Agilent U3400 Series 4½ and 5½ Digit
Digital Multimeters
Data Sheet
Basic + Good = Elegant Simplicity
Features
Up to 119,999 counts resolution
Up to 0.012% basic DCV accuracy
11 basic measurements and up to
six built-in math functions
Dual display on bright VFD
Selectable resolutions for variable
measurement speeds*
Kensington lock slot security
Simply right for your basic needs,
without compromising quality
The U3400 Series digital multimeters
(DMMs) come with functions vital for your
everyday measurements: DC, AC and AC+
DC voltage and current, 2- and 4-wire
resistance, frequency, continuity and
diode tests.
Math functions dBm, Min/Max, Relative,
Compare, Hold and Percentage** are also
available to ease your measurement
analysis.
Not only are the U3400 Series built robust
for lasting reliability over the long haul,
these DMMs also ensure dependable
measurements with up to 0.012% basic
DCV accuracy.
Efficient testing with dual display
and selectable resolutions*
Dual display on the U3400 Series lets you
view two parameters simultaneously as you
measure, and further enhances your
troubleshooting tasks. For example, the
ability to see both AC voltage and
frequency simultaneously would help you
to measure the frequency response of
amplifier circuits more efficiently and
effectively. Refer to Page 3 for more typical
dual display combinations and applications.
Up to three measurement speeds are
available on the U3402A: Slow, Medium and
Fast. This means greater flexibility in
catering to different testing needs: a faster
measurement speed at lower resolution or
a higher resolution at slower measurement
speed.
Physical security
Your instruments may be at risk of theft or
misplacement whenever you leave them
unattended on the bench. With the U3400
Series rear Kensington lock slot, you can
secure your DMM and be assured that it is
where you expect it to be for your continued
testing the next day.
* U3402A only
** U3401A only
Take a closer look
Primary displaySecondary display
Input terminals
and current fuse
Measurement
function
keypads
Autoranging,
manual range
and comparator
operation
operation
keypads
ShiftMath
Power
on/off
switch
Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with slight differences in the position of certain functions. Please
refer to the U3401As Users and Service Guide U3401-90001 for details.
AC line
voltage selector
Line voltage fuse
selection table
AC power
connector
Figure 2 Rear panel of the U3401A/U3402A.
AC line fuseChassis ground lugRS232 interface connector
(For calibration use only)
2
Typical dual display combinations and applications
Primary display
DCV
ACV + DCV
DCV
DCV
ACI + DCI
ACV
ACI + DCI
ACV
ACV
ACI
DCI
ACI + DCI
dBm
dBm
dBm
dBm
Secondary display
ACV
DCV
DCI
ACI
DCV
DCI
ACV
ACI
Hz
Hz
ACI
DCI
Reference W
DCV
ACV
Hz
Application
Testing of DC-to-AC or AC-to-DC converter circuit
Power supply DC level and AC ripple measurement
Testing of power supply load regulation
Loop current and voltage drop level checking
Testing of line and load regulation
Testing of AC-to-DC or DC-to-AC converter
Power supply DC level and AC ripple measurement
Transformer testing
Amplifier circuits AC frequency response measurement
Adjustment of AC motor control
Power supply AC ripple and DC current measurement
Current dissipation measurement
RF frequency measurement
Frequency response checking
3
U3400 Series specifications
DC voltage
DCV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
1000.0 V
[1] Input impedance is in parallel with capacitance <100 pF.
[2] In VDC 1000 V range, 1200 V is readable with audio warning.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
1200.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
1000.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
1000.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
1000 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
[2]
Accuracy (1 year; 23 ºC ± 5 ºC)
0.02% + 4
0.02% + 4
0.02% + 4
0.02% + 4
0.02% + 4
Maximum reading
119.999
1.19999
11.9999
119.999
[3]
1000.00
399.99
3.9999
39.999
399.99
[3]
1000.0
399.9
3.999
39.99
399.9
[3]
1000
Accuracy
(1 year; 23 ºC ± 5 ºC)
0.012% + 8
[2]
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 5
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
0.012% + 2
Typical input impedance
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
Typical input
impedance
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
10.0 MW
11.1 MW
10.1 MW
10.0 MW
10.0 MW
[1]
[1]
[1] Input impedance is in parallel with capacitance <120 pF.
[2] Relative (REL) operation is used.
[3] In VDC 1000 V range, 1050 V is readable.
4
AC voltage (true RMS, AC coupling mode)
ACV resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
750.0 V
[1] Accuracy specified at input >5% of full scale.
[2] Input voltage <200 V RMS.
[3] For 5 kHz to 10 kHz, accuracy is 0.7% + 15.
[4] In VAC 750 V range, 1000.0 V RMS is readable with audio warning.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
1000.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
750 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
Maximum
reading
119.999
1.19999
11.9999
119.999
750.00
399.99
3.9999
39.999
399.99
750.0
399.9
3.999
39.99
399.9
750
Accuracy (1 year; 23 ºC ± 5 ºC)
30 Hz to 50 Hz
1% + 40
1% + 20
1% + 20
Not specified
Not specified
Accuracy (1 year; 23 ºC ± 5 ºC)
20 Hz to 45 Hz
1% + 100
1% + 100
1% + 100
1% + 100
[4]
1% + 100
[2]
1% + 40
1% + 40
1% + 40
[2]
1% + 40
[2]
1% + 40
1% + 5
1% + 5
1% + 5
[2]
1% + 5
[2]
1% + 5
[1]
50 Hz to 10 kHz
0.05% + 40
0.35% + 15
0.35% + 15
0.5% + 15
0.5% + 15
[3]
[1]
45 Hz to 10 kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 40
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
0.2% + 5
10 kHz to 30 kHz
2% + 60
1% + 20
1% + 20
[2]
1% + 20
[2]
1% + 20
10 kHz to 30 kHz
1.5% + 100
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 80
1% + 40
1% + 40
1% + 40
1% + 40
1.5% + 10
1% + 5
1% + 5
1% + 5
1% + 5
30 kHz to 100 kHz
3% + 20
3% + 50
3% + 50
3% + 50
[2]
Not specified
30 kHz to 100 kHz
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
5% + 120
3% + 80
3% + 80
3% + 80
3% + 80
5% + 15
3% + 10
3% + 10
3% + 10
3% + 10
[2]
[2]
[2]
[2]
[3]
[2]
[2]
[2]
[2]
[3]
[2]
[2]
[2]
[2]
[3]
[1] Accuracy specified at input >5% of full scale.
[2] Input voltage <200 V RMS.
[3] Input voltage <500 V RMS.
[4] In VAC 750 V range, 787.5 V RMS is readable.
5
AC voltage (true RMS, AC+DC coupling mode)
AC+DC voltage resolution, full-scale reading and accuracy [± (% of reading + count)]
U3401A 4.5-digit DMM
Range
500.00 mV
5.0000 V
50.000 V
500.00 V
750.0 V
[1] Accuracy specified at input >5% of full scale.
[2] Input voltage <200 V RMS.
[3] In VAC 750 V range, 1000.0 V RMS is readable with audio warning.
[4] For 5 kHz to 10 kHz, accuracy is 0.7% + 25.
Resolution
10 µV
100 µV
1 mV
10 mV
100 mV
Maximum reading
510.00
5.1000
51.000
510.00
[3]
1000.0
U3402A 5.5-digit DMM
Rate
Slow
Medium
Fast
Range
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
400.00 mV
4.0000 V
40.000 V
400.00 V
750.0 V
400.0 mV
4.000 V
40.00 V
400.0 V
750 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
10 µV
100 µV
1 mV
10 mV
100 mV
100 µV
1 mV
10 mV
100 mV
1 V
Maximum
reading
119.999
1.19999
11.9999
119.999
750.000
399.99
3.9999
39.999
399.99
750.00
399.9
3.999
39.99
399.9
750.0
Accuracy (1 year; 23 ºC ± 5 ºC)
50 Hz to 10 kHz
0.5% + 50
0.5% + 25
0.5% + 25
0.5% + 25
0.5% + 25
[4]
Accuracy (1 year; 23 ºC ± 5 ºC)
45 Hz to 10 kHz
0.2% + 100
0.2% + 100
0.2% + 100
0.2% + 100
[2]
0.2% + 100
0.2% + 45
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 43
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
0.2% + 7
[1]
10 kHz to 30 kHz
2% + 70
1% + 30
1% + 30
1% + 30
1% + 30
[2]
[2]
[1]
10 kHz to 30 kHz
1.5% + 300
1% + 100
1% + 100
1% + 100
1% + 100
1.5% + 83
1% + 43
1% + 43
1% + 43
1% + 43
1.5% + 12
1% + 7
1% + 7
1% + 7
1% + 7
30 kHz to 100 kHz
3% + 130
3% + 60
3% + 60
[2]
3% + 60
Not specified
30 kHz to 100 kHz
5% + 300
3% + 200
3% + 200
3% + 200
3% + 200
[3]
5% + 125
3% + 83
3% + 83
3% + 83
3% + 83
[3]
5% + 18
3% + 12
3% + 12
3% + 12
[3]
3% + 12
[1] Accuracy specified at input >5% of full scale.
[2] In VAC 750 V range, 787.5 V RMS is readable.
[3] Input voltage <500 V RMS.
6
Loading...
+ 13 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.