This manual may contain references to HP or Hewlett-Packard. Please note that Hewlett-Packard's former test
and measurement, semiconductor products and chemical analysis businesses are now part of Agilent
Technologies. To reduce potential confusion, the only change to product numbers and names has been in the
company name prefix: where a product number/name was HP XXXX the current name/number is now Agilent
XXXX. For example, model number HP8648 is now model number Agilent 8648.
Ce manuel peut contenir des références à <<HP>> ou <<Hewlett-Packard.>> Veuillez noter que les produits de
test et mesure, de semi-conducteur et d'analyse chimique qui avaient fait partie de la société Hewlett-Packard
sont maintenent une partie de la société Agilent Technologies. Pour reduire la confusion potentielle, le seul
changement aux noms de reference a été dans le préfixe de nom de société : là où un nom de référence était HP
XXXX, le nouveau nom de référence est maintenant Agilent XXXX. Par example, le HP 8648 s'appelle
maintenent Agilent 8648.
Diese Gebrauchsanweiseung kann Bezug nehmen auf die Namen HP oder Hewlett-Packard. Bitte beachten Sie,
dass ehemalige Betriebsbereiche von Hewlett-Packard wie HP-Halbleiterprodukte, HP-chemische Analysen
oder HP-Test- und Messwesen nun zu der Firma Agilent Technology gehören. Um Verwirrung zu vermeiden
wurde lediglich bei Produktname und - Nummer der vo laufende Firmenname geändert: Produkte mit dem
Namen/Nummer HP XXXX lauten nun mehr Agilent XXXX. Z.B, das Modell HP 8648 heißt nun Agilent 8648.
Questo manuale potrebbe contenere riferimenti ad HP o Hewlett-Packard. Si noti che le attività precedentemente
gestite da Hewlett-Packard nel campo di Test & Misura, Semiconduttori, ed Analisi Chimica sono ora diventate
parte di Agilent Technologies. Al fine di ridurre il rischio di confusione, l'unica modifica effettuata sui numeri di
prodotto e sui nomi ha riguardato il prefisso con il nome dell'azienda : dove precedentemente compariva "HP
XXXX" compare ora "Agilent XXXX". Ad esempio: il modello HP8648 è ora indicato come Agilent 8648.
Este manual puede hacer referencias a HP o Hewlett Packard. Las organizaciones de Prueba y Medición (Test
and Measurement), Semiconductores (Semiconductor Products) y Análisis Químico (Chemical Analysis) que
pertenecían a Hewlett Packard, ahora forman parte de Agilent Technologies. Para reducir una potencial
confusión, el único cambio en el número de producto y nombre, es el prefijo de la compañía: Si el producto solía
ser HP XXXX, ahora pasa a ser Agilent XXXX. Por ejemplo, el modelo HP8648 es ahora Agilent 8648.
The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this
material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in
connection with the furnishing, performance, or use of this material.
Notice - Hewlett-Packard to Agilent
Technologies Transition
This documentation supports a product that previously shipped under the HewlettPackard company brand name. The brand name has now been changed to Agilent
Technologies. The two products are functionally identical, only our name has
changed. The document still includes references to Hewlett-Packard products, some
of which have been transitioned to Agilent Technologies.
2
Safety Information
Safety Information
WARNING:Warning denotes a hazard. It calls attention to a procedure which, if not
correctly performed or adhered to, could result in injury or loss of life. Do not
proceed beyond a warning note until the indicated conditions are fully
understood and met.
CAUTION:Caution denotes a hazard. It calls attention to a procedure that, if not correctly
performed or adhered to, would result in damage to or destruction of the instrument.
Do not proceed beyond a caution sign until the indicated conditions are fully
understood and met.
CAUTION:This product is designed for use in Installation Category II and Pollution Degree 2.
This instrument has been designed and tested in accordance with IEC Publication 348, Safety Requirements for Electronic Measuring Apparatus, and
has been supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the instrument in a safe condition.
“CSA” The CSA mark is a registered trademark of the Canadian Standards
Association.
Notice for Germany: Noise Declaration LpA < 70 dB am Arbeitsplatz (operator position) normaler Betrieb (normal position) nach DIN 45635 T. 19 (per
ISO 7779).
3
Safety Information
4
1
Operation and Service Guide
5
Operation and Service Guide
Figure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A
and ECP-E26A, Respectively)
6
Operation and Service Guide
General Information
General Information
This operation and service guide contains information about initial inspection, operation, performance tests, troubleshooting and repair of the E4412A
and E4413A power sensors.
Warranty
The power sensors are warranted and certified as indicated on the inside
cover of this manual.
Instruments Covered by Manual
These instruments have a two-part serial number: the prefix (two letters and
the first four numbers), and the suffix (the last four numbers). The two letters
identify the country in which the unit was manufactured. The four numbers
of the prefix are a code identifying the date of the last major design change
incorporated in your Agilent Technologies product. The four-digit suffix is a
sequential number and, coupled with the prefix, provides a unique identifica
tion for each unit produced. The contents of this manual apply directly to all
serial numbers unless otherwise indicated
-
Description
The E4412A and E4413A power sensors are diode power sensors. They are
intended for measurement of CW microwave power levels in a wide
dynamic range from −70 dBm to +20 dBm (100 pW to 100 mW). The
E4412A measures at frequencies from 10 MHz to 18.0 GHz. The E4413A
measures at frequencies from 50 MHz to 26.5 GHz. These are high-speed
power sensors, and do not incorporate narrow-bandwidth averaging used in
average-power sensors. Signals with digital, pulse, or other forms of ampli
tude modulation may introduce measurement errors. Multi-tone signals
(containing multiple frequency components), or signals with significant harmonic content (> −45 dBc) may introduce measurement errors at high power
levels. (Specifications for the power sensors are in Table 1.)
-
7
Operation and Service Guide
General Information
These power sensors measure CW power, displayed on a compatible power
meter in logarithmic (dBm or dB) or linear (Watts or %) measurement units.
The E4413A is shipped with a 3.5-mm to Type-N adapter, part number
08485-60005.
NOTE:The E4412A and E4413A power sensors are compatible ONLY with the newer
E44XX-Series power meters. They are NOT compatible with the earlier 430-Series,
E1416A, or 70100A power meters.
CAUTION:The E4412A and E4413A power sensors are extremely static- sensitive. Do not open
the power sensor unless you and the power sensor are at a static-free workstation.
Specifications
The specifications listed in Table 1 are the performance standards or limits
against which the power sensor may be tested. These specifications are valid
ONLY after proper calibration of the power meter. Refer to the Calibration
Procedure Using Agilent E-Series Power Sensors in the E4418A or the
Agilent E4419A Power Meter User’s Guide.
8
Operation and Service Guide
Table 1 E4412A and E4413A Specifications
Limits Comments
Frequency RangeE4412A: 10 MHz to 18 GHz
E4413A: 50 MHz to 26.5 GHz
Power Range−70 dBm to +20 dBm (100 pW to
100 mW)
Impedance50 ohmnominal
Connector TypeE4412A: Type-N (male)
E4413A: 3.5-mm (male)
General Information
Maximum Standing Wave Ratio
(SWR) and Reflection Coeffi
cient (Rho)b
SWR Rho
Return Loss (dB)
E4412A
10 MHz to <30 MHz1.22 0.09920.08
30 MHz to <2 GHz1.15 0.07023.13
2 GHz to <6 GHz1.17 0.07822.12
6 GHz to <11 GHz1.20 0.09120.82
11 GHz to <18 GHz1.27 0.11918.42
E4413A
50 MHz to <100 MHz1.21 0.09520.45
100 MHz to <8 GHz1.19 0.08721.23
8 GHz to <18 GHz1.21 0.09520.45
18 GHz to 26.5 GHz1.26 0.11518.79
Maximum Power200 mW peak (+23 dBm)
200 mW average (+23 dBm)
AC Coupled Inputmax: 20 VDC
Zero Set±50 pW