Keysight (Agilent) N2772A Users Guide

User’s Guide
Publication number N2772-92001 March 2000
s1
For Safety information, Warranties, Regulatory information, and publishing information, see the pages at the back of this book.
© Copyright Agilent Technologies 1999-2000
All Rights Rese rved.
Contents
Inspect the Probe 3 Compatibility 4 Probe Parts Supplied 5 Characteristics and Specifications 6 Electrical Characteristics 6 Input Characteristics 7 Output Characteristics 7 Environmental Specifications 7 Safety Specifications 8 Using the N2772A Probe 10 Powering the N2772A 10 External Power Supply 10 Install or Replace the Battery 11 Oscilloscope Input Characteristic 11 Select the Correct Attenuation 12 Measurement Tips 12 Calibration 13 Cleaning the Probe 13
2
N2772A Differential Voltage Probe

Inspect the Probe

Inspect the Probe
Inspect the shipping container for damage.
Keep a damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the probe has been checked mechanically and electrically.
Check the accessories.
Any accessories that were supplied with the probe are listed in “Probe Parts Supplied” this manual.
• If the contents are incomplete or damaged, notify your Agilent Technologies Sales Office.
Inspect the instrument.
• If there is mechanical damage or defect, or if the probe does not operate properly or pass performance tests, notify your Agilent Technologies Sales Office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as your Agilent Technologies Sales Office. Keep the shipping materials for the carrier’s inspection. The Agilent Technologies office will arrange for repair or replacement at Agilent Technologies’ option without waiting for claim settlement.
3
N2772A Differential Voltage Probe
Inspect the Probe
N2772A Differential Voltage Probes
The N2772A differential probe allows you to safely measure floating circuits, as high as 600V dc peak to ground, with the oscilloscope grounded. It is ideal for many applications such as motor speed controls, power supply designs and electronic high-power converters.
With 20 MHz bandwidth, switchable attenuation of 20:1 and 200:1, and a maximum voltage input of 600V, it provides the versatility for a broad range of applications including high-voltage circuits.
Each probe comes with sharp probe tips for use on small components and in tight places and alligator clips for connecting to larger cables.
The probe is powered by a 9V battery or by using a separate power supply, the N2773A.

Compatibility

The N2772A differential probe is compatible with any oscilloscope that has the following features:
• 1 M input impedance
• Can display 30V peak input
• Has BNC input
4
N2772A Differential Voltage Probe
Inspect the Probe

Probe Parts Supplied

The following diagram and table show the parts supplied with the N2772A Probe.
3,
1
Parts fo r N 2772A Probe Item Description
1 Differential voltage probe 2 Probe tips 3 Banana jack alliga tor - black 4 Banana jack alligator - red
2
2, 3
5
N2772A Differential Voltage Probe

Characteristics and Specifications

Characteristics and Specifications
Characteristics and specifications for the N2772A Differential Voltage Probe are shown below.
Electrical Characteristics
Attenuation ratios 20:1 and 200:1 (Selectable via switch on probe.) Bandwidt h (into 1 M , 50 pF) 20 MHz Accuracy 2.5% into 1 M Rise Time 200x: 17.5 ns
High CMRR 80 dB @ 60Hz, 50 dB @ 1 MHz Input im pedence Between inputs: 10 M , 5 pF Output im pedence 50 Noise 200x: <2 mVrms
Offset <=10 mV into 1 M Switch positions OFF, 200x, 20x External power Via power adapter N2773A (optional) Internal power Battery power: Alkaline 9V, IEC6LR61
Power indicators Green LED: ON at normal operation. Blinks at
Auto stan d by After 30 minutes, only when battery operated.
*Battery life measured @ 25 C with Duracell® akaline battery. (Delivered with probe.)
±Ω
20x: 17.5 ns
20x: <3 mVrms
Battery life: 8 hour operation, 400 hour in auto standby*
standby. Red LED: ON when battery needs to be replaced.
°