The information contained in this document is subject to change
without notice.
Agilent Technologies makesno warranty of any kind with regard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Warranty
This Agilent Technologies instrument product is warranted against
defects in material and workmanship for a period of three years from
date of shipment. During the warranty period, Agilent Technologies
Company will, at its option, either repair or replace products that prove
to be defective.
For warranty service or repair, this product must be returned to a
service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent Technologies and Agilent Technologies
shall pay shipping charges to return the product to Buyer. However,
Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware
designated by Agilent Technologies for use with an instrument will
execute its programming instructions when properly installed on that
instrument. Agilent Technologies does not warrant that the operation
of the instrument, or software, or firmware will be uninterrupted or
error-free.
LIMITATION OF WARRANTY
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse, operation
outside of the environmental specifications for the product, or improper
site preparation or maintenance.
ii
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
EXCLUSIVE REMEDIES
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT
BE LIABLE FOR ANY DIRECT,INDIRECT, SPECIAL, INCIDENTAL,
OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON
CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
iii
Safety Information
The following safety notes are used throughout this manual.
Familiarize yourself with these notes before operating this instrument.
WARNINGWarning denotes a hazard. It calls attention to a procedure
which, if not correctly performed or adhered to, could result in
injury or loss of life. Do not proceed beyond a warning note
until the indicated conditions are fully understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a procedure that, if not
correctly performed or adhered to, could result in damage to or
destruction of the instrument. Do not proceed beyond a caution sign
until the indicated conditions are fully understood and met.
WARNINGThis is a Safety Class 1 Product (provided with a protective
earth ground incorporated in the power cord). The mains plug
shall be inserted only in a socket outlet provided with a
protected earth contact. Any interruption of the protective
conductor inside or outside of the product is likely to make the
product dangerous. Intentional interruption is prohibited.
WARNINGNo operator serviceable parts inside. Refer servicing to
qualified personnel. To prevent electrical shock do not remove
covers.
CAUTIONAlways use the three-prong AC power cord supplied with this product.
Failure to ensure adequate grounding may cause product damage.
This chapter introduces you to basics of cdmaOne technology and the
general functionality of the ESA with the cdmaOne measurement
personality installed. Sources for additional information on digital
communications are also listed.
1-1
Understanding cdmaOne
cdmaOne Standards
cdmaOne Standards
The cdmaOne communication system personality is defined in the
following Electronics Industry Association (EIA), Telecommunications
Industry Association (TIA), American National Standards Institute
(ANSI), Association ofRadio Industries and Businesses (ARIB)(Japan),
and Korean standards documents:
IS-95-A:
TIA/EIA-IS-95-AMobile Station-Base Station Compatibility Standard for
Dual-Mode Wideband Spread Spectrum Cellular System.
May 1995
TIA/EIA-IS-97-ARecommended Minimum Performance Standards for Base
Stations Supporting Dual-Mode WidebandSpread Spectrum
Cellular Mobile Stations. July 1996
TIA/EIA-IS-98-ARecommended Minimum Performance Standards for
Dual-Mode Wideband Spread Spectrum Cellular Mobile
Stations. July 1996
TIA/EIA-95-B Cell and TIA/EIA-95-B PCS:
TIA/EIA-95-B
TIA/EIA-97-B
TIA/EIA-98-B
Mobile Station-Base Station Compatibility Standard for
Dual-Mode Spread Spectrum Systems. (SP-3693-1)
July 17, 1998
Recommended Minimum PerformanceStandardsfor Base Stations
Supporting Dual-Mode Spread Spectrum Cellular Mobile Stations.
August 1998
Recommended Minimum Performance Standards for Dual-Mode
Spread Spectrum Cellular Mobile Stations. August 1998
95-C Cell and 95-C PCS:
TIA/EIA-95-B
Mobile Station-Base Station Compatibility Standard for
Dual-Mode Spread Spectrum Systems. (SP-3693-1)
July 17, 1998
TIA/EIA-97-CRecommended Minimum Performance Standards for Base
Recommended Minimum Performance Standards for Dual-Mode
Spread Spectrum Mobile Stations. (SP-4383) Ballot Version:
March. 19, 1999
Understanding cdmaOne
cdmaOne Standards
ANSI JSTD-008:
ANSI J-STD-008Personal Station-Base Station Compatibility Requirements
for 1.8 to 2.0 GHz Code Division Multiple Access (CDMA)
Personal Communications Systems. August 29, 1995.
ANSI J-STD-018Recommended Minimum Performance Requirements for 1.8
to 2.0 GHz Code Division Multiple Access (CDMA) Personal
Stations. (SP-3385) January 16, 1996
ANSI J-STD-019Recommended Minimum Performance Requirements for
Base Stations Supporting 1.8 to 2.0 GHz Code Division
Multiple Access (CDMA) Personal Stations. (SP-3383)
January 12, 1996
Korea Cell
TTA.KO-06.0003
Air Interface for CDMA Cellular System in the band 800 MHz.
Edition 1.0 July 3, 1997.
Korea PCS
TTA.KO-06.0013Air Interface between Personal Station and Base Station for
1.7 GHz to 1.9 GHz CDMA PCS System. Edition 1.0
July 30, 1997.
ARIB-T53 (Japan Cell)
ARIB STD-T53 Ver. 1.2CDMA Cellular System Part I. Mobile Station-Base Station
Compatibility Standard. January 1998.
ARIB STD-T53 Ver. 1.2CDMA Cellular System Part II. Minimum Performance
Standard for Base Stations. January 1998.
ARIB STD-T53 Ver. 1.2CDMA Cellular System Part III. Minimum Performance
Standard for Mobile Stations. January 1998.
Chapter 11-3
Tuning Plan Channel and Band Parameters
The cdmaOne standards use various tuning plans. The channel and frequency parameters of the various tuning plans are
shown in the following table. To look at the entire frequency range for a standard and tuning plan, block can be set to “Full”.
For example, Block = Full for J-STD-008 would tune the analyzer from Block A 1930 MHZ (lower edge) to Block C 1990 MHz
(upper edge).
What Does the cdmaOne Measurement Personality and Hardware Do?
What Does the cdmaOne Measurement
Personality and Hardware Do?
The Agilent ESA-E Series Spectrum Analyzer with cdmaOne
measurement personality and hardware can help determine if a
cdmaOne transmitter is working correctly. When configured for
cdmaOne, you can use the instrument to test a cdmaOne transmitter,
according to a variety of industry standards. These documents define
complex, multi-part measurements used to maintain an
interference-free environment. For example, the documents include
methods for measuring the power in a channel. Refer to “cdmaOne
Standards” on page 1-2, for more information on the standards
supported by the measurement personality.
When you select a standard format, the ESA-E Series Spectrum
Analyzer is configured for one-button measurements based on the
measurement methods and limits defined in the standards. The
instrument displays detailed results of the measurements allowing you
to analyze cdmaOne base station system transmitter performance.
The standards specify the test method and settings. However, you can
customize measurements for analyses by changing limits and
measurement parameters.
For cellular and PCS infrastructure tests, the instrument will test
base-station transmitters in a non-interfering manner with of a coupler
or power splitter.
This ESA with the cdmaOne personality includes these measurements:
❏ Channel Power
❏ Code Domain Power
❏ Modulation Accuracy (Rho)
❏ Monitor Channel/Band
❏ Occupied Bandwidth
❏ Out of Band Spurious
❏ Receive Channel Power
❏ Receiver Spurious
❏ Spur Close (In Band Spurious)
❏ Spurs at Harmonics
Chapter 11-5
Understanding cdmaOne
What Does the cdmaOne Measurement Personality and Hardware Do?
In addition to the measurements listed above, the cdmaOne mode
provides or uses the following supplemental functions:
❏ Selection between format specifications
❏ External attenuation and gain compensation
❏ Automatic frequency/channel determination based on user-entered
channel number or channel frequency
❏ Automatic signal level detection and analyzer setup
❏ External reference configuration and control
❏ Save and recall mode state (Mode is the operation mode of the
instrument. For example SA = Spectrum Analyzer or
cdmaOne = cdmaOne measurement personality.)
❏ Storing/printing of results internally or directly to a floppy disk in
spreadsheet (.csv) format
1-6Chapter1
Understanding cdmaOne
Other Sources of Measurement Information
Other Sources of Measurement Information
Additional measurement application information is available through
your local Agilent Technologies sales and service office. The following
application notes provide more detail on digital communications and
measurements.
• Application Note 1298
Digital Modulation in Communications Systems - An Introduction
HP/Agilent part number 5965-7160E
• Application Note 1311
Understanding CDMA Measurements for Base Stations and Their
Components
HP/Agilent part number 5968-0953E
• Application Note 1313
Testing and Troubleshooting Digital RF Communications
Transmitter Designs
HP/Agilent part number 5968-3578E
• Application Note 1314
Testing and Troubleshooting Digital RF Communications Receiver
Designs
HP/Agilent part number 5968-3579E
• Application Note 150
Spectrum Analyzer Basics
HP/Agilent part number 5952-0292
Chapter 11-7
Understanding cdmaOne
Other Sources of Measurement Information
1-8Chapter1
2Getting Started
This chapter introduces you to basic features of the instrument,
including the front panel keys, rear panel connections, and display
annotation. You will also find equipment required for cdmaOne
functions, available documentation, and processes for installing and
uninstalling applications.
2-1
Getting Started
Instrument Overview
Instrument Overview
This section provides information on only cdmaOne mode features. For
those features not described here, refer to the Agilent ESA SpectrumAnalyzers User’s Guide.
Front-Panel Features
For additional information on those features that are described here,
refer to Chapter 5 , “Front-Panel Key Reference.”
Figure 2-1Front-Panel Feature Overview
1Mode keysThese keys allow you to select the measurement mode and mode
parameters such as input and trigger settings.
• MODE accesses menu keys to select the instrument mode. Each mode is
independent of all other modes.
• Mode Setup accesses menu keys that allow you to configure the parameters
specific to the current mode and affect all measurements within that mode.
2-2Chapter2
Rear-Panel Features
This section provides information on only cdmaOne mode features. For
those features not described here, refer to the Agilent ESA SpectrumAnalyzers User’s Guide.
Figure 2-2Rear-Panel Feature Overview
Getting Started
Instrument Overview
1DSP and Fast
ADC
DSP and Fast ADC (Option B7D) provides digital signal processing
and fast ADC required for many of the digital demodulation
measurements in the GSM and cdmaOne measurement
personalities.It must beordered with Option B7E and Option 1D5.
2RF Comms
Hardware
RF Communications Hardware (Option B7E) provides the RF down
convertor hardware required for digital demodulation
measurements. It must be ordered with Option B7D and
Option 1D5
3Ext Ref InAccepts an external 1 MHz to 30 MHz reference frequency source.
410 MHz REF INAccepts an external frequency source to provide the 10 MHz, −15 to +10
dBm frequency reference used by the analyzer.
Chapter 22-3
.
Getting Started
Instrument Overview
510 MHz OutProvides a 10 MHz, 0 dBm minimum, timebase reference signal phase
locked to the Ext Ref In.
610 MHz REF
OUT
7Ext Frame
Sync
Provides a 10 MHz, 0 dBm minimum, timebase reference signal.
Accepts an external 0 to 5 V TTL Even Second Clock (ESEC) to provide
the synchronization signal for the frame timer.
Display Annotation
For the screen annotations not described here, refer to the Agilent ESA
Spectrum Analyzers User’s Guide.
Below is an example of the annotation that may appear on an analyzer
display. The display annotation is referenced by numbers which are
listed below.
For examples of particular measurement displays, refer to the results
section for that measurement in the cdmaOne measurement guide.
2-4Chapter2
Getting Started
Instrument Overview
1
Active Function Area
2
Measurement Bar
3Current
Measurement
4
Base/Mobile
5
CH Frequency
6
MaxP
7
Standard
8
Averaging Indicator
9
Ext Att/Ext Gain
Displays parameter entries. If you press a key which activates a
function, the parameter description and value will appear in the
active function area.
Displays information about measurements including some mode
setup parameters.
Displays the name of the current measurement.
Indicates which digital communications system platform you are
testing, base station systems or a mobile unit.
Indicates the frequency of the channel you have selected for
measurements.
Indicates the maximum RF input power you have indicated will
be applied to the analyzer.
Indicates which digital communications standard measurements
are being tested to.
Indicates the number of the current average (averagingprogress)
when averaging is turned on.
Indicates the user input value for external attenuation or
preamplifier gain at the RF input.
10
Pass/Fail Status
11
Trig
12
PN Offset
13Sync ESECIndicates you have set the Time Reference Sync to ESEC (even
Indicates the statusof the measurementlimit pass/fail test when
one or more limit test function is turned on.
Indicates external triggering mode you have selected for your
measurement: External, Free, RF Burst, or Frame Timer.
Indicates PN offset value you have set for demod measurements.
second clock) for demod measurements. If ESEC is displayed in
red, the even second clock signal has not been connected to the
Ext Frame Sync rear panel connector on the DSP and Fast ADC
module. If ESEC is displayed in black, the even second clock
signal is correctly connected.
Chapter 22-5
Getting Started
Options Required
Options Required
One of the following Agilent ESA-E Series Spectrum Analyzers with an
options combination is required to make specific measurements.
Table 2-1 lists the spectrum analyzer models that are compatible and
the upper frequency range of each. You must have one of these
instruments to use the cdmaOne option.
Table 2-1Compatible Agilent ESA-E Series Spectrum Analyzers
The cdmaOne measurement personality will have partial functionality
in the ESA without the installation of all additional hardware options
listed. In addition, some measurements require certain options and
other measurements are enhanced by certain options.
Table 2-2 lists the measurements provided by the BAC option and the
options required or recommended for the measurements. For optimum
performance of cdmaOne measurements, Option B74 should be
installed in your Agilent ESA-E Series Spectrum Analyzer.
Not all of the options can be installed by the user. Some of the options
require that the instrument be returned to the factory or an Agilent
Technologies service center. In addition, some of the options require
Performance Verification and Adjustments to be performed after
installation. Refer to Table 2-2 for option specific information.
NOTEWhen transporting the instrument, use the original packaging or
comparable packaging. If theshipping container is damaged orany part
is missing, notify Agilent Technologies (refer to the ESA-E Series
Personality
Occupied Bandwidth
Receive Channel Power
Memory ExtensionB72
Monitor Band/Channel
Spur Close
Out of Band Spurious
Rx Spur
Channel Power
Spurs at Harmonics
RF and Digital Communication
Hardware Option bundle
Occupied Bandwidth
Modulation Quality (Rho)
Code Domain
Receive Channel Power
Monitor Band/Channel
Spur Close
Out of Band Spurious
Rx Spur
Includes the following options:
1D6
B72
1D5
B7D
B7E
1DS
1DR
Modulation Quality (Rho)
DSP and Fast ADC
a
Code Domain
RF Communications Hardware
High Stability Frequency
Reference
b
OptionRemarks
BACRequired options for
any cdmaOne
measurements.
B74This is the preferred
a
option to install with
Option BAC. It
containsallhardware
necessary to perform
all measurements
andbundle packaging
makes it easy to
order.
B7DThese options enable
digital demodulation
a
B7E
measurements and
must be ordered
1D5
together.
Receive Channel Power
Preamplifier
b
Monitor Band/Channel
Spur Close
Out of Band Spurious
10 Hz Narrow Resolution
Bandwidth Option
b
Rx Spur
Distance to Fault
50 Ohm Tracking Generator
b
Distance to FaultHDF
a. Factory or Service Center installation, calibration required.
b. Factory installation only
Chapter 22-7
.
1DSThese options
enhance the
1DR
sensitivity for these
measurements and
can be ordered
together with
Option 1D5 as the
Performance Option
Bundle B75.
1DNThese options allow
Distance to Fault
measurement.
Option 1DNgivesyou
theabilityto measure
SWR and Flatness.
Getting Started
Options Required
When you select the cdmaOne mode, the spectrum analyzer performs a
hardware check. The available hardware is compared to the cdmaOne
measurement personality hardware requirements. For example, if
Option B7D and Option B7E hardware is not present, the
Mod Accuracy (Rho) and Code Domain Power keys will be grayed out on
the
MEASURE menu and will not be available for use.
2-8Chapter2
cdmaOne Documentation for the ESA-E Series Spectrum Analyzers
cdmaOne Documentation for the ESA-E Series
Spectrum Analyzers
Spectrum Analyzers with cdmaOne Installed
When you purchase your instrument with the cdmaOne measurement
personality, you receive the following materials:
Table 2-3Personality Documentation
Part NumberPart DescriptionNotes
Getting Started
Refer to the ESA
WEB site for the
current part
number.
User’s Guide
ESA-E Series Spectrum Analyzers
cdmaOne Measurement Personality
Quick Reference Card
ESA-E Series Spectrum Analyzers
cdmaOne Measurement Personality
Measurement Guide
ESA-E Series Spectrum Analyzers
cdmaOne Measurement Personality
Programming Commands
ESA-E Series Spectrum Analyzers
cdmaOne Measurement Personality
Agilent ESA Spectrum Analyzers
Specifications Guide
Agilent ESA Spectrum Analyzers
Documentation and Instrument
Driver CD-ROM
BAC Option manuals
BAC Option manuals
BAC Option manuals
BAC Option manuals
Includes specifications for
all optional measurement
personalities
Does not include service
documentation or software
Chapter 22-9
Getting Started
cdmaOne Documentation for the ESA-E Series Spectrum Analyzers
Spectrum Analyzers without cdmaOne Installed
If your instrument is ordered without measurement personalties
installed, you can order the cdmaOne measurement personality
Option BAC and Option B74
purchase of a personality option upgrade, you receive the documents
listed in Table 2-3 and the documents shown below in Table 2-4.
Table 2-4Personality Documentation
Part NumberPart Description Notes
1
instrument upgrade package. With the
Refer to the ESA
WEB site for the
current part
number
Understanding Digital Communications
Measurements
Additional measurement application information is available through
your local Agilent Technologies sales and service office. Refer to “Other
Sources of Measurement Information” on page 1-7 for application notes.
ESA Spectrum Analyzers Update
For the latest information about this instrument, including software
upgrades, application information, and product information, please
visit the URL listed below.
Updating the Firmware and Software
Updated versions of the AgilentESA Spectrum Analyzers firmware and
software will be available via several sources. Information on the latest
firmware and software revision can be accessed through the following
URL.
RF and Digital Communications
Hardware Installation Note
B74 Option manual
URL to Obtain Update Information
http://www.agilent.com/find/esa/
1. Refer to Table 2-2 for option details.
2-10Chapter2
Getting Started
Installing Optional Measurement Personalities
Installing Optional Measurement
Personalities
Active License Key
The measurement personality software you have purchased with your
instrument has been installed and the license key has been enabled at
the factory. With any future purchase of a new personality software,
you will receive a certificate that displays the unique license key
number. The license key enables you to install, or reinstall, any
measurement personality you have purchased.
You will want to keep a copy of your license key number in a secure
location. Please enter your license key numbers in the box provided
below for future reference. If you should lose your license key number,
get in touch with your local Agilent Technologies service or sales office
for assistance. For the location of these offices, refer to Table 6-1 on
page 6-5.
Active License Key Numbers for Instrument with Serial # ________
For Option______________ the license number is ________________________
For Option______________ the license number is ________________________
For Option______________ the license number is ________________________
You will need to use a license key number only under the following
conditions:
• If you purchase an additional measurement software package.
• If the controller board is repaired or replaced.
Installing the Licensing Key
If you are installing a new option, follow these steps to install the
unique license key number for the measurement personality software
that you want to install in your instrument:
1. Press
When you press
instructions on using the alpha editor, refer to the ESA SpectrumAnalyzers User’s Guide.
System, More, Licensing, Option.
Option the alpha editor will be activated. For
2. Use the alpha editor to enter the three letter designation for the
software option that you wish to install in the instrument.
3. Press
Chapter 22-11
Done on the alpha editor menu.
Getting Started
Installing Optional Measurement Personalities
4. Press License Key.
When you press
Licensing Key the alpha editor will be activated. For
instructions on using the alpha editor, refer to the ESA Spectrum
Analyzers User’s Guide.
5. Use the alpha editor to enter the 12 character licensing key number
for the software option that you wish to install in the instrument.
6. Press
7. Press
Done on the alpha editor menu.
Activate to turn on the licensing key. You may now install the
measurement personality option software.
Using Install Key
You may want to install a software revision, install new measurement
software, reinstall measurement software that you have previously
uninstalled, or uninstall measurement software. Before you can install
software, you will need a set of installation diskettes.
If you have ordered a measurement personality, you will receive the
installation disk set in the option upgrade package. If you are updating
an existing, previously installed measurement option, you may order
the diskettes from Agilent Technologies or create a set from the Agilent
internet site shown in “ESA Spectrum Analyzers Update” on page 2-10.
When you order the updated software disk set, you will need to order
Option UE2 with an cdmaOne revision update. (Option UE2 is a
firmware update and is needed to ensure that the firmware and the
software are compatible.) To order a set of diskettes, get in touch with
your local Agilent Technologies service or sales office.Forthe locationof
these offices, refer to Table 6-1 on page 6-5. To create a disk set refer to
“Creating Software Installation Disks” below.
Creating Software Installation Disks
To create the installation disks on-line, visit the Agilent internet site
shown under “ESA Spectrum Analyzers Update” on page 2-10. Follow
the instructions provided on the internet site for downloading the
current measurement personalty software and creating the installation
disk set. The instructions for creating the disk set will step you through
the process to create a firmware disk set when you create the
measurement personalty software disk set. (A firmware update may be
needed to ensure that the firmware and the software are compatible.)
After you have created the disk set, follow the on-line instructions to
install the firmware. After successfully installing the firmware update,
proceed with the following instructions for installing the measurement
personalty software in your instrument.
2-12Chapter2
Getting Started
Installing Optional Measurement Personalities
Installing Personality/Software Options
This procedure gives steps to install a new software option in an ESA-E
Series Spectrum Analyzer using the internal floppy drive of the
instrument. Screen messages display the update progress and give
directions. The instrument will not need to be re-calibrated after this
procedure since no changes are made to calibration or adjustment files.
If you have a problem with the installation process, refer to
“Troubleshooting the Installer” on page 2-14.
NOTEWhen the installer starts up, it examines the instrument to ensure that
all the required software and hardware options are present. If they are
not, the installer will generate and error and you will not be able to
install the personality.
1. If this is the installation of new personality option software, you
must enter the License Key for the new option. For instructions on
entering the License Key, refer to the “Installing the Licensing Key”
on page 2-11.
When you have completed entering the license key number, continue
with the next step.
2. Insert disk one of the installation disk set into the disk drive located
on the right side of the ESA front panel.
3. Press
System, More, Personalities, and Install. The instrument will
then load the installer off of the floppy drive. If there is no floppy in
the drive, the incorrect diskis inserted,or thereis no installer on the
disk, the error “No install disk present in disk drive” will be shown.
Once the instrument has loaded the installer, the screen will change
to the installer screen and the
Install Pers. menu will be shown. For
more information on the installer screen and menu, refer to
“Installer Screen and Menu” on page 2-15.
4. When the installer first starts up, it will show a popup message.
Select
NOTEOnce the installer has begun installing a personality, any error will
Verify Disks.
cause the whole personality (including a previously installed version) to
be removed from the instrument. Because of this, it is very important
that you verify the disks prior to installing them. If any of the disks or
files are bad, you will not be able to use the personality until you obtain
a new installation disk set and run the install using them.
5. When prompted, insert the next disk and press
When Verify Disks is running, the
Install Now and Exit Install keys
Verify Disks again.
will be greyed out.
Chapter 22-13
Getting Started
Installing Optional Measurement Personalities
6. When the verification is complete, press Install Now and the
installation of the personality will begin. Some of the disks may take
only a short time to load or be skipped entirely, while others can take
up to about 30 minutes to load.
When installer is running, the
Verify Disks and Exit Install keys will
be greyed out.
7. When prompted, insert the next disk and press Install Now again.
8. Once the installation is complete, press
Exit Install.
Troubleshooting the Installer
If the installation process stalls or fails in another way, follow these
steps to resolve your problem.
1. If the instrument stops the update process before all the disks are
loaded proceed as follows:
a. Press
Exit Install to abort the process.
b. Return to step 2 on the previous page and start the installation
process again.
2. If the instrument fails after repeating the installation procedure, get
in touch with your nearest Agilent sales and service office listed in
Table 6-1 on page 6-5 for assistance. Please provide the following
information:
Model Number:
Serial Number:
State that you are having trouble installing a software option
update.
2-14Chapter2
Loading...
+ 112 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.