Keysight (Agilent) N2802A Data Sheet

Keysight Technologies
InniiMax III/III+ Probing System
The world’s highest performing probe system keeps getting better
Data Sheet
02 | Keysight | InniiMax III/III+ Probing System – Data Sheet
Unmatched performance
– Full 30 GHz bandwidth to the probe tip – InniiMode probing for making dif ferential, single-ended and common mode
– Industry’s highest delity and accuracy due to bandwidth and extremely low loading – Probe ampliers loaded with measured s-parameters for more accurate response
correction
– Bandwidth upgradable (InniiMax III only) – Variety of probe heads for different use models with maximum usability
Industry-leading probe performance
The InniiMax III probing system of fers you the highest performance available for measuring differential and single-ended signals, with exible connectivity solutions for today’s high-density ICs and circuit boards. Four different InniiMax III probe ampliers ranging from 16 GHz to 30 GHz are available for matching your probing solution to your performance and budget requirements. A proprietary 200 GHz f T InP (indium phosphide) IC process with backside ground vias and novel thick lm technology is utilized to accommodate your highest performance needs and is unmatched by any product in the market.
InniiMax III probe amplier
The N2830A /N7000A Series InniiMax III+ probing system is the next generation of InniiMax probing, greatly expanding the measurement capabilities and usability of a probe capable of measuring all components of a differential signal. The built-in InniiMode technology allows you to switch to differential, single-ended, and common mode without adjusting probe tip connections. The InniiMax III+ probe’s InniiMode provides the following modes of operation.
– A - B (differential) – A - ground (single-ended A) – B - ground (single-ended B) – (A+B)/2 - ground (common mode)
– The InniiMax III+ works with the full array of InniiMax III probe heads and supports the full bandwidth of the InniiMax III+.
InniiMax III+ probe amplier (N2830A Series)
InniiMax III+ probe amplier (N7000A Series)
03 | Keysight | InniiMax III/III+ Probing System – Data Sheet
Correction
Frequency Hz
4E10
12
Highest delity and accuracy
The InniiMax III provides the highest bandwidth and incredibly low loading to allow for a new level of signal delity and accuracy. Continuing the probe head topology pioneered by Keysight Technologies, Inc. in the InniiMax I and II probe systems, ve probe heads are provided to accommodate multiple use models: a 30 GHz browser that is extremely usable, a 25/28 GHz ZIF probe head with economical replaceable/removable ZIF tips, a 28 GHz
2.92 mm probe head which allows cabled measurements using
2.92 mm, 3.5 mm, or SMA coax cables, an economical 16/26 GHz solder-in probe head for demanding measurements, and a 16 GHz QuickTip for a quick and secure connection.
More accurate probe correction
Each individual InniiMax III and III+ probe amp contains its unique S-parameters, and this frequency response data is used with the S-parameters of the various probe heads to further atten the probe’s magnitude and phase response for accuracy. Traditionally, probe correction uses a nominal model based on a typical probe amplier instead of the specic amplier. Generally, the biggest variation between probing systems is a result of the probe amplier. The ability to correct a specic probe amplier’s response results in a more accurate probe correction, which yields a more accurate measurement.
Uncompromised Usability
Extensive line-up of probe heads and accessories
Keysight’s InniiMax III/III+ probes support a wide variety of high-speed applications with an ex tensive line-up of probe heads and accessories.
9
6
3
0
-3
-6
No Correction
-9
-12
-15
-18 1E8
1E9
1E10
– The N5445A browser head (30 GHz) is the best choice for
quick general-purpose troubleshooting of differential signals with its z-axis compliance and variable spacing from 20 mil – 125 mil (or 0.5 mm – 3.1 mm). The span between the signal tips is easily adjusted with a thumb wheel on the browser. Integrated LED lighting at the tip illuminates the probing area for better visibility. Order N5476A for replacement browser tips (set of 4).
04 | Keysight | InniiMax III/III+ Probing System – Data Sheet
– The N2848A /49A QuickTip offers the industry’s rst
magnetically-engaged probe head and tip for a quick and secure connection, pushing the usability to the next level. The N2848A QuickTip probe head quickly snaps to the N2849A probe tip, utilizing magnets to connect to the two sides of the differential signal and ground. Multiple N2849A probe tips can be installed on a DUT, allowing quick and reliable measurement of many probe points. The QuickTip supports the InniiMode probing when used with the InniiMax III+ probe amp and supports differential probing with the InniiMax III probe amp. The N2849A QuickTip tips can also be used with the InniiMax I/II probe amps when used in conjunction with the N2851A QuickTip probe head for InniiMax I/II.
– The N5439A ZIF probe head provides 28 GHz bandwidth in
an economical replaceable tip form factor. For differential measurement, the N5439A ZIF probe head with the N5440A or N5447A ceramic ZIF tip provides the industry’s lowest signal loading. The ZIF tips can be left on the DUT as the probe head is moved from one probing site to the next. Order N5440A (450 Ω ceramic), N5447A (200 Ω ceramic) or the new N2838A (450 Ω PC board) for a set of 5 ZIF tips with plastic sporks to aid in soldering the tips to your DUT. The N2838A PC board ZIF tip increased the robustness of the ZIF tip signicantly while maintaining the bandwidth performance up to 25 GHz when used in conjunction with the N2803A and N5439A. Variable spacing from 5 mil – 80 mil (or 0.127 mm – 2 mm). (The N5447A 200 Ω ZIF tip is not compatible with InniiMax III+ probes.)
– The N5444A 2.92 mm/3.5 mm/SMA probe head (28 GHz)
allows you to connect two 2.92 mm, 3.5 mm or SMA cables to make a differential measurement on a single oscilloscope channel. Order N5448A 2.92 mm head ex cables (10” or 25 cm) to extend the cable length and add convenience.
– The N5441A solder-in probe head is an economical,
semi-permanent connection that provides up to 16 GHz of bandwidth. Variable span of the leads ranges from 5 mil – 80 mil (or 0.127 mm – 2 mm).
Loading...
+ 8 hidden pages