Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Keysight 53150A/151A/152A
Microwave Frequency
Counter
This guide describes how to use the Keysight 53150A, 53151A, and
53152A Microwave Frequency Counters. The information in this guide
applies to instruments having the number prefix listed below, unless
accompanied by a “Manual Updating Changes” package indicating
otherwise.
SERIAL PREFIX NUMBER: 3735A, US3925, and US4050 (53150A)
3736A, US3926, and US4051 (53151A)
3737A, US3927, and US4052 (53152A)
duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Manual Part Number
53150-90015
Edition
Edition 3, December 1, 2017
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hardware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, perform, display, or disclose commercial
computer software or commercial computer software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight 53150A/151A/152A Assembly Level Service Guide
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Caution, risk of danger (refer to this
manual for specific Warning or Caution
information)
Earth (ground) terminalFrame or chassis (ground) terminal
Alternating current (AC)Direct current (DC)
Caution, risk of electric shock
Keysight 53150A/151A/152A Assembly Level Service Guide3
Detailed Warranty Information
Keysight does not warrant that the operation of Keysight products will be
uninterrupted or error free. If Keysight is unable, within a reasonable time, to
repair or replace any product to a condition as warranted, customer will be
entitled to a refund of the purchase price upon prompt return of the product.
Keysight products may contain remanufactured parts equivalent to new in
performance or may have been subjected to incidental use.
The warranty period begins on the date of delivery or on the date of installation if
installed by Keysight. If customer schedules or delays Keysight installation more
than 30 days after delivery, warranty begins on the 31st day from delivery.
Warranty does not apply to defects resulting from (a) improper or inadequate
maintenance or calibration, (b) software, interfacing, parts or supplies not
supplied by Keysight, (c) unauthorized modification or misuse, (d) operation
outside of the published environmental specifications for the product, or
(e) improper site preparation or maintenance.
TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES ARE
EXCLUSIVE AND NO OTHER WARRANTY OR CONDITION, WHETHER WRITTEN
OR ORAL, IS EXPRESSED OR IMPLIED AND KEYSIGHT SPECIFICALLY DISCLAIMS
ANY IMPLIED WARRANTIES OR CONDITIONS OF MERCHANTABILITY,
SATISFACTORY QUALITY, AND FITNESS FOR A PARTICULAR PURPOSE.
Keysight will be liable for damage to tangible property per incident up to the
greater of $300,000 or the actual amount paid for the product that is the subject
of the claim, and for damages for bodily injury or death, to the extent that all such
damages are determined by a court of competent jurisdiction to have been
directly caused by a defective Keysight product.
TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS WARRANTY
STATEMENT ARE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS
INDICATED ABOVE, IN NO EVENT WILL KEYSIGHT OR ITS SUPPLIERS BE LIABLE
FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL
(INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN
CONTRACT, TORT, OR OTHERWISE.
For consumer transactions in Australia and New Zealand: the warranty terms
contained in this statement, except to the extent lawfully permitted, do not
exclude, restrict or modify and are in addition to the mandatory statutory rights
applicable to the sale of this product to you.
4Keysight 53150A/151A/152A Assembly Level Service Guide
Safety Considerations
WARNING
WARNING
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
INSTRUCTIONS FOR ADJUSTMENTS WHILE COVERS ARE REMOVED AND
FOR SERVICING ARE FOR USE BY SERVICE-TRAINED PERSONNEL ONLY. TO
AVOID DANGEROUS ELECTRIC SHOCK, DO NOT PERFORM SUCH
ADJUSTMENTS OR SERVICING UNLESS QUALIFIED TO DO SO.
ANY INTERRUPTION OF THE PROTECTIVE GROUNDING CONDUCTOR
(INSIDE OR OUTSIDE THE PRODUCT’S CIRCUITRY) OR DISCONNECTING
THE PROTECTIVE EARTH TERMINAL WILL CAUSE A POTENTIAL SHOCK
HAZARD THAT COULD RESUL T IN PERSONAL INJURY. (GROUND ING ONE
CONDUCTOR OF A TWO CONDUCTOR OUTLET IS NOT SUFFICIENT
PROTECTION.)
Keysight 53150A/151A/152A Assembly Level Service Guide5
Whenever it is likely that the protection has been impaired, the instrument must
be made inoperative and be secured against any unintended operation.
If this instrument is to be energized via an autotransformer (for voltage reduction),
make sure the common terminal is connected to the earthed pole terminal
(neutral) of the power source.
Instructions for adjustments while covers are removed and for servicing are for use
by trained personnel only. To avoid dangerous electric shock, do not perform such
adjustments or servicing unless qualified to do so.
For continued protection against fire, replace the line fuse(s) with fuses of the
same current rating and type (for example, normal blow, time delay). Do not use
repaired fuses or short-circuited fuseholders.
Safety Earth Ground
WARNING
An uninterruptible safety earth ground must be maintained from the mains power
source to the product’s ground circuitry.
WHEN MEASURING POWER LINE SIGNALS, BE EXTREMELY CAREFUL AND
ALWAYS USE A STEP-DOWN ISOLATION TRANSFORMER WHICH OUTPUT IS
COMPATIBLE WITH THE INPUT MEASUREMENT CAPABILITIES OF THIS
PRODUCT. THIS PRODUCT’S FRONT AND REAR PANELS ARE TYPCIALLY AT
EARTH GROUND. THUS, NEVER TRY TO MEASURE AC POWER LINE
SIGNALS WITHOUT AN ISOLATION TRANSFORMER.
Acoustic Noise Emissions
LpA<47 dB at operator position, at normal operation, tested per EN 27779. All
data are the results from type test.
Geräuschemission
LpA<47 dB am Arbeits platz, normaler Betrieb, geprüft nach EN 27779. Die
Angagen beruhen auf Ergebnissen von Typenprüfungen.
Electrostatic Discharge Immunity Testing
When the product is tested with 8kV AD, 4kV CD and 4kV ID according to
IEC801-2, a system error may occur that may affect measurement data made
during these disturbances. After these occurrences, the system self-recovers
without user intervention.
6Keysight 53150A/151A/152A Assembly Level Service Guide
Waste Electrical and Electronic Equipment (WEEE) Directive
This instrument complies with the WEEE Directive marking requirement. This
affixed product label indicates that you must not discard this electrical or
electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
Keysight 53150A/151A/152A Assembly Level Service Guide7
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
8Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Assembly Level Service Guide15
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
16Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Microwave Frequency Counter
Assembly Level Service Guide
In This Guide
How to Use This Guide 18
Description of the Microwave Frequency Counter 21
Options 22
Accessories Supplied and Available 23
This guide provides assembly-level service information for the Keysight 53150A,
53151A, and 53152A Counters.
17
In This Guide
NOTE
NOTE
How to Use This Guide
Repair Strategy
This service guide is designed to isolate failures to the assembly level only.
The Keysight 53150A/151A/152A Counter can be returned to Keysight for all
service work, including troubleshooting, and verifying specifications. Contact your
nearest Keysight Sales and Service Office for more details.
ISD (Instrument Service Division) Emergency Response or Express Calibration
Service is available for Keysight customers in the USA. If downtime is critical,
you can receive your repaired Counter via overnight shipment. Call
1-800-403-0801, and ask for Emergency Response or Express Calibration
Service. When your Counter is repaired, it is returned via overnight shipment at
no extra charge.
If you decide to service the Counter yourself, use the troubleshooting procedures
in Chapter 2 (Service) and the disassembly and reassembly procedures in
Chapter 3 (Replacing Assemblies - Disassembly and Reassembly). Then, use the
calibration instructions in Chapter 2 to calibrate the Counter for
peak-performance operation, and finally, perform all of the performance tests in
Chapter 1 to verify that the Counter is operating to the specifications.
If the motherboard, the sampler, or the front-panel assembly are defective, the
instrument must be returned to an Keysight Service Center for repair and
recalibration.
18Keysight 53150A/151A/152A Assembly Level Service Guide
Instrument Identification
The instrument is identified by the serial number on the rear panel. Keysight uses
a two-part serial number with the first part (prefix) identifying a series of
instruments and the second part (suffix) identifying a particular instrument within
a series. Keysight-assigned alpha characters before the prefix identify the country
in which the instrument was manufactured.
Instruments Covered by this Guide
This guide applies directly to Keysight 53150A, 53151A, and 53152A Counters
that have the same serial number prefix(es) shown on the title page. If the serial
number prefix of your Counter differs from that listed on the title page of this
guide, there may be differences between this guide and your instrument.
Instruments having a higher serial prefix are covered (when required) by one or
more manual-change sheets included with this guide. If a required change sheet
is missing, contact your nearest Keysight Sales Office listed at the back of this
guide.
Assembly-Level Service Guide Organization
In This Guide
This Assembly-Level Service Guide consists of a table of contents, a preface, six
chapters, and an index. The page headers identify the chapters and sections of
this manual. The chapter contents are summarized as follows:
that verify the Counter operates properly and meets the Keysight 53150A/151A/
152A specifications given in Chapter 6, “Specifications” in this guide.
Chapter 2, “Service” is divided into seven main sections that provide instructions
for returning the Counter to Keysight for service, calibrating the Counter, and
troubleshooting the assemblies in the Counter.
Chapter 3, “Replacing Assemblies - Disassembly and Reassembly” provides
procedures for replacing defective assemblies and/or modules in the Counter.
Keysight 53150A/151A/152A Assembly Level Service Guide19
In This Guide
How to Order Guides
Chapter 4, “Replaceable Parts” lists the replaceable parts contained in the
Counter, and explains how to order replacement parts for your Counter.
Chapter 5, “Backdating” contains information required to adapt this manual for
older instruments (to be provided when required).
Chapter 6, “Specifications” lists all the specifications and operating
characteristics for the Keysight 53150A/151A/152A Counter.
The part number for this guide is listed on the Certification and Warranty page
(page 4) and on the back cover of this guide.
20Keysight 53150A/151A/152A Assembly Level Service Guide
Description of the Microwave Frequency Counter
The Keysight 53150A, 53151A, and 53152A Microwave Frequency Counters are
capable of measuring frequencies from 10 Hz to 125 MHz on Channel 1 and from
50 MHz to 20 GHz (53150A), 26.5 GHz (53151A), and 46 GHz (53152A) on
Channel 2. These frequency counters are also capable of measuring power on
Channel 2 (in the same frequency ranges). All three Counters have a maximum
frequency resolution of 1 Hz.
The Keysight 53150A/151A/152A provides GPIB and RS-232 serial interfaces and
are suitable for bench-top and ATE operation.
The basic measurement functions of the Keysight 53150A/151A/152A include
Frequency, Relative Frequency, Frequency Offset, and Power (including Power
Offset and Relative Power). All of these features are accessible from the front panel
and over the GPIB and RS-232 interfaces.
The Keysight 53150A/151A/152A includes the following additional measurement
functions and features that are designed specifically for manufacturing and
service applications:
– 1, 2, 5, and 10 MHz external reference capability
– Optional high-stability oven oscillator for high-accuracy needs and lengthened
calibration cycles
– Frequency and power offset capabilities for relative measurements
– SCPI programming capability
In This Guide
– Battery and dc input option for operation in locations where AC power is
unavailable
– Optional soft carrying case for safe transportation and mobile use
Programmable control is performed via an GPIB or an RS-232 serial interface. The
GPIB and RS-232C ports are standard for the Keysight 53150A, 53151A, and
53152A.
Keysight 53150A/151A/152A Assembly Level Service Guide21
In This Guide
NOTE
Options
Hardware
The options available for the Keysight 53150A/151A/152A are listed following this
paragraph. Specifications for the options are listed in Chapter 6, “Specifications”.
Options ordered with the Counter are installed at the factory and are ready for
operation on delivery.
– High Stability Oven Timebase, Option 001
– Battery/DC Power Input, Option 002
– Rack Mount Kit, Option 1CM (not installed)
– Soft Carrying Case, Option 007 (not installed)
Hardware options can be retrofitted at an Keysight Service Center. However, the
cost to retrofit options is considerably higher than the cost of the same options
when purchased with the instrument.
Support
–3-year Return to Keysight for Repair, Option W30
– 3-year Return to Keysight for Calibration, Option W32
– 3-year Return to Keysight for Standards Compliant Calibration,
Option W34
–5-year Return to Keysight for Repair, Option W50
– 5-year Return to Keysight for Calibration, Option W52
– 5-year Return to Keysight for Standards Compliant Calibration,
Option W54
22Keysight 53150A/151A/152A Assembly Level Service Guide
Accessories Supplied and Available
Accessories Supplied
– Power cord, 2.3 meters (part number dependent upon destination country)
– Fuse (Keysight P/N 2110-0007)
Accessories Available
– Accessories for use with Option 002 (Battery/DC Power Input):
– Automotive Power Adapter (Keysight P/N 53150-60214)
– Battery (Keysight P/N 53150-80010)
– Battery Charger (Keysight P/N 53150-60217) 115V
(includes Automotive Power Adapter)
– Battery Charger (Keysight P/N 53150-60218) 230V
(includes Automotive Power Adapter)
– GPIB Cables (Keysight P/N 10833A/B/C/D)
– RS-232 Cable (Keysight P/N 53150-60215)
In This Guide
– Soft Carrying Case (Keysight P/N 53150-80016) — same as Option 007
– Rack Mount Kit (Keysight P/N 53150-67001) — same as Option 1CM
Manuals Supplied
Keysight 53150A/151A/152A Operating Guide
(Keysight P/N 53150-90013)
Keysight 53150A/151A/152A Programming Guide
(Keysight P/N 53150-90014)
Keysight 53150A/151A/152A Assembly-Level Service Guide
(Keysight P/N 53150-90015)
Keysight 53150A/151A/152A Assembly Level Service Guide23
In This Guide
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
24Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Microwave Frequency Counter
Assembly Level Service Guide
1Performance Tests -
Verifying Specifications
Introduction 26
Equipment Required 28
Operational Verification 29
Complete Performance Tests 35
Test 1: External Timebase Input Test 36
Test 2: Channel 1 Frequency Sensitivity 40
Test 3: Channel 2 Frequency Sensitivity 45
Test 4: Channel 2 Power Measurement 49
Performance Test Record (Page 1 of 4) 54
Performance Test Record (Page 2 of 4) 55
Performance Test Record (Page 3 of 4) 56
Performance Test Record (Page 4 of 4) 57
25
1Performance Tests - Verifying Specifications
Introduction
This chapter provides procedures to test the electrical performance of the
Keysight 53150A/151A/152A. These procedures are based on the Counter
specifications in Chapter 6, “Specifications”.
Two types of testing are provided:
– Operational Verification
– Complete Performance Tests
This chapter is organized as follows:
– Introductionpg. 26
– Equipment Requiredpg. 28
– Operational Verificationpg. 29
– Complete Performance Testspg. 35
– Performance Test Recordpg. 54
Operational Verification
Operational Verification is an abbreviated series of tests that you can perform
(instead of performing the Complete Performance Tests) to provide a high degree
of confidence that the instrument is operating properly. Operational Verification is
useful for incoming inspection, routine maintenance, and after instrument repair.
26Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
Complete Performance Tests
The Complete Performance Tests verify the specifications listed in Chapter 6,
“Specifications”. All tests can be performed without opening the instrument.
Recommended Calibration Cycle
The Counter requires periodic verification of operation. Depending on the type of
use, environmental conditions, aging, and measurement accuracy required, the
Counter should be checked using the operational verification procedure at least
once every year after the unit is first placed in operation. A full Calibration and
Performance Test should be performed each time the Counter changes
environment or if an assembly or module has been replaced.
Test Record
The results of the Operational Verification and the Complete Performance Tests
should be recorded on a copy of the Performance Test Record, located at the end
of the Complete Performance Test section in this chapter.
Keysight 53150A/151A/152A Assembly Level Service Guide27
1Performance Tests - Verifying Specifications
Equipment Required
Tabl e 1-1 lists the test equipment and accessories needed to perform the tests.
Tab le 1- 1Recommended Test Equipment and Accessories
Model
Instrument TypeRequired Characteristics
Synthesized Signal Generator
Microwave Synthesized
Signal Generator
Microwave Amplifier45 MHz to 50 GHzKeysight 83051A1P, T
Power Meter
Power Sensor
Power Sensor Cable2.4 mm to 2.4 mm, 5 ft.Keysight 11730A1P, T
50 W Feedthrough Terminator50 W (m to f) BNCKeysight 10100C1OV, P, T
[a] Equivalent equipment can be substituted for all instruments and accessories.
[b] OV = Operational Verification P = Performance Tests C = Calibration T = Troubleshooting
28Keysight 53150A/151A/152A Assembly Level Service Guide
Operational Verification
NOTE
NOTE
NOTE
Operational Verification is an abbreviated series of tests that you can perform
(instead of performing the Complete Performance Tests) to provide a high degree
of confidence that the instrument is operating properly. Operational Verification is
useful for incoming inspection, routine maintenance, and after instrument repair.
If you are not familiar with operating the Counter, you should review Chapter 1,
“Getting Started,” in the Keysight 53150A/151A/152A Operating Guide. However,
the procedures in this chapter are written so that little experience is necessary.
These procedures should be followed in the order in which they appear.
Power-On Self Tests
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Inspect the Counter for damage.
2 Make sure no cables are connected to the Counter’s inputs.
3 Connect the power cord to the Counter and the power source.
Performance Tests - Verifying Specifications1
Since the Counter’s power supply automatically senses the line voltage, there is
no AC input-voltage setting.
4 Turn on the Main ~ Powerswitch on the Counter’s rear panel.
It is normal for the fan in the Counter to run when the Counter is in Standby
mode. Power is supplied to the timebase whenever the
Main ~ Power switch is on to maintain long-term measurement reliability, and
the fan helps to maintain the timebase’s temperature stability.
5 Press and release the Power button on the front panel.
Keysight 53150A/151A/152A Assembly Level Service Guide29
1Performance Tests - Verifying Specifications
NOTE
6 Verify that the front-panel display shows the following:
– All segments of the front-panel display are temporarily activated.
– TESTING is displayed.
– SELF TEST OK is displayed.
– The model number of the Counter is displayed (a four-digit hex number is
also displayed).
– GPIB ADDR nn (nn = a two digit number from 0 to 30) is displayed.
– CH2 NO SIGNAL is displayed.
7 If an error message is displayed, refer to the troubleshooting section in
Chapter 2, “Service”.
8 Mark Pass or Fail in the Performance Test Record on page 54, Test 1.
To ensure that the test results are valid, the Counter and the test equipment
should be powered on for at least 30 minutes prior to beginning the tests. This
allows the internal temperatures of the equipment and the timebase to stabilize.
10 MHz Test
1 Connect a BNC-to-BNC cable from the Reference 10MHz connector on the
Counter’s rear panel to the CHANNEL 1 input connector on the front panel.
2 Press and release the Chan Select key (CH1 is displayed).
3 Verify that the display reads 10,000,000 Hz.
4 Mark Pass or Fail in the Performance Test Record on page 54, Test 2.
30Keysight 53150A/151A/152A Assembly Level Service Guide
Run Self Test
NOTE
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Disconnect all signal cables from the input and Reference 10MHz connectors.
2 Press and release the Shift key, and then press and release the Menu (Reset/
Local) key.
3 Press the up- and/or down-arrow key(s) as many times as necessary until
DO SELF TEST is displayed.
4 Press and release the Enter key.
5 If no errors are detected, SELF TEST OK is briefly displayed when the self test
is completed. If any error messages are displayed, refer to the troubleshooting
section in Chapter 2, “Service”.
6 Mark Pass or Fail in the Performance Test Record on page 54, Test 3.
Performance Tests - Verifying Specifications1
Keysight 53150A/151A/152A Assembly Level Service Guide31
1Performance Tests - Verifying Specifications
NOTE
External Timebase Test
This test verifies the 10 MHz external timebase specifications of the Counter by
verifying that setting REF OSC to EXT allows the Counter to be synchronized to
the external reference signal.
Equipment Required
Keysight 3325A/B Synthesizer (2)
Keysight 10100C 50W Feedthrough Terminator
Keysight 10503 Series Coaxial Cables (BNC m to m) (2)
– To ensure that the test results are valid, the Counter and the test equipment
should be powered on for at least 30 minutes prior to beginning the tests.
This allows the internal temperatures of the equipment to stabilize.
– These tests are appropriate for the Keysight 53150A, 53151A, and 53152A
Counters.
10 MHz External Timebase Input Test
1 Connect an Keysight 10503 series coaxial cable between the output of the first
Keysight 3325A/B Synthesizer and the Reference 10MHz connector on the
Counter’s rear panel (see Figure 1-2).
2 Connect the Keysight 10100C 50W Feedthrough Terminator to the Counter’s
CHANNEL 1 input connector.
3 Connect an Keysight 10503 series coaxial cable between the output of the
second Synthesizer and the Keysight 10100C 50W Feedthrough Terminator on
the Counter’s CHANNEL 1 input connector.
32Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
Keysight
53150A/51A/52A
Counter
Keysight 3325A/B Synthesizer
Keysight 3325A/B Synthesizer
To Reference 10 MHz
Connector
(Rear Panel)
Output
Output
Channel 1
Figure 1-1External Timebase Test Setup
4 Cycle the POWER button to preset the Counter.
5 Press the Shift key on the Counter’s front panel.
Keysight 53150A/151A/152A Assembly Level Service Guide33
6 Press the Menu key on the Counter’s front panel.
7 If necessary, use the up- and down-arrow keys to cycle the display until it
reads REF OSC> INT.
8 Press the right-arrow key once. INT begins to flash off and on.
9 Press the up- or down-arrow key once. INT changes to EXT.
10 Press the Enter key.
11 Press the Chan Select key once to select the CHANNEL 1 input connector.
1Performance Tests - Verifying Specifications
NOTE
12 Set the first Synthesizer (the one connected to the Counter’s rear-panel
Reference 10MHz connector) to output a 10 MHz, 1V rms, sine-wave signal.
13 Set the second Synthesizer (the one connected to the CHANNEL 1 input) to
output a 2 MHz, 100 mV rms, sine-wave signal.
14 Press and release the Counter’s Reset/Local key.
15 Verify that the Counter displays a reading of approximately 2 MHz and that the
Ext Ref annunciator is activated.
16 Mark Pass or Fail for Test 4 in the Performance Test Record on page 54.
This completes the Operational Verification.
– If you intend to perform the Complete Performance Tests next, do not
disconnect the cable from the Counter’s Reference 10 MHz connector. This
connection is used in the Complete Performance Tests.
– Do not turn off the Counter if you intend to perform the Complete
Performance tests. Turning the Counter off and back on resets the reference
oscillator (REF OSC) selection to internal (INT) and the channel selection to
CHANNEL 2.
34Keysight 53150A/151A/152A Assembly Level Service Guide
Complete Performance Tests
NOTE
NOTE
The Complete Performance Tests verify the specifications of the Keysight 53150A/
151A/152A Counter listed in Chapter 6, “Specifications”. All of these tests can be
performed without opening the instrument. Tab le 1-2 lists a summary of the
performance tests.
Record the results of the performance tests in the appropriate place on the
Performance Test Record, which starts on page 54.
To ensure that the test results are valid, the Counter and the test equipment
should be powered on for at least 30 minutes prior to beginning the tests. This
allows the internal temperatures of the equipment and the timebase to stabilize.
Table 1-2Complete Performance Tests
Page NumberTest Description
Page 36Test 1: External Timebase Input Test
Page 36Test 2: Channel 1 Frequency Sensitivity
Page 45Test 3: Channel 2 Frequency Sensitivity
Performance Tests - Verifying Specifications1
Page 49Test 4: Channel 2 Power Measurement
Other Counter measurement functions (e.g., Averaging) are mathemati-cally
derived by the microprocessor from the parameters verified by these
performance tests. If the Counter passes the performance tests, the other
measurement functions are also functioning to specifications.
Keysight 53150A/151A/152A Assembly Level Service Guide35
1Performance Tests - Verifying Specifications
NOTE
Test 1: External Timebase Input Test
This test verifies the 10 MHz external timebase specifications of the Counter by
verifying that setting REF OSC to EXT allows the Counter to be synchronized to
the external reference signal.
Equipment Required
Keysight 3325A/B Synthesizer (2)
Keysight 10100C 50W Feedthrough Terminator
Keysight 10503 Series Coaxial Cables (BNC m to m) (2)
– To ensure that the test results are valid, the Counter and the test equipment
should be powered on for at least 30 minutes prior to beginning the tests.
This allows the internal temperatures of the equipment to stabilize.
– These tests are appropriate for the Keysight 53150A, 53151A, and 53152A
Counters.
1 MHz External Timebase Input Test
1 Connect an Keysight 10503 series coaxial cable between the output of the first
Keysight 3325A/B Synthesizer and the Reference 10MHz connector on the
Counter’s rear panel (see Figure 1-2).
2 Connect the Keysight 10100C 50W Feedthrough Terminator to the Counter’s
CHANNEL 1 input connector.
3 Connect an Keysight 10503 series coaxial cable between the output of the
second Synthesizer and the Keysight 10100C 50W Feedthrough Terminator on
the Counter’s CHANNEL 1 input connector.
36Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
Keysight
53150A/51A/52A
Counter
Keysight 3325A/B Synthesizer
Keysight 3325A/B Synthesizer
To Reference 10 MHz
Connector
(Rear Panel)
Output
Output
Channel 1
Figure 1-2External Timebase Test Setup
4 Cycle the POWER button to preset the Counter.
5 Press the Shift key on the Counter’s front panel.
Keysight 53150A/151A/152A Assembly Level Service Guide37
6 Press the Menu key on the Counter’s front panel.
7 If necessary, use the up- and down-arrow keys to cycle the display until it
reads REF OSC> INT.
8 Press the right-arrow key once. INT begins to flash off and on.
9 Press the up- or down-arrow key once. INT changes to EXT.
10 Press the Enter key.
11 Press the Chan Select key once to select the CHANNEL 1 input connector.
1Performance Tests - Verifying Specifications
12 Set the first Synthesizer (the one connected to the Counter’s rear-panel
Reference 10MHz connector) to output a 1 MHz, 1V rms, sine-wave signal.
13 Set the second Synthesizer (the one connected to the CHANNEL 1 input) to
output a 2 MHz, 100 mV rms, sine-wave signal.
14 Press and release the Counter’s Reset/Local key.
15 Verify that the Counter displays a reading of approximately 2 MHz and that the
Ext Ref annunciator is activated.
16 Mark Pass or Fail for Test 1a in the Performance Test Record on page 54.
2 MHz External Timebase Input Test
1 Change the frequency of the first Synthesizer (connected to the Counter’s
rear-panel Reference 10 MHz connector) to 2 MHz.
2 Press and release the Counter’s Reset/Local key.
3 Verify that the Counter displays a reading of approximately 2 MHz and that the
Ext Ref annunciator is activated.
4 Mark Pass or Fail for Test 1b in the Performance Test Record on page 54.
5 MHz External Timebase Input Test
1 Change the frequency of the first Synthesizer (connected to the Counter’s
rear-panel Reference 10 MHz connector) to 5 MHz.
2 Press and release the Counter’s Reset/Local key.
3 Verify that the Counter displays a reading of approximately 2 MHz and that the
Ext Ref annunciator is activated.
4 Mark Pass or Fail for Test 1c in the Performance Test Record on page 54.
38Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
10 MHz External Timebase Input Test
1 Change the frequency of the first Synthesizer (connected to the Counter’s
rear-panel Reference 10 MHz connector) to 10 MHz.
2 Press and release the Counter’s Reset/Local key.
3 Verify that the Counter displays a reading of approximately 2 MHz and that the
Ext Ref annunciator is activated.
4 Mark Pass or Fail for Test 1d in the Performance Test Record on page 54.
Keysight 53150A/151A/152A Assembly Level Service Guide39
1Performance Tests - Verifying Specifications
NOTE
Test 2: Channel 1 Frequency Sensitivity
This set of tests verifies the frequency-sensitivity specifications of Channel 1 of the
Keysight 53150A/151A/152A Counters.
Equipment Required
Keysight 3325A/B Synthesizer
Keysight 83650B Synthesizer
Keysight 437B Power Meter
Keysight 8487A/D Power Sensor
Keysight 11730A Power Sensor Cable
Keysight 11667C Power Splitter
Keysight 8490D (Opt. 020) 20 dB Attenuator
Keysight 1250-2015 BNC (m) to SMA (f) Adapter
Keysight 11904A 2.92 mm (m) to 2.4 mm (m) Adapter
Keysight 10503 Series Coaxial Cable (BNC m to m)
2 mm (m) to 2 mm (f) Cable (2)
Test 2a: 10 Hz to 20 MHz Frequency Sensitivity
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Connect an Keysight 10503 Series Coaxial Cable between the RF output
connector on the Keysight 3325A/B Synthesizer and Counter’s CHANNEL 1
connector.
40Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
Keysight 3325A/B Synthesizer
Keysight
53150A/51A/52A
Counter
Output
Channel 1
Reference
Output
(Rear Panel)
To Reference 10 MHz
Connector (Rear Panel)
Figure 1-310 Hz to 20 MHz Frequency Sensitivity Test Setup
2 Verify that the Counter is still set to use an external reference signal (refer to
steps 5 through 10 on page 37).
3 Set the output of the Synthesizer to the first frequency specified for Test 2a in
the Performance Test Record on page 54 (10 Hz).
4 Set the power output of the Synthesizer to the first power value specified for
Test 2a in the Performance Test Record on page 54 (–14.9 dBm) (40 mV).
5 Record the frequency value read on the Counter (±1 count) for Test 2a in the
Performance Test Record on page 54.
6 Repeat steps 3, 4, and 5 for each of the remaining frequencies specified for
Test 2a. Don’t forget to change the power level when performing the 1 KHz
test (–19.2 dBm) (25 mV).
7 If all of the frequency values you recorded for Test 2a are correct, mark Pass in
the Performance Test Record for Test 2a. If any of the frequency readings you
recorded are incorrect, mark Fail in the Performance Test Record.
Keysight 53150A/151A/152A Assembly Level Service Guide41
1Performance Tests - Verifying Specifications
NOTE
NOTE
– Do not disconnect the cable from the Counter’s Reference 10 MHz
connector. This connection is used in the following test.
– Do not turn off the Counter. Turning the Counter off and back on resets the
reference oscillator (REF OSC) selection to internal (INT) and the channel
selection to CHANNEL 2.
Test 2b: 50 MHz to 125 MHz Frequency Sensitivity
1 Connect an Keysight 10100C 50 W Feedthrough Terminator to the CHANNEL 1
input connector on the Counter’s front panel (refer to Figure 1-4 for steps 1
through 9).
Keysight 83650B Synthesizer
Reference
Output
(Rear Panel)
Keysight
53150A/51A/52A
Counter
To Reference 10 MHz
(Rear Panel)
(See Note Below)
Keysight 10100C 50
Feedthrough
Channel 1
RF Output
Keysight 8490D 20 dB Attenuator
Keysight 11904A 2.92 mm to 2.4 mm Adapter
Keysight 1250-2015 BNC to SMA Adapter
Keysight 437B Power Meter
Keysight 11667C
Power Splitter
Keysight 8487A/D
Power Sensor
Figure 1-450 MHz to 125 MHz Frequency Sensitivity Test Setup
If the 10 MHz reference output of the 53650B Synthesizer is not sufficient to
drive the 53150A/51A/52A, reverse the test setup so that you drive the 83650B
from the Counter.
input
42Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
2 Connect the Keysight 1250-2015 BNC (m) to SMA (f) Adapter to the
Keysight 10100C Feedthrough Terminator on the Counter’s CHANNEL 1
connector.
3 Connect the Keysight 11904A 2.92 mm (m) to 2.4 mm (m) Adapter to the
Keysight 1250-2015 Adapter.
4 Connect the Keysight 8490D 20 dB Attenuator to the Keysight 11904A
Adapter.
5 Connect one of the output connectors on the Keysight 11667C Power Splitter
to the Keysight 8490D Attenuator
6 Connect the Keysight 11900B 2.4 mm (f) to 2.4 mm (f) Adapter to the RF
output of the Synthesizer.
7 Connect a 2.4 mm (m) to 2.4 mm (m) cable between the Adapter on the
Synthesizer’s RF output connector and the input connector on the
Keysight 11667C Power Splitter.
8 Connect the Keysight 8487A/D Power Sensor to the remaining output
connector on the Keysight 11667C Power Splitter.
9 Connect the Keysight 11730A Power Sensor Cable between the Power Sensor
and the sensor connector on the Keysight 437B Power Meter.
10 Verify that the Counter is still set to use an external reference signal (refer to
steps 5 through 10 on page 37).
11 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for Test 2b in the Performance Test Record on page 54 (50 MHz).
12 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for Test 2b in the Performance Test
Record on page 54 (+.8 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
Keysight 53150A/151A/152A Assembly Level Service Guide43
1Performance Tests - Verifying Specifications
NOTE
13 Record the frequency value read on the Counter (±1 count) for Test 2b in the
Performance Test Record on page 54.
14 Set the output of the Synthesizer to 125 MHz, and repeat steps 11
through 13.
15 If both of the frequency values you recorded for Test 2b are correct, mark Pass
in the Performance Test Record for Test 2b. If either of the frequency readings
you recorded are incorrect, mark Fail in the Performance Test Record.
16 Disconnect the Keysight 10100C Feedthrough Terminator, the
Keysight 1250-2015 Adapter, and the Keysight 11904A Adapter from the
Counter’s CHANNEL 1 connector.
– Do not disconnect the rest of the test setup, as these connections are also
used in the Channel 2 Frequency Sensitivity test.
– Do not turn off the Counter. Turning the Counter off and back on resets the
reference oscillator (REF OSC) selection to internal (INT).
44Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
NOTE
Test 3: Channel 2 Frequency Sensitivity
This set of tests verifies the frequency-sensitivity specifications of Channel 2 of the
Keysight 53150A/151A/152A Counters.
Equipment Required
Keysight 83650B Synthesizer
Keysight 437B Power Meter
Keysight 8487A/D Power Sensor
Keysight 11667C Power Splitter
Keysight 11730A Power Sensor Cable
Keysight 8490D (Opt. 020) 20 dB Attenuator
Keysight 11904D 2.92 mm (m) to 2.4 mm (f) Adapter
Keysight 10503 series Coaxial Cable (BNC m to m)
2.4 mm (m) to 2.4 mm (m) Cable (2)
Test 3a: 50 MHz to 20 GHz Frequency Sensitivity
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Connect the Keysight 11904D 2.92 mm (m) to 2.4 mm (f) Adapter to the
Counter’s CHANNEL 2 input connector (refer to Figure 1-5 for steps 1 through
4).
2 Connect the Keysight 8490D Attenuator to the Keysight 11904D Adapter,
as shown in Figure 1-5 (the other end of the Attenuator should remain
connected to the Keysight 11667C Power Splitter).
3 Verify that the cable (with Adapter) is still connected between the RF output of
the Keysight 83650B Synthesizer and the Power Splitter input connector.
4 Verify that the Power Sensor is still connected to the Power Splitter and that
the Power Sensor Cable is still connected between the Power Sensor and the
sensor connector on the Power Meter.
Keysight 53150A/151A/152A Assembly Level Service Guide45
1Performance Tests - Verifying Specifications
Keysight 11667C Power Splitter
Keysight 8487A/D Power Sensor
Keysight 8490D 20 dB Attenuator
(50 MHz to 40 GHz tests only)
Keysight 83650B Synthesizer
Keysight 437B Power Meter
Keysight z11904D Adapter
Keysight
53150A/51A/52A
Counter
input
Channel 2
RF Output
To Reference
Output
(Rear Panel)
To Reference 10 MHz
(Rear Panel)
Figure 1-550 MHz to 20 GHz Frequency Sensitivity Test Setup
5 Verify that the Counter is still set to use an external reference signal (refer to
steps 5 through 10 on page 37).
6 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for Test 3a in the Performance Test Record on page 55 (50 MHz).
7 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for Test 3a in the Performance Test
Record on page 55 (0 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
8 Record the frequency value read on the Counter (±1 count) for Test 3a in the
Performance Test Record on page 55.
9 Repeat steps 6, through 8 for each of the remaining frequencies specified for
Test 3a. Don’t forget to change the power level when performing the 400 MHz
and 19 GHz tests (53150A, 53151A, and 53152A) and the 16 GHz and 19 GHz
tests (53152A only).
46Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
NOTE
10 If all of the frequency values you recorded for Test 3a are correct, mark Pass in
the Performance Test Record. If any of the frequency readings you recorded
are incorrect, mark Fail in the Performance Test Record.
Test 3b: 22 GHz to 26.5 GHz Frequency Sensitivity
This test is appropriate for the Keysight 53151A and 53152A Counters.
1 Use the same test setup as described for Test 3a on page 45.
2 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for Test 3b in the Performance Test Record on page 55 (22 GHz).
3 Set the power output of the Synthesizer (read on the Power Meter) to
a power value 20 dBm above the value specified for Test 3b in the
Performance Test Record on page 55 (–5 dBm for the Keysight 53151A;
–7 dBm for the Keysight 53152A). (The added 20 dBm compensates for the
20 dB Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for Test 3b in the
Performance Test Record on page 55.
5 Repeat steps 2 through 4 for each of the remaining frequencies specified for
Test 3b.
6 If all of the frequency values you recorded for Test 3b are correct, mark Pass in
the Performance Test Record for Test 3b. If any of the frequency readings you
recorded are incorrect, mark Fail in the Performance Test Record.
Keysight 53150A/151A/152A Assembly Level Service Guide47
1Performance Tests - Verifying Specifications
NOTE
NOTE
Test 3c: 30 GHz to 46 GHz Frequency Sensitivity
This test is appropriate for the 53152A Counter only.
1 Use the same test setup as described for Test 3a on page 45.
2 Set the output of the Keysight 83650B Synthesizer to the first frequency
specified for Test 3c in the Performance Test Record on page 55 (30 GHz).
3 Set the power output of the Synthesizer (read on the Power Meter) to a power
value 20 dBm above the value specified for Test 3c in the Performance Test
Record on page 55 (–3 dBm). (The added 20 dBm compensates for the 20 dB
Attenuator.)
4 Record the frequency value read on the Counter (±1 count) for Test 3c in the
Performance Test Record on page 55.
5 Repeat steps 2, through 4 for each of the remaining frequencies specified for
Test 3c. Don’t forget to change the power level for the 42, 44, and 46 GHz
tests.
6 If all of the frequency values you recorded for Test 3c are correct, mark Pass in
the Performance Test Record for Test 3c. If any of the frequency readings you
recorded are incorrect, mark Fail in the Performance Test Record.
– Do not disconnect the Power Sensor or the Power Sensor Cable from the
power meter or the Power Splitter. These connections are reused in the
following tests.
– Do not turn off the Counter. Turning the Counter off and back on resets the
reference oscillator (REF OSC) selection to internal (INT).
48Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
NOTE
Test 4: Channel 2 Power Measurement
This set of tests verifies the power-measurement specifications of Channel 2 of
the Keysight 53150A/151A/152A Counters.
Equipment Required
Keysight 83650B Synthesizer
Keysight 83051A Microwave Amplifier
Keysight 437B Power Meter
Keysight 8487A/D Power Sensor
Keysight 11730A Power Sensor Cable
Keysight 11667C Power Splitter
Keysight 11904D 2.92 mm (m) to 2.4 mm (f) Adapter
Keysight 11900A 2.4 mm (m) to 2.4 mm (m) Adapter
Keysight 11900B 2.4 mm (f) to 2.4 mm (f) Adapter
2.4 mm (m) to 2.4 mm (m) Cable (2)
Test 4a: 50 MHz to 20 GHz Power Measurement (–20 dBm)
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Verify that the cable (with Adapter) is still connected to the RF output of the
Keysight 83650B Synthesizer.
2 Connect the other end of the 2.4 mm cable to the input connector on the
Keysight 83051A Microwave Amplifier.
3 Connect a 2.4 mm cable from the output of the Microwave Amplifier to the
input connector on the Keysight 11667C Power Splitter.
4 Verify that the Power Sensor is still connected to the Power Splitter and that
the Power Sensor Cable is still connected between the Power Sensor and the
sensor connector on the Power Meter.
Keysight 53150A/151A/152A Assembly Level Service Guide49
1Performance Tests - Verifying Specifications
5 Connect the Keysight 11900A 2.4 mm (m) to 2.4 mm (m) Adapter to the
Counter’s CHANNEL 2 input connector.
6 Connect the unused output of the Microwave Splitter to the Adapter on the
Counter’s CHANNEL 2 input connector.
Keysight 83650B Synthesizer
Keysight
Microwave Amplier
Keysight
53150A/51A/52A
Counter
83051A
Channel 2
Keysight
RF Output
Keysight
11900B Adapter
11667C Power Splitter
Keysight
Keysight
8487A/D Power Sensor
437B Power Meter
Figure 1-650 MHz to 20 GHz (–20 dBm) Power Measurement Test
Setup
7 Verify that the Counter is still set to CHANNEL 2 (if necessary, use the Chan
Select key on the Counter’s front panel to select CHANNEL 2).
8 Adjust the power output of the Synthesizer so that the power output of the
Microwave Amplifier is –20 dBm.
9 Set the Synthesizer to the first frequency value listed for Test 4a in the
Performance Test Record on page 56 (50 MHz).
10 Record the actual power reading in the Performance Test Record.
Input
50Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
NOTE
11 Repeat steps 8 through 10 for each of the remaining frequency values listed for
Test 4a.
12 Mark Pass or Fail for Test 4a in the Performance Test Record.
Test 4b: Channel 2 22 GHz to 26.5 GHz Power Measurement
(–20 dBm)
This test is appropriate for the Keysight 53151A and 53152A Counters.
1 Use the same test setup as described for Test 4a on page 49.
2 With the power output of the Microwave Amplifier still at –20 dBm (adjust the
power output of the Keysight 83650B Synthesizer, if necessary), set the
Synthesizer to the first frequency value listed for Test 4b in the Performance
Test Record on page 56 (22 GHz).
3 Record the actual power reading in the Performance Test Record.
4 Repeat steps 2 and 3 for each of the remaining frequency values listed for Test
4b.
5 Mark Pass or Fail for Test 4b in the Performance Test Record.
Keysight 53150A/151A/152A Assembly Level Service Guide51
1Performance Tests - Verifying Specifications
NOTE
NOTE
Test 4c: Channel 2 26.5 GHz to 46 GHz Power Measurement
(–20 dBm)
This test is appropriate for the Keysight 53152A Counter only.
1 Use the same test setup as described for Test 4a on page 49.
2 With the power output of the Microwave Amplifier still at –20 dBm (adjust the
power output of the Synthesizer, if necessary), set the Synthesizer to the first
frequency value listed for Test 4c in the Performance Test Record on page 56
(30 GHz).
3 Record the actual power reading in the Performance Test Record.
4 Repeat steps 2 and 3 for each of the remaining frequency values listed
for Test 4c.
5 Mark Pass or Fail for Test 4c in the Performance Test Record.
Test 4d: Channel 2 50 MHz to 20 GHz Power Measurement (0 dBm)
This test is appropriate for the Keysight 53150A, 53151A, and 53152A Counters.
1 Use the same test setup as described for Test 4a on page 49.
2 Adjust the power output of the Synthesizer so that the power output of the
Microwave Amplifier is 0 dBm.
3 Set the Synthesizer to the first frequency value listed for Test 4d in the
Performance Test Record on page 57 (50 MHz).
4 Record the actual power reading in the Performance Test Record.
5 Repeat steps 2 through 4 for each of the remaining frequency values listed for
Test 4d.
6 Mark Pass or Fail for Test 4d in the Performance Test Record.
52Keysight 53150A/151A/152A Assembly Level Service Guide
Performance Tests - Verifying Specifications1
NOTE
NOTE
Test 4e: Channel 2 22 GHz to 26.5 GHz Power Measurement (0 dBm)
This test is appropriate for the Keysight 53151A and 53152A Counters.
1 Use the same test setup as described for Test 4a on page 49.
2 With the power output of the Microwave Amplifier still at 0 dBm
(adjust the power output of the Synthesizer, if necessary), set the Synthesizer
to the first frequency value listed for Test 4e in the Performance Test Record
on page 57 (22 GHz).
3 Record the actual power reading in the Performance Test Record.
4 Repeat steps 2 and 3 for the remaining frequency values listed for Test 4e.
5 Mark Pass or Fail for Test 4e in the Performance Test Record.
Test 4f: Channel 2 30 GHz to 46 GHz Power Measurement (0 dBm)
This test is appropriate for the Keysight 53152A Counter only.
1 Use the same test setup as described for Test 4a on page 49.
2 With the power output of the Microwave Amplifier still at 0 dBm
(adjust the power output of the Synthesizer, if necessary), set the Synthesizer
to the first frequency value listed for Test 4f in the Performance Test Record on
page 57 (30 GHz).
3 Record the actual power reading in the Performance Test Record.
4 Repeat steps 2 and 3 for each of the remaining frequency values listed
for Test 4f.
5 Mark Pass or Fail for Test 4f in the Performance Test Record.
Keysight 53150A/151A/152A Assembly Level Service Guide53
1Performance Tests - Verifying Specifications
Performance Test Record (Page 1 of 4)
Keysight 53150A/151A/152A Microwave Frequency Counter
Serial Number: _________________________________________
Test Performed By: _____________________________________
Date: __________________________________________________
Keysight 53150A/151A/152A Assembly Level Service Guide57
1Performance Tests - Verifying Specifications
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
58Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Microwave Frequency Counter
Assembly Level Service Guide
2Service
Introduction 60
Returning the Counter to Keysight Technologies for Service 61
Calibration Procedure 63
Pre-Troubleshooting Information 65
Troubleshooting the Counter 70
59
2Service
Introduction
This chapter provides service information for your Keysight 53150A/151A/152A. It
is divided into four major sections:
– Returning the Counter to Keysight Technologies for Service
(page 61). This section provides you with step-by-step instructions on how to
return the instrument for service.
– Calibration Procedure (page 63). This section provides step-by-step
procedures for calibrating the Keysight 53150A/151A/152A.
– Pre-Troubleshooting Information (page 65). This section provides you with
pertinent information such as safety considerations, recommended test
equipment, repair and after-service considerations, and assembly
identification and location.
– Troubleshooting the Counter (page 70). This section provides you with
troubleshooting procedures that isolate the faulty assembly or module.
Replacement and recalibration of most modules can only be performed at an
authorized Keysight Technologies Service Center.
If the instrument is under warranty, return it to Keysight for service.
Refer to “Returning the Counter to Keysight Technologies for Service” on page 61.
If you decide to troubleshoot the instrument yourself, refer to the section titled
“Troubleshooting the Counter” on page 70.
60Keysight 53150A/151A/152A Assembly Level Service Guide
Service2
NOTE
Returning the Counter to Keysight Technologies for Service
Providing Repair Information
Before shipping the Counter to an Keysight office for service or repair,
call the nearest Keysight Sales Office to make arrangements. Then, tag and
package the Keysight 53150A/151A/152A for shipment.
ISD (Instrument Service Division) Emergency Response or Express Calibration
Service is available for Keysight customers in the USA. If downtime is critical,
you can receive your repaired Counter via overnight shipment. Call
1-800-403-0801, and ask for Emergency Response or Express Calibration
Service. When your Counter is repaired, it is returned via overnight shipment at
no extra charge.
1 Write the following information on a tag:
– Owner’s name and address
– Counter model number
– Complete serial number
– Description of service required or failure indications
2 Attach the tag to the instrument.
3 Pack the instrument.
If the original packaging materials are available, use the procedure titled
“Packing the Counter in the Original Packaging Materials” If the original
packaging materials are not available, you can order new packaging materials
from an Keysight Sales Office. The new packaging materials are identical to
those used by the factory when packaging new instruments. To use
commercially available packaging materials, use the procedure titled “Packing
the Counter in Commercially Available Packaging Materials.” Both procedures
are on the following page.
Keysight 53150A/151A/152A Assembly Level Service Guide61
2Service
Packing the Counter in the Original Packaging Materials
Packing the Counter in Commercially Available Packaging Materials
To prevent shipping damage, it is best to repack the Counter in its original
packaging. In any correspondence, refer to the Counter by the model number and
the complete serial number.
1 Disconnect the power cord, probes, cables, or other accessories attached to
the instrument.
2 Make sure the folded corrugated spacer (which normally contains the
manuals) is in the box to ensure proper fitting.
3 Make sure one of the polystyrene blocks is in its proper position in the box.
4 Place the instrument, front panel end first, on the polystyrene block.
5 Place the other polystyrene block on top of the rear end of the instrument to
secure it.
6 Do not return the manuals with the instrument. Return an accessory only when
it is a part of the failure symptoms.
7 Seal the shipping container securely.
If the original factory packaging materials are not available, use the following
procedure to pack the Counter in commercially available shipping materials.
1 Wrap the instrument in heavy paper or plastic.
2 Place the instrument in a strong shipping container. Use a double-walled
carton made of at least 350 lb. test material.
3 Protect the control panel with cardboard.
4 Add a layer of shock-absorbing material that is 3 to 4 inches (70 to 100 mm)
thick around all sides of the instrument cushion it and to prevent it from
moving within the container
5 Seal the shipping container securely.
6 Mark the shipping container FRAGILE.
62Keysight 53150A/151A/152A Assembly Level Service Guide
Calibration Procedure
NOTE
Calibration
Access Hole
There is only one calibration procedure that can be performed outside of an
Keysight authorized service facility. The following procedure describes how to
calibrate the Counter’s timebase.
The Counter should be powered on for at least 30 minutes prior to beginning the
calibration procedure. This allows the internal temperature of the Counter and
its timebase to stabilize.
Equipment Required
10 MHz Reference Standard
Non-metallic adjustment tool for slotted screw head
Keysight 10503 series Coaxial Cable
1 Connect an Keysight 10503 series coaxial cable from the output of your
10 MHz frequency standard to the Channel 1 input on the Counter’s
front panel.
2 If there is a label covering the calibration adjustment access hole,
remove it (see Figure 2-1).
Service2
Figure 2-1Calibrating the Counter
Keysight 53150A/151A/152A Assembly Level Service Guide63
2Service
3 Insert a non-metallic slotted adjustment tool through the hole in the
lower-right side of the Counter’s cover and into the slot in the timebase
calibration screw.
4 Turn the calibration screw clockwise to increase the Counter’s reading or
counter-clockwise to decrease the reading until the frequency value displayed
on the Counter’s front panel is exactly 10,000,000 Hz.
64Keysight 53150A/151A/152A Assembly Level Service Guide
Pre-Troubleshooting Information
WARNING
This section contains the following pertinent troubleshooting information:
– Safety Considerations
– Recommended Test Equipment
– Repair Considerations
– After Service Considerations
– Assembly Identification and Location
Safety Considerations
Although this instrument has been designed in accordance with international
safety standards, this guide contains information, cautions, and warnings which
must be followed to ensure safe operation and to retain the instrument in a safe
condition. Service instructions, and adjustment procedures requiring removal of
the instrument cover, are for use by service-trained personnel only. To avoid
dangerous electric shock, do not perform any servicing or make any adjustments
with the cover removed, unless qualified to do so.
Service2
Keysight 53150A/151A/152A Assembly Level Service Guide65
BEFORE APPLYING AC POWER, THE INSTRUMENT AND ALL PROTECTIVE
EARTH TERMINALS, EXTENSION CORDS, AUTO TRANSFORMERS, AND
DEVICES CONNECTED TO THE INSTRUMENT SHOULD BE CONNECTED TO A
PROTECTIVE EARTH GROUNDED SOCKET.
ANY INTERRUPTION OF THE PROTECTIVE GROUNDING CONDUCTOR INSIDE
OR OUTSIDE THE INSTRUMENT OR DISCONNECTION OF THE PROTECTIVE
EARTH TERMINAL WILL CAUSE A POTENTIAL SHOCK HAZARD THAT COULD
RESULT IN PERSONAL INJURY. INTENTIONAL INTERRUPTION IS
PROHIBITED.
Any maintenance and repair of the opened instrument under voltage should be
avoided as much as possible and, if necessary, should be carried out only by a
skilled person who is aware of the hazards involved
(for example, fire and electric shock).
2Service
Recommended Test Equipment
Repair Considerations
Test equipment recommended for testing and troubleshooting the
Keysight 53150A/151A/152A is listed in Chapter 1, “Performance Tests - Verifying
Specifications”. Substitute equipment may be used if it meets or exceeds the
required characteristics listed in Tab le 2-1.
Electrostatic Discharge
Electronic components and assemblies in the Keysight 53150A/151A/152A can
be permanently degraded or damaged by electrostatic discharge.
Use the following precautions when servicing the instrument:
1ENSURE that static-sensitive devices or assemblies are serviced at static-safe
work stations providing proper grounding for service personnel.
2ENSURE that static-sensitive devices or assemblies are stored in static
shielding bags or containers.
3DO NOT wear clothing subject to static-charge buildup, such as wool
or synthetic materials.
4DO NOT handle components or assemblies in carpeted areas.
5DO NOT remove an assembly or component from its static shielding protection
until you are ready to install it.
6AVOID touching component leads (handle by packaging only).
Surface Mount Repair
None of the assemblies in the Counter that use surface-mount components can
be repaired outside of authorized Keysight Service Centers. Any attempt to repair
these assemblies voids the Keysight factory warranty and makes the assemblies
ineligible for return credit.
Disassembly and Reassembly Specifics
Refer to Chapter 3 of this guide, “Replacing Assemblies - Disassembly and
Reassembly,” for complete disassembly and reassembly instructions, and
Chapter 4, “Replaceable Parts” for an exploded view of the instrument’s parts.
66Keysight 53150A/151A/152A Assembly Level Service Guide
After Service Considerations
WARNING
Product Safety Checks
The following safety checks must be performed after any troubleshooting and
repair procedures have been completed to ensure the safe operation of the
instrument.
THE RESISTANCE CHECKS DESCRIBED IN THE FOLLOWING TEXT REQUIRE
THAT THE POWER CORD BE CONNECTED TO THE INSTRUMENT AND THAT
AC POWER BE DISCONNECTED. BE SURE THAT THE POWER CORD IS NOT
CONNECTED TO AN AC POWER SOURCE BEFORE PERFORMING ANY
SAFETY CHECKS.
1 VISUAL INSPECTION. Visually inspect the interior of the instrument for any
signs of abnormal internally generated heat, such as discolored printed circuit
boards or components, damaged insulation, or evidence of arcing. Determine
and remedy the cause of any such condition.
2 GROUND CONTINUITY TEST. Plug the power cord into the rear- panel power
receptacle. (DO NOT connect the instrument to AC power at this time.) Using a
suitable ohmmeter, check resistance from the instrument's metallic
connection (such as the rear panel or BNC ground collar) to the ground pin on
the power cord plug. The reading must be less than 1Ω. Flex the power cord
while making this measurement to determine whether intermittent
discontinuities exist.
3 Check any indicated front- or rear-panel ground terminals marked,
using the above procedure.
Service2
Keysight 53150A/151A/152A Assembly Level Service Guide67
4 INSULATION RESISTANCE TEST. Tie the line and neutral pins of the power
cord plug together. Measure the resistance from the instrument enclosure
(chassis) to the line and neutral pins of the power cord plug. The minimum
acceptable resistance is 2 MΩ. Replace any component which results in a
failure.
Product Performance Checks
After replacement of any functional assembly, perform the operational verification
test in Chapter 1 of this guide, “Performance Tests - Verifying Specifications.”
2Service
Assembly Identification and Location
The assembly number, name, and Keysight part number for each of the Keysight
53150A, 53151A, and 53152A assemblies are listed in Table 2-1. Figure 2-2
illustrates the replaceable assemblies in the Keysight 53150A/151A/152A.
Tab le 2- 1Keysight 53150A/151A/152A Assembly Identification
Assembly NameKeysight Part No.
A1 Adapter (53152A only)53152-80001
A2 Cooling Fan Assembly53150-80012
A3 Standby Power Switch Assembly53150-80013
A4 Battery Housing Assembly
A5 DC/DC Converter PCB
A6 Battery
A7 Battery Sled Assembly
A8 Battery Cover Assembly53150-00003
[a]
[a]
[a]
[a]
53150-00002
53150-60004
53150-80010
53150-00004
A9 Rear Panel Assembly—
A10 Power Entry Module Assembly53150-40007
A11 Power Supply Assembly0950-3299
A12 Rear Panel Power Switch Assembly53150-60220
[a] With Option 002 (Battery Option) only.
68Keysight 53150A/151A/152A Assembly Level Service Guide
Service2
A4
A2
A3
A5
A11
A9
A6
A7
A8
A10
A12
A1
Figure 2-2Replaceable Assembly Locations — Top Internal View
Keysight 53150A/151A/152A Assembly Level Service Guide69
2Service
WARNING
NOTE
Neutral
TB1 (AC In)
Line
Pin 8
TB2 (DC Out)
Pin 1
Approx. location of R23
Troubleshooting the Counter
Power Supply Check
HAZARDOUS VOLTAGES ARE PRESENT ON THE POWER SUPPLY ASSEMBLY.
ONLY TRAINED AND QUALIFIED SERVICE PERSONNEL SHOULD PERFORM
THE FOLLOWING PROCEDURE.
1 Remove the power cord from the back of the instrument.
2 Remove the cover as described in Chapter 3.
The DC Power Supply Assembly is fused for safety reasons. It must be assumed
that, if the fuses are blown, the power supply has probably suffered damage and
has therefore become unreliable. If the power supply fails, it must be replaced,
even if the fuse in the power supply appears to be all that has failed.
3 Connect the Counter to the power source.
4 Connect the negative lead of a voltmeter to the chassis and measure each of
the voltages listed in Tab le 2-2 on Power Supply connectors TB1 and TB2 (see
Figure 2-3).
Figure 2-3A4 DC Power Supply Test-Point Locations
70Keysight 53150A/151A/152A Assembly Level Service Guide
Service2
Table 2-2 lists A4 Power Supply's input and output voltages, the corresponding
wire color for each input and output, and the tolerances specified for these
voltages.
Table 2-2A4 Power Supply Inputs and Outputs
Input/Output VoltageConnector and Pin DesignationWire ColorTolerance
AC-input (neutral) 0 VAC
[a]
TB1, pin Nblue—
AC-input (line in)
85 - 264 VAC
–15.0 VDC @ 0.7 A max.TB2, pin 1lavender±3%
0.0 VDC (ground for –15 VDC)TB2, pin 2brown—
N/A (unused)TB2, pin position 3 (no pin)N/A—
N/A (unused)TB2, pin 4
+15.0 VDC @ 2 A max.TB2, pin 5blue±5%
0.0 VDC (ground for +15 VDC)TB2, pin 6brown—
+5.0 VDC @ 5A max.TP 10 (bottom of motherboard)red±1%
0.0 VDC (ground for +5 VDC)TB2, pin 8black—
[a] AC-input values are referenced to chassis ground.
[a]
TB1, pin L
brown—
Adjusting the +5 VDC Output
The +5 VDC output of the power supply is adjustable. If this voltage is found to be
outside of the tolerance range listed in Table 2 -2, it can be adjusted using R23,
which is visible on the top surface of the power supply’s circuit board. When
adjusting the +5 VDC output, use a non-metallic or insulated adjustment tool.
Keysight 53150A/151A/152A Assembly Level Service Guide71
2Service
Self-Test
The Counter performs a Self-Test procedure when it is powered on.
The Self-Test can also be invoked from the Counter’s Menu.
Running the Self Test
1 Disconnect all cables connected to the Counter’s front and rear signal
connectors.
2 Turn on the Main ~ Power switch on the Counter’s rear panel.
3 Press and release the Power button on the front panel.
4 Observe the front-panel display and note any error messages displayed.
5 When CH 2 NO SIGNAL is displayed, press and release the Shift key,
and then press and release the Menu key.
6 Press the up-arrow key or the down-arrow key repeatedly until DO SELF TEST
is displayed, and then press and release the Enter key.
7 Observe the front-panel display and note any error messages displayed.
72Keysight 53150A/151A/152A Assembly Level Service Guide
Service2
Self Test Error Messages
Table 2-3 lists and describes messages that are generated by the Counter during
Self-Test to indicate whether a component passed or failed its test. These
messages are sent via the RS-232 serial output only—they do not appear on the
Counter’s front-panel display.
Table 2-3Self-Test Messages
MessageDescription
ROM TEST FAIL
ROM TEST OK
ROM failed read test.
ROM passed read test.
RAM DATA LINES OK
RAM DATA ERROR
RAM ADDR LINES OK
RAM ADDR ERROR
RAM TEST OK
EEPROM FAIL - CONFIGURATION DATA
ROM FAIL; Computed checkbyte does not match
the value stored in EEPROM.
EEPROM FAIL - CONFIGURATION DATA; Needs to
be (re)initialized.
EEPROM FAIL - POWER CAL DATA; Using default
data
EEPROM FAIL - SAVED SETTINGS; Using default
data
EEPROM FAIL - SAVED SETTINGS; Invalid EEPROM
SAV n Data.
GPIB FAIL; Conf. TestThe GPIB hardware failed its confidence test.
RAM data lines passed test.
RAM data lines failed test.
RAM address lines passed test.
RAM address lines failed test.
RAM tests completed with no errors detected.
The configuration data saved in EEPROM memory
is defective.
The checksum of the ROM data does not match the
value stored in EEPROM.
The EEPROM org code does not verify with current
revision of ROM code.
The checksum of the EEPROM power-calibration
table is bad. Factory default calibration data will be
used.
The checksum of the user settings stored in
EEPROM is bad. Factory default settings will be
used.
The checksum of one set of user settings (1 – 9)
stored in EEPROM is bad.
Keysight 53150A/151A/152A Assembly Level Service Guide73
2Service
Tabl e 2-4 and Table 2-5 list and describe the messages that are generated by the
Counter during Self-Test or during operation to indicate that a problem has been
detected. These messages are displayed on the Counter’s front-panel display and
are also sent via the RS-232 serial output (in many cases, the exact message text
that is displayed on the front panel is a condensed form of the message that is
sent via RS-232).
Tab le 2- 4Front Panel Display Error Messages
MessageDescription
12V FAILThe +12 VDC output from the power supply is not within specifications.
–12V FAILThe –12 VDC output from the power supply is not within specifications.
–5V FAILThe –5 VDC output from the power supply is not within specifications.
ADC FAILFailure detected in the ADC.
PATH FAILFailure detected in the Channel 1 signal path.
THRS FAILFailure detected in the Channel 1 threshold circuit.
THRS FAILFailure detected in the Channel 2 RF threshold circuit.
THRS THRUFailure detected in the Channel 2 through-path threshold circuit.
THRS HETFailure detected in the Channel 2 heterodyne-path threshold circuit.
EEP WRT FAILFailure detected while writing to EEPROM.
FPANEL FAILThe front panel or its interconnecting cable is defective or not properly connected.
FPGA FAILFailure detected in the FPGA (Field Programmable Gate Array).
GPIB FAILFailure detected in the GPIB hardware.
HET PATH FAILFailure detected in the heterodyne-path circuit.
IIC FAILAn attempt to write to the LCD display failed.
INSTCFG FAILThe instrument’s configuration data is missing or has become corrupted.
OVERTEMPThe Counter’s internal temperature is above the acceptable limit.
PWR CAL FAILThe instrument’s power-calibration data is missing or corrupted.
PWR CKT FAILFailure detected in the Channel 2 power-measurement circuit.
RAM FAILFailure detected in RAM.
ROM FAILFailure detected in ROM.
74Keysight 53150A/151A/152A Assembly Level Service Guide
SAV SET FAILOne or more of the sets of user settings is missing or is corrupted.
SERVICE FAILThe instrument’s service data is missing or has become corrupted.
THRU FAILFailure detected in the through-path circuit of Channel 2.
VCO/CNT FAILFailure detected in the VCO or the Count Chain.
Table 2 -5RS-232 Error Messages
MessageDescription
12V FAILThe +12 VDC output from the power supply is not within specifications.
–12V FAILThe –12 VDC output from the power supply is not within specifications.
–5V FAILThe –5 VDC output from the power supply is not within specifications.
ADC FAILFailure detected in the ADC.
B1 SIGNAL PATH FAILFailure detected in the Channel 1 signal path.
B1 THRESHOLD FAILFailure detected in the Channel 1 threshold circuit.
Service2
B2 RF THRESHOLD FAILFailure detected in the Channel 2 RF threshold circuit.
B2 THROUGH-PATH
THRESHOLD FAIL
B2 HETERODYNE PATH
THRESHOLD FAIL
EEPROM FAIL - WRITEFailure detected while writing to EEPROM.
FRONT PANEL FAILThe front panel or its interconnecting cable are defective or not properly connected.
FPGA FAILFailure detected in the FPGA (Field Programmable Gate Array).
GPIB FAILFailure detected in the GPIB hardware.
HETERODYNE PATH FAILFailure detected in the heterodyne-path circuit.
IIC FAILAn attempt to write to the LCD display failed.
INSTCFG FAILThe instrument’s configuration data is missing or corrupted.
OVER TEMPERATUREThe Counter’s internal temperature is above the limit.
Failure detected in the Channel 2 through-path threshold circuit.
Failure detected in the Channel 2 heterodyne-path threshold circuit.
Keysight 53150A/151A/152A Assembly Level Service Guide75
2Service
Tab le 2- 5RS-232 Error Messages (continued)
MessageDescription
PWR CAL FAILThe instrument’s power-calibration data is missing or corrupted.
POWER METER FAILFailure detected in the Channel 2 power-measurement circuit.
RAM FAILFailure detected in RAM.
ROM FAILFailure detected in ROM.
SAV SET FAILOne or more of the sets of user settings is missing or corrupted.
SERVICE FAILThe instrument’s service data is missing or corrupted.
THRU PATH FAILFailure detected in the through-path circuit of Channel 2.
VCO/COUNT CHAIN FAILFailure detected in the VCO or the Count Chain.
76Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Microwave Frequency Counter
Assembly Level Service Guide
3Replacing Assemblies -
Disassembly and
Reassembly
Introduction 78
Tools Required 79
Do This First 79
Removing the Bumpers 80
Removing the Handle and Handle Mounts 81
Removing the Rear Bezel 82
Removing the Shroud 84
Removing the Front Bezel 87
Removing the Rear Panel Assembly 88
Removing the Cooling Fan 91
Replacing the Power Supply 92
77
3Replacing Assemblies - Disassembly and Reassembly
WARNING
CAUTION
NOTE
Introduction
This chapter provides instructions for removing major assemblies in the Keysight
53150A/151A/152A. To install an assembly, apply the removal instructions in
reverse.
The following disassembly and repair procedures are provided:
– Removing the Bumperspg. 80
– Removing the Handle and Handle Mountspg. 81
– Removing the Rear Bezelpg. 82
– Removing the Shroudpg. 84
– Removing the Rear Panel Assemblypg. 88
– Removing the Cooling Fanpg. 91
– Replacing the Power Supplypg. 92
HAZARDOUS VOLTAGES ARE PRESENT ON THE POWER SUPPLY
ASSEMBLY. DISCONNECT THE COUNTER FROM THE POWER SOURCE, AND
WAIT AT LEAST SIX MINUTES BEFORE WORKING INSIDE THE INSTRUMENT.
Do not replace assemblies when the Counter is operating to prevent damage
to components.
Unless otherwise stated, directional referents in the procedures in this chapter
are accurate when viewing the Counter from the front.
Part numbers for all replaceable parts and assemblies are provided in Chapter 4,
“Replaceable Parts”.
78Keysight 53150A/151A/152A Assembly Level Service Guide
Tools Required
CAUTION
The following tools are required for the removal/replacement procedures in this
chapter:
– #1 Phillips screwdriver
– #2 Phillips screwdriver
–TORX
– 1/4-inch open-end or box wrench
– 1/4-inch nut driver
– 5/16-inch nut driver
– Needle-nose pliers
Do This First
Perform the following steps before beginning any of the removal and replacement
procedures:
1 If the Counter is on, press and release the POWER button on the front panel,
2 Disconnect the AC power cord from the AC input socket on the rear panel.
Replacing Assemblies - Disassembly and Reassembly3
®
15 (T15) hand screwdriver
and then set the Main ~ Power switch on the rear panel to “0” (off).
3 If the Counter has the Battery Option (002) installed, remove the battery sled,
and disconnect any DC input cable from the EXT DC connector on the rear
panel.
While most of the hardware used in the Counter is SAE, the four TORX
screws that attach the handle mounts to the Counter are metric. To avoid
damage to the threaded holes in the chassis that handle mount screws fit in, be
sure to use the correct screws.
Keysight 53150A/151A/152A Assembly Level Service Guide79
3Replacing Assemblies - Disassembly and Reassembly
NOTE
Removing the Bumpers
Use the following procedure to remove the front and rear bumpers:
1 To remove the front bumper, lift the center of the top of the bumper upward,
stretch a corner of the bumper diagonally, then slide it off the front of the
Counter.
Figure 3-1Removing the Bumpers
2 Use the equivalent procedure for the rear bumper.
The bumpers can be used to support and protect the Counter after the shroud is
removed. Place both bumpers on their inner edges on your work surface, and
then place the Counter on top of them. The rear bumper can also serve as a
holder for screws and other hardware temporarily removed from the Counter.
OPTIONS
80Keysight 53150A/151A/152A Assembly Level Service Guide
Replacing Assemblies - Disassembly and Reassembly3
TORX Screws
TORX Screws
Handle Mount
Handle Mount
Counter
NOTE
Removing the Handle and Handle Mounts
Use the following procedure to remove the carrying handle and handle mounts:
1 Rotate the handle to the vertical position above the Counter, pull the ends of
the handle outward at the mounts to detach it, and set it aside.
Figure 3-2Removing the Handle
2 Remove the front bumper (see “Removing the Bumpers” on page 80).
3 Use a T15 TORX screwdriver to remove the two screws that attach each of the
handle mounts to the Counter, and remove the mounts.
Figure 3-3Removing the Handle Mounts
When reattaching the handle mounts, limit the torque on the TORX screws to 4
to 5 inch-pounds.
Keysight 53150A/151A/152A Assembly Level Service Guide81
3Replacing Assemblies - Disassembly and Reassembly
NOTE
WARNING:
To avold electric shock,
do not remove covers.
No user-serviceable parts inside.
Refer all servicing to qualified personnel.
OPTIONS
001 Oven Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with domestic and foreign content
11 TO 18 VDC
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232
Main ~ Power
100 – 130 VAC, 50/60/400 Hz 75 VA
220 – 240 VAC, 50/60 Hz 75 VA250 V
FUSE
1.0 A T
AC POWER
ISM 1-A
Removing the Rear Bezel
Use the following procedure to remove the rear bezel:
1 Remove the rear bumper (see “Removing the Bumpers” on page 80).
The rear bumper can be used as a holder for hardware and other small parts
while they are temporarily removed from the Counter.
2 Use a #2 Phillips screwdriver to remove the two 6-32 pan-head machine
screws that attach the rear bezel to the rear panel of the Counter
(see Figure 3-4).
Figure 3-4Rear Bezel Mounting Screws
3 Use a #1 Phillips screwdriver to remove the three 4-40 x 3/16-inch
countersunk shroud screws on the top of the Counter near the back edge
of the shroud (see Figure 3-5).
82Keysight 53150A/151A/152A Assembly Level Service Guide
Replacing Assemblies - Disassembly and Reassembly3
WARNING:
To avold electric shock,
do not remove covers.
No user-serviceable parts inside.
Refer all servicing to qualified personnel.
OPTIONS
001 Ove
n Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with domestic and foreign content
AC POWER
100 – 130 VAC, 50/60/400 Hz
75 VA
220 – 240 VAC, 50/60 Hz
75 VA
250 V
FUSE
11 TO 18 VD
C
E
XT D
C
R
efere
nce
10 M
H
z
In
or
O
ut
A
u
xillary
R
S-232
M
ain ~ P
o
w
er
1.0 A T
ISM 1-A
Shroud Screws
WARNING:
To avold electric shock,
do not remove covers.
No user-serviceable parts inside.
Refer all servicing to qualified personnel.
OPTIONS
001 Oven Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with domestic and foreign c
ontent
AC POWER
100 – 130 VAC, 50/60/400 Hz 75 VA
220 – 240 VAC, 50/60 Hz 7
5 VA
250 V
FUSE
11 TO 18 VDC
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232
Main ~ Power
1.0 A T
ISM 1-A
Shroud Screws
Figure 3-5Top Rear Shroud Screws
4 Turn the Counter upside-down, and remove the three 4-40 x 3/16-inch
countersunk shroud screws on the bottom of the Counter near the back edge
of the shroud (see Figure 3-6).
Figure 3-6Bottom Rear Shroud Screws
5 Carefully work the rear bezel out of the back end of the shroud by alternately
pulling opposite corners of the bezel away from the Counter a small amount at
a time. Do not use tools to pry the bezel out of the shroud, since this could
damage it.
Keysight 53150A/151A/152A Assembly Level Service Guide83
3Replacing Assemblies - Disassembly and Reassembly
WARNING
NOTE
Removing the Shroud
Use the following procedure to remove the shroud:
WHEN THE SHROUD IS REMOVED FROM THE COUNTER, DANGEROUS LINE
VOLTAGES THAT CAN CAUSE SERIOUS INJURY ARE EXPOSED. ALWAYS
DISCONNECT THE POWER CORD BEFORE REMOVING THE SHROUD.
1 Remove the front and rear bumpers (see “Removing the Bumpers” on
page 80).
The bumpers can be used to support and protect the Counter after the shroud is
removed. Place both bumpers on their inner edges on your work surface, and
then place the Counter on top of them. The rear bumper can also serve as a
holder for screws and other hardware temporarily removed from the Counter.
2 Remove the handle and the handle mounts (see “Removing the Handle and
Handle Mounts” on page 81).
3 Remove the rear bezel (see “Removing the Rear Bezel” on page 82).
4 Use a #1 Phillips screwdriver to remove the three 4-40 x 3/16-inch
countersunk shroud screws from the top of the Counter near the front edge of
the shroud (see Figure 3-7).
84Keysight 53150A/151A/152A Assembly Level Service Guide
Replacing Assemblies - Disassembly and Reassembly3
W
ARNING:
To avold electric shock,
do not remove covers.
No user-serviceable parts inside.
Refer all servic
ing to qualified personnel.
OPTIONS
001 Oven Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with domestic and foreign content
AC PO
W
E
R
100 – 130 VA
C
, 50/6
0/400 Hz 75 V
A
220 – 240 VAC
, 50/60 H
z 75 VA
250 V
FUSE
11 TO 18 VD
C
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232
Main ~ Power
1.0 A
T
ISM 1-A
Shroud Screws
WARNING:
To avold ele
ctric sh
ock,
do not remove covers.
No user-serviceable parts insid
e.
Refer all serv
icing to q
ualified personnel.
OPTIONS
001 Oven Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with domestic and foreign content
A
C
PO
W
ER
100 – 130 VA
C
, 50/60/400 H
z 75 V
A
220 – 240 VA
C
, 50/60 H
z 75 VA
250 V
FU
S
E
11 TO 18 VDC
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232
Main ~ Power
1.0 A T
ISM 1-A
Shroud Screws
Figure 3-7Top Front Shroud Screws
5 Turn the Counter upside-down, and remove the three 4-40 x 3/16-inch
countersunk shroud screws on the bottom of the Counter near the front edge
of the shroud (see Figure 3-8).
Keysight 53150A/151A/152A Assembly Level Service Guide85
Figure 3-8Bottom Front Shroud Screws
3Replacing Assemblies - Disassembly and Reassembly
WARNING:
To avold electric shock,
do not rem
ove covers.
No user-serviceable parts inside.
Refer all servicing to qualified personnel.
OPTIONS
001 Oven Time Base
002 Battery
This unit must be earth grounded.
Made in U.S.A.
with dom
estic and foreign c
ontent
AC POWER
100 – 130VAC, 50/60/400 Hz
75VA
220–240 VAC, 50/60 Hz 75VA
250 V
250 V
FUSE
11 TO 18 VDC
EXT DC
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232 RS-232
Main ~ Power
1.0 AT
1.0AT
ISM 1-A
6 With the Counter resting upside-down on your work surface, grip the front
edge of the shroud in the cutout of the front bezel with the fingers of one hand,
and push the Counter’s rear panel toward the front of the shroud with your
other hand. Reduce the amount of pressure you apply to the rear panel when
the shroud begins to slide off the Counter, and carefully remove the shroud the
rest of the way (see Figure 3-9).
Figure 3-9Removing the Shroud
86Keysight 53150A/151A/152A Assembly Level Service Guide
Removing the Front Bezel
Front
Bezel
Screws
Front
Bezel
Screws
CAUTION
Use the following procedure to remove the front bezel:
1 Remove the shroud hardware and slide the shroud back far enough to fully
expose the front bezel, or remove the shroud completely (see “Removing the
Shroud” on page 84).
2 Remove the four 4-40 x 3/16-inch self-locking flathead machine screws that
attach the front bezel to the front panel (two on top and two on
the bottom as shown in Figure 3-10).
Replacing Assemblies - Disassembly and Reassembly3
Figure 3-10Front Bezel Screws
3 Carefully slide the front bezel off the front panel.
When reattaching the front bezel to the front-panel assembly, do not tighten
the screws excessively to prevent damage to the bezel.
Keysight 53150A/151A/152A Assembly Level Service Guide87
3Replacing Assemblies - Disassembly and Reassembly
11 TO 18 VDC
EXT DC
Reference 10 MHz
In
or
Out
Auxillary
RS-232
Main ~ Power
Rear Panel Screws
Rear Panel Screws
Battery
Option
Screws
Removing the Rear Panel Assembly
Use the following procedure to remove the rear panel assembly:
1 Remove the shroud (see “Removing the Shroud” on page 84).
2 If the Battery Option is not installed, remove the four 4-40 self-locking hex
nuts that attach the battery opening cover to the rear panel (these nuts are
accessible from inside the chassis). Then, remove the battery opening cover,
and skip to step 5.
3 If the Battery Option is installed, remove the battery sled from the Counter (for
instructions on removing the battery sled, see Appendix C of the Keysight 53150A/151A/152A Operating Guide).
4 If the Battery Option is installed, remove the four 4-40 x 3/16-inch
self-locking flathead machine screws that attach the rear panel assembly to
the Battery Option chassis as shown in Figure 3-11 (two screws on each side of
the opening, and two more screws at the bottom of the opening).
Figure 3-11Rear Panel Assembly Screws
5 Remove the six 4-40 x 3/16-inch self-locking flathead machine screws that
attach the rear panel assembly to the chassis (see Figure 3-11).
88Keysight 53150A/151A/152A Assembly Level Service Guide
Replacing Assemblies - Disassembly and Reassembly3
NOTE
NOTE
The screw in the lower-left corner of the rear panel assembly (as viewed from the
rear of the Counter) may be covered by the serial-number sticker. If it is,
carefully cut away the lower-left corner of the sticker to expose the screw head.
6 Disconnect the seven-wire cable that runs between the connector at the back
end of the power supply (TB2) and connector PMAIN1 on the main board to
provide better access to the cables that are attached to the back panel
assembly.
7 Disconnect the GPIB ribbon cable from connector JP3 on the main board, and
disengage the ribbon cable from the ribbon-cable clamps that are attached to
the power supply and the inside of the left wall of the chassis. Position this
cable so that it cannot catch on any other cables or other components of the
Counter.
8 Disconnect the external-reference input/output cable (coaxial) from connector
J8 on the main board. J8 is directly under the back edge of the power supply
near the chassis wall.
9 A two-wire AC-input cable (brown and blue wires) runs between the Main ~
Power switch on the rear-panel assembly and the connector that is accessible
through the cutout in the right-front corner of the power supply enclosure
(TB1). Disconnect this cable from the power supply.
When reinstalling the Rear Panel Assembly, be sure to route the two-wire
AC-input cable (brown and blue wires) underneath the power supply.
10 If the Battery Option is installed, there is a two-wire cable (brown and white
wires) that runs between the Main ~ Power switch on the rear panel assembly
and connector J3 on the Battery Option circuit board (J3 is the third connector
from the back edge of the board). Disconnect this cable from J3 on the Battery
Option circuit board.
Keysight 53150A/151A/152A Assembly Level Service Guide89
3Replacing Assemblies - Disassembly and Reassembly
WARNING
CAUTION
11 Use a 5/16-inch spin tight to remove the self-locking hex nut that attaches the
rear panel assembly grounding wire (green/yellow) to the right wall of the
chassis. Slide the grounding lug off the welded stud on the chassis wall, and
thread the lock nut back onto the stud a few turns until it is finger-tight.
AN ELECTRICAL SHOCK HAZARD EXISTS IF THE COUNTER IS OPERATED
WITH THE REAR PANEL ASSEMBLY GROUNDING WIRE NOT PROPERLY
CONNECTED TO THE CHASSIS. TO PREVENT SERIOUS INJURY AND/OR
DAMAGE TO THE EQUIPMENT, NEVER OPERATE THE COUNTER UNLESS
THIS GROUND WIRE IS SECURELY CONNECTED TO THE CHASSIS WALL
AND THE REAR PANEL ASSEMBLY.
12 Extract the ends of all of the cables that you disconnected in Steps 7 through 9
from under the power supply, so that they are completely exposed and cannot
catch on other cables and components when you remove the rear panel
assembly.
13 Carefully pull the rear panel assembly away from the back end of the chassis.
Guide the cables that remain attached to the rear panel assembly as you
remove it to prevent them from catching on other cables or other components
of the Counter.
When reinstalling the Rear Panel Assembly, position it carefully to avoid
damage to the power LED, the serial-port connector (RJ-12), and the main
board. If these components are not correctly aligned with the openings in the
back panel, they may be damaged.
90Keysight 53150A/151A/152A Assembly Level Service Guide
Removing the Cooling Fan
Use the following procedure to remove the cooling fan:
1 Remove the shroud (see “Removing the Shroud” on page 84).
2 Disconnect the two-wire cooling-fan power cable (blue and red wires) from
J16 on the main board. J16 is the connector that is closest to the rear panel
assembly near the right chassis wall and behind the power supply.
3 Open the two cable clamps that are attached to the floor of the chassis
immediately behind the front panel assembly, and remove the cooling-fan
cable from the cable clamps.
4 Extract the end of the cooling-fan cable from under the power supply by
pulling it toward the front of the Counter with one hand while guiding the
connector with your other hand.
5 Use a #1 Phillips screwdriver and a 1/4-inch wrench or spin tight to remove the
four 4-40 x 1-inch panhead machine screws and 1/4-inch
self-locking hex nuts that attach the cooling fan to the chassis (see
Figure 3-12).
Replacing Assemblies - Disassembly and Reassembly3
Figure 3-12Cooling Fan Screws
6 While removing the cooling fan from the chassis, guide the cable to be sure it
does not catch on other cables or other components of the Counter.
Keysight 53150A/151A/152A Assembly Level Service Guide91
3Replacing Assemblies - Disassembly and Reassembly
Replacing the Power Supply
Use the following procedure to replace the power-supply circuit board:
1 Remove the shroud (see “Removing the Shroud” on page 84).
2 Disconnect the two power-supply connectors (one at the front and one at the
rear of the power supply).
3 Use a #1 Phillips screwdriver to remove the two pan-head screws and the two
star washers that fasten the perforated power-supply cover to the
power-supply chassis. The screws are accessible through the rectangular
cutout in the side of the Counter’s chassis.
4 Use a small flat-bladed screwdriver to pry the power-supply cover toward the
opening in the chassis to loosen it from the chassis.
5 Remove the power-supply cover by sliding it through the rectangular opening
in the side of the counter’s chassis.
6 Use a
7 Remove the two flat-head screws that are on both sides of the power-supply
8 Lift the circuit board up until it clears the power-supply cover mounting tabs,
9 Disassemble the replacement power supply in the same manner as described
10 Insert the circuit board from the new power supply into position in the
11 Insert the two screws removed in step 7 into the holes on both sides of the
12 Use a pair of needle-nose pliers to position one of the
¼-inch nut driver to remove the two nuts used to fasten the circuit
board to the power-supply chassis. You may have to hold the screws to prevent
them from turning while removing the nuts. Do not discard the nuts, lock
washers,
nameplate.
and remove it from the chassis. Discard the defective circuit board.
in steps 3 through 8.
power-supply chassis inside the Counter.
power-supply nameplate. Do not tighten these screws at this time.
¼-inch spacers, and 5/8-inch flat-head screws.
¼-inch spacers
removed in step 6 between the circuit board and the power-supply chassis,
and align the spacer with the mounting holes in the circuit board and the
chassis.
92Keysight 53150A/151A/152A Assembly Level Service Guide
Replacing Assemblies - Disassembly and Reassembly3
13 Insert one of the 5/8-inch, flat-head screws removed in step 6 up through the
circuit-board mounting hole in the bottom of the power-supply chassis,
through the spacer, and through the hole in the circuit board.
14 Place a lockwasher over the screw on top of the circuit board.
15 Thread a nut onto the screw, but do not tighten it fully at this time.
16 Repeat steps 12 through 15 for the other circuit-board mounting hole.
17 Tighten the two screws next to the power-supply nameplate.
18 Tighten the nuts on the circuit-board mounting screws.
19 Insert the power-supply cover through the rectangular hole in the side of the
counter’s chassis, and position in properly on the power-supply.
20 Insert the two screws that fasten the power-supply cover to the chassis and
tighten them.
21 Reconnect the two power-supply connectors and the front and rear ends of
the power supply.
22 Turn the Counter on, and perform the “Power Supply Check” procedure on
page 70.
23 If necessary, adjust the +5 VDC output of the power supply using the
procedure titled “Adjusting the +5 VDC Output” on page 71.
24 Reassemble the Counter.
Keysight 53150A/151A/152A Assembly Level Service Guide93
3Replacing Assemblies - Disassembly and Reassembly
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
94Keysight 53150A/151A/152A Assembly Level Service Guide
Keysight 53150A/151A/152A Microwave Frequency Counter
Assembly Level Service Guide
4Replaceable Parts
Introduction 96
Replaceable Parts 97
How To Order A Part 98
Parts Identification 100
95
4Replaceable Parts
Introduction
This chapter contains parts-ordering information for the Keysight 53150A/151A/
152A Counter. Table 4-1 lists the reference designations used in the parts list and
throughout this guide, Table 4 -2 lists all replaceable assemblies and parts.
Figure 4-1 is an exploded view of the Keysight 53150A/151A/152A’s internal
parts, Figure 4-2 is an exploded view of the cabinet and other external parts, and
Figure 4-3 is a top view of the chassis with the shroud removed. These illustrations
are provided to help you locate and identify parts.
96Keysight 53150A/151A/152A Assembly Level Service Guide
Replaceable Parts
Table 4-2 contains a list of replaceable parts that is organized as follows:
1 Electrical assemblies in alphanumeric order by reference designation.
2 Chassis-mounted electrical parts in alphanumeric order by reference
designation.
3 Chassis-mounted mechanical parts in alphanumeric order by reference
designation.
The information given for each part consists of the following:
1 Reference designation.
2 Keysight part number.
3 The quantity (QTY) of the part used in the instrument (field-replaceable items
only).
4 Description of the part.
5 Typical manufacturer's part number for the part (when available).
Replaceable Parts4
Keysight 53150A/151A/152A Assembly Level Service Guide97
4Replaceable Parts
How To Order A Part
Keysight Technologies wants to keep your parts ordering process as simple and
efficient as possible. To order parts, perform the following steps:
1 Identify the part and the quantity you need.
2 Determine the ordering method to be used, and contact Keysight.
If the part you need is not identified in the guide, you can call Keysight for
assistance (see the following section titled Contacting Keysight Technologies).
Please have the following information available when you contact Keysight:
– Instrument Model Number (example, “Keysight 53151A”).
– Complete instrument Serial Number (example, “1234A56789”). Information
about where to find the serial number is provided in the Preface portion of
this guide in the section titled How to Use This Guide.
– Description of the part(s) and its use.
– Quantity of the part(s) required.
98Keysight 53150A/151A/152A Assembly Level Service Guide
Contacting Keysight Technologies
Depending on where you are in the world, there are one or more ways to obtain
parts or parts information from Keysight Technologies.
– Outside the United States, contact your local Keysight Technologies sales
office.
– Within the United States, we encourage you to order replacement parts or
request parts information directly by telephone or mail from Keysight using the
telephone numbers or address listed on the next page. (You can also contact
your local Keysight sales office. Keysight sales offices are listed in “Sales and
Technical Support” on page 7.)
By Telephone:
– For Parts Ordering, use our toll-free number, (800) 227-8164,
Monday through Friday (except Holidays), 6 a.m. to 5 p.m.
(Pacific Time).
– For Parts Identification Assistance, call us at (916) 783-0804.
Our Parts Identification hours are from 6 a.m. to 3 p.m. (Pacific Time) on
Monday through Thursday and from 6 a.m. to 2 p.m. (Pacific Time) on Friday.
For Mail Correspondence, please use the address below:
Keysight Technologies
P.O. Box 1145
Roseville, CA 95661-1145
Replaceable Parts4
Keysight 53150A/151A/152A Assembly Level Service Guide99
4Replaceable Parts
Parts Identification
To identify the part(s) you need, refer to the exploded views in Figure 4-1 and
Figure 4-2 and the top view of the chassis in Figure 4-3.
When ordering from Keysight Technologies, the important numbers to note from
the Parts List are the Keysight Part Number, the part-number check digit (in the
“CD” column), and the quantity of the part(s) you need.
Reference Designations
Tabl e 4-1 lists the reference designations used in the parts lists.
Tab le 4- 1Reference Designations
A
=
H
=
J
=
MP
=
P
=
assembly
hard ware
electrical connector
(stationary portion); jack
miscellaneous mechanical part
electrical connector
(movable portion); plug
P/0
S
T
TP
U
W
=
part of
=
switch
=
transformer
=
test point
=
integrated circuit; microcircuit
=
cable; transmission path; wire
Cabinet Parts and Hardware
To locate and identify cabinet and external parts and mounting hardware, refer to
Figure 4-2 (internal parts and mounting hardware are shown in Figure 4-1). These
figures provide different views of the instrument, with the parts identified by
reference designations; the reference designations correspond with the ones in
Tabl e 4-2 .
Accessories and Miscellaneous Items
Tabl e 4-3 lists all of the accessories available for the Counter and also
lists some other miscellaneous items, such as power cords, rack mounts, and
manuals.
100Keysight 53150A/151A/152A Assembly Level Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.