E4432B-02A
SERVICE NOTE
SUPERSEDES: NONE
ESG "B" Series RF Signal Generator
Serial Numbers: See below:
Duplicate Service Notes:
E4400B-03A Serial Numbers: US38440487/US39340699
GB38450200/GB393402 69
E4420B-02A Serial Numbers: US38440464/US39340716
GB38450272/GB393403 13
E4421B-02A Serial Numbers: US38440492/US39340834
GB38450457/GB393408 27
E4422B-02A Serial Numbers: US38440311/US39340523
GB38450127/GB393402 28
E4430B-02A Serial Numbers: US3844021 0/US39340287
GB38450163/GB393401 78
E4431B-02A Serial Numbers: US38440236/US39340393
GB38450249/GB393402 75
E4432B-02A Serial Numbers: US38441373/US39342028
GB38450432/GB393406 94
E4433B-01A Serial Numbers: US38440715/US39341027
GB38450413/GB393405 93
Family Reference Board Module Replacement due to failure of 1 GHz
VCO causing front panel display error 513 "Het band VCO Unlock" and
frequency inaccur acy bel ow 250 MHz.
DATE: February 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: Support Life
AUTHOR: JE ENTITY: 5331 ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR 1.0 assy/6.0 automated adjust Hours
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
Continued
RETURN
SCRAP
SEE TEXT
© 2000 AGILENT TECHNOLOGIES
PRINT ED IN U.S.A.
1 of 2

Page 2 Service Note E4432B-02A
To Be Per formed By: Agilent Technologies- Q ua lif ie d Personnel
Parts Required:
P/N Description Quantity
E4400-60010 Bd Assy Reference, TCXO(NEW) 1
E4400-60009 Bd Assy Reference, OCXO(Option 1E5)(New) 1
Situation:
The ESG E4400B instrume nt se r ies Het ban d VCO unl oc ks due to a non-linearity in the tuning
charactere ristic s of the sur f ace a coustic wave oscill at or use d to ge ne rat e th e hete r odyne band
(below 250 MHz) in the ESG. This VC O is loca ted on the Ref erenc e Boar d Asse m bly Module
(A11). Instrument front panel err or 513 r e sult s from this failure and f requency inaccura c y be low
250 MHz. Inform at ion fr om the VCO vender indica te s an exp ected fa ilur e rate of 8% to 12%
may be experienced over the specific ESG population.
Solution / Action:
The resolution plan is to repair on failure. Replace the referen ce board assembly and recalibrate
the ESG A/D RF Signal Generator in accordance to the Post-Repair Procedures; located in the
ESG Service Guide (P/N: E4400-90155/Print Da te Sept 99) . Shi p failed units to Agilent
Technologies c ustomer service ce nte r as desc r ibe d in the routing guide . Othe r w ise , post- r e pa ir
may be performed using the ESG Servi ce So ftware along with the ESG A/D Calibration Guide
(P/N: E4400-90325/Print Date Nov 99). All defe ctive Re fe renc e Boar d Asse m b lies ha ve be e n
purged from inventory and ALL failed E4400-60139/69139 and E4400-60140/69140
assemblies should be scrappe d.
Changes to effec te d r an ge of ser ial numbers or repair stra te gy will be communicat ed in revi sions
to this servic e note