3
Integration Time Normal mode rejection
(1)
100 PLC/1.67s (2s) 60 dB
(2)
10 PLC/167 ms (200 ms) 60 dB
(2)
1 PLC/16.7 ms (20 ms) 60 dB
(2)
<1 PLC/3 ms (800 µs) 0 dB
(1) For power-line frequency ± 0.1%.
(2) For power-line frequency ± 1%, subtract 20 dB; for ± 3%, subtract 30 dB.
System Speed
(Speeds are for 4.5 digits, Delay 0 and Autozero OFF. Includes measurement and
data transfer over VXI backplane.)
Function change: 30/s
Range change: 65/s
Autorange time: <30 ms
Max. internal trigger rate: 1000/s
Max. external trigger rate to memory: 1000/s
dc:dc Ratio
Measurement method: Input HI-LO/Reference HI-LO
Apply “Reference HI-LO” signal to
Ohms 4-Wire Sense terminals.”
Input HI to Input LO: 100 mV to 300 V
Reference HI to Input LO: <12 V on 100 mV to 10 V ranges
(autoranged)
Reference LO to Input LO: <2 V
Additional Error with Autozero OFF
Following instrument warm-up at calibration temperature ± 1° C and <10 minutes.
100mV-100V ranges: add 0.0002% reading + 5 µV
300 V range: add 0.0006% reading
dc Operating Characteristics
Readings speeds for 60 Hz and (50 Hz) operation, Autozero OFF.
Function NPLC Digits Reading/s Additional Noise
Error
DCV, DCI and Ω 100 6.5 0.6 (0.5) 0% of range
DCV, DCI and Ω 10 6.5 6 (5) 0% of range
DCV, DCI and Ω 1 5.5 60 (50) 0.001% of range*
DCV, DCI and Ω .2 5.5 300 0.001% of range*
DCV, DCI and Ω .02 4.5 1,000 0.01% of range*
*For 300 V range: use 0.003% of range for 5.5 digits and 0.030% of range for 4.5 digits.
For all ranges: add 20 µV for dc Volts, 4 µA for dc current, or 20 mΩ for resistance.
Considerations
Settling considerations: Reading settling times are affected by
source impedance, cable dielectric
characteristics, and input signal
changes.
Measurement considerations: Agilent recommends the use of Teflon
or other high impedance, lowdielectric absorption wire insulation
for these measurements.
Measurement method: Continuously integrating, multi-slope
III A/D converter
A/D linearity: 0.0002% of reading + 0.0001% of
range
2/4-wire Ω: 100 MΩ
Input resistance:
0.1 V, 1 V, 10 V ranges: Selectable 10 MΩ or 10 GΩ
100 V, 300 V ranges: 10 MΩ ± 1%
Input bias current: <30 pA at 25° C
Input terminals: Copper alloy
Input protection: 300 V on all ranges
Resistance
Measurement method: Selectable 4-wire or 2-wire Ω
(Current
source referenced to low input)
Max. lead resistance: (4-wire Ω) 10% of range per lead for
100 Ω and 1 kΩ per lead on all other
ranges
Input protection: 300 V on all ranges
dc Current
Shunt resistor: 0.1 Ω for 1 A and 3 A, 5 Ω for 10 mA
and 100 mA
Input protection: Externally accessible 3 A, 250 V fuse
Measurement Noise Rejection
60 Hz (50 Hz) (For 1 kΩ unbalance in LO lead.)
dc CMMR: 140 dB
dc Voltage Characteristics