translation without prior written permission is prohibited, except as allowed under the
copyright laws.
Notices
Subject Matter
The material contained in this document is subject to change without
notice. AGILENT TECHNOLOGIES MAKES NO WARRANTY OF ANY
KIND WITH REGARD TO THIS MATERIAL, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY
AND FITNESS FOR A PARTICULAR PURPOSE. Agilent Technologies
shall not be liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing, performance or
use of this material.
Agilent Technologies assumes no responsibility for the use or reliability of
its software on equipment that is not furnished by Agilent Technologies.
Restricted Rights Legend
If Software is for use in the performance of a U.S. Government contract or
subcontract, Software is delivered and licensed as "Comercial computer
software" as defined in DFAR 252.227-7014 (June 1995), or as a
"commercial item" as defined in FAR 52.227-19 (June 1987) or any
equivalent agency regulation or contract clause. Use, duplication of
Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Department and Agencies of the U.S. Government
will receive no greater than Restricted Rights as defined in FAR
Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent Technologies further certifies that its calibration
measurements are traceable to the United States National Bureau of Standards, to the
extent allowed by the Bureau’s calibration facility, and to the calibration facilities of other
International Standards Organization members.
Assistance
Product maintenance agreements and other customer assistance agreements are available
for Agilent Technologies products.
For any assistance, contact your nearest Agilent Technologies Sales and Service Office.
Statement of Compliance
Electromagnetic Compatibility (EMC) Information
This product has been designed to meet the protection requirements of the
European Communities Electromagnetic Compatibility (EMC) directives:
In order to preserve the EMC performance of the product, any cable which
becomes worn or damaged, must be replaced with the same type and
specification.
EN55011:1991 (Group 1, Class A)
EN50082-1:1992
Safety Information
- IEC 1000-4-2 (1995) ESD
- IEC 1000-4-3 (1995) Radiated Susceptibility
- IEC 1000-4-4 (1995) EFT
This product conforms to EN61010-1(1993) / IEC 1010-1(1990) +A1(1992)
+A2(1994) /CSA C22.2 No. 1010.1(1993) Safety requirements for Electrical
Equipment for Measurement, Control and Laboratory Use, and has been
supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure
safe operation and to maintain the instrument in a safe condition.
Warranty
This Agilent Technologies instrument product is warranted against defects in material
and workmanship for a period of one year from date of shipment. During the warranty
period, Agilent Technologies will, at its option, either repair or replace products which
prove to be defective.
For warranty service or repair, this product must be returned to a service facility
designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent
Technologies and Agilent Technologies shall pay shipping charges to return the product to
Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products
returned to Agilent Technologies from another country.
Agilent Technologies warrants that its software and firmware designated by Agilent
Technologies for use with an instrument will execute its programming instructions when
properly installed on that instrument. Agilent Technologies does not warrant that the
operation of the instrument, or software, or firmware will be uninterrupted or error-free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper
or inadequate maintenance by Buyer, Buyer-supplied software or
interfacing, unauthorized modification or misuse, operation outside of the
environmental specifications for the product, or improper site preparation
or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT
TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE.
Limitation of Remedies and Liability
THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND
EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,
TORT, OR ANY OTHER LEGAL THEORY.
Radio Frequency Interference Statement Deutsche
Bundespost
(Federal Republic of Germany only)
Herstellerbescheinigung
Hiermit wird bescheinigt, daß dieses Gerät, System Agilent 83220A/E in
Ûbereinstimmung mit den Bestimmungen von Postverfügung 526/527
funkentstört ist.
Der Deutschen Bundespost wurde das Inverkehrbringen dieses
Gerätes/Systems angezeigt und die Berechtigung zur Ûberprüfungder
Serie auf Einhaltung der Bestimmungen eingeräumt.
Zusatzinformation für Meß- und Testgeräte:
Werden Meß - und Testgeräte mit ungeschirmten Kabeln und/oder in
offenen Meßaufbauten verwendet, so ist vom Betreiber sicherzustellen,
daß die Funk-Entstörbestimmungen unter Betriebsbedingungen an seiner
Grundstücksgrenze eingehalten werden.
Manufacturer’s Declaration
This is to certify that the equipment Agilent 83220A/E meets the radio
frequency interference requirements of Directive FTZ 526/527. The
German Bundespost has been notified that this equipment was put into
circulation and has been granted the right to check the product type for
compliance with these requirements.
Additional Information for Test and Measurement Equipment:
If test and measurement equipment is operated with unscreened cables
and/or used for measurements on open set-ups, the user has to assure that
under operating conditions the Radio Interference Limits are still met at
the border of the user’s premises.
Sound Emission
Manufacturer’s Declaration
This statement is provided to comply with the requirements of the
German Sound Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position)
< 70 dB(A).
• Sound Pressure Lp < 70 dB(A).
• At Operator Position.
• Normal Operation.
• According to ISO 7779:1988/EN 27779:1991 DIN 45635 pt. 19 (Type
Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlärminformationsverordnung vom 18 Januar 1991.
• Schalldruckpegel Lp < 70 dB(A).
• Am Arbeitsplatz.
• Normaler Betrieb.
• Nach ISO 7779:1988/EN 27779:1991 DIN 45635 pt. 19 (Typprüfung).
Responsibilities of the Customer
The customer shall provide;
1. Access to the products during the specified periods of coverage to perform maintenance.
2. Adequate working space around the products for servicing by Agilent Technologies
personnel.
3. Access to and use of all information and facilities determined necessary by Agilent
Technologies to service and/or maintain the products. (In so far as these items may
contain proprietary or classified information, the customer shall assume full
responsibility for safeguarding and protection from wrongful use.)
4. Routine operator maintenance and cleaning as specified in the Agilent Technologies
Operating and Service Manuals.
5. Consumables such as paper, disks, magnetic tapes, ribbons, inks, pens, gases, solvents,
lamps, filters, fuses, seals, etc.
General Safety Considerations
The following general safety precautions must be observed during all phases of operation,
service, and repair of this instrument. Failure to comply with these precautions or with
specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent Technologies Inc. assumes no
liability for the customer’s failure to comply with these requirements.
WARNING
DO NOT operate the product in an explosive atmosphere or in the presence of flammable
gasses or fumes.
DO NOT use repaired fuses or short-circuited fuseholders: For continued protection
against fire, replace the line fuse(s) only with fuse(s) of the same voltage and current
rating and type.
DO NOT perform procedures involving cover or shield removal unless you are qualified to
do so: Operating personnel must not remove equipment covers or shields. Procedures
involving the removal of covers and shields are for use by service-trained personnel only.
DO NOT service or adjust alone: Under certain conditions, dangerous voltages may exist
even with the equipment switched off. To avoid dangerous electrical shock, service
personnel must not attempt internal service or adjustment unless another person, capable
of rendering first aid and resuscitation, is present.
DO NOT operate damaged equipment: Whenever it is possible that the safety protection
features built into this product have been impaired, either through physical damage,
excessive moisture, or any other reason, REMOVE POWER and do not use the product
until safe operation can be verified by service-trained personnel. If necessary, return the
product to an Agilent Technologies Sales and Service Office for service and repair to
ensure the safety features are maintained.
This is a Safety Class I instrument (provided with a protective
earthing ground, incorporated in the power cord). The mains plug
shall only be inserted in a socket outlet provided with a protective
earth contact. Any interruption of the protective conductor inside or
outside of the instrument is likely to make the instrument
dangerous. Intentional interruption is prohibited.
DO NOT substitute parts or modify equipment: Because of the danger of introducing
additional hazards, do not install substitute parts or perform any unauthorized
modification to the product. Return the product to a Agilent Technologies Sales and
Service Office for service and repair to ensure the safety features are maintained.
WARNING
There are many points in the instrument which can, if contacted,
cause personal injury. Be extremely careful.
Any adjustments or service procedures that require operation of the
instrument with protective covers removed should be performed
only by trained service personnel.
Safety Symbols
The following symbols on the instrument and in the manual indicate precautions which
must be taken to maintain safe operation of the instrument
Safety Symbols
The Instruction Documentation Symbol. The product is
marked with this symbol when it is necessary for the user
to refer to the instructions in the supplied documentation.
Indicates the field wiring terminal that must be connected
to earth ground before operating the equipment - protects
against electrical shock in case of fault.
Frame or chassis ground terminal - typically connects to
the equipment's metal frame.
Alternating current (AC)
Direct current (DC)
Warning,risk of electric shock.
WARNINGWarning denotes a hazard. It calls attention to a
procedure, which if not correctly performed or adhered to
could result in injury or loss of life. Do not proceed beyond
a warning note until the indicated conditions are fully
understood and met.
CAUTIONCaution denotes a hazard. It calls attention to a
procedure, which if not correctly performed or adhered to
could result in damage to or destruction of the instrument.
Do not proceed beyond a warning note until the indicated
conditions are fully understood and met.
The CE mark shows that the product complies with all
relevant European Legal Directives.
ISM 1-AThis is a symbol of an Industrial, Scientific, and Medical
Group 1 Class A product.
The CSA mark is a registered trademark of the Canadian
Standards Association, and indicates compliance to the
standards defined by them.
Indicates that a laser is fitted. The user must refer to the
manual for specific Warning or Caution information to
avoid personal injury or damage to the product.
ASIA PACIFIC
Agilent Technologies
19/F,Cityplaza One,
1111 King’s Road,
Taikoo Shing,
Hong Kong, SAR
(tel) (852) 2599 7889
(fax) (852) 2506 9233
CANADA
Agilent Technologies
Canada Inc.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414
JAPAN
Agilent Technologies Japan Ltd
Measurement Assistance
Center
9-1, Takakura-Cho,
Hachioji-Shi, Tokyo, 192-8510
(tel) (81) 426 56 7832
(fax) (81) 426 56 7840
AUSTRALIA/
NEW ZEALAND
Agilent Technologies Australia
Pty Ltd
347 Burwood Highway
Forest Hill, Victoria 3131
(tel) 1-800 629 485 (Australia)
(fax) (61 3) 9272 0749
(tel) 0 800 738 378 (New Zeal’d)
LATIN AMERICA
Agilent Technologies
Latin America Region Headquarters
5200 Blue Lagoon Drive,
Suite #950
Miami, Florida 33126
U.S.A.
(tel) (305) 267 4245
(fax) (305) 267 4286
EUROPE
Agilent Technologies
Test & Measurement
European Marketing Org.
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (31 20) 547 9999
UNITED STATES
Agilent Technologies
Test and Measurement Call
Center
P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 488
ii
Agilent 83220A/E Description
Agilent 83220A/E Capability
The Agilent 83220A/E expands the capabilities of the HP/Agilent 8922 GSM Test Set
family to comprehensively test DCS1800 mobile radios across the band 1710 MHz to 1880
MHz; and PCS1900 mobile radios across the band 1850 MHz to 1990 MHz. All the features
of the RF generator, RF analyzer and general instrumentation of the HP/Agilent 8922
series are retained.
DCS1800 Base Station
The Agilent 83220A/E, when configured with the HP/Agilent 8922, can now emulate a
DCS1800/PCS1900 base station and establish and maintain a DCS1800/PCS1900 link.
Once a link has been established, comprehensive tests can be sequenced across the full
DCS1800/PCS1900 band without interrupting the link.
NOTE
PCS1900 is configured as a standard on all Agilent 83220Es with serial prefix
3524U00479 or greater. Prior to this serial break PCS1900 was available on
the Agilent 83220E as options K19.
PCS1900 is configured as a standard on all Agilent 83220As with serial prefix
3524U00424 or greater. Prior to this serial break PCS1900 was available on
the Agilent 83220A as options K19.
Your HP/Agilent 8922 must also be configured for PCS1900 operation:
HP/Agilent 8922A/B require firmware revision B.05.04 or greater.
HP/Agilent 8922E/G do not support PCS1900.
HP/Agilent 8922F require firmware revision B.05.02 or greater to support
PCS1900.
HP/Agilent 8922H require firmware revision B.06.00 or greater to support
PCS1900.
iii
How to Use This Manual
Where to Start
If you have just received the Agilent 83220A/E and want to get it ready to use for the first
time:
Thoroughly read the “Installing YourAgilent 83220A/E” chapter in this manual and follow
its instructions for:
• Unpacking the Test Set.
• Preparing it for use.
• Installation Overview.
• System Calibration.
• DCS1800/PCS1900 Configuration.
If you need to verify the unit is operating within its specifications, perform the confidence
tests in the “Verifying Performance” chapter in this manual.
Making measurements in DCS1800/PCS1900 mode is identical to making measurements
in GSM900 mode. Follow the instructions in the “Making Measurements” chapter in the
appropriate HP/Agilent 8922 Series User’s Guide to learn how to make measurements
using the DCS1800 Test System.
To find more detailed information about the Test Set, its applications, and key
descriptions, consult the “Screens”, “Connectors”, and “GPIB” chapters in this manual, and
in the appropriate HP/Agilent 8922 Series User’s Guide.
To learn how to program the Test Set, read the “GPIB” chapter in this manual, and the
“GPIB”, “Programming Examples”, and “Programmer’s Reference” chapters of the
appropriate Agilent 8922M/S Programming Reference Guide
NOTE
WARNING
If you have the Agilent 83220A/E Option 010 Multi-Band Test System, refer
to the appropriate HP/Agilent 8922 Multi-Band User’s Guide for more details
on connection and operating differences.
If this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must only be used
in a normal condition (in which all means of protection are intact).
iv
v
1Installing Your Agilent 83220A/E
1
Installing Your Agilent 83220A/E
Introduction
Introduction
This section provides installation instructions for the DCS1800 Test System (Agilent
83220A/E and Agilent/HP 8922).
The Agilent 83220A/E User’s Guide covers the operation of the Agilent 83220A/E in
conjunction with an appropriate HP/Agilent 8922 Series GSM Test Set, extending the
capabilities of the HP/Agilent 8922 Series to cover the DCS1800/PCS1900 frequency
bands. The Agilent 83220A/E and HP/Agilent 8922 together constitute a
DCS1800/PCS1900 Test System.
Using this Chapter
Use the following procedure to power-up the Agilent 83220A/E, HP/Agilent 8922 DCS1800
TestSystem correctly. After completion of this procedure, refer to “Making Measurements”
in the appropriate HP/Agilent 8922 Series User’s Manual for an introduction to operating
the appropriate HP/Agilent 8922. Select the test required, follow the instructions as in the
appropriate HP/Agilent 8922 manual. Make the measurement.
The information contained in this guide includes:
❏ Accessories supplied.
❏ Initial inspection.
❏ Installation check list.
❏ Preparation for use.
❏ Fuses and power cords.
❏ Installation overview
❏ Running the System Calibration routine.
❏ DCS1800/PCS1900 configuration.
❏ General information.
The interface between the HP/Agilent 8922 and the Agilent 83220A/E is via a 15 way
D-Type connector and five BNC coaxial connectors.
NOTE
DCS1800 Test System setup is shown for the HP/Agilent 8922H. DCS1800
Test System setup using any of the other HP/Agilent 8922 Series will be
similar. The HP/Agilent 8922 rear panel layout is slightly different for each
model in the series.
NOTE
WARNING
2Chapter1
If you have the Agilent 83220A/E Option 010 Multi-Band Test System, refer
to the appropriate HP/Agilent 8922 Multi-Band User’sGuide for more details
on connection and operating differences.
If this instrument is not used as specified, the protection provided by
the equipment could be impaired. This instrument must only be used
in a normal condition (in which all means of protection are intact).
Accessories Supplied
Table 1-1 Accessories Supplied with the DCS1800 Test System
DescriptionAgilent Part Number Comments
Installing Your Agilent 83220A/E
Introduction
User’s Guide83220-90027
Interconnecting CablesSee Table 1-2Shipped with DCS1800 Test System.
Power CordSee Figure 1-2Shipped with DCS1800 Test System.
Rackmount Flange Kit 5062-4071Shipped with Option AX4
GSM/DCS1800 Mobile
Station Test Software
Table 1-2 Supplied Cables
TypePartNumber
BNC8120-1838CONNECT TO 8922
BNC8120-1838CONNECT TO 8922
BNC83220-6102510 MHz IN
83212DAvailable separately
Model
Number
AUX RF OUT
(Front panel)
AUX RF IN
(Front panel)
(Rear panel)
1 copy shipped with Agilent 83220A/E.
Agilent 83220A/E
Port
HP/Agilent 8922Port
AUX RF OUT
(Front panel)
AUX RF IN
(Front panel)
10 MHz OUT
(Rear panel)
BNC83220-61025SCOPE
(Rear panel)
BNC83220-61025AM
(Rear panel)
D-type08922-61083PCN INTERFACE
(Rear panel)
CAUTION
It is advisable to connect all the cables as shown in the Installation
SCOPE
(Rear panel)
AM
(Rear panel)
PCN INTERFACE
(Rear panel)
Overview section before switching on the DCS1800 Test System.
Initial Inspection
Inspect the shipping container for damage. If the shipping container or cushioning
material is damaged, it should be kept until the contents of the shipment have been
checked for completeness and the Agilent 83220A/E has been checked mechanically and
electrically. Procedures for checking the electrical performance of the DCS1800 Test
System are given in the “Verifying Performance” chapter of this manual. Procedures for
checking the electrical performance of the HP/Agilent 8922 are given in the “Verifying
Performance” chapter of the appropriate HP/Agilent 8922 Series User’s Guide. If the
Chapter 13
Installing Your Agilent 83220A/E
Introduction
contents are incomplete, if there is mechanical damage or defect, or if the DCS1800 Test
System does not pass the electrical performance tests, notify the nearest Agilent
Technologies office.
If the shipping container is damaged, or the cushioning material shows signs of stress,
notify the carrier as well as the Agilent Technologies office. Keep the shipping material for
the carrier’s inspection.
WARNING
To avoid hazardous electrical shock, do not apply power to the
Agilent 83220A/E when there are any signs of shipping damage to any
part of the Agilent 83220A/E.
If the shipping materials are in good condition, retain them for possible future use. You
may wish to ship the Agilent 83220A/E to another location or to return it to Agilent
Technologies for service. See “How to Return Your Agilent 83220A/E for Servicing,” in this
chapter.
Installation Check List
Before connecting the Agilent 83220A/E or HP/Agilent 8922 to the line (mains) voltage,
ensure the following steps are taken:
1. Read the warning in this chapter, see Preparation for Use.
2. Check that the line voltage setting on the rear panel power line modules are set to the
correct voltage, see Fuses and Power Cords.
3. Ensure that the rating of the line (mains) fuse is appropriate for the line voltage being
used, see Table 1-3 on page 6.
4. Ensure that the power cable is the correct type, see Power Cords.
Preparation for Use
WARNING
The Agilent 83220A/E and HP/Agilent 8922 Series are Safety Class 1
products (provided with a protective earth terminal). An
uninterruptible safety earth ground must be provided from the
mains power source to the product input wiring terminals, power
cord, or supplied power cord set. Whenever it is likely that the
protection has been impaired, the DCS1800 Test System must be
made inoperative and be secured against any unintended operation.
If the Agilent 83220A/E or HP/Agilent 8922 are to be energized via an
external autotransformer for voltage reduction, make sure that the
common terminal is connected to the neutral pole of the power
source.
Before connecting the Agilent 83220A/E or HP/Agilent 8922, the
protective earth terminal of the Agilent 83220A/E and HP/Agilent
8922 must be connected to the protective conductor of the (mains)
power cord. The mains plug shall only be inserted in a socket outlet
provided with a protective earth contact. The protective action must
not be negated by the use of an extension cord (power cable) without
a protective conductor (grounding). Grounding one conductor of a
two-conductor outlet is not sufficient protection.
4Chapter1
Fuses and Power Cords
Line Voltage Selection and Fuse Selection
Installing Your Agilent 83220A/E
Fuses and Power Cords
CAUTION
A rear-panel, line power module permits operation from 100, 120, 220, or 240 Vac. The
HP/Agilent 8922 power module is documented in the Fuses and Power Cords section of
“Installing YourHP/Agilent 8922” in the appropriate HP/Agilent 8922 Series User’s guide.
On the Agilent 83220A/E, the number visible in the window (located on the module)
indicates the nominal line voltage to which the Agilent 83220A/E must be connected.
Verify that the Line VoltageSelection Cam is matched to the power source (see Figure 1-1).
Table 1-3 lists the ratings and the Agilent part numbers of the line fuses.
Before connecting the DCS1800 Test System to the mains (line) voltage,
ensure that the correct operating voltages and fuses have been selected for
each instrument.
Internal Fuses
The fuses used within this instrument are listed below, and are non operator replaceable.
These fuses are fast blow type (F).
FuseCurrentVoltageType
F13.0A250VF
F26.0A250VF
NOTE
The line voltage and line fuse ratings are selected at the factory according to
the line voltage specified when the Agilent 83220A/E and HP/Agilent 8922
were ordered. If the voltage was not specified, the line voltage and line fuse
rating are selected according to the country of destination.
Line Voltage Selection
1. Open the Cover Door by placing a small standard screwdriver between the Cover Door
and the Line Power Module, and pry open.
2. Remove the Line Voltage Selection Cam with fingers. Rotate cam so that the desired
line voltage appears at the Cover Door Window when the cam is re-inserted.
Fuse Replacement
1. Remove Fuse Holder with fingers. Replace Fuse in Fuse Holder and re-insert into the
Line Power Module.
2. Close the Cover Door by pressing it firmly.
Chapter 15
Installing Your Agilent 83220A/E
Fuses and Power Cords
Figure 1-1 Fuse Replacement and Line Voltage Selection
Cover Door Window
Cover Door
Line Voltage Selection Cam
Line Power Module
Fuse
Fuse Holder
WARNING
For protection against fire hazard, the line fuse should only be a
250V fuse with the correct current rating. (See Table 1-3).
Table 1-3 Line Fuse Rating and HP Part Number
ModelLine VoltageRatingTypePart Number
HP/Agilent
8922
Agilent
83220A/E
100, 120 Vac
220, 240 Vac
100, 120 Vac
220, 240 Vac
5.0A, 250V
2.5A, 250V
2.0AT, 250V
1.5AT, 250V
Normal blow
Normal blow
Time delay
Time delay
2110-0010
2110-0083
2110-0303
2110-0304
Power Cords
The Agilent 83220A/E and HP/Agilent 8922 are each supplied with a three-wire power
cable. When connected to an appropriate ac power receptacle, these cables ground the
DCS1800 Test System cabinets. The type of Mains plug shipped with each instrument
depends on the country of destination. Refer to Table 1-4 on page 7 for the part numbers of
the power cables and Mains plugs available.
6Chapter1
Table 1-4 Line Cords
Installing Your Agilent 83220A/E
Fuses and Power Cords
PlugType
a
Length,
inches
(mm)
90 (229)
90 (229)
Cable
Color
Mint Gray
Mint Gray
For Use In
Country
United Kingdom,
Cyprus, Nigeria,
Cable Part
NumberCD
8120-1351
8120-170304
Plug Description
90/Straight BS1363A
90
Rhodesia,
Singapore
8120-1369
8120-0696
04Straight
NZSS198/ASC112
Straight/90
79 (201)
87 (221)
a
Gray
Gray
Australia,
Argentina,
New Zealand,
Mainland China
8120-1689
8120-1692
72Straight
90
79 (201)
79 (201)
Mint Gray
Mint Gray
East and West
Europe, Central
African Republic,
Arabia, Egypt
8120-1378
8120-4753
8120-1521
8120-4754
8120-1348
8120-153823
8120-2104
1
Straight
6
NEMA5-15P
Straight
1
90
90
Straight
90
3
Straight SEV 1011
80 (203)
a
90 (230)
80 (203)
90 (230)
Jade Gray
Jade Gray
United States,
Canada, Mexico,
Philippines,
Taiwan, Japan
a. Part number shown for plug is industry identifier for plug only. Number shown
for cable is Agilent Part Number for complete cable including plug. E = Earth
Ground; L = Line; N = Neutral.
Chapter 17
Installing Your Agilent 83220A/E
Operating Environment
Operating Environment
Temperature:
from 0 to +55
Humidity:
95% (non-condensing). However, protect the Agilent 83220A/E from temperature extremes
which can cause condensation within the Agilent 83220A/E.
Cooling:
its rear panel.
CAUTION
The Agilent 83220A/E may be operated in environments with temperatures
°C.
The Agilent 83220A/E may be operated in environments with humidity from 0 to
The Agilent 83220A/E obtains cooling airflow by forced ventilation from the fan on
Ensure that all airflow passages at the rear and side of the Agilent 83220A/E
are clear before installing the Agilent 83220A/E. This is especially important
in a rack mount configuration.
Rack Mounting the Agilent 83220A/E
If the Agilent 83220A/E is to be rack mounted, ensure that all airflow passages at the rear
and sides of the Agilent 83220A/E are not obstructed. Full instructions for rack mounting
the Agilent 83220A/E are included with the rack mounting kit.
8Chapter1
Installation Overview
Installing Your Agilent 83220A/E
Installation Overview
NOTE
1. Unpack the Agilent 83220A/E.
2. On the front panel, connect the Agilent 83220A/E to the HP/Agilent 8922 as shown
using the cables provided. (Part No. 8120-2682).
DCS1800 Test System setup is shown for the HP/Agilent 8922H and Agilent
83220A. DCS1800 Test System setup using any of the other HP 8922 Series
will be similar.The HP/Agilent 8922 rear panel layout is slightly different for
each model in the series.
3. On the rear panel, connect AM to AM and SCOPE to SCOPE as shown using the cables
provided. (Part No. 83220-61025).
Chapter 19
Installing Your Agilent 83220A/E
Installation Overview
4. On the rear panel, connect Agilent 83220A/E 10 MHz IN to HP/Agilent 8922 10 MHz
OUT and connect the PCN INTERFACE cable as shown using the cables provided.
(Part No. 83220-61025 and 08922-61083).
5. Connect a 1, 2, 5, 10, or 13 MHz signal to HP/Agilent 8922 REF IN. If you have OPT
001, connect as shown using the cable provided. Refer to the Installation Overview
section in the “Installing YourHP/Agilent 8922” chapter of the appropriate HP/Agilent8922 Series User’s Guide for information on configuring the reference.
6. Connect the supplied power cords to the Agilent 83220A/E and HP/Agilent 8922 and
power up the instruments.
10Chapter1
Installing Your Agilent 83220A/E
Running the System Calibration Routine (Agilent 83220A only)
Running the System Calibration Routine (Agilent
83220A only)
The System Calibration is a user-initiated calibration routine performed between the
HP/Agilent 8922 and the Agilent 83220A.
NOTE
To perform a System Calibration:
1. Powerdown the DCS1800 Test System (Agilent 83220A and HP/Agilent 8922), achieved
by powering down the HP/Agilent 8922.
2. On the Agilent 83220A, using a short cable (<30cm) (Part No. 10502A or similar) and 2
BNC to N-type adapters (Part No. 1250-0780 or similar), link the AUX RF OUT to the
RF IN/OUT port.
3. On the HP/Agilent 8922, using a short cable (<30cm) (Part No. 10502A or similar), link
the AUDIO OUT port to the TRIGGER IN port.
4. Power up the DCS1800 Test System.
5. From the HP/Agilent 8922 front panel, select the MESSAGE Screen.
6. Look for any error message.
7. There should be no error messages before starting the System Calibration Procedure, if
any errors occur, error messages are listed in the “Messages” chapter in this manual,
and in the appropriate HP/Agilent 8922 Series User’s Guide.
The System Calibration routine should be run whenever an Agilent 83220A is
first connected to an HP/Agilent 8922, or whenever a different Agilent
83220A is connected to an HP/Agilent 8922, and after a RAM Initialize from
the SERVICE screen.
SHIFT, MSSG.
8. From the HP/Agilent 8922 front panel, select the TESTS Screen.
9. Using the knob, move the cursor to the Location field, and select ROM from the
Choices field.
10.Move the cursor to the Procedure field and select SYS_CAL from the Choices field.
11.Using the knob, select Run Test.
12.Follow the instructions on the screen.
13.When the System Calibration Routine has finished, Cycle the DCS1800 Test System
power.
Chapter 111
TESTS
Installing Your Agilent 83220A/E
RAM Initialization (Agilent 83220E Only)
RAM Initialization (Agilent 83220E Only)
NOTE
To perform a RAM Initialize:
1. From the home screen, select the ``CONFIG'' Screen.
2. From the ``CONFIG'' Screen, select the ``SERVICE'' Screen.
3. Move the cursor next to the RAM Initialize field.
4. Select YES.
5. The instruments will then follow a restart cycle, after which, calibration information
will be down loaded, and the instruments will be ready to use.
A RAM Initialize from the ``SERVICE'' screen should be run whenever an
Agilent 83220E is first connected to an HP/Agilent 8922 or whenever a
different Agilent 83220E is connected to an HP/Agilent 8922.
12Chapter1
Installing Your Agilent 83220A/E
DCS1800/PCS1900 Configuration
DCS1800/PCS1900 Configuration
1. Access, then set up the CONFIGURE Screen with the cursor. Refer to the “Screens”
chapter in this manual and in the appropriate HP/Agilent 8922 Series User’s Guide.
2. Select the Radio Type field. From the list of choices, select either DCS1800 or PCS1900
depending on the radio type you are testing.
NOTE
CAUTION
Chapter 113
When making DCS1800/PCS1900 measurements, use the RF IN/OUT port,
AUX RF OUT port or AUX RF IN port on the Agilent 83220A/E. Making
measurements in DCS1800 mode is identical to making measurements in
GSM900 mode. Follow the instructions in the “Making Measurements”
chapter in the appropriate HP/Agilent 8922 Series User’s Guide.
When making GSM900 measurements, disconnect the connections to the
Agilent 83220A/E and use the RF IN/OUT port, AUX RF OUT port or AUX
RF IN port on the HP/Agilent 8922. The maximum Agilent 83220A/E RF
IN/OUT port input power is 2W. Never connect a GSM900 radio to the
Agilent 83220A/E or irreparable damage may occur.
Installing Your Agilent 83220A/E
How to Return Your Agilent 83220A/E for Service
How to Return Your Agilent 83220A/E for Service
If you need to return your Agilent 83220A/E to Hewlett-Packard, first obtain the correct
Service Center shipping address from your local sales office and carry out the following
procedure.
CAUTION
1. Please send the following information with the returned instrument:
a. Type of service required.
b. Description of the problem and whether it is constant or intermittent.
c. Name and phone number of technical contact person.
d. Return address.
e. Model number of returned instrument.
f. Full serial number of returned instrument.
g. List of any accessories returned with instrument
2. Send copies of any performance data recorded for the instrument.
3. Pack the instrument in the original shipping materials or the equivalent.
4. If the original shipping materials are not available, follow these steps to repackage the
instrument for shipment:
Instrument damage can result from using packaging materials other than the
original shipping materials or equivalent. Never use styrene pellets as
packaging materials. They do not adequately cushion the instrument, do not
prevent it from shifting in the carton, and can cause instrument damage due
to ESD.
a. Wrap the instrument in antistatic plastic to reduce the possibility of ESD-caused
damage.
b. Obtain a double-walled, corrugated cardboard carton of 159 kg (350 lb) test strength.
The carton must be large enough and strong enough to accommodate the
instrument. Allow at least 7 to 10 cms on all sides of the instrument for packing
material.
c. Surround the instrument with 7 to 10 cms of packing material to protect it and
prevent it from moving in the carton.
5. Seal the carton with strong nylon adhesive tape.
6. Mark the carton "FRAGILE, HANDLE WITH CARE."
7. Retain copies of all shipping papers.
14Chapter1