Keysight (Agilent) E3634A Service Note

E3634A-02
SERVICE NOTE
SUPERSEDES: NONE
E3634A DC Power Supply
Serial Numbers: See below.
Duplicate Service Notes: E3633A-02: KR82600051/KR92800741
E3634A-02: KR84900051/KR92900529
Over Current Protection (OCP) is not functional
To Be Performed By:
Parts Required: P/N Description Quantity
E3633-60034 Programmed ROM 1 E3634-60034 Programmed ROM 1
Situation:
1. If the value of resistive load is less than 0.4 ohm when the OCP crcuit is tripped, a pulse current occurs in the output terminal. The pulse width is 10 msec, period is 30 sec and amplitude is the value of current limit setting.
Continued
DATE: November 1999
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
ACTION CATEGORY:
LOCATION CATEGORY:
AVAILABILITY: PRODUCT’S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: November 2002
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
STANDARDS: LABOR 0.5 Hours
SERVICE RETURN INVENTORY: SCRAP
SEE TEXT
USED RETURN PARTS: SCRAP
SEE TEXT
AUTHOR: NKP ENTITY: MR00 ADDITIONAL INFORMATION:
© 1999 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 2
Page 2 Service Note E3634A-02
2. If the value of resistive load is greater than 0.4 ohm when the OCP crcuit is tripped, the out­put voltage is 1 V and the output current is set by the Ohm’s law.
3. If clearing the OCP condition when setting the OCP trip level to 0.2 A, OCP function is not operational.
4. When setting the OCP trip level to 0.2 A, OCP function is not operational intermittently.
Solution / Action:
1. Replace the programmed ROM with REV. 1.4. The part numbers are E3633-60034 for the E3633A power supply and E3634-60034 for the E3634A power supply
2. Note that you have to recalibrate the power supply after changing the programmed ROM.