E3631A/32A/33A/34A Programmable dc Power Supply Specifications
Model Number E3631A E3632A E3633A E3634A
123
dc Output 0 to +6 V, 0 to +25 V, 0 to -25 V, 0 to 15 V / 7A or 0 to 8 V / 20 A or 0 to 25 V / 7 A or
Rating (0°C to 40°C) 0 to 5 A 0 to 1 A 0 to 1 A 0 to 30 V / 4 A 0 to 20 V / 10 A 0 to 50 V / 4 A
Load Regulation
1
<0.01% + 2 mV
±(% of output + offset) <0.01% + 250 uA
Line Regulation <0.01% + 2 mV
±(% of output + offset) <0.01% + 250 uA
Ripple and Noise (20 Hz to 20 MHz)
Normal Mode Voltage <350 uVrms / 2 mVpp <350 uVrms / 3 mVpp <500 uVrms / 3 mVpp
Normal Mode Current <2 mArms <500 uArms <2 mArms
Common Mode Current <1.5 uArms
Accuracy 12 Mos (25°C + 5°C), ± (% output + offset)
Programming
Voltage 0.1% + 5 mV 0.05% + 20 mV 0.05% + 10 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.2% + 10 mA
Readback
Voltage 0.1% + 5 mV 0.05% + 10 mV 0.05% + 5 mV
Current 0.2% + 10 mA 0.15% + 4 mA 0.15% + 5 mA
Resolution
Program 0.5 mV / 0.5 mA 1.5 mV / 0.1 mA 1 mV / 0.5 mA 1 mV /1 mA 3 mV / 0.5 mA
Readback 0.5 mV / 0.5mA 1.5 mV / 0.1mA 0.5 mV / 0.1mA 0.5 mV / 1mA 1.5 mV / 0.5 mA
Meter 1 mV / 1mA 10 mV / 1mA 1 mV / 1mA 1 mV / 1 mA (<10 A), 10 m A (≥10 A)
Transient Response Less than 50 usec for output to recover to within 15 mV following a change in output current from full load to half load or vice versa.
Command Processing Time
2
<100 msec
OVP/OCP
Accuracy, N/A 0.5% + 0.5 V / 0.5% + 0.5 A
±(% output + offset)
Activation time N/A 1.5 msec, OVP ≥3 V / <10 msec, OVP <3 V and OCP
Temperature Coefficient per °C ± (% output + offset)
Voltage 0.01% + 2 mV 0.01% + 3 mV
Current 0.02% + 3 mA 0.02% + 0.5 mA 0.02% + 3 mA
Stability, constant output & temperature ± (% of output + offset), 8 hrs
Voltage 0.03% + 1 mV 0.02% + 2 mV 0.02% + 1 mV
Current 0.1% + 3 mA 0.05% + 1 mA 0.1% + 1 mA
Remote Sense Max voltage in each load lead 1 V 0.7 V
Voltage Programming Speed, to within 1% of total excursion
Up Full Load 11 msec 50 msec 50 msec 95 msec 80 msec
No Load 10 msec 45 msec 20 msec 45 msec 100 msec
Down Full Load 13 msec 20 msec 45 msec 30 msec 30 msec
No Load 200 msec 400 msec 400 msec 450 msec 450 msec
AC Input (47 Hz - 63 Hz) 100 Vac ±10% (Opt 0E9) / 115 Vac ±10% (Std) / 230 Vac ±10% (Opt 0E3)
Dimension/Net Weight 213 mm W x 133 mm H x 348 mm D 213 mm W x 133 mm H x 348 mm D
(8.4 x 5.2 x 13.7 in) / 8.2 kg (18 lbs) (8.4 x 5.2 x 13.7 in) / 9.5 kg (21 lbs)
Warranty 1 year
Product Regulation Certified to CSA 22.2 No. 231 (for E3631A), No. 1010.1 (for E3632A /33A/34A); conforms to IEC 1010-1; carries CE mark;
complies with CISPR-11, Group 1, Class A
Note: 1. With sense terminal connected for E3632A/33A/34A.
2. Maximum time for output to change after receipt of commands.
2