Keysight Technologies 34136A User Manual

Keysight 34136A High Voltage Probe
User’s Guide
Notices
CAUTION
WARNING
Copyright Notice
duced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technol­ogies as governed by United States and international copyright laws.
Manual Part Number
34136-92000
Edition
Edition 6, July 2018
Printed in:
Printed in Germany
Published by:
Keysight Technologies Bayan Lepas Free Industrial Zone, 11900 Penang, Malaysia
Keysight Technologies Europe: Keysight Technologies Herrenberger Strasse 130 71034 Boeblingen, Germany
Technology Licenses
The hard ware and/or software described in this document are fur­nished under a license and may be used or copied only in accordance with the terms of such license.
Declaration of Conformity
Declarations of Conformity for this product and for other Keysight prod­ucts may be downloaded from the Web. Go to http://www.keysight.com/
go/conformity. You can then search by
product number to find the latest Dec­laration of Conformity.
U.S. Government Rights
The Software is “commercial computer software,” as defined by Federal Acqui­sition Regulation (“FAR”) 2.101. Pursu­ant to FAR 12.212 and 27.405-3 and Department of Defense FAR Supple­ment (“DFARS”) 227.7202, the U.S. government acquires commercial com­puter software under the same terms by which the software is customarily provided to the public. Accordingly, Keysight provides the Software to U.S. government customers under its stan­dard commercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.keysight.com/
find/sweula. The license set forth in the
EULA represents the exclusive authority by which the U.S. government may use, modify, distribute, or disclose the Soft­ware. The EULA and the license set forth therein, does not require or per­mit, among other things, that Keysight: (1) Furnish technical information related to commercial computer soft­ware or commercial computer software documentation that is not customarily provided to the public; or (2) Relinquish to, or otherwise provide, the govern­ment rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, per­form, display, or disclose commercial computer software or commercial com­puter software documentation. No additional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursu­ant to the FAR and the DFARS and are set forth specifically in writing else­where in the EULA. Keysight shall be under no obligation to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR
12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR
227.7103-5 (c), as applicable in any technical data.
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DIS­CLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED, WITH REGARD TO THIS MANUAL AND ANY INFORMA­TION CONTAINED HEREIN, INCLUD­ING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MER­CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSE­QUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR OF ANY INFORMATION CON­TAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WAR­RANTY TERMS COVERING THE MATE­RIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT SHALL CONTROL.
Safety Information
A CAUTION notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not pro­ceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating proce­dure, practice, or the like that, if not correctly performed or adhered to, could resul t in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
2 Keysight 34136A User’s Guide
Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU
This product complies with the WEEE Directive (2012/19/EU) marking requirements. This affixed product label indicates that you must not discard this electrical or electronic product in domestic household waste.
Product category:
With reference to the equipment types in the WEEE directive Annex 1, this product is classified as a “Monitoring and Control Instrument” product.
The affixed product label is as shown below.
Do not dispose in domestic household waste.
To return this unwanted product, contact your nearest Keysight Service Center, or visit http://about.keysight.com/en/companyinfo/environment/takeback.shtml for more information.
Keysight 34136A User’s Guide 3
WARNING
This high voltage probe is designed to prevent accidental shock to the operator when properly used. This operating note must be read and understood prior to using the probe. Improper procedures or incorrect analysis of the measurement situation can result in serious shock.
General Information
The Keysight 34136A High Voltage Probe is an accessory to be used with analog or digital multimeters having an input resistance of 10 MΩ (±1%). The 34136A is a 1000:1 divider which extends a voltmeter’s measurement capability to 28 kV rms ac (DC+AC p-p 40 kV).
Specifications
Maximum input voltage
DC voltage: 28 kV rms ac (DC+AC p-p 40 kV) AC voltage: (0 to 300 Hz) 28 kV rms
Typical accuracy
DC voltage: ±2% (1 kV to 20 kV) ±3% (20 kV to 40 kV)
Pollution degree 2
Other Characteristics
Input resistance
1000 MΩ
Cable length
1 meter
Altitude
Up to 2000 meters (6561 ft)
Typical accuracy
AC voltage: 5% at 60 Hz
Temperature coefficient
Less than 200 ppm/ °C
Typical division ratio accuracy
1000:1 ±1% when terminated in 10 M 2000:1 ±6% when terminated in 1 M Note: For the 34401A, 34405A, L4411A, 34410A/11A, and 3458A multimeters, the ACV input impedance function has a 1 M input impedance.
Operating temperature
0 °C to +50 °C
Storage temperature
–20 °C to +70 °C
Humidity
Up to 80% relative humidity at +40 °C
4 Keysight 34136A User’s Guide
AC output derates with frequency to –3 dB at 300 Hz. The graph below shows
0
–3
Atten (dB)
DC
Frequency
300 Hz
typical AC accuracy.
Actual AC accuracy varies with the amount of shunted input capacitance.
Safety Precautions
– This high voltage probe must only be used by personnel who are trained,
experienced, or otherwise qualified to recognize hazardous situations and who are trained in the safety precautions that are necessary to avoid possible injury
when using such a device. – Do not work alone when working with high voltage circuits. – For your own safety, inspect the probes for cracks and frayed or broken leads
before each use. If defects are noted, DO NOT use the probe. – Hands, shoes, floor and work bench must be dry. Avoid making measurements
under humid, damp or other environmental conditions that might effect the
safety of the measurement situation. – It is advisable to turn the high voltage source off before connecting or
disconnecting the probe. – The probe body should be kept clean and free of any conductive
contamination. Refer to the section on cleaning.
Operation
– Connect the plugs to the volts [Hi] and com [Lo] terminals of your voltmeter. – Select the desired voltage function and range; do not use autoranging. – Whenever possible, turn the high voltage source off before making any
connections. – Connect the divider probe common lead (alligator clip) to a good earth ground
or reliable chassis ground.
Keysight 34136A User’s Guide 5
WARNING
– Do not attempt to take measurements from sources where the chassis or
return lead is not grounded.
– The ground lead is critical to the safe operation of the probe. Failure to
make this connection when making high voltage measurements may result in personal injury or damage to the probe or voltmeter. This connection must always be made BEFORE the probe tip comes in contact with the high voltage and must not be removed until the probe tip has been removed from the high voltage source.
– Do not connect the ground clip lead to the high voltage source for any
reason.
– Before turning the high voltage on, make sure that no part of your body is
in contact with the device under test.
– Measure the voltage remembering that the voltage being measured is
1000 times greater than the voltmeter reading.
– Turn the high voltage off.
– Disconnect the probe tip from the high voltage source BEFORE removing
the ground clip lead.
– This probe assembly is intended for use within measurement category I.
Do not use the probe assembly for measurements within the other measurement categories or transient overvoltages greater than 1500 V.
Cleaning
– Clean only the exterior probe body and cables. Use a soft cotton cloth lightly
moistened with a mild solution of detergent and water. Do not allow any portion of the probe to be submerged at any time.
– Dry the probe thoroughly before attempting to make voltage measurements.
– Do not subject the probe to solvents or solvent fumes as these can cause
deterioration of the probe body and cables.
6 Keysight 34136A User’s Guide
This information is subject to change without notice. Always refer to the Keysight website for the latest revision.
© Keysight Technologies 2011 - 2018 Edition 6, July 2018
Printed in Germany
*34136-92000*
34136-92000
www.keysight.com