Keysight (Agilent) E4413A Operation and Service Guide

Errata

This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-Packard's former test and measurement, semiconductor products and chemical analysis businesses are now part of Agilent Technologies. To reduce potential confusion, the only change to product numbers and names has been in the company name prefix: where a product number/name was HP XXXX the current name/number is now Agilent XXXX. For example, model number HP8648 is now model number Agilent 8648.
Ce manuel peut contenir des références à <<HP>> ou <<Hewlett-Packard.>> Veuillez noter que les produits de test et mesure, de semi-conducteur et d'analyse chimique qui avaient fait partie de la société Hewlett-Packard sont maintenent une partie de la société Agilent Technologies. Pour reduire la confusion potentielle, le seul changement aux noms de reference a été dans le préfixe de nom de société : là où un nom de référence était HP XXXX, le nouveau nom de référence est maintenant Agilent XXXX. Par example, le HP 8648 s'appelle maintenent Agilent 8648.
Diese Gebrauchsanweiseung kann Bezug nehmen auf die Namen HP oder Hewlett-Packard. Bitte beachten Sie, dass ehemalige Betriebsbereiche von Hewlett-Packard wie HP-Halbleiterprodukte, HP-chemische Analysen oder HP-Test- und Messwesen nun zu der Firma Agilent Technology gehören. Um Verwirrung zu vermeiden wurde lediglich bei Produktname und - Nummer der vo laufende Firmenname geändert: Produkte mit dem Namen/Nummer HP XXXX lauten nun mehr Agilent XXXX. Z.B, das Modell HP 8648 heißt nun Agilent 8648.
Questo manuale potrebbe contenere riferimenti ad HP o Hewlett-Packard. Si noti che le attività precedentemente gestite da Hewlett-Packard nel campo di Test & Misura, Semiconduttori, ed Analisi Chimica sono ora diventate parte di Agilent Technologies. Al fine di ridurre il rischio di confusione, l'unica modifica effettuata sui numeri di prodotto e sui nomi ha riguardato il prefisso con il nome dell'azienda : dove precedentemente compariva "HP XXXX" compare ora "Agilent XXXX". Ad esempio: il modello HP8648 è ora indicato come Agilent 8648.
Este manual puede hacer referencias a HP o Hewlett Packard. Las organizaciones de Prueba y Medición (Test and Measurement), Semiconductores (Semiconductor Products) y Análisis Químico (Chemical Analysis) que pertenecían a Hewlett Packard, ahora forman parte de Agilent Technologies. Para reducir una potencial confusión, el único cambio en el número de producto y nombre, es el prefijo de la compañía: Si el producto solía ser HP XXXX, ahora pasa a ser Agilent XXXX. Por ejemplo, el modelo HP8648 es ahora Agilent 8648.
䖭Ͼ᠟ݠ䞠䴶ৃ㛑৿᳝ᚴ᱂݀ৌⱘ䌘᭭DŽ䇋⊼ᛣᚴ᱂݀ৌҹࠡⱘ⌟䆩ˈᇐԧѻ કˈ࣪ᄺߚᵤ䚼䮼⦄೼ሲѢᅝ᥋Ӻ݀ৌDŽЎњޣᇥৃ㛑ⱘ䇃㾷ˈѻકোⷕৡᄫ াᬍব᳔ࠡ䴶ⱘ݀ৌৡᄫDŽབᵰϔϾѻકⱘোⷕˋৡᄫҹࠡᰃ+3;;;;ˈ⦄೼ⱘ োⷕˋৡᄫᰃᅝ᥋Ӻ;;;;DŽ՟བ῵ൟোⷕᰃᚴ᱂˔˒ː˔DŽ⦄೼ᰃ῵ൟোⷕᅝ ᥋Ӻ˔˒ː˔DŽ 
A
マニュアル・チェンジ
変更
本文中の「HP(YHP)」、または「(横河)ヒューレット・パッカード株式会社」とい
う語句を、「Agilent」、または「アジレント・テクノロジー株式会社」と変更して
ください。
ヒューレット・パッカード社の電子計測、半導体製品、化学分析ビジネス部門は分
離独立し、アジレント・テクノロジー社となりました。
社名変更に伴うお客様の混乱を避けるため、製品番号の接頭部のみ変更しておりま
す。
(例: 旧製品名 HP 4294A は、現在 Agilent 4294A として販売いたしておりま
す。)
Operation and Service Guide
Agilent E4412A and E4413A
Power Sensors
HP Part No.
Part No. E4412-90013
Printed in USA
June 13, 2007
© Copyright 1999-2007 Agilent Technologies, Inc.
Legal
Legal
The information contained in this document is subject to change without no­tice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantabil­ity and fitness for a particular purpose. Agilent Technologies shall not be lia­ble for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Notice - Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the Hewlett­Packard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.
2
Safety Information
Safety Information
WARNING: Warning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met.
CAUTION: Caution denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, would result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met.
CAUTION: This product is designed for use in Installation Category II and Pollution Degree 2.
This instrument has been designed and tested in accordance with IEC Publi­cation 348, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
“CSA” The CSA mark is a registered trademark of the Canadian Standards Association.
Notice for Germany: Noise Declaration LpA < 70 dB am Arbeitsplatz (oper­ator position) normaler Betrieb (normal position) nach DIN 45635 T. 19 (per ISO 7779).
3
Safety Information
4
1

Operation and Service Guide

5
Operation and Service Guide
Figure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A
and ECP-E26A, Respectively)
6
Operation and Service Guide

General Information

General Information
This operation and service guide contains information about initial inspec­tion, operation, performance tests, troubleshooting and repair of the E4412A and E4413A power sensors.
Warranty
The power sensors are warranted and certified as indicated on the inside cover of this manual.
Instruments Covered by Manual
These instruments have a two-part serial number: the prefix (two letters and the first four numbers), and the suffix (the last four numbers). The two letters identify the country in which the unit was manufactured. The four numbers of the prefix are a code identifying the date of the last major design change incorporated in your Agilent Technologies product. The four-digit suffix is a sequential number and, coupled with the prefix, provides a unique identifica tion for each unit produced. The contents of this manual apply directly to all serial numbers unless otherwise indicated
-
Description
The E4412A and E4413A power sensors are diode power sensors. They are intended for measurement of CW microwave power levels in a wide dynamic range from 70 dBm to +20 dBm (100 pW to 100 mW). The E4412A measures at frequencies from 10 MHz to 18.0 GHz. The E4413A measures at frequencies from 50 MHz to 26.5 GHz. These are high-speed power sensors, and do not incorporate narrow-bandwidth averaging used in average-power sensors. Signals with digital, pulse, or other forms of ampli tude modulation may introduce measurement errors. Multi-tone signals (containing multiple frequency components), or signals with significant har­monic content (> 45 dBc) may introduce measurement errors at high power levels. (Specifications for the power sensors are in Table 1.)
-
7
Operation and Service Guide
General Information
These power sensors measure CW power, displayed on a compatible power meter in logarithmic (dBm or dB) or linear (Watts or %) measurement units. The E4413A is shipped with a 3.5-mm to Type-N adapter, part number 08485-60005.
NOTE: The E4412A and E4413A power sensors are compatible ONLY with the newer
E44XX-Series power meters. They are NOT compatible with the earlier 430-Series, E1416A, or 70100A power meters.
CAUTION: The E4412A and E4413A power sensors are extremely static- sensitive. Do not open
the power sensor unless you and the power sensor are at a static-free workstation.
Specifications
The specifications listed in Table 1 are the performance standards or limits against which the power sensor may be tested. These specifications are valid ONLY after proper calibration of the power meter. Refer to the Calibration
Procedure Using Agilent E-Series Power Sensors in the E4418A or the Agilent E4419A Power Meter User’s Guide.
8
Operation and Service Guide
Table 1 E4412A and E4413A Specifications
Limits Comments
Frequency Range E4412A: 10 MHz to 18 GHz
E4413A: 50 MHz to 26.5 GHz
Power Range −70 dBm to +20 dBm (100 pW to
100 mW)
Impedance 50 ohm nominal
Connector Type E4412A: Type-N (male)
E4413A: 3.5-mm (male)
General Information
Maximum Standing Wave Ratio (SWR) and Reflection Coeffi cient (Rho)b
­ SWR Rho
Return Loss (dB)
E4412A
10 MHz to <30 MHz 1.22 0.099 20.08
30 MHz to <2 GHz 1.15 0.070 23.13
2 GHz to <6 GHz 1.17 0.078 22.12
6 GHz to <11 GHz 1.20 0.091 20.82
11 GHz to <18 GHz 1.27 0.119 18.42
E4413A
50 MHz to <100 MHz 1.21 0.095 20.45
100 MHz to <8 GHz 1.19 0.087 21.23
8 GHz to <18 GHz 1.21 0.095 20.45
18 GHz to 26.5 GHz 1.26 0.115 18.79
Maximum Power 200 mW peak (+23 dBm)
200 mW average (+23 dBm)
AC Coupled Input max: 20 VDC
Zero Set ±50 pW
9
Loading...