E1411B-01
SUPERSEDES: None
DATE: 1 June 1991
E1411B 5 1/2 Digit Multimeter (VXI C-size)
Serial Numbers: 3002A00202 / 3002A00408
Modifications to prevent erroneous readings and incorrect delay settings.
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
Part No. Qty. Description Ref Des
E1326-88832 1 IC-PRGMD 87C51 U104
E1411-88811 1 IC-PRGMD 87C51 U23
Situation:
If the E1411B is set to an aperture of 10 uSec and the number of triggers (n) in
“TRIG:COUN n” is greater than 1, erroneous readings can be returned by the multimeter.
Also, if the trigger delay (x) in “TRIG:DEL x” is set to uSec, then the multimeter defaults
to the maximum delay of 16.7 seconds.
1 of 2
AUTHOR: KD
ENTITY: 0900
© 1991 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION
CATEGORY:
LOCATION
CATEGORY:
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: June 1992
ADDITIONAL INFORMATION:
Internal reference PCO 09-28504
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN
SCRAP
SEE TEXT
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
STANDARDS:
Labor 0.25 Hours
Continued

Page 2 Service Note E1411B-01
Solution:
The two conditions noted above are corrected by replacing the programmed EPROMs.
Modification can be performed at the next scheduled service or at customer’s convenience. If customer is operating, or intends to operate, the E1411B as described above,
modification should be performed as soon as possible.
Action:
CAUTION:
The EPROMs are static-sensitive devices. Perform the modification at an ESD safe work area!
Remove the top shield from the E1411B (7 screws, Torx T10).
On the E1411-66511 board, remove the E1326-88822 EPROM (U104) and replace it
with the E1326-88832 EPROM. Discard the old EPROM.
Next replace the E1411-88801 EPROM (U23) with the E1411-88811 EPROM. Discard
the old EPROM. Reassemble the E1411B.