Agilent Technologies certifies that this product met its pub lished specificat ions at the time of shipment from the factory. Agilent
Technologie s further certifies tha t its calibr ation measure ments are trac eable to the Unit ed States Nati onal Ins titute of Stan dards and
Technology (formerly Nati onal Bureau of S tandards), to the ex tent allowed by that organizat ion’s calibrat ion facility , and to the calibration
facilities of other International Standards Organization members.
Warranty
This Agilent Technologies product is warranted against defects in materials and workmanship for a period of one (1) year from date of
shipment. Duration and conditions of warranty for this product may be superseded when the product is integrated into (becomes a part
of) other Agilent products. During the warranty period, Agilent Technologies will, at its option , e ith er repair or r ep la ce p ro d ucts which
prove to be def ec tive.
For warrant y servic e or repair, t hi s product m ust be returned to a service facility designated by Agilent Technologies. Buyer shall prepay
shipping charges to Agilent and Agil ent shall pay shippi ng charge s to retur n t he product to Buy er . H ow ever, Buye r shall pay a ll shipping
charges, duties, and ta xes for product s re t urned to Agil ent fr om another cou n t ry.
Agilent warrants that its software and firmware designated by Agilent for use with a product will execute its programming instructions
when properly installed on that product. Agilent does not warrant that the operation of the product, or software, or firmware will be
uninterrupt ed or error free .
Limitation Of Warranty
The foregoi ng warranty shall n ot apply to defects res ul ting from im pr oper or inadequate maintenance by Buye r, B uyer-suppli ed products
or interfacing, unauthori zed modific ation or mis use, operat ion outs ide of the envi ronmenta l specific ations for the prod uct, or improper site
preparation or mainte na nce.
The design and implementation of any circuit on this product is the sole responsibility of the Buyer. Agilent does not warrant the Bu yer ’s
circuitry or malfunctions of Agilent products that result from the Buyer’s circuitry. In addition, Agilent does not warrant any damage th at
occurs as a r es ul t of th e B uyer’s circu it or any defects tha t res ul t fro m Bu yer-supplie d products.
NO OTHER WARRANTY IS EXPRESSED OR I MPLIED. Agilen t SPECIFICALLY DISCLAIM S THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Exclusive Remedies
THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE RE MEDIES. Agilent SHALL NOT BE LIABLE
FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Notice
The information contained in this document is subject to change without notice. Agilent Technologies MAKES NO WARRANTY OF
ANY KIND WITH REGARD TO THIS MATERIAL, INCL UDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Agilent shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing, performance or use of this material. This document contains
proprietar y information whi ch is protected by copy right. All rights are reserved. No part of this docume nt may be phot ocopied, rep roduced,
or translat ed t o anot her l anguag e wi thout the prior w ritten consent of Agilent Technologies, Inc. Agilent assumes no responsibility for the
use or reliability of its software on equipment that is not furnished by Agilent.
U.S. Government Restricted Rights
The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as "commercial
computer software" as defined in DFARS 252.227- 7013 (Oct 1988), DFARS 252.211-7015 (May 1991) or DFARS 252.227-7014 (Jun
1995), as a "com mercial ite m" as defined in FAR 2.101(a), or as "Restri cted comput er software" as defi ned in FAR 52.227-19 (Jun 1987)(or
any equival ent agen cy reg ulati on or contra ct cla use ), which ever i s appli cabl e. You hav e only t hose r ight s provide d for suc h Software and
Documen ta t io n by t h e a pplicable FA R or DF A R S cl ause or the Agi l ent standard sof tware agree m ent for the prod uct involved.
The Printing History shown below lists all Editions and Updates of this manual and the printing date(s). The first printing of the manual
is Edition 1. The Editi on number increments b y 1 when eve r th e ma nua l is revised. Updates, wh ich ar e is sued between Editions, contain
replacement pages to correct the current Edition of the manual. Updates are numbered sequentially starting with Update 1. When a new
Edition is crea ted, it cont ains all the Upda te i nfor mat ion for the p revious Ed ition. E ach ne w E dition or U pdat e a lso incl udes a re vised copy
of this prin ti ng history page. M an y pr oduct updat es or rev is i ons do not requ ir e manual changes and, conver sely, manu al correc tions may
be done without accompanying product changes. Therefore, do not expect a one-to-one correspondence between product updates and
manual updates.
Edition 1 (P ar t Num ber E1418 -90 001). . . . . . . . . . . . . . . . . . . . . . October 1995
Edition 2 (P ar t Num ber E1418 -90 002). . . . . . . . . . . . . . . . . . . . . . October 1996
Edition 2 Rev 2 (Pa rt N um ber E1418-9000 2) . . . . . . . . . . . . . . . . . . . July 2006
Safety Symbols
Instructi on manual sy mbol affixed to product.
Indic ates that the user m u st refe r to the manual for specific WARNING or CAUTION
informatio n to avoi d personal injury or damage to the pr oduct.
Alternating current (AC).
Direct current (DC).
Indicates t he field wir ing termina l that must
be connec ted to earth gr ound before op erating
the equipment—protects against electrical
shock in case of fault.
or
Frame or chassis ground terminal—typically
connects to the equipment’s metal frame.
WARNING
CAUTION
Indicate s ha zardous vol ta ges.
Calls a tt e ntion to a procedure, practi ce, or c ondition that could cause bodi l y i nj ury or death.
Calls attention to a proce dur e, practice, or con dition that could possibly cause damage to
equipme nt or pe rm anent loss of dat a.
WARNINGS
The following general safety precautions must be observed during all phases of operation, service, and repair of this product.
Failure to comply with these precau tions or w ith specific warnings elsewher e in this manual viol ates safe ty standard s of design,
manufacture, and intended use of the product. Agilent Technologies assumes no liability for the customer’s failure to comply with
these re qu i remen ts .
Ground the equipment: For Safety Class 1 equipment (equipment having a protective earth terminal), an uninterruptible safety earth
ground must be provided from t he m ains power s our ce to the product i nput wiring ter m i nals or suppli ed power cable .
DO NOT operate the p r od uct in an ex pl osive atmo sp here or in the pres ence of flammable ga ses or fume s.
For continued protection against fire, replace the line fuse(s) only with fuse(s) of the same voltage and current rating and type.
DO NOT use re pai r ed fuses or sho rt-c ircuited fu se holders.
Keep away fr om live circuit s: Ope ra t in g personnel mus t not remove equi pm ent covers or s hi el ds . Pr oc edures involvi ng the remova l of
covers or shields are for use by service-trained personnel only. Under certain conditions, dangerous voltages may exist even with the
equipmen t switch ed off. To avoid da ngerou s elec trica l shoc k, DO NOT perform proced ures invo lvin g cover or shield re mova l unless you
are qualified to do so.
DO NOT operate damaged equipment: Whenever it is possib le that the safety prote ction features built into this prod uct have been
impaired, either through physical damage, excessive moisture, or any other reason, REMOVE POWER and do not use the product until
safe oper at i on can be verified by service-trained personnel . If nece ss ar y, return th e product to an Agilent Technologies Sales and Ser vice
Office for service and repair to ensure that safety features are maintained.
DO NOT serv ice or adjust al one: Do not at tempt inte rnal ser vice or adjus tment unles s anot her per son, ca pabl e of renderi ng first aid and
resuscitatio n, is present.
DO NOT substitute parts or modify equipm ent: Because of the danger of introducing additional hazards, do not install substitute parts
or perform any unauthorized modification to the product. Return the product to an Agilent Technologies Sales and Service Office for
service and repair to ensure that safety features are maintained.
According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Manufacturer’s Name:
Manufacturer’s Address:
Agilent Technologies, Incorporated
815 – 14th St. SW
Loveland, Colorado 80537
USA
Declares, that the product
Product Name:
Model Number:
16 Channel 16 Bit D/A Converter
E1418A, E1523A, E1524A, E1525A
Product Options: This declaration covers all options of the above product(s).
Conforms with the following European Directives:
The product herewith complies with the requirements of the Low Voltage Directive 73/23/EEC and the EMC Directive 89/336/EEC
(including 93/68/EEC) and carries the CE Marking accordingly.
Conforms with the following product standards:
EMC Standard
IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998
CISPR 11:1990 / EN 55011:1991
IEC 61000-4-2:1995+A1:1998 / EN 61000-4-2:1995
IEC 61000-4-3:1995 / EN 61000-4-3:1995
IEC 61000-4-4:1995 / EN 61000-4-4:1995
IEC 61000-4-5:1995 / EN 61000-4-5:1995
IEC 61000-4-6:1996 / EN 61000-4-6:1996
IEC 61000-4-11:1994 / EN 61000-4-11:1994
12 Agilent E1 41 8A 8/16 D/A Conv er te r Module User ’ s M an ua l
Using This Chapter
This chapter provi de s ge ne ral m odule information and tasks you must
perform to install a nd pr ep are your module. A procedu re to ve rif y your
installatio n is a lso gi ve n. The ch a pte r is divided into the fol lowing sections:
The Agilent E1418A is an 8 or 16 cha nne l digital-to-ana log converter
module for us e in a VXI bus C-size mainf r am e . The module is a
register-base d de vic e . The module can be progra mm e d via dir e ct regis te r
access or, with th e appr opriate driver, by high level comm ands. This
manual describes programmi ng the module using SCPI ( Sta ndard
Commands for Programmable Instruments) and the SC PI dr iver.
Each Agilent E1418A module is a uniqu e instrument havin g its own output
buffer and error qu eu e. Multiple modules cannot be combin ed into a single
instrument.
Each channel can be co nf igur e d to e ither voltage or cur r en t output mode.
When configu red f or volt ag e ou tput, voltages in the range of -16.0 to
+16.0 Volts can be set. When configured for current output, current in the
range of -0.02 to + 0.0 2 Amps can be set. The channe l output mode can be
programmatic a lly se t, or , can be fo rced to e ither voltage or cur r en t by
mechanica l jum p e r s on the te r minal module.
Chapter 1Module Setup and Instal lation 13
Each output ch annel is individ u a lly c onfigurable to be eit he r an isolated
output or a non-isolated output. Cha nne l c onfiguration to isolate d or
non-isolated is made by individual plug-on modules for each channel.
Figure 1-1 shows the module.
Figure 1-1. The E1418A 8/16-Channel D/A Converter
with Terminal Module
14 Module Setup and Installati onCha pter 1
Orderin g OptionsThe Agilent E1418A may be order e d from Agilent Technologies in a
variety of configura tions. As ordered, the module has the following options:
Description
E1418A8-Channel D/A Converter with Non-Isolat ed O ut pu ts
E1418A Option 00116-Channel D/A Converter with Non-Isolated Outputs
E1418A Option 0028-Channel D/ A C onve rter with I sol ated O ut puts
E1418A Option 00316-Channel D/A Converter with Isolated Outputs
Field Ki tsThe mod ule can a lso be us er c onf igured as described be ginning on page 167
of this manual. The following field expansion and configuration kits are
available:
DescriptionUse
E1523ASingle Channel Isola te d
plug-on module
E1524AExpansion kit, 8-C han nel
Non-Isolated Output s
E1525AExpansion kit, 8-C han nel
Isolated Outputs
To change a single channel from
non-isolated to isolat ed ou tp ut .
To add 8 additional non-isolated
channels to an existi ng 8- cha nnel
module.
To add 8 additional isol at ed cha nnel s t o
an existing 8-channel m odule.
Terminal ModulesThe standard Agilent E1418A Terminal Module provides screw terminals
for connect ions. Two other ter minal optio ns a re a va ila ble with the
Agilent E1418A:
– Crimp and Inser t (Option A3E)
– Ribbon Cable (Option A3H)
– Terminal Modul e Hou sing without a termina l module PC board or
connectors inc lude d (O pti on 135)
Chapter 1Module Setup and Instal lation 15
Functional Description
The Agilent E1418A prov ide s gr e at flexibility in config ur at ion. This
functional de sc ri ption will help you be c ome f am iliar with the v arious
configurations and how the module is struc tur ed . A functi ona l block
diagram is shown in Figure 1-2.
Figure 1-2 . A Funct ional Block Di a gr a m
Each channel MUST have either an isol ated or non- is olat ed pl ug- on m odu le ins talled. Channels may be co nf ig ure d
*
using any combinatio n of Isola te d and Non-Isolated plu g-o n mo dules.
16 Module Setup and Installati onCha pter 1
As shown, there are four major assemblies that make up an Agilent E1418A:
– Channels 1 through 8 main board and VXIbus backplane circuitry.
– Channels 9 through 16 expansion board.
– Isolated or Non-isolated plug- on modules.
– Terminal Module.
Channels 1–8 main board and VXIbus b ac kplane cir cuitr y.
This assembly contain s all the VXIbus interf ace an d decodin g circui t ry .
Isolated and Non -Iso la te d plug-on module s fo r chan nels 1–8 connect to this
assembly. The panel connectors for the terminal module are als o mounted
on this assembly.
The optional Channels 9–16 expansion board is electrically connected and
mechanically mounted to the main board. Isolated and Non-Isolated
plug-on module s f or ch an ne ls 9–16 connect to thi s as se mbly.
Isolated or Non-isolat ed plug -on module s a re re quir e d fo r ever y c ha nne l.
Eight plug-on modules mount to the main board a nd 8 plug-on module s
mount to the exp an sion board. Any com bina ti on of isolated or non-isol at ed
plug-on module s may be used, but ev ery c ha nne l must have a plug-on
module insta lle d.
The Terminal Module shown in Figure 1-2 is the standa r d sc rew-type
terminal module. This module provides screw connections for each channel
output, the cal ibra tion connections, and the extern a l trigger in connecti ons.
This module also contains two jumpers for each channel; one, called the P/J
Jumper, to enabl e or disable VXIbus progra mm ing of the c ha nne l mode
(either voltage or cu rre nt) a nd one , called the V/I Jumpe r , to se t the ch an ne l
mode when the P/J Jumper is in the Jumper posi tion.
Chapter 1Module Setup and Instal lation 17
Front Panel Connectors
Figure 1-3 shows th e connections at the f ro nt pa ne l c onne c tors of the ma in
module.
Figure 1-3. E1418A Front Pan e l Connect or Pinout
18 Module Setup and Installati onCha pter 1
Setting the Logical Address Switch
Figure 1-4 shows the logical address switch location and setting. The logical
address switch factory setting is 72. Valid addresses range f rom 1 to 255.
When used with an Agilent Command Module the factory set address of
decimal 72 results in a module address of 9 (72 divided by 8). The module
address is used as a se co nda ry ad dr es s whe n using GPIB and a Command
Module. For exam ple , the module addres s ma y be 7 09 09, whe re 7 is the
GPIB interface address, 09 is the command module address, and 09 is the
Agilent E1418A address .
The Agilent E1418A su pports dynamic addre ss c onfiguration. Whe n you
set the logical add ress to 255, the res ourc e m an ager sets the logic a l ad dr es s
programmatically.
NoteWhen using an Agilent C-size mainframe with the Agilent E1405/E1406
Command Modules, the logical address setting must be a multiple of 8.
Figure 1-4. Setting the Logical Address
Chapter 1Module Setup and Instal lation 19
Module Installation
To install the module:
1. Verify the position of the CAL S tor e Enab le Jump er f or your
application. The Ju mpe r , as sh ippe d, is in the CAL position. In this
position, a cal ibr a tion may be performe d a nd the results store d in
non-volatil e m emor y. In the Secured pos ition, a calibration may be
performed, but may NOT be stored in non-volatile memor y. This
jumper is descri be d in more detail on page 157.
2. Verify or set the logical address switch a s shown in Figure 1-4 .
3. Install the module in a mainframe as shown in Figure 1-5.
Installat io n in a
Mainframe
The Agilent E1418A m ay be in sta lle d in any slot (exc ep t slot 0 ) in a C-si ze
VXIbus mainf ram e. To install in a mainfr am e :
1. Set the extr a ction levers out. Slide the mo dule into any slot (except
slot 0) until the backplan e connectors touch.
2. Seat the module by moving the levers toward each other.
3. Tighten the top a nd bottom screws to secure the mod u le in the
mainframe.
Figure 1-5. Installing the E1418A in a VXIbus Mainframe
20 Module Setup and Installati onCha pter 1
To Remove a Module:To remove a module from a mainframe:
1. Remove any te rm ina l m odules.
2. Loosen the top and bottom screws secur ing th e m odule in the
mainframe.
3. Move the extraction levers away from each other. As the levers are
moved, the module will detach from the ba ck pla ne co nne c tor s.
4. Slide the module out.
NoteThe extracti on levers will not seat and unseat the backplane connec tors on
older Agilent VXIbus ma inf r a mes a nd non-Agilent m ai nframes. You must
manually seat the connectors by pushing the module into the mainframe
until the front pa ne l is f lush with the front of the mainframe . The e xtr a ct ion
levers may be used to guide or remove th e m odul e.
Chapter 1Module Setup and Instal lation 21
Terminal Modules
Four terminal mo dule s a re ava ila ble for the Agilent E1418A; a screw-type,
a crimp-and-inse r t type, a ribbon cable type , an d an em pty te r mi na l housing
(without a PC board or c onne c tors) .
Screw-Type
Terminal Module
(Standard)
The standard te r minal module use s sc rew - type terminals to make
connections. Figu re 1-6 ma y be photocopied and used for wiring layout.
Figure 1-6. Screw-Type Terminal Module Layout
22 Module Setup and Installati onCha pter 1
Wiring the Terminal Module
Figure 1-7 shows how to open a nd wir e the E1418A te rminal module.
Figure 1-7. Wiring the E1418A Terminal Module
(
Chapter 1Module Setup and Instal lation 23
continued on next page)
Figure 1-7. Wiring the E1418A Terminal Module
(
continued from pr evious page)
24 Module Setup and Installati onCha pter 1
Attaching the Terminal Module
Figure 1-8 shows how to attach the E1418A terminal module.
Figure 1-8. Attaching the E1418A Terminal Module
Chapter 1Module Setup and Instal lation 25
Removing the Terminal Module
Figure 1-9 shows how to remove the E1418A terminal module.
Figure 1-9. Removing the E1418A Terminal Module
26 Module Setup and Installati onCha pter 1
Terminal Module Options
Besides the standa r d sc rew-type terminal mo dule , the Agile nt E1418A can
be ordered with the following two options. One opt ion (A 3E) pr ovides
crimp-and-in se rt con ne ct or s an d te rmina l housing. The other option (A3H)
provides a ribbon cab le con ne cto r and term ina l housing.
Option A3EAgilent E1418A Option A3E can be ordered if a crim p- an d- inse rt te r minal
module is desi red. This allows you to crim p co nne c tors onto wires which
are then inserted directly into the Agilent E1418A faceplate connector.
Refer to the pin-out diagr am on pa ge 30 to m ak e th e connections. The
crimp-and-insert connector is shown in Figure 1-10.
NoteThe pin numbering on the crimp-and-insert connector may not agree with the
pin numbering on the Agilent E1418’s faceplate connector. Use the pin
numbering on the faceplate connector to wire the crimp-and-insert connector.
The following access ories are necessary for use with crimp-and-i nsert
Option A3E:
Single-C onductor and
Contact
Shielded-Twisted-Pair and
Contacts
A crimp-and-ins ert cont act is crimped onto one end of a wir e. The other
end is not terminate d. Ord er Agil en t 91510A.
A crimp-and-insert contact is crimped onto each conductor at one end of a
shielded-twisted-pair cable. The other end is not terminated. Order
Agilent 91511A.
Jumper Wire and ContactsA crimp-and-insert contact is crimped onto each end of a single-conductor
jumper wire. Th is jumper is typically used to tie two p ins together in a
single crimp-a nd- insert connecto r. Order Agilent 91512A.
Crimp-and-Insert ContactsThese contacts may be crimped on to a con duc tor a nd the n inserted into a
crimp-and-in se rt con n e cto r . The crimp tool kit is required to crimp the
contacts onto a conductor and remove the c onta c t fro m the c onne c tor .
Order Agilent 91515A.
Crimp-a nd-Insert ToolsThe hand crimp tool ( pa rt number Agilent 9151 8A) is used f or crim ping
contacts onto a con duc tor . The pi n ex tr actor tool (part number
Agilent 91519A) is requ ir ed for r emoving contacts fr om the
crimp-and-insert connector. These products are not included with
Option A3E or with the terminal option accessories listed earlier.
28 Module Setup and Installati onCha pter 1