duced in any form or by any means
(including electronic storage and
retrieval or translation into a foreign
language) without prior agreement and
written consent from Keysight Technologies as governed by United States and
international copyright laws.
Trademarks
Microsoft® and Windows® are U.S.
registered trademarks of Microsoft Corporation. All other brand and product
names are trademarks or registered
trademarks of their respective companies.
Manual Part Number
34420-90010
Edition
Edition 3, September 2018
Printed in:
Printed in Malaysia
Published by:
Keysight Technologies
Bayan Lepas Free Industrial Zone,
11900 Penang, Malaysia
Technology Licenses
The hard ware and/or software
described in this document are furnished under a license and may be
used or copied only in accordance with
the terms of such license.
Declaration of Conformity
Declarations of Conformity for this
product and for other Keysight products may be downloaded from the
Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Declaration of Conformity.
U.S. Government Rights
The Software is “commercial computer
software,” as defined by Federal Acquisition Regulation (“FAR”) 2.101. Pursuant to FAR 12.212 and 27.405-3 and
Department of Defense FAR Supplement (“DFARS”) 227.7202, the U.S.
government acquires commercial computer software under the same terms
by which the software is customarily
provided to the public. Accordingly,
Keysight provides the Software to U.S.
government customers under its standard commercial license, which is
embodied in its End User License
Agreement (EULA), a copy of which can
be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority
by which the U.S. government may use,
modify, distribute, or disclose the Software. The EULA and the license set
forth therein, does not require or permit, among other things, that Keysight:
(1) Furnish technical information
related to commercial computer software or commercial computer software
documentation that is not customarily
provided to the public; or (2) Relinquish
to, or otherwise provide, the government rights in excess of these rights
customarily provided to the public to
use, modify, reproduce, release, perform, display, or disclose commercial
computer software or commercial computer software documentation. No
additional government requirements
beyond those set forth in the EULA
shall apply, except to the extent that
those terms, rights, or licenses are
explicitly required from all providers of
commercial computer software pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in the EULA. Keysight shall be
under no obligation to update, revise or
otherwise modify the Software. With
respect to any technical data as
defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government
acquires no greater than Limited Rights
as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any
technical data.
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER
EXPRESS OR IMPLIED, WITH REGARD
TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE. KEYSIGHT
SHALL NOT BE LIABLE FOR ERRORS
OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT
OR OF ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT
AND THE USER HAVE A SEPARATE
WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT
CONFLICT WITH THESE TERMS, THE
WARRANTY TERMS IN THE SEPARATE
AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
2Keysight 34420A Service Guide
Safety Symbols
The following symbols on the instrument and in the documentation indicate
precautions which must be taken to maintain safe operation of the instrument.
Earth (ground) terminalFrame or chassis (ground) terminal
Keysight 34420A Service Guide3
Safety Considerations
WARNING
WARNING
NOTE
Read the information below before using this instrument.
The following general safety precautions must be observed during all phases of
operation, service, and repair of this instrument. Failure to comply with these
precautions or with specific warnings elsewhere in this manual violates safety
standards for design, manufacture, and intended use of the instrument. Keysight
Technologies assumes no liability for the customer’s failure to comply with these
requirements.
Only qualified, service-trained personnel who are aware of the hazards
involved should remove the cover from the instrument.
For continued protection against fire, replace the line fuse only with a fuse of
the specified type and rating.
Do not install substitute parts or perform any unauthorized modification to the
product. Return the product to an Keysight Technologies Sales and Service
Office for service and repair to ensure that safety features are maintained.
4Keysight 34420A Service Guide
Environmental Conditions
The 34420A is designed for indoor use and in an area with low condensation. The
table below shows the general environmental requirements for this instrument.
Environmental conditionRequirement
Temperature
Humidity
Operating condition
– 0 °C to 55 °C
Storage condition
– –40 °C to 75 °C
Operating condition
– Up to 80% RH at 40°C (non-condensing)
Keysight 34420A Service Guide5
Waste Electrical and Electronic Equipment (WEEE) Directive 2002/
96/EC
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affixed product label indicates that you must not discard this
electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this
instrument is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted instrument, contact your nearest Keysight Service Center,
or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml
for more information.
Sales and Technical Support
To contact Keysight for sales and technical support, refer to the support links on
the following Keysight websites:
– www.keysight.com/find/34420A
(product-specific information and support, software and
documentation updates)
– www.keysight.com/find/assist
(worldwide contact information for repair and service)
6Keysight 34420A Service Guide
The 34420A is a 71/2 digit, high performance nanovolt, micro-ohm meter. Its
combination of bench-top and system features makes this meter a versatile
solution for your testing requirements now and in the future.
Convenient bench-top features
– Built-in math operations including thermistor, thermocouple and RTD
temperature measurements
– Two channel input allows ratio and difference functions for voltage
measurements
– Highly visible vacuum-fluorescent display
– Portable, ruggedized case with non-skid feet
Flexible system features
– GPIB (IEEE-488) interface and RS-232 interface are standard
– SCPI (Standard Commands for Programmable Instruments) and Keithley 181
compatibility
– Readings at up to 250 per second
Keysight 34420A Service Guide7
The Front-Panel at a Glance
1 Input channel select
2 Measurement function keys
3 Math operation keys
4 Filter select key
8Keysight 34420A Service Guide
5 Menu operation keys
6 Range/number of digits displayed keys
7 Single trigger/autotrigger/auto hold key
8 Shift / local key
The Front-Panel Menu at a Glance
The menu is organized in a top-down tree structure with three levels.
A: MEASurement MENU
1: DIG FILTER ) 2: INTEGRATE ) 3: OCOMP W ) 4: LOW POWER W ) 5: LOW VOLT W )
6: LoV LIMIT W
B: TEMPerature MENU
1:PROBE TYPE ) 2: UNITS ) 3: RTD TYPE ) 4:RTD Ro ) 5:T/C TYPE ) 6: COLD JUNCT ) 7:JUNCT
C: MATH MENU
1: STATS ) 2: NULL VALUE ) 3: SCALE GAIN ) 4 : SCALE OFST
D: TRIGger MENU
1: READ HOLD ) 2: TRIG DELAY ) 3: N SAMPLES
Keysight 34420A Service Guide9
E: SYStem MENU
NOTE
1: RDGS STORE ) 2: SAVED RDGS ) 3: ERROR ) 4: TEST ) 5: CHART OUT ) 6: CHART SPAN )
The commands enclosed in square brackets ([ ]) in the CAL MENU are "hidden"
unless the meter is UNSECURED for calibration.
10Keysight 34420A Service Guide
Display Annunciators
*
Adrs
Rmt
Man
Trig
OC Off
Null
Stats
Scale
ERROR
Shift
Ch1
Ch2
Ch1 - Ch2
Ch1 / Ch2
W
LP
2W
LoV
Hold
Filt
Mem
Turns on during a measurement.
Meter is addressed to listen or talk over the GPIB interface.
Meter is in remote mode (using remote interface).
Meter is using manual ranging (autorange is disabled).
Meter is waiting for a single trigger or external trigger.
Offset compensation is turned off.
A null value is being used.
Math statistics operations are being used.
Math scaling operations are being used.
Hardware or remote interface command error(s) detected.
"Shift" key has been pressed.
Meter input is on Channel 1.
Meter input is on Channel 2.
Meter is indicating the difference between inputs on channel 1 and channel 2.
Meter is indicating the ratio of the inputs on channel 1 and channel 2.
Meter is measuring resistance (Ohms).
Meter is using low power Ohms.
Meter is using 2-wire Ohms (annunciator off indicates a 4-wire measurement).
Meter is using voltage limited Ohms.
Automatic reading hold is enabled.
The analog and/or the digital filter is enabled.
Turns on when reading memory is enabled.
To review the display annunciators, hold down the Shift key as you turn on the
meter.
Keysight 34420A Service Guide11
The Rear Panel at a Glance
1 Chassis ground
2 Power-line fuse-holder assembly
3 Power-line voltage setting
4 Chart recorder output terminal
(Analog out)
Use the front-panel Input / Output Menu to:
– Select the GPIB or RS-232 interface
– Set the GPIB bus address
– Set the RS-232 baud rate and parity
12Keysight 34420A Service Guide
5 Voltmeter complete output terminal
6 External trigger input terminal
7 GPIB (IEEE-488) interface
connector
8 RS-232 interface connector
In This Book
NOTE
Quick Start Chapter 1 prepares the meter for use and helps you get familiar
with a few of its front-panel features.
Front-Panel OperationChapter 2 introduces you to the front-panel menu and
describes some of the meter's menu features.
Calibration Procedures Chapter 3 provides calibration, verification, and
adjustment procedures for the meter.
Theory of Operation Chapter 4 describes block and circuit level theory related
to the operation the meter.
Service Chapter 5 provides guidelines for returning your meter to Keysight
Technologies for servicing, or for servicing it yourself.
Replaceable Parts Chapter 6 contains a detailed parts lists of the meter.
Backdating Chapter 7 describes the differences between this manual and
older issues or this manual.
Schematics Chapter 8 contains the meter's block diagram, schematics,
disassembly drawings, and component locator drawings.
Characteristics and Specifications Chapter 9 lists the meter's specifications
and describes how to interpret these specifications.
If you have questions relating to the operation of the meter, call
1-800-452-4844 in the United States, or contact your nearest Keysight
Technologies Sales Office.
Keysight 34420A Nano Volt/Micro Ohm Meter
Service Guide
1Quick Start
Quick Start 26
To Prepare the Meter For Use 27
If the Meter Does Not Turn On 28
To Adjust the Carrying Handle 30
To Make Input Connections 31
To Measure Voltage 33
To Measure Resistance 34
To Measure Temperature With RTDs 36
To Measure Temperature With Thermocouples 37
To Select a Range 38
To Set the Number of Digits 39
To Set the Integration Time 40
Front-Panel Display Formats 41
To Rack Mount the Meter 42
25
1Quick Start
Shift
Shift
DCV
Shift
NOTE
Quick Start
This chapter helps you prepare the meter for use and contains exercises designed
to get you started with the meter, its menus, and the front panel.
The front panel has two rows of keys to select various functions and operations.
Most keys have a shifted function printed in blue above the key. To perform a
shifted function, press (the Shift display annunciator will turn on). Then,
press the key that has the desired label above it. For example, to select the
temperature measurement function, press .
If you accidentally press , just press it again to turn off the Shift annunciator.
The rear cover of this book is a fold-out Quick Reference Guide. On this cover
you will find a quick summary of various meter features. Inside the rear cover is a
diagram of the front panel menu options.
26Keysight 34420A Service Guide
To Prepare the Meter For Use
Shift
Shift
The following steps help you verify that the meter is ready for use.
1 Check the list of supplied items.
Verify that you have received the following items with your meter. If any item is
missing, contact your nearestKeysight Technologies Sales Office.
– One low thermal input cable.
– One low thermal four-wire shorting plug.
– One power cord.
–This User’s Guide.
–One Service Guide.
–One folded Quick Reference card.
– Certificate of Calibration with removable calibration label.
– One bottle of DeoxIT™ contact cleaner.
2 Connect the power cord and turn the meter on.
The front panel display will light up while the meter performs its power-on
self-test. The GPIB bus address is displayed. The meter is left in 61/2 digit
resolution, channel 1 input, and digital filter on.
Quick Start1
[1]
To view the display with all annunciators turned on, hold down as you
turn on the meter.
3 Perform a complete self-test.
The complete self-test performs a more extensive series of tests than those
performed at power-on. Hold down Shift as you press the power switch to turn
on the meter; continue to holdfor more than 5 seconds. The self-test will
begin when you release the key.
If the self-test is successful, “PASS” is displayed on the front panel. If the test is
not successful, “FAIL” is displayed and the ERROR annunciator turns on. See
the Service Guide for instructions on returning the meter to Keysight
Technologies for service.
[1] DeoxIT™ is a trademark of CAIG Laboratories, Inc., San Diego, California.
Keysight 34420A Service Guide27
1Quick Start
NOTE
If the Meter Does Not Turn On
Use the following steps to help solve problems you might experience when
turning on the meter. If you need more help, see the Service Guide for instructions
on returning the meter to Keysight Technologies for service.
1 Verify there is ac power to the meter.
First verify that the meter’s power switch is in the “On” position. Make sure the
power cord is firmly plugged into the power module on the meter’s rear panel.
Verify that the power source is energized.
2 Verify the power-line voltage setting.
The line voltage is set to the proper value for your country when the meter is
shipped from the factory. Change the voltage setting if it is not correct. The
settings are: 100, 120, 220, or 240 Vac (for 230 Vac operation, use the 220 Vac
setting).
See the diagram on the next page to verify or change the setting.
To replace the 250 mAT 250 V fuse, order part number 2110-0817. This fuse is
used for all power line voltage settings.
28Keysight 34420A Service Guide
Quick Start1
1 Remove the power cord. Remove
the fuse holder assembly from the
rear
3 Rotate the line-voltage selector
until the correct voltage appears in
the window.
2 Install the fuse. Remove the
line-voltage selector from the
assembly.
4 Replace the fuse-holder assembly
in the rear panel.
Install the correct fuse and verify that the correct line voltage appears in the
window.
Keysight 34420A Service Guide29
1Quick Start
To Adjust the Carrying Handle
To adjust the position, grasp the handle by the sides and pull outward. Then,
rotate the handle to the desired position.
Bench-top viewing positionsCarrying Position
30Keysight 34420A Service Guide
To Make Input Connections
NOTE
WARNING
Using the Cable Provided
Align the front panel connector and cable conductors, press in.Tighten coupling
nut.
Quick Start1
Keysight 34420A Service Guide31
The connector and cable are an integral part of the measurement system. For
the highest accuracy, use the copper cable and connectors supplied by Keysight
Technologies.
The conductors may require occasional cleaning to remove oxides. Cleaning the
conductors is described on page 137.
The connector body and cable shield are connected to earth ground.
1Quick Start
Using a Custom Cable
Refer to page 137 for additional information about building custom input cables.
32Keysight 34420A Service Guide
To Measure Voltage
NOTE
Channel 1 Ranges: 1 mV, 10 mV, 100 mV, 1 V, 10 V, 100 V
Maximum resolution: 0.1 nV (on 1 mV range)
Channel 2 Ranges: 1 mV, 10 mV, 100 mV, 1 V, 10 V
Maximum resolution: 0.1 nV (on 1 mV range)
Independent NULL for each channel
Channel 1 LO to Channel 2 LO isolated to 150 Vpeak.
Quick Start1
Scan mode alternates between input channels every 3 seconds or every
measurement, whichever is longer.
When using the internal thermistor as the temperature reference, you should
make a custom cable to connect the thermocouple wire directly to the input
terminals (see page 137).
Keysight 34420A Service Guide37
1Quick Start
NOTE
To Select a Range
You can let the meter automatically select the range using autoranging or you can
select a fixed range using manual ranging.
– Autoranging is selected at power-on and after a remote interface reset.
–Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
– If the input signal is greater than the present range can measure, the meter will
give an overload indication (“OVLD”).
– When making difference or ratio measurements, an overload on either channel
will give an overload indication (“OVLD”).
For voltage measurements, ranging is local to the selected channel. This means
that you can select the ranging method (auto or manual) for each channel
independently. When manually ranging, the selected range is local to the active
channel; the meter remembers the range when you switch between channels.
38Keysight 34420A Service Guide
To Set the Number of Digits
You can set the display to show 4½, 5½, 6½, or 7½ digits. In this book, the most
significant digit (leftmost on the display) is referred to as the “½” digit, since it can
only be a “0” or “1”. The number of digits displayed also depends upon the
integration time set, see page 40.
Quick Start1
– The number of digits is set to 6½ digits at power-on and after a remote
interface reset.
– The number of digits shown is dependent upon integration time and filter
settings. The meter will not allow you to show more digits than the practical
measurement capability of the meter. You can, however, reduce the number of
digits shown.
Keysight 34420A Service Guide39
1Quick Start
NOTE
To Set the Integration Time
Integration time is specified in Number of Power Line Cycles (NPLC). You can set
the NPLC to 0.02, 0.2, 1, 2, 10, 20, 100, or 200.
– You can set the integration time to one of three fixed values by choosing the
number of digits displayed, see page 39.
– You can set the integration time in the MEASure menu using the INTEGRATE
command.
– The Integration Time is directly related to the maximum number of digits the
meter will display.
NPLC
0.024½5½
0.25½6½
16½7½
26½7½
106½7½
207½7½
1007½7½
2007½7½
Filter Off
Max Digits Shown
Filter On
Max Digits Shown
– You can always show fewer digits than the maximum allowed (the minimum
number of digits shown is 4½).
Integration Time is local to the selected function. This means that you can select
the integration time for each function independently. The meter remembers
integration time when you switch between functions.
40Keysight 34420A Service Guide
1.021,651 V
-045.230 mV
0.133,256,7 W
O.VLD mV
-H.DDD,DDD,D EFF
6 digits
“½” digit
5 digits
“½” digit
7 digits
“½” digit
Front-panel display format
This is the 1 V range,
This is the 100 mV range,
This is the 1 ohm range,
This is an overload indication on the 1 mV range.
6½ digits are displayed.
5½ digits are displayed.
7½ digits are displayed.
–
H
D
E
F
Negative sign or blank (positive)
“½” digit (0 or 1)
Numeric Digits
Exponent (m, k, M)
Measurement units
Front-Panel Display Formats
Quick Start1
Keysight 34420A Service Guide41
1Quick Start
To Rack Mount the Meter
You can mount the meter in a standard 19-inch rack cabinet using one of three
optional kits available. Instructions and mounting hardware are included with
each rack-mounting kit. Any System II instrument of the same size can be
rack-mounted beside the 34420A meter.
Remove the carrying handle, and the front and rear rubber bumpers, before
rack-mounting the meter.
To remove the handle, rotate it to the vertical position and pull the end s
outward.
To remove the rubber bumper, stretch a corner and then slide it off.
42Keysight 34420A Service Guide
Quick Start1
To rack mount a single instrument, order adapter kit 5063-9240.
To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and
flange kit 5063-9212.
To install one or two instruments in a slid ing support shel f, order shel f
5063-9255, and slide kit 1494-0015 (for a single instrument, also order filler
panel 5002-3999).
Keysight 34420A Service Guide43
1Quick Start
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
44Keysight 34420A Service Guide
Keysight 34420A Nano Volt/Micro Ohm Meter
Service Guide
2Front-Panel Operation
Front-Panel Operation 46
Front-Panel Menu Reference 47
A Front-Panel Menu Tutorial 51
To Select an Input Channel 60
To Select a Range 61
To Set or Change the Number of Digits 62
To Set the Integration Time 63
To Trigger the Meter 64
45
2Front-Panel Operation
Front-Panel Operation
This chapter assumes you are familiar with the meter and menu operations. You
should also understand how to make connections for the various types of
measurements. If you are not familiar with this information, please read Chapter 1,
"Quick Start" starting on page 25.
This chapter does not give a detailed description of every front-panel key or menu
operation. It does, however, give you a good overview of the front-panel menu
and the most common front-panel operations. See Chapter 3, “Features and Functions”, of the Keysight 34420A User’s Guide for a complete discussion of the
meter’s capabilities and operation.
46Keysight 34420A Service Guide
Front-Panel Menu Reference
A: MEASurement MENU
1: DIG FILTER ) 2: INTEGRATE ) 3: OCOMP W ) 4: LOW POWER W ) 5: LOW VOLT W )
6: LoV LIMIT W
1: DIG FILTERSelects the digital filter speed. Can be set to FAST,
MEDIUM, or SLOW.
2: INTEGRATESets the measurement integration time.
3: OCOMP W Enables or disables offset compensation for resistance
measurements.
4: LOW POWER WEnables or disables low power ohms measurements.
5: LOW VOLT WEnables or disables voltage limited resistance
measurements.
6: LoV LIMIT W Sets the voltage limit for voltage limited resistance
measurements.
Front-Panel Operation2
B: TEMPerature MENU
1:PROBE TYPE ) 2: UNITS ) 3: RTD TYPE ) 4:RTD Ro ) 5:T/C TYPE ) 6: COLD JUNCT ) 7:JUNCT
1: PROBE TYPE Selects thermocouples, thermistors, or RTDs as the
temperature probe.
2: UNITSSets the measurement units to either °C, °F, or Kelvins.
3: RTD TYPE Sets the type of RTD in use.
4: RTD Ro Sets the nominal (0°C) value for the RTD in use.
5: T/C TYPESelects the type of thermocouple in use.
6: COLD JUNCTSets the source of the temperature reference junction.
7: JUNCT TEMP Allows direct entry of reference junction temperature.
Keysight 34420A Service Guide47
2Front-Panel Operation
C: MATH MENU
1: STATS ) 2: NULL VALUE ) 3: SCALE GAIN ) 4 : SCALE OFST
1:STATS Recalls the values of minimum, maximum, standard
2: NULL VALUE Selects independent null settings for voltage on channels 1
3: SCALE GAIN Sets the gain for linear scaling of the data.
4: SCALE OFSTSets the offset for linear scaling of data.
D: TRIGger MENU
1: READ HOLD ) 2: TRIG DELAY ) 3: N SAMPLES
1: READ HOLD Selects the reading hold sensitivity band.
2: TRIG DELAY Specifies a time delay between the trigger and the start of
3: N SAMPLES Sets the number of samples taken per trigger.
deviation, average, peak-to-peak, and number of readings.
and 2, resistance, and temperature functions.
the measurement.
48Keysight 34420A Service Guide
Front-Panel Operation2
E: SYStem MENU
1: RDGS STORE ) 2: SAVED RDGS ) 3: ERROR ) 4: TEST ) 5: CHART OUT ) 6: CHART SPAN )
1: SECURED The meter is secured against calibration; enter code to
unsecure.
1: UNSECURED The meter is unsecured for calibration; enter code to
secure.
2: CALIBRATEPerforms a calibration of meter; must be UNSECURED.
3: CHART ZERO Performs a calibration of the chart recorder output zero
level.
4: CHART GAIN Performs a calibration of the chart recorder output gain.
5: INJECTED I Performs a calibration to minimize the injected current.
6: CAL COUNT Reads the total number of times the meter has been
calibrated.
7: MESSAGE Reads the first 11 characters of a calibration string, if any,
entered from the remote interface.
[1] The commands enclosed in square brackets ( [ ] ) are “hidden” unless the meter is UNSECURED for
calibration.
50Keysight 34420A Service Guide
A Front-Panel Menu Tutorial
∧
∧
∧
∧
Shift
∧
Shift∧Auto/Man
Shift
∧
NOTE
This section is a step-by-step tutorial which shows how to use the front-panel
menu. We recommend that you spend a few minutes with this tutorial to get
comfortable with the structure and operation of the menu.
The menu is organized in a top-down tree structure with three levels (menus,
commands, and parameters). You move down or up the menu tree to
get from one level to the next. Each of the three levels has several horizontal
choices which you can view by moving left or right .
Front-Panel Operation2
–To turn on the menu, press
–To turn off the menu, press
(Menu On/Off).
(Menu On/Off), or press any of the
function or math keys on the top row of front-panel keys.
–To execute a menu command, press .
–To recall the last menu command that was executed, press
(Menu
Recall).
–To turn off the menu at any time without saving changes, press any function
key
If you become confused or get lost at any point during the tutorial, simply turn
off the menu and start over again with step 1 for that example.
Keysight 34420A Service Guide51
2Front-Panel Operation
∧
∧
∧
∧
∧
∧
∧∧∧∧∧
∧
Shift∧∧
Auto/Man
Shift
∧
Auto/Man
Messages Displayed During Menu Use
TOP OF MENU -You pressed while on the menus level; this is the top level of the menu and you
cannot go any higher.
To turn off the menu, press Shift (Menu On/Off). To move across the choices on a level,
press or . To move down a level, press .
MENUS - You are on the menus level. Press or view the choices.
COMMANDS - You are on the commands level. Press or to view the command choices
within the selected menu group.
PARAMETER - You are on the parameter level. Press or to view and edit the parameter
for the selected command.
MENU BOTTOM - You pressed while on the parameter level; this is the bottom level of the
menu and you cannot go any lower.
To turn off the menu, press (Menu On/Off). To move up a level, press .
ENTERED - The change made on the parameter level is saved. This is displayed after you press Auto/
Man (Menu Enter) to execute the command.
MIN VALUE - The value you specified on the parameter level is too small for the selected command.
The minimum value allowed is displayed for you to edit.
MAX VALUE - The value you specified on the parameter level is too large for the selected command.
The maximum value allowed is displayed for you to edit.
EXITING - You will see this message if you turn off the menu by pressing (Menu On/Off) or
a front-panel function/math key. You did not edit any values on the parameter level and changes were
NOT saved.
NOT ENTERED - You will see this message if you turn off the menu by pressing (Menu
On/Off) or a front-panel function/math key. You did some editing of parameters but the changes were
NOT saved. Press
VOLTS ONLY - The requested function is only available for voltage measurements.
UNAVAILABLE - The action requested is NOT allowed in the present configuration.
(Menu Enter) to save changes made on the parameter level.
52Keysight 34420A Service Guide
Front-Panel Operation2
-0.824,153 V-0.824153 V
With comma separator
(factory setting)
Without comma separator
Shift
∧
On/Off
A: MEAS MENU
∧∧∧
∧
E: SYS MENU
Menu Example 1
The following steps show you how to turn on the menu, move up or down
between levels, move across the choices on each level, and turn off the menu.
In this example, you will turn off the display comma separator.
The meter can display readings on the front panel with or without a comma
separator. The following steps show how to disable the comma.
1 Turn on the menu.
You enter the menu on the menus level. The MEAS MENU is your first choice
on this level.
2 Move across to the SYS MENU choice on this level.
There are six menu group choices available on the menus level. Each choice
has a letter prefix for easy identification (A: , B: , etc.).
Keysight 34420A Service Guide53
2Front-Panel Operation
∧
1: RDGS STORE
∧∧∧∧∧∧∧
∧
9: COMMA
∧
ON
∧
OFF
3 Move down to the commands level within the SYS MENU.
The RDGS STORE command is your first choice on this level.
4 Move across to the COMMA command on the command s level.
There are eleven command choices available in the SYS MENU. Each choice on
this level has a number prefix for easy identification (1: , 2: , etc.).
5 Move down a level to the COMMA parameter choices.
The first parameter choice is ON for the COMMA command (the comma
setting is stored in non-volatile memory and ON is the factory setting).
6 Move across to the OFF choice.
There are two parameter choices for COMMA.
54Keysight 34420A Service Guide
Front-Panel Operation2
Auto/Man
ENTER
ENTERED
Shift
∧
Recall
9: COMMA
∧
OFF
7 Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now in
effect. You are then exited from the menu. The meter will no longer show
commas in the display.
Menu Example 2
The following exercise demonstrates how to use the menu recall feature as a
shortcut to set the COMMA command back to its original setting. You must
perform the steps in Menu Example 1 before you start this example.
1 Use menu recall to return to the COMMA command.
This returns you to the COMMA command, which was the last command used
before you exited the menu in the Example 1.
2 Move down to the COMMA parameter choices.
The first parameter choice is OFF (the current setting from Example 1).
Keysight 34420A Service Guide55
2Front-Panel Operation
∧
ON
Auto/Man
ENTER
ENTERED
Shift
∧
On/Off
A: MEAS MENU
3 Move across to the ON choice.
Set the parameter back to its original value.
4 Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now in
effect. You are then exited from the menu.
Menu Example 3
Some commands in the menu require that you enter a numeric parameter value.
The following steps show you how to enter a number in the menu. For this
example, you will set the null value for channel 1 to –30 millivolts.
Make sure the meter has channel 1 selected, is in the dc voltage function, and
has 6½ digits displayed. Connect the four-wire short to the meter input.
1 Turn on the menu.
You enter the menu on the menus level. The MEAS MENU is your first choice
on this level.
56Keysight 34420A Service Guide
Front-Panel Operation2
∧
∧
C: MATH MENU
∧
1: STATS
∧
2: NULL VALUE
∧
˄
0.000,000,0 V
∧
2 Move across to the MATH MENU choice on this level.
There are six menu group choices available on this level.
3 Move down to the commands level within the MATH MENU.
The STATS command is your first choice on this level.
4 Move across to the NULL VALUE command.
5 Move down to edit the NULL VALUE parameter.
The null value should be 0.000000 V when you come to this point in the menu
for the first time. For this example, you will first set the null value to –0.300000
volts.
When you see the flashing “
” on the left side of the display, you can abort the
^
edit and return to the “commands” level by pressing .
Keysight 34420A Service Guide57
2Front-Panel Operation
∧∧– 0.000,000,0 V
∧
∧
– 0.000,000,0 V
∧∧∧
– 0.300,000,0 V
∧∧∧
∧
∧∧∧
– 0.300,000,0 V
6 Make the number negative.
The left most character on the display toggles between +, –, and
.
^
7 Move the flashing cursor over to edit the second digit.
Notice that the digit to the right of the decimal point is flashing.
8 Increment the digit until “3” is displayed.
You decrement or increment each digit independently. Neighboring digits are
not affected.
9 Move the flashing cursor over to the units location.
58Keysight 34420A Service Guide
Notice that the units are flashing on the right side of the display.
Front-Panel Operation2
∧
– 030.000,00 mV
Auto/Man
ENTER
ENTERED
Null
Null
Shift
∧
∧
NOTE
10 Decrease the displayed number by a factor of 10.
Notice that the position of the decimal point changes and the displayed
number increases by a factor of 10 and the mV annunciator is on.
11 Save the change and turn off the menu.
The meter beeps and displays a message to show that the change is now in
effect. You are then exited from the menu.
You will be returned to the normal measurement display, and, if you have
installed the four-wire short, the display will show a positive measurement in
the approximate amount of the null value just entered.
The Null annunciator is on in the display.
Keep in mind that null is turned on and - 0.03 volts is used as the null value for
measurements. To clear the null value, press .
Hint You can use the menu recall feature to save steps when entering the null
value. First, press on the front panel and then press
(Menu
Recall). The meter will jump to the 2: NULL VALUE command. Press . The
null value in the display is the first reading taken after you turned NULL on. You
can then edit this number as described.
This is the end of the front-panel menu tutorial. The remainder of the chapter
discusses several of the most common front-panel operations.
Keysight 34420A Service Guide59
2Front-Panel Operation
Channel
Shift
Channel
Channel
NOTE
To Select an Input Channel
For voltage measurements, you have two independent input channels.
–Pressing toggles the display between Channel 1 and Channel 2.
– Pressing (Scan) creates an alternating Channel 1 and Channel 2
display. Press again to cancel.
Scan mode alternates between input channels every 3 seconds or every
measurement, whichever is longer. If you are using the scan mode, turn off all
the input filters (both digital and analog).
60Keysight 34420A Service Guide
To Select a Range
NOTE
You can let the meter automatically select the range using autoranging or you can
select a fixed range using manual ranging.
Front-Panel Operation2
– Autoranging is selected at power-on and after a remote interface reset.
– Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
– If the input signal is greater than the present range can measure, the meter will
give an overload indication (“OVLD”).
– When making difference or ratio measurements, an overload on either channel
will give an overload indication (“OVLD”).
For voltage measurements, ranging is local to the selected channel. This means
that you can select the ranging method (auto or manual) for each channel
independently. When manually ranging, the selected range is local to the active
channel; the meter remembers the range when you switch between channels.
Keysight 34420A Service Guide61
2Front-Panel Operation
To Set or Change the Number of Digits
You can set the display to show 4½, 5½, 6½, or 7½ digits. In this book, the most
significant digit (leftmost on the display) is referred to as the “½” digit, since it can
only be a “0” or “1”. The number of digits displayed also depends upon the
integration time set, see page 63.
– The number of digits is set to 6½ digits at power-on and after a remote
interface reset.
– The number of digits shown is dependent upon integration time and filter
settings. The meter will not allow you to show more digits than the practical
measurement capability of the meter. You can, however, reduce the number of
digits shown.
62Keysight 34420A Service Guide
To Set the Integration Time
NOTE
Integration time is specified in Number of Power Line Cycles (NPLC). You can set
the NPLC to 0.02, 0.2, 1, 2, 10, 20, 100, or 200.
– You can set the integration time to one of three fixed values by choosing the
number of digits displayed, see page 62.
– You can set the integration time in the MEASure menu using the INTEGRATE
command.
– The Integration Time is directly related to the maximum number of digits the
meter will display.
Front-Panel Operation2
NPLC
0.024½5½
0.25½6½
16½7½
26½7½
106½7½
207½7½
1007½7½
2007½7½
Filter Off
Max Digits Shown
Filter On
Max Digits Shown
– You can always show fewer digits than the maximum allowed (the minimum
number of digits shown is 4½).
Integration Time is local to the selected function. This means that you can select
the integration time for each function independently. The meter remembers
integration time when you switch between functions.
Keysight 34420A Service Guide63
2Front-Panel Operation
Single
Single
Single
To Trigger the Meter
You can trigger the meter from the front panel using single trigger or autotrigger.
– Autotriggering is enabled when you turn on the meter. Notice that the
* (sample) annunciator turns on during each measurement.
– Single triggering takes one reading each time you press and then waits
for the next trigger. Continue pressing this key to trigger the meter.
Using an External Trigger
The external trigger mode is also enabled by pressing . It is like the single
trigger mode except that you apply a trigger pulse to the rear-panel Ext Trig
terminal. The meter is triggered on the negative edge of a TTL pulse.
The front-panel key is disabled when in remote.
64Keysight 34420A Service Guide
Null (Relative) Operation
Null
Null
When making null measurements, also called relative, each reading is the
difference between the input signal and a stored null value. You could, for
example, make a more accurate two-wire ohms measurement by shorting the test
leads and pressing to remove the test lead resistance.
Reading = measurement – null value
– The null value is adjustable and you can set it to any value between 0 and ±
120% of the highest range, for the present function.
– The null value is stored in volatile memory; the value is cleared when power
has been off or after a remote interface reset.
– Each function and channel has an independent null value. For voltage
measurements, each input channel has an independent null. Temperature
measurement null is applied to all temperature measurements (if you change
the probe type, null is still enabled and applied). The null for resistance
measurements applies to both 2-wire and 4-wire measurements.
– The null value is stored in the meter’s Null Register. There are two ways you
can specify the null value. First, you can enter a specific number into the
register from the front-panel menu or from the remote interface. Any previously
stored value is replaced with the new value. If you are operating the meter from
the front panel, entering a null value also turns on the null function.
The second way to enter the null value is to let the meter store the first reading
in the register. Press the key to make the displayed measurement the null
value and enable the Null operation. After you enable null, the first reading
displayed will be zero. If you entered a number into the register, as described
in the paragraph above, the first reading does not overwrite the stored value.
– Since an independent null value is stored for each input channel, difference or
ratio functions also use these null values. Additionally, when using the
difference function, an additional difference null can be applied (the difference
null is only available from the front panel).
Front-Panel Operation2
Keysight 34420A Service Guide65
2Front-Panel Operation
Stats
Shift
∧
Stats
Shift
∧
∧
To Use Math (Stats)
There are two math operations available, only one of which can be enabled at a
time. The meter can either scale the readings as they are taken, or can keep
statistics on a group of reading. The selected math operation remains in effect
until you disable it, change functions, turn off the power, or perform a remote
interface reset.
The scaling math operation is described in the Keysight 34420A User's Guide. This
discussion applies to stats (statistics) operations only.
Enable the statistics by pressing the key. Read the statistics gathered in the
MATH menu under the STATS command.
After viewing the stored statistics, press (Menu On/Off) to return to the
measurement and continue taking statistics.
After you enable statistics, the first reading that the meter takes is stored as both
the minimum and maximum value. The minimum is replaced with any subsequent
value that is less. The maximum is replaced with any subsequent value that is
greater.
All values are stored in volatile memory; the meter clears the values when stats is
turned on, when power has been off, or after a remote interface reset or preset.
When enabled, the following statistics are available:
MIN minimum reading
MAX maximum reading
AVER arithmetic average of all readings
STD DEVstandard deviation of all readings
PEAK-PEAKpeak to peak value of minimum and maximum
COUNTthe total number of readings
Hint You can use the menu recall feature to save steps when using the math
operations. After pressing on the front panel, press (Menu
Recall). The meter will jump to the 1:STATS command and you can press to
66Keysight 34420A Service Guide
Front-Panel Operation2
Shift
∧
view the statistics taken. After viewing the stored statistics, press
(Menu On/Off) to return to the measurement and continue taking statistics.
Keysight 34420A Service Guide67
2Front-Panel Operation
To Set the Number of Samples
Normally, the meter takes one reading (or sample) each time it receives a trigger
from the selected trigger source (if the meter is in the wait-for-trigger state). You
can, however, instruct the meter to take multiple readings for each trigger
received.
– Number of samples: 1 to 50,000. The default is 1 sample per trigger.
– The sample count is stored in volatile memory; the meter sets the sample
count to 1 when power has been off or after a remote interface reset. The
CONFigure and MEASure? commands automatically set the sample count to 1.
– Front Panel operation: set the number of samples per trigger in the
You can use the Chart Output (Analog Output) connector on the rear panel to run
a strip chart recorder or similar instrument. The connector provides an output
voltage proportional to the measured voltage.
– Enable the chart output under the 5: CHART OUT command in the SYStem
menu.
Set the chart span under the 6: CHART SPAN command in the SYStem menu.
Set the chart offset under the 7: CHART NULL command in the SYStem menu.
You can also press (Chart Null) to null the chart output voltage
based upon the displayed reading. This action places a null value in the Chart
Offset register. Pressing
7: CHART NULL command where you can manually edit the null value.
– The output can range from –3.00 V to +3.00 V.
– When chart output is disabled (Off), the output is set to 0 V.
– The output impedance is 1 kW.
– The output voltage is related to the input signal as follows:
Front-Panel Operation2
(Menu Recall) will take you to the
Note that reading includes the results of measurement unit conversion,
filtering, null, and any math operation performed on the measurement.
– The chart output state (including the span and null values) is stored in
non-volatile memory.
– span cannot have a negative value.
– The chart output has a voltage “rollover” to prevent the loss of data, see the
Keysight 34420A User’s Guide.
Keysight 34420A Service Guide69
2Front-Panel Operation
A: MEAS MENU
∧
G: CAL MENU
1: SECURED
Shift
∧
∧
NOTE
^000000: CODE
^034420: CODE
To Unsecure the Meter for Calibration
The meter can use a calibration security code to prevent unauthorized or
accidental calibration. This procedure shows you how to unsecure the meter for
calibration.
1 Turn on the menu.
2 Move across to the CAL MENU choice on this level.
3 Move down a level to the SECURED command.
If the display shows UNSECURED, you do not need to perform this procedure to
perform calibration.
4 Move down to the ''parameters” level.
5 Enter the security code.
6 Save the change and exit.
70Keysight 34420A Service Guide
Front-Panel Operation2
NOTE
The security code is set to “KT34420” when the meter is shipped from the factory.
The security code is stored in non-volatile memory and does not change when the
power has been off or after a remote interface reset.
To enter the security code from the front panel, enter only the six digits. To enter
the security code from the remote interface, you may use up to 11 characters.
Use the arrow keys to move left or right between digits. Use the up or down arrow
keys to change the digits.
To re-secure the meter (following calibration), perform this procedure again.
Additional information about the calibration security feature is given on page 91.
Keysight 34420A Service Guide71
2Front-Panel Operation
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.
72Keysight 34420A Service Guide
Keysight 34420A Nano Volt/Micro Ohm Meter
Service Guide
3Calibration Procedures
Keysight Technologies Calibration Services 75
Recommended Test Equipment 77
Test Considerations 78
Performance Verification Tests 79
Zero Offset Verification 81
Gain Verification 84
1 mV and 10 mV Gain Verification (Optional) 86
Injected Current Verification (Optional) 88
Chart Output Verification (Optional) 89
Calibration Security Code 91
Calibration Count 94
Calibration Message 95
Calibration Procedure 96
Aborting a Calibration in Progress 97
Zero Calibration 98
Gain Calibration 99
Injected Current Calibration (Optional) 101
Internal Temperature Reference Calibration (Optional) 103
Chart Output Calibration (Optional) 104
Error Messages 105
This chapter contains procedures for verification of the meter's performance and
adjustment (calibration).
Closed-Case Electronic Calibration The meter features closed-case electronic
calibration. No internal mechanical adjustments are required. The meter
calculates correction factors based upon the input reference value you set. The
73
3Calibration Procedures
new correction factors are stored in non-volatile memory until the next calibration
adjustment is performed. (Non-volatile memory does not change when power has
been off or after a remote interface reset.)
74Keysight 34420A Service Guide
Keysight Technologies Calibration Services
When your meter is due for calibration, contact your local Keysight Technologies
Service Center for a low-cost recalibration. The Keysight 34420A is supported on
automated calibration systems which allow Keysight Technologies to provide this
service at competitive prices. Calibrations to MIL-STD-45662 are also available at
competitive prices.
Calibration Interval
The meter should be calibrated on a regular interval determined by the
measurement accuracy requirements of your application. A 1-year or 2-year
interval is adequate for most applications. Keysight Technologies does not
recommend extending calibration intervals beyond 2 years for any application.
Whatever calibration interval you select, Keysight Technologies recommends that
complete re-adjustment should always be performed at the calibration interval.
This will increase your confidence that the Keysight 34420A will remain within
specification for the next calibration interval. This criteria for re-adjustment
provides the best long-term stability. Performance data measured using this
method can be used to extend future calibration intervals.
Calibration Procedures3
Time Required for Calibration
The Keysight 34420A can be automatically calibrated under computer control.
With computer control you can perform the complete calibration procedure and
performance verification tests in less than 50 minutes. Manual calibrations using
the recommended test equipment will take approximately 2 hours.
Automating Calibration Procedures
You can automate the complete verification and adjustment procedures outlined
in this chapter if you have access to programmable test equipment. You can
program the instrument configurations specified for each test over the remote
interface. You can then enter readback verification data into a test program and
compare the results to the appropriate test limit values.
You can also enter calibration constants from the remote interface. Remote
operation is similar to the local front-panel procedure. You can use a computer to
Keysight 34420A Service Guide75
3Calibration Procedures
perform the adjustment by first selecting the required setup. The calibration value
is sent to the meter and then the calibration is initiated over the remote interface.
The meter must be unsecured prior to initiating the calibration procedure. For
further information on programming the meter, see Chapter 4 in the Keysight 34420A User's Guide.
76Keysight 34420A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below. If the exact instrument is not available,
substitute calibration standards of equivalent accuracy.
Calibration Procedures3
InstrumentRequirementsRecommended model
Low thermal 4-wire shorting plugLow EMF four-wire short
Low thermal input connector and cableLow EMF, all copper
Multifunction Calibrator
Capacitor1000 pFnone0
1000:1 Voltage DividerLow Thermal Noisenone0
Temperature Measurement Device (RTD or
thermistor)
6 1/2 Digit Digital Multimeter (DMM)10 Vdc ± 0.01%Keysight 34401AO,T
[a] Q = Quick verification
P = Performance verification Tests
O = Optional Verification Tests
T = Troubleshooting
DC Volts O to 100 V resistance o to 1
MΩ
± 0.2° Cnone0
Keysight 34103A
(provided with meter)
Keysight 34102A
(provided with meter)
Fluke 5700Q,P
Use
Q,P,O,T
Q,P,O,T
[a]
Keysight 34420A Service Guide77
3Calibration Procedures
Test Considerations
To ensure proper instrument operation, verify that you have selected the correct
power line voltage prior to attempting any test procedure in this chapter. See “To
Prepare the Meter For Use” on page 27 for more information.
For optimum performance, all test procedures should comply with the following
recommendations:
– Assure that the calibration ambient temperature is stable and between 18° C
and 28° C.
– Assure ambient relative humidity is less than 80%.
– Allow a 3-hour warm-up period with the low thermal shorting plug connected
before verification or adjustment.
– Assure that low noise connections are made to the meter and calibrator.
Chapter 7 of the Keysight 34420A User's Guide describes low noise input
connections.
– Allow 5 minutes for thermal offset stabilization after handling the input
connector or shorting plug before making measurements or performing
calibrations.
– Ensure the input connector or low thermal shorting plug coupling nut is
properly tightened (torque to 20 in-lb maximum).
78Keysight 34420A Service Guide
Performance Verification Tests
Shift
The performance verification tests use the meter's specifications listed in
datasheet.
You can perform four different levels of performance verification tests:
– Self-Test A series of internal verification tests that give a high confidence that
the meter is operational.
– Quick Verification A combination of the internal self-tests and selected
verification tests.
– Performance Verification Tests An extensive set of tests that are
recommended as an acceptance test when you first receive the meter or aft.er
performing adjustments.
– Optional Verification Tests Tests not performed with every calibration.
Perform these tests to verify additional specifications or functions of the meter.
Self-Test
Calibration Procedures3
A brief power-on self-test occurs automatically whenever you turn on the meter.
This limited test assures that the meter is capable of operation.
To perform a complete self-test hold down the key as you press the Power
switch to turn on the meter; hold down the key for more than 5 seconds (a
complete description of these tests can be found in chapter 6). The meter will
automatically perform the complete self-test procedure when you release the key.
The self-test will complete in approximately 10 seconds.
You can perform many tests individually (or all tests at once) using the TEST
command in the SYS MENU. You can also perform a self-test from the remote
interface. See Chapter 4 in the Keysight 34420A User's Guide.
– If the self-test is successful, “PASS” is displayed on the front panel.
– If the self-test fails, “FAIL” is displayed and the ERROR annunciator turns on. If
repair is required, see Chapter 5, "Service" for further details.
– If all tests pass, you have a high confidence (90%) that the meter is
operational.
Keysight 34420A Service Guide79
3Calibration Procedures
NOTE
NOTE
NOTE
Quick Performance Check
The quick performance check is a combination of internal self-test and an
abbreviated performance test (specified by the letter Q in the performance
verification tests). This test provides a simple method to achieve high confidence
in the meter's ability to functionally operate and meet specifications. These tests
represent the absolute minimum set of performance checks recommended
following any service activity. Auditing the meter's performance for the quick
check points (designated by a Q) verifies performance for “normal” accuracy drift
mechanisms.
This test does not check for abnormal component failures.
To perform the quick performance check, do the following:
– Perform a complete self-test. A procedure is given on page 72.
– Perform only the performance verification tests indicated with the letter Q.
If the meter fails the quick performance check, adjustment or repair is required.
If the meter fails the quick performance check, adjustment or repair is
required.
Performance Verification Tests
The performance verification tests are recommended as acceptance tests when
you first receive the meter. The acceptance test results should be compared
against the 90 day test limits. You should use the 24-hour test limits only for
verification within 24 hours after performing the adjustment procedure. After
acceptance, you should repeat the performance verification tests at every
calibration interval.
If the meter fails performance verification, ad justment or repair is
required
80Keysight 34420A Service Guide
Zero Offset Verification
CAUTION
This procedure verifies the zero performance of the meter. Verification checks are
only performed for those functions and ranges with unique zero calibration
constants. A low thermal 4-wire shorting plug is applied to the input of the meter
for these checks.
Zero Offset Verification Procedure
1 Set the meter to the following configuration for each function, channel, and
range in the table on the next page:
7 ½ digits (100 PLC)
Filters OFF
2 Read “Test Considerations” on page 78.
3 Apply the low thermal 4-wire shorting plug (Keysight PN 34103A) to the
meter's input terminal. Ensure the connector coupling nut is properly
tightened.
4 Allow 5 minutes for the thermal shorting plug and input connector to
temperature stabilize.
5 Select each function and range in the order shown in the table on the next
page. Compare each measurement result to the test limits shown in the table.
Calibration Procedures3
– Zero offset calibration using a multifunction calibrator is NOT possible.
– Use only the low thermal 4-wire shorting plug provided.
Keysight 34420A Service Guide81
3Calibration Procedures
Table 3-1Zero Offset Verification Procedure
Step#Function/ChannelRange24 hour90day1 year
1Voltage / Channel 11 mV
120 nV
± Error From Zero
[a]
120 nV
[a]
120 nV
[a]
Q2Voltage / Channel 110mV
120 nV
[a]
120 nV
[a]
130 nV
3Voltage / Channel 1100 mV300 nV400 nV400 nV
4Voltage / Channel 11 V3 µV4 µV4 µV
5Voltage / Channel 110V10 µV40 µV40 µV
Q6Voltage / Channel 1100V400 µV500 µV500 µV
1Voltage /Channel 21 mV
Q2Voltage / Channel 210 mV
120 nV
120 nV
[a]
[a]
120 nV
120 nV
[a]
120 nV
[a]
130 nV
3Voltage / Channel 2100mV300 nV400 nV400 nV
4Voltage / Channel 21 V3 µV4 µV4 µV
5Voltage I Channel 210V10 µV40 µV40 µV
74-wire Ohms1 Ω2 µΩ2 µΩ2 µΩ
84-wire Ohms10 Ω20 µΩ20 µΩ20 µΩ
94-wire Ohms100 Ω200 µΩ200 µΩ200 µΩ
Q104-wire Ohms1 kΩ2 mΩ2 mΩ2 mΩ
[a]
[a]
[a]
114-wire Ohms10 kΩ20 mΩ20 mΩ20 mΩ
124-wire Ohms100 kΩ300 mΩ400 mΩ400 mΩ
134-wire Ohms1 MΩ3 Ω4 Ω4 Ω
14Low Power Ohms1 Ω2 µΩ2 µΩ2 µΩ
82Keysight 34420A Service Guide
Calibration Procedures3
Table 3-1Zero Offset Verification Procedure
± Error From Zero
Step#Function/ChannelRange24 hour90day1 year
15Low Power Ohms10 Ω20 µΩ20 µΩ20 µΩ
16Low Power Ohms100 Ω200 µΩ200 µΩ200 µΩ
17Low Power Ohms1 kΩ2 mΩ2 mΩn2 mΩ
18Low Power Ohms10 kΩ40 mΩ40 mΩ40 mΩ
19Low Power Ohms100 kΩ1.2 Ω1.5 Ω1.5 Ω
Q20Low Power Ohms1 MΩ3 Ω4 Ω4 Ω
21Voltage Limited Ohms10 Ω20 µΩ20 µΩ20 mΩ
22Voltage Limited Ohms100 Ω200 µΩ200 µΩ200 µΩ
Optional
[a] This value includes 100 nV for NULL off, see the specifications in datasheet.
[b] For 2-wire Ohms with Null ON (within ± 1°C of NULL), add 2 mΩ to the listed specification. If null is not used, add 0.2 Ω to the listed
specification.
[b]
2-wire OhmsAll
Use 4-wire Ohms specifications
[b]
Keysight 34420A Service Guide83
3Calibration Procedures
NOTE
Gain Verification
This procedure is used to check the “full scale” reading calibration of the meter.
Verification checks are only performed for those functions, channels, and ranges
with unique gain calibration constants.
Gain Verification Procedure
1 Set the meter to the following configuration:
7 ½ digits (100 PLC)
Filters OFF
2 Read “Test Considerations” on page 78.
3 Connect a multifunction calibrator to the meter's input terminals using the low
thermal input cable provided (Keysight PN 34102A). Ensure the connector
coupling nut is properly tightened.
4 Allow 5 minutes for the input connections to temperature stabilize.
5 Select each function and range on the meter in the order shown in the table on
the next page. At each function and range apply a full scale (input with the
multifunction calibrator. Compare each measurement result to the test limits
shown in the table.
The 1 mV and 10 mV gain verification is an optional verification test (see
page 86). These range gains are internally derived from the 1OOmV, 1V,
and 1OV range calibrations.
Table 3-2Gain Verification Procedure
± Error From Full Scale
Step#Cal ibrator OutputFunction/ChannelRange24 hours90day1 year
1100mVVoltage/ Channel 1100 mV1.8 µV3.4 µV4.4 µV
21 VVoltage/ Channel 11 V13 µV29 µV39 µV
Q310 VVoltage/ Channel 110 V30 µV240 µV340 µV
84Keysight 34420A Service Guide
Calibration Procedures3
Table 3-2Gain Verification Procedure
± Error From Full Scale
Step#Cal ibrator OutputFunction/ChannelRange24 hours90day1 year
[a] Channel 2 can be verified only, no channel 2 gain calibration exists. Channel 2 gain calibration is derived from channel 1 gain calibration.
[b] For 2-wire Ohms with Null ON (within± 1°C of NULL), add 2 mΩ to the listed specification. If null is not used, add 0.2 Ω to the listed
specification.
Use 4-wire Ohms specifications
[b]
Keysight 34420A Service Guide85
3Calibration Procedures
NOTE
Calibrator
Meter
Source +
Sense +
Source -
Sense -
1 k
1
Isothermal
environment
HI
LO
1 mV and 10 mV Gain Verification (Optional)
This optional procedure checks the “full scale” gain for the 1 mV and 10 mV
ranges. Because of the low voltage levels used to verify these ranges and the
inherent inaccuracies of multifunction calibrators at these levels, a voltage divider
is used to check these ranges.
These ranges can only be verified. The gain calibrations for these ranges
are derived from the calibrations on the 100mV, JV, and 10 V ranges.
1 mV and 10 mV gain verification procedure
1 Set the meter to the following configuration:
Channel 1 input
7 ½ digits (100 PLC)
Filters OFF
2 Connect the calibrator, resistive divider, and meter as shown.
3 Measure or calculate the divider ratio.
86Keysight 34420A Service Guide
Calibration Procedures3
NOTE
NOTE
4 Set the calibrator to fixed range (2.2 Volt range on range lock) and O Volt
output.
5 Null the reading on the meter.
6 Set the calibrator to 1 Volt output. Adjust the calibrator output from nominal
by the divider ratio to produce a 1 mV input to the meter.
Step 4, step 5 and step 6 should be performed quickly to minimize
thermal offsets.
7 Verify the meter reading against the table below.
8 Repeat step 4 through step 7 using the calibrator 11 Volt range and applying a
10 V output to produce a 10 mV input on the meter.
You should repeat step 4 through step 8 several times and average the
results to reduce the effects of measurement noise.
Table 3-31 mV and 10 mV gain verification procedure
± Error From Full Scale
Function/ChannelRange24 hour90day1 year
Voltage/ Channel 11 mV45nV60 nV70 nV
Voltage/ Channel 110mV270nV420 nV530nV
Keysight 34420A Service Guide87
3Calibration Procedures
NOTE
DMM
1 Mohm
+
-
1000 pF
Channel 1 LO
Keysight 34420A
Meter
LO
Common
Power Line
Injected Current Verification (Optional)
This optional procedure checks the injected current level. Injected current varies
with the power line configuration and frequency. You should, therefore, verify
injected current using the same power line configuration as the intended use.
Injected Current Verification Procedure
1 Connect the DMM and the meter to the same power line receptacle.
2 Set the DMM to measure AC Volts rms on the 100 mV range.
The DMM input impedance must be 1 MQ to perform this procedure.
3 Set the Keysight 34420A to the following configuration:
Channel 1 input
7½ digits (100 PLC)
Filters OFF
4 Connect the DMM, capacitor, and meter as shown.
5 Measure the rms voltage on the DMM. With the setup shown, 1 mV rms is
equivalent to 1 nA rms of injected current. Verify the measured voltage is <18
88Keysight 34420A Service Guide
mV rms (for peak-to-peak values, multiple the measured rms voltage by 2.8).
Chart Output Verification (Optional)
Shift
∧
On/Off
∧
∧∧∧
∧
∧
∧
∧
∧
Auto/Man
ENTER
Shift
∧
Recall
∧
∧
∧
Auto/Man
ENTER
This optional procedure checks the chart output (analog output) for zero and gain
calibration. Before beginning this procedure, you should check and record any
settings for the chart span and chart null (see page 63) so that these values can
be restored at the end of the procedure.
Chart Output Verification Procedure
1 Set the meter to the following configuration:
6 1/2 digits (10 PLC)
Channel 1 DCV
Math OFF
Null OFF
2 Read “Test Considerations” on page 78.
3 Connect the low thermal 4-wire shorting plug to the meter's input terminals.
4 Connect the rear panel chart output terminal to a DMM.
5 Turn on the menu.
6 Select the E:SYS menu.
Calibration Procedures3
7 Move down to the commands level and across to the 6:CHART SPAN
command.
8 Move down to the parameters level and set the value to 1.000,000 V.
9 Execute the command.
10 Use menu recall to return to the 6:CHART SPAN command.
11 Move to the 7:CHART NULL command.
12 Move down to the parameters level and set the value shown to 0.000,000 V.
13 Execute the command.
Keysight 34420A Service Guide89
3Calibration Procedures
Shift
∧
Recall
∧∧∧
Auto/Man
ENTER
Shift
∧
Recall
∧
∧
∧
Auto/Man
ENTER
14 Use menu recall to return to the 7: CHART NULL command.
15 Move to the 5:CHART OUT command.
16 Move down to the parameters level and set the value to ON.
17 Execute the command.
18 Measure the chart output with the DMM and verify that the output is:
0 Volts ±l mV.
19 Use menu recall to return to the 5: CHART OUT command.
20 Move to the 7:CHART NULL command.
21 Move down to the parameters level and set the value to -1.000,000 V.
22 Execute the command.
23 Measure the chart output with the DMM and verify that the output is:
1.000 Volt ±2 mV.
24 Restore chart null and chart span to the values set before beginning this
procedure.
90Keysight 34420A Service Guide
Calibration Security Code
NOTE
This feature allows you to enter a security code (electronic key) to prevent
accidental or unauthorized calibrations of the meter. When you first receive your
meter, it is secured. Before you can adjust calibration constants you must
unsecure the meter by entering the correct security code. A procedure to
unsecure the meter is given on “To Unsecure the Meter for Calibration” on
page 70.
– The security code is set to “KT034420” when the meter is shipped from the
factory. The security code is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
– To secure the meter from the remote interface, the security code may contain
up to 11 alphanumeric characters as shown below. The first character must be
a letter, but the remaining characters can be letters or numbers. You do not
have to use all 11 characters but the first character must always be a letter.
A _ _ _ _ _ _ _ _ _ _ _ (11 characters)
– To secure the meter from the remote interface but allow it to be unsecured
from the front panel, use the eight-character format shown below. The first two
characters must be “KT” and the remaining characters must be numbers. Only
the last six characters are recognized from the front panel, but all eight
characters are required. (To unsecure the meter from the front panel, omit the
“KT” and enter the remaining numbers.)
K T _ _ _ _ _ _ _ _ (8 characters)
Calibration Procedures3
If you secure the meter from the remote interface and the secure code
you use does not have the first two characters “KT” and the next 6
characters as digits, you cannot unsecure the meter from the front panel.
– If you forget your security code, you can disable the security feature by adding
a jumper inside the meter, and then entering a new code. See the procedure
on the following page.
Keysight 34420A Service Guide91
3Calibration Procedures
WARNING
To Unsecure the Meter Without the Security Code
To unsecure the meter without the correct security code, follow the steps below. A
procedure to unsecure the meter is given on “To Unsecure the Meter for
Calibration” on page 70. Also see Chapter 5, "Electrostatic Discharge (ESD)
Precautions"before beginning this procedure.
1 Disconnect the power cord and all input connections (both front and rear
terminals).
2 Remove the instrument cover. Refer to the disassembly drawing on page 155.
3 Connect the power cord and turn on the meter. Be careful not to touch the
power line connections.
Do not touch exposed mains.
4 Apply a short between the two exposed metal pads on JM500 as shown in
Figure 3-1.
Figure 3-1Applying short on JM500
92Keysight 34420A Service Guide
Calibration Procedures3
5 While maintaining the short, enter any unsecure code. The meter is now
unsecured.
6 Remove the short at JM500.
7 Turn off the meter and remove the power cord. Reassemble the meter.
8 The meter is now unsecured and you can enter a new security code. Be sure
you take note of the new security code.
Keysight 34420A Service Guide93
3Calibration Procedures
Calibration Count
The calibration count feature provides an independent “serialization” of your
calibrations. You can determine the number of times that your meter has been
calibrated. By monitoring the calibration count, you can determine whether an
unauthorized calibration has been performed.
Since the value increments by one for each calibration, a complete calibration
increases the value by approximately 24 counts.
– The calibration count is stored in non-volatile memory and does not change
when power has been off or after a remote interface reset. Your meter was
calibrated before it left the factory. When you receive your meter read the
calibration count to determine its value.
– The calibration count increments up to a maximum of 32,767 after which it
wraps around to 0. There is no way provided to program or reset the calibration
count. It is an independent electronic calibration “serialization” value.
94Keysight 34420A Service Guide
Calibration Message
You can use the calibration message feature to record calibration information
about your meter. For example, you can store such information as the last
calibration date, the next calibration due date, the meter's serial number, or even
the name and phone number of the person to contact for a new calibration.
You can record information in the calibration message only from the remote
interface. You can read the message from either the front-panel menu or the
remote interface.
– The calibration message may contain up to 40 characters. The meter can
display up to 11 characters of the message on the front panel; any additional
characters are truncated.
– The calibration message is stored in non-volatile memory, and does not
change when power has been off or after a remote interface reset.
Calibration Procedures3
Keysight 34420A Service Guide95
3Calibration Procedures
Calibration Procedure
The adjustment procedures use the CALibrate menu to initiate internal procedures
to generate and set calibration constants. The general menu procedure is the
same for all calibration setups.
Before beginning any calibration procedure the meter must be unsecured for
calibration (see page 70). The following procedure is the recommended method to
complete a meter calibration.
–Read “Test Considerations” on page 78.
– “To Unsecure the Meter for Calibration” on page 70.
– Perform the Zero Calibration procedure (page 81).
– Perform the Gain Calibration procedure (page 84).
– Secure the meter against calibration.
– Note the new security code and calibration count in the meter's maintenance
records.
96Keysight 34420A Service Guide
Aborting a Calibration in Progress
Shift
Sometimes it may be necessary to abort a calibration after the procedure has
already been initiated. You can abort a calibration at any time by pressing any
front-panel key (except ). When performing a calibration from the remote
interface, you can abort a calibration by issuing a remote interface device clear
message or by pressing the front-panel LOCAL key.
Calibration Procedures3
Keysight 34420A Service Guide97
3Calibration Procedures
Shift
∧
On/Off
∧
∧
∧
∧
Auto/Man
ENTER
Zero Calibration
A new set of zero offset correction constants are stored each time a zero
calibration is performed. Each channel, function, and range has a unique offset
correction. Once the procedure is started, the meter will automatically sequence
through all the required input channels, functions, and ranges and store a new
zero offset constant for each.
You may not perform a zero calibration for a single input channel, function or
range. All zero offset constants are determined when a zero calibration is begun.
As a part of the Zero Calibration, the input amplifier bias current is minimized. This
error current is dependent upon temperature. You should perform the Zero
Calibration procedures whenever the meter is used in environments with
temperatures greater than ±5° C from the calibration temperature.
Zero Calibration Procedure
The automatic zero calibration takes approximately 10 minutes to complete.
The procedure is outlined below. You should review “Test Considerations” on
page 78 before beginning this procedure.
1 Set the meter to the following configuration:
61/2 digits (10 PLC)
Channel 1 DCV
Filters OFF
2 Connect the low thermal 4-wire shorting plug to the meter's input terminals.
Ensure the connector coupling nut is properly tightened.
3 Allow 5 minutes for the thermal shorting plug and input connector to
temperature stabilize.
4 Turn on the menu and select G:CAL menu.
5 Move to the commands level and select 2:CALIBRATE.
6 Move down to the parameters level and set the value shown to 0.000,000,0 V.
7 Execute the command to begin the automatic zero calibration. The calibration
will take approximately 10 minutes to complete. The display will indicate the
calibration progress as each channel, range, and function is calibrated.
8 Perform the Zero Offset Verification procedure (page 81) to check the results.
98Keysight 34420A Service Guide
Gain Calibration
NOTE
The meter stores new gain correction constants each time this procedure is
followed. The gain constants are calculated from calibration values entered and
from measurements made automatically by the meter.
The 1 mV and 10 mV ranges use an internally derived gain constant and do not
have a separate gain calibration. The 1 mV and 10 mV gain constants are based
upon measurements performed during the 100 mV, 1 V and 10 V gain calibrations.
Channel 1 gain is used for both channel 1 and channel 2. No separate channel 2
gain constants are stored.
It is recommended that the calibration procedure be followed in the order
given here. If you are performing adjustments for an individual function,
perform the adjustments ONLY in the order shown.
Gain Adjustment Considerations
– The zero calibration procedure must have been recently performed before
beginning these gain adjustment procedures.
– When performing the 4-wire ohms gain adjustment, new gain constants are
also stored for the 2-wire ohms gain constants.
– Gain adjustments can be made with calibrator input values that are within 0.9
to 1.1 of Full Scale.
Calibration Procedures3
Gain Calibration Procedure
Perform the adjustments for each function only in the order shown.
1 Read “Test Considerations” on page 78.
2 Set the meter to the following configuration for each function, channel, and
range in Table 3-4:
7 1/2 digits (100 PLC)
Null OFF
Filters OFF
Keysight 34420A Service Guide99
3Calibration Procedures
3 Connect a multifunction calibrator to the meter's input terminals using the
provided cable (Keysight PN 34102A). Ensure the connector coupling nut is
properly tightened.
4 Allow 5 minutes for the input connector to temperature stabilize.
5 Select each function and range in the order shown in Table 3-4. Apply the
input signal shown.
6 Turn on the menu and select G:CALIBRATE menu.
7 Move to the commands level and select 2:CALIBRATE.
8 Move down to the parameters level and set the value in the display to the
calibrator input value.
9 Execute the command to begin the gain calibration.
10 Repeat step 1 through step 9 for each function and gain listed in the table.
11 Perform the “Gain Verification Procedure” on page 84 to check the results.
Table 3-4Gain calibration procedure
Step#
1
2
3
4100.00000 VDCV / Channel 1100V
51.000,0000 Ω4-wire Ohms1 Ω
610.000000 Ω4-wire Ohms10 Ω
7100.00000 Ω4-wire Ohms100 Ω
81,000.0000 Ω4-wire Ohms1 kΩ
910,000.000 Ω4-wire Ohms10 kΩ
10100,000.00 Ω4-wire Ohms100 kΩ
111,000,000.0 Ω4-wire Ohms1 MΩ
[a] Gain calibration can be performed with an input that is within 0.9 to 1.1 of range.
[b] The constants calculated for these ranges are used to derive the gain calibration constants for
the 1 mV and 10 mV ranges.
Calibrator Output
100.00000 mV
1.0000000 V
10.000000 V
[a]
[b]
[b]
[b]
Function/ChannelRange
DCV / Channel 1100mV
DCV / Channel 11 V
DCV / Channel 110 V
100Keysight 34420A Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.