Keysight (Agilent) U3401A Data Sheet

Agilent U3400 Series 4½ and 5½ Digit Digital Multimeters
Data Sheet
Basic + Good = Elegant Simplicity
Features
Up to 119,999 counts resolution
Up to 0.012% basic DCV accuracy
11 basic measurements and up to
six built-in math functions
Dual display on bright VFD
Selectable resolutions for variable
measurement speeds*
Kensington lock slot security
Simply right for your basic needs, without compromising quality
The U3400 Series digital multimeters (DMMs) come with functions vital for your everyday measurements: DC, AC and AC+ DC voltage and current, 2- and 4-wire resistance, frequency, continuity and diode tests.
Math functions dBm, Min/Max, Relative, Compare, Hold and Percentage** are also available to ease your measurement analysis.
Not only are the U3400 Series built robust for lasting reliability over the long haul, these DMMs also ensure dependable measurements with up to 0.012% basic DCV accuracy.
Efficient testing with dual display and selectable resolutions*
Dual display on the U3400 Series lets you view two parameters simultaneously as you
measure, and further enhances your troubleshooting tasks. For example, the ability to see both AC voltage and frequency simultaneously would help you to measure the frequency response of amplifier circuits more efficiently and effectively. Refer to Page 3 for more typical dual display combinations and applications.
Up to three measurement speeds are available on the U3402A: Slow, Medium and Fast. This means greater flexibility in catering to different testing needs: a faster measurement speed at lower resolution or a higher resolution at slower measurement speed.
Physical security
Your instruments may be at risk of theft or misplacement whenever you leave them unattended on the bench. With the U3400 Series rear Kensington lock slot, you can secure your DMM and be assured that it is where you expect it to be for your continued testing the next day.
* U3402A only ** U3401A only
Take a closer look
Primary display Secondary display
Input terminals and current fuse
Measurement function keypads
Autoranging, manual range and comparator operation
operation keypads
ShiftMath
Power on/off switch
Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with slight differences in the position of certain functions. Please refer to the U3401As Users and Service Guide U3401-90001 for details.
AC line voltage selector
Line voltage fuse selection table
AC power connector
Figure 2 Rear panel of the U3401A/U3402A.
AC line fuse Chassis ground lug RS232 interface connector
(For calibration use only)
2
Typical dual display combinations and applications
Primary display
DCV
ACV + DCV
DCV
DCV
ACI + DCI
ACV
ACI + DCI
ACV
ACV
ACI
DCI
ACI + DCI
dBm
dBm
dBm
dBm
Secondary display
ACV
DCV
DCI
ACI
DCV
DCI
ACV
ACI
Hz
Hz
ACI
DCI
Reference W
DCV
ACV
Hz
Application
 Testing of DC-to-AC or AC-to-DC converter circuit
 Power supply DC level and AC ripple measurement
 Testing of power supply load regulation
 Loop current and voltage drop level checking
 Testing of line and load regulation
 Testing of AC-to-DC or DC-to-AC converter
 Power supply DC level and AC ripple measurement
 Transformer testing
 Amplifier circuits AC frequency response measurement
 Adjustment of AC motor control
 Power supply AC ripple and DC current measurement
 Current dissipation measurement
 RF frequency measurement
 Frequency response checking
3
U3400 Series specifications
DC voltage
DCV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
[1] Input impedance is in parallel with capacitance <100 pF. [2] In VDC 1000 V range, 1200 V is readable with audio warning.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
1200.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
[2]
Accuracy (1 year; 23 ºC ± 5 ºC)
0.02% + 4
0.02% + 4
0.02% + 4
0.02% + 4
0.02% + 4
Maximum reading
119.999
1.19999
11.9999
119.999
[3]
1000.00
399.99
3.9999
39.999
399.99
[3]
1000.0
399.9
3.999
39.99
399.9
[3]
1000
Accuracy (1 year; 23 ºC ± 5 ºC)
0.012% + 8
[2]
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
Typical input impedance
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
Typical input impedance
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
[1]
[1]
[1] Input impedance is in parallel with capacitance <120 pF. [2] Relative (REL) operation is used. [3] In VDC 1000 V range, 1050 V is readable.
4
AC voltage (true RMS, AC coupling mode)
ACV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
750.0 V
[1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] For 5 kHz to 10 kHz, accuracy is 0.7% + 15. [4] In VAC 750 V range, 1000.0 V RMS is readable with audio warning.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
1000.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
750 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
Maximum reading
119.999
1.19999
11.9999
119.999
750.00
399.99
3.9999
39.999
399.99
750.0
399.9
3.999
39.99
399.9
750
Accuracy (1 year; 23 ºC ± 5 ºC)
30 Hz to 50 Hz
1% + 40
1% + 20
1% + 20
Not specified
Not specified
Accuracy (1 year; 23 ºC ± 5 ºC)
20 Hz to 45 Hz
1% + 100
1% + 100
1% + 100
1% + 100
[4]
1% + 100
[2]
1% + 40
1% + 40
1% + 40
[2]
1% + 40
[2]
1% + 40
1% + 5
1% + 5
1% + 5
[2]
1% + 5
[2]
1% + 5
[1]
50 Hz to 10 kHz
0.05% + 40
0.35% + 15
0.35% + 15
0.5% + 15
0.5% + 15
[3]
[1]
45 Hz to 10 kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
10 kHz to 30 kHz
2% + 60
1% + 20
1% + 20
[2]
1% + 20
[2]
1% + 20
10 kHz to 30 kHz
1.5% + 100
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 80
1% + 40
1% + 40
1% + 40
1% + 40
1.5% + 10
1% + 5
1% + 5
1% + 5
1% + 5
30 kHz to 100 kHz
3% + 20
3% + 50
3% + 50
3% + 50
[2]
Not specified
30 kHz to 100 kHz
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
5% + 120
3% + 80
3% + 80
3% + 80
3% + 80
5% + 15
3% + 10
3% + 10
3% + 10
3% + 10
[2]
[2]
[2]
[2]
[3]
[2]
[2]
[2]
[2]
[3]
[2]
[2]
[2]
[2]
[3]
[1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] Input voltage <500 V RMS. [4] In VAC 750 V range, 787.5 V RMS is readable.
5
AC voltage (true RMS, AC+DC coupling mode)
AC+DC voltage resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
750.0 V
[1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] In VAC 750 V range, 1000.0 V RMS is readable with audio warning. [4] For 5 kHz to 10 kHz, accuracy is 0.7% + 25.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
[3]
1000.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
750 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
Maximum reading
119.999
1.19999
11.9999
119.999
750.000
399.99
3.9999
39.999
399.99
750.00
399.9
3.999
39.99
399.9
750.0
Accuracy (1 year; 23 ºC ± 5 ºC)
50 Hz to 10 kHz
0.5% + 50
0.5% + 25
0.5% + 25
0.5% + 25
0.5% + 25
[4]
Accuracy (1 year; 23 ºC ± 5 ºC)
45 Hz to 10 kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
[2]
0.2% + 100
0.2% + 45
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
[1]
10 kHz to 30 kHz
2% + 70
1% + 30
1% + 30
1% + 30
1% + 30
[2]
[2]
[1]
10 kHz to 30 kHz
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 83
1% + 43
1% + 43
1% + 43
1% + 43
1.5% + 12
1% + 7
1% + 7
1% + 7
1% + 7
30 kHz to 100 kHz
3% + 130
3% + 60
3% + 60
[2]
3% + 60
Not specified
30 kHz to 100 kHz
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
[3]
5% + 125
3% + 83
3% + 83
3% + 83
3% + 83
[3]
5% + 18
3% + 12
3% + 12
3% + 12
[3]
3% + 12
[1] Accuracy specified at input >5% of full scale. [2] In VAC 750 V range, 787.5 V RMS is readable. [3] Input voltage <500 V RMS.
6
Loading...