THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL
DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR
ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
TROL.
CON
DFARS/Restricted Rights Notice
If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is
delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in
FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or
disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights
as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited
Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Certification
Keysight Technologies certifies that this product met its published specifications at the time of shipment
from the factory. Keysight Technologies further certifies that its calibration measurements are traceable to
the United States National Institute of Standards and Technology, to the extent allowed by the Institute's
calibration facility, and to the calibration facilities of other International Standards Organization members.
85038A/F/M
Assistance
Product maintenance agreements and other customer assistance agreements are available for Keysight
products.
For any assistance, contact Keysight Technologies. Refer to "Contacting Keysight," on page 5-5.
Printing Copies of Documentation from the Web
To print copies of documentation from the Web, download the PDF file from the Keysight web site:
•Go to www.keysight.com.
•Enter the document’s part number (located on the title page) in the Se
Class Assignments and Standard Definitions Values are Available on the Web . . . . . . . . . . . . . . . . . . . . . . . . A-2
Contents-ii85038A/F/M
1 General Information
General Information
Calibration Kit Overview
Calibration Kit Overview
The Keysight 85038 7-16 calibration kits are used to calibrate Keysight network analyzers up to 7.5 GHz for
measurements of components with 7-16 connectors.
Kit Contents
Use the Contents List in the shipping container to verify the completeness of your shipment. Although this
list is the most accurate, you can also use the illustration in Chapter 7 to verify the items in your shipment.
If your shipment is not complete, contact Keysight Technologies - refer to
5-5.
Refer to Ta b le 6- 1 for a list of kit contents part numbers.
Calibration Definitions
"Contacting Keysight," on page
The calibration kit must be selected and the calibration definitions for the devices in the kit installed in the
network analyzer prior to performing a calibration. Refer to your network analyzer user’s guide for
instructions on selecting the calibration kit and performing a calibration.
The calibration definitions can be:
•resident within the analyzer
•entered from the front panel
Class assignments and standard definitions may change as more accurate model and calibration methods
are developed. You can download the most recent class assignments and standard definitions from
Keysight’s Calibration Kit Definitions Web page at
The calibration definitions for the kit may be permanently installed in the internal memory or hard disk of
the network analyzer.
If the calibration definitions for the kit are not permanently installed in the network analyzer, they must be
manually entered. Refer to your network analyzer user’s guide for instructions.
Equipment Required but Not Supplied
Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are not supplied
with the calibration kit but are required to ensure successful operation of the kit. Refer to
6-5 for ordering information.
1- 285038A/F/M
Table 6-2 on page
General Information
Incoming Inspection
Incoming Inspection
Check for damage. The foam-lined storage case provides protection during shipping. Verify that this case
and its contents are not damaged.
If the case or any device appears damaged, or if the shipment is incomplete, contact Keysight
Technologies. For contact information, see
repair or replacement of incomplete or damaged shipments without waiting for a settlement from the
transportation company.
When you send the kit or device to Keysight, include a service tag (found near the end of this manual) with
the following information:
•your company name and address
•the name of a technical contact person within your company, and the person's complete phone number
•the model number and serial number of the kit
•the part number and serial number of the device
“Contacting Keysight” on page 5-5. Keysight will arrange for
•the type of service required
•a detailed description of the problem
85038A/F/M 1-3
General Information
Serial Numbers
Serial Numbers
A serial number is attached to this calibration kit. The first four digits followed by a letter comprise the
serial number prefix; the last five digits are the suffix, unique to each calibration kit.
Recording the Device Serial Numbers
In addition to the kit serial number, the devices in the kit are individually serialized (serial numbers are
labeled onto the body of each device). Record these serial numbers in
numbers will prevent confusing the devices in this kit with similar devices from other kits.
Ta bl e 1-1 Serial Number Record for the 85038A/F/M
DeviceSerial Number
Tab le 1 -1 . Recording the serial
Calibration kit
Open -m-
Open -f-
Short -m-
Short -f-
Load -m-
Load -f-
Adapters (7-16)
-m- to -m-
-f- to -f-
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
_______________________________
1- 485038A/F/M
General Information
Clarifying the Terminology of a Connector Interface
Clarifying the Terminology of a Connector Interface
In this document and in the prompts of the PNA calibration wizard, the sex of cable connectors and
adapters is referred to in terms of the center conductor. For example, a connector or device designated as
1.85 mm –f– has a 1.85 mm female center conductor.
8510-series, 872x, and 875x ONLY: In contrast, during a measurement calibration, the network analyzer
softkey menus label a 1.85 mm calibration device with reference to the sex of the analyzer’s test port
connector—not the calibration device connector. For example, the label SHORT(F) refers to the short that is
to be connected to the female test port. This will be a male short from the calibration kit.
Ta bl e 1-2 Clarifying the Terminology of Connectors: Examples
TerminologyMeaning
Short –f–
Short (f)Male short (male center conductor) to be connected to female port
A connector gage is referred to in terms of the connector that it measures. For instance, a male connector
gage has a female connector on the gage so that it can measure male devices.
Female short (female center conductor)
Preventive Maintenance
The best techniques for maintaining the integrity of the devices in your kit include:
•routine visual inspection
•cleaning
•proper connection techniques
All of these are described in Chapter 3 , “Use, Maintenance, and Care of the Devices.” Failure to detect and
remove dirt or metallic particles on a mating plane surface can degrade repeatability and accuracy and can
damage any connector mated to it. Improper connections, resulting from pin depth values being out of limit
(see
Table 2-2 on page 2-4) or from bad connection techniques, can also damage these devices.
When to Calibrate
A network analyzer calibration remains valid as long as the changes in the systematic error are
insignificant. This means that changes to the uncorrected leakages (directivity and isolation), mismatches
(source match and load match), and frequency response of the system are small (<10%) relative to
85038A/F/M 1-5
General Information
When to Calibrate
accuracy specifications.
Change in the environment (especially temperature) between calibration and measurement is the major
cause in calibration accuracy degradation. The major effect is a change in the physical length of external
and internal cables. Other important causes are dirty and damaged test port connectors and calibration
standards. If the connectors become dirty or damaged, measurement repeatability and accuracy is affected.
Fortunately, it is relatively easy to evaluate the general validity of the calibration. To test repeatability,
remeasure one of the calibration standards. If you can not obtain repeatable measurements from your
calibration standards, maintenance needs to be performed on the test port connectors, cables and
calibration standards. Also, maintain at least one sample of the device under test or some known device as
your reference device. A verification kit may be used for this purpose. After calibration, measure the
reference device and note its responses. Periodically remeasure the device and note any changes in its
corrected response which can be attributed to the test system. With experience you will be able to see
changes in the reference responses that indicate a need to perform the measurement calibration again.
1- 685038A/F/M
2 Specifications
Specifications
Environmental Requirements
Environmental Requirements
Ta bl e 2-1 Environmental Requirements
Parameter Limits
Operating temperature
Error-corrected temperature range
Storage temperature40 C to +75 C (40 F to +167 F)
Relative humidityType tested, 0% to 95% at 40 C, non-condensing
a. The temperature range over which the calibration standards maintain conformance to their specifications.
b. The allowable network analyzer ambient temperature drift during measurement calibration and during
measurements when the network analyzer error correction is turned on. Also, the range over which the
network analyzer maintains its specified performance while correction is turned on.
a
b
+15 C to +35 C (+59 F to +95 F)
1 C of measurement calibration temperature
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating temperature is a
critical factor in performance. During a measurement calibration, the temperature of the calibration devices
must be stable and within the range shown in
IMPORTANTAvoid unnecessary handling of the devices during calibration because your fingers are a
heat source.
Tab le 2 -1 .
2-285038A/F/M
Specifications
Pn42m
MALE PIN DEPTH
OUTER CONDUCTOR
MATING PLANE
FEMALE PIN DEPTH
OUTER CONDUCTOR
MATING PLANE
OUTER
CONDUCTOR
CENTER
CONDUCTOR
FEMALEMALE
Mechanical Characteristics
Mechanical Characteristics
Mechanical characteristics such as pin depth are not performance specifications. They are, however,
important supplemental characteristics related to electrical performance.
Pin Depth
Pin depth is the distance the center conductor mating plane differs from being flush with the outer
conductor mating plane. See
Ta b l e 2 -2 .
The pin depth of a connector can be in one of two states: either protruding or recessed.
Protrusion is the condition in which the center conductor extends beyond the outer conductor mating
plane.
Recession is the condition in which the center conductor is set back from the outer conductor mating
plane.
Figure 2-1. Typical pin depth values for the devices in the kits are listed in
Figure 2-1 Connector Pin Depth
85038A/F/M 2-3
Specifications
Electrical Specifications
Table 2-2 Pin Depth Typical Values
Device Typical Pin Depth
Inches
Millimeters
Male devices-0.068 to -0.069
-1.73 to -1.75
Female devices-0.069 to -0.071
-1.75 to -1.80
Electrical Specifications
The electrical specifications in Ta b l e 2 -3 apply to the devices in your calibration kit when connected with a
Keysight precision interface.
Ta bl e 2-3 Electrical Specifications for 85038A/F/M Devices
DeviceSpecificationFrequency (GHz)
Loads Return loss 4ddc to 4
(male and female)Return loss 36 d> 4 to 7.5
a
Opens
(male and female) 1.25 deviation from nominal phase> 4 to 7.5
a
Shorts
(male and female) 0.85 deviation from nominal phase> 4 to 7.5
Adapters
b
a. The specifications for the opens and shorts are given as allowed deviation from the nominal model as defined in
the standard definitions (see
b. Applies only to 85038F/M kits.
1 deviation from nominal phase
0.6 deviation from nominal phase
Return loss 38 d
Appendix A , “Standard Definitions.”).
dc to 4
dc to 4
dc to 7.5
2-485038A/F/M
Specifications
Electrical Specifications
Certification
Keysight Technologies certifies that this product met its published specifications at the time of shipment
from the factory. Keysight further certifies that its calibration measurements are traceable to the United
States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s
calibration facility, and to the calibration facilities of other International Standards Organization members.
See
"How Keysight Verifies the Devices in Your Kit," on page 4-2 for more information.
Connector Standards
The 7-16 connectors in these calibration kits conform to or exceed the requirements for:
•IEC 169-4 reference grade
•DIN 47223
•EN 122190 grade 0 (European standard)
•BSEN 122190 (British standard)
85038A/F/M 2-5
Specifications
Electrical Specifications
2-685038A/F/M
3 Use, Maintenance, and Care of the Devices
Use, Maintenance, and Care of the Devices
Electrostatic Discharge
Electrostatic Discharge
Protection against ESD (electrostatic discharge) is essential while connecting, inspecting, or cleaning
connectors attached to a static-sensitive circuit (such as those found in test sets).
Static electricity can build up on your body and can easily damage sensitive internal circuit elements when
discharged. Static discharges too small to be felt can cause permanent damage. Devices such as
calibration components and devices under test (DUTs), can also carry an electrostatic charge. To prevent
damage to the test set, components, and devices:
•always wear a grounded wrist strap having a 1 M resistor in series with it when handling components
and devices or when making connections to the test set.
•always use a grounded, conductive table mat while making connections.
•always wear a heel strap when working in an area with a conductive floor. If you are uncertain about
the conductivity of your floor, wear a heel strap.
•always ground yourself before you clean, inspect, or make a connection to a static-sensitive device or
test port. You can, for example, grasp the grounded outer shell of the test port or cable connector briefly.
•always ground the center conductor of a test cable before making a connection to the analyzer test port
or other static-sensitive device. This can be done as follows:
1. Connect a short (from your calibration kit) to one end of the cable to short the center conductor to
the outer conductor.
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector.
3. Connect the other end of the cable to the test port.
4. Remove the short from the cable.
Refer to Chapter 6 , “Replaceable Parts,” for part numbers and instructions for ordering ESD protection
devices.
Figure 3-1 ESD Protection Setup
3-285038A/F/M
Use, Maintenance, and Care of the Devices
Visual Inspection
Visual Inspection
Visual inspection and, if necessary, cleaning should be done every time a connection is made. Metal
particles from the connector threads may fall into the connector when it is disconnected. One connection
made with a dirty or damaged connector can damage both connectors beyond repair.
Magnification is helpful when inspecting connectors, but it is not required and may actually be misleading.
Defects and damage that cannot be seen without magnification generally have no effect on electrical or
mechanical performance. Magnification is of great use in analyzing the nature and cause of damage and in
cleaning connectors, but it is not required for inspection.
Look for Obvious Defects and Damage First
Examine the connectors first for obvious defects and damage: badly worn plating on the connector
interface, deformed threads, or bent, broken, or misaligned center conductors. Connector nuts should move
smoothly and be free of burrs, loose metal particles, and rough spots.
What Causes Connector Wear?
Connector wear is caused by connecting and disconnecting the devices. The more use a connector gets,
the faster it wears and degrades. The wear is greatly accelerated when connectors are not kept clean, or
are connected incorrectly.
Connector wear eventually degrades performance of the device. Calibration devices should have a long life
if their use is on the order of a few times per week. Replace devices with worn connectors.
The test port connectors on the network analyzer test set may have many connections each day, and are
therefore more subject to wear. It is recommended that an adapter be used as a test port saver to minimize
the wear on the test set’s test port connectors.
Inspect the Mating Plane Surfaces
Flat contact between the connectors at all points on their mating plane surfaces is required for a good
connection. See
particles on the connector mating plane surfaces. Also look for signs of damage due to excessive or
uneven wear or misalignment.
Light burnishing of the mating plane surfaces is normal, and is evident as light scratches or shallow
circular marks distributed more or less uniformly over the mating plane surface. Other small defects and
cosmetic imperfections are also normal. None of these affect electrical or mechanical performance.
Figure 2-1 on page 2-3. Look especially for deep scratches or dents, and for dirt and metal
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven
wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the
cause of damage before connecting a new, undamaged connector in the same configuration.
85038A/F/M 3-3
Use, Maintenance, and Care of the Devices
Cleaning Connectors
Inspect Female Connectors
Pay special attention to the contact fingers in the female center conductor. These can be bent or broken,
and damage to them is not always easy to see. A connector with damaged contact fingers will not make
good electrical contact and must be replaced.
NOTEThis is particularly important when mating nonprecision to precision devices.
Cleaning Connectors
Clean connectors are essential for ensuring the integrity of RF and microwave coaxial connections.
1. Use Compressed Air or Nitrogen
WARNINGAlways use protective eyewear when using compressed air or nitrogen.
Use compressed air (or nitrogen) to loosen particles on the connector mating plane surfaces. Clean air
cannot damage a connector or leave particles or residues behind.
You can use any source of clean, dry, low-pressure compressed air or nitrogen that has an effective
oil-vapor filter and liquid condensation trap placed just before the outlet hose.
Ground the hose nozzle to prevent electrostatic discharge, and set the air pressure to less than 414 kPa
(60 psi) to control the velocity of the air stream. High-velocity streams of compressed air can cause
electrostatic effects when directed into a connector. These electrostatic effects can damage the device.
Refer to
2. Clean the Connector Threads
WARNINGKeep isopropyl alcohol away from heat, sparks, and flame. Store in a tightly closed
“Electrostatic Discharge” on page 3-2 for additional information.
container. It is extremely flammable. In case of fire, use alcohol foam, dry chemical, or
carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with eyes, skin, and
clothing. It causes skin irritation, may cause eye damage, and is harmful if swallowed or
inhaled. It may be harmful if absorbed through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local
environmental regulations.
3-485038A/F/M
Use, Maintenance, and Care of the Devices
Connections
Use a lint-free swab or cleaning cloth moistened with isopropyl alcohol to remove any dirt or stubborn
contaminants on a connector that cannot be removed with compressed air or nitrogen. Refer to
6-2 on page 6-5 for part numbers for cleaning swabs.
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the connector threads.
c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
3. Clean the Mating Plane Surfaces
a. Apply a small amount of isopropyl alcohol to a lint-free cleaning swab.
b. Clean the center and outer conductor mating plane surfaces. Refer to Figure 2-1 on page 2-3. When
cleaning a female connector, avoid snagging the swab on the center conductor contact fingers by
using short strokes.
c. Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure
compressed air or nitrogen. Always completely dry a connector before you reassemble or use it.
Ta bl e
4. Reinspect
Inspect the connector again to make sure that no particles or residue are present.
Connections
Good connections require a skilled operator. The most common cause of measurement error is bad
connections. The following procedures illustrate how to make good connections.
How to Make a Connection
Preliminary Connection
1. Ground yourself and all devices. Wear a grounded wrist strap and work on a grounded, conductive table
mat. Refer to
2. Visually inspect the connectors. Refer to “Visual Inspection” on page 3-3.
3. If necessary, clean the connectors. Refer to “Cleaning Connectors” on page 3-4.
4. Carefully align the connectors. The male connector center pin must slip concentrically into the contact
finger of the female connector.
“Electrostatic Discharge” on page 3-2 for ESD precautions.
5. Push the connectors straight together.
CAUTIONDo not turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is twisted.
Do not twist or screw the connectors together. As the center conductors mate, there is usually a slight
85038A/F/M 3-5
Use, Maintenance, and Care of the Devices
Connections
resistance.
6. The preliminary connection is tight enough when the mating plane surfaces make uniform, light
contact. Do not overtighten this connection.
A connection in which the outer conductors make gentle contact at all points on both mating surfaces
is sufficient. Very light finger pressure is enough to accomplish this.
7. Make sure the connectors are properly supported. Relieve any side pressure on the connection from
long or heavy devices or cables.
Final Connection Using a Torque Wrench
1. Use a torque wrench to make a final connection. Tabl e 3 -1 provides information about the torque
wrench recommended for use with these calibration kits. A torque wrench is included in the 85038A
calibration kit but not in the 85038F/M calibration kits. Refer to
information.
Ta bl e 3-1 Torque Wrench Information
Connector TypeTorque SettingTor q u e Tol e ran c e
7-16226 N-cm (20 in-lb)22.6 N-cm (2.0 in-lb)
Chapter 6 for part number and ordering
Using a torque wrench guarantees that the connection is not too tight, preventing possible connector
damage. It also guarantees that all connections are equally tight each time.
2. Prevent the rotation of anything other than the connector nut that you are tightening. It may be possible
to do this by hand if one of the connectors is fixed (as on a test port). In all situations, however, it is
recommended that you use an open-end wrench to keep the body of the device from turning. Refer to
Chapter 6 for part number and ordering information.
3. Position both wrenches within 90 degrees of each other before applying force. See Figure 3-2.
Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can
misalign and stress the connections of the devices involved. This is especially true when several
devices are connected together.
3-685038A/F/M
Use, Maintenance, and Care of the Devices
Connections
Figure 3-2 Wrench Positions
4. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-3.
Figure 3-3 Using the Torque Wrench
5. Apply force perpendicular to the wrench handle. This applies torque to the connection through the
wrench.
Do not hold the wrench so tightly that you push the handle straight down along its length rather than
pivoting it, otherwise you apply an unknown amount of torque.
6. Tighten the connection just to the torque wrench break point. The wrench handle gives way at its
internal pivot point. See
Figure 3-3. Do not tighten the connection further.
CAUTIONYou don’t have to fully break the handle of the torque wrench to reach the specified torque;
doing so can cause the handle to kick back and loosen the connection. Any give at all in the
handle is sufficient torque.
85038A/F/M 3-7
Use, Maintenance, and Care of the Devices
Handling and Storage
Do not pivot the wrench handle on your thumb or other fingers, otherwise you apply an unknown amount of
torque to the connection when the wrench reaches its break point.
Do not twist the head of the wrench relative to the outer conductor mating plane. If you do, you apply more
than the recommended torque.
How to Separate a Connection
To avoid lateral (bending) force on the connector mating plane surfaces, always support the devices and
connections.
CAUTIONDo not turn the device body. Only turn the connector nut. Damage to the center conductor
can occur if the device body is twisted.
1. Use an open-end wrench to prevent the device body from turning.
2. Use another open-end wrench to loosen the connector nut.
3. Complete the separation by hand, turning only the connector nut.
4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors.
Handling and Storage
•Install the protective end caps and store the calibration devices in the foam-lined storage case when
not in use.
•Never store connectors loose in a box, or in a desk or bench drawer. This is the most common cause of
connector damage during storage.
•Keep connectors clean.
•Do not touch mating plane surfaces. Natural skin oils and microscopic particles of dirt are easily
transferred to a connector interface and are very difficult to remove.
•Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces
can be damaged if the interface comes in contact with any hard surface.
3-885038A/F/M
4 Performance Verification
Performance Verification
Introduction
Introduction
The performance of your calibration kit can only be verified by returning the kit to Keysight Technologies for
recertification. The equipment required to verify the specifications of the devices in the kit has been
specially manufactured and is not commercially available.
How Keysight Verifies the Devices in Your Kit
Keysight verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test system are verified with precision airlines and shorts
that are directly traced to the National Institute of Standards and Technology (NIST). The airline and
short characteristics are developed from mechanical measurements. The mechanical measurements
and material properties are carefully modeled to give very accurate electrical representation. The
mechanical measurements are then traced to NIST through various plug and ring gages and other
mechanical measurements.
2. Each calibration device is electrically tested on this system. For the initial (before sale) testing of the
calibration devices, Keysight includes the test measurement uncertainty as a guardband to guarantee
each device meets the published specification. For recertifications (after sale), no guardband is used
and the measured data is compared directly with the specification to determine the pass or fail status.
The measurement uncertainty for each device is, however, recorded in the calibration report that
accompanies recertified kits.
These two steps establish a traceable link to NIST for Keysight to the extent allowed by the institute’s
calibration facility. The specifications data provided for the devices in your kit is traceable to NIST through
Keysight Technologies.
4-285038A/F/M
Performance Verification
Recertification
Recertification
The following will be provided with a recertified kit:
•a new calibration sticker affixed to the case
•a certificate of calibration
•a calibration report for each device in the kit listing measured values, specifications, and uncertainties
NOTEA list of NIST traceable numbers may be purchased upon request to be included in the
calibration report.
Keysight Technologies offers a Standard calibration for the recertification of your kit. For more information,
contact Keysight Technologies. For contact information, see
How Often to Recertify
“Contacting Keysight” on page 5-5.
The suggested initial interval for recertification is 12 months or sooner. The actual need for recertification
depends on the use of the kit. After reviewing the results of the initial recertification, you may establish a
different recertification interval that reflects the usage and wear of the kit.
NOTEThe recertification interval should begin on the date the kit is first used after the
recertification date.
Where to Send a Kit for Recertification
Contact Keysight Technologies for information on where to send your kit for recertification. For contact
information, see “Contacting Keysight” on page 5-5.
When you return the kit, complete and attach a service tag. Refer to "Returning a Kit or Device to
Keysight," on page 5-4 for details.
85038A/F/M 4-3
Performance Verification
Recertification
4-485038A/F/M
5 Troubleshooting
Troubleshooting
Clean Devices
Torque 7-16 Devices
to 20 in-lb, 226 N-cm
Load Correct Cal Constants
Recertify Kit
Recalibrate System
Start
(Bad Cal Suspected)
Bad
Measurement?
Devices
Cleaned?
Proper
Torque?
Correct
Cal Constants
Loaded?
Mating
Surfaces Smooth &
Unmarred?
Replace Device(s)
Everything
is Okay
NO
NO
NO
NO
YES
NO
YES
YES
YES
YES
pn43m
Troubleshooting Process
Troubleshooting Process
If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow
the steps in
Figure 5-1 Troubleshooting Flowchart
Figure 5-1.
5-285038A/F/M
Troubleshooting
Where to Look for More Information
Where to Look for More Information
This manual contains limited information about network analyzer system operation. For detailed
information on using a VNA, ENA or PNA series network analyzer, refer to the appropriate user guide or
online Help.
•To view the ENA or PNA online Help, press the Help key on the front panel of the network analyzer.
•To view an online VNA user guide, use the following steps:
1. Go to www.keysight.com.
2. Enter your VNA model number (Ex: N5242A) in the Search box and click Search.
3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF you want to
view.
If you need additional information, see “Contacting Keysight” on page 5-5.
85038A/F/M 5-3
Troubleshooting
Returning a Kit or Device to Keysight
Returning a Kit or Device to Keysight
If your kit or device requires service, contact Keysight Technologies for information on where to send it. See
"Contacting Keysight," on page 5-5. Include a service tag (located near the end of this manual) on which
you provide the following information:
•your company name and address
•a technical contact person within your company, and the person's complete phone number
•the model number and serial number of the kit
•the part number and serial number of each device
•the type of service required
•a detailed description of the problem and how the device was being used when the problem occurred
(such as calibration or measurement)
5-485038A/F/M
Troubleshooting
Contacting Keysight
Contacting Keysight
Assistance with test and measurements needs and information on finding a local Keysight office are
available on the Web at:
www.keysight.com/find/assist
If you do not have access to the Internet, please contact your Keysight field engineer.
NOTE In any correspondence or telephone conversation, refer to the Keysight product by its
model number and full serial number. With this information, the Keysight representative can
determine whether your product is still within its warranty period.
85038A/F/M 5-5
Troubleshooting
Contacting Keysight
5-685038A/F/M
6 Replaceable Parts
Replaceable Parts
Introduction
Introduction
Ta b l e 6 -1 lists the replacement part numbers for items included in the 85038A/F/M calibration kits and
Figure 6-1 illustrates each of these items.
Table 6-2 on page 6-5 lists the replacement part numbers for items not included in the calibration kits that
are either required or recommended for successful operation of the kits. The 7-16 Adapter Kits can be
purchased from Maury Microwave Corp. Get ordering information at
To order a listed part, note the description, the part number, and the quantity desired. Telephone or send
your order to Keysight Technologies. For contact information, see
www.maurymw.com
“Contacting Keysight” on page 5-5.
6-285038A/F/M
Figure 6-1 Replaceable Parts for the 85038A/F/M Calibration Kits
Replaceable Parts
Introduction
85038A/F/M 6-3
Replaceable Parts
Introduction
Ta bl e 6-1 Replaceable Parts for the 85038A/F/M Calibration Kits
Item
No.
1Female open
2Female short
3Female load
4Male open
5Male short
6Male load
7Male to male
8Female to female
9
User’s and service guide
DescriptionPart of
85038A
Calibration Devices (7-16)
✔✔
✔✔
✔✔
✔✔
✔✔
✔✔
Adapters (7-16)
Purchase from Maury Microwave Corp.
Get ordering information
Calibration Kit Manual
b
✔✔✔
Part of
85038F
a
Part of
85038M
a
at www.maurymw.com
✔
✔
Qty per kitKeysight Part
Number
185038-80002
185038-80004
185038-80006
185038-80003
185038-80005
185038-80007
1-
1-
185038-90010
Wrenches
10Open-end wrench
11Tor q ue w r en c h
Protective End Caps for Connectors
12For male devices
13For female devices
a. See“Clarifying the Terminology of a Connector Interface” on page 1-5.
b. Refer to “Printing Copies of Documentation from the Web” on page -iii.
✔
✔
✔✔
✔✔
18710-2174
18710-2175
as required1401-0418
as required1401-0417
6-485038A/F/M
Ta bl e 6-2 Replaceable Parts—Items Not Included in the Calibration Kits
Description QtyKeysight Part Number
Adapter Kits
Purchase from Maury Microwave Corp.
Get ordering information
7-16 to 7-16 (includes the following three adapters:)1-
7-16-male to 7-16-male1-
7-16-female to 7-16-female1-
7-16-male to 7-16-female1-
7-16 to type-N (includes the following four adapters:)1-
7-16-male to type-N-male1-
7-16-female to type-N-female1-
7-16-male to type-N-female1-
7-16-female to type-N-male1-
at www.maurymw.com
Replaceable Parts
Introduction
7-16 to 7-mm (includes the following two adapters:)1-
7-16-male to 7-mm1-
7-16-female to 7-mm1-
7-16 to 3.5-mm (includes the following four adapters:)1-
7-16-male to 3.5-mm-male1-
7-16-female to 3.5-mm-male1-
7-16-male to 3.5-mm female1-
7-16-female to 3.5-mm female1-
ESD Protection Devices
Grounding wrist strap19300-1367
5 ft grounding cord for wrist strap19300-0980
2 ft by 4 ft conductive table mat with 15 ft grounding wire19300-0797
ESD heel strap19300-1308
Connector Cleaning Supplies
Anhydrous isopropyl alcohol (92% pure)
Foam-tipped cleaning swabs 1009301-1243
a
--
--
a. Keysight can no longer safely ship isopropyl alcohol, so customers should purchase it locally.
85038A/F/M 6-5
Replaceable Parts
Introduction
6-685038A/F/M
A Standard Definitions
Standard Definitions
Class Assignments and Standard Definitions Values are Available on the Web
Class Assignments and Standard Definitions Values are Available on the
Web
Class assignments and standard definitions may change as more accurate model and calibration methods
are developed. You can download the most recent class assignments and standard definitions from
Keysight’s Calibration Kit Definitions Web page at
For a detailed discussion of calibration kits, refer to the Keysight Application Note, “Specifying Calibration
Standards and Kits for Keysight Vector Network Analyzers.” This application note covers calibration
standard definitions, calibration kit content and its structure requirements for Keysight vector network
analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an
existing calibration kit definition file. To download a free copy, go to
number 5989-4840EN in the Search window.
www.keysight.com and enter literature
A-285038A/F/M
Index
Numerics
8510 network analyzer, 1-2
A
adapter kits
part numbers, 6-4
adapters
part numbers, 6-3
agreements
customer assistance, -ii
maintenance, -ii
alcohol
isopropyl
as cleaning solvent, 3-4
precautions for use of, 3-4
altitude
specifications, 2-2
assistance
customer, -ii
C
cal kit
performance
verifying, 4-2
calibration
bad, 5-2
certificate of, 4-3
constants, 1-2
permanently stored, 1-2
frequency, 1-5
MIL-STD 45662, 4-3
report, 4-3
temperature, 2-2
when to perform, 1-5