Keysight (Agilent) 8922P Service Note

8922P-03
SERVICE NOTE
SUPERSEDES: None
8922P GSM MS Test Set
Serial Numbers: 0000A00000 / 9999Z99999
Flash firmware download will overwrite cyphering (#H05) ROM.
Duplicate Service Notes:
8922M-06 8922P-03 8922X-03
To Be Performed By: Agilent-Qualified Personnel
Situation:
A flash firmware upgrade of the instrument will overwrite the contents of the #H05 ROM erasing the contents permanently.
#H05 is a licenced encryption facility for GSM and is not a common option in 8922 (Currently under 50 units world wide). It is not on CPL.
This option consists of one ROM, GSM Controller board U17. #H05 Encryption ROM for firmware B.07.07 or below is 08922-87249 (8922H/M). #H05 Encryption ROM for firmware B.07.08 or above is 8922M 08922-87391, #H05 Encryption ROM for firmware C.01.04 or above is 8922P/X (ie 8922M#010) 08922-87392.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
AS
ENTITY:
E600
Continued
DATE: January 1999
ADDITIONAL INFORMATION:
Warning also visible on firmware web page
© 1999 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
1 of 2
Page 2 Service Note 8922P-03
To date (Sept 98) the #H05 option has NOT been identified on the instrument - this option will be identified on new instruments and retro-fit labels are being distributed to current instruments which will include the serial number of the instrument to which the licence applies.
How to check if #H05 is fitted : a) On MS INFO screen, select CYPHERING. If options are OFF and DISABLED,
#H05 is not fitted. If ENABLED is also an option, then #H05 is fitted.
b) On GSM Controller board, if U17 is one of above ROMs, then #H05 is fitted.
Solution / Action:
Having identified that #H05 is fitted, there are 3 scenarios to consider :
1) Upgrading from below B.07.07 to B.07.07.
The #H05 ROM 08922-87249 needs removed from the instrument and a ’dummy’ ROM inserted for the duration of theupgrade. ’Dummy’ ROM part no is 1818-5035. Remove ’dummy’ and replace with #H05 ROM after upgrade is complete.
2) Upgrading from B.07.08/C.01.04 to later versions The #H05 ROM 08922-87391or 08922-87392 needs removed from the instrument and a ’dummy’ ROM inserted for the duration of the upgrade. ’Dummy’ ROM part no is 1818-5035. Remove ’dummy’ and replace with #H05 ROM after upgrade is complete. (ROM part numbers also advised at firmware web page)
3) Upgrading from B.07.07 or below to B.07.08/C.01.04 or above. This is the transition which will require ROM upgrading. Please contact Linda Keir, QMD Specials Engineering, with the instrument serial number. The original #H05 ROM will need to be returned to QMD after completion of the upgrade to comply with the licence conditions.