a pin depth offset of approximately 5.26 mm (0.207 inch), not permitting these
planes to be flush. The male center conductors are recessed by the offset value while the
female center conductors compensate by protruding the same amount. This offset necessitates
the redefining of pin depth with regard to protrusion and recession.
Protrusion refers to a male type-N connector center conductor having a pin depth value less
than 5.26 mm (0.207 inch), or a female type-N connector center conductor having a pin depth
value greater than 5.26 mm (0.207 inch).
Recession refers to a male type-N connector center conductor having a pin depth value
greater than 5.26 mm (0.207 in), or a female type-N connector center conductor having a pin
depth value less than 5.26 mm (0.207 inch).
Figure 2-1 Connector Pin Depth
85054B 2-3
Specifications
Mechanical Characteristics
NOTEThe gages for measuring type-N connectors compensate for the designed offset
of 5.26 mm (0.207 inch), therefore, protrusion and recession readings are in
relation to a zero reference plane (as if the inner and outer conductor planes were
intended to be flush). Gage readings can be directly compared with the observed
values listed in Table 2-2.
The pin depth value of each calibration device in this kit is not specified, but is an important
mechanical parameter. The electrical performance of the device depends, to some extent, on
its pin depth. The electrical specifications for each device in this kit take into account the effect
of pin depth on the device’s performance. Table 2-2 lists the typical pin depths and
measurement uncertainties, and provides observed pin depth limits for the devices in the kit. If
the pin depth of a device does not measure within the observed pin depth limits, it may be an
indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for an
illustration of pin depth in type-N connectors.
Table 2-2 Pin Depth Limits
Device
Typical Pin Depth
micrometers
–4
inches)
(10
Opens0 to 12.7
(0 to 5.0)
Shorts0 to 12.7
(0 to 5.0)
Lowband
loads
Sliding loads0 to 7.6
Adapters
(7 mm end)
Adapters
(type-N end)
a. Approximately +2 sigma to 2 sigma of gage uncertainty based on studies done at the factory
according to recommended procedures.
b. Observed pin depth limits are the range of observation limits seen on the gage reading due to
measurement uncertainty. The depth could still be within specifications.
0 to 50.8
(0 to 20.0)
(0 to 3.0)
0 to 50.8
(0 to 20.0)
0 to 12.7
(0 to 5.0)
Measurement
Uncertainty
a
micrometers
–4
inches)
(10
+3.8 to 3.8
(+ 1.5 to 1.5)
+3.8 to 3.8
(+ 1.5 to 1.5)
+3.8 to 3.8
(+ 1.5 to 1.5)
+3.8 to 3.8
(+ 1.5 to 1.5)
+3.8 to 3.8
(+ 1
.5 to 1.5)
+3.8 to 3.8
(+ 1.5 to 1.5)
Observed Pin Depth
b
Limits
micrometers
–4
inches)
(10
+3.8 to 16.5
(+ 1.5 to 6.5)
+3.8 to 16.5
(+ 1.5 to 6.5)
+3.8 to 54.6
(+ 1.5 to 21.5)
+3.8 to 11. 4
(+ 1.5 to 4.5)
+3.8 to 54.6
(+ 1.5 to 21.5)
+3.8 to 16.5
(+ 1.5 to 6.5)
NOTEWhen measuring pin depth, the measured value (resultant average of three or
more measurements) is not the true value. Always compare the measured value
with the observed pin depth limits in Table 2-2 to evaluate the condition of device
connectors.
2- 485054B
Specifications
Electrical Specifications
Electrical Specifications
The electrical specifications in Table 2-3 apply to the devices in your calibration kit when
connected with a Keysight precision interface.
Table 2-3 Electrical Specifications
DeviceFrequency (GHz)ParameterSpecification
Lowband loads DC to 2Return Loss48d0.00398)
Sliding loads
Adapters
(both styles)
Offset Opens
Offset Shorts
a. Assuming proper usage, the specifications for the residual return loss after calibration for the
b. The specifications for the opens and shorts are given as allowed deviation from the nominal
a
b
b
sliding load termination include:
• the quality of the airline portions within the sliding load, combined with
• the effective stability of the sliding element.
Proper usage includes the following practices:
Connector mating surfaces are clean.
•
• The changes in slide positioning are NOT done in equal steps since this results in very poor
calibration for some portions of the frequency range.
• The center conductor of test port connectors are nominally set back from the outer
conductor.
Sliding loads are designed to allow the center conductor to be moved. The position of the
sliding load center conductor should be set by a reference block and not positioned flush
against the center conductor of the test port.
model as defined in the standard definitions (see
Definitions Values are Available on the Web” on page A-2
> 2 to 18Return Loss42d0.00794)
DC to 8
> 8 to 18
at 18Deviation from Nominal
at 18Deviation from Nominal
Return Loss
Return Loss
Phase
Phase
34d0.0200)
28d0.0398)
1.5
1.0
“Class Assignments and Standard
).
Certification
Keysight Technologies certifies that this product met its published specifications at the time of
shipment from the factory. Keysight further certifies that its calibration measurements are
traceable to the United States National Institute of Standards and Technology (NIST) to the
extent allowed by the Institute's calibration facility, and to the calibration facilities of other
International Standards Organization members. See “How Keysight Verifies the Devices in This
Kit” on page 4-2 for more information.
85054B 2-5
Specifications
Electrical Specifications
2- 685054B
3 Use, Maintenance, and Care of the
Devices
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or
cleaning connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit
elements when discharged. Static discharges too small to be felt can cause permanent
damage. Devices such as calibration components and devices under test (DUTs), can also
carry an electrostatic charge. To prevent damage to the test set, components, and devices:
•always wear a grounded wrist strap having a 1 M resistor in series with it when handling
components and devices or when making connections to the test set.
•always use a grounded, conductive table mat while making connections.
•always wear a heel strap when working in an area with a conductive floor. If you are
uncertain about the conductivity of your floor, wear a heel strap.
•always ground yourself before you clean, inspect, or make a connection to a static-sensitive
device or test port. You can, for example, grasp the grounded outer shell of the test port or
cable connector briefly.
•always ground the center conductor of a test cable before making a connection to the
analyzer test port or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center
conductor to the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 6 , “Replaceable Parts,” for part numbers and instructions for ordering ESD
protection devices.
Figure 3-1 ESD Protection Setup
3- 285054B
Use, Maintenance, and Care of the Devices
Visual Inspection
Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made.
Metal particles from the connector threads may fall into the connector when it is disconnected.
One connection made with a dirty or damaged connector can damage both connectors beyond
repair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be
misleading. Defects and damage that cannot be seen without magnification generally have no
effect on electrical or mechanical performance. Magnification is of great use in analyzing the
nature and cause of damage and in cleaning connectors, but it is not required for inspection.
Look for Obvious Defects and Damage First
Examine the connectors first for obvious defects and damage: badly worn plating on the
connector interface, deformed threads, or bent, broken, or misaligned center conductors.
Connector nuts should move smoothly and be free of burrs, loose metal particles, and rough
spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a
connector gets, the faster it wears and degrades. The wear is greatly accelerated when
connectors are not kept clean, or are connected incorrectly.
Connector wear eventually degrades performance of the device. Calibration devices should
have a long life if their use is on the order of a few times per week. Replace devices with worn
connectors.
The test port connectors on the network analyzer test set may have many connections each
day, and are therefore more subject to wear. It is recommended that an adapter be used as a
test port saver to minimize the wear on the test set’s test port connectors.
Inspect the Mating Plane Surfaces
Flat contact between the connectors at all points on their mating plane surfaces is required for
a good connection. See Figure 2-1 on page 2-3. Look especially for deep scratches or dents,
and for dirt and metal particles on the connector mating plane surfaces. Also look for signs of
damage due to excessive or uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or
shallow circular marks distributed more or less uniformly over the mating plane surface. Other
small defects and cosmetic imperfections are also normal. None of these affect electrical or
mechanical performance.
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces,
or uneven wear, clean and inspect it again. Devices with damaged connectors should be
discarded. Determine the cause of damage before connecting a new, undamaged connector in
the same configuration.
85054B 3-3
Use, Maintenance, and Care of the Devices
Cleaning Connectors
Inspect the Precision Slotless Connectors (female)
Precision slotless female connectors are used to improve accuracy. The slotless contacts are
not affected by the slight variations in male contact pin diameter. However, it is still advisable to
inspect them regularly for damage.
NOTEThis is particularly important when mating nonprecision to precision devices.
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial
connections.
1. Use Compressed Air or Nitrogen
WARNINGAlways use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane
surfaces.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an
effective oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less
than 414 kPa (60 psi) to control the velocity of the air stream. High-velocity streams of
compressed air can cause electrostatic effects when directed into a connector. These
electrostatic effects can damage the device. Refer to “Electrostatic Discharge” earlier in this
chapter for additional information.
2. Clean the Connector Threads
WARNINGKeep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly
closed container. It is extremely flammable. In case of fire, use alcohol
foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with
eyes, skin, and clothing. It causes skin irritation, may cause eye damage,
and is harmful if swallowed or inhaled. It may be harmful if absorbed
through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal,
state, and local environmental regulations.
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or
stubborn contaminants on a connector that cannot be removed with compressed air or
nitrogen. Refer to Table 6-2 on page 6-3 for a part number for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
3- 485054B
Use, Maintenance, and Care of the Devices
Cleaning Connectors
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on
page 2-3. When cleaning a female connector, avoid snagging the swab on the center
conductor contact fingers by using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean,
low-pressure compressed air or nitrogen. Always completely dry a connector before you
reassemble or use it.
4. Inspect
Inspect the connector again to make sure that no particles or residue are present.
85054B 3-5
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Connectors
The gages available from Keysight Technologies are intended for preventive maintenance and
troubleshooting purposes only. (See Table 6-1 on page 6-2 for part number information.) They
are effective in detecting excessive center conductor protrusion or recession, and conductor
damage on DUTs, test accessories, and the calibration kit devices.Do not use the gages for
precise pin depth measurements.
Connector Gage Accuracy
The connector gages are only capable of performing coarse measurements. They do not
provide the degree of accuracy necessary to precisely measure the pin depth of the kit devices.
This is partially due to the repeatability uncertainties that are associated with the measurement.
Only the factory—through special gaging processes and electrical testing— can accurately
verify the mechanical characteristics of the devices.
With proper technique, however, the gages are useful in detecting gross pin depth errors on
device connectors. To achieve maximum accuracy, random errors must be reduced by taking
the average of at least three measurements having different gage orientations on the
connector. Even the resultant average can be in error by as much as 0.0001 inch due to
systematic (biasing) errors usually resulting from worn gages and gage masters. The
information in Table 2-2 on page 2-4 assumes new gages and gage masters. Therefore, these
systematic errors were not included in the uncertainty analysis. As the gages undergo more
use, the systematic errors can become more significant in the accuracy of the measurement.
The measurement uncertainties (see Table 2-2 on page 2-4) are primarily a function of the
assembly materials and design, and the unique interaction each device type has with the gage.
Therefore, these uncertainties can vary among the different devices. For example, note the
difference between the uncertainties of the opens and shorts in Table 2-2.
The observed pin depth limits in Table 2-2 add these uncertainties to the typical factory pin
depth values to provide practical limits that can be referenced when using the gages. See “Pin
Depth” on page 2-3. Refer to “Kit Contents” on page 1-2 for more information on the design of
the calibration devices in this kit.
NOTEWhen measuring pin depth, the measured value (resultant average of three or
more measurements) contains measurement uncertainty and is not necessarily
the true value. Always compare the measured value with the observed pin depth
limits (which account for measurement uncertainties) in Table 2-2 on page 2-4 to
evaluate the condition of device connectors.
3- 685054B
Use, Maintenance, and Care of the Devices
Gaging Connectors
When to Gage Connectors
Gage a connector at the following times:
•Prior to using a device for the first time: record the pin depth measurement so that it can be
compared with future readings. (It will serve as a good troubleshooting tool when you
suspect damage may have occurred to the device.)
•If either visual inspection or electrical performance suggests that the connector interface
may be out of typical range (due to wear or damage, for example).
•If a calibration device is used by someone else or on another system or piece of equipment.
•Initially after every 100 connections, and after that as often as experience indicates.
Reading the Connector Gage
The gage dial is divided into increments of 0.0001 inch and major divisions of 0.001 inch (see
Figure 3-2). For each revolution of the large dial, the smaller dial indicates a change of 0.01
inch. Use the small dial as the indicator of multiples of 0.01 inch. In most connector measuring
applications, this value will be zero.
When making a measurement, the gage dial indicator will travel in one of two directions. If the
center conductor is recessed from the zero reference plane, the indicator will move
counterclockwise to indicate the amount of recession, which is read as a negative value. If the
center conductor protrudes, the indicator will move clockwise to indicate the amount of
protrusion, which is read as a positive value. Refer to “Pin Depth” on page 2-3 for definitions
of protrusion and recession.
Figure 3-2 Reading the Connector Gage
85054B 3-7
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Procedures
Gaging Male Type-N Connectors
NOTEAlways hold a connector gage by the gage barrel, below the dial indicator. This
gives the best stability, and improves measurement accuracy.
1. Select the proper gage for your connector. (Refer to Table 6-2 for the gage set part number).
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-3):
a. While holding the gage by the barrel, and without turning the gage or the gage master,
screw the gage master connecting nut onto the male gage, just until you meet
resistance. Connect the nut finger tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
135 N-cm (12 in-lb). Refer to “Connections” on page 3-16 for more information.
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero.
d. Tighten the dial lock screw and remove the gage master.
e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to Figure 3-3):
a. While holding the gage by the barrel, and without turning the gage or the device, screw
the connecting nut of the device being measured onto the gage, just until you meet
resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
135 N-cm (12 in-lb). Refer to “Connections” on page 3-16 for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor is
protruding by an amount indicated by the black
numbers. If the needle has moved
counterclockwise, the center conductor is recessed by an amount indicated by the red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take an
average of the readings. After each measurement, rotate the gage a quarter-turn to
reduce measurement variations that result from the gage or the connector face not being
exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page
2-4.
3- 885054B
Figure 3-3 Gaging Male Type-N Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85054B 3-9
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging Female Type-N Connectors
NOTEAlways hold a connector gage by the gage barrel, below the dial indicator. This
gives the best stability, and improves measurement accuracy.
1. Select the proper gage for your connector. (Refer to Table 6-2 for the gage set part number).
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” and “Cleaning Connectors” earlier in this chapter.
3. Zero the connector gage (refer to Figure 3-4):
a. While holding the gage by the barrel, and without turning the gage or the gage master,
screw the gage connecting nut onto the female gage master, just until you meet
resistance. Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
135 N-cm (12 in-lb). Refer to “Connections” on page 3-16 for more information.
c. Loosen the dial lock screw on the gage and rotate the gage dial so that the pointer
corresponds to the correction value noted on the gage master. Do not adjust the gage
dial to zero, unless the correction value on the gage master is zero.
d. Tighten the dial lock screw and remove the gage master.
e. Attach and torque the gage master to the gage once again to verify that the setting is
repeatable. Remove the gage master.
4. Gage the device connector (refer to Figure 3-3 on page 3-9):
a. While holding the gage by the barrel, and without turning the gage or the device, screw
the gage connecting nut onto the device being measured, just until you meet resistance.
Connect the nut finger-tight. Do not overtighten.
b. Use the torque wrench recommended for use with this kit to tighten the connecting nut to
135 N-cm (12 in-lb). Refer to “Connections” on page 3-16 for more information.
c. Gently tap the barrel of the gage with your finger to settle the gage reading.
d. Read the gage indicator dial. If the needle has moved clockwise, the center conductor is
protruding by an amount indicated by the black
numbers. If the needle has moved
counterclockwise, the center conductor is recessed by an amount indicated by the red
numbers.
For maximum accuracy, measure the connector a minimum of three times and take an
average of the readings. After each measurement, rotate the gage a quarter-turn to
reduce measurement variations that result from the gage or the connector face not being
exactly perpendicular to the center axis.
e. Compare the average reading with the observed pin depth limits in Table 2-2 on page
2-4.
3- 1085054B
Figure 3-4 Gaging Female Type-N Connectors
Use, Maintenance, and Care of the Devices
Gaging Connectors
85054B 3-11
Use, Maintenance, and Care of the Devices
Gaging Connectors
Gaging the Sliding Loads
Gage a sliding load before each use. If the sliding load pin depth is out of the observed pin
depth limits listed in Table 2-2 on page 2-4, refer to “Adjusting the Sliding Load Pin Depth” on
page 3-14.
NOTEAlways hold a connector gage by the gage barrel, below the dial indicator. This
gives the best stability, and improves measurement accuracy. (Cradling the gage
in your hand or holding it by the dial applies stress to the gage plunger
mechanism through the dial indicator housing.)
NOTEThe sliding load uses a plastic centering bead to support its center conductor
when pin depth is adjusted and gaged and when the load is stored. Remove this
support bead from the sliding load before you connect the load for an electrical
calibration. Reinsert this support bead when you’ve finished using the sliding
load.
1. Select the proper gage for your connector. Refer to Table 6-1 on page 6-2 for gage part
numbers.
2. Inspect and clean the gage, gage master, and device to be gaged. Refer to “Visual
Inspection” on page 3-3 and “Cleaning Connectors” on page 3-4 earlier in this chapter.
3. Zero the connector gage as described in either (for a male gage) or (for a female gage).
4. Remove the center conductor protective cap from the sliding load.
5. Loosen the center conductor pull-back nut completely, and press the center-conductor cap
to extend the center conductor beyond the end of the connector. With the sliding ring pulled
back approximately 0.5 inch, install a centering bead (if not already installed) in the
connector end of the sliding load.
6. Continue to press the center conductor cap and mate the center conductor of the sliding
load with the gage’s center conductor.
CAUTIONThe sliding load center conductor can be damaged if the sliding load is not in line
when mating the load to a connector. Always line-up the sliding load when
connecting or removing it from a connector.
7. Mate the outer conductor of the sliding load with the outer conductor of the gage. Torque the
connection with a 3/4 inch torque wrench to approximately 135 N-cm (12 in-lb). Re-tighten
the center conductor pull-back nut. It will “click” when it is tight.
8. Gently tap the barrel of the gage with your finger to settle the gage reading.
9. Read the gage indicator dial. If the needle had moved clockwise, the center conductor is
protruding and the value is determined by the black numbers. If the needle had moved
counterclockwise, the center conductor is recessed by an amount determined by the red
numbers.
3- 1285054B
Use, Maintenance, and Care of the Devices
Gaging Connectors
10.For maximum accuracy, measure the connector a minimum of three times and take an
average of the readings.
NOTEWhen performing pin depth measurements, use different orientations of the gage
within the connector. Averaging a minimum of three readings, each taken after a
quarter-turn rotation of the gage, reduces measurement variations that result
from the gage or the connector face not being exactly perpendicular to the center
axis.
11.Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4. If
the pin depth is outside the limits, it must be adjusted before proceeding. Refer to “Adjusting
the Sliding Load Pin Depth” on page 3-14.
12.Loosen the connection between the gage and sliding load, and remove the sliding load from
the gage.
13.Leave the centering bead installed on the sliding load if you are going to adjust the pin
depth. Carefully remove the centering bead from the sliding load if you’re going to use it for
an electrical calibration. If the centering bead does not come out of the sliding load easily,
loosen the center conductor pull-back nut, and press the center conductor cap to extend the
center conductor. This should expose the centering bead so that it may be removed. Retract
the center conductor and retighten the pull-back nut.
If the centering bead still will not come out, hold the sliding load with the connector end
pointed down. Move the sliding element up, then quickly down. The trapped air behind the
centering bead helps eject it.
CAUTIONDamage can occur to the sliding load during the removal of a centering bead that
has slipped too far into the sliding load. If you’re going to perform an electrical
calibration, prevent damage by removing the centering bead immediately after
aligning the sliding load pin depth. The sliding load will not perform to its
specifications if the centering bead is not removed from the sliding load before an
electrical calibration.
Figure 3-5 Gaging the Sliding Loads
85054B 3-13
Use, Maintenance, and Care of the Devices
Gaging Connectors
Adjusting the Sliding Load Pin Depth
The sliding loads in this kit have a setback mechanism that allows the pin depth to be set to any
desired value. The pin depth of the sliding load is preset at the factory. The pin depth should not have to be reset each time the sliding load is used, but it should be checked before each
use.
If the pin depth is outside the observed limits listed in Table 2-2 on page 2-4, use the following
procedure to reset it. Always measure the sliding load pin depth before attaching it to any
connector.
This procedure assumes that you were directed here from “Gaging the Sliding Loads” on
page 3-12. If not, perform the steps in that procedure before performing this procedure.
1. The gage should be attached to the sliding load. The sliding load should have its centering
bead installed. Refer to “Gaging the Sliding Loads” on page 3-12 if necessary.
2. With a 0.050 inch hex key, loosen the two largest hex screws by turning them 1/4 turn
clockwise. Refer to Figure 3-6.
CAUTIONDo not loosen any hex screws other than the two largest hex screws pointed out
in Figure 3-6.
3. Gently turn the center conductor pin depth adjustment knob on the sliding load until the
gage pointer reads
3.81 micrometers (1.5 x 10
4
inches). Refer to Figure 3-6.
4. Tighten the two hex screws just until they are finger tight (do not overtighten).
5. Wait approximately five minutes to allow the temperature to stabilize. Do not touch either the
gage or the sliding load during this time.
6. Note the gage reading. If it is no longer within the allowable range, perform steps 2
5 again.
7. Loosen the connecting nut and remove the gage from the sliding load. If you’re gong to
store the sliding load, leave the centering bead installed. If you’re going to use the sliding
load for an electrical calibration, remove the centering bead.
NOTEWhen performing pin depth measurements, use different orientations of the gage
within the connector. Averaging a minimum of three readings, each taken after a
quarter-turn rotation of the gage, reduces measurement variations that result
from the gage or the connector face not being exactly perpendicular to the center
axis.
The sliding load pin depth is now in specification and the load is ready to use. Once the sliding
load pin depth is set it rarely needs to be adjusted. However, the pin depth should be
rechecked before each use. Replace the protective plastic caps on the sliding load and gage
connectors when these devices are not in use.
3- 1485054B
Figure 3-6 Adjusting the Sliding Load Pin Depth
Use, Maintenance, and Care of the Devices
Gaging Connectors
85054B 3-15
Use, Maintenance, and Care of the Devices
Connections
Connections
Good connections require a skilled operator. The most common cause of measurement error is
bad connections. The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded,
conductive table mat. Refer to “Electrostatic Discharge” on page 3-2 for ESD precautions.
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4.
4. Use a connector gage to verify that all center conductors are within the observed pin depth
values in Table 2-2 on page 2-4. Refer to “Gaging Connectors” on page 3-6.
5. Carefully align the connectors. The male connector center pin must slip concentrically into
the contact finger of the female connector.
6. Push the connectors straight together.
CAUTIONDo not turn the device body. Only turn the connector nut. Damage to the center
conductor can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is
usually a slight resistance.
7. The preliminary connection is tight enough when the mating plane surfaces make uniform,
light contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating
surfaces is sufficient. Very light finger pressure is enough to accomplish this.
8. Make sure the connectors are properly supported. Relieve any side pressure on the
connection from long or heavy devices or cables.
Final Connection Using a Torque Wrench
Use a torque wrench to make a final connection. Table 3-1 provides information about the
torque wrench recommended for use with this calibration kit. A torque wrench is not included in
the calibration kit. Refer to Chapter 6 for part number and ordering information.
Table 3-1 Torque Wrench Information
Connector TypeTorque SettingTorque Tolerance
Type-N135 N-cm (12 in-lb)13.5 N-cm (1.2 in-lb)
Using a torque wrench guarantees that the connection is not too tight, preventing possible
3- 1685054B
Use, Maintenance, and Care of the Devices
Connections
connector damage. It also guarantees that all connections are equally tight each time.
Prevent the rotation of anything other than the connector nut that you are tightening. It may be
possible to do this by hand if one of the connectors is fixed (as on a test port). In all situations,
however, it is recommended that you use an open-end wrench to keep the body of the device
from turning. Refer to Chapter 6 for part number and ordering information.
1. Position both wrenches within 90 degrees of each other before applying force. See Figure
3-7. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting
action which can misalign and stress the connections of the devices involved. This is
especially true when several devices are connected together.
Figure 3-7 Wrench Positions
2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See
Figure 3-8.
Figure 3-8 Using the Torque Wrench
85054B 3-17
Use, Maintenance, and Care of the Devices
Connections
3. Apply downward force perpendicular to the wrench handle. See Figure 3-8. This applies
torque to the connection through the wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length
rather than pivoting it, otherwise you apply an unknown amount of torque.
4. Tighten the connection just to the torque wrench break point. The wrench handle gives way
at its internal pivot point. See Figure 3-8. Do not tighten the connection further.
CAUTIONYou don’t have to fully break the handle of the torque wrench to reach the
specified torque; doing so can cause the handle to kick back and loosen the
connection. Any give at all in the handle is sufficient torque.
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an
unknown amount of torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do,
you apply more than the recommended torque.
Connecting the Sliding Load
Use this procedure to connect the sliding load to a test port or a type-N cable connector.
NOTEThe sliding load uses a plastic centering bead to support its center conductor
when pin depth is adjusted and gaged and when the load is stored. Remove this
support bead from the sliding load before you connect the load for an electrical
calibration. Reinsert this support bead when you’ve finished using the sliding
load.
CAUTIONCircuitry inside the test set at the test ports may be destroyed if precautions are
not taken to avoid electrostatic discharge (ESD). During this procedure, the
center conductor of the sliding load is connected to the exposed center conductor
of the test port. Ground yourself to prevent electrostatic discharge.
CAUTIONThe sliding load center conductor can be damaged if the sliding load is not in line
when mating the load to a connector. Always line-up the sliding load when
connecting or removing it from a connector.
1. Refer to Figure 3-9. Loosen the center conductor pull-back nut completely. Press the center
conductor cap to extend the center conductor of the sliding load beyond the end of the
connector.
2. Continue to press the center conductor cap and mate the center conductor of the sliding
load with the cable/test port connector’s center conductor.
3. Release pressure on the center conductor and mate the outer conductor of the sliding load
with the outer conductor of the cable/test port connector. Torque the connection with a 3/4
3- 1885054B
Use, Maintenance, and Care of the Devices
Connections
inch torque wrench to approximately 135 N-cm (12 in-lb). Refer to “Final Connection Using a
Torque Wrench” on page 3-16 for additional information.
4. Retighten the center conductor pull-back nut. It will “click” when it is tight.
Figure 3-9 Connecting the Sliding Load
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces, always support the
devices and connections.
CAUTIONTurn the connector nut, not the device body. Major damage to the center
conductor can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use the torque wrench to loosen the connector nut.
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the
connectors.
85054B 3-19
Use, Maintenance, and Care of the Devices
Using the Sliding Load
Using the Sliding Load
When performing a sliding load calibration, it is recommended that the sliding ring be set at the
marked positions (rings) along the sliding load body. Using the set marks ensures that a broad
distribution of phase angles is selected, thereby optimizing the calibration.
The set marks function as detents so that the internal center of the sliding ring can mate with
them. Because of this, the set mark being used cannot be seen but is felt as the sliding ring is
moved from mark to mark during a calibration. Moving the sliding ring with only the index
fingers of both hands will increase your ability to detect the sliding ring detent at each position.
NOTEThe sliding load uses a plastic bead to support its center conductor when it is
stored. Remove this support bead from the sliding load before you connect the
load for use. Reinsert this support bead when you are finished using the sliding
load.
Preferred Method:
1. Move the sliding ring forward as far as possible toward the connector end of the load.
2. Move the sliding ring back until you feel it detent at the first set mark.
3. Follow the network analyzer instructions to measure the sliding load at each detent position.
NOTEAfter a calibration has begun, always move the sliding ring toward the center
conductor pullback end of the sliding load. If you slightly overshoot the desired
mark by less than 0.5 mm (0.02 inch), do not move the sliding ring, but continue
with the calibration as if the sliding ring is set to the proper position. If the sliding
ring is moved toward the connector end of the load during the calibration
sequence, the calibration may be unstable and poor measurements may result. If
the desired position is overshot by more than 0.5 mm (0.02 inch), restart the
calibration sequence from step 1.
It does not matter if you start with the slide at the front and move it backwards, or
start at the back and move it forward. To minimize stability errors, it is important to
start at one end and move it in the same direction for each of the measurement
steps.
This manual contains limited information about a sliding load calibration. For detailed
information on this calibration using a VNA, ENA or PNA series network analyzer, refer to the
appropriate user guide or online Help.
•To view the ENA or PNA online Help, press the Help key on the front panel of the network
analyzer.
•To view an online VNA user guide, use the following steps:
1. Go to www.keysight.com.
2. Enter your VNA model number (Ex: N5242A) in the Search box and click Search.
3- 2085054B
Use, Maintenance, and Care of the Devices
Using the Sliding Load
3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF
you want to view.
If you need additional information, see “Where to Look for More Information” on page 5-3.
Figure 3-10 Using the Sliding Load (Preferred Method)
85054B 3-21
Use, Maintenance, and Care of the Devices
Handling and Storage
Handling and Storage
•Install the protective end caps and store the calibration devices in the foam-lined storage
case when not in use.
•Never store connectors loose in a box, desk, or bench drawer. This is the most common
cause of connector damage during storage.
•Keep connectors clean.
•Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are
easily transferred to a connector interface and are very difficult to remove.
•Do not set connectors contact-end down on a hard surface. The plating and the mating
plane surfaces can be damaged if the interface comes in contact with any hard surface.
3- 2285054B
4 Performance Verification
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Keysight
Technologies for recertification. The equipment required to verify the specifications of the
devices in the kit has been specially manufactured and is not commercially available.
How Keysight Verifies the Devices in This Kit
Keysight verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and
shorts that are directly traced to NIST (National Institute of Standards and Technology). The
airline and short characteristics are developed from mechanical measurements. The
mechanical measurements and material properties are carefully modeled to give very
accurate electrical representation. The mechanical measurements are then traced to NIST
through various plug and ring gages and other mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale)
testing of the calibration devices, Keysight includes the test measurement uncertainty as a
guardband to guarantee each device meets the published specification. For recertifications
(after sale), no guardband is used and the measured data is compared directly with the
specification to determine the pass or fail status. The measurement uncertainty for each
device is, however, recorded in the calibration report that accompanies recertified kits.
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the
Institute's calibration facility. The specifications data provided for the devices in this kit is
traceable to NIST through Keysight Technologies.
4- 285054B
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
•a new calibration sticker affixed to the case
•a certificate of calibration
•a calibration report for each device in the kit listing measured values, specifications,
and uncertainties
NOTEA list of NIST traceable numbers may be purchased upon request to be included
in the calibration report.
Keysight Technologies offers a Standard calibration for the recertification of this kit. For more
information, contact Keysight Technologies. For contact information, see page 5-5.
How Often to Recertify
The suggested initial interval for recertification is 12 months or sooner. The actual need for
recertification depends on the use of the kit. After reviewing the results of the initial
recertification, you may establish a different recertification interval that reflects the usage and
wear of the kit.
NOTEThe recertification interval should begin on the date the kit is first used after the
recertification date.
Where to Send a Kit for Recertification
Contact Keysight Technologies for information on where to send your kit for recertification. For
contact information, refer to page 5-5.
When you return the kit, complete and attach a service tag. Refer to “Returning a Kit or Device
to Keysight” on page 5-4 for details.
85054B 4-3
Performance Verification
Recertification
4- 485054B
5 Troubleshooting
Troubleshooting
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance
verification, follow the steps in Figure 5-1.
Figure 5-1 Troubleshooting Flowchart
5- 285054B
Troubleshooting
Where to Look for More Information
Where to Look for More Information
This manual contains limited information about network analyzer system operation. For
detailed information on using a VNA, ENA or PNA series network analyzer, refer to the
appropriate user guide or online Help.
•To view the ENA or PNA online Help, press the Help key on the front panel of the network
analyzer.
•To view an online VNA user guide, use the following steps:
1. Go to www.keysight.com.
2. Enter your VNA model number (Ex: N5242A) in the Search box and click Search.
3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF
you want to view.
If you need additional information, see “Contacting Keysight” on page 5-5.
85054B 5-3
Troubleshooting
Returning a Kit or Device to Keysight
Returning a Kit or Device to Keysight
If your kit or device requires service, contact Keysight Technologies for information on where to
send it. See “Contacting Keysight” on page 5-5 for contact information. Include a service tag
(located near the end of this manual) on which you provide the following information:
•your company name and address
•a technical contact person within your company, and the person's complete telephone
number
•the model number and serial number of the kit
•the part number and serial number of each device
•the type of service required
•a detailed description of the problem and how the device was being used when the problem
occurred (such as calibration or measurement)
5- 485054B
Troubleshooting
Contacting Keysight
Contacting Keysight
Assistance with test and measurements needs and information on finding a local Keysight
office are available on the Web at:
www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
NOTEIn any correspondence or telephone conversation, refer to the Keysight product
by its model number and full serial number. With this information, the Keysight
representative can determine whether your product is still within its warranty
period.
85054B 5-5
Troubleshooting
Contacting Keysight
5- 685054B
6 Replaceable Parts
Replaceable Parts
Introduction
Introduction
Table 6-1 lists the replacement part numbers for items included in the 85054B calibration kit
and Figure 6-1 illustrates each of these items.
Table 6-2 lists the replacement part numbers for items recommended or required for successful
operation but not included in the calibration kit.
To order a listed part, note the description, the part number, and the quantity desired. Refer to
“Contacting Keysight” on page 5-5.
Table 6-1 Replaceable Parts for the 85054B Calibration Kit
Description Qty Per KitKeysight Part Number
Calibration Devices (50 Type-N)
Type-N -m- sliding load185054-60035
Type-N -f- sliding load185054-60036
Type-N -m- lowband load100909-60011
a
Type-N -f- lowband load100909-60012
Type-N -m- offset short185054-60025
Type-N -f- offset short185054-60026
Type-N -m- offset open185054-60027
Type-N -f- offset open185054-60028
Adapters
Type-N -m- to Type-N -m-185054-60038
Type-N -f- to Type-N -f-185054-60037
Type-N -f- to 7 mm
Type-N -m- to 7 mm285054-60032
Wrenches
3/4 in., 135 N-cm (12 in-lb) Torque
Spanner
Calibration Kit Storage Kit
Storage case assembly (includes items listed below)
285054-60031
18710-1766
108513-20014
1
85054-60039
Storage Box1
Foam pad (bottom)185054-80023
Foam pad (lid)15181-5543
5180-7900
Gages
Type-N gage set (includes items listed below)185054-60049
Type-N gage –f–185054-60050
6- 285054B
Replaceable Parts
Introduction
Table 6-1 Replaceable Parts for the 85054B Calibration Kit
Description Qty Per KitKeysight Part Number
Type-N gage master –f–185054-60052
Type-N gage –m–185054-60051
Type-N gage master –m–185054-60053
Centering beads185054-80028
Miscellaneous Items
User’s and service guide1
Protective End Cap, 0.812-ID, Blackas required1401-0214
Protective End Cap, 0.625-ID, Blackas required1401-0225
a
85054-90049
b
a. Refer to “Clarifying the Terminology of a Connector Interface” on page 1-7.
b. Refer to “Printing Copies of Documentation from the Web” on page -ii
Table 6-2 Replaceable Parts—Items Not Included in the Calibration Kit
Description QtyKeysight Part Number
1/2 in and 9/16 in open-end wrench18710-1770
a
Adapter: Type-N to 7 mm 50 ohm -m- (extendable/retractable
sleeve)
Adapter: Type-N to 7 mm 50 ohm -f- (extendable/retractable
sleeve)
Grounding wrist strap19300-1367
5 ft grounding cord for wrist strap19300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire19300-0797
Anhydrous isopropyl alcohol (92% pure)
Cleaning swabs1009301-1243
b
185054-60009
185054-60001
----
a. Refer to “Clarifying the Terminology of a Connector Interface” on page 1-7.
b. Keysight can no longer safely ship isopropyl alcohol, so customers should
purchase it locally.
85054B 6-3
Replaceable Parts
Introduction
Figure 6-1 Replaceable Parts for the 85054B Calibration Kit
6- 485054B
Figure 6-2 More Replaceable Parts for the 85054B Calibration Kit
Replaceable Parts
Introduction
85054B 6-5
Replaceable Parts
Introduction
6- 685054B
A Standard Definitions
Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web
Class Assignments and Standard Definitions Values are
Available on the Web
Class assignments and standard definitions may change as more accurate model and
calibration methods are developed. You can download the most recent class assignments and
standard definitions from Keysight’s Calibration Kit Definitions Web page at
For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying
Calibration Standards and Kits for Keysight Vector Network Analyzers.” This application note
covers calibration standard definitions, calibration kit content and its structure requirements for
Keysight vector network analyzers. It also provides some examples of how to set up a new
calibration kit and how to modify an existing calibration kit definition file. To download a free
copy, go to www.keysight.com and enter literature number 5989-4840EN in the Search window.
A- 285054B
Index
Numerics
8510 network analyzer
A
adapters
part numbers, 6-2
alcohol
isopropyl
altitude, 2-2
C
cal kit
contents
overview, 1-2
calibration
bad, 5-2
certificate of, 4-2
constants, See calibration definitions
definitions
frequency, 1-8
report, 4-2
standards, 2-5
sticker, 4-3
temperature, 2-2
when to perform, 1-8