THE MA TERIAL CONT AINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT
TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE
MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL
WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE
IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR
PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN.
SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH
WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT
WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL
CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the per formance of a U.S. Government prime contract or
subcontract, Software is delivered and licensed as “Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer sof tware” as defined in FAR 52.227-19 (June 1987) or
any equivalent agency regulation or contract clause. Use, duplication or disclosure of
Software is subject to Agilent Technologies’ standard commercial l icense terms, and
non-DOD Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987) . U.S. Government users
will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or
DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Certification
Agilent T echnol ogies, Inc. certifies that this p roduct met its publi shed sp ecific atio ns at the
time of shipment from the factory. Agilent Technologies, Inc . further certifies that its
calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute's calibration facility, and
to the calibration facilities of other Int ernational Standards Organization members.
ii
Assistance
Product maintenance agreements and other cus tomer assistanc e agreements are a vailable
for Agilent Technologies, Inc. products. For information about these agreements and for
other assistance, contact Agilent. Refer to “Contacting Agilent” on page 1-15.
Safety and Regulatory Information
The safety a n d regulator y informati o n p e rtaining to th i s product is l ocated in the s e ction
“Safety and Regulatory Information” on page 1-18.
Safety Notes
The following safety notes are used in this manual. Familiarize yourself with each of the
notes and its meaning before operating this instrument. All pertinen t saf ety notes for
using this product are located in “Safety and Regulatory Information” on page 1-18
WARNINGWarning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss
of life. Do not proceed beyond a warning note until the indicated
conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of
the instrument. Do not proceed beyond a caution sign until the indicated
conditions are fully understood and met .
Technology Licenses
The hardware and / or software described in this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
iii
Printing Copies of On-Line Documentation
To print copies of documentation from the Web, download the PDF file from the Agilent
web site:
•Go to http://www.agilent.com. En t e r the docume n t’s pa rt nu mber (loca t ed on the titl e
page) in the Search box.
The purpose of this manual is to help you use your Electronic Calibration (ECal) module
confidently and effectively. ECal is a precision, single-connection calibration technique
which uses fully traceable and verifiable electronic standards. Each module has unique
S-parameter data that is stored in the module's memory. During calibration, ECal
transfers this data to the network analyzer to calculate the error terms for your network
analyzer.
This manual provides instructions for operating and maintai ning your ECal module. Also
included are calibration procedures (for supported analyzers), mechanical specifications,
and replaceable parts available for each model.
RF Modules
The 8509x and N4431B series provide two-port and four-port calibration solutions with
operating frequencies starting at 300 kHz and ending as high as 13.5 GHz.
Microwave Modules
The N469xA and N469xB series (which replace the 8506x seri es) an d the N4432A and
N4433A series provide two-port and four-port calibration solutions with operating
frequencies starting as low as 300 kHz and ending as high as 67 GHz.
Prerequisite Knowledge
This manual assumes you are trained in proper connector care. Because an accurate
calibration depends on the integrity and c leanliness of the connector interface, a damaged
connector will invalidate the calibration achieved with that module. Refer to the
“Principles of Microwave Connector Care Quick Reference Card” located at the back of this
manual.
Clarifying the Terminology of a Connector Interface
In this manual, adapters, ECal modules, and gage masters are referred to by way of their
interface connector. For example:
• A female adapter has a female interface.
• A male adapter has a male interface.
A connector gage is referred to by way of the connector it measures. For example:
• A male connector gage has a female interface so it can measure male devices.
• A female connector gage has a male interface so it can measure female devices .
Accuracy of Electronic Calibration versus Mechanical Calibration
As the numb er of DUT test port s increases, the re quired number of calib r ation
measurements also increases. Thi s boosts the probability of bad connections and bad
calibrations when mechanical calibratio n stand ards are used rather than ECal. Also, test
port cable repeatability and stability errors are amplified by the increased number of
1-2Chapter 1
General Information
Manual Overview
connections and disconnections. To per form an SOLT 4-port calibration, a minimum of
twelve 1-port standards and three 2-p ort (thru) standards are measured, requiring a total
of eighteen connections. The same calibration using a 4-port ECal module (N4431B,
N4432A and N4433A) requires only four connections .
Ignoring connection related errors, the following table compares the accuracy of ECal
versus mechanical calibration in various calibration options.
Table 1-1 Accuracy of ECal versus Mechanical Calibration
a. Flush thru should NOT be used when the test port connectors are not the same type as
the calibration standards. SMA and 3.5 mm connectors are mateable but are NOT the
same type.
Chapter 11-3
General Information
Models and Options
Models and Options
This section provides the model number, operating characteristics and connector options
available with Agilent RF and microwave ECal modules. You can order ECal modules by
selecting the model number followed by the desired options.
8509x Modules (RF Two-Port)
ECal modules from this series are available wit h the connector types and fr equency ranges
shown in Table 1-2 on page 1-5. Mixed connector options allow you to conf igure the module
with a different connector type at each port (as shown in Figure 1-2).
00AAdds one -m- to -m- and one -f- to -f- adapter.
UK6Adds commercial calibration certificate with measured data.
Port A OptionPort B Option
Type-f--m-Type-f--m-Type-f--m-
7-16
a
1051063.5 mm201202Type-N 50Ω203204
00M
85099-60010
a
7-16
a
7-16
00F
85099-60011
205206
205206
Chapter 11-5
General Information
Models and Options
N4431B Modules (RF Four-Port)
NOTEThe N4431A is no longer available for purchase, but may be returned to
Agilent for service and recertification.
ECal modules from this series are available wit h the connector types and fr equency ranges
shown in Table 1-4. Mixed connector options (shown in Table 1-5) allow you to configure
the module with a different connector type at each port.
Figure 1 -3 Model N443 1B Op tio n 010
Table 1-4 N4431B Modules
Model
Number
N4431BN4431-60007Type-N 50Ω -f-300 kHz to
Module
Number
N4431-600063.5 mm -f-300 kHz to
N4431-60008Mixed connector types – see
Connector Type Port
A,B,C,D
Table 1-5 on page 1-7
Operating
Frequency
13.5 GHz
13.5 GHz
300 kHz to
13.5 GHz
Option
010
020
See Table 1-5
on page 1-7
1-6Chapter 1
General Information
Models and Options
Table 1-5 N4431B Mixed Connector Options
Connector TypePort A OptionPort B OptionPort C OptionPort D Option
3.5 mm -f-101201301401
3.5 mm -m-102202302402
Type-N 50Ω -f-103203303403
Type-N 50Ω -m-104204304404
7-16 -f7-16 -m-
a
a
105205305405
106206306406
a. Limits ECal module high frequency to 7.5 GHz.
Other OptionsDescription
UK6Adds commercial calibration certificate with measured data.
Chapter 11-7
General Information
Models and Options
N469xA/B Modules (Microwave Two-Port)
NOTEThe N4690/1/6A are no longer available for purchase, but may be ret urned to
Agilent for service and recertification.
ECal modules from this series are available wit h the connector types and fr equency ranges
shown in Table 1-6. Mixed connector options are not av ailable with this series.
Figure 1-4 Model N4693A Option M0F
1-8Chapter 1
Table 1-6 N469xA/B Modules
General Information
Models and Options
Model
Number
N4690BType-N 50Ω300 kHz to 18 GHzM0F00M00F
N4691B3.5 mm300 kHz to 26.5 GHzM0F00M00F
N4692A2.92 mm10 MHz to 40 GHzM0F00M00F
N4693A2.4 mm10 MHz to 50 GHzM0F00M00F
N4694A1.85 mm10 MHz to 67 GHzM0F00M00F
Connector
Type
Operating
Frequency
Port A and B Option
-m- and -f--m- and -m--f- and -f-
Module Part Number
N4690-60004N4690-60005N4690-60006
Module Part Number
N4691-60004N4691-60005N4691-60006
Module Part Number
N4692-60001N4692-60002N4692-60003
Module Part Number
N4693-60001N4693-60002N4693-60003
Module Part Number
N4694-60001N4694-60002N4694-60003
N4696B7 mm300 kHz to 18 GHzNo Connector Option
Module Part Number
N4696-60004
Other OptionsDescription
00AAdds one -m- to -m- and one -f- to -f- adapter (not available with N4696B).
UK6Adds commercial calibration certificate with measured data.
Chapter 11-9
General Information
Models and Options
8506x Modules
ECal modules in this series are no longer available for purchas e, but may be returned to
Agilent for service and recertification. The 8506x series has been replaced by the
N469xA/B series. See “N469xA/B Modules (Microwave Two-Port)” on page 1-8.
Figure 1-5 Model 85062B
Table 1-7 8506x Modules
Model
Number
85060B7 mm1 to 18 GHz
85062B3.5 mm1 to 26.5 GHz
85064BType-N 50Ω1 to 18 GHz
Connector
Type
Operating
Frequency
8506xA series modules with serial numbers below 800 need to be retrofitted for operation
with the 85097B VNA interface kit. The retrofitting must be done by a qualified Agilent
Technologies service office. See “Contacting Agilent” on page 1-15.
1-10Chapter 1
General Information
N4432A and N4433A Modules (Microwave Four-Port)
N4432A and N4433A Modules (Microwave Four-Port)
ECal modules from this series are available wit h the connector types and fr equency ranges
shown in Table 1-8. Mixed connector options (shown in Table 1-9) allow you to configure
the module with a different connector type at each port.
Figure 1-6 Model N4433A Option 010
Table 1-8 N4432A and N4433A Modules
Model
Number
N4432AN4432-60003Type-N 50Ω -f-300 kHz to 18 GHz020
N4433AN4433-600033.5 mm -f-300 kHz to 20 GHz010
Module
Number
N4432-60005Mixed connector types – see
N4433-60004Mixed connector types – see
Connector Type Port
A,B,C,D
Table 1-9 on page 1-12
Table 1-10 on page 1-12
Operating
Frequency
300 kHz to 18 GHzSee Table 1-9
300 kHz to 20 GHzSee
Option
on page 1-12
Table 1-10 on
page 1-12
Chapter 11-11
General Information
N4432A and N4433A Modules (Microwave Four-Port)
Table 1-9 N4432A Mixed Connector Options
Connector TypePort A OptionPort B OptionPort C OptionPort D Option
a. Frequency range of ECal module limited by other connectors.
Other OptionsDescription
UK6Adds commercial calibration certificate with measured data.
1A7Adds ISO 17025 compliant calibration.
A6JAdds ANSI Z540 compliant calibration.
Table 1-10 N4433A Mixed Connector Options
Connector TypePort A OptionPort B OptionPort C OptionPort D Option
3.5 mm -f-101201301401
3.5 mm -m-1 02202302402
Other OptionsDescription
UK6Adds commercial calibration certificate with measured data.
1A7Adds ISO 17025 compliant calibration.
A6JAdds ANSI Z540 compliant calibration.
1-12Chapter 1
General Information
Compatible Network Analyzers
Compatible Network Analyzers
ECal modules operate with the following network analyzer models, as shown in Table 1-11.
Not all ECal module and network analyzer combinations have the same features or
capabilities. See your network analyzer user guide or online Help.
a. Only ports A and B available when operated with 8753/8720-series analyzers.
b. 8506x modules are no longer available for purchase.
c. 8753/8720-series firmware revision 7.68 or higher required.
d. 8753/8720-series firmware revision 7.74 or higher and adapter cable (part no.
8121-1047) required.
e. PNA firmware revision 3.0 or higher required with N4692/3A, and N4694A.
NOTEFor 8753 network analyzers, RF ECal modules (8509x) may be used below 300 kHz.
However, calibration accuracy is not specified. The display t race connecting the two
points on either side of the 300 kHz boundary may misrepresent actual performance.
Chapter 11-13
General Information
Service and Recertification
Service and Recertification
If your ECal module requires service or recertifi cation, contact the Agilent office nearest
you for information about where to send i t. The perf ormance of your ECal mo dule can only
be verified by specially manufactured equipment and calibration s tand ards from Agilent.
1-14Chapter 1
General Information
Service and Recertification
Contacting Agilent
Assistance with test and measurements needs and information on finding a local Agilent
office are available on the Web at:
http://www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE In any correspondence or telephone conversation, refer to the Agilent product
by its model number and full serial number. With this information, the
Agilent representative can determine whether your product is still within its
warranty period.
Chapter 11-15
General Information
Service and Recertification
Returning Devices to Agilent
If you are returning the product to Agilent, please include a service tag (located near the
end of this manual) on which you provide the following information:
• your company name and address
• a technical contact person within your company, and the person’s complete telephone
number including country code and area code
• the model number and serial number of the ECal module
• type of service required
• a detailed description of the problem and how the device was being used when the
problem occurred (such as calibration or measurement)
NOTEWhen transporting the module, use original or comparable packaging.
1-16Chapter 1
General Information
Service and Recertification
Recertification of ECal Modules
The suggested interval for recertification is 12 months. After reviewing the results of the
initial recertification, you may es tablish a shorter interval that reflects greater use and
wear of the module.
Where to Send a Module for Recertification
Contact Agilent for information on where to send your kit for recertification. See
“Contacting Agilent” on page 1-15. Refer to “Returning Devices to Agilent” o n page 1-16 for
instructions on the preparation of returning the device.
How Agilent Verifies Your ECal Module
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test sys tem are verified with precision
airlines and shorts or low frequency resistance that are directly traced to the National
Institute of Standards and Technology (NIST). The airline and short c haract eri stics are
developed from mechanical measurements. The mechanical measurements and
material properties are carefully modeled to give very accurate electrical
representation. The mechanical measurement s are then traced to NIST through
various plug and ring gages and other mechanical measurements.
2. Each module is electrically tested on this system to the specifi cation listed in Chapte r 5,
“Specifications and Characteristics.”
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the
institute’s calib ration facility. The specifications d ata provided for the module are t raceable
to NIST through Agilent Technologies.
What Is Provided with a Recertification
• a new calibra t i on st i cke r a ff i x e d to th e m od u l e
• a certificate of calibration
• a list of United States National Institut e of Standa rds and Technology (NIST) traceable
numbers
• a calibration report for each traceable module listing measured values, specifications,
and uncertainties
• a new set of S-parameter data (embedded in module memory) if the old set of
S-parameters data no longer allows for a calibratio n that me ets all performance
specifications
Agilent Technologies offers different types of calibration for the recertification of the
module. For more information, contact Agilent. See “Contacting Agilent” on page 1-15.
Chapter 11-17
General Information
Safety and Regulatory Information
Safety and Regulat o ry Inf ormation
Review this section to familiarize yourself with safety markings and instructions before
you operate the ECal module. This produc t has been designed and tested in accordance
with international standards.
WARNINGThe WARNING notice denotes a hazard. It calls attention to a
procedure, practice, or the like, that, if not correctly performed or
adhered to, could result in personal injury. Do not proceed beyond a
WARNING notice until the indicated conditions are fully understood
and met.
CAUTIONThe CAUTION noti ce denotes a hazard. It calls attention to an operating
procedure, practice, or the like, that, if not correctly performed or adhered to,
could result in damage to the product or loss of important data. Do not
proceed beyond a CA UTION notice until the indicated conditions are fully
understood and met.
Safety Notices
WARNINGTo prevent electrical shock, discon nect from instrument before
cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
WARNINGIf this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a
normal condition (in which all means for protection are intact) only.
Installation Notices
CAUTIONThis product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
CAUTIONWhen installing the product in a cabinet, the convection in and out of the
product must not be restricted. The ambient temperature (outside the
cabinet) must be less than the maximum operating temperature of the s ystem
by 4 °C for every 100 watts dissipated in the cabinet. If the total power
dissipated in the cabinet is greater than 800 w atts, then forced convection
must be used.
1-18Chapter 1
Instrument Markings
When you see this symbol on your instrument, you should
!
1SM1-AThis text indicates that the instrument is an Industrial
refer to the instrument’s instruction manual for important
information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have
a laser output.
This symbol indicates that the instrument requires
alternating current (ac) input.
The CE mark is a registered trademark of the European
Community. If it is accompanie d by a year, it indicates th e
year the design was proven.
The CSA mark is a registered trademark of the Canadian
Standards Association.
Scientific and Medical Group 1 Class A product (CISPER
11, Clause 4).
General Information
Safety and Regulatory Information
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or
in STANDBY position.
The C-Tick mark is a registered trademark of the
Australian Spectrum Management Agency.
This product complies with the WEEE Directive
(2002/96/EC) marking requirements. The affixed label
indicates that you must not discard this electrical/
electronic product in domestic household waste.
Product Category: With reference to the equipment types in
the WEEE Directive Annex I, this product is classed as a
"Monitoring and Control instrumentation" product.
Do not dispose in domestic household waste.
To return unwanted products, contact your local Agilent
office, or see http://www.agilent.com /envi ronment/p rod uct/
for more information.
Chapter 11-19
General Information
Safety and Regulatory Information
Compliance Notices
This product has been designated and tested in accordance with the standards in the
Declaration of Conformity (see below), and has bee n supp lied in a safe condition. The
documentation contains information and warnings that must be followed by the user to
ensure safe operation and to maintain the product in a safe conditions.
Compliance IEC 61000-4-4: 1995 (electrical transients) and IEC 61000-4-2: 1995
(electrostatic discharge): Passes using Criterion C.
Calibration cycle may be interrupted by power line transients applied to external power
supply or by electrostatic discharge applied to module. Normal function is restored by
power-cycling the module.
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme a la norme NMB du Canada.
Declaration of Conformity
A declaration of conformity is on file for the PNA models, and a copy is available upon
request.
1-20Chapter 1
2Preparing to Use the ECal Module
2-1
Preparing to Use the ECal Module
Inspecting ECal Module Kit
Inspecting ECal Module Kit
Inspect the Kit Contents
Verify the case and its contents are not damaged and that all p arts ar e includ ed (s ee it ems
listed in T able 2- 2, Table 2-3, and Table 2-4). If the case or any device appears damaged, or
if the shipment is incomplete, contact Agilent . Agilent will arrange for repair or
replacement of incomplete or damaged shipments without waiting for settlement from the
transportation company. See “Contacting Agilent” on page 1-15.
Record the Serial Number
ECal modules are individually serialized. Record the se rial number and the date of initial
use as a reminder for annual recertification. See “Recertification of ECal Modules” on
page 1-17.
Table 2-1 Table for Tracking ECal Module Serial Numbers
Module Number Serial NumberDate
Verify Electrical Performance
To verify electrical performance of the module, see “Validating a Calibration” on page 3-7.
Equipment Required but Not Supplied
VNA Interface Kit (85097B)
The VNA Interface Kit allows you to connect an ECal module to supported vector network
analyzers. Microw ave ECal modules (N469xA/B series) require an adapter cable
(part no. 8121-1047) to operat e wit h the 85097B. The adapter cable is included as an
option with the interfa ce kit or can be ordered separately. Refer t o Chapter 6 , “Replaceable
Parts.”
Pin Depth Gages
Gages for measuring recession or protrusion are not provided in the kit. See Chapter 6 ,
“Replaceable Parts.” or contact Agilent to order gages and other tes t accessories.
2-2Chapter 2