E4401B-03
SERVICE NOTE
SUPERSEDES: NONE
E4401B Spectrum Analyzer
Serial Numbers: See Below
Frequency Response (less than 50kHz) MAY fail the +/- 0.5dB
table 1.1 specificat i on.
Duplicate Service Notes:
E4401B-03 S/N US39050101 / US39440309
E4402B-02 S/N US39010101 / US39440782
E4403B-02 S/N US39010101 / US39440355
E4404B-02 S/N US39050101 / US39440353
E4405B-02 S/N US39050101 / US39440265
E4407B-02 S/N US39010101 / US39440687
E4408B-02 S/N US39050101 / US39440252
E4411B-02 S/N US39050101 / US39440759
To Be Performed By:
Agilent-Qualified Perso nn el who ha ve the abi lity to run th e ESA PC base d Per form anc e Tes t
& Adjustment software. Agilent Technologies Service Centers have this software and the
necessary test equipment available at their facility. Customers may have purchased the
N2717A software product, which will allow them the perform the necessary testing if the
failure occurs .
DATE: February 2000
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: February 2002
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUS/MER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR 0.5 Hours
SERVICE
INVEN/RY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
Continued
RETURN
SCRAP
SEE TEXT
AUTHOR: BD ENTITY: 5330 ADDITIONAL INFORMATION: If the customer has the
N2717A, they can perform the test.
© 2000 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
1 of 2

Page 2 Service Note E4401B-03
Parts Required to verify failure:
Part Number Description
E4401-90132 Calibration Guide
(*) 8663A Synthesized signal generator
(+) 83630/40/50B
(+) 83620/30/40/50B Synthesized Sweeper
33120A or 3325B Function Generator
E4419A Power Meter
8482A Power Sensor, 2 required
8485A Microwave Power Sensor
(-) 11667A Power Splitter
($)11667B Microwave Power Splitter
3458A Digital Multimeter
1250-1749 Adapter, APC 3.5(f) to APC 3.5(f)
1250-1475 Adapter, Type N(m) to Type N(m)
1251-2277 Dual banana plug to BNC(f)
1250-0781 BNC Tee (BNC f,m,f)
10503A Cable, BNC, (48-inch) 2 required
11500D Cable, Type-N (60 inch)
8120-4921 Cable, APC 3.5mm
11593A 50-Ohm Termination
(*) E4401B and E4411B ONLY
(+) All models except E440 1B an d E44 11 B
(-) E4401B/02B/03B/11B ONLY
($) E4404B/05B/07B/08B ONLY
Situation:
Instruments le av in g th e fa ct or y ca n fa il the low Fre qu ency Response spe cifi ca tion less than
50kHz. Manufacturing discovered a software problem where incorrect low frequency flatness
correction values were being loaded into the instrument memory.
Solution / Action:
Perform the Frequency Response test using the automated software (N2717A) or run the test
manually as per the ESA Calibration Guide. If the test fails, you will need to load new flatness
correction values into the analyzer by using the N2717A Performance Test Adjustment software
or sending the instrument into an Agilent Technologies customer service center. The instrument
will be adjusted at no charge.