E3616A-02
SERVICE NOTE
SUPERSEDES: NONE
E3616A Lab Bench DC Power Supply
Serial Numbers: KR51300867 / KR61301249
To eliminate possible open circuit in the pattern trace of the
printed circuit.
Duplicate Service Notes: E3617A-02
Parts Required: 20 AWG (40 mm length)
To be performed by: Agilent-Qualified Personnel
Situation:
n:
No output in the output terminals of the power supply due to the open circuit connection
caused by improper pad design between C8 and PCB pattern trace or C7 and PCB pattern
trace.
Soution / Action:
1. Connect a jumper wire between (+) terminal of C8 and (+) terminal of C7.
2. Connect a jumper wire between (-) terminal of C8 and (-) terminal of C7.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION RECOMMENDED
DATE: January 1997
ACTION
CATEGORY:
LOCATION
CATEGORY:
AVAILABILITY:PRODUCT'S SUPPORT LIFE AGILENT RESPONSIBLE UNTIL: January 1999
AUTHOR: NKP ENTITY: Y300 ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
STANDARDS:
LABOR 1.0 Hours
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
© 1997 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
RETURN
SCRAP
SEE TEXT
1 of 1