E4419B-08-S
SERVICE NOTE
SUPERS EDES: None
Agilent E4419B EPM Power Meter
Serial Numbers: GB00000000 / GB99999999
MY00000000 / MY99999999
US00000000 / US99999999
Failure to tak e hee d of the information c onta ined in this safety se r v ice note may le a d to a repa ir e d unit failing to meet the creepage
and creapage requirements for the altitude specificati on of 3,000m.
In this instanc e , the unit will fail to comply with the following
standards; IEC 61010-1 :92, EN 61010:93 and CSA C22.2 No
1010-1:92. Agilent currently claim compliance with these standards in the unit specifications.
Additional Regulatory Information For Power Supply Replacement
Duplicate Service Notes: E4418B-08-S
To Be Per forme d By: Agilent Qua lif ie d Personnel or Customer
DATE: October 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
SAFETY
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY: ALWAYS AGILENT RESPONSIBLE UNTIL : ALWAYS
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
0.25 hours LABOR
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
Continued
RETURN
SCRAP
SEE TEXT
AUTHOR: FC ENTITY: E600 ADDITIONAL INFORMATION:
© 2000 AGILENT TECHNOLOGIES
PRINT ED IN U.S.A.
1 of 2

Page 2 Service Note E4419B-08-S
Situation:
It has been highlighted that the force-fit connector connecting the earth wire to the Power
Supply Unit can be fitted in two different ways .
By fitting the connector with the bottom (flat) side facing the large capacitor on the Power
Supply Unit, the ga p be twe en the m is m ax imised (approxim at ely 2mm ) .
By fitting the connector with the top (rounded) si de facing the large capacitor on the Power
Supply Unit, the ga p be twe en the m is m inim ise d ( a ppr oxi ma te ly 1. 2mm).
The gap should be ma ximised at all time s to e nsur e the unit meets the design and regulatory
requireme nts relating to ele ct ri cal insulation. This r equirement is su c h tha t the ga p should
be at least 1.71mm.
Solutio n / A ction:
During the replacement of the Power Supply Unit, the force-fit connector will be
disconnected.
Ensure that th is c onne c tor is r e -fitted such that th e bo ttom (flat) side of it is facing the larg e
capacitor on the Power Supply Unit.
Figure 1: Correct orientation of force-fit connector.