Keysight Technologies 1130B, 1130B Series, 1131B, 1132B, 1134B User Manual

Keysight 1130B-Series Differential and Single-Ended Probes
User’s Guide
Notices
CAUTION
WARNING
© Keysight Technologies, 2016-2019
No part of this manual may be reproduced in any form or by any means (including elec­tronic storage and retrieval or translation into a foreign language) without prior agree­ment and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
01130-97014
Edition
Ninth Edition, January 2019
Available in electronic format only
Published by: Keysight Technologies, Inc.
1900 Garden of the Gods Road Colorado Springs, CO 80907 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maxi­mum extent permitted by applicable law, Keysight disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any infor­mation contained herein. Should Keysight and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the war­ranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accor­dance with the terms of such license.
U.S. Government Rights
The Software is "commercial computer soft­ware," as defined by Federal Acquisition Regulation ("FAR") 2.101. Pursuant to FAR
12.212 and 27.405-3 and Department of Defense FAR Supplement ("DFARS")
227.7202, the U.S. government acquires commercial computer software under the same terms by which the software is cus­tomarily provided to the public. Accordingly, Keysight provides the Software to U.S. gov­ernment customers under its standard com­mercial license, which is embodied in its End User License Agreement (EULA), a copy of which can be found at http://www.key-
sight.com/find/sweula. The license set forth
in the EULA represents the exclusive author­ity by which the U.S. government may use, modify, distribute, or disclose the Software. The EULA and the license set forth therein, does not require or permit, among other things, that Keysight: (1) Furnish technical information related to commercial computer software or commercial computer software documentation that is not customarily pro­vided to the public; or (2) Relinquish to, or otherwise provide, the government rights in excess of these rights customarily provided to the public to use, modify, reproduce, release, perform, display, or disclose com­mercial computer software or commercial computer software documentation. No addi­tional government requirements beyond those set forth in the EULA shall apply, except to the extent that those terms, rights, or licenses are explicitly required from all providers of commercial computer software pursuant to the FAR and the DFARS and are set forth specifically in writing elsewhere in the EULA. Keysight shall be under no obliga­tion to update, revise or otherwise modify the Software. With respect to any technical data as defined by FAR 2.101, pursuant to FAR 12.211 and 27.404.2 and DFARS
227.7102, the U.S. government acquires no greater than Limited Rights as defined in FAR 27.401 or DFAR 227.7103-5 (c), as applicable in any technical data. 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any tech­nical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAU- TION notice until the indicated con­ditions are fully understood and met.
A WARNING notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Contents

1 Getting Started
2Using Probe Heads
Introduction 8 Probe Handling 12 Using Offset With InfiniiMax Active Probes 16 Slew Rate Requirements for Different Technologies 18 Available Accessories 20 Safety Information 31 Troubleshooting 33
Recommended Configurations at a Glance 36
1. MX0100A InfiniiMax Micro Probe Head 38
2. E2677B Differential Solder-in Head with Full BW Resistors 49
3. E2678B Differential Socketed Head 51
4. E2675B Differential Browser 54
5. N5380B SMA Head 57
6. N2851A QuickTip Probe Head 60
7. E2679B Single-Ended Solder-In Head 64
8. E2676B Single-Ended Browser 65
9. E2677B Differential Solder-In Head with Medium BW Resistors 67
10. E2679B Single-Ended Solder-In Head with Long Wire 69
11. E2678B Differential Socketed Head with Damped Wire Accessory 71
12. E2678B Differential Socketed Head with Header Adapter 73 Replacing Resistors on E2677B/9B Solder-In Probe Heads 74
3 Calibrating Probes
Calibration for Solder-In and Socketed Probe Heads 80 Calibration for Hand-held Browser Probe Heads 87 Calibration for N5380A/B SMA Probe Heads 89
4 Characteristics and Specifications
General 92 Environmental 94 Regulatory 95 Probe Dimensions 96 MX0100A Probe Head Dimensions 97
5 1130B Performance Data Plots
E2675B Differential Browser 100 E2676B Single-Ended Browser 103
3
E2677B Differential Solder-in Probe Head (Full BW) 106 E2677B Differential Solder-in Probe Head (Medium BW) 109 E2678A/B Differential Socketed Probe Head (Full BW) 112 E2678A/B Differential Socketed Probe Head with Damped Wire Accessory 115 E2679B Single-Ended Solder-in Probe Head (Full BW) 118 E2679B Single-Ended Solder-in Probe Head (Medium BW) 121 N2851A QuickTip Head with N2849A QuickTip 124
6 1131B Performance Data Plots
E2675B Differential Browser 126 E2676B Single-Ended Browser 129 E2677B Differential Solder-in Probe Head (Full BW) 132 E2677B Differential Solder-in Probe Head (Medium BW) 135 E2678A/B Differential Socketed Probe Head (Full BW) 138 E2678A/B Differential Socketed Probe Head with Damped Wire Accessory 141 E2679B Single-Ended Solder-in Probe Head (Full BW) 144 E2679B Single-Ended Solder-in Probe Head (Medium BW) 147 N2851A QuickTip Head with N2849A QuickTip 150
7 1132B Performance Data Plots
E2675B Differential Browser 152 E2676B Single-Ended Browser 155 E2677B Differential Solder-in Probe Head (Full BW) 158 E2677B Differential Solder-in Probe Head (Medium BW) 161 E2678A/B Differential Socketed Probe Head (Full BW) 164 E2678A/B Differential Socketed Probe Head with Damped Wire Accessory 167 E2679B Single-Ended Solder-in Probe Head (Full BW) 170 E2679B Single-Ended Solder-in Probe Head (Medium BW) 173 N2851A QuickTip Head with N2849A QuickTip 176
8 1134B Performance Data Plots
E2675B Differential Browser 178 E2676B Single-Ended Browser 181 E2677B Differential Solder-in Probe Head (Full BW) 184 E2677B Differential Solder-in Probe Head (Medium BW) 187 E2678A/B Differential Socketed Probe Head (Full BW) 190 E2678A/B Differential Socketed Probe Head with Damped Wire Accessory 193 E2679B Single-Ended Solder-in Probe Head (Full BW) 196 E2679B Single-Ended Solder-in Probe Head (Medium BW) 199 N5380B SMA Probe Head 202 N2851A QuickTip Head with N2849A QuickTip 204
9 Performance Verification
Using the 8720ES VNA successfully 206 Procedure 207 Performance Test Record 214
4 1130B-Series Probes User’s Guide
10 SPICE Models
SPICE Model for Differential Probe Heads 217 SPICE Model for Single-Ended Probe Heads 218 SPICE Deck and Measured/Modeled Data Matching 219
11 Replacement Parts
E2675B Differential Browser Probe Head 230 E2677B Differential Solder-In Probe Head 230 E2678A/B Differential Socketed Probe Head 231 E2679B Single-Ended Solder-in Probe Head 231 Other Accessories 232
Index
1130B-Series Probes User’s Guide 5
6 1130B-Series Probes User’s Guide
Keysight InfiniiMax I Series Probes
WARNING
CAUTION
User’s Guide

1 Getting Started

Introduction 8 Probe Handling 12 Using Offset With InfiniiMax Active Probes 16 Slew Rate Requirements for Different Technologies 18 Available Accessories 20 E2668B Single-Ended Connectivity Kit 20 E2669B Differential Connectivity Kit 21 N5450B Extreme Temp Cable Extension Kit 23 N2880A InfiniiMax In-Line Attenuator Kit 24 N2881A InfiniiMax DC Blocking Caps 28 MX0102A Soldering Toolkit 29 Safety Information 31 Troubleshooting 33
The 1130/1/2/4B InfiniiMax active probes are designed for probing differential and single-ended high-frequency signals. The probes are compatible with the Infiniium AutoProbe Interface which completely configures the Infiniium series of oscilloscopes for the probes.
Before using the probe, refer to “Safety Information" on page 31.
Before using the probes, refer to “Probe Handling" on page 12.
7
1 Getting Started

Introduction

Before you can use the probe, you must connect one of the available probe heads to an 1130/1/2/4B probe amplifier.
Figure 1 Probe Amplifier with Attached Head
Probe Heads
The available probe heads are shown in Figure 2 on page 9 and are documented in
Chapter 2, “Using Probe Heads". The InfiniiMax I probe heads are designed
specifically for the 1130/1/2/4B probes amplifiers. The InfiniiMax II probe heads are designed for 1168/9B probe amplifiers, but can also be used with the 1130/1/2/4B probe amplifiers.
The probe heads can be ordered at the same time as 1130/1/2/4B probe amplifiers. The E2669B and E2668B connectivity kits, described on page 21 and
page 20 conveniently package multiple probe heads and their accessories.
Differential probe heads offer easy measurement of differential signals and greatly improve the measurement of single-ended signals. Single-ended probe heads offer extremely small size for probing single-ended signals in confined spaces.
Compatible Oscilloscopes
Table 1 on page 10 lists the oscilloscopes that are compatible with the
1130/1/2/4B probes. Is Your Oscilloscope Software Up-to-Date? Keysight periodically releases Oscilloscope software updates to support your probe, fix known defects, and incorporate product enhancements. To download the latest firmware, go to www.keysight.com and search for your oscilloscope’s topic. Click on the “Drivers, Firmware & Software” tab.
8 1130B-Series Probes User’s Guide
Getting Started 1
NOTE
Figure 2 Available Probe Heads and Accessories
N2849A QuickTips are also compatible with the N2848A InfiniiMode probe heads which are designed for N2830/1/2A InfiniiMax III+ and N2800/1/2/3A InfiniiMax III probes.
1130B-Series Probes User’s Guide 9
1 Getting Started
Table 1 Compatible Oscilloscopes
Oscilloscope Adapter Required
Infiniium Oscilloscopes
S-Series none
90000 X-, and Q-Series N5442A
90000A Series none
86100C/D Series N1022A/B
9000 H-Series none
9000A-Series none
8000A-Series none
InfiniiVision Oscilloscopes
6000 X-Series none
4000 X-Series none
3000 X-Series none
7000A Series none
6000A Series (350 MHz — 1 GHz) none
5000A Series none
Channel Identification Rings
When multiple probes are connected to the oscilloscope, use the channel identification rings to associate the channel inputs with each probe. Place one colored ring near the probe’s channel connector and place an identical color ring near the probe head.
Inspecting the Probe
Figure 3 on page 11 shows the accessories that are shipped with the 1130/1/2/4B
probe amplifiers. The probe amplifiers do not come with a probe head unless selected at the time of order. Any head shown in Figure 2 on page 9 can be ordered at any time for any 1130/1/2/4B probes.
Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has been checked mechanically and electrically.
Check the accessories.
10 1130B-Series Probes User’s Guide
Getting Started 1
If the contents are incomplete or damaged, notify your Keysight Technologies
Sales Office.
Inspect the probe. If there is mechanical damage or defect, or if the probe does not operate properly or pass calibration tests, notify your Keysight Technologies Sales Office.
If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Keysight Technologies Sales Office. Keep the shipping materials for the carrier’s inspection. The Keysight Technologies office will arrange for repair or replacement at Keysight Technologies’ option without waiting for claim settlement.
Figure 3 Accessories Supplied With the Probe Amplifier
Cleaning the probe
If the probe requires cleaning, disconnect it from the oscilloscope and clean it with a soft cloth dampened with a mild soap and water solution. Make sure the probe is completely dry before reconnecting it to the oscilloscope.
1130B-Series Probes User’s Guide 11
1 Getting Started
CAUTION

Probe Handling

This probe has been designed to withstand a moderate amount of physical and electrical stress. However, with an active probe, the technologies necessary to achieve high performance do not allow the probe to be unbreakable. Treat the probe with care. It can be damaged if excessive force is applied to the probe tip. This damage is considered to be abuse and will void the warranty when verified by Keysight Technologies service professionals.
Exercise care to prevent the probe end from receiving mechanical shock.
Store the probe in a shock-resistant case such as the foam-lined shipping case
which came with the probe.
Connecting and Disconnecting Probe Heads
When disconnecting a probe head from an amplifier, pull the probe head connectors straight out of the sockets as shown in Figure 4. When connecting a probe head to an amplifier, push straight in. Always grasp the indentations located on the sides of the amplifier as shown in Figure 4. There are also indentations on many of the probe head sockets so you have a convenient place to grasp there as well.
Figure 4 Properly Pulling the Probe Head Straight Out
Avoid damaging the connection pins. Never bend the probe head in order to “pop” it loose from the amplifier. Do not wiggle the probe head up and down or twist it to remove the connectors from the sockets.
12 1130B-Series Probes User’s Guide
Figure 5 Improperly Disconnecting a Probe Head From an Amplifier
CAUTION
CAUTION
Handling the Probe Cable
Getting Started 1
Avoid degrading the probe’s performance. Do not twist, kink, or tightly bend the probe’s cable.
When the probe is attached to an oscilloscope, avoid letting object hit the probe cable where the cable exits the probe amplifier and bend it well beyond its limit.
When storing the probe, coil the cable in a large loops and avoid twisting the cable. Coil the cable in a similar manner to how garden hoses or extension cords are typically coiled. You can start by wrapping the cable around your thumb as shown in Figure 6. Then continue to circle your thumb, but provide a slight twist with each rotation. This allows the cable rotations to lie flat against each other and will eliminate the net twisting of the cable in the end.
Figure 6 Recommended Coil for Storage
1130B-Series Probes User’s Guide 13
1 Getting Started
CAUTION
Connecting the Probe to an Oscilloscope
Handling the Probe Amplifier
Make the coil’s radius fairly large so it does not induce kinking or bending.
The probes are only meant to be plugged into gold plated BNCs (like those on Infiniium oscilloscopes). To connect the probe to the oscilloscope, simply push the probe into the BNC connector and the locking mechanism automatically engages. To disconnect the probe, push and hold the locking lever to the left and then remove the probe.
The probe amplifier contains a delicate circuit board. Treat it carefully and take standard precautions (for example, not dropping it repeatedly or from large heights, not getting it wet, not smashing it with heavy objects, etc.). These probes are sensitive ESD devices so standard precautions need to be used to not ruin the probe from the build-up of static charges.
Securing Probe Heads and Amplifiers to Your DUTs
When soldering a probe head to a circuit, first provide strain relief by using low temperature hot glue (use as little as possible) or non-conductive double-sided tape. Do not use super glue and do not get the low temperature hot glue on the actual probe head tip as this can damage the precision components of your probing system (only use the low temperature hot glue on the probe head cables). The provided velcro pads can be used to secure your probe amplifier casing to the board.
Once strain relief has been provided, solder the probe tip to the circuit board and then plug the probe head into the probe amplifier.
Figure 7 Correct Securing Methods
14 1130B-Series Probes User’s Guide
Getting Started 1
Figure 8 Incorrect Securing Method Because Glue is Placed on the Probe Head Tip
The velcro dots can be used to secure the probe amplifier to a circuit board removing the weight of the probe from the circuit connection. Attach a Velcro dots to both the probe amplifier and the circuit board as shown in Figure 9 on page 15.
Figure 9 Using the Velcro Dots
1130B-Series Probes User’s Guide 15
1 Getting Started

Using Offset With InfiniiMax Active Probes

It is important to understand how the 113xA probes behave with respect to offset when different probe head / signal combinations are used.
The purpose of offset in active probes or oscilloscope front ends is to allow the subtraction of most or all of the dc component of the input signal so the signal can better utilize the dynamic range of the input. When using an InfiniiMax probe with an Infiniium oscilloscope, you can select the case (see the three cases described below) that applies for your measurement by selecting the Probes button under the channel setup menu. This allows you to select which type of probe head is being used and, if it is a differential probe head, allows you to select whether you are probing a differential or single-ended signal. With these inputs, the oscilloscope will use the proper type of offset for your measurement case. The specifics for each case are discussed below.
As an important side note, whenever adjusting the offset for a particular probe head, make sure to have a triggered signal.
Case 1. A single-ended probe head probing a single-ended signal
For this case, the offset control on the oscilloscope controls the probe offset and the channel offset is set to zero. This allows the offset voltage to be subtracted from the input signal before the signal gets to the differential amplifier. Since this subtraction is done before any active circuits, the offset range is large (±12V for the 113X amplifiers and 25-kW probe heads). Note that the minus probe tip is not present when using a single-ended probe head which means nothing is plugged into the "–" input of the probe amp. This is normal and causes no problems.
Case 2. A differential probe head probing a single-ended signal
For this case, the offset control on the oscilloscope controls the probe offset and the channel offset is set to zero. This allows the offset voltage to be subtracted from the input signal before the signal gets to the differential amplifier. Since this subtraction is done before any active circuits, the offset range is large (±12V for the 113X amplifiers and 25-kW probe heads). A differential probe can make higher bandwidth and more accurate measurements on single-ended signals than a single-ended probe and this method of applying offset to only the plus side of a differential probe means there is no sacrificing of offset range.
Case 3. A differential probe head probing a differential signal
For this case, the offset control on the oscilloscope controls the oscilloscope channel offset. The probe offset is not used and set to zero. Since the plus and minus sides of differential signals have the same dc component, it will be subtracted out and the output of the probe will by definition be centered around ground.
16 1130B-Series Probes User’s Guide
Getting Started 1
The channel offset allows the waveform seen on screen to be moved as desired. The allowable dc component in the plus and minus signals is determined by the common mode range of the probe which for the 113x probe amps and 25 kW probe heads is ±6.75 V.
1130B-Series Probes User’s Guide 17
1 Getting Started

Slew Rate Requirements for Different Technologies

The following table shows the slew rates for several different technologies. The maximum allowed input slew rate is 18 V/ns for single-ended signals and 30 V/ns for differential signals. Table 2 shows that the maximum required slew rate for the different technologies is much less that of the probe.
Table 2 Slew Rate Requirements
Max Single-Ended
Differential
Name of Technology
PCI Express (3GIO) YES 9.6 19.2 50 1.6
RapidIO Serial 3.125Gb YES 8.0 16.0 60 1.6
10GbE XAUI (4x3.125Gb) YES 8.0 16.0 60 1.6
1394b YES 8.0 16.0 60 1.6
Fibre Channel 2125 YES 8.0 16.0 75 1
Gigabit Ethernet 1000Base-CX YES 7.8 15.5 85 2.2
RapidIO 8/16 2Gb YES 7.2 14.4 50 1.2
Infiniband 2.5Gb YES 4.8 9.6 100 1.6
HyperTransport 1.6Gb YES 4.0 8.0 113 1.5
SATA (1.5Gb) YES 1.3 2.7 134 0.6
USB 2.0 YES 0.9 1.8 375 1.1
DDR 200/266/333 NO 7.2 n/a 300 3.6
PCI NO 4.3 n/a 500 3.6
Signal
Slew Rate (V/ns)
*
Max Differential Slew Rate (V/ns)
Driver Min Edge
Rate (20%-80% ps)
Max Transmitter Level (Diff V)
AGP-8X NO 3.1 n/a 137 0.7
* The probe specification is 18 V/ns
† The probe specification is 30 V/ns
18 1130B-Series Probes User’s Guide
Getting Started 1
Slew Rates of Popular Technologies Compared to Maximum Probe Slew Rates
0.0
5.0
10.0
15.0
20.0
25.0
30.0
)
1394b
Hype
rT
rans
p
o
rt
1
.
6Gb
S
AT
A (
1
.5
Gb
)
Popular Technologies
Edge Slew Rates (V/nS) +
Diff erential Slew Rates
Maximum Probe Differ ential Slew Rate (30 V /nS)
Slew Rates of Popular Te chnologies Compared to Maximum Probe Slew Rates
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
18.0
20.0
Popular Technologies
Edge Slew Rates (V/nS) +
Single-ended Slew Rates
Maximum Probe Single-ended Slew Rate (18 V/nS)
**
*
**
*
* Measurement of one side of differential signal
Maximum Edge Amplitude x 0.6
Minimum 20% to 80% Rise Time
------------------------ ----------------------------- ----------------------------
PCI Expr
PCI
Rap
ess
Exp
(3
id
re
Ra
GIO)
IO Seria
GIO
(3
3
s
l
s
a
ri
e
IO S
d
i
p
10GbE XAUI (4x3.125Gb)
b
G
.125
3
l
XAUI (4x3.125Gb)
10GbE
.125Gb
*
4b
39
1
e Cha
r
Fib
gabit Etherne
i
G
Ch
ibre
F
it Ethernet 1000Base-CX
b
a
g
Gi
25
1
nel 2
Base-CX
n
0
100
t
Gb
idIO 8/1
p
Ra
Gb
2.5Gb
band
i
HyperTransport 1.6
2Gb
6
2.5
d
n
a
b
i
in
f
In
*
*
)
b
G
5Gb
1.
(
A
AT
S
U
SB 2.0
DD
R 200/2
6/333
6
2125
l
e
n
n
a
*
6 2
/1
IO 8
pid
Infin
Ra
USB
CI
P
.0
2
X
AGP-8
1130B-Series Probes User’s Guide 19
Figure 10 Slew Rates of Popular Technologies Compared to Maximum Probe Slew Rates
1 Getting Started

Available Accessories

This section lists accessories that are available in addition to the probe heads described in Chapter 2, “Using Probe Heads".
E2668B Single-Ended Connectivity Kit
The optional E2668B single-ended connectivity kit is an accessory that provides the three probe heads shown in Figure 11. A single-ended socket-tip probe head was not developed since it did not offer a significant size advantage. The kit can be ordered at the same time as 1130/1/2/4B probe amplifiers.
Figure 11 E2668B Single-Ended Connectivity Kit (not to scale)
Table 3 Supplied Accessories (Sheet 1 of 2)
Qty
Description
E2679B Single-Ended Solder-In Head 1
E2678B Differential Socketed Head 1
E2676B Single-Ended Browser 1
91W resistor for full bandwidth 16
150W resistor for medium bandwidth 8
0W resistor for full and medium bandwidth 24
91W resistor template 2
150W resistor template 2
160W damped wire accessory 6
82W resistor for full bandwidth 48
Socket for 25 mil (25/1000 inch) square pins, female on both ends
Supplied
4
Used With
E2679B E2678B E2676B
Part Number
1NC3-1091
1NC3-1150
1NC3-1000
01131-94311
01131-94308
01130-21303
01130-81506
01131-85201
*
20 1130B-Series Probes User’s Guide
Table 3 Supplied Accessories (Sheet 2 of 2)
Getting Started 1
Qty
Description
25 mil female socket w/20 mil round male pin on other end 4
Heat shrink tubing for square-pin socket accessory 4
Header adapter, 91W 2
82W resistor template 1
Resistive tip (blue), 91W 10
Ergonomic handle 1
Ground collar assembly for single-ended browser 2
Socketed ground lead 6 inches 1
* Not orderable.
Supplied
Used With
E2679B E2678B E2676B
E2669B Differential Connectivity Kit
The optional E2669B differential connectivity kit provides multiple quantities of the three probe heads as shown in Figure 12. These probe heads allow full bandwidth probing of differential and single-ended signals. The kit can be ordered at the same time as 1130/1/2/4B probe amplifiers.
Part Number
01131-85202
01130-41101
01130-63201
01131-94309
01131-62107
01130-43202
01130-60012
E2676-21301
*
Figure 12 E2669B Differential Connectivity Kit (not to scale)
1130B-Series Probes User’s Guide 21
1 Getting Started
NOTE
Table 4 Supplied Accessories
Qty
Description
E2678B Differential Socketed Head 2
E2677B Differential Solder-In Head 4
E2675B Differential Browser 1
160W damped wire accessory 12
82W resistor for full bandwidth 96
Socket for 25 mil (25/1000 inch) square pins, female on both ends
25 mil female socket w/20 mil round male pin on other end
Heat shrink socket accessory 8
Header adapter, 91W 4
82W resistor template 1
91W resistor for full bandwidth 80
150W resistor for medium bandwidth 40
91W resistor template 1
Supplied
8
8
Used With
E2678B E2677B E2675B
Part Number
01130-21303
01130-81506
01131-85201
01131-85202
01130-41101
01130-63201
01131-94309
1NC3-1091
1NC3-1150
01131-94311
*
150W resistor template 1
Resistive tip (blue), 91W 20
Ergonomic handle 1
* Not orderable.
Resistor performance. The S2 resistors were changed from 100W to 91W for slightly better performance. Either value produces a response that is well within specifications.If you have some of the older 100W resistors, ensure that you use either two 100W or two 91W resistors. Do not mix them.
01131-94308
01131-62107
01131-43201
22 1130B-Series Probes User’s Guide
N5450B Extreme Temp Cable Extension Kit
CAUTION
CAUTION
CAUTION
The extreme temperature cable extension kit is an accessory that allows an oscilloscope probe to be used to monitor a device in a temperature chamber. Keysight’s Infiniimax probe amplifiers have a specified operating temperature range from 5 can be operated over a much larger range of temperatures. Use the extension cables to physically separate the amplifier from the probe head which allows you to operate the probe head inside a temperature chamber while the probe amplifier remains outside the chamber.
To ensure a high-quality measurement, the N5450B cable set have been phase-matched at the factory. A coupling tag is included with the cables to ensure the cables stay as a matched pair. To install the coupling tag, slip the small end of each cable through the holes in the tag. The tag can be positioned anywhere along the length of the cable and can withstand the temperature ranges specified.
Table 5 Probing Temperature Ranges
o
C to 40o C, but the probe heads
Getting Started 1
Probe Head Configuration
E2677B –25 to +80 > 1000
E2678A/B –25 to +80 > 1000
N5425B + N5426A –40 to +85 > 500
N5451A –25 to +80 > 1000
MX0100A -55° C dwell, 1000 hours minimum
* Refers to the probe head or tip that is attached to the cable extension kit.
*
Operating Temperature Range (oC)
+150° C dwell, 1000 hours minimum
-55° C to 150° C cycles, 1000 cycles minimum (as per JEDEC JESD22-A104 revision E)
Avoid rapid changes in temperature that can lead to moisture accumulating in the form of condensation on the probe components, as well as the DUT. If this occurs, wait until the moisture has evaporated before making any measurements.
Additional care must be taken when handling probe heads used during extreme temperature cycling because this process makes the probe heads less robust.
Expected Lifetime of the
Probe Head (cycles)
> 1000
Secure the ends of the extension cable near the probe head in the temperature chamber such that the probe head legs are not tugged or moved around significantly.
1130B-Series Probes User’s Guide 23
1 Getting Started
CAUTION
CAUTION
NOTE
NOTE
N2880A InfiniiMax In-Line Attenuator Kit
Prevent abrasion and tears in the cable’s jacket, do not rest the extension cables on any metal objects or objects with sharp edges.
Do not kink the cables. The cables are designed to be flexible, but are not designed to be bent sharply.
Keep your extreme temperature testing probes separate from the probes they use under milder conditions. This is because cycling probe heads through extreme temperature ranges has a marked affect on their lifetimes as listed in Table 5. Only the lifetime of the probe head is affected by temperature cycling. The extension cables and probe amplifier should not need to be replaced with extended temperature cycling.
Discoloration or texture changes are possible with the extension cables. These changes do not, however, affect the performance or the quality of a measurement.
The in-line attenuators are an accessory for probes. The maximum input range of the 1130B probes are 5 V design allows you to add the N2880A InfiniiMax in-line attenuators between the probe head and probe amplifier to increase the maximum input range (as listed in
Table 6 on page 25). Additionally, these attenuators enable you to increase the
offset range of the probe as specified in Table 6 on page 25. When using the N2880A In-Line Attenuators, the bandwidth and rise time of your probing system is not affected. There is, however, a trade-off in noise (refer to Table 6) and in the accuracy of DC offset relative to the input.
The N2880A provides a pair of 6 dB, 12 dB, and 20 dB attenuators. The attenuators come as matched pairs and should only be used with each other. Each attenuator has a serial number. The pair of matching attenuators in each set will have the same four digit numeric prefix and will differ by the last letter (one attenuator in the matched pair will be labeled A and the other will be labeled B).
. If you need to measure larger signals, the probe’s
p-p
Figure 13 Placement of Attenuators Between Probe Amplifier and Head
24 1130B-Series Probes User’s Guide
Compatibility with probe heads and amplifiers
NOTE
All InfiniiMax I probe heads and amplifiers are compatible with the N2880A In-line attenuators. However, the following two limitations should be considered when planning to use N2880A attenuators.
The maximum input voltage of the InfiniiMax probe heads is ±30 Vdc (depending on the frequencies of your signal, the maximum allowed slew rate (see table below) may require that the maximum input voltage magnitude be less than 30V), so they should not be used to measure signals that exceed this range. This places a practical limit of 20 dB on the attenuators used with the InfiniiMax probing system. Larger attenuation ratios will only degrade the noise performance and gain of the system.
Due to the N5380B dual-SMA probe head’s maximum input voltage specification of 2.28 V large enough to require an added attenuator.
.
Table 6 N2880A With 1130B-Series Probe Amplifiers
RMS
Getting Started 1
, the N5380B is not suitable for measuring signals
Maximum Input Range
Added Attenuator
None 5 Vp-p ±12 V 3 mV RMS se: 18 V/ns, diff: 30 V/ns 10:1
6 dB (2:1) 10 Vp-p ±24 V 7.8 mV RMS se: 36 V/ns, diff: 60 V/ns 20:1
12 dB (4:1) 20 Vp-p ±30 V
20 dB (10:1) 50 Vp-p ±30 V 41.7 mV RMS se: 180 V/ns, diff: 300 V/ns 100:1
* These slew rate do not apply when the N5380B SMA probe head is used with the InfiniiMax amplifiers.
† The actual range of DC voltage for these attenuators is greater than +/- 30 V, but the usable range of DC voltage at the probe input is limited
to ±30 Vdc. Also, depending on the frequencies of your signal, the maximum allowed slew rate may require that the maximum input voltage magnitude be less than 30 V.
(mains isolated circuits only)
Offset Range
Typical Noise Referred to
16.7 mV RMS se: 72 V/ns, diff: 120 V/ns 40:1
Maximum Allowed Input Slew Rate (se = single-ended) (diff = differential)
*
Nominal DC Attenuation of Probe System
The values shown above do not apply to the N5380B dual-SMA probe head. Due to the maximum input voltage specification of 2.28 VRMS (mains isolated circuits only) for the N5380B, it is not suitable for measuring signals large enough to require an added attenuator.
Calibrating and Configuring Attenuators on an Infiniium Series Oscilloscope
The software in the Infiniium and InfiniiVision oscilloscopes will detect a probe when it is connected and by default will assume that no additional attenuators are installed. If you want to scale readings and settings on the oscilloscope so they are correct with the attenuators installed, refer to the procedures below for your specific oscilloscope series.
Configuring Attenuators on a Infiniium Scope
1130B-Series Probes User’s Guide 25
1 Getting Started
You cannot calibrate your InfiniiMax probes with the attenuators attached. Calibrate the InfiniiMax probes as you normally would (with no attenuators), configure the attenuators as discussed in the next section, and begin probing.
First, plug your InfiniiMax probe amplifier / probe head into one of the oscilloscope channels with the attenuators connected. Then enter the Probe Setup dialog box (can be reached via Setup > Probes on the oscilloscope menu). Press the Configure Probing System button. A pop-up window will appear where you can select External Scaling. Click the Decibel radio button under the External Scaling section and then set the Gain field to either –6 dB, –12 dB, or –20 dB depending on the attenuator you are using (be sure to include the negative sign). Finally, you will need to manually set the Offset field in this dialog box to zero out the signal.
Calibrating Probe with Attenuators on a InfiniiVision Scope
The following instructions only apply if you have InfiniiVision software release 5.25 or newer installed on your oscilloscope.
1 Plug your InfiniiMax probe amplifier / probe head into one of the oscilloscope
channels with the attenuators attached.
2 Press the Channel on/off key to turn the channel on (if the channel is off).
3 Press the Probe softkey in the Channel menu. A series of probe related softkeys
will appear.
4 Repeatedly press the second softkey from the left softkey until the probe head
selection matches the attenuator you are using. The choices are:
10:1 single-ended browser
10:1 differential browser
10:1 (+6 dB Atten) single-ended browser
10:1 (+6 dB Atten) differential browser
10:1 (+12 dB Atten) single-ended browser
10:1 (+12 dB Atten) differential browser
10:1 (+20 dB Atten) single-ended browser
10:1 (+20 dB Atten) differential browser
Once the probe head configuration has been selected, you can press the Calibration key in the same probe menu and follow the on-screen instructions to calibrate the probe/attenuator setup.
Frequency Response Plots
Figure 14 on page 27 shows the frequency response plots for four setups: the
probe without any attenuators, the probe with the 6 dB attenuators, the probe with the 12 dB attenuators, and the probe with the 20 dB attenuators.
26 1130B-Series Probes User’s Guide
Getting Started 1
Red = dB(Vout/Vin) + 10.8 dB of probe
Black = dB(Vout/Vin) + 6dB attenuator + 10.8 dB
Blue = dB (Vout/Vin) + 12 dB attenuator + 10.8 dB of probe
Green = dB(Vout/Vin) + 20 dB attenuator + 10.8 dB of probe
Figure 14 Frequency Response
1130B-Series Probes User’s Guide 27
1 Getting Started
N2881A InfiniiMax DC Blocking Caps
The DC blocking capacitors are an accessory for the probes. The architecture of the InfiniiMax probing system allows you to place the N2881A DC Blocking Caps in between the probe amplifier and the probe head (as shown in the picture below). These N2881A InfiniiMax DC Blocking Caps block out the DC component of the input signal (up to 30 Vdc). The N2881A InfiniiMax DC Blocking Caps can be used with the N2880A In-Line Attenuators. The order of the two products in the probing system (i.e. which one is closest to the probe amplifier) does not matter.
Figure 15 Placement of DC Blocking Caps Between Probe Amplifier and Head
Figure 16 on page 28 shows the frequency response plot of the N2881A DC
Blocking Caps (no probe included).
Figure 16 Graph of DC Blocking Cap insertion loss (S2,1) versus frequency (DC Blocking Cap
only)
28 1130B-Series Probes User’s Guide
MX0102A Soldering Toolkit
Straight
Tweezers
Cutting
Tweezer s
Double-sided Foam Tape
Low Temperature
Solder Wire
Regular Solder Wire
Probe Tip Wire
Kapton Tape
The optional MX0102A soldering toolkit provides tools that can make soldering tasks easier. For instance, you can use the tools available in this kit while soldering the lead wires of the MX0100A Micro probe head to a DUT (see page 41).
Getting Started 1
Table 7 Accessories supplied in the soldering toolkit
Description
Straight Tweezers
Qty Supplied
1 8710-2837 (Anti-magnetic straight pointed tip 120mm) For general purpose manipulation / movement of com-
ponents such as probe tip wires and probe head.
Cutting Tweezers
1 8710-2838 (Narrow oblique head 115mm) To cut a probe tip wire to a desired length.
Kapton Tape (36 yards roll)
1 0460-3121 To provide strain-relief to the neck portion of the probe
head by taping it to a flat surface (such as a DUT circuit board).
1130B-Series Probes User’s Guide 29
Part Number
*
1 Getting Started
Description
Double-sided Foam Tape
To provide strain-relief to either the neck portion of the probe head or the plastic housings by taping it to a flat surface such as a tabletop or a DUT circuit board.
Regular Solder Wire
Lead free, .009" diameter, 2 feet long To attach the probe tip wires to a DUT using standard lead-free soldering temperatures (330 °C to 350 °C). (NOTE: This alloy melts at 217
o
C.)
Low Temperature Solder Wire
Lead free, .010" diameter, 2 feet long To attach the probe tip wires to a DUT using a low tem-
perature setting on your soldering iron. (NOTE: This alloy melts at 138
o
C.)
Probe Tip Wire
.004" diameter, 2 feet long To add ground wires to your probe tip if InfiniiMode mea-
surements (differential, single ended, and common mode signals with a single probe tip) are desired. Clip as short as possible using the cutting tweezers included in the kit.
Qty
Supplied
Part Number
*
10 0460-3122
1 MX0102-21302
1 MX0102-21303
1 MX0102-21301
* You can reorder these items using the part numbers included in the table above.
30 1130B-Series Probes User’s Guide
Loading...
+ 204 hidden pages