Keysight (Agilent) E4433B Service Note

E4433B-02
SERVICE NOTE
SUPERSEDES: NONE
ESG "B" Series RF Signal Generators
EMI suppression filters on the reference boards with part numbers E4400-60009 and E4400-60010 (A11 - Assembly 11), may short power supplies or cause no digital control on the ESG.
SERIAL NUMBERS:
E4400B:-04 US40050929-US40051015 E4420B:-03 US40050877-US40050964 E4421B:-03 US40051418-US40051600 E4422B:-03 US40050701-US40050809 E4430B:-03 US40050341-US40050387 E4431B:-03 US40050515-US40050527 E4432B:-03 US40052793-US40053733 E4433B:-02 US40051623-US40052040
TO BE PER F OR M ED BY:Ag il ent Technolog ie s qua lif ie d pe r sonnel.
PARTS REQUIRED: P/N Description Qty
E4400-6O009 Bd Assy-Reference, OCXO (Option 1E5) 1 or E4400-60010 Bd Assy-Reference, TCXO 1
DATE: October 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: Support Life
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 4.0 Hours
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
Continued
RETURN SCRAP SEE TEXT
AUTHOR: SN ENTITY: 5331 ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES PRINT ED IN U.S.A.
1 of 2
Page 2 Service Note E4433B-02
SITUATION:
EMI suppressio n filte r s (pn 9135-0461) on referenc e bo ards with pa rt num be rs E4400-60009 and E4400-60010 ( A11) , may shor t power supplies or cau se no digital control on the ESGs mentioned above. During of some of the A11 boards, corrosive flux came in contact with the filters. As a result, when the instr ument is in norma l ope r a tion, the filters m ay shor t to gr ound. This process has now be e n ch anged to prevent the c or ro sion problem. If th ere are an y failures, they should oc cu r within two to six months of nor ma l oper a tion. A high tempera ture and low humidity environment will dry out the remaining flux and prevent the filters from being shorted to ground. It is e stimated that approximately 4.0 % ( 12 to 16 units) will fail.
SOLUTION:
Maintenance is required on the specific failure. Service will be provided at Agilent Service Centers (for loc a tions see the ESG Fa mily Signal Gen era tor s U se r ’ s G uide ) . The resolution will be to replace the reference board A11 and do all the necessary performance test in accordance with the ESG Fa mily Signal Gene r at or s Ca libration Guide.