Keysight (Agilent) 8960 Quick Referance

How Do I Turn Off a Measurement?
Agilent Technologies 8960 Series 10 E5515A,B Wireless Communications Test Set E1960A GSM Mobile Test Application - Revision A.04
2 3
4
1
1. Press the MEASUREMENT SELECTION key.
2. Highlight the measurement you want to turn off.
3. Press Close Measurement (F4).
4. Press Close Menu (F6).
How Do I Change Call Parameters?
Quick Guide
How Do I Make Measurements on a Mobile?
A Establish a Call.
1
5
2
1. Press the SHIFT key.
2. Press the PRESET key.
3. Connect the Mobile.
Note: Is the Mobile Camped? PGSM is default Cell Band setting.
4. On the Mobile press 1, 2, 3, and then press send.
5. Check for “Connected” in the Active Cell Status field.
3
1
3
1. Press F7, F8, or F9.
2. Enter a value or highlight a selection and press the knob.
Note: For a dual-band handover, change Traffic Band selection (F7 on Call Parms menu 2 of 4).
E1960-90004 (Rev. E) www.Agilent.com/find/8960support Printed in U.S.A. 3/00
B Select Measurements.
2
Indicates measurement being made, result is not being displayed.
1
1. Press the MEASUREMENT SELECTION key.
2. Highlight a measurement and press the knob.
3. Repeat steps 1 and 2 to add measurements.
How Do I Change Measurement Setup?
How Do I Change Cell Parameters?
A Select Measurement.
2
1
1. Press the MEASUREMENT SELECTION key.
2. Highlight a measurement to setup and press the knob.
B Set Up Measurement.
1
2
A Select Cell Parameters Menu.
3
2
1. Press the CALL SETUP key.
2. Press Cell Info (F6).
3. Press Cell Parameters (F2).
B Set a Cell Parameter.
1
4
2
1
3 4
1. Press the setup key (F1).
2. Highlight a parameter and press the knob.
3. Enter a value or selection and press the knob.
Note: For Statistical measurement results, change the Multi-Measure-
ment Count parameter from “Off” to a number >1.
4. Press Close Menu (F6).
3
To change “Network” cell parameters follow the instructions below. For all other cell parameters:
Highlight the parameter, press the knob, enter a value, and press the knob.
To Change “Network” Cell Parameters:
1. Highlight Cell Activated State and press the knob.
2. Set Cell Activated State to “Off”.
Highlight “Off” and press the knob.
3. Set “Network” Cell Parameter to the desired value.
Highlight the parameter, press the knob, enter a value, and press the knob.
4. Set Cell Activated State to “On”.