Agilent E8285A Programming Manual

Agilent Technolo gies
E8285A CDMA Mobile Station Test Set
Condensed Programming Reference Guide
Firmware Version A.05.00 and Above
Agilent Part Number: E8285-90020
Printed in U. S. A.
June 2000
1
© Copyright Agilent Technologies 1999, 2000
Notice Information contained in this document is subject to change without notice.
All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws.
This material may be reproduced by or for the U.S. Government pursuant to the Copyright License under the clause at DFARS 52.227-7013 (APR 1988).
Agilent Technologies Learning Products Department 24001 E. Mission Liberty Lake, WA 99019-9599 U.S.A.
2
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
Sound Pressure Lp < 70 dB(A).
• At Operator Position.
• Normal Operation.
• According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlärminformationsverordnung vom 18 Januar 1991.
• Schalldruckpegel Lp < 70 dB(A).
• Am Arbeitsplatz.
• Normaler Betrieb.
• Nach ISO 7779:1988/EN 27779:1991 (Typprüfung).
3
Safety Considerations
WARNING
GENERAL This product and related documentation must be reviewed for familiarization with
safety markings and instructions before operation. This product has been designed and tested in accordance with IEC Publication
1010, "Safety Requirements for Electronic Measuring Apparatus," and has been supplied in a safe condition. This instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
SAFETY EARTH GROUND A uninterruptible safety earth ground must be provided from the main power
source to the product input wiring terminals, power cord, or supplied power cord set.
CHASSIS GROUND TERMINAL To prevent a potential shock hazard, always connect the rear-panel chassis ground
terminal to earth ground when operating this instrument from a dc power source. SAFETY SYMBOLS Indicates instrument damage can occur if indicated operating limits are exceeded.
!
Indicates hazardous voltages. Indicates earth (ground) terminal
A WARNING note denotes a hazard. It calls attention to a procedure, practice, or the like, which, if not correctly performed or adhered to, could result in personal injury. Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met.
CAUTION
4
A CAUTION note denotes a hazard. It cal ls attention to a n operation p rocedure, practice, or the like, which, if not correctly performed or adhered to, could result in damage to or destructi on of pa rt or al l of the produ ct. Do not p roceed bey ond an CAUTION note until the indicated conditions are fully understood and met.
Safety Considerations for this Instrument
WARNING This product is a Safety Class I instrument (provided with a protective
earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited..
Whenever it is likely that the protection has been impaired, the instrument must be made inoperative and be secured against any unintended operation.
If this instrument is to be energized via an auto transformer (for voltage reduction), make sure the common terminal is connected to the earth terminal of the power source.
If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.
No operator serviceable parts in this product. Refer servicing to qualified personnel. To prevent electrical shock, do not remove covers.
Servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so.
The opening of covers or removal of parts is likel y to expose dangerous voltages. Disconnect the product from all voltage sources while it is being opened.
Adjustments descr ibed in the manual are performed with power supplied to the instrument while protective covers are removed. Energy availa ble at many points may, if contacted, result in perso nal injury.
The power cord is connected to internal capacitors that my remain live for 5 seconds after disconnecting the plug from its power supply.
For Continued protection against fire hazard, replace the line fuse(s) only with 250 V fuse(s) or the same current rating and type (for example, normal blow or time delay). Do not use repaired fuses or short circuited fuseholders.
5
CAUTION: Always use the three-prong ac power cord supplied with th is product. Failur e to ensure
adequate earth grounding by not using this cord may cause product damage. This product is designed for use in Installati on C ategory I I and Pollu tion Deg ree
2 per IEC 1010 and IEC 664 respectively. FOR INDOOR USE ONLY. This product has autoranging line voltage input, be sure the supply voltage is
within the specified range.
To prevent electrical shock, disconnect instrument from mains (line) before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
Ventilation Requirements: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating
temperature of the product by 4° C for every 100 watts dissipated in the cab inet. If the total power dissipated in the cabin et is greater than 800 w atts, then forced convection must be used.
Product Markings CE - the CE mark is a registered trademark of the European Community. A CE
6
mark accompanied by a year indicated the year the design was proven. CSA - the CSA mark is a registered trademark of the Canadian Standards
Association.
Agilent Technologies Warranty Statement for Commercial Products
Agilent Technologies E8285A CDMA Mobile Station Test Set
Duration of Warranty: 1 year
1. Agilent Technologies warrants Agilent Technologies hardware, accessories and
supplies against defects in materials and workmanship for the period specified above. If Agilent Technologies receives notice of such defects during the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective. Replacement products may be either new or like-new.
2 Agilent Technologies warrants that Agilent Technologies software will not fail to exe-
cute its programming instructions, for the period specified above, due to defects in ma­terial and workmanship when properly installed and used. If Agilent Technologies receives notice of such defects during the warranty period, Agi len t Technologies will replace software media which does not execute its programming instructions due to such defects.
3. Agilent Technologies does not warrant that the operation of Agilent Technologies
products will be uninterrup ted or erro r free. If Agil ent Tec hnologi es is unabl e, within a reasonable time, to repair or replace any product to a condition as warranted, customer will be entitled to a refund of the purch a se price upon prompt return of the product.
4 Agilent Technologies product s may contai n remanufac tured parts equiva lent to new in
performance or may have been subject to incidental use.
5. The warranty period begins on the date of delivery or on the date of installation if
installed by Agilent Technologies. If customer schedules or delays Agilent Technologies installat ion more than 30 days afte r delive ry, w arranty begins on th e 31st day from delivery.
6 Warranty does not apply to defects resulting from (a) improper or inadequate mainte-
nance or calibration, (b) software, interfacing, parts or supplies not supplied by Agilent Technologies, (c) unauthorized modification or misuse, (d) operation outside of the published environmenta l specifi cations for the product , or (e) imp roper sit e preparat ion or maintenance.
7 TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES
ARE EXCLUSIVE AND NO OTHER WARRANTYOR CONDITION, WHETHER WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND AGILENT TECHNOL­OGIES SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDI­TIONS OR MERCHANTABILITY, SATISFACTOR Y QUALITY, AND FITNESS FOR A PARTICUL A R PU RPOSE.
8 Agilent Technologies will be liable for damage to tangible property per incident up to
the greater of $300,000 or the actual amount paid for the product that is the subject of the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused by a defective Agilent Technologies product.
7
9. TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS
WARRANTY STATEMENT ARE CUSTOMER’S SOLE AND EXCLUSIVE REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL AGILENT TECHNOLOGIES OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT, TORT, OR OTHERWISE.
FOR CONSUMER TRANSACTIONS IN AU STRALIA AND NEW ZEALAND: THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE RESTRICT OR MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
ASSISTANCE Product maintenance agreements and other customer assistance agreements are
8
available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
DECLARATION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014
Manufacturer’s Name:
Manufacturer’s Address:
declares that the product
Product Name: Model Number: Product Options:
conforms to the following Product specifications:
Safety: IEC 61010-1:1990+A1+A2 / EN 61010-1:1993+A2 EMC: CISPR 11:1990 / EN 55011:1991- Group 1, Class A
IEC 61000-3-3:1995 / EN 61000-3-3:1994 EN 50082-1 : 1992 IEC 801-2:1991 - 4kV CD, 8kV AD IEC 801-3:1984 - 3 V/m
1 kV Power Lines
Supplementary Information:
This product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC and carries the CE-marking accordingly.
Spokane, Washington USA June 16, 1999
European Contact : Your local Agilent Technologies Sales and Service Office or Agilent Technologies G m bH Department ZQ/Standards Europe, Herrenberger Strasse 130, D-71034 Böblinger, Germany (FAX+49-7031-14-3143)
IEC 61000-3-2:1995 / EN 61000-3-2:1995
IEC 801-4:1988 - 0.5 kV Signal Lines,
Agilent Technologies, Inc. 24001 E. Mission Avenue
Liberty Lake, Washington 99019-9599 USA
CDMA Mobile Station Test Set Agilent Technologies E8285A
All
Vince Roland Reliability & Regulatory Engineering Manager
9
Agilent Technologies E8285A Support Contacts
The documentation supplied with your test set is an excellent source of reference, applications, and service information. Please use these manuals if you are experiencing technical problems:
• Applications information is included in the Agilent Technologies E8285A CDMA Mobile Station Test Set Application Guide (Agilent P/N E8285-90019)
• Calibration and repair information are in the Agilent Technologies E8285A CDMA Mobile Station Test Set Assembly Level Repair Manual - this manual (Agilent P/N E8285-90033).
If you have used the manuals and still have application questions, contact your local Agilent Technologies Sales Representative.
Repair assistance is available for the Agilent E8285A CDMA Mobile Test Set from the factory by phone and e-mail.
When calling or writing for repair assistance, please have the following information ready:
• Instrument model number (Agilent E8285A)
• Instrument Serial Number (tag located on the rear panel).
• Installed options - if any (tag located on the rear panel).
• Instrument firmware revision (displayed at the top of the screen when th e Test S e t is powered up, and is also displayed on the CONFIGURE screen).
Support Telephone
Electronic mail (Internet): Spokane_Service@spk.agilent.com
Numbers:
1 800 827 3848 (Spokane Division Service Assistance, U.S. only) 1 509 921 3848 (Spokane Division Service Assistance, International) 1 800 227 8164 (Agilent Technologies Direct Parts Ordering, U.S. only) 1 916 783 0804 (Agilent Technologies Service Parts Identification, U.S. & Intl.)
10
Table 1 Regional Sales and Service Offices
United States of America: Agilent Technologies Test and Measurement Call Center P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 4844
Japan:
Agilent Technologies Japan Ltd. Measurement Assistance Center 9-1 Takakura-Cho, Hachioji-Shi, Tokyo 192-8510, Japan
(tel) (81) 456-56-7832 (fax) (81) 426-56-7840
Asia Pacific:
Agilent Technologies 24/F, Cityplaza One, 111 Kings Road, Taikoo Shing, Hong Kong
(tel) (852) 3197 7777 (fax) (852) 2506 9233
Canada: Agilent Technologies Canada Inc. 5150 Spectrum Way Mississauga, Ontario L4W 5G1
(tel) 1 877 894 4414
Latin America:
Agilent Technologies Latin America Region Headquarters 5200 Blue Lagoon Drive, Suite #950 Miami, Florida 33126 U.S. A .
(tel) (305) 267 4245 (fax) (305) 267 4286
Europe: Agilent Tec hnologies European Marketing Organization P.O. Box 999 1180 AZ Amstelveen The Netherlands
(tel) (3120) 547 9999
Australia/New Zealand:
Agilent Technologies Australia Pty Ltd. 347 Burwood Highway Forest Hill, Victoria 3131
(tel) 1 800 629 485 (Australia) (fax) (61 3) 9272 0749 (tel) 0 800 738 378 (New Zealand) (fax) (64 4) 802 6881
11
Documentation Map
12
All of the following literature, with the exception of the Instrument BASIC User’s Handbook, is shipped with the Agilent Technologies E8285A on a CD-ROM. The Agilent part number of the CD-ROM is E8285-10004.
Unless a delete option is specified, paper versions of the Application Guide and Condensed Programming Reference Guide are also shipped with each Test Set.
If option OBW is ordered, paper versions of the Reference Guide and the User’s Guide will also be included with the Test Set.
Reference Guide (E8285-90016)
This guide describes the functions performed by each front panel key, front and rear panel connector, and display screen and field. GPIB command examples for each dis­play field are included.
User’s Guide (E8285-90018)
This guide provides a t utorial-style overview of opera t ing the Test Set, incl udi ng a s e c­tion designed to help you ge t sta rt ed. S t atu s repo rt i ng, IB ASIC c ontrol l er i nfo rmat io n, and error message descriptions are also included.
Application Guide (E8285-90019)
This guide contains step-by-step procedures and programming examples for calibrating the Test Set, setting up a call, and making measurements on CDMA and AMPS mobile stations. Tips for increasing measurement throughput are also included, as well as a procedure for logging protocol messages.
Condensed Programming Reference Guide (E8285-90020)
This pocket-sized guide contains a complete listing of GPIB commands, along with a cross-reference between front-panel display fields and the corresponding commands.
Assembly Level Repair (E8285-90033)
This guide includes procedures for performing periodic adjustments, verifying performance, troubleshooting, and repairing the Test Set. Block diagrams and a list of replaceable parts are also included.
Instrument Basic User’s Handbook (E2083-90000)
This guide contains a complete listing of IBASIC commands. This guide is not shipped with the Test Set. For ordering information, contact your nearest re­gional sales office.
Specifications (5968-8839E)
This document provides a short description of the Agilent E8285A and lists the operating specifications.
This document also includes the specifications for Agilent Technologies 83217A Option 001, 003, and 004 software.
1. Part numbers listed are Agilent Technologies part numbers unless otherwise stated.
1
Contents
Remote/Local Modes
Control Annunciators ........................................................................................................................ 15
Remote Mode .................................................................................................................................... 15
Local Mode ....................................................................................................................................... 15
Remote or Local Mode ...................................................................................................................... 15
Local To Remote Transitions ............................................................................................................ 16
Remote To Local Transitions ............................................................................................................ 16
Local Lockout ................................................................................................................................... 17
Clear Lockout/Set Local .................................................................................................................... 17
GPIB Command Syntax
Adjacent Channel Power.................................................................................................................... 23
AF Analyzer................................................. ..................... ..................... ..................... ...................... . 25
AF Generator 1........................ ..................... ..................... ..................... ..................... ...................... . 31
AFGenerator2|Encoder............ ..................... ..................... ..................... ..................... ...................... . 33
AF Generator 2 Pre-Modulation Filters ............................................................................................. 41
CALL Process .................................................................................................................................... 43
CDMA................................................................................................................................................ 83
CELL.................................................................................................................................................. 111
Configure............................................................................................................................................ 113
CSPectrum.......................................................................................................................................... 125
Decoder .............................................................................................................................................. 129
Display ............................................................................................................................................... 135
Measure .............................................................................................................................................. 139
Oscilloscope ....................................................................................................................................... 157
Program .............................................................................................................................................. 163
RF Analyzer ....................................................................................................................................... 165
RF Generator...................................................................................................................................... 169
Spectrum Analyzer............................................................................................................................. 173
Save/Recall Registers......................................................................................................................... 177
Special (GPIB Only Commands) ....................................................................................................... 179
Status.................................................................................................................................................. 181
13
Contents
System.................................................................................................................................................189
Tests ....................................................................................................................................................191
Trigger
.............................................................................................................................................................195
Integer Number Setting Syntax...........................................................................................................197
Real Number Setting Syntax...............................................................................................................199
Multiple Real Number Setting Syntax ................................................................................................201
Number Measurement Syntax.............................................................................................................203
Multiple Number Measurement Syntax..............................................................................................205
GPIB Common Commands
IEEE 488.2 Common Commands ......................................................................................................207
Common Command Descriptions ......................................................................................................208
Remote Capabilities
Remote Operating Capabilities ..........................................................................................................219
Remote Interface Functions ................................................................................................................220
GPIB Examples
Alphabetical list of fields with GPIB example ...................................................................................225
Index
14

Remote/Local Modes

Control Annunciators

The letters and symbols at the top right corner of the display indicate these conditions:
•R indicates the Test Set is in remote mode. The Test Set can be put into the remote mode by an external controller or by an IBASIC program running on the built-in IBASIC con­troller.
•L indicates the Test Set has been addressed to Listen.
•T indicates the Test Set has been addressed to Talk.
•S indicates the Test Set has sent the Require Service message by setting the Service Re­quest (SRQ) bus line true.
•C indicates the Test Set is currently the Active Co nt roller on the bus.
•* indicates an IBASIC program+ is running.
•? indicates an IBASIC program is waiting for a user response.
•- indicates an IBASIC program is paused.

Remote Mode

In Remote mode all front panel keys are disabled (except for the LOCAL key, POWER switch, Volume control and Squelch control). The LOCAL key is only disabled by the Local Lockout bus command. When in Remote mode and addressed to Listen the Test Set responds to the Data, Remote, Local, Clear(SDC),
and Trigger messages. When the Test Set is in Remote mode, the “R” annunciator will be displayed in the upper right corner of the display screen and triggering is set to the state it was last set to in Remote mode (if no previous setting the default is FULL SETTling and REPetitive RETRiggering). When the Test Set is being addressed to Listen or Talk the “L” or “T” annunciators will be displayed in the upper right corner of the display screen.

Local Mode

In Local mode the Test Set’s front panel controls are fully operational. The Test Set uses FULL SETTling and REPetitive RETRiggering in Local mode.

Remote or Local Mode

When addressed to Talk in Remote or Local mode , the Test Set can issue the Data and Status Byte messages and responds to the Take Control message. In addition the Test Set can issue the Service Request Message (SRQ). Regardless of whether it is addressed to talk or listen, the Test Set will respond to the Clear(DCL), Local Lockout, Clear Lockout/Set Local, and Abort messages.
15

Local To Remote Transitions

Local To Remote Transitions
The Test Set switches from Local to Remote mode upon receipt of the Remote message (REN bus line true and Test Set is addressed to listen). No instrument settings are changed by the transition from Local to Remote mode, but triggering is set to the state it was last set to in Remote mode (if no previous setting the
default is FULL SETTling and REPetitive RETRiggering). The “R” annunciator in the upper right corner of the display is turned on.
When the Test Set makes a transition from local to remote mode all currently active measurements are flagged as invalid causing any currently available measurement results to become unavailable. If the GPIB trigger mode is RETR:REP then a new measurement cycle is started and measurement results will be available for all active measurements when valid results have been obtained. If the GPIB trigger mode is RETR:SING then a measurement cycle must be started by issuing a trigger event.
Refer to “Triggering Analog Measurements in Remote Mode (GPIB Operation)” or “Triggering CDMA Measurements in Remote Mode (GPIB Operation)” in the Measurement Triggering Process chapter of the Test Set’s User’s Guide.

Remote To Local Transitions

The Test Set switches from Remote to Local operation (full front panel control) upon receipt of the Local message (Go To Local (GTL) bus message and Test Set is addressed to listen) or the Clear Lockout/Set Local message (REN bus line false). No instrument settings are changed by the transition from Remote to Local mode, but triggering is reset to FULL SETTling and REPetitive RETRiggering. The “R” annunciator in the upper right corner of the display is turned off.
If it is not in Local Lockout mode the Test Set switches from Remote to Local mode whenever the front­panel LOCAL key is pressed.
If the Test Set was in Local Lockout mode when the Local message was received, front-panel control is returned, but Local Lockout mode is not cleared. Unless the Test Set receives the Clear Lockout/Set Local message, the Test Set will still be in Local Lockout mode the next time it goes to the Remote mode.
16
16
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\remmodes.fb

Local Lockout

Local Lockout
The Local Lockout mode disables the front-panel LOCAL key and allows return to Local mode only by commands from the System Controller (Clear Lockout/Set Local message).
When a data transmission to the Test Set is interrupted, which can happen if the LOCAL key is pressed, the data being transmitted may be lost. This can leave the Test Set in an unknown state. The Local Lockout mode prevents loss of data or system control due to someone unintentionally pressing front-panel keys.
NOTE: Return to Local mode can also be accomplished by setting the POWER switch to OFF and
back to ON. However, returning to Local mode in this way has the following disadvantages:
1. It defeats the p urpo se of the Local Lockout mode i n tha t the Ac tiv e Controller will loose
control of the Test Set,.
2. Instrument configuration is reset to the power up condition thereby loosing the
instrument configuration set by the Active Controller.

Clear Lockout/Set Local

The Test Set returns to Local mode (full front panel control) when it receives the Clear Lockout/Set Local message. No instrument settings are changed by the transition from Remote mode with Local Lockout to Local mode but triggering is reset to FULL SETTling and REPetitive RETRiggering.
17
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\remmodes.fb
Clear Lockout/Set Local
18
18
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\remmodes.fb

GPIB Command Syntax

Instrument Comma nd Syntax Listings
"Adjacent Channel Power" on page 23. "AF Analyzer" on page 25 "AF Generator 1" on page 31 "AF Generator 2 Pre-Modulation Filters" on page 41 "AFGenerator2|Encoder" on page 33 "CALL Process" on page 43 "CDMA" on page 83 "CELL" on page 111 "CSPectrum" on page 125 "Decoder" on page 129 "Oscilloscope" on page 157 "RF Analyzer" on page 165 "RF Generator" on page 169 "Spectrum Analyzer" on page 173
Instrument Numb e r Se tting Command Syntax Listings
"Integer Number Set ting Syntax" on page 197. "Real Number Setting Syntax" on page 199 "Multiple Real Number Setting Syntax" on page 201
Measurement Command Syntax Listings
"Measure" on page 139. "Trigger" on page 195
Measurement Numbe r Se tting Command Syntax Li stings
"Number Measurement Syntax" on page 203. "Multiple Nu mb e r Measurement Syn tax" on page 205
Instrument Funct i on Command Syntax Listings
"Configure" on page 113. "Display" on page 135 "Program" on page 163 "Save/Recall Registers " on page 177 "Status" on page 181 "System" on page 189 "Tests" on page 191
GPIB Only Command Syntax Listings
"Special (GPIB Only Commands)" on page 179.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
19
Command Listing Conventions
Command listings are used to define the Test Set’s GPIB commands. The listing shows the commands, their hierarchical relationships, related parameters (if any), and associated notes (if any).
The Test Set’s commands are based upon a hierarchical structure, also known as a tree system. In such a system, associated commands are grouped together under a common node in the hierarchy, analogous to the way leaves at a same level are connected at a common branch. This and similar branches are connected to fewer and thicker branches, until they meet at the root of the tree. The closer to the root, the higher a node is considered in the hierarchy.
The command listing is divided into columns, as indicated by light gray vertical lines. The root node is the leftmost column. Lower nodes in the hierarchy are indented one position to the right, below the root node. To obtain a particular command, the full path to it must be specified. For example: to set the low side of the audio input on the Audio Analyzer to float you would generate the following command:
:AFANalyzer: AI N ’F loat’
:AFANalyzer
:AIN ’Gnd’
’Float’ ’600 to Hi’
:AIN? (Returns quoted string)
:ELResistor (See Real Number Setting Syntax, :STATe not included)
Directs the user to a specific Instrument Command, Measurement Command, or Number Setting Command syntax listing. Notes indicate which, if any, Number Setting Commands are not supported by this particular path.
Square brackets([]) are used to enclose a keyword that is optional when programming the command; that is, the Test Set will process the command to have the same effect whether the option keyword is omitted by the programmer or not.
Letter case (uppercase or lowercase) in listings is used to differentiate between the accepted short form (the uppercase characters) and the long form (the whole keyword). The Test Set accepts only the exact short and the exact long forms. Sending a keyword that is not the exact short form or the exact long form will generate an error.
In the parameter section of the listing a number of characters have special significance. Square brackets ([]) are used to enclose one or more parameters that are optional when controlling the Test Set. Braces({}), or curly brackets, are used to enclose one or more parameters that may be included zero or more times. The vertical bar (|) can be read as “or” and is used to separate alternative parameter options.
20 20
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\syntax2.fb
The query form of a command is generated by appending a question mark to the last keyword. However, not all commands have a query form, and some commands exist only in the query form. The listings include, where applicable, the command form and the query form of each command.
CAUTION: When changing a field’s setting, a space must always precede the setting value in the
command string, regardless of the field type. For example:
:RFG:FREQ<space>850MHZ :EFG:ATT<SPACE>’On’
Improper punctuation will result in the following error:
HP-IB Error: -102 Syntax Error.
21
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\syntax2.fb
22 22
S:\agilent\e8285\QK_REF\BOOK\CHAPTERS\syntax2.fb

Adjacent Channel Power

Adjacent Channel Power
:ACPower
:CBW|CBANdwidth (See "Real Number Setting Syntax" on page 199, :STATe not
:COFFset (See "Real Number Setting Syntax" on page 199, :STATe not included)
included)
HIS COMMAND SETS THE BANDWIDTH OF THE CARRIER AND ADJACENT
T
CHANNELS TO BE MEASURED IN ADJACENT CHANNEL POWER MEASUREMENTS.
HE CHANNEL BW FIELD IS FOUND ON THE ADJACENT CHANNE L PO WER
T
SCREEN.
HIS COMMAND SETS THE FREQUENCY DIFFERENCE BETWEEN THE TUNE FREQ OR RF
T
HANNEL FIELD SETTINGS AND THE CENTER OF THE ADJACENT CHANNEL TO BE
C
MEASURED IN ADJACENT CHANNEL POWER MEASUREMENTS. THE CH OFFSET FIELD IS FOUND ON THE ADJACENT CHANNEL POWER SCREEN.
:MEASurement ’Ratio’
:MEASurement? (Returns quoted string)
:RBW|RBANdwidth ’300 Hz’
:RBW?|RBANdwidth? (Returns quoted string)
:RMODulation ’Unmod’
:RMODulation? (Returns quoted string)
’Level’
HESE COMMANDS SET/QUERY THE FORMAT FOR DISPLAYING ADJACENT
T
CHANNEL POWER MEASUREMENTS. THE ACP MEAS FIELD IS FOUND ON THE
ADJACENT CHANNEL POWER SCREEN
’1 kHz’
HESE COMMANDS SET/QUERY THE RESOLUTION BANDWIDTH FOR ADJACENT CHANNEL POWER
T
MEASUREMENTS. THE RES BW FIELD IS FOUND ON THE ADJAC ENT CHANNEL POWER SCREEN.
’Mod’
HESE COMMANDS SET/QUERY THE CARRIER REFERENCE SETTING, INDICATING WHETHER
T
THE CARRIER BEING MEASURED DURING ADJACENT CHANNEL POWER REFERENCE MEASUREMENTS WILL BE UNMODULATED OR MODULATED. THE CARRIER REF FIELD IS FOUND ON THE ADJACENT CHANNEL POWERSCREEN.
23
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\acpower.sec
Adjacent Channel Power
24
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\acpower.sec

AF Analyzer

:AFANalyzer
:AIN ’Gnd’
’Float’ ’600 to Hi’
:AIN? (Returns quoted string)
:CURRent
[:ZERO]
:DEMPhasis ’750 us’
:DEMPhasis? (Returns quoted string)
:GAIN ’0 dB’
:GAIN? (Returns quoted string)
HESE COMMANDS SET/QUERY THE INPUT STATE OF THE AUDIO IN (LO)
T
CONNECTOR. THE AUDIO IN LO FIELD IS FOUND ON THE AF ANALYZER screen.
HIS COMMAND ZEROES THE DC CURRENT MEASUREMENT. THE DC CURRENT
T
FIELD IS FOUND ON THE AF ANALYZER SCREEN.
’Off’
HESE COMMANDS SET/QUERY THE STATE OF DE-EMPHASIS NETWORKS IN THE AUDIO
T
ANALYZER AND SPEAKER CIRCUITRY. THE DE-EMPHASIS FIELD IS FOUND ON THE AF
NALYZER screen.
A
’10 dB’ ’20 dB’ ’30 dB’
HESE COMMANDS SET/QUERY THE AF ANALYZER DE-EMPHASIS AMPLIFIER GAIN. THE
T DE-EMP GAIN FIELD IS FOUND ON THE AF ANALYZER screen.
AF Analyzer
25
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
AF Analyzer
:AFAN
:DETector ’RMS’
:DETector? (Retur ns qu ot ed string)
’RMS*SQRT2’ ’PK+’ ’PK-’ ’PK+-/2’ ’PK+-MAX’ ’PK+ HOLD’ ’PK- HOLD’ ’PK+-/2 Hd’ ’PK+-MX Hd’
HESE COMMANDS SET/QUERY THE TYPE OF DETECTOR USED WHEN MEASURING AND
T
DISPLAYING AF SIGNAL LEVELS. THE DETECTOR FIELD IS FOUND ON THE AF
NALYZER screen.
A
:PKLocation ’Filters’
:PKLocation? (Returns quoted string)
:SETTling ’Fast’
:SETTling? (Returns qu ot ed string)
:ELResistor (See "Real Number Setting Syntax" on page 199, :STATe not
’De-Emp’
HESE COMMANDS SET/QUERY THE SIGNAL SOURCE FOR THE PEAK DETECTOR
T
MEASUREMENTS. THE PK DET TO FIELD IS FOUND ON THE AF ANALYZER
screen.
’Slow’
HESE COMMANDS SET/QUERY THE SETTLING TIME FOR AUDIO MEASUREMENTS.
T
SE “FAST” WHEN MEASURING SIGNALS GREATER THAN 200 HZ). THE
(U
SETTLING FIELD IS FOUND ON THE AF ANALYZERscreen.
included, valid range 1 ohm to 1 megohm)
HIS COMMAND SETS THE EXTERNAL LOAD RESISTANCE FOR MEASUREMENTS USING
T
THE AUDIO IN CONNECTORS. THE EXT LOAD R FIELD IS FOUND ON THE AF
NALYZER screen.
A
26
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
:AFAN
:FILTER1|FILT1 ’<20Hz HPF’
:FILTER1?|FILT1? (Returns quoted string)
’50HZ HPF’ ’300Hz HPF’ ’C MESSAGE’
HESE COMMANDS SET/QUERY THE AUDIO FILTER 1 SELECTION. THE
T FILTER 1 FIELD IS FOUND ON THE AF ANALYZERscreen.
AF Analyzer
:FILTER2|FILT2 ’300Hz LPF’
:FILTER2?|FILT2? (Returns quoted string)
:GTIMe (See "Real Number Setting Syntax" on page 199, :STATe not included, valid
:INPut ’FM Demod’
:INPut? (Returns quoted string)
’3kHz LPF’ ’15kHz LPF’ ’>99kHz LP’ ’6khz BPF’
HESE COMMANDS SET/QUERY THE AUDIO FILTER 2 SELECTION. THE
T FILTER 2 FIELD IS FOUND ON THE AF ANALYZERscreen.
HIS COMMAND SETS THE GATE TIME (HOW LONG THE AF COUNTER SAMPLES THE INPUT
T
SIGNAL) FOR THE AUDIO FREQUENCY COUNTER. THE AF CNT GATE FIELD IS FOUND ON THE AF ANALYZER scre en .
’AM Demod’ ’SSB Demod’ ’Audio In’ ’Ext Mod’ ’FM Mod’ ’AM Mod’ ’Audio Out’
HESE COMMANDS SET/QUERY THE INPUT TO THE AUDIO ANALYZER. THE
T AF ANL IN FIELD IS FOUND ON THE AF ANALYZERscreen.
range 10 mS to 1 S)
27
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
AF Analyzer
:AFAN
:GAIN ’0 dB’
’20 dB’ ’40 dB’
:GAIN? (Returns quoted string)
HESE COMMANDS SET/QUERY THE INPUT GAIN SETTING FOR THE AUDIO ANALYZER. THE
T INPUT GAIN FIELD IS FOUND ON THE AF ANALYZERscreen.
:SMPoint ’De-Emp’
:SMPoint? (Returns quoted string)
:NOTCh :GAIN ’0 dB’
:GAIN? (Returns quoted string)
:FREQuency (See "Real Number Setting Syntax" on page 199, :STATe not
:RANGing 'Auto'
:RANGing? (Returns quoted string)
’Filters’ ’Input’ ’Notch’
HESE COMMANDS SET/QUERY THE SIGNAL SOURCE FOR THE OSCILLOSCOPE. THE
T SCOPE TO FIELD IS FOUND ON THE AF ANALYZERscreen.
’10 dB’ ’20 dB’ ’30 dB’ ’40 dB’
HESE COMMANDS SET/QUERY THE GAIN OF THE AF ANALYZER’S NOTCH FILTER
T
AMPLIFIER (USED FOR MAKING SINAD MEASUREMENTS). THE NOTCH GAIN FIELD IS FOUND ON THE AF ANALYZERscreen.
included, valid range 300 Hz to 10 kHz)
HIS COMMAND SETS THE CENTER FREQUENCY FOR THE VARIABLE FREQUENCY NOTCH
T
FILTER. THE NOTCH FREQ FIELD IS FOUND ON THE AF ANALYZERscreen.
'Hold'
HESE COMMANDS SET/QUERY THE GAIN CONTROL MODE OF THE AF ANALYZER’S
T
NOTCH FILTER AMPLIFIER (USED FOR MAKING SINAD MEASUREMENTS). THE NOTCH
GAIN FIELD IS FOUND ON THE AF ANALYZERscreen.
28
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
:AFAN
:SPEaker :MODE ’On’
:MODE? (Returns quoted string)
:VOLume 'Pot'
:VOLume? (Returns quoted string)
’Off’
HESE COMMANDS SET/QUERY THE GAIN CONTROL MODE OF THE AF ANALYZERS
T
NOTCH FILTER AMPLIFIER (USED FOR MAKING SINAD MEASUREMENTS). THE NOTCH
GAIN FIELD IS FOUND ON THE AF ANALYZERscreen.
'Off'
HESE COMMANDS SET/QUERY THE SPEAKER VOLUME, WHICH IS CONTROLLED BY THE
T VOLUME KNOB WHEN “POTIS SELECTED. THESPEAKER VOL FIELD IS FOUND ON THE AF ANALYZER screen.
AF Analyzer
29
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
AF Analyzer
30
S:\agilent\e8285\QK_REF\BOOK\SECTIONS\afanalyr.sec
Loading...
+ 265 hidden pages