Keysight (Agilent) E1411B Data Sheet

Agilent E1411B
5.5-Digit Multimeter, C-Size
Data Sheet
• 1-Slot, C-size, register based
• DCV, ACV, 2- & 4-wire Ω, temperature
• 5.5-digit low-noise integrating A/D
• 13 kHz high-speed sampling A/D
• Software calibration
Description
The Agilent Technologies E1411B 5.5-Digit Multimeter is a C-size,1-slot, register-based VXI module. It is identical in electri­cal design to the E1326B, differing only in size. You can use the integrating A/D to make 5.5-digit, low-noise measurements, or switch to the sampling A/D to make 14-bit readings at rates up to 13 kHz.
When combined with any Agilent VXI relay or FET multiplexer, you can create a multichannel scanning multimeter. For example, by sending just one SCPI command to the E1406A, you can program the multimeter and the channels of your multiplexers all at one time. The E1411B provides flexible triggering with built-in timer pacer, also.
Product functions for this DMM include Vdc/ac, 2- and 4-wire
, offset-compensated , thermocouples thermistors,and
RTDs. This autoranging DMM is especially well suited for data acquisition and computer-aided test applications.
Refer to the Agilent Technologies Website for instrument driver availability and downloading instructions, as well as for recent product updates, if applicable.
2
Reading Rate
Auto zero off, fixed range, default trigger delay, offset comp off, Sample Source “TIMER” for rates >15 readings/s.
Maximum reading rate: 13 K
Typical Reading Rates (rdgs/s)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc voltage: 3 3.5 49 59 365 3125 13000 Four-wire
resistance: 3 3.5 49 59 365 3125 13000 ac voltage: 1.3 1.4 1.9 1.9 1.9 1.9 1.9
Resolution (bits/digit)
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
Binary bits: ±22 ±22 ±20 ±20 ±18 ±15 ±14 Decimal
digits: 6.5 6.5 6 6 5.5 4.5 4
Noise Rejection (dB) for DC voltage and Resistance functions.
Noise Rejection Conditions: CMR measured with 1 kin both High and Low leads with a 10% imbalance. Low connted to Common at source,
measured with respect to earth ground. NMR is for specified frequencies ±0.1%
APERTURE
320 ms 267 ms 20 ms 16.7 ms 2.5 ms 100 µs 10 µs
dc: Common mode rejection 150 dB 150 dB 150 dB 150 dB 150 dB 150 dB 150 dB ac (0-400): Common mode rejection* 70 dB 70 dB 70 dB 70 dB 70 dB 70 dB 70 dB
50 Hz: Power line cycles (NPLC) 16 1 ––––
Normal mode rejection 84 dB 0 dB 60 dB 0 dB 0 dB 0 dB 0 dB Effective Common Mode ECMR* 154 dB 70 dB 130 dB 70 dB 70 dB 70 dB 70 dB
60 Hz: Power line cycles (NPLCs) 16 1
Normal mode (60 Hz) rejection 0 dB 84 dB 0 dB 60 dB 0 dB 0 dB 0 dB Effective Common Mode Rejection* 70 dB 154 dB 70 dB 130 dB 70 dB 70 dB 70 dB
400 Hz: Power line cycles (NPLCs) 128 8 1
Normal mode (400 Hz) rejection 84 dB 0 dB 84 dB 0 dB 60 dB 0 dB 0 dB Effective Common Mode ECMR* 154 dB 70 dB 154 dB 70 dB 130 dB 70 dB 70 dB
*
64 & 300 volt ranges reduced by 36 dB
dc Voltage
Accuracy Conditions: Auto zero on,one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
(Volts) ± (% of Reading + Volts)
Range Input Resistance 20/16.7 ms 10 µs 20/16.7 ms 10 µs
125 mV >100 M 120 nV 7.6 µV 0.023% + 5 µV 0.115% + 60 µV 1 V >100 M 1.0 µV 61 µV 0.013% + 15 µV 0.1% + 200 µV 8 V >100 M 7.6 µV 488 µV 0.01% + 50 µV 0.1% + 1.5 µV 64 V 10 M+ 5% 61 µV 3.9 mV 0.015% + 1 mV 0.1% + 20 µV 300 V 10 M+ 5% 488 µV 31 mV 0.015% + 5 mV 0.1% + 80 µV
Product Specifications
3
Four-Wire Resistance
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 18-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
Source Maximum Open () ± (% of Reading + )
Range Current Circuit Voltage 20/16.7 ms 10 µs 20/16.7 ms 10 µs
256 488 µA 11.5 V 250 µ 15 m 0.035% + 10 m 0.12% + 50 m 2 k 488 µA 11.5 V 2 m 125 m 0.025% + 20 m 0.1% + 200 m 16 k 61 µA 11.5 V 15 m 1 0.025% + 200 m 0.1% + 2 131 k 61 µA 11.5 V 125 m 8 0.025% + 1 0.1% + 16 1 M 7.6 µA 11.5 V 1 64 0.015% + 10 0.1% + 120
NOTE: With offset compensation on, accuracy is the same as for the voltmeter alone.
True RMS ac Voltage (ac coupled)
Crest Factor: 7 at 0% full scale; 5 at full scale. Accuracy Conditions: Sine wave inputs >0% of full scale. dc component <0% of ac component. Auto-zero on, one hour warmup. Temperature within ± 5°C of calibration temperature (module calibrated at 8-28°C).
RESOLUTION vs APERTURE 90-DAY ACCURACY vs APERTURE
Range Input (Volts) ± (% of Reading + Volts) (RMS) Impedance Frequency 320/267 ms 10 µs 320/267 ms All other apertures
87.5 mV >100 M, 20-50 Hz 30 nV 7.6 µV 2.175% +200 µV 2.175% +1 mV
<100 pF 50 Hz-1 kHz 0.675% +200 µV 0.675% +200 µV
5-10 kHz 3.175% +200 µV 3.175% +200 µV
700 mV >100 M, 20-50 Hz 0.24 µV 61 µV 2.125% +1.5 mV 2.125% +8 mV
<100 pF 50 Hz-1 kHz 0.625% +1.5 mV 0.625% +1.5 mV
1-5 kHz 0.625% +1.5 mV 0.625% +1.5 mV
5-10 kHz 3.125% +1.5 mV 3.125% +1.5 mV
5.6 V >100 M, 20-50 Hz 2.0 µV 488 µV 2.125% +15 mV 2.125% +80 mV
<100 pF 50 Hz-1 kHz 0.625% +15 mV 0.625% +15 mV
1-5 kHz 1.125% +15 mV 1.125% +15 mV
5-10 kHz 10.125% +15 mV 10.125% +15 mV
44.8 V 10 M± 5%, 20-50 Hz 15 µV 3.9 mV 2.125% +100 mV 2.125% + 500 mV
<100 pF 50 Hz-1 kHz 0.625% +100 mV 0.625% +100 mV
1-5 kHz 1.125% +100 mV 1.125% +100 mV
5-10 kHz 10.125% +100 mV 10.125% +100 mV
300 V 10 M± 5%, 20-50 Hz 122 µV 31 mV 2.125% + 500 mV 2.125% + 2.5 V
<100 pF 50 Hz-1 kHz 0.625% + 500 mV 0.625% + 500 mV
1-5 kHz 1.125% + 500 mV 1.125% + 500 mV
5-10 kHz 10.125% + 500 mV 10.125% + 500 mV
Timing/Synchronization
Timer/pacer
Timer range: 76 µs to 65.5 ms Resolution: 2 µs
Programmable delay
Delay range: 40 µs to 16 s Resolution: 2 µs
External trigger
Minimum pulse width: 100 ns Maximum trigger rate: 5 kHz (Trigger Condition, negative
edge; Fixed range,10 µs aperture)
Isolation 450 Vpk between any terminal
and chassis.
dc Voltage Accuracy with Relay Multiplexers
Range 20/16.7 ms 10 µs 20/16.7 ms 10 µs
125 mV 0.023%+9 µV 0.115%+64 µV 0.023%+55 µV 0.115%+110 µV 1 V 0.013%+19 µV 0.1%+204 µV 0.013%+65 µV 0.1%+250 µV 8 V 0.01%+54 µV 0.1%+1.5 mV 0.01%+100 µV 0.1%+1.55 mV 64 V 0.015%+1 mV 0.1%+20 mV 0.015%+1.05 mV 0.1%+20 mV 300 V 0.015%+5 mV 0.1%+80 mV 0.015%+5.05 mV 0.1%+80 mV
Accuracy Conditions: Auto zero on, one hour warmup. Temperature within ±5°C of calibration temperature (module calibrated at 8-28°C).