8922P-02
SERVICE NOTE
SUPERSEDES: None
8922P Multiband Test Set
8922M #010 A33 HOP Controller Board 08922-60202
Serial Numbers: 0000A00000 / 9999Z99999
8922 Hop controller board U62 - component part number transition
Duplicate Service Notes:
8922M-05
8922X-02
To Be Performed By: Agilent-Qualified Personnel.
Situation:
8922P options H09/K09 and H17/K17 require an updated component U62 fitted on the A33
Hop Controller board 08922-60202. The new ROM is 08922-87390. Options
H09/K09/H17/K17 are currently being trialled with selected customers prior to release and
may be installed on some units returned for repair.
Solution / Action:
If it is necessary to replace the Hop controller board during a repair, check which U62 ROM
is fitted. If 08922-87390 is in the faulty board it should be installed in the new board during
the repair and the ROM that comes with the new board (08922-87057) discarded.
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
INFORMATION ONLY
AUTHOR:
AS
ENTITY:
E600
DATE: December 1998
ADDITIONAL INFORMATION:
None
© 1998 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
1 of 1