To our 4284A/4279A LCR Users, the E4980A Provides Even Greater Value!
E4980A Precision LCR 4284A Precision LCR 4279A C-V meter
Frequency 20 Hz to 2 MHz 20 Hz to 1 MHz 1 MHz
Test signal level 0 to 2 Vrms/0 to 20 mArms 0 to 2 Vrms/0 to 20 mArms 20 m, 50 m, 100 m,
0 to 20 Vrms/0 to 100 mArms10 to 20 Vrms/0 to 200 mArms2200 m, 500 m, 1 Vrms
Auto level control (ALC) Ye s Ye s N o
DC bias capability Built-in 1.5 V, 2 V Built in 1.5 V, 2 V Built-in ± 38 V
± 40 V
1
± 40 V
2
External bias input
DC source ± 10 V
1
No No
Programmable list sweep 201 points 10 points 51 points
Remote control GPIB, LAN, USB GPIB GPIB
Web browser control Ye s No No
Control commands 4284A compatible 4284A unique 4279A unique
Basic accuracy 0.1 % @ SHORT 0.1 % @ SHORT 0.1 % @ SHORT
0.05 % @ MED/LONG 0.05 % @ MED/LONG
Parameters Cp-D/Q/G/Rp, Cs-D/Q/Rs, Cp-D/Q/G/Rp, Cs-D/Q/Rs, C-D/Q/ESR/G
Lp-D/Q/G/Rp, Ls-D/Q/Rs, R-X, Lp-D/Q/G/Rp, Ls-D/Q/Rs, R-X,
Z-Ød/Ør, G-B, Y-Ød/Ør Z-Ød/Ør, G-B, Y-Ød/Ør
Lp-Rdc1, Ls-Rdc1, Vdc-Idc
1
Measurement time mode Standard model
(SHORT/MED) 330 ms/380 ms @ 20 Hz 1500 ms/1500 ms @ 20 Hz 10 ms @ 1 MHz (List
100 ms/180 ms @100 Hz 270 ms/400 ms @ 100 Hz sweep mode, bias ON)
20 ms/110 ms @ 1 kHz 40 ms/190 ms@ 1 kHz
7.7 ms/92 ms @ 10 kHz 30 ms/180 ms @ 10 kHz
5.7 ms/89 ms@ 100 kHz 30 ms/180 ms @ 100 kHz
5.6 ms/88 ms @ 1 MHz 30 ms/180 ms @ 1 MHz
5.6 ms/88 ms @ 2 MHz
Entry model (E4980A-005)
1040 ms/1150 ms @ 20 Hz
240 ms/380 ms @100 Hz
37 ms/200 ms @ 1 kHz
25 ms/180 ms @ 10 kHz
23 ms/180 ms @ 100 kHz
23 ms/180 ms @ 1 MHz
23 ms/180 ms @ 2 MHz
Storage devices Internal/USB memory Internal/memory card Internal
Cable lengths 0, 1, 2, 4 m 0, 1, 23, 43m 0, 1, 2 m
Cabinet dimensions (mm) 370 (W) x 105 (H) x 390 (D) mm 426 (W) x 177 (H) x 498 (D) mm 426 (W) x 177 (H) x 498 (D) mm
Weight 5.3 kg 15 kg 15 kg
1. Option E4980A-001 required.
2. Option 4284A-001 required.
3. Option 4284A-006 required.
Enhanced test efficiency
Both the Agilent 4284A Precision LCR
Meter and 4279A 1 MHz C-V Meter have
long been recognized as industry standard
equipment for a wide range of impedance
measurement applications.
The E4980A LCR combines fast, accurate
measurements with powerful features to
enhance your measurement efficiency and
make your job easier.
High compatibility to ease migration
Almost all E4980A functions are compatible
with the 4284A and 4279A, enabling users
to migrate to the E4980A with ease. For
detailed migration information, refer to
the technical overview, Migrating from
an Agilent 4284A to an Agilent E4980A
Precision LCR Meter and Migrating from an
Agilent 4279A to an Agilent E4980A
Precision LCR Meter available on our Web
site. www.agilent.com/find/E4980A
Key specifications and function compatibility