No part of this manual may be reproduced in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Keysight
Technologies as governed by United States
and international copyright laws.
Manual Part Number
16860-97007
Edition
Second edition, September 2018
Available in electronic format only
Keysight Technologies
1900 Garden of the Gods Road
Colorado Springs, CO 80907 USA
Warranty
THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED "AS IS," AND IS SUBJECT TO BEING CHANGED, WITHOUT
NOTICE, IN FUTURE EDITIONS. FURTHER,
TO THE MAXIMUM EXTENT PERMITTED BY
APPLICABLE LAW, KEYSIGHT DISCLAIMS
ALL WARRANTIES, EITHER EXPRESS OR
IMPLIED WITH REGARD TO THIS MANUAL
AND ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION
WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY
INFORMATION CONTAINED HEREIN.
SHOULD KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer software," as defined by Federal Acquisition
Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of
Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires
commercial computer software under the
same terms by which the software is cu temporarily provided to the public. Accordingly,
Keysight provides the Software to U.S. government customers under its standard commercial license, which is embodied in its
End User License Agreement (EULA), a copy
of which can be found at http://www.key-
sight. com/find/sweula. The license set
forth in the EULA represents the exclusive
authority by which the U.S. government may
use, modify, distribute, or disclose the Software. The EULA and the license set forth
therein, does not require or permit, among
other things, that Keysight: (1) Furnish technical information related to commercial
computer software or commercial computer
software documentation that is not customarily provided to the public; or (2) Relinquish
to, or otherwise provide, the government
rights in excess of these rights customarily
provided to the public to use, modify, reproduce, release, perform, display, or disclose
commercial computer software or commercial computer software documentation. No
additional government requirements beyond
those set forth in the EULA shall apply,
except to the extent that those terms, rights,
or licenses are explicitly required from all
providers of commercial computer software
pursuant to the FAR and the DFARS and are
set forth specifically in writing elsewhere in
the EULA. Keysight shall be under no obligation to update, revise or otherwise modify
the Software. With respect to any technical
data as defined by FAR 2.101, pursuant to
FAR 12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no
greater than Limited Rights as defined in
FAR 27.401 or DFAR 227.7103-5 (c), as
applicable in any technical data. 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It
calls attention to an operating procedure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not proceed beyond a CAUTION notice until
the indicated conditions are fully
understood and met.
A WARNING notice denotes a hazard.
It calls attention to an operating procedure, practice, or the like that, if
not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
2Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 3
Additional Safety Notices
!
This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety
Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety
Class I instrument (provided with terminal for protective earthing). Before applying power, verify that
the correct safety precautions are taken (see the following warnings). In addition, note the external
markings on the instrument that are described under “Safety Symbols.”
Warnings
•Before turning on the instrument, you must connect the protective earth terminal of the
instrument to the protective conductor of the (mains) power cord. The mains plug shall only be
inserted in a socket outlet provided with a protective earth contact. You must not negate the
protective action by using an extension cord (power cable) without a protective conductor
(grounding). Grounding one conductor of a two-conductor outlet is not sufficient protection.
•Only fuses with the required rated current, voltage, and specified type (normal blow, time delay,
etc.) should be used. Do not use repaired fuses or short-circuited fuseholders. To do so could
cause a shock or fire hazard.
•If you energize this instrument by an auto transformer (for voltage reduction or mains isolation),
the common terminal must be connected to the earth terminal of the power source.
•Whenever it is likely that the ground protection is impaired, you must make the instrument
inoperative and secure it against any unintended operation.
•Service instructions are for trained service personnel. To avoid dangerous electric shock, do not
perform any service unless qualified to do so. Do not attempt internal service or adjustment
unless another person, capable of rendering first aid and resuscitation, is present.
•Do not install substitute parts or perform any unauthorized modification to the instrument.
•Capacitors inside the instrument may retain a charge even if the instrument is disconnected from
its source of supply.
•Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any
electrical instrument in such an environment constitutes a definite safety hazard.
•Do not use the instrument in a manner not specified by the manufacturer.
Safety Symbols
Instruction manual symbol: the product is marked with this symbol when it is necessary for you to
refer to the instruction manual in order to protect against damage to the product.
Indicates hazardous voltages and potential for electrical shock.
Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
The C-tick mark is a registered trademark of the Spectrum Management Agency of Australia. This
signifies compliance with the Australia EMC Framework regulations under the terms of the Radio
Communication Act of 1992.
CE compliance marking to the EU Safety and EMC Directives.
ISM GRP-1A classification according to the international EMC standard.
ICES/NMB-001 compliance marking to the Canadian EMC standard.
Keysight 16860 Series Portable Logic Analyzer Service Guide3
Page 4
KC certification mark to demonstrate compliance with the South Korean EMC requirements.
South Korean Class A EMC declaration:
This equipment is Class A suitable for professional use and is for use in electromagnetic
environments outside of the home.
CSA is the Canadian certification mark to demonstrate compliance with the Safety requirements.
Notice for the European Community: This product complies with the WEEE Directive (2002/96/EC)
marking requirements. The affixed label indicates that you must not discard this electrical/electronic
product in domestic household waste. Product Category: With reference to the equipment types in
the WEEE Directive Annex I, this product is classed as a “Monitoring and Control instrumentation”
product.
Do not dispose in domestic household waste. To return unwanted products, contact your local
Keysight office or get more information from www.keysight.com/environment/product.
4Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 5
Keysight 16860 Series Logic Analyzer - At a Glance
The Keysight Technologies 16860 series logic analyzers are portable logic analyzers that range from
34 to 136 logic acquisition channels, depending on the model.
The 16860-series models provide features such as:
• deeper memory
• ¼ channel 10GHz Timing mode
• deskew of timing traces by individual channels
• single as well as multiple clocks support
• advanced clocking capabilities
• clock hysteresis
• various clock modes such as Master, Dual Sample, Master/Slave, and Demultiplex.
This series of logic analyzers can be used:
• at a higher speed (700MHz to 12.5 MSps) for debug, validation and analysis of DDR and LPDDR
memory systems.
• at a lower speed (350MHz to 0 MSps) as a medium to high performance general purpose logic
analyzer.
Model Comparison
Keysight Model Number 16861A16862A16863A16864A
Logic acquisition channels34 68102136
Features, Logic Acquisition
• 2 M to 128 M samples of memory depth per channel (depending on memory option), software
upgradeable.
• 350 MHz or 700 MHz maximum state clock rate (depending on state speed option), software
upgradeable.
• Full Channel Timing Mode at 2.5 GHz sampling with 12.5 GHz Timing Zoom.
• Half Channel Timing Mode at 5.0 GHz sampling with 12.5 GHz Timing Zoom
Keysight 16860 Series Portable Logic Analyzer Service Guide5
Page 6
• Quarter Channel Timing Mode at 10 GHz sampling with 12.5 GHz Timing Zoom
• Eye scan (automatic threshold and sample position setup) feature.
• Single-ended and differential probing support
Features, Mainframe
• Built-in 15 inch color touch screen with 1024 x 768 resolution
• Removable hard drive
• 10/100/1000 Base-T LAN port
• USB 2.0 ports (four total, two on front, two on back)
• USB 3.0 ports (two total, both ports are on the back)
• One PCI expansion slot
• One PCI Express x1 expansion slot
• Windows 7 installation
• Keysight Logic and Protocol Analyzer application which takes the complexity out of making
logic analyzer measurements. You can perform all operations directly from one window.
To know about these features in detail, refer to the 16860 series logic analyzer Data Sheet
(5992-1723EN) on www.keysight.com and the Logic and Protocol Analyzer online help installed with
the Logic and Protocol Analyzer software.
Supplied Accessories
• USB mouse
• USB keyboard
• Accessory pouch and power cord
Snap the accessories pouch to the top of the 16860 series logic analyzer. Use it to store probe leads,
accessories, or manuals. Use the tie-down straps under the flap to conveniently hold pod cables not
in use or during transport.
Optional Accessories
• Probes
Service Strategy
The service strategy for this instrument is the replacement of defective assemblies/parts. This service
guide contains information for finding a defective assembly by testing and returning it to Keysight
Technologies for all service work, including troubleshooting. Contact your nearest Keysight
Technologies Sales Office for more details.
Contacting Keysight Technologies
To locate a sales or service office near you, go to www.keysight.com/find/contactus.
6Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 7
In This Service Guide
This book is the service guide for the 16860 series logic analyzer.
This service guide has seven chapters.
Chapter 1, “General Information” contains information about the instrument, lists accessories for the
module, gives specifications and characteristics of the instrument, and provides a list of the
equipment required for servicing the instrument.
Chapter 2, “Preparing for Troubleshooting or Performance Testing” tells how to prepare the
instrument for use.
Chapter 3, “Testing 16860 Series Logic Analyzers Performance” tells how to verify the 16860 series
logic analyzer performance with specifications.
Chapter 4, “Calibrating” contains calibration instructions for the instrument (if required).
Chapter 5, “Troubleshooting” contains explanations of self-tests and flowcharts for troubleshooting
the instrument.
Chapter 6, “Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies” describes
how to replace the instrument and assemblies of the instrument, and how to return these to Keysight
Technologies.
Chapter 7, “Replaceable Parts” contains a list of replaceable parts available for 16860 Series Logic
Analyzers and how to order these parts.
Keysight 16860 Series Portable Logic Analyzer Service Guide7
Page 8
8Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 9
Contents
1 General Information
Keysight 16860 Series Logic Analyzer - At a Glance5
Model Comparison5
Features, Mainframe6
Supplied Accessories6
Optional Accessories6
In This Service Guide7
Accessories14
Probes, Cables, and Accessories14
Software14
Specifications15
Characteristics16
2 Preparing for Troubleshooting or Performance Testing
Operating Environment18
To apply power19
To clean the instrument20
To start the user interface21
To test the 16860 series logic analyzer22
3 Testing 16860 Series Logic Analyzers Performance
Perform the Self-Tests25
Equipment Required for the Performance Test26
Assemble the SMA/Flying Lead Test Connectors27
Set Up the Test Equipment30
Connect the Test Equipment33
Connect the Logic Analyzer Pod to the Pulse Generator33
Test the 16860 series Logic Analyzer35
Keysight 16860 Series Portable Logic Analyzer Service Guide9
Page 10
Contents
Create Test Configuration Files36
Create Configuration File 1 - Multiple Clocks (no Marker)36
Create Configuration File 2 - Multiple Clocks (with Marker)39
Create Configuration File 3 - Single Clock (no Marker)42
Create Configuration File 4 - Single Clock with Marker43
Determine Maximum Data Rate for Multiple Clocks Mode45
Pod 1 Clock - Maximum Clock Rate45
Pod 1 Clock - Setup for Maximum Data Rate46
Pod 1 Clock - Measuring Maximum Data Rate50
Pod 2 Clock - Maximum Clock Rate51
Pod 2 Clock - Setup for Maximum Data Rate53
Pod 2 Clock - Measuring Maximum Data Rate54
Pod 3 Clock - Maximum Clock Rate55
Pod 3 Clock - Setup for Maximum Data Rate56
Pod 3 Clock - Measuring Maximum Data Rate57
Pod 4 Clock - Maximum Clock Rate59
Pod 4 Clock - Setup for Maximum Data Rate60
Pod 4 Clock - Measuring Maximum Data Rate61
Determine Maximum Data Rate for Single Clock Mode63
Single Clock Rising Edge - Maximum Clock Rate63
Single Clock Rising Edge - Pod 1 Data - Setup for Maximum Data Rate65
Single Clock Rising Edge - Pod 1 Data - Measuring Maximum Data Rate66
Single Clock Falling Edge - Maximum Clock Rate67
Single Clock Falling Edge - Pod 1 Data - Setup for Maximum Data Rate68
Single Clock Falling Edge - Pod 1 Data - Measuring Maximum Data Rate69
Single Clock Both Edges - Maximum Clock Rate70
Single Clock Both Edges - Pod 1 Data - Setup for Maximum Data Rate72
Single Clock Both Edges - Pod 1 Data - Measuring Maximum Data Rate72
Single Clock Both Edges - Pod 2 Data - Setup for Maximum Data Rate73
Single Clock Both Edges - Pod 2 Data - Measuring Maximum Data Rate74
Single Clock Both Edges - Pods 3 to 875
Conclude the State Data Rate Tests76
Performance Test Record77
4Calibrating
Calibration Strategy80
5 Troubleshooting
To use the system troubleshooting flowcharts82
To use the logic acquisition troubleshooting flowcharts87
10Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 11
To troubleshoot system power problems89
Power Supplies89
To run the self-tests90
Logic Acquisition Self-Test Descriptions91
To exit the test system92
To restore the system software93
Contacting Keysight Service/Support96
To test the logic acquisition cables97
6 Removing, Replacing, or Returning 16860 Series Logic Analyzer Assemblies
Before you Start106
Repair Tools Needed106
Back Up Data and License Files107
Prepare the Logic Analyzer for Disassembly107
Removing and Replacing the Hard Drive, Acquisition Bezel, and Cover109
Contents
Removing and Replacing the Front Panel112
Disassembling the Front Panel114
Removing and Replacing the Fan Deck118
Removing and Replacing the Acquisition Tray119
Removing and Replacing the Duct121
Removing and Replacing the Motherboard122
Removing and Replacing the FIB Board124
Removing and Replacing the Power Supply Shroud and Power Supply126
Returning the 16860 Series Logic Analyzer or its Assemblies129
7 Replaceable Parts
Ordering Replaceable Parts132
Exchanging Assemblies132
Replaceable Parts List133
Index
Keysight 16860 Series Portable Logic Analyzer Service Guide11
Page 12
Contents
12Keysight 16860 Series Portable Logic Analyzer Service Guide
This chapter lists the accessories and some of the specifications and characteristics for testing and
servicing the 16860 series logic analyzer. See the 16860 Series logic analyzer’s online help for a full
listing of all specifications and characteristics.
Page 14
1General Information
Accessories
Probes, Cables, and Accessories
Software
One or more of the following accessories, sold separately, are required to set up and operate the
16860 series logic analyzer for testing and servicing it.
Refer to the 16860 series logic analyzer Data Sheet (5992-1723EN) on www.keysight.com to get a
list of supported interposers, probes, and cables for 16860 series logic analyzer. Details of these
interposers, probes, and cables are in their respective user guides available on www.keysight.com.
The 16860 series logic analyzer requires:
• Keysight Logic Analyzer software version 06.30 or higher to configure, control, and use the 16860
series logic analyzer.
14Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 15
Specifications
General Information1
The specifications are the performance standards against which the product is tested.
Table 1 Maximum State Data Rate Specification for the 16860 Series Logic Analyzers
Maximum state data rate
1
(spec)
350 MHz
(Base configuration)
Single ClockSingle Clock Multiple Clocks
• Captures data up to
350 Mbps on either
edge of a clock up to
350 MHz
• Captures data up to
700 Mbps on either
edge of the clock up
to 700 MHz
700 MHz
(Option 700)
• Captures data up to
700 Mbps on both
edges of a clock up to
350 MHz
• Captures data up to
1400 Mbps on both
edges of the clock up
to 700 MHz
350 MHz
(Base configuration)
• Captures data up to
700 Mbps on any
combination of
multiple clocks up to
350 MHz
1.Specification (spec): Represents warranted performance of a calibrated instrument that has been
stored for a minimum of 2 hours within the operating temperature range of 5 to 40 °C, unless otherwise
stated, and after a 45-minute warm-up period. The specifications include measurement uncertainty.
Keysight 16860 Series Portable Logic Analyzer Service Guide15
Page 16
1General Information
Characteristics
For a full listing of all specifications and characteristics, see the Keysight 16860 series logic analyzer
Data Sheet, literature part number 5992-1723EN available on Keysight’s web site
(www.keysight.com).
16Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 17
Keysight 16860 Series Portable Logic Analyzer
Service Guide
2Preparing for Troubleshooting
or Performance Testing
Operating Environment / 18
To apply power / 19
To clean the instrument / 20
To start the user interface / 21
To test the 16860 series logic analyzer / 22
This chapter provides instructions for preparing the 16860 series logic analyzer for troubleshooting or
servicing it.
Page 18
2Preparing for Troubleshooting or Performance Testing
Operating Environment
The operating environment specifications are listed in the Keysight 16860 series logic analyzer Data Sheet, literature part number 5992-1723EN available on Keysight’s web site (www.keysight.com).
Note the non-condensing humidity limitation. Condensation within the instrument can cause poor
operation or malfunction. Provide protection against internal condensation.
18Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 19
To apply power
Preparing for Troubleshooting or Performance Testing2
1 Connect the supplied power cord to the instrument and to the power source.
This instrument autodetects the line voltage from 100 VAC to 240 VAC. It is equipped with a
three-wire power cable. When connected to an appropriate AC power outlet, this cable grounds
the instrument cabinet. The type of power cable plug shipped with the instrument depends on
the model ordered and the country of destination.
2 Turn on the power switch located on the front panel.
For first-time power up considerations and setup steps, refer to the Installation Guide that came
with your instrument. To get the most up-to-date installation guide:
• Go to www.keysight.com.
• Search for 16860 Series Logic Analyzers.
• Look under Technical Support and then Document Library.
Keysight 16860 Series Portable Logic Analyzer Service Guide19
Page 20
2Preparing for Troubleshooting or Performance Testing
To clean the instrument
If the instrument requires cleaning:
1 Remove power from the instrument.
2 Clean the external surfaces of the instrument with a soft cloth dampened with water.
3 Make sure that the instrument is completely dry before reconnecting it to a power source.
20Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 21
To start the user interface
Start the Keysight Logic and Protocol Analyzer application from the Start menu or using a shortcut.
On the desktop, the Keysight Logic Analyzer icon looks like:
Refer to the Keysight Logic and Protocol Analyzer application’s on-line help for information on how
to operate the user interface.
Preparing for Troubleshooting or Performance Testing2
Keysight 16860 Series Portable Logic Analyzer Service Guide21
Page 22
2Preparing for Troubleshooting or Performance Testing
To test the 16860 series logic analyzer
The 16860 series logic analyzer does not require an operational accuracy calibration or adjustment.
After installing the instrument, you can test and use the instrument.
• If you require a test to verify logic analyzer’s performance with the specifications, see “Testing
16860 Series Logic Analyzers Performance" on page 23.
• If you require a test to verify correct instrument operation, see “To run the self-tests" on page 90.
• If the instrument does not operate correctly, see “Troubleshooting" on page 81.
22Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 23
Keysight 16860 Series Portable Logic Analyzer
NOTE
Service Guide
3Testing 16860 Series Logic
Analyzers Performance
Perform the Self-Tests / 25
Equipment Required for the Performance Test / 26
Assemble the SMA/Flying Lead Test Connectors / 27
Set Up the Test Equipment / 30
Connect the Test Equipment / 33
Test the 16860 series Logic Analyzer / 35
Create Test Configuration Files / 36
Determine Maximum Data Rate for Multiple Clocks Mode / 45
Determine Maximum Data Rate for Single Clock Mode / 63
Conclude the State Data Rate Tests / 76
Performance Test Record / 77
This chapter provides information on how to test the performance of the 16860 series logic analyzers
module against the specifications listed on page 15.
To ensure the 16860 series logic analyzer is operating as specified, software tests (self-tests) and a
manual performance test is done. The logic analyzer is considered performance-verified if all of the
software tests pass and the manual performance test meets specifications.
The specifications for the 16860 series logic analyzers define a maximum state data rate at which
data can be acquired in state mode. The manual performance test (maximum state data rate test)
measures the performance and verifies that the logic analyzer meets these specifications.
This logic analyzer supports 2 modes: Multiple Clocks and Single Clock. This chapter provides
information on measuring performance in both of these modes.
Test Strategy
Only specified parameters are tested. Specifications are listed on page 15. The test conditions
defined in this procedure ensure that the specified parameters are as good as or better than
specifications. Not all channels of the logic analyzer will be tested; a sample of channels is tested.
The calibration laboratory may choose to elaborate on these tests and test all channels at their
discretion.
A 16860 series analyzer that is licensed with the Base state clock option (350MHz) needs to be
tested at a Keysight Service Center. The Service Center has the capability to test the analyzer at up
to the 1.4Gb/s state speed to ensure that the calibration will remain valid even after upgrading it
to the 700MHz license.
Eye Scan is used to adjust the sampling position on every channel. Eye scan must be used to achieve
maximum state data rate performance.
Page 24
3Testing 16860 Performance
The 16860 series logic analyzer supports 2 modes: Multiple Clocks and Single Clock. Performance of
the two clocking modes of the Logic Analyzer will be measured separately. In the multiple clock
mode, performance of all 4 clocks will be measured, except for the 16861A only 2 pods. In the single
clocking mode, data from up to 8 pods (depending on model) will be verified, one pod at a time.
When the logic analyzer acquires data on both edges of the clock, the generator test frequency is set
to half of the acquisition speed.
Test Interval
Test the performance of the 16860 series logic analyzer against its specifications at two-year
intervals.
Test Record Description
A Performance Test Record for recording the results of each procedure is provided in this chapter.
You may want to make copies of the form, and fill-in a copy each time you test the logic analyzer.
Test Equipment
A list of the recommended test equipment is provided. You can use any equipment that satisfies the
specifications given. However, the instructions are written with the presumption that you are using
the recommended test equipment.
24Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 25
Perform the Self-Tests
Power the 16860 series analyzer on and start the Logic Analyzer application.
1 Before performing the self- tests, disconnect all probes from the logic analyzer.
2Select Help->Self-Test... from the main menu. The Analysis System Self Tests window appears.
Testing 16860 Performance3
3From the Select suites list, select the option displayed for 16860. Then, select All from the Select
tests list.
4Select Start to start a complete module self-test. The progress of the self tests is displayed in the
Results area of the window.
5 When the self-tests are complete, check the Results window to ensure that the Result Summary
says that all tests passed. If all tests did not pass, refer to “To use the logic acquisition
troubleshooting flowcharts" on page 87.
6 Select the Close button to close the Analysis System Self Tests window.
7 If all self- tests pass, then record “PASS” in the “Logic Analysis System Self-Tests” section of the
Performance Test Record (page 77).
Keysight 16860 Series Portable Logic Analyzer Service Guide25
Page 26
3Testing 16860 Performance
Equipment Required for the Performance Test
The following equipment is required for the performance test procedure.
150 ps Transition Time Converter (Qty 4)Required if pulse generator's rise time is less than
Flying Lead Probe Set (Qty 2)A combination of U4201A & E5382A can be used.Keysight U4203A
SMA/Flying Lead test connectors, (f) SMA to (f) SMA
to Flying Lead Probe (Qty 4)
≥ 800 MHz, two channels, differential outputs,
150-180 ps rise/fall time (if faster, use transition
time converters)
150 ps. (Pulse generator conditions: Voffset=1V,
δV=250 mV.) Required for 81134A
no substituteSee “Assemble the SMA/Flying Lead Test
Keysight 81134A or equivalent
Keysight 15435A
Connectors" on page 27
26Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 27
Testing 16860 Performance3
Assemble the SMA/Flying Lead Test Connectors
The SMA/Flying Lead test connectors provide a high-bandwidth connection between the logic
analyzer and the test equipment. The following procedure explains how to fabricate the four required
test connectors.
Table 3 Materials Required for SMA/Flying Lead Test Connectors
SMA Board Mount Connector (Qty 8)Cinch 142-0701-801
Pin Strip Header (Qty 1, which will be separated)0.100" X 0.100" Pin Strip Header, right angle, pin
length 0.230", two rows, 0.110" solder tails, 2 X 40
contacts
SMA 50 ohm terminators (Qty 4)Minimum bandwidth 2 GHzCinch 142-0801-866 50 ohm Dummy Load Plug or
SMA m-m adapter (Qty 4)Cinch 142-0901-811 SMA Plug to Plug or similar
(see www.cinch.com )
3M 2380-5121TG or similar 2- row with 0.1" pin
spacing
similar
1 Prepare the pin strip header:
a Cut or cleanly break a 2 x 2 section from the pin strip.
b Trim about 1.5 mm from the pin strip inner leads and straighten them so that they touch the
outer leads.
c Trim about 2.5 mm from the outer leads.
d Using a very small amount of solder, tack each inner lead to each outer lead at the point where
they are touching.
Keysight 16860 Series Portable Logic Analyzer Service Guide27
Page 28
3Testing 16860 Performance
2 Solder the pin strip to the SMA board mount connector:
3 Attach the second SMA board mount connector:
a Solder the leads on the left side of the pin strip to the center conductor of the SMA connector
as shown in the diagram below.
b Solder the leads on the right side of the pin strip to the inside of the SMA connector's frame as
shown in the diagram below. Use a small amount of solder.
a Re-heat the solder connection made in the previous step, and attach the second SMA
connector, as shown in the diagram below. Note that the second SMA connector is
upside-down, compared to the first. Add a little solder to make a good connection.
b Solder the center conductor of the second SMA connector to the center conductor of the first
SMA connector and the leads on the left side of the pin strip.
c Rotate the assembly 180 degrees and solder the two SMA board mount connector frames
together.
4 Check your work:
a Ensure that the following four points have continuity between them: The two pins on the left
side of the pin strip, and the center conductors of each SMA connector.
b Ensure that there is continuity between each of the two pins on the right side of the pin strip,
and the SMA connector frames.
c Ensure that there is NO continuity between the SMA connector center conductor and the SMA
connector frame (ground).
5 Finish creating the test connectors:
a Attach an SMA m-m adapter to one end of each of the four SMA/Flying Lead test connectors.
b Attach a 50 ohm terminator to the other end of the four SMA/Flying Lead test connectors.
28Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 29
c The finished test connector is shown in the pictures below.
Testing 16860 Performance3
Keysight 16860 Series Portable Logic Analyzer Service Guide29
Page 30
3Testing 16860 Performance
Set Up the Test Equipment
This section explains how to set up the test equipment for the maximum state data rate test.
1 Connect Transition Time Converters (if required) to each of the four outputs of the pulse
2 Connect the four SMA/Flying Lead test connectors (see “Assemble the SMA/Flying Lead Test
Connectors" on page 27) with 50 ohm terminators to the Transition Time Converters on the 4
pulse generator outputs. (If Transition Time Converters are not required, connect the SMA/Flying
Lead test connectors directly to the pulse generator outputs.)
3 Turn on the Pulse Generator. Let all of the test equipment and the logic analyzer warm up for 30
minutes before beginning any test.
4 Load the default configuration into the 81134A Pulse Generator.
•Select Main
•Hit Recall
• Press 0
5 Setup the pulse generator according to the following.
a Set the frequency of the pulse generator:
In this test procedure, the logic analyzer uses both edges of the clock to acquire data. The test
frequency is half the test clock rate because data is acquired on both the rising edge and the
falling edge of the clock. For the multiple clocks mode, set the frequency to:
•Base Option: Temporarily License the analyzer to full speed and test at 700 Mb/s plus 2%
(714 Mb/s). Set the generator to half this or 357 MHz.
30Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 31
•Option 700: Test at 700 Mb/s plus 2% (714 Mb/s). Set the generator to half this or 357
MHz.
This includes the frequency uncertainty of the pulse generator, cabling, and a test margin. If
you are using an 81134A pulse generator, the frequency accuracy is ±0.005% of setting.
b Set the rest of the pulse generator parameters to the values shown in the following tables.
Clock InternalDelay Ctrl Input OffDelay Ctrl Input Off
Delay 0 psDelay 0 ps
Pulse Perf: NormalPulse Perf: Normal
Deskew: 0 psDeskew: 0 ps
Levels: Normal, CustomLevels: Normal, Custom
Ampl: 450 mVAmpl: 450 mV
Testing 16860 Performance3
Offset: 0 mVOffset: 0 mV
Term Voltage: 0 mVTerm Voltage: 0 mV
Limit to current Levels: unselectedLimit to current Levels: unselected
Output: Enable (LED on)Output: Enable (LED on)
Output: Enable (LED on)Output: : Enable (LED on)
Keysight 16860 Series Portable Logic Analyzer Service Guide31
Page 32
3Testing 16860 Performance
32Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 33
Connect the Test Equipment
ground clip
NOTE
Connect the Logic Analyzer Pod to the Pulse Generator
1 Connect one U4203A Flying Lead Probe Set to Pods 1/2 of the 16860 series logic analyzer.
2 If not a 16861A, connect one U4203A Flying Lead Probe Set to Pods 3/4 of the analyzer.
3 Connect the Pod 1 U4203A Flying Lead Probe Set's CLK lead to the pin strip of the SMA/Flying
Lead connector at the pulse generator's Channel 1 OUTPUT.
Testing 16860 Performance3
On all connections, be sure to use the black ground clip (supplied with the U4203A Flying Lead
Probe Set) and orient the leads so that the black clip is connected to one of the SMA/Flying
Lead connector's ground pins!
4 Connect the Pod 1 U4203A Flying Lead Probe Set's CLK (NOT) lead to the SMA/Flying Lead
connector at the pulse generator's Channel 1 OUTPUT (NOT). Again, be sure to use the black
ground clip and orient the leads so that the black clip is connected to ground.
5 Connect the Pod 1 U4203A Flying Lead Probe Set's bits 2 and 10 to the SMA/Flying Lead test
connector's pin strip connector at the pulse generator's Channel 2 OUTPUT.
6 Connect the Pod 1 U4203A Flying Lead Probe Set's bits 6 and 14 to the SMA/Flying Lead test
connector's pin strip connector at the pulse generator's Channel 2 OUTPUT (NOT).
Keysight 16860 Series Portable Logic Analyzer Service Guide33
Page 34
3Testing 16860 Performance
34Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 35
Test the 16860 series Logic Analyzer
NOTE
The following sections explain how to measure the maximum state data rate.
1 Record the logic analyzer's model and serial number in the Performance Test Record (see
page 77). Record your work order number (if applicable) and today's date.
2 Record the test equipment information in the “Test Equipment Used” section of the Performance
Tes t Record .
3 Turn on the frame.
Before testing the performance of the logic analyzer, warm-up the logic analyzer and the test
equipment for 30 minutes.
a Plug in the power cord to the power connector on the rear panel of the logic analyzer.
b Connect a key board and a mouse to the frame.
c Press the Standby (Power) button on the front panel of the logic analyzer to power on the logic
analyzer.
While the logic analyzer is booting, observe for the following:
• Ensure all of the installed memory is recognized.
• Any error messages.
• Interrupt of the boot process with or without error message.
4 During initialization, check for any failures.
If an error or an interrupt occurs, refer to the chapter “Troubleshooting" on page 81.
5 Start the Keysight Logic Analyzer application if it is not started already.
Testing 16860 Performance3
Keysight 16860 Series Portable Logic Analyzer Service Guide35
Page 36
3Testing 16860 Performance
Create Test Configuration Files
The test consists of multiple setups to test all the various analyzer modes. To make this easier, 4
configuration files will be created. The configuration files will then be used as starting points for the
measurement setups.
The analyzer has two state modes: Multiple Clocks and Single Clock. Two configuration files will be
created for each of these modes. One for finding the maximum clock frequency (no Marker) and
another to verify that the data is correct (with Marker).
Create Configuration File 1 - Multiple Clocks (no Marker)
1 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods also.
2 In the Keysight Logic and Protocol Analyzer application, choose File > New. This puts the logic
analyzer into its initial state.
3 From the main drop down menu, select Setup then select Bus/Signal. The Analyzer Setup dialog is
displayed.
4 In the Analyzer Setup dialog, click the Threshold button for Pod 1.
5The Threshold Setting dialog is displayed. Set the Threshold to 0 V.
36Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 37
Testing 16860 Performance3
6 Apply this to all Pods and Clocks by clicking the Apply to All Other Pods and Clocks button.
7Click Done to close the Threshold dialog.
8 Select the Sampling tab.
9In the Sampling dialog, set the Acquisition mode to State - Synchronous Sampling.
10 Set the State Options to Multiple Clocks.
Keysight 16860 Series Portable Logic Analyzer Service Guide37
Page 38
3Testing 16860 Performance
11 Set Pod1 Clock to Both Edges. Set other clocks to Don't Care.
12 Set the Trigger Position to 100% Poststore. Set Acquisition Depth to 256K.
13 Close the Sampling dialog by clicking OK.
14 Select the Listing Window by selecting Window and then Listing.
38Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 39
15 Save the configuration by selecting File then Save As ….
16 In the Save As dialog, type the file name MultipleClocks_NoMarker.ala. For file options, select Setup
1 Start with the MultipleClocks_NoMarker configuration file. Select File then Open. In the Open dialog
select the file and then open. If asked about saving the current configuration, click No.
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods. The Pod 1 leads should be connected to the generator as described
above.
3 From the main drop down menu, select Setup and then select Bus/Signal. The Analyzer Setup dialog
is displayed.
4 The activity indicators now show activity on the channels that are connected to the pulse
generator. Un-assign all channels. You can do this quickly by clicking on the left-most check
mark and dragging to the right across all of the other check marks.
5 Assign bits that are connected, Bits 2, 6, 10, 14.
Keysight 16860 Series Portable Logic Analyzer Service Guide39
Page 40
3Testing 16860 Performance
6 Close the Analyzer Setup Dialog Window by clicking OK.
7Click the Run button. The analyzer needs to have data before the Marker can be added.
8 Add a new marker by selecting Marker and then New.
9In the New Marker dialog, change the Position information. Select Value in the drop down.
10 Open the Value dialog by clicking the Occurs button.
11 In the Value dialog, create a marker that looks like the following.
40Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 41
Testing 16860 Performance3
12 In the Value dialog, select the Properties... button.
13 In the Value Properties dialog, select Stop repetitive run when value is not found.
14 Close the Marker dialogs by clicking OK several times.
15 Save the configuration by selecting File then Save As ….
16 In the Save As dialog, type the file name MultipleClocks_WithMarker.ala. For file options select Setup
Only. Then click Save.
Keysight 16860 Series Portable Logic Analyzer Service Guide41
Page 42
3Testing 16860 Performance
Create Configuration File 3 - Single Clock (no Marker)
1 Start with the MultipleClocks_NoMarker configuration file. Select File then Open. In the Open dialog
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
3Click the Analyzer Setup > Sampling tab.
select the file and then open. If asked about saving the current configuration, click No.
U4203A to these pods also. The Pod 1 leads should be connected to the generator as described
above.
4In the State Options section, change the Sampling options to Single Clock.
5 In the clock selection area, change the Pod 1 clock to Both Edges.
6 Close the dialog by clicking OK.
7 Save the configuration by selecting File then Save As ….
42Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 43
8In the Save As dialog, type the file name SingleClock_NoMarker.ala. For file options, select Setup
Only. Then click Save.
Create Configuration File 4 - Single Clock with Marker
1 Start with the MultipleClocks_WithMarker configuration file. Select File and then Open. In the Open
dialog select the file then open. If asked about saving the current configuration, click No.
2 Ensure that there is a U4203A connected to Pods 1 and 2. If Pods 3 and 4 exist, connect a
U4203A to these pods. The Pod 1 leads should be connected to the generator as described
above.
3Open the Analyzer Setup Sampling tab.
Testing 16860 Performance3
4In the State Options section, change the Sampling options to Single Clock.
5 In the clock selection area, change the Pod 1 clock to Both Edges.
6 Close the dialog by clicking OK.
7 Save the configuration by selecting File then Save As ….
Keysight 16860 Series Portable Logic Analyzer Service Guide43
Page 44
3Testing 16860 Performance
8In the Save As dialog, type the file name SingleClock_WithMarker.ala. For file options, select Setup
Only. Then click Save.
44Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 45
Determine Maximum Data Rate for Multiple Clocks Mode
The multiple clocks test measures the maximum data rate of each of the four clocks used for this
mode: Pod1, Pod2, Pod3, Pod4. Each Pod / clock is measured separately. The measurement is done
in two parts:
• The first part is to determine the maximum clock rate that the analyzer will run at.
• The second part is to verify that the data captured by the analyzer is correct; markers are used to
verify the data patterns.
Pod 1 Clock - Maximum Clock Rate
For this test, Pod 1 clock should be connected to Channel 1 of the 81134A.
1Load the MultipleClocks_NoMarker.ala configuration file. If asked to save the current configuration
file, click No.
Testing 16860 Performance3
2 Verify that the Generator is set to 357 MHz as a starting point.
3Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly.
4 Acquired data will start appearing in the Listing window.
5 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
6 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
7 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
Keysight 16860 Series Portable Logic Analyzer Service Guide45
Page 46
3Testing 16860 Performance
8 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
9 Continue changing the frequency and running unit the frequency can no longer be increased and
10 Set the generator to the highest frequency that did not cause an error.
11 Click Run Repetitive .
12 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
13 Click the Stop toolbar button to stop the data acquisition.
14 Note the generator frequency setting this will be used in the next section to verify the data rate.
Pod 1 Clock - Setup for Maximum Data Rate
To measure the maximum data rate, the starting frequency will be the frequency found in the section
above. A new configuration file will be used that has a marker setup to verify that the data patterns
are correct. To ensure that the data is sampled correctly, the sample position needs to be set. Eye
Scan is used to set the sample points of all the data to the same eye., The following section describes
how to do use Eye Scan to do this. This same process will be used during all the other data
verification steps later in the procedure.
1Load the MultipleClocks_WithMarker.ala configuration file. If asked to save the current configuration
does not display an error, increase the frequency by 1 MHz and run again.
the error message not displayed.
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
file click No.
2 Verify that the Generator is set to the frequency found in the last section as a starting point.
3 Adjust the sample position using the procedure in the following section.
Adjust Sampling Positions using Eye Scan
This Eye Scan procedure will be used before each of the data rate verification steps. This procedure
should be followed any time the setup or frequency is changed.
1 Open the sampling setup window.
2 Select the Eye Scan: Sample Positions and Thresholds... button. The Eye Scan - Sample Positions and
Threshold Settings dialog appears.
46Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 47
Testing 16860 Performance3
3 Select the Assign Busses/Signals.
4 In the Buses/Signals section of the dialog, ensure that the check box next to My Bus 1 is checked
and the 4 data bits 0-3. Click OK.
5Select Full time/voltage scan.
Keysight 16860 Series Portable Logic Analyzer Service Guide47
Page 48
3Testing 16860 Performance
6 In the dialog, set Do full time/voltage scan and Show time/voltage diagram.
7 Run Eye Scan by clicking the Run This Measurement button.
8 If the scan does not fill the scan area, right click in the scan area and select Autoscale all.
The waveform should now fill the Eye Scan area.
9 Move the sample position of all 4 data bits into the Eye between 0 and -1ns. Using the mouse,
grab the vertical lines and move them into the correct eye.
48Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 49
Testing 16860 Performance3
10 If all sample positions are not in the same eye, then click the Plus sign to expand My Bus 1.
11 Use the mouse to grab the bits that are not in the correct eye and move them into the right eye.
12 After all bit sample positions are in the correct eye, right-click in the My Bus 1 Scan area and
select Set to Suggested.
Keysight 16860 Series Portable Logic Analyzer Service Guide49
Page 50
3Testing 16860 Performance
13 After setting to suggested, all bits should be centered in their own eye near -.5 ns.
14 Close Eye Scan by clicking OK.
15 Close the Sampling dialog by clicking OK.
Pod 1 Clock - Measuring Maximum Data Rate
After setting the sample positions, the data capture can now be verified at the frequency found in the
clock rate section. The configuration file that was loaded earlier with markers is set up to verify the
data pattern. If bad data is found, the marker will cause the repetitive run to stop.
1 Select the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
50Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 51
3 If the clock rate is too high, this error message may occur.
NOTE
Testing 16860 Performance3
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan run and
the test rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
If bad data is found or an error message occurs, decrease the pulse generator frequency by 1
MHz and rerun Eye Scan to find the sample position. Run the analyzer again looking for the
appropriate number of A's and 5's. After 1 minute, stop the analyzer. Repeat these steps until
you get acquisitions without any error display.
6 After the analyzer runs for about 1 minute, select the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For this Pod's Clock, record the generator frequency and the Data Rate in the Maximum State
Data Rate section of the “Performance Test Record" on page 77. Remember that the data rate is
twice the generator frequency.
Pod 2 Clock - Maximum Clock Rate
1Load the MultipleClocks_NoMarker.ala configuration file. If asked to save the current configuration
file, click No.
2 Disconnect the U4203A Flying Lead Probe Set from channels 1 & 2 of the 81134A pulse
generator output (Bits 2, 6, 10, 14) and clock leads.
3 Connect the probe set from Pod 2 of logic analyzer to the pulse generator channels 1 & 2 output.
• Clock to Channel 1 Output
• Clock (NOT) Channel 1 Output (not)
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
Keysight 16860 Series Portable Logic Analyzer Service Guide51
Page 52
3Testing 16860 Performance
4 Open the Sampling Tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
5 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 2 Clock to Both Edges.
6 Close the dialog by clicking OK.
7 Set the generator frequency to 357 MHz as a starting point.
8Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
9 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
acquiring data repeatedly.Acquired data will start appearing in the Listing window.
observing the logic analyzer data acquisition status.
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
11 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error increase the frequency by 1 MHz and run again.
13 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
14 Set the generator to the highest frequency that did not cause an error.
15 Click Run Repetitive .
16 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
52Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 53
17 Click the Stop toolbar button to stop the data acquisition.
18 Note the generator frequency setting this will be used in the next section to verify the data rate.
Pod 2 Clock - Setup for Maximum Data Rate
1 Verify that the Generator is set to the frequency found in the last section as a starting point.
2Load the MultipleClocks_WithMarker.ala configuration file. If asked to save the current configuration
file, click No.
3Open the Bus/Signal Setup dialog by clicking the Setup Icon in the tool bar.
4 Unassign the data bits from Pod1.
5 Assign bits 2, 6, 10, and 14 of Pod 2.
Testing 16860 Performance3
6Open the Sampling tab.
7 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 2 Clock to Both Edges.
8 Adjust the sample position using the procedure described earlier by clicking the Eye Scan: Sample
Position and Thresholds button.
Keysight 16860 Series Portable Logic Analyzer Service Guide53
Page 54
3Testing 16860 Performance
9 Close the dialog boxes by clicking OK.
Pod 2 Clock - Measuring Maximum Data Rate
1Click the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window appears. The data in the listing window should be
A's & 5's.
3 If the clock rate is too high, this error message may occur.
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
be rerun, and the test should be rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, click the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
54Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 55
7 For this Pod's Clock, record the generator frequency and the Data Rate in the "Maximum State
NOTE
Data Rate" section of the “Performance Test Record" on page 77. Remember that the data rate is
twice the generator frequency.
Pod 3 Clock - Maximum Clock Rate
1Load the MultipleClocks_NoMarker.ala configuration file. If asked to save the current configuration
file, click No.
2 Disconnect the U4203A Flying Lead Probe Set from channels 1 & 2 of the 81134A pulse
generator output (Bits 2, 6, 10, 14) and clock leads.
3 Connect the probe set from Pod 3 of logic analyzer to the pulse generator channels 1 & 2 output.
• Clock to Channel 1 Output
• Clock (NOT) Channel 1 Output (not)
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
4Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
Testing 16860 Performance3
This section is only applicable to models: 16862A, 16863A & 16864A.
5 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 3 Clock to Both Edges.
6 Close the dialog by clicking OK.
7 Set the generator frequency to 357 MHz as a starting point.
8Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
9 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
Keysight 16860 Series Portable Logic Analyzer Service Guide55
Page 56
3Testing 16860 Performance
NOTE
11 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
13 Continue changing the frequency and running unit the frequency can no longer be increased and
14 Set the generator to the highest frequency that did not cause an error.
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
does not display an error, increase the frequency by 1 MHz and run again.
the error message not displayed.
15 Click Run Repetitive .
16 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
17 Click the Stop toolbar button to stop the data acquisition.
18 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Pod 3 Clock - Setup for Maximum Data Rate
This section is only applicable to models: 16862A, 16863A & 16864A.
1 Verify that the Generator is set to the frequency found in the last section as a starting point.
2Load the MultipleClocks_WithMarker.ala configuration file. If asked to save the current
configuration file, click No.
3Open the Bus/Signal Setup dialog by clicking the Setup Icon in the tool bar.
56Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 57
4 Unassign the data bits from Pod1.
5 Assign bits 2, 6, 10, and 14 of Pod 3.
6Open the Sampling tab.
Testing 16860 Performance3
7 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 3 Clock to Both Edges.
8 Adjust the sample position using the procedure described earlier by clicking the Eye Scan:
Sample Position and Thresholds button.
9 Close the dialog windows by clicking OK.
Pod 3 Clock - Measuring Maximum Data Rate
Keysight 16860 Series Portable Logic Analyzer Service Guide57
Page 58
3Testing 16860 Performance
NOTE
1 Select the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
2 If incorrect data is found, the following window will appear. The data in the listing window should
This section is only applicable to models: 16862A, 16863A & 16864A.
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
be A's & 5's.
3 If the clock rate is too high, this error message may occur.
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
be rerun, and the test should be rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, click the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For this Pod's Clock, record the generator frequency and the Data Rate in the "Maximum State
Data Rate" section of the “Performance Test Record" on page 77. Remember that the data rate is
twice the generator frequency.
58Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 59
Pod 4 Clock - Maximum Clock Rate
NOTE
1Load the MultipleClocks_NoMarker.ala configuration file. If asked to save the current configuration
file, click No.
2 Disconnect the U4203A Flying Lead Probe Set from channels 1 & 2 of the 81134A pulse
generator output (Bits 2, 6, 10, 14) and clock leads.
3 Connect the probe set from Pod 4 of logic analyzer to the pulse generator channels 1 & 2 output.
• Clock to Channel 1 Output
• Clock (NOT) Channel 1 Output (not)
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
4Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
5 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 4 Clock to Both Edges.
Testing 16860 Performance3
This section is only applicable to models: 16862A, 16863A & 16864A.
6 Close the dialog window by clicking OK.
7 Set the generator frequency to 357 MHz as a starting point.
8Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
9 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
11 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error, increase the frequency by 1 MHz and run again.
13 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
14 Set the generator to the highest frequency that did not cause an error.
15 Click Run Repetitive .
Keysight 16860 Series Portable Logic Analyzer Service Guide59
Page 60
3Testing 16860 Performance
NOTE
16 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
17 Click the Stop toolbar button to stop the data acquisition.
18 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Pod 4 Clock - Setup for Maximum Data Rate
1 Verify that the Generator is set to the frequency found in the last section as a starting point.
2Load the MultipleClocks_WithMarker.ala configuration file. If asked to save the current
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
This section is only applicable to models: 16862A, 16863A & 16864A.
configuration file, click No.
3Open the Bus/Signal Setup dialog by clicking the Setup Icon in the tool bar.
4 Unassign the data bits from Pod1.
5 Assign bits 2, 6, 10, and 14 of Pod 4.
6Open the Sampling tab.
7 In the clock assignment area, set Pod 1 Clock to Don't Care and set Pod 4 Clock to Both Edges.
60Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 61
8 Adjust the sample position using the procedure described earlier by selecting "Eye Scan: Sample
NOTE
Position and Thresholds".
9 Close the dialog windows by clicking OK.
Pod 4 Clock - Measuring Maximum Data Rate
This section is only applicable to models: 16862A, 16863A & 16864A.
Testing 16860 Performance3
1 Select the Run Repetitive icon . Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window appears. The data in the listing window should be
A's & 5's.
3 If the clock rate is too high, the following error message may occur.
Keysight 16860 Series Portable Logic Analyzer Service Guide61
Page 62
3Testing 16860 Performance
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, select the Stop button to stop the acquisition. If
7 For this Pod's Clock, record the generator frequency and the Data Rate in the "Maximum State
be rerun and the test should be rerun.
the "can't find occurrence" window does not appear, then the analyzer has found good data.
Data Rate" section of the “Performance Test Record" on page 77. Remember that the data rate is
twice the generator frequency.
62Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 63
Determine Maximum Data Rate for Single Clock Mode
The single clock test measures the maximum data rate for the single clock mode. This test measures
data rates for Rising, Falling, and Both Edge modes. Using Both Edge clocking, it verifies data rates
on all pods. The measurement is done in two parts:
• The first part is to determine the maximum clock rate that the analyzer will run at.
• The second part is to verify that the data captured by the analyzer is correct; markers are used to
verify the data patterns.
Single Clock Rising Edge - Maximum Clock Rate
1Load the SingleClock_NoMarker.ala configuration file. If asked to save the current configuration file,
click No.
Testing 16860 Performance3
2 Verify that the Generator is set to 714 MHz as a starting point.
3 Disconnect the U4203A Flying Lead Probe Set from channels 1 & 2 of the 81134A pulse
generator output (Bits 2, 6, 10, 14) and clock leads.
4 Connect the probe set from Pod 1 of logic analyzer to the pulse generator channels 1 & 2 output.
• Clock to Channel 1 Output
• Clock (NOT) Channel 1 Output (not)
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
5 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
6 In the clock assignment area, set Pod 1 Clock to Rising Edge.
7 Close the dialog windows by clicking OK.
Keysight 16860 Series Portable Logic Analyzer Service Guide63
Page 64
3Testing 16860 Performance
8Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
9 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
10 Continue increasing the generator frequency until the logic analyzer displays an error that the
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
observing the logic analyzer data acquisition status.
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
11 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
12 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error, increase the frequency by 1 MHz and run again.
13 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
14 Set the generator to the highest frequency that did not cause an error.
15 Click Run Repetitive .
16 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
17 Click the Stop toolbar button to stop the data acquisition.
18 Note the generator frequency setting this will be used in the next section to verify the data rate.
64Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 65
Single Clock Rising Edge - Pod 1 Data - Setup for Maximum Data Rate
To measure the maximum data rate, the starting frequency will be the frequency found in the section
above. A new configuration file will be used that has a marker setup to verify that the data patterns
are correct. To ensure that the data is sampled correctly, the sample position needs to be set. Eye
Scan is used to set the sample points of all the data to the same eye.
1 On the 81134A, select the channel 2 setup screen. This is the data channel.
2 Set the Channel 2 Freq. Divider to 2.
3Load the SingleClock_WithMarker.ala configuration file. If asked to save the current configuration
file, click No.
Testing 16860 Performance3
4 Verify that the Generator is set to the frequency found in the last section as a starting point.
5 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
6 In the clock assignment area, set Pod 1 Clock to Rising Edge.
7 Adjust the sample position using the procedure described earlier by clicking the Eye Scan: Sample
Position and Thresholds button.
8 Close the dialog windows by clicking OK.
Keysight 16860 Series Portable Logic Analyzer Service Guide65
Page 66
3Testing 16860 Performance
Single Clock Rising Edge - Pod 1 Data - Measuring Maximum Data Rate
1 Select the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
2 If incorrect data is found, the following window will appear. The data in the listing window should
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
be A's & 5's.
3 If the clock rate is too high, this error message may occur.
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan rerun
and the test rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute select the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For Single Clock for Rising Edge record the generator frequency and the Data Rate in the
"Maximum State Data Rate" section of the “Performance Test Record. Note: For Single Edge
Clocking, the data rate is the same as the generator frequency.
66Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 67
Single Clock Falling Edge - Maximum Clock Rate
1Load the SingleClock_NoMarker.ala configuration file.If asked to save the current configuration file,
click No.
2 Verify that the Generator is set to 714 MHz as a starting point.
3 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
Testing 16860 Performance3
4 In the clock assignment area, set Pod 1 Clock to Falling Edge.
5 Close the dialog windows by clicking OK.
6Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly. Acquired data will start appearing in the Listing window.
7 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
8 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
Keysight 16860 Series Portable Logic Analyzer Service Guide67
Page 68
3Testing 16860 Performance
9 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
10 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
11 Continue changing the frequency and running unit the frequency can no longer be increased and
12 Set the generator to the highest frequency that did not cause an error.
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
does not display an error, increase the frequency by 1 MHz and run again.
the error message not displayed.
13 Click Run Repetitive .
14 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
15 Click the Stop toolbar button to stop the data acquisition.
16 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Single Clock Falling Edge - Pod 1 Data - Setup for Maximum Data Rate
1 On the 81134A, verify that Channel 2 Freq Divide is set to 2.
2Load the SingleClock_WithMarker.ala configuration file. If asked to save the current configuration
file, click No.
3 Verify that the Generator is set to the frequency found in the last section as a starting point.
68Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 69
Testing 16860 Performance3
4 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
5 In the clock assignment area, set Pod 1 Clock to Falling Edge.
6 Adjust the sample position using the procedure described earlier by clicking the Eye Scan: Sample
Position and Thresholds button.
7 Close the dialog windows by clicking OK.
Single Clock Falling Edge - Pod 1 Data - Measuring Maximum Data Rate
1 Select the Run Repetitive icon. Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
3 If the clock rate is too high, this error message may occur.
Keysight 16860 Series Portable Logic Analyzer Service Guide69
Page 70
3Testing 16860 Performance
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan rerun
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute select the Stop button to stop the acquisition. If
7 For Single Clock for Falling Edge, record the generator frequency and the Data Rate in the
and the test rerun.
the "can't find occurrence" window does not appear, then the analyzer has found good data.
"Maximum State Data Rate" section of the “Performance Test Record. Note: For Single Edge
Clocking, the data rate is the same as the generator frequency.
Single Clock Both Edges - Maximum Clock Rate
1Load the SingleClock_NoMarker.ala configuration file. If asked to save the current configuration file,
click No.
2 Verify that the Generator is set to 714 MHz as a starting point.
3 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
4 In the clock assignment area, set Pod 1 Clock to Both Edges.
70Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 71
Testing 16860 Performance3
5 Close the dialog windows by clicking OK.
6Click the Run Repetitive toolbar button to start a repetitive run on the logic analyzer for
acquiring data repeatedly.Acquired data will start appearing in the Listing window.
7 Start increasing the frequency on the pulse generator by 1 MHz increments while simultaneously
observing the logic analyzer data acquisition status.
8 Continue increasing the generator frequency until the logic analyzer displays an error that the
data could not be displayed, or the clock is too fast. It is possible that the error is caused by noise
introduced as the generator frequency is changed.
9 Without changing the frequency, clear the message by clicking OK. Then click the Run Repetitive
toolbar button to start a repetitive run again at the same frequency. Running at the same
frequency checks to see if the error message was caused by noise due to the changing of the
generator frequency.
10 If the analyzer displays an error again, lower the frequency by 1MHz and run again. If the test
does not display an error, increase the frequency by 1 MHz and run again.
11 Continue changing the frequency and running unit the frequency can no longer be increased and
the error message not displayed.
12 Set the generator to the highest frequency that did not cause an error.
13 Click Run Repetitive.
14 Wait for logic analyzer to complete 100 acquisitions at the new pulse generator frequency
without displaying any error. If an error is displayed, decrease the pulse generator frequency by 1
MHz and then again wait for 100 acquisitions at this new frequency without any error. Repeat this
step until you get 100 acquisitions without any error display.
Keysight 16860 Series Portable Logic Analyzer Service Guide71
Page 72
3Testing 16860 Performance
15 Click the Stop toolbar button to stop the data acquisition.
16 Note the generator frequency setting. This will be used in the next section to verify the data rate.
Single Clock Both Edges - Pod 1 Data - Setup for Maximum Data Rate
1 On the 81134A, select the channel 2 setup screen. This is the data channel.
2 Set the Channel 2 Freq. Divider to 1.
3Load the SingleClock_WithMarker.ala configuration file. If asked to save the current configuration
file, click No.
4 Verify that the Generator is set to the frequency found in the last section as a starting point.
5 Open the Sampling tab in the Analyzer Setup dialog by clicking the Sampling Setup Icon.
6 Adjust the sample position using the procedure described earlier by clicking the Eye Scan: Sample
Position and Thresholds button. Select the eye closest to -1 ns.
7 Close the dialog windows by clicking OK.
Single Clock Both Edges - Pod 1 Data - Measuring Maximum Data Rate
1 Select the Run Repetitive icon . Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window will appear. The data in the listing window should
be A's & 5's.
72Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 73
3 If the clock rate is too high, this error message may occur.
NOTE
Testing 16860 Performance3
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
be rerun and the test should be rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, select the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For Single Clock for Both Edges Pod1, record the generator frequency and the Data Rate in the
"Maximum State Data Rate" section of the “Performance Test Record. Note: For Both Edges
Clocking, the data rate is twice the generator frequency.
Single Clock Both Edges - Pod 2 Data - Setup for Maximum Data Rate
The next set of tests use the Pod 1 clocks to verify the data rate on the other pods in Single Clock
mode.
The starting frequency for the following tests (Pods 2-8) will be the frequency
found in the "“Single Clock Both Edges - Maximum Clock Rate" section
above.
1 Verify that the Generator is set to the frequency found in “Single Clock Both Edges - Maximum
Clock Rate section above.
2 Disconnect the U4203A Flying Lead Probe Set from channels 2 of the 81134A pulse generator
output (Bits 2, 6, 10, 14). Do not disconnect the clock leads from Pod 1.
3 Connect the probe set from Pod 2 of logic analyzer to the pulse generator channels 2 outputs.
Keysight 16860 Series Portable Logic Analyzer Service Guide73
Page 74
3Testing 16860 Performance
4Open the Bus/Signal Setup dialog by clicking the Setup Icon in the toolbar.
5 Unassign the data bits from Pod1.
6 Assign bits 2, 6, 10, and 14 of Pod 2.
7Open the Sampling tab.
• Bits 2 & 10 to Channel 2 Output
• Bits 6 & 14 to Channel 2 Output (not)
8 Adjust the sample position using the procedure described earlier by clicking the Eye Scan: Sample
Position and Thresholds button. Select the eye closest to -1 ns.
9 Close the dialog windows by clicking OK.
Single Clock Both Edges - Pod 2 Data - Measuring Maximum Data Rate
1 Select the Run Repetitive icon . Let the logic analyzer run for about 1 minute. The analyzer will
acquire data and the Listing Window will continuously update. A marker search window will
appear and show progress.
2 If incorrect data is found, the following window appears. The data in the listing window should be
A's & 5's.
74Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 75
3 If the clock rate is too high, the following error message may occur.
Testing 16860 Performance3
4 If either of these messages occur, the generator frequency should be lowered, Eye Scan should
be rerun, and the test should be rerun.
5 Lower the generator frequency by 1 MHz and rerun the test.
6 After the analyzer runs for about 1 minute, select the Stop button to stop the acquisition. If
the "can't find occurrence" window does not appear, then the analyzer has found good data.
7 For Single Clock for Both Edges Pod 2, record the generator frequency and the Data Rate" in the
"Maximum State Data Rate" section of the “Performance Test Record. Note: For Both Edges
Clocking, the data rate is twice the generator frequency.
Single Clock Both Edges - Pods 3 to 8
For the product models that have more Pods, the data can be verified using the same procedure that
was used above for "Single Clock Both Edges - Pod 2 Data". The Pod 1 clock will be used for all other
pods. The U4203A cable will need to be moved from Pod 3 & 4 to the other pods as needed.When
this is done, the Thresholds will need to be set to 0V. Follow both the Setup and Measurement
sections for each pod, connecting to the generator and verifying the data patterns. Record each
measured frequency in the “Performance Test Record table for each Pod.
Keysight 16860 Series Portable Logic Analyzer Service Guide75
Page 76
3Testing 16860 Performance
Conclude the State Data Rate Tests
Perform the following steps to properly shut down the logic analyzer session after completing the
state mode tests.
End the test.
1 From the Main menu, select File > Exit. In the dialog "Do you want to save the current
configuration?", click No.
2 Disconnect all cables and adapters from the pulse generator.
3 Power down the Analyzer frame.
76Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 77
Performance Test Record
LOGIC ANALYZER MODEL NO.: 16861A, 16862A, 16863A, 16864A
Logic Analyzer Serial No.Work Order No.
Date:Recommended Test Interval - 2 Year/4000 hours
Recommended next testing:
TEST EQUIPMENT USED
Pulse Generator Model No.
Pulse Generator Serial No.
Pulse Generator Calibration Due Date:
MEASUREMENT UNCERTAINTY
Clock Rate
Testing 16860 Performance3
Pulse Generator Frequency Accuracy:
81134A: ±0.005% of setting.
Approx. Cabling Accuracy
±0.005% of setting.
2.01% = ±0.010% Uncertainty + 2% Test Margin.
Setting
Base option: (should be tested to the Option 700 level at the service
center)
Option 700:
• 700 MHz + 2% = 714 MHz (Single Clock mode)
• 350 MHz + 2% = 357 MHz (Multiple Clocks mode)
TEST RESULTS
Logic Analysis System Self-Tests (Pass/Fail):
TEST RESULTS - Maximum State Data Rate
Mode: Multiple Clocks (Both Edges)
Data Rate
(Spec)
(Mb/s)
Pod 1 Clock700
Pod 2 Clock700
Generator Frequency
(MHz)
Measured
Data Rate
(Mb/s)
Pod 3 Clock700
Pod 4 Clock700
Keysight 16860 Series Portable Logic Analyzer Service Guide77
Page 78
3Testing 16860 Performance
TEST RESULTS - Maximum State Data Rate
Mode: Single Clock
Data Rate
(Spec)
(Mb/s)
Clock - Rising Edge (Data Pod 1)700
Clock - Falling Edge (Data Pod 1)700
Clock - Both Edges (Data Pod 1)1400
Clock - Both Edges (Data Pod 2)1400
Clock - Both Edges (Data Pod 3)1400
Clock - Both Edges (Data Pod 4)1400
Clock - Both Edges (Data Pod 5)1400
Clock - Both Edges (Data Pod 6)1400
Clock - Both Edges (Data Pod 7)1400
Clock - Both Edges (Data Pod 8)1400
Generator Frequency
(MHz)
Measured
Data Rate
(Mb/s)
78Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 79
Keysight 16860 Series Portable Logic Analyzer
Service Guide
4Calibrating
Calibration Strategy / 80
This chapter provides instructions for calibrating the16860 series logic analyzer.
Page 80
4Calibrating
Calibration Strategy
The 16860 series logic analyzer does not require any periodic adjustments or calibration by the user
to ensure operational accuracy.
However, Keysight recommends that performance of the 16860 series logic analyzer be tested
against its specifications at two-year intervals. This testing is required in order to obtain calibration
certification.
You can refer to Chapter 3, “Testing 16860 Series Logic Analyzers Performance" to find detailed
information on how to test the performance of the 16860 series logic analyzer.
80Keysight 16860 Series Portable Logic Analyzer Service Guide
Page 81
Keysight 16860 Series Portable Logic Analyzers
CAUTION
Service Guide
5Troubleshooting
To use the system troubleshooting flowcharts / 82
To use the logic acquisition troubleshooting flowcharts / 87
To troubleshoot system power problems / 89
To run the self-tests / 90
To restore the system software / 93
To test the logic acquisition cables / 97
This chapter provides instructions for troubleshooting a 16860 logic analyzer that is not operating
correctly.
The troubleshooting consists of flowcharts, self-test instructions, a cable test, and how to restore
system software.
If you suspect a problem, start at the top of the first flowchart. During the troubleshooting
instructions, the flowcharts will direct you to perform the self-tests or the cable test.
The service strategy for the 16860 logic analyzer is the replacement of defective assemblies. You can
send this logic analyzer to Keysight Technologies for all service work, including troubleshooting.
Contact your nearest Keysight Technologies Sales Office for more details.
Keysight requires that the 16860 logic analyzer unit be repaired only at qualified repair facilities such
as the Keysight Repair Centers.
Electrostatic discharge can damage electronic components. Use grounded wrist
straps and mats when you perform any troubleshooting procedures to this
instrument.
Page 82
5Troubleshooting
To use the system troubleshooting flowcharts
Flowcharts are the primary tool used to isolate defective assemblies. The flowcharts refer to other
tests to help isolate the trouble. The circled references on the charts indicate connections with the
other flowcharts or other parts within the same flowchart. Start your troubleshooting at the top of
the first flowchart (Figure 1 on page 83).
82Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 83
Troubleshooting5
Figure 1System Troubleshooting Flowchart
Keysight 16860 Series Portable Logic Analyzers Service Guide83
Page 84
5Troubleshooting
Figure 2System Power Troubleshooting Flowchart
84Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 85
Troubleshooting5
Figure 3System Display Troubleshooting Flowchart
Keysight 16860 Series Portable Logic Analyzers Service Guide85
Page 86
5Troubleshooting
Figure 4System Boot Up Troubleshooting Flowchart
86Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 87
To use the logic acquisition troubleshooting flowcharts
Flowcharts are the primary tool used to isolate defective assemblies. The flowcharts refer to other
tests to help isolate the trouble. The circled numbers on the charts indicate connections with the
other flowchart. Start your troubleshooting at the top of the first flowchart.
If the instrument still doesn't work correctly after completing all the procedures described in the
flowchart, return it to Keysight Technologies for repair. Be sure to include a note describing the
problem in detail.
Troubleshooting5
Keysight 16860 Series Portable Logic Analyzers Service Guide87
88Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 89
To troubleshoot system power problems
If the system warns you it is powering down before it powers down, it is a fan/overtemp problem. If it
just powers down, it is a power supply problem.
If the lights do not come on and if the system powers up momentarily when you plug it in, make sure
the power button hasn't become jammed or stuck in the pushed-in position.
Power Supplies
All 16860 Series logic analyzers have the same 600 W power supply and a second 15 W power
supply. The power supplies must remain connected in order to test their output voltages. There are
power supply test points on the Frame Interface Board (FIB).
Troubleshooting5
Keysight 16860 Series Portable Logic Analyzers Service Guide89
Page 90
5Troubleshooting
To run the self-tests
The self-tests check the functional operation of the logic analyzer. Perform the self-tests as an
acceptance test when receiving the logic analyzer or when the logic analyzer is repaired.
1In the Keysight Logic and Protocol Analyzer application, click Help>Self Test...
2In the Analysis System Self Test dialog, double-click the test you want to run.
90Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 91
Logic Acquisition Self-Test Descriptions
The self-tests for the logic analyzer identify the correct operation of major functional areas in the
module.
Interface FPGA Version Test
This test verifies that the FPGA program is a version that the software can use. This is necessary
because new features will be added to the 16860 logic analyzer that will require both new software
and new FPGA bits.
Interface FPGA Register Test
The purpose of this test is to verify that the backplane interface can communicate with the backplane
FPGA. The FPGA must be working before any of the other circuits on the board will work. Also, the
FPGA generates the board ID code that is returned to identify the module and slot.
FPGA to FPGA Communication Test
This test is only run if there are two or more logic analyzer installed and connected together with the
flex cables. The purpose of this test is to verify that the FPGAs can drive and receive the signals
correctly.
SPI Bus Communication Test
Troubleshooting5
The purpose of this test is to verify communications over the SPI bus from the Interface FPGA to
various devices attached to the SPI bus.
EEPROM Test
The purpose of this test is to verify:
• The address and data paths to the EEPROM.
• That each cell in the EEPROM can be programmed high and low.
• That individual locations can be independently addressed.
• The EEPROM can be block erased.
Probe ID Read Test
The purpose of this test is to verify that the Probe ID values can be correctly read and to verify the
functionality of the Digital to Analog Converter by testing the two Probe ID DAC outputs at various
voltage levels.
Chip Registers Read/Write Test
The purpose of this test is to verify that each bit in each register of the Analysis chip can be written
with a 1 and 0 and read back again. The test also verifies that a chip reset sets all registers to their
reset condition (all 0s for most registers).
Freq Synth Lock Detect Test
This test determines if all the voltage-controlled oscillators (VCOs) are working properly.
Acquisition Chip BIST Test
Tests the Timing Zoom memory and other internal memories on the acquisition chip.
Resource Bus Connection Test
The purpose of this test is to verify global resources.
Keysight 16860 Series Portable Logic Analyzers Service Guide91
Page 92
5Troubleshooting
Comparator Programming Test
The purpose of this test is to verify the programming path to each of the comparators.
Comparator/DAC Test
This test is executed only if all probes are detached.
This test uses the pod, bonus, and calibration DACs, the calibration oscillator (implemented in the
interface FPGA), the comparators, the connections between the comparators and the Analysis chip,
and the activity indicators in the Analysis chip. We verify that we can use the DACs to control the
data input to the comparators. We verify that each comparator data channel produces output. We
verify that each comparator output is connected to each ASIC data input.
Comparator Delay Test
The comparator delay test verifies the integrity of all the delay line elements for each delay line in the
comparators. Each delay line consists of 11 delay elements. When set for maximum delay, all 11
elements are connected in series. If any element is faulty, then data will not propagate through the
comparator. If this is the only test failing, then it is almost certainly a bad comparator.
Comparator Zero-Hold Cal Test
Tests the delay elements for each delay line in the comparators. It tests that each delay line can
increase its delay in a linear way through a range of delay values.
Comparator Calibrations Test
The purpose of this test is to verify that each of the comparator one-time calibrations can
successfully be performed. This verifies that all of the calibration circuitry and components are within
the tolerance limits required for proper calibration. This test is executed only if all probes are
detached.
Acquisition Memory Write/Read Test
This test checks that each acquisition chip can write data to DDR acquisition memory and read the
same data back.
Acquisition Memory Cell Test
Tests every bit of the DDR acquisition memory. The test verifies that every bit can be written to 0 and
written to 1 and read back accurately.
ATB (AXIe Trigger Bus) Test
This test verifies the ATB signal connections between the acquisition chips, the interface FPGA and
the two 8-bit transceiver chips.
To exit the test system
1 Close the self-test dialog. No additional actions are required.
92Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 93
To restore the system software
NOTE
CAUTION
Restoring your system software might be necessary for the following reasons:
• Hard drive failure.
• Virus in the system or unstable system.
• Intentional disk clean - for example if you are passing the system to another team or returning it
to a rental company and you do not want any data left on it.
The 16860 series of Logic Analyzers have the Windows 7 operating system installed. These systems
also have the Keysight Logic and Protocol Analyzer software version 6.30 or later preinstalled.
If you need to restore the logic analysis system software, you run the recovery process on the hard
drive of the system. This recovery process uses the hidden partition on the hard drive to restore the
hard disk drive back to its original state in which it was shipped. When you run this process, the
recovered hard disk drive contains:
• the Windows 7 operating system
• the version of the Logic Analyzer software which was installed when the system was originally
shipped and not the latest or upgraded version of the software that might be available at the time
of system recovery.
• the license files of any licensed optional products that you had purchased with the Logic Analyzer
and that were installed when the system was originally shipped.
Troubleshooting5
When you run the recovery process, the software licensing Host ID of your logic analyzer
may change. If this happens, the restored license files will not work with the new host ID
and you will need rehosted license files. To get these files, you can contact your nearest
Keysight sales/service office. To locate a sales or service office, go to
www.keysight.com/find/contactus.
Running the recovery process reformats your logic analyzer’s hard disk drive to the state
in which it was originally shipped to you. All user data files and programs are overwritten
when the recovery process runs. Therefore, save your data to a CD or to another machine
before you start recovering the system software.
To run the recovery process
1 Shut down and then restart your Logic Analyzer.
While booting up, the Logic Analyzer displays the Windows Boot Manager screen. This screen
provides you the following options:
• Microsoft Windows 7 - This is the default selected option. This option starts the Logic Analyzer
with the Windows 7 operating system. This is the normal startup of the system.
• Keysight Recovery system - You should select this option when you want to restore/repair your
Logic Analyzer software by running the Keysight recovery process.
Keysight 16860 Series Portable Logic Analyzers Service Guide93
Page 94
5Troubleshooting
2 After a few seconds, Logic Analyzer automatically starts with the option selected in the Windows
Boot Manager. Select the Keysight Recovery system option in the Windows Boot Manager screen
using the arrow keys and then press Enter.
The recovery process starts preparing the system for recovery and displays the following screen
with options to choose. You can enter:
• 1 to run Check Disk on the Logic Analyzer’s hard disk drive. If the recovery process encounters
any problems while running Check Disk, it reports these problems else it returns to the
Keysight Recovery system prompt on completion of Check Disk.
• 2 to restore your Logic Analyzer software back to its original state in which it was shipped.
• 3 to view a document that provides information on the recovery process.
• 4 to repair the Logic Analysis system hard disk drive.
• 5 to exit the recovery process and restart the Logic Analyzer in the normal mode.
3 Select the second option in the above screen by entering 2 in the text box and clicking OK.
4 A warning message is displayed stating that the recovery process overwrites the data on C: drive.
If you want to save your data to a CD or to another machine before you proceed further, then exit
the recovery process and save your data. Else, click OK to proceed with the recovery process.
94Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 95
Troubleshooting5
The recovery process starts.
If the recovery process is able to restore the system software successfully, then the following
screen is displayed.
5Click OK to proceed. You Logic Analysis system now reboots to Windows 7.
6 The Windows Boot Manager screen is displayed with Microsoft Windows7 as the default selection.
The system automatically starts with this default selection. Alternatively, you can press Enter to
proceed with the default selection.
Keysight 16860 Series Portable Logic Analyzers Service Guide95
Page 96
5Troubleshooting
NOTE
Contacting Keysight Service/Support
There are situations when you are not able to run the recovery process, (for instance,
when the hard disk drive of your system fails) or when running the recovery process does
not recover your system software. In such situations, you can send your Logic Analysis
system for hard disk repair/replacement to Keysight. Alternatively, you can contact your
nearest Keysight sales/service office. To locate a sales or service office, go to
www.keysight.com/find/contactus.
To locate a sales or service office near you, go to: http://www.keysight.com/find/contactus
96Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 97
To test the logic acquisition cables
This test allows you to functionally verify the logic analyzer cable and the flying lead probe of any of
the logic analyzer pods. Only one probe and cable can be tested at a time. Repeat this test for each
probe and cable to be tested. Two Flying Lead Probes are required if you need to test pods other
than Pod 1 because the clock from Pod 1 will be used to acquire data.
This test allows you to functionally verify U4201A logic analyzer cables and Keysight E5379A probes.
Table 5 Equipment Required to Test Cables
EquipmentCritical SpecificationRecommended Part
Stimulus BoardNo Substitute16760-60001
Differential ProbesNo SubstituteE5379A (Qty 2)
1 Connect the 16860 logic analyzer to the stimulus board.
a Connect the Keysight E5379A 90-pin differential probes to the logic analyzer cable (also called
“Pods”) to be tested. Start with Pods 1 and 2.
b Connect the E5379A probe from logic analyzer Pod 1 to connector “Pod 4” on the stimulus
board.
c Connect the E5379A probe from logic analyzer Pod 2 to connector “Pod 5” on the stimulus
board.
d Connect the stimulus board power supply output to the stimulus board power supply
connector J82.
e Plug in the stimulus power supply to line power. The green LED DS1 should illuminate showing
that the stimulus board is active.
Troubleshooting5
Keysight 16860 Series Portable Logic Analyzers Service Guide97
Page 98
5Troubleshooting
Pod 1
Pod 2
Pod 4
connector
Pod 5
connector
2 Set up the stimulus board
a Configure the oscillator select switch S1 according to the following settings:
•S1 0 (Off).
•S2 1 (On).
•S3 0 (Off).
•Int.
b Configure the data mode switch S4 according to the following settings:
•Even.
•Count.
c Press the Resynch VCO button, then the Counter RST (Counter Reset) button.
3In the Keysight Logic and Protocol Analyzer application, choose File→New. This puts the logic
analysis system into its initial state.
98Keysight 16860 Series Portable Logic Analyzers Service Guide
Page 99
Troubleshooting5
4 Disable all analyzers except the one being tested. This simplifies the instructions and makes
module initialization faster.
a Select the Overview tab at the bottom of the main window.
5Set up the bus:
a In the Overview window, select Setup -> Bus/Signal... from the module’s drop-down menu.
b Scroll if necessary to view the pods you are testing.
c Verify that the pod threshold buttons say “Threshold: Differential”. If they don’t, make sure the
correct probes (E5379A) are attached to pods 1 and 2. The threshold is set to Differential
automatically when E5379A probes are attached.
Keysight 16860 Series Portable Logic Analyzers Service Guide99
Page 100
5Troubleshooting
d Channels 7 through 0 are already assigned by default. Assign pod 2 channels 15 through 0 and
pod 1 channels 15 through 8 by clicking and dragging from the left-most channel box to the
right-most channel box. Your display should look like the lower picture when you are done.
6 Select the State sampling mode and set the State Clock options:
a Select the Sampling tab of the Analyzer Setup window.
b Select State - Synchronous Sampling.
c For State Clock, select CLK1 Clock and Both edges.
7 Set the trigger position and acquisition memory depth:
a Set the Trigger Position to 100% Poststore.
b Set the Acquisition Depth to 8K.
100Keysight 16860 Series Portable Logic Analyzers Service Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.