Keysight (Agilent) E7402A Service Note

E7402A-02
SERVICE NOTE
SUPERSEDES: None
E7401A EMC Analyzer E7402A EMC Analyzer E7403A EMC Analyzer E7404A EMC Analyzer E7405A EMC Analyzer
Serial Numbers:
E7401A US39110101 / US39150240 E7402A US39110101 / US39150158 E7403A US39110101 / US39430122 E7404A US39110101 / US39150116 E7405A US39110101 / US39430154
Frequency Response (less than 50 kHz) may fail the +/- 0.5 dB specification
Duplicate Service Notes:
E7401A-02 E7402A-02 E7403A-02 E7404A-02 E7405A-02
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION AVAILABLE
ACTION CATEGORY:
AGREEABLE TIME
Continued
DATE: January 2000
PERFORMANCE ENHANCEMENT SERVICE/RELIABILITY ENHANCEMENT
LOCATION CATEGORY:
AUTHOR: JMS
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
ENTITY: 5340
© 2000 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
AVAILABLE UNTIL: January 2001
ADDITIONAL INFORMATION:
1 of 2
Page 2 Service Note E7402A-02
To Be Performed By: Agilent-Qualified Personnel
Situation:
Instruments leaving the factory can fail the Low Frequency Response specification less than 50kHz. Manufacturing discovered a software problem where incorrect low frequency flatness correction values were being loaded into the instrument memory.
Solution / Action:
As of January 7th, 2000, the Field Software for the E7400A-series EMC Analyzers IS NOT AVAILABLE at the CSC’s. If a customer sends in an E7400A-series EMC Analyzer for this specific problem, the analyzer will have to be sent back to the Factory for re-alignment.
When the Field Software becomes available, and an E7400A-series EMC Analyzer is sent in for this problem, the following should be done:
Perform the Frequency Response Test using the automated software, or run the test manually as per the E7400A-series EMC Analyzer Calibration Guide. If the test fails, you will need to load new flatness correction values into the analyzer using the Adjustment Software.