Keysight (Agilent) V8486A Data Sheet

Your New Choice for Accurate Power Measurements from 50 to 75 GHz
Compatible with Your Current Agilent Power Meter
The Agilent Technologies V8486A V-band power sensor is fully compatible with Agilent 435B, 436A, 437B, 438A, 70100A, E1416A, E4418A/B, and E4419A/B power meters.
In RF and microwave power measurements, the largest single source of error is usually sensor and source mismatch. To minimize any measurement uncertainty caused by mismatch, the V8486A offers a specified SWR of 1.06 (>30 dB return loss).
Agilent V8486A
V-Band Power Sensor
Profile
2
V8486A Calibration Factor Uncertainty at 1 mW (0 dBm)
Frequency (GHz) Uncertainty %
50 4.8 51 6.1 52 5.9 53 5.9 54 5.9 55 4.6 56 6.1 57 6.1 58 6.2 59 6.2 60 4.7 61 6.2 62 6.1 63 6.0 64 6.0 65 4.5 66 6.6 67 6.7 68 6.7 69 6.6 70 4.4 71 6.7 72 6.8 73 7.0 74 7.3 75 5.1
50 MHz Calibration
For easy calibration with the power meter, the Agilent V8486A incorporates a 50 MHz calibration port. This eliminates the variance in making meas­urements with different meter/sensor combina­tions and the uncertainties due to temperature changes. It also provides traceability to the U.S. National Institute of Standards and Technology (NIST) at millimeter-wave frequencies.
High Reliability
The V8486A uses the same proven Modified Barrier Integrated Diode (MBID) technology found in other Agilent sensors, and always operates in the square-law characteristic for true-average detec­tion.
The uncertainty analysis for the calibration of the V8486A is car­ried out in accordance with the ISO/TAG4 Guide. The uncertainty data, reported on the calibration certificate, is the expanded uncertainty with a 95% confidence level and a coverage factor (k) of 2.