Keysight (Agilent) E7402A Data Sheet

E7402A-05
SERVICE NOTE
SUPERSEDES: NONE
E4401B, E4402B, E4403B, E4404B, E4405B, E4407B, E4408B, and E4411B ESA Spectrum Analyzer family. Also E7401A, E7402A, E7403A, E7404A, and E7405A EMC Pre-Compliance Analyzers.
Intermittent screen "flicker" due to a cold solder joint
ESA Spectrum Analyzer Serial Numbers:
E4401B-07 US39010101 / US40240 610 E4402B-07 US39010101 / US40241 334 E4403B-06 US39010101 / US40240 745 E4404B-06 US39010101 / US40240 716 E4405B-06 US39010101 / US40240 518 E4407B-06 US39010101 / US40241 203 E4408B-06 US39010101 / US40240 473 E4411B-07 US39010101 / US40241 566
E7401A-05 US39 110101 / US40240361 E7402A-05 US39 110101 / US40240208 E7403A-05 US39 110101 / US40240139 E7404A-05 US39 110101 / US40240128 E7405A-05 US39 110101 / US40240198
To Be Performed By: Agil ent Service Ce nte r s O nly
Continued
DATE: January 2 001
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: End of GMS period
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 1.0 Hours
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
RETURN SCRAP SEE TEXT
AUTHOR: BD ENTITY: 5320 ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES PRINT ED IN U.S.A.
1 of 3
Page 2 Service Note E7402A-05
Parts Required:
No specific pa r t numbers, just the too ls to ta ke a p a rt the ESA Fr ont Frame assembly . You w il l also need scis sor s, solder, and a solder ing ir on.
Situation:
Instruments within the above serial range can experience an intermittent screen "flic ke r" due to a cold solder joint on pin 7 of the transf or mer on the backlight inverter board. The backlight inverter board a tta c he s to the f ro nt pa ne l interface board (A1A1) on the a na lyz e r. The atta ch ed figures below in the Sol ution/Action section c a n be refere nc e d on the proc e dur e to use to co rrec t for this pro b le m.
The backlight inverter board and LCD displa y a re proc ur e d toge ther as an assembly from NEC. Manufactu ring ha s tr oubleshot screen flic ke r is sue s to poor solder qualit y on pin 7 of the transformer on the backlight inve rt er boar d. NEC has been notified and an in spe c tion process implemen te d in ma nufacturing to cur e th is pr oblem.
Solution / Action:
If an ESA or EMC analyze r is returne d a nd the fail description r el ates to a di spla y problem (no display or intermittent "flicker"), Agilent Service Centers should remove the backlight inverter board and perfor m the f ollowing procedure:
1) Peel back the plastic shield on the backlight inverter board. See Figure 1.
Figure 1
2) Cut the plasti c wit h a pair of sc issors. See Figure 2.
Continued