No part of this document may be
photocopied, reproduced, or translated to
another language without the prior
agreement and written consent of Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Warranty
The material contained in this document
is provided “as is,” and is subject to
being changed, without notice, in future
editions. Further, to the maximum extent
permitted by applicable law, Agilent
disclaims all warranties, either express or
implied, with regard to this manual and
any information contained herein,
including but not limited to the implied
warranties of merchantability and fitness
for a particular purpose. Agilent shall not
be liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance
of this document or of any information
contained herein. Should Agilent and the
user have a separate written agreement
with warranty terms covering the material
in this document that conflict with these
terms, the warranty terms in the separate
agreement shall control.
Manual Editions
Manual Part Number: 5969-2908
Fifth Edition, January, 2005
Printed in Malaysia.
Reprints of this manual containing minor
corrections and updates may have the
same printing date. Revised editions are
identified by a new printing date.
Certification
Agilent Technologies certifies that this
product met its published specifications
at time of shipment from the factory.
Agilent Technologies further certifies that
its calibration measurements are
traceable to the United States National
Institute of Standards and Technology, to
the extent allowed by the Institute's
calibration facility, and to the calibration
facilities of other International Standards
Organization members.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE
THE CUSTOMER'S SOLE AND EXCLUSIVE
REMEDIES. AGILENT TECHNOLOGIES
SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER
BASED ON CONTRACT, TORT, OR ANY
OTHER LEGAL THEORY.
Assistance
This product comes with the standard
product warranty. Warranty options,
extended support contacts, product
maintenance agreements and customer
assistance agreements are also available.
Contact your nearest Agilent
Technologies Sales and Service office for
further information on Agilent
Technologies' full line of Support
Programs.
Technologies Licenses
The hardware and or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
U.S. Government Restricted
Rights
Software and technical data rights
granted to the federal government include
only those rights customarily provided to
end user customers. Agilent provides this
customary commercial license in
Software and technical data pursuant to
FAR 12.211 (Technical Data) and 12.212
(Computer Software) and, for the
Department of Defense, DFARS 252.2277015 (Technical Data – Commercial
Items) and DFARS 227.7202-3 (Rights in
Commercial Computer Software or
Computer Software Documentation).
Trademarks
Microsoft and Windows are U.S.
registered trademarks of Microsoft
Corporation.
2 Series N6700 User’s Guide
Safety Notices
The following general safety precautions
must be observed during all phases of
operation of this instrument. Failure to
comply with these precautions or with
specific warnings or instructions
elsewhere in this manual violates safety
standards of design, manufacture, and
intended use of the instrument. Agilent
Technologies assumes no liability for the
customer's failure to comply with these
requirements.
General
Do not use this product in any manner
not specified by the manufacturer. The
protective features of this product may be
impaired if it is used in a manner not
specified in the operation instructions.
Before Applying Power
Verify that all safety precautions are
taken. Make all connections to the unit
before applying power. Note the
instrument's external markings described
under "Safety Symbols"
Ground the Instrument
This product is a Safety Class 1
instrument (provided with a protective
earth terminal). To minimize shock
hazard, the instrument chassis and cover
must be connected to an electrical
ground. The instrument must be
connected to the ac power mains through
a grounded power cable, with the ground
wire firmly connected to an electrical
ground (safety ground) at the power
outlet. Any interruption of the protective
(grounding) conductor or disconnection of
the protective earth terminal will cause a
potential shock hazard that could result in
personal injury.
Fuses
The instrument contains an internal fuse,
which is not customer accessible.
Do Not Operate in an Explosive
Atmosphere
Do not operate the instrument in the
presence of flammable gases or fumes.
Do Not Remove the Instrument
Cover
Only qualified, service-trained personnel
who are aware of the hazards involved
should remove instrument covers. Always
disconnect the power cable and any
external circuits before removing the
instrument cover.
Do Not Modify the Instrument
Do not install substitute parts or perform
any unauthorized modification to the
product. Return the product to an Agilent
Sales and Service Office for service and
repair to ensure that safety features are
maintained.
In Case of Damage
Instruments that appear damaged or
defective should be made inoperative and
secured against unintended operation
until they can be repaired by qualified
service personnel.
CAUTION
A CAUTION notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that, if
not correctly performed or adhered to,
could result in damage to the product
or loss of important data. Do not
proceed beyond a CAUTION notice
until the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or the
like that, if not correctly performed
or adhered to, could result in
personal injury or death. Do not
proceed beyond a WARNING notice
until the indicated conditions are
fully understood and met.
Safety Symbols
Direct current
Alternating current
Both direct and alternating
current
Three phase alternating
current
Earth (ground) terminal
Protective earth ground
terminal.
Frame or chassis
terminal
Terminal is at earth
potential.
Neutral conductor on
permanently installed
equipment
Line conductor on
permanently installed
equipment.
On supply
Off supply
Standby supply. Unit is not
completely disconnected
from ac mains when switch
is off
In position of a bi-stable
push switch
Out position of a bi-stable
push switch
Caution, risk of electric
shock
Caution, hot surface
Caution, refer to
accompanying
description
Series N6700 User’s Guide 3
In this Book
Specific chapters in this manual contain the following information:
Quick Reference – Chapter 1 is a quick reference section that
helps you quickly become familiar with your Agilent N6700
Modular Power System. It describes the differences between the
various modules in the power system.
Installation – Chapter 2 describes how to install your power
system. It describes how to connect various loads to the output. It
discusses remote sensing as well as parallel and series operation.
Getting Started – Chapter 3 describes how to set the voltage,
current, over-voltage protection, and turn on the output. It also
describes how to configure the remote interface.
Operating the Power System – Chapter 4 describes how to use
the advanced features of the power system using the front panel
menus and the corresponding SCPI commands.
Introduction to Programming – Chapter 5 gives a brief overview
of the SCPI command structure and basic programming concepts.
Language Dictionary – Chapter 6 describes all of the SCPI
example programs that illustrate some common applications.
Specifications – Appendix A describes specifications and
supplemental characteristics.
Verification and Calibration Procedures – Appendix B explains
the verification and calibration procedures.
Using the Digital Port – Appendix C describes how to configure
and use the digital port on the back of the instrument.
Service – Appendix D describes what to do if service is required.
Compatibility – Appendix E documents the compatibility
commands of the Agilent Series 662xA DC power supplies that
are supported by the Agilent N6700 Modular Power System.
You can contact Agilent Technologies at one of the following telephone
numbers for warranty, service, or technical support information.
In the United States: (800) 829-4444
In Europe: 31 20 547 2111
In Japan: 0120-421-345
Or use our Web link for information on contacting Agilent in your country or
specific location: www.agilent.com/find/assist
Or contact your Agilent Technologies Representative.
The web contains the most up to date version of the manual. Go to
http://www.agilent.com/find/N6700
to get the latest version of the manual.
4 Series N6700 User’s Guide
Contents
1 Quick Reference
The Agilent N6700 Modular Power System – At a Glance..........................8
The Front Panel - At a Glance......................................................................... 10
The Rear Panel – At a Glance.........................................................................10
Front Panel Display – At a Glance .................................................................11
Front Panel Keys – At a Glance...................................................................... 12
Front Panel Menu Reference ..........................................................................13
2 Installation
General Information.......................................................................................... 16
Inspecting the Unit ........................................................................................... 17
Installing the Unit..............................................................................................17
Connecting the Line Cord ................................................................................ 19
Connecting the Outputs................................................................................... 20
Remote Sense Connections............................................................................ 24
Differences in Earlier Agilent N6700A Mainframes..................................197
Index ...........................................................................................................................................................201
6 Series N6700 User’s Guide
1
Quick Reference
The Agilent N6700 Modular Power System – At a Glance.......................... 8
The Front Panel - At a Glance......................................................................... 10
The Rear Panel – At a Glance......................................................................... 10
Front Panel Display – At a Glance .................................................................11
Front Panel Keys – At a Glance...................................................................... 12
Front Panel Menu Reference.......................................................................... 13
NOTE
This chapter concisely describes the operation of the Agilent N6700
Modular Power System (MPS).
This chapter does not describe every operating feature in detail. It is
simply a quick reference guide to quickly become familiar with the
essential operating features of the power system.
A quick reference programming command chart is included in the
beginning of chapter 6.
Unless otherwise noted, the Agilent N6700 Modular Power System will also be
referred to as “MPS” and “power system” throughout this manual.
Series N6700 User’s Guide 7
1 Quick Reference
The Agilent N6700 Modular Power System – At a Glance
The Agilent N6700 Modular Power System is a configurable, 1U (rack
unit) high platform that lets you mix and match power modules to
create a power system optimized for your test system requirements.
Up to four power modules can be installed in each Agilent N6700A/B
MPS mainframe. Power modules come in power levels of 50 and 100
Watts, have various voltage and current combinations, and provide
the following output performance levels:
The N675xA High-Performance, Autoranging DC Power Modules
provide low noise, high accuracy, fast programming times, and
advanced programming and measurement capabilities to speed
test throughput.
The N676xA Precision DC Power Modules provide precise control
and measurements in the milli- and micro-ampere region with
the ability to simultaneously digitize voltage and current and
capture those measurements into an oscilloscope-like data buffer.
The N673xA/B and N674xA/B DC Power Modules provide
programmable voltage and current, measurement, and protection
features, making these economical modules suitable for powering
the device-under-test or system resources such as fixture
controls.
Output Features
Programmable
voltage and
current
Fast command
processing
Fast up/down
programming
Fast transient
response
Low output
noise
Autoranging
capability
Full programming capability is provided for the entire range of output voltage and
current. O
sources.
Command processing time of less than 1 millisecond per command.
1.5 millisecond response time from 10% to 90% of the output rating for autoranging
and precision power modules. Refer to Appendix A for details.
Transient response is less than 100 microseconds for autoranging and precision
power modules. Refer to Appendix A for details.
Output noise is typically 4 mV peak-to-peak for autoranging and precision power
modules, which is comparable to linear supplies. Refer to Appendix A for details.
Autoranging produces the maximum rated power over a wide and continuous range
of voltage and current settings for autoranging and precision power modules. Refer
to Appendix A for details.
The output and system features are described in the following
sections. Not all output features are available on every power
module. The “Model Differences” section describes the features that
apply only to specific power modules.
utputs can operate as either constant voltage (CV) or constant current (CC)
Output On/Off
sequencing
8 Series N6700 User’s Guide
A turn-on/turn-off delay capability for each output allows output on/off sequencing.
Quick Reference 1
Remote voltage
sensing
Two remote sensing terminals are provided for each output. When shipped from the
factory, the remote sense jumpers are included in a separate bag. Refer to Chapter 2
for details.
Voltage
All power modules can measure their own output voltage and current.
and current
measurements
Voltage, current,
and temperature
protection
Each output has over-voltage, over-current, and over-temperature protection. Overvoltage and over-current protection are programmable. When activated, the
protection circuits cause the voltage to go to zero, the output to be disabled, and the
protection status to be reported.
System Features
SCPI language The instrument is compatible with the Standard Commands for Programmable
Instruments (SCPI).
Choice of three
interfaces
Front panel I/O
setup
GPIB (IEEE-488), LAN, and USB (mini B) remote programming interfaces are
built in.
Menus let you set up GPIB and LAN parameters from the front panel. Refer to
Chapter 3 for details.
Built-in Web
server
Real-time status
information
Module
identification
A built-in Web server lets you control the instrument directly from an internet
browser on your computer. Refer to Chapter 3 for details.
The front panel indicates the status of each output. It also indicates when a
protection shut-down has occurred.
Each module has identifying data stored in non-volatile memory. Information
includes model number, serial number, and options. This information can be
displayed on the front panel.
Model Differences
Feature DC Power Modules (A+B) Autoranging Modules Precision Modules
N6731 -
N6736
Output power rating 50 W1 100 W2 50 W 100 W 50 W 100 W
Autoranging output capability NO NO YES YES YES YES
Precision output and measurement capability NO NO NO NO YES YES
Low voltage output and measurement range NO NO NO NO YES YES
Low current output and measurement range NO NO NO NO YES YES
Simultaneous voltage and current measurement NO NO NO NO YES YES
Output list capability (Test Extensions) NO NO Option Option YES YES
Array readback capability (Test Extensions) NO NO Option Option YES YES
Programmable sample rate (Test Extensions) NO NO Option Option YES YES
1
Model N6735A has a maximum output of 40 W.
2
Models N6742A and N6745A have a maximum output of 80 W.
N6741 N6746
N6751A N6752A N6761A N6762A
Series N6700 User’s Guide 9
1 Quick Reference
The Front Panel - At a Glance
Display
Turns off after 1 hour of
inactivity. Press any key to
restore the display.
N6700A Modular Power System
20.007V 4.004A
CV Set: 20.000V 5.500A
o -
1
On/Off switch and LED
LED indicates power is on.
Green = normal operation.
Amber = display is screen saver mode.
The Rear Panel – At a Glance
Navigation keys
Move the cursor to a menu item.
Select the highlighted menu item.
Menu
Meter
Channel
Help
Back
Error
On/Off
Voltage
Sel
Current
System keys
Toggle between meter mode and the
command menus.
Exit a menu and return to meter mode.
Select an output channel to display.
Output keys
Turn the outputs on or off.
Enter voltage or current.
987
456
2
1
3
E
0
.
Enter
+/-
Numeric entry keys
Enter values.
Increment or decrement
values.
+s + -s
+s + -s
4-pin output connector.
Includes +/−output and
+/− sense terminals.
WARNING
SHOCK HAZARD The power cord provides a chassis ground through a third
conductor. Be certain that your power outlet is of the three-conductor type
with the correct pin connected to earth ground.
GPIB connector
+s + -s +s + -s
8-pin digital control
connector
Connector function is
user-configurable.
(N6700A mainframes
use a 4-pin connector.)
Chassis ground
binding post
1 2 3 4 5 6 7
3-pin IEC 320 AC
input connector
Power cord requires
ground conductor.
USB connector LAN connector
10/100 Base-T
Left LED indicates
activity. Right LED
indicates link integrity.
10 Series N6700 User’s Guide
Front Panel Display – At a Glance
Single-channel View Voltage measurement Current measurement
Press the Meter key
to toggle between
views
Operating mode (CV =
constant voltage mode)
Multiple-channel View Voltage and Current measurements
Press the Meter key
to toggle between
views
Voltage and Current
settings
Quick Reference 1
Remote interface status
(ALL, SRQ, ERR, IO)
The highlighted channel is the active channel
Grouped-channel View Channels 2 through 4 are connected in parallel and have been
configured or grouped to act as a single, higher-power channel
Refer to Chapter 4,
under “System Related
Operations” for more
information
Grouped channels are addressed using the channel number of the
lowest channel in the group
Series N6700 User’s Guide 11
1 Quick Reference
Front Panel Keys – At a Glance
System Keys
Meter returns the display to metering mode.
Back
Sel
On/Off
Voltage
Current
Menu
Menu accesses the command menu.
Channel selects or highlights a channel to control.
Back backs out of a menu without activating any changes.
Help accesses information about the displayed menu control.
Error displays any error messages in the error queue.
The Arrow keys let you move around in the command menus.
The Select key lets you make a selection in the command menus.
It also lets you enter edit mode for numeric parameters.
On/Off controls the selected output (or all outputs when ALL is lit).
This key is only active in Single- channel or Multiple-channel view.
Voltage lets you change the voltage setting of the selected channel.
Current lets you change the current setting of the selected channel.
The Enter key enters a value. If you exit a field without pressing the
Enter key, the value is ignored.
12 Series N6700 User’s Guide
Front Panel Menu Reference
Quick Reference 1
NOTE
Menu commands that appear grayed-out on the display are either not available
for the power module that is being programmed, or are password protected.
Refer to appendix E for information about the front panel menu commands for
firmware revisions prior to B.00.00
Menu Command Control Description
Output Voltage Programs voltage setting and range.
Current Programs current setting and range.
Delay Programs Turn-on /Turn off delay.
Slew Programs voltage slew rate.
Measure Range Selects voltage and current measurement range.
Sweep Specifies measurement points, time interval, and trigger offset.
Series Connections........................................................................................... 26
This chapter describes how to install your power system. It discusses
installation, rack mounting, and line cord connections.
This chapter discusses how to connect your load to the output
terminals. It discusses what you need to know about wire sizes and
how to compensate for voltage drops in the load leads. It also
discusses various loads configurations and how to connect the output
terminals in series and parallel.
Before installing the instrument, check the list under “Items
Supplied” and verify that you have received these items with your
instrument. If anything is missing, please contact your nearest
Agilent Sales and Support Office.
Series N6700 User’s Guide 15
2 Installation
General Information
Models
Agilent Model Description
N6700A / N6700B MPS Mainframe - without DC Power Modules
N6751A / N6752A 50 W / 100 W High-Performance Autoranging DC Power Module
N6761A / N6762A 50 W / 100 W Precision DC Power Module
N6731B / N6741B 50 W / 100 W 5 V DC Power Module
N6732B / N6742B 50 W / 100 W 8 V DC Power Module
N6733B / N6743B 50 W / 100 W 20 V DC Power Module
N6734B / N6744B 50 W / 100 W 35 V DC Power Module
N6735B / N6745B 50 W / 100 W 60 V DC Power Module
N6736B / N6746B 50 W / 100 W 100 V DC Power Module
Filler Modules Provided if there are less than four power modules installed. N6708A
1253-5826 or
1253-6211 (model N6741B only)
8120-8821 or
0360-2935 (model N6741B only)
1253-6408 (8-pin) or
1253-5830 (4-pin)
Options
Option Description
054 High-speed Test Extensions. Includes digitized output measurements and output list capability.
Available for Agilent Models N6751A/N6752A. Included with Agilent Models N6761A/N6762A.
761 Output relays. Includes 2 - DPDT galvanic-disconnect relays. Disconnects both output and sense
terminals. Available for all Agilent Models.
908 Rack Mount Kit for mounting in a 19-inch EIA rack cabinet. Also available as Model N6709A.
16 Series N6700 User’s Guide
Inspecting the Unit
Installing the Unit
Safety Considerations
Installation 2
When you receive your power system, inspect it for any obvious
damage that may have occurred during shipment. If there is damage,
notify the shipping carrier and nearest Agilent Sales and Support
Office immediately. Refer to Appendix D for more information.
Until you have checked out the power system, save the shipping
carton and packing materials in case the unit has to be returned.
This power system is a Safety Class 1 instrument, which means it has
a protective earth terminal. That terminal must be connected to
earth ground through a power source equipped with a ground
receptacle.
Refer to the Safety Summary page at the beginning of this guide for
general safety information. Before installation or operation, check
the power system and review this guide for safety warnings and
instructions. Safety warnings for specific procedures are located at
appropriate places throughout this Guide.
Environment
WARNING
Do not operate the instrument in the presence of flammable gasses or fumes
The environmental conditions of the instrument are documented in
Appendix A. Basically, the instrument should only be operated
indoors in a controlled environment.
The dimensions of your instrument as well as an outline diagram are
given in Appendix A. A fan cools the power system by drawing air
through the sides and exhausting it out the side and back. The
instrument must be installed in a location that allows sufficient space
at the sides and back of the unit for adequate air circulation.
Rack Installation
Do not block the air intake and exhaust at the sides of the unit or
the exhaust at the rear of the unit. Refer to the outline diagram in
Appendix A.
CAUTION
You cannot use support rails for rack mounting your instrument as they would
block the airflow needed for cooling. Use the Rack Mount kit (Option 908) to
rack mount your instrument. The Rack Mount Kit is also available by ordering
part number N6709A.
Series N6700 User’s Guide 17
2 Installation
The Agilent N6700 MPS can be mounted in a 19-inch EIA rack
cabinet. It is designed to fit in one rack unit (1U) of space. Install the
rack mount kit as shown in the following figure.
Step 1. Install eight clip-nuts on the rack frame (2 in each corner) where
your instrument will be located
Step 2. Install the two front ears and the two rear extender supports on the
instrument as shown in the figure. Use six M3 x 8mm screws (a) for
the front ears and four M3 x 6mm screws (b) for the extender
supports. If the standard extender supports are either too short or
too long, use the longer supports (c). Cut the supports if required (d).
Step 3. Install the two rear ears on the back of the instrument rack as shown
in the figure. Use the four plain 10-32 screws to install the rear ears.
Step 4. Slide the instrument into the rack, making sure that the rear
extender supports are aligned inside the rear ears.
Step 5. Attach the front ears to the front of the instrument rack using the
four dress 10-32 screws provided.
Step 6. This is optional. Insert a plain 10-32 screw through the slot of the
rear ear and extender support. Attach it with a clip-nut. Note that
this will prevent the unit from being slid out of the front of the rack.
4
3
1
6
2b
2d
2c
2a
5
1
18 Series N6700 User’s Guide
Bench Installation
Do not block the air intake and exhaust at the sides, or the exhaust
at the rear of the unit. Refer to the outline diagram in Appendix A.
Minimum clearances for bench operation are 2 inches (51 mm) along
the sides and back.
Channel Number
The channel number of a power module is determined by the location
of that module in the mainframe. When viewed from the rear, the
module next to the GPIB connector is always output channel one.
Numbering continues sequentially to the left, from one to four.
If there are less than four modules, channel numbering corresponds
to the actual number of installed power modules. Unused channel
slots contain filler modules to ensure proper airflow for cooling.
Installation 2
NOTE
Cleaning
WARNING
Connecting the Line Cord
WARNING
Power modules that are connected in parallel and have been configured or
grouped to act as a single, higher-power channel, are addressed using the
channel number of the lowest channel in the group.
SHOCK HAZARDTo prevent electric shock, unplug the unit before cleaning.
Use a dry cloth or one slightly dampened with water to clean the
external case parts. Do not attempt to clean internally.
FIRE HAZARD Use only the power cord that was supplied with your
instrument. Using other types of power cords may cause overheating of the
power cord, resulting in fire.
SHOCK HAZARD The power cord provides a chassis ground through a third
conductor. Be certain that your power outlet is of the three-conductor type
with the correct pin connected to earth ground.
Connect the power cord to the IEC 320 connector on the rear of the
unit. If the wrong power cord was shipped with your unit, contact
your nearest Agilent Sales and Support Office.
The AC input on the back of your unit is a universal AC input. It
accepts nominal line voltages in the range of 100 VAC to 240 VAC.
The frequency can be 50 Hz, 60 Hz, or 400 Hz.
NOTE
Series N6700 User’s Guide 19
The detachable power cord may be used as an emergency disconnecting
device. Removing the power cord will disconnect ac input power to the unit.
2 Installation
400 Hz Operating Considerations
Connecting the Outputs
Power Factor
At 400 Hz operation, the unit’s power factor is affected as follows:
Under full load at 400 Hz, power factor drops from 0.99 (@120
VAC) to as low as 0.76 (@ 265 VAC)
Power factor degrades further under no load conditions.
Redundant Ground Requirement
At 400 Hz operation, the leakage current of the unit exceeds 3.5 mA.
This requires the installation of a permanent, redundant ground from
the instrument chassis to earth ground. This ensures that ground will
always be connected and that any leakage current will be diverted to
ground. Appendix D describes how to connect the redundant ground.
WARNING
SHOCK HAZARD Turn off AC power before making rear panel connections.
All wires and straps must be properly connected with the terminal block
screws securely tightened.
Disconnect the connector plug to make your wire connections. The
connector accepts wires sizes from AWG 12 to AWG 30. Note that
wire sizes smaller than AWG 20 are not recommended. Each
connector has four openings for attaching wires (see the figure
below). Load connections are made at the + and - terminals. Sense
connections are made on the +s and -s terminals. Securely fasten the
wires by tightening the screw terminals.
After your wires are securely connected, insert the connector plug
into the back of the unit and secure it by tightening the locking
screws. A chassis ground binding post is available next to the AC
input connector for your convenience.
TIGHTEN SCREWS
LOCKING SCREW
INSERT WIRES
CONNECTOR
PLUG SHOWN
+S + -S
SENSE JUMPERS
INSTALLED FOR
LOCAL SENSING
TWIST LEADS
+
LOAD
20 Series N6700 User’s Guide
Wire Size
Installation 2
WARNING
FIRE HAZARD Select a wire size large enough to carry short-circuit current
without overheating. To satisfy safety requirements, load wires must be
heavy enough not to overheat while carrying the short-circuit output current
of the unit (refer to the following chart).
Along with conductor temperature, you must also consider voltage
drop when selecting wire sizes. The following chart lists the
resistance for various wire sizes and the maximum lengths to limit
the voltage drop to 1.0 volts for various currents.
Wire size Current-carrying capacity (Amps) Resistance Max. Length to Limit Voltage to 1 V/Lead
2. Capacity for metric wires are derived from IE Publication 335-1.
3. Capacity of aluminum wire is approximately 84% of that listed for copper wire.
4. Because of wire inductance considerations, it is recommended that you keep your load leads twisted, tie
wrapped, or bundled together and less than 50 feet (14.7 meters) in length per lead.
Note that the minimum wire size required to prevent overheating as
shown in the above chart may not be large enough to prevent OV trip
and to maintain good regulation. Under most conditions, the load
wires should be heavy enough to limit the voltage drop to no more
than l.0 V per lead.
NOTE
To help prevent nuisance tripping of the over-voltage circuit, select a wire size
sufficient to handle the FULL output current of the unit no matter what the
intended load current or current limit setting
Load lead resistance is an important factor relating to the CV
stability of the instrument when remote sensing capacitive loads. If
high capacitance loads are expected, you should not use wire gauges
heavier than 12 to 14 AWG for long runs of load lead.
Series N6700 User’s Guide 21
2 Installation
Multiple Loads
If you are using local sensing and are connecting multiple loads to
one output, connect each load to the output terminals using separate
connecting wires (see the figure below). This minimizes mutual
coupling effects and takes full advantage of the power system's low
output impedance. Each pair of wires should be as short as possible
and twisted or bundled to reduce lead inductance and noise pickup.
If load considerations require the use of distribution terminals that
are located away from the instrument, connect the output terminals
to the remote distribution terminals by a pair of twisted or bundled
wires. Connect each load to the distribution terminals separately.
Remote voltage sensing is recommended under these circumstances.
Sense either at the remote distribution terminals or, if one load is
more sensitive than the others, directly at the critical load.
+S + -S
SENSE JUMPERS
INSTALLED FOR
LOCAL SENSING
TWIST LEADS
+ +
LOAD
LOAD
Positive and Negative Voltages
Either positive or negative voltages can be obtained from the output
by grounding (or "commoning") one of the output terminals. Always
use two wires to connect the load to the output regardless of where
or how the system is grounded. The instrument can be operated with
any output terminal ± 240 VDC including output voltage from ground.
Response Time with an External Capacitor
When programming with an external capacitor, voltage response time
may be longer than that specified in Appendix A. Use the following
formula to estimate the additional response time for up
programming:
Response Time = (Added Output Capacitor)X(Change in Vout)
Current Limit Setting
Note that programming into an external output capacitor may cause
the power system to briefly enter constant current or constant power
operating mode, which adds additional time to the estimation.
22 Series N6700 User’s Guide
Remote Voltage Sensing
Because of the unavoidable voltage drop developed in the load leads,
the terminal block strapping patterns discussed thus far do not
provide the best possible voltage regulation at the load. The remote
sensing connections shown in the figure below improve the voltage
regulation at the load by monitoring the voltage there instead of at
the output terminals. This allows the power system to automatically
compensate for the voltage drop in the load leads.
Remote sensing is especially useful for CV operation with load
impedances that vary or have significant lead resistance. It has no
effect during CC operation. Because sensing is independent of other
power system functions, remote sensing can be used regardless of
how the power system is programmed. Note that with remote
sensing, the voltage readback circuit monitors the load voltage
through the sense terminals.
Installation 2
NOTE
The OVP circuit senses at the main output terminals and not through the sense
terminals. Due to the voltage drop in the load leads, the voltage sensed by the
OVP circuit could be higher than the voltage being regulated at the load.
Therefore, you must take into account the additional voltage drop in the load
leads when setting the over-voltage trip point.
+S + -S
TWIST LEADS
TWIST PAIR
+
LOAD
Series N6700 User’s Guide 23
2 Installation
Remote Sense Connections
Remember to turn off the power system before making or changing
any connections on the rear panel terminal blocks. Connect the unit
for remote sensing by first disconnecting the straps between sense
and load terminals. Make your connections as shown in the previous
figure. Connect the sense leads as close to the load as possible. Refer
to the “Wire Size” section for information about selecting the proper
wire size. Best results are obtained by using the shortest load leads
practical. It is recommended that you keep your load leads under
14.7 meters (50 feet) per lead because of inductance effects.
The sense leads carry only a few milliamperes of current and
therefore, can be lighter gauge than the load leads. However, note
that any voltage drop in the sense leads can degrade the voltage
regulation of the instrument. Try to keep the sense lead resistance
less than about 0.5Ω per lead (this requires 20 AWG or heavier for a
50 foot length).
Open Sense Leads
The sense leads are part of the output's feedback path. Connect them
in such a way so that they do not inadvertently become open
circuited. The power system includes protection resistors that reduce
the effect of open sense leads during remote-sensing operation. If the
sense leads open during operation, the power system returns to the
local sensing mode, with the voltage at the output terminals
approximately 1% higher than the programmed value.
Output Noise Considerations
Any noise picked up on the sense leads will appear at the output
terminals and may adversely affect CV load regulation. Twist the
sense leads or use a ribbon cable to minimize the pickup of external
noise. In extremely noisy environments it may be necessary to shield
the sense leads. Ground the shield at the power system end only; do
not use the shield as one of the sensing conductors.
The noise specifications in Appendix A apply at the output terminals
when using local sensing. However, voltage transients may be
produced at the load by noise induced in the leads or by load current
transients acting on the inductance and resistance of the load lead. If
it is desirable to keep voltage transient levels to a minimum, place an
aluminum or a tantalum capacitor, with an approximate value of 10
µF per foot (30.5cm) of load lead, right across the load.
24 Series N6700 User’s Guide
Parallel Connections
Installation 2
CAUTION
Only connect outputs that have identical voltage and current ratings in parallel.
Connecting outputs in parallel provides a greater current capability
than can be obtained from a single output.
The following figures show how to connect two outputs in parallel.
The figure on the left illustrates local sensing. If voltage drop in the
load leads is a concern, the figure on the right shows how to connect
the sense leads directly at the load. Note that in both cases, the
remote sense terminals must be connected together.
OUTPUT 2OUTPUT 1
+S + -S
TWIST LEADS
+
LOAD
+S + -S
SENSE
JUMPERS
INSTALLED
OUTPUT 2
+S + -S
OUTPUT 1
+S + - S
TWIST LEADS
+
LOAD
WITH LOCAL SENSING
Grouping the Outputs
Once outputs have been connected in parallel, they can be configured
or “grouped” to act as a single, higher-power channel. This applies
when programming via the front panel or using SCPI commands.
Information about how to group output channels that have been
connected in parallel is provided in Chapter 4 under “System Related
Operations” as well as Chapter 6 under ”System Commands”.
NOTE
The ability to group outputs is only available on Agilent N6700 MPS mainframes
with firmware revision B.00.00 and up. Almost all instrument functionality is
supported by grouped channels, including voltage and current programming,
measurements, status, step transients, and list transients.
To program paralleled outputs on units with earlier version
firmware, first program both outputs to the desired output voltage.
Then program the current limit point of each output. The current
limit of the paralleled outputs will be the sum of both individual
current limit points.
WITH REMOTE SENSING
Series N6700 User’s Guide 25
2 Installation
Effect on Specifications
Specifications for outputs operating in parallel can be obtained from
the specifications for single outputs. Most specifications are
expressed as a constant or as a percentage (or ppm) plus a constant.
For parallel operation, the percentage portion remains unchanged
while constant portions or any constants are changed as indicated
below. For current readback accuracy and temperature coefficient of
current readback, use the minus current specifications:
Current All parallel specifications referring to current are twice the single output
specification except for programming resolution, which is the same for both
single output and parallel output operation.
Voltage All parallel specifications referring to voltage are the same as for a single
output except for CV load effect, CV load cross regulation, CV source effect,
and CV short term drift. These are all twice the voltage programming accuracy
(including the percentage portion) at all operating points.
Load Transient
Recovery Time
Series Connections
WARNING
CAUTION
Load transient specifications are typically twice the single output.
SHOCK HAZARD Floating voltages must not exceed 240 VDC. No output
terminal may be more than 240 VDC from chassis ground.
Only connect outputs that have identical voltage and current ratings in series.
Each output has reverse voltage protection diodes across its output terminals.
The current conducted by this diode is not internally limited by the output.
Never connect an output in such a way that the diodes will conduct current in
excess of the rated current of the output since damage could result.
Connecting outputs in series provides a greater voltage capability
than can be obtained from a single output. Because the current is the
same through each element in a series circuit, outputs connected in
series must have equivalent current ratings.
The following figure shows an example of how to connect two
outputs in series to a single load with local sensing.. Connecting the +
S terminal of output 2 to the - S terminal of output 1 and removing
the sense jumper (between + S and + V) on output 2 compensates for
the IR drop in the load lead from output 2 to output 1.
If voltage drop in the load leads is a concern, connect the sense leads
of output 1 and output 2 for remote sensing as shown in the figure on
the right. Note that the + sense lead of output 2 must remain
connected to the -sense terminal of output 1. The outputs may be set
as previously described.
26 Series N6700 User’s Guide
Installation 2
OUTPUT 2
+S + -S
TWIST LEADS
+
LOAD
WITH LOCAL SENSING
Setting the Outputs
Outputs connected together in series cannot be grouped.
To program outputs connected in series, first program the current
limit of each output to the total desired current limit point. Then
program the voltage of each output so that the sum of both voltages
equals the total desired operating voltage. The simplest way to
accomplish this is to program each output to one half of the total
desired operating voltage.
OUTPUT 1
+S + -S
SENSE
JUMPERS
INSTALLED
OUTPUT 2
+S + -S
TWIST LEADS
+S + -S
+
LOAD
WITH REMOTE SENSING
OUTPUT 1
INSTALLED
SENSE
JUMPER
NOTE
The operating mode of each output channel is determined by the channel’s
programmed settings, operating point, and load condition. Because these
conditions may change during parallel operation, the status annunciators on the
front panel will reflect these changes. This is normal. Momentary status
changes are also normal.
Effect on Specifications
Specifications for outputs operating in series can be obtained from
the specifications for single outputs. Most specifications are
expressed as a constant or a percentage (or ppm) plus a constant.
For series operation, the percentage portion remains unchanged
while constant portions or any constants are changed as indicated.
Voltage All series specifications referring to voltage are twice the single output
specification except for programming resolution, which is the same as for a
single output.
Current All series specifications referring to current are the same as for a single
output except for CC load effect, CC load cross regulation, CC source effect,
and CC short term drift which are twice the current programming accuracy
(including the percentage portion).
Load Transient
Load transient specifications are typically twice the single output.
Recovery Time
Series N6700 User’s Guide 27
3
Getting Started
Turning the Unit On.......................................................................................... 30
Selecting an Output Channel ..........................................................................30
Entering an Output Voltage Setting ...............................................................30
Entering a Current Limit Setting..................................................................... 31
Enabling the Output.......................................................................................... 31
Using the Front Panel Menu ........................................................................... 32
Connecting to the Interfaces ..........................................................................34
NOTE
This chapter describes how to get started using your power system. It
discusses turning the unit on, using the front panel controls, and
navigating the front panel command menu. A map of the front panel
menu structure is found in chapter 1.
This chapter also contains information on how to configure the three
remote interfaces that are provided on the back of the instrument.
Detailed information on configuring the remote interfaces is included in the
Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide, which is
available on the Automation-Ready CD-ROM included with this product
.
Series N6700 User’s Guide 29
3 Getting Started
Turning the Unit On
After you have connected the line cord, turn the unit on with the
front panel power switch. The front panel display will light up after a
few seconds.
A power-on self-test occurs automatically when you turn the unit on.
This test assures you that the instrument is operational. If the selftest fails, the Err annunciator comes on. Press the Error key to
display the list of errors on the front panel. Refer to Appendix D for
further information.
When the front panel display appears, you can use the front panel
controls to enter voltage and current values.
Selecting an Output Channel
Channel
Press the Channel key to select the output channel that you wish to program.
Use the left and right navigation keys to navigate to the setting that you wish
to change. In the display below, channel 1’s voltage setting is selected. Enter
a value using the numeric keypad. Then press Enter.
You can also use the arrow keys to adjust the value up or down. When the
output is on and the unit is operating in CV mode, the output voltage
changes immediately. Otherwise, the value will become effective when the
output is turned on.
Use the Voltage key to select the voltage entry field. In the display below,
channel 1’s voltage setting is selected. Enter the desired setting using the
numeric keypad. Then press Enter.
NOTE
30 Series N6700 User’s Guide
If you make a mistake, either use the § backspace key to delete the number,
press Back to back out of the menu, or press Meter to return to meter mode.
Loading...
+ 176 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.