Keysight (Agilent) N4691B Reference Guide

Reference Guide
Agilent Technologies
Electronic Calibration Modules
RF Two-P ort
85091C 85092C 85093C 85096C 85098C 85099C
RF Four-Port
N4431B
Microwave Two-Port
N4693A N4694A N4696B
Microwave Four-Port
N4432A N4433A
Agilent Part Number: N4693-90001
Printed in USA
Print Date: March 2006
Supersedes: February 2006
© Copyright Agilent Technologies, Inc. 2000–2006
WARRANTY STA TEMENT
THE MA TERIAL CONT AINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, AGILENT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. AGILENT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD AGILENT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL.
DFARS/Restricted Rights Notice
If software is for use in the per formance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer sof tware” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial l icense terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987) . U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Certification
Agilent T echnol ogies, Inc. certifies that this p roduct met its publi shed sp ecific atio ns at the time of shipment from the factory. Agilent Technologies, Inc . further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other Int ernational Standards Organization members.
ii
Assistance
Product maintenance agreements and other cus tomer assistanc e agreements are a vailable for Agilent Technologies, Inc. products. For information about these agreements and for other assistance, contact Agilent. Refer to “Contacting Agilent” on page 1-15.
Safety and Regulatory Information
The safety a n d regulator y informati o n p e rtaining to th i s product is l ocated in the s e ction
“Safety and Regulatory Information” on page 1-18.
Safety Notes
The following safety notes are used in this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument. All pertinen t saf ety notes for using this product are located in “Safety and Regulatory Information” on page 1-18
WARNING Warning denotes a hazard. It calls attention to a procedure which, if
not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met .
Technology Licenses
The hardware and / or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
iii
Printing Copies of On-Line Documentation
To print copies of documentation from the Web, download the PDF file from the Agilent web site:
•Go to http://www.agilent.com. En t e r the docume n t’s pa rt nu mber (loca t ed on the titl e page) in the Search box.
• Click Search.
• Click on the hyperlink for the document.
• Click the printer icon located in the tool bar.
iv
Contents
1 General Information
Manual Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Prerequisite Knowledge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Clarifying the Terminology of a Connector Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Accuracy of Electronic Calibration versus Mechanical Calibration. . . . . . . . . . . . . . . . . . .1-2
Models and Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
8509x Modules (RF Two-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
N4431B Modules (RF Four-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
N469xA/B Modules (Microwave Two-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
8506x Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-10
N4432A and N4433A Modules (Microwave Four-Port) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Compatible Network Analyzers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-13
Service and Recertification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-14
Contacting Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 -15
Returning Devices to Agilent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-16
Recertification of ECal Modules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-17
Safety and Regulatory Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
Safety Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
Installation Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-18
Instrument Markings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-19
Compliance Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-20
2 Preparing to Use the ECal Module
Inspecting ECal Module Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Inspect the Kit Contents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Record the Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Verify Electrical Performance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Equipment Required but Not Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
8509x Series Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
8509x Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
N4431B Series Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
N4431B Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
N469xA/B Series Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
N469xA/B Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
N4432A and N4433A Series Kit Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
N4432A Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
N4433A Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Operating and Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
ESD Damage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Input Power Level. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Operating Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Handling and Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Connecting and Disconnecting ECal Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Connect ECal Module to PNA or ENA Analyzer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Connect ECal Module to VNA Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12

Table of Contents Contents-v

Contents
3 Operating the ECal Module
Setting Up a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Required Procedure for All Calibrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Calibration Types Available . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
When to Correct for Isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Calibration Methods for Non-Insertable Devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Performing a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Validating a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Why Validate a Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Measure Known Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Perform ECal Confidence Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
4 Use, Maintenance, and Care of the Devices
Electrostatic Discharge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Visual Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Look for Obvious Defects and Damage First . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Inspect the Mating Plane Surfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Inspect Female Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Cleaning Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning Coax Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning Precision 7 mm Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Gaging Connectors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Connector Gage Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
When to Gage Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Recognizing Gage Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Reading the Connector Gage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Gaging Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-13
Making Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
3.5 mm, 2.92 mm, 2.4 mm, 1.85 mm, Type-N, 7-16, and Type-F Devices. . . . . . . . . . . . . 4-15
7 mm Devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Final Connection Using a Torque Wrench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Separating Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-18
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-19
5 Specifications and Characteristics
Specifications: Terminology and Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Environmental Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Operating Temperature and Accuracy Enhancement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Barometric Pressure and Relative Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Characteristic Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-4
8509x Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
N4431B (3.5 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
N4431B (Type-N 50 Ω). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
N4432A (Type-N 50 Ω). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
N4433A (3.5 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
N469xA/B Series. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Mechanical Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Pin Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Contents-vi Table of Contents
Contents
Type-N Characteristics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Typical Pin Depth Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Supplemental Mechanical Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Measurement Uncertainty Calculator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-18
Download the Vector Network Analyzer Uncertainty Calculator . . . . . . . . . . . . . . . . . . .5-18
6 Replaceable Parts
Description of This Chapter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
7 mm Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2
Type-N (50 ohm) Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
3.5 mm Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
7-16 Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
2.92 mm Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
2.4 mm Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
1.85 mm Modules and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
Type-N (75 ohm) Modules and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Type-F Modules and Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-10
Other ECal Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
Table of Contents Contents-vii
Contents
Contents-viii Table of Contents

1 General Information

1-1
General Information

Manual Overview

Manual Overview
The purpose of this manual is to help you use your Electronic Calibration (ECal) module confidently and effectively. ECal is a precision, single-connection calibration technique which uses fully traceable and verifiable electronic standards. Each module has unique S-parameter data that is stored in the module's memory. During calibration, ECal transfers this data to the network analyzer to calculate the error terms for your network analyzer.
This manual provides instructions for operating and maintai ning your ECal module. Also included are calibration procedures (for supported analyzers), mechanical specifications, and replaceable parts available for each model.
RF Modules
The 8509x and N4431B series provide two-port and four-port calibration solutions with operating frequencies starting at 300 kHz and ending as high as 13.5 GHz.
Microwave Modules
The N469xA and N469xB series (which replace the 8506x seri es) an d the N4432A and N4433A series provide two-port and four-port calibration solutions with operating frequencies starting as low as 300 kHz and ending as high as 67 GHz.

Prerequisite Knowledge

This manual assumes you are trained in proper connector care. Because an accurate calibration depends on the integrity and c leanliness of the connector interface, a damaged connector will invalidate the calibration achieved with that module. Refer to the “Principles of Microwave Connector Care Quick Reference Card” located at the back of this manual.

Clarifying the Terminology of a Connector Interface

In this manual, adapters, ECal modules, and gage masters are referred to by way of their interface connector. For example:
• A female adapter has a female interface.
• A male adapter has a male interface.
A connector gage is referred to by way of the connector it measures. For example:
• A male connector gage has a female interface so it can measure male devices.
• A female connector gage has a male interface so it can measure female devices .

Accuracy of Electronic Calibration versus Mechanical Calibration

As the numb er of DUT test port s increases, the re quired number of calib r ation measurements also increases. Thi s boosts the probability of bad connections and bad calibrations when mechanical calibratio n stand ards are used rather than ECal. Also, test port cable repeatability and stability errors are amplified by the increased number of
1-2 Chapter 1
General Information
Manual Overview
connections and disconnections. To per form an SOLT 4-port calibration, a minimum of twelve 1-port standards and three 2-p ort (thru) standards are measured, requiring a total of eighteen connections. The same calibration using a 4-port ECal module (N4431B, N4432A and N4433A) requires only four connections .
Ignoring connection related errors, the following table compares the accuracy of ECal versus mechanical calibration in various calibration options.
Table 1-1 Accuracy of ECal versus Mechanical Calibration
Mechanical Calibration Type ECal Thru
Type
1-Port; calibration constant (not
NA Reflection ++
Measurement ECal Mechanical
Calibration
data-based) SOL 1-Port; short, open, sliding load NA Reflection + 1-Port; data-based, expanded NA Reflection + 2-Port TRL All types Reflection
Transmission
characterized Reflection
Transmission
2-Port; calibration constant SOL; flush Thru
a
internal unknown
external flush
Reflection Transmission
a
Reflection Transmission
2-Port; calibration constant SOL; unknown Thru
characterized Reflection
Transmission
internal unknown
Reflection Transmission
++
++
++
+
++
++
==
++ ++
++
++
2-Port; data-based, expanded, flush Thru
a
2-Port; data-based, expanded, unknown Thru
characterized Reflection
Transmission
internal unknown
external flush
Reflection Transmission
a
Reflection Transmission
characterized Reflection
Transmission
internal unknown
Reflection Transmission
+
++
+ +
+ +
+
++
+ +
++ better; + slightly better; = same
a. Flush thru should NOT be used when the test port connectors are not the same type as
the calibration standards. SMA and 3.5 mm connectors are mateable but are NOT the same type.
Chapter 1 1-3
General Information

Models and Options

Models and Options
This section provides the model number, operating characteristics and connector options available with Agilent RF and microwave ECal modules. You can order ECal modules by selecting the model number followed by the desired options.

8509x Modules (RF Two-Port)

ECal modules from this series are available wit h the connector types and fr equency ranges shown in Table 1-2 on page 1-5. Mixed connector options allow you to conf igure the module with a different connector type at each port (as shown in Figure 1-2).
Figure 1-1 Model 85092C Option M0F
Figure 1-2 Model 85092C Op tion 104, 201
1-4 Chapter 1
Table 1-2 8509x Modules
General Information
Models and Options
Model
Number
Connector
Type
Operating
Frequency
Port A and B Option
-m- and -f- -m- and -m- -f- and -f-
85091C 7 mm 300 kHz to 9 GHz No Connector Option
85091-60004
85092C
Type-N 50 ohm 300 kHz to 9 GHz M0F
85092-60008
Mixed
300 kHz to 9 GHz See Table 1-3
connector types
00M
85092-60009
85092-60004
– see Table 1-3
85093C
3.5 mm 300 kHz to 9 GHz M0F 85093-60008
Mixed
300 kHz to 9 GHz See Table 1-3
connector types
00M
85093-60009
85092-60004
– see Table 1-3
85096C Type-N 75 ohm 300 kHz to 3 GHz M0F
85096-60007
7-16 300 kHz to 7.5 GHz M0F
85098C
Mixed
300 kHz to 7.5 GHz See Table 1-3
85098-60007
connector types
00M
85096-60008
00M
85098-60008
85092-60004
– see Table 1-3
00F
85092-60010
00F
85093-60010
00F
85096-60009
00F
85098-60009
85099C Type-F 300 kHz to 3 GHz M0F
85099-60009
Table 1-3 8509x Mixed Connector Options
Model
Number
85092C Type-N 50 103 104 3.5 mm 201 202 85093C 3.5 mm 101 102 Type-N 50 203 204 85098C
a. Limits ECal module high frequency to 7.5 GHz.
Other Options Description
00A Adds one -m- to -m- and one -f- to -f- adapter.
UK6 Adds commercial calibration certificate with measured data.
Port A Option Port B Option
Type -f- -m- Type -f- -m- Type -f- -m-
7-16
a
105 106 3.5 mm 201 202 Type-N 50 203 204
00M
85099-60010
a
7-16
a
7-16
00F
85099-60011
205 206 205 206
Chapter 1 1-5
General Information
Models and Options

N4431B Modules (RF Four-Port)

NOTE The N4431A is no longer available for purchase, but may be returned to
Agilent for service and recertification.
ECal modules from this series are available wit h the connector types and fr equency ranges shown in Table 1-4. Mixed connector options (shown in Table 1-5) allow you to configure the module with a different connector type at each port.
Figure 1 -3 Model N443 1B Op tio n 010
Table 1-4 N4431B Modules
Model Number
N4431B N4431-60007 Type-N 50 -f- 300 kHz to
Module Number
N4431-60006 3.5 mm -f- 300 kHz to
N4431-60008 Mixed connector types – see
Connector Type Port A,B,C,D
Table 1-5 on page 1-7
Operating Frequency
13.5 GHz
13.5 GHz 300 kHz to
13.5 GHz
Option
010
020
See Table 1-5
on page 1-7
1-6 Chapter 1
General Information
Models and Options
Table 1-5 N4431B Mixed Connector Options
Connector Type Port A Option Port B Option Port C Option Port D Option
3.5 mm -f- 101 201 301 401
3.5 mm -m- 102 202 302 402 Type-N 50 -f- 103 203 303 403 Type-N 50 -m- 104 204 304 404
7-16 -f­7-16 -m-
a
a
105 205 305 405 106 206 306 406
a. Limits ECal module high frequency to 7.5 GHz.
Other Options Description
UK6 Adds commercial calibration certificate with measured data.
Chapter 1 1-7
General Information
Models and Options

N469xA/B Modules (Microwave Two-Port)

NOTE The N4690/1/6A are no longer available for purchase, but may be ret urned to
Agilent for service and recertification.
ECal modules from this series are available wit h the connector types and fr equency ranges shown in Table 1-6. Mixed connector options are not av ailable with this series.
Figure 1-4 Model N4693A Option M0F
1-8 Chapter 1
Table 1-6 N469xA/B Modules
General Information
Models and Options
Model
Number
N4690B Type-N 50 300 kHz to 18 GHz M0F 00M 00F
N4691B 3.5 mm 300 kHz to 26.5 GHz M0F 00M 00F
N4692A 2.92 mm 10 MHz to 40 GHz M0F 00M 00F
N4693A 2.4 mm 10 MHz to 50 GHz M0F 00M 00F
N4694A 1.85 mm 10 MHz to 67 GHz M0F 00M 00F
Connector
Type
Operating
Frequency
Port A and B Option
-m- and -f- -m- and -m- -f- and -f-
Module Part Number
N4690-60004 N4690-60005 N4690-60006
Module Part Number
N4691-60004 N4691-60005 N4691-60006
Module Part Number
N4692-60001 N4692-60002 N4692-60003
Module Part Number
N4693-60001 N4693-60002 N4693-60003
Module Part Number
N4694-60001 N4694-60002 N4694-60003
N4696B 7 mm 300 kHz to 18 GHz No Connector Option
Module Part Number
N4696-60004
Other Options Description
00A Adds one -m- to -m- and one -f- to -f- adapter (not available with N4696B).
UK6 Adds commercial calibration certificate with measured data.
Chapter 1 1-9
General Information
Models and Options

8506x Modules

ECal modules in this series are no longer available for purchas e, but may be returned to Agilent for service and recertification. The 8506x series has been replaced by the N469xA/B series. See “N469xA/B Modules (Microwave Two-Port)” on page 1-8.
Figure 1-5 Model 85062B
Table 1-7 8506x Modules
Model
Number
85060B 7 mm 1 to 18 GHz 85062B 3.5 mm 1 to 26.5 GHz 85064B Type-N 50 1 to 18 GHz
Connector
Type
Operating
Frequency
8506xA series modules with serial numbers below 800 need to be retrofitted for operation with the 85097B VNA interface kit. The retrofitting must be done by a qualified Agilent Technologies service office. See “Contacting Agilent” on page 1-15.
1-10 Chapter 1
General Information

N4432A and N4433A Modules (Microwave Four-Port)

N4432A and N4433A Modules (Microwave Four-Port)
ECal modules from this series are available wit h the connector types and fr equency ranges shown in Table 1-8. Mixed connector options (shown in Table 1-9) allow you to configure the module with a different connector type at each port.
Figure 1-6 Model N4433A Option 010
Table 1-8 N4432A and N4433A Modules
Model Number
N4432A N4432-60003 Type-N 50 -f- 300 kHz to 18 GHz 020
N4433A N4433-60003 3.5 mm -f- 300 kHz to 20 GHz 010
Module Number
N4432-60005 Mixed connector types – see
N4433-60004 Mixed connector types – see
Connector Type Port A,B,C,D
Table 1-9 on page 1-12
Table 1-10 on page 1-12
Operating Frequency
300 kHz to 18 GHz See Table 1-9
300 kHz to 20 GHz See
Option
on page 1-12
Table 1-10 on page 1-12
Chapter 1 1-11
General Information
N4432A and N4433A Modules (Microwave Four-Port)
Table 1-9 N4432A Mixed Connector Options
Connector Type Port A Option Port B Option Port C Option Port D Option
3.5 mm -f-
3.5 mm -m-
a
a
101 201 301 401 102 202 302 402
Type-N 50 -f- 103 203 303 403 Type-N 50 -m- 104 204 304 404
a. Frequency range of ECal module limited by other connectors.
Other Options Description
UK6 Adds commercial calibration certificate with measured data.
1A7 Adds ISO 17025 compliant calibration. A6J Adds ANSI Z540 compliant calibration.
Table 1-10 N4433A Mixed Connector Options
Connector Type Port A Option Port B Option Port C Option Port D Option
3.5 mm -f- 101 201 301 401
3.5 mm -m- 1 02 202 302 402
Other Options Description
UK6 Adds commercial calibration certificate with measured data.
1A7 Adds ISO 17025 compliant calibration. A6J Adds ANSI Z540 compliant calibration.
1-12 Chapter 1
General Information

Compatible Network Analyzers

Compatible Network Analyzers
ECal modules operate with the following network analyzer models, as shown in Table 1-11. Not all ECal module and network analyzer combinations have the same features or capabilities. See your network analyzer user guide or online Help.
Table 1-11 Compatible Network Analyzers
Network
Analyzers
8509xC
N4431B
ECal Module Series
a
N4432/3A N469xA/B
8753/8720-Series Network Analyzers
8719ES/ET 8720ES/ET 8722ES/ET 8753ES/ET
Yes Yes Yes Yes
c
c
c
c
Yes Yes Yes Yes
c
c
c
c
No No No No
PNA Network Analyzers
E8356/7/8A Yes Yes E8801/2/3A Yes Yes N3381/2/3A Yes Yes E8362/3/4A Yes Yes E8362/3/4B Yes Yes E8361A Yes Yes
Yes Yes Yes Yes Yes Yes
N5230A Yes Yes Yes
Yes Yes Yes Yes
Yes Yes Yes Yes Yes Yes Yes
Interface Required
b
8506x
d
d
d
d
e
e
e
e
e
e
e
Yes 85097B Yes 85097B Yes 85097B Yes 85097B
No USB No USB No USB No USB No USB No USB No USB
ENA Network Analyzers
E5070/1A/B Yes Y es
Yes
Yes No USB
E5061/2A Yes Yes Yes Yes No USB
a. Only ports A and B available when operated with 8753/8720-series analyzers. b. 8506x modules are no longer available for purchase. c. 8753/8720-series firmware revision 7.68 or higher required. d. 8753/8720-series firmware revision 7.74 or higher and adapter cable (part no.
8121-1047) required.
e. PNA firmware revision 3.0 or higher required with N4692/3A, and N4694A.
NOTE For 8753 network analyzers, RF ECal modules (8509x) may be used below 300 kHz.
However, calibration accuracy is not specified. The display t race connecting the two points on either side of the 300 kHz boundary may misrepresent actual performance.
Chapter 1 1-13
General Information

Service and Recertification

Service and Recertification
If your ECal module requires service or recertifi cation, contact the Agilent office nearest you for information about where to send i t. The perf ormance of your ECal mo dule can only be verified by specially manufactured equipment and calibration s tand ards from Agilent.
1-14 Chapter 1
General Information
Service and Recertification

Contacting Agilent

Assistance with test and measurements needs and information on finding a local Agilent office are available on the Web at:
http://www.agilent.com/find/assist
If you do not have access to the Internet, please contact your Agilent field engineer.
NOTE In any correspondence or telephone conversation, refer to the Agilent product
by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period.
Chapter 1 1-15
General Information
Service and Recertification

Returning Devices to Agilent

If you are returning the product to Agilent, please include a service tag (located near the end of this manual) on which you provide the following information:
• your company name and address
• a technical contact person within your company, and the person’s complete telephone number including country code and area code
• the model number and serial number of the ECal module
• type of service required
• a detailed description of the problem and how the device was being used when the problem occurred (such as calibration or measurement)
NOTE When transporting the module, use original or comparable packaging.
1-16 Chapter 1
General Information
Service and Recertification

Recertification of ECal Modules

The suggested interval for recertification is 12 months. After reviewing the results of the initial recertification, you may es tablish a shorter interval that reflects greater use and
wear of the module.
Where to Send a Module for Recertification
Contact Agilent for information on where to send your kit for recertification. See
“Contacting Agilent” on page 1-15. Refer to “Returning Devices to Agilent” o n page 1-16 for
instructions on the preparation of returning the device.
How Agilent Verifies Your ECal Module
Agilent verifies the specifications of these devices as follows:
1. The residual microwave error terms of the test sys tem are verified with precision airlines and shorts or low frequency resistance that are directly traced to the National Institute of Standards and Technology (NIST). The airline and short c haract eri stics are developed from mechanical measurements. The mechanical measurements and material properties are carefully modeled to give very accurate electrical representation. The mechanical measurement s are then traced to NIST through various plug and ring gages and other mechanical measurements.
2. Each module is electrically tested on this system to the specifi cation listed in Chapte r 5,
“Specifications and Characteristics.”
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calib ration facility. The specifications d ata provided for the module are t raceable to NIST through Agilent Technologies.
What Is Provided with a Recertification
• a new calibra t i on st i cke r a ff i x e d to th e m od u l e
• a certificate of calibration
• a list of United States National Institut e of Standa rds and Technology (NIST) traceable numbers
• a calibration report for each traceable module listing measured values, specifications, and uncertainties
• a new set of S-parameter data (embedded in module memory) if the old set of S-parameters data no longer allows for a calibratio n that me ets all performance specifications
Agilent Technologies offers different types of calibration for the recertification of the module. For more information, contact Agilent. See “Contacting Agilent” on page 1-15.
Chapter 1 1-17
General Information

Safety and Regulatory Information

Safety and Regulat o ry Inf ormation
Review this section to familiarize yourself with safety markings and instructions before you operate the ECal module. This produc t has been designed and tested in accordance with international standards.
WARNING The WARNING notice denotes a hazard. It calls attention to a
procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.
CAUTION The CAUTION noti ce denotes a hazard. It calls attention to an operating
procedure, practice, or the like, that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CA UTION notice until the indicated conditions are fully understood and met.

Safety Notices

WARNING To prevent electrical shock, discon nect from instrument before
cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
WARNING If this product is not used as specified, the protection provided by
the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.

Installation Notices

CAUTION This product is designed for use in INSTALLATION CATEGORY II and
POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.
CAUTION When installing the product in a cabinet, the convection in and out of the
product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the s ystem by 4 °C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 w atts, then forced convection must be used.
1-18 Chapter 1

Instrument Markings

When you see this symbol on your instrument, you should
!
1SM1-A This text indicates that the instrument is an Industrial
refer to the instrument’s instruction manual for important information.
This symbol indicates hazardous voltages.
The laser radiation symbol is marked on products that have a laser output.
This symbol indicates that the instrument requires alternating current (ac) input.
The CE mark is a registered trademark of the European Community. If it is accompanie d by a year, it indicates th e year the design was proven.
The CSA mark is a registered trademark of the Canadian Standards Association.
Scientific and Medical Group 1 Class A product (CISPER 11, Clause 4).
General Information
Safety and Regulatory Information
This symbol indicates that the power line switch is ON.
This symbol indicates that the power line switch is OFF or in STANDBY position.
The C-Tick mark is a registered trademark of the Australian Spectrum Management Agency.
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical/ electronic product in domestic household waste.
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a "Monitoring and Control instrumentation" product.
Do not dispose in domestic household waste. To return unwanted products, contact your local Agilent
office, or see http://www.agilent.com /envi ronment/p rod uct/ for more information.
Chapter 1 1-19
General Information
Safety and Regulatory Information

Compliance Notices

This product has been designated and tested in accordance with the standards in the Declaration of Conformity (see below), and has bee n supp lied in a safe condition. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe conditions.
Compliance IEC 61000-4-4: 1995 (electrical transients) and IEC 61000-4-2: 1995 (electrostatic discharge): Passes using Criterion C.
Calibration cycle may be interrupted by power line transients applied to external power supply or by electrostatic discharge applied to module. Normal function is restored by power-cycling the module.
Compliance with Canadian EMC Requirements
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB du Canada.
Declaration of Conformity
A declaration of conformity is on file for the PNA models, and a copy is available upon request.
1-20 Chapter 1

2 Preparing to Use the ECal Module

2-1
Preparing to Use the ECal Module

Inspecting ECal Module Kit

Inspecting ECal Module Kit

Inspect the Kit Contents

Verify the case and its contents are not damaged and that all p arts ar e includ ed (s ee it ems listed in T able 2- 2, Table 2-3, and Table 2-4). If the case or any device appears damaged, or if the shipment is incomplete, contact Agilent . Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for settlement from the transportation company. See “Contacting Agilent” on page 1-15.

Record the Serial Number

ECal modules are individually serialized. Record the se rial number and the date of initial use as a reminder for annual recertification. See “Recertification of ECal Modules” on
page 1-17.
Table 2-1 Table for Tracking ECal Module Serial Numbers
Module Number Serial Number Date

Verify Electrical Performance

To verify electrical performance of the module, see “Validating a Calibration” on page 3-7.

Equipment Required but Not Supplied

VNA Interface Kit (85097B)
The VNA Interface Kit allows you to connect an ECal module to supported vector network analyzers. Microw ave ECal modules (N469xA/B series) require an adapter cable (part no. 8121-1047) to operat e wit h the 85097B. The adapter cable is included as an option with the interfa ce kit or can be ordered separately. Refer t o Chapter 6 , “Replaceable
Parts.”
Pin Depth Gages
Gages for measuring recession or protrusion are not provided in the kit. See Chapter 6 ,
“Replaceable Parts.” or contact Agilent to order gages and other tes t accessories.
2-2 Chapter 2