Keysight (Agilent) 8922P Service Note

8922P-04
SERVICE NOTE
SUPERSEDES: None
8922P Multiband Test Set
Serial Numbers: 0000A00000 / 9999Z99999
Service Information for verifying signals at System Bus connector
Duplicate Service Notes:
8922X-04
Situation:
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
AS
© 1999 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
ENTITY:
E600
Continued
DATE:March 1999
ADDITIONAL INFORMATION:
This will not be a high volume option.
The user manual for #H09/K09 is 08922-90218
1 of 2
Page 2 Service Note 8922P-04
Solution / Action:
Service Information for verifying signals at System Bus connector. Output at pin 5 using DMM (Pin 5 normally TTL hi level).
Check 1
Set service latch ’g_run_reset_cntl = 0’ on service screen. Pin 5 should be at UL 0.8 VDC, LL 0 VDC. Check 2 Set service latch ’g_run_reset_cntl = 1’. Confirm Pin 5 high, ie voltage UL = 5 VDC, LL= 2.4 VDC.
Input level to pin 3 using DC level Check 3 Apply lo level (0.5VDC or ground) to Pin 3. Confirm error ’Call disconnected:Ph error:0x009f’ is seen on instrument screen Check 4 Apply high (5VDC) to Pin 3 Confirm no recurrance of error ’Call disconnected:Ph error:0x009f’ on screen
Output at pin 24 using DMM (Pin 24 normally TTL hi level). Check 5 Set service latch ’g_ext_trig_out = 0’. Pin 24 should be at UL 0.8VDC, LL 0 VDC. Check 6 Set service latch ’g_ g_ext_trig_out = 1’. Confirm Pin 24 high, ie voltage UL = 5 VDC, LL= 2.4 VDC.
These checks confirm operation of the signals at the system bus connector.