Safety SummaryThe following general safety precautions must be observed during all phases of
operation of this instrument. Failure to comply with these precautions or with
specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the instrument. Agilent Technologies Inc.
assumes no liability for the custome r’s failure to comply with thes e requireme nts.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth
terminal). The protectiv e features of this product may be impaired if it is used in a
manner not specified in the operat ion instru ctions.
All Light Emitting Diodes (L EDs) used in this product are Class 1 LEDs as per IEC
60825-1.
This product has been designed and tested in accordance with IEC Publication 1010, "Safety Requirements for Electronic Measuring Apparatus," and has been
supplied in a s afe c ondi tion. This instruction documentation cont ai ns information
and warnings which must be followed by the user to ensure safe operation and to
maintain the product in a safe condition.
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution
degree 2 environment. It i s designed to opera te at a maximum relati ve humidity of
95% and at altitudes of up to 2000 meters. Re fer to the s pecificati ons tables f or the
ac mains voltage requirements and ambient operating temperature range.
V entilation Requirements: When installing the product in a cabinet, the convection
into and out of the product must not be restricted. The ambient temperature
(outside the cabin et) must be les s than t he maximum oper atin g temper atu re of t he
product by 4° C for every 100 watts dissipated in the cabinet. If the total power
dissipated in th e cabi net is great er th an 8 00 watts , then force d con vecti on must b e
used.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is
installed, and all safety precautions are taken . Note the instrument's external
markings described under Safety Symbols.
3
GROUND THE INSTRUMENT
T o mi nimize s hock hazard , the inst rument chas sis and c over must be connec ted to
an electrical protective earth ground. The instrument must be connected to the ac
power mains through a grounded power cable, with the ground wire firmly
connected to an electrical ground (safety ground) at the power outlet. Any
interruption of the protective (grounding) conductor or disconnection of the
protective earth terminal will caus e a potential shock haza rd that could result in
personal injury.
FUSES
Only fuses with the required rate d current, voltage, and specified type (normal
blow , time de lay, etc.) should be use d. Do not use re paired fus es or short -circuit ed
fuse holders. To do so could cause a shock or fire hazard.
DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE
Do not operate the instrument in the presence of flammable gases or fumes.
DO NOT REMOVE THE INSTRUMENT COVER
Operating personnel mus t not remove in strument cover s. Component repl acement
and internal adjustments must be made only by qualified service personnel.
Instruments that appear damaged or defective should be made inoperative and
secured against unintended operation until they can be repaired by qualified
service personnel.
WARNING:The WARNING sign denotes a hazard. I t calls at tention to a procedure, practice, o r
the like, which, if not correctly performed or adhered to, could result in personal
injury. Do not proceed beyond a WARNING sign until the in dicated conditions are
fully understood and met.
CAUTION:The CAUTION sign denotes a hazard. It calls attention to an operating procedure, or the
like, which, if not correctly performed or adhered to, could result in damage to or
destruction of part or all of the product. Do not proceed beyond a CAUTION sign until the
indicated conditions are fully understood and met.
4
S:\agilent\8920\ALR\MANUAL\front.fb
Safety Symbols
Caution, refer to accompanying documents
Warning, risk of electric shock
Earth (ground) terminal
Alternating current
Frame or chassis terminal
Standby (suppl y). Units with this symbol ar e not completel y disconnect ed from ac
mains when this switch is off.
T o completely disconnect the unit from ac mains, either disconnec t the power cord,
or have a qualified electrician install an external switch.
Product MarkingsCE - the CE mark is a registered trademark of the European Community. A CE
mark accompanied by a year indicated the year the design was proven.
CSA - the CSA mark is a re gistered trademark of the Canadian Standards
Association.
CERTIFICATIONAgilent Tech nol ogi es certifies that this product met it s publ is hed sp ec if ic ations at
the time of shipment fr om the fac tory . Agilent Technologies further cer tifies that it s
calibration measurements are traceable to the United States National Institute of
Standards and Technology, to the extent allowed by the Institute’s calibration
facility, and to the calibration facilities of other International Standards
Organization members
5
Agilent Technologies 08920 RF Communications Test Set
Duration of
Warranty: 1 year
1. Agilent Technologies warrants Agilent Technologies hardware, accessories and
supplies against defects in materials and workmanship for the period specified above.
If Agilent Technologies receives notice of such defects during the warranty period,
Agilent Technologies will, at its option, either repair or replace products which prove
to be defective. Replacement products may be either new or like-new.
2Agilent Technologies warrants that Agilent Technologies software will not fail to exe-
cute its programming instructions, for the period specified above, due to defects in material and workmanship when properly installed and used. If Agilent Technologies
receives notice of such defects during the warranty period, Agilent Technologies will
replace software media which does not execute its programming instructions due to
such defects.
3. Agilent Technologies does not warrant that the operation of Agilent Technologies
products will be uninterrupted or error free. If Agilent Technologies is unable, within
a reasonable time, to repair or replace any product to a condition as warranted,
customer will be entitled to a refund of the purchase price upon prompt return of the
product.
4Agilent Technologies products may contain remanufactured par ts equivalent to new in
performance or may have been subject to incidental use.
5. The warranty period begins on the date of delivery or on the date of installation if
installed by Agilent Technologies. If customer schedules or delays Agilent
Technologies installation more than 30 days after delivery, warranty begins on the 31st
day from delivery.
6Warranty does not apply to defects resulting from (a) improper or in adequate mainte-
nance or calibration, (b) software, interfacing, parts or supplies not supplied by Agilent
Technologies, (c) unauthorized modification or misuse, (d) operation outside of the
published environmental specifications for the product, or (e) improper site preparation
or maintenance.
7TO THE EXTENT ALLOWED BY LOCAL LAW, THE ABOVE WARRANTIES
ARE EXCLUSIVE AND NO OTHER WARRANTYOR CONDITION, WHETHER
WRITTEN OR ORAL IS EXPRESSED OR IMPLIED AND AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS ANY IMPLIED WARRANTIES OR CONDITIONS OR MERCHANTABILITY, SATISFACTORY QUALITY, AND FITNESS
FOR A PARTICULAR PURPOSE.
6
S:\agilent\8920\ALR\MANUAL\front.fb
8. Agilent Technologies will be liable for damage to tangible property per incident up to
the greater of $300,000 or the actual amount paid for the product that is the subject of
the claim, and for damages for bodily injury or death, to the extent that all such damages are determined by a court of competent jurisdiction to have been directly caused
by a defective Agilent Technologies product.
9. TO THE EXTENT ALLOWED BY LOCAL LAW, THE REMEDIES IN THIS
WARRANTY STATEMENT ARE CUSTOMER’S SOLE AND EXCLUSIVE
REMEDIES. EXCEPT AS INDICATED ABOVE, IN NO EVENT WILL AGILENT
TECHNOLOGIES OR ITS SUPPLIERS BE LIABLE FOR LOSS OF DATA OR FOR
DIRECT, SPECIAL, INCIDENTAL, CONSEQUENTIAL (INCLUDING LOST
PROFIT OR DATA), OR OTHER DAMAGE, WHETHER BASED IN CONTRACT,
TORT, OR OTHERWISE.
FOR CONSUMER TRANSACTIONS IN AUSTRALIA AND NEW ZEALAND:
THE WARRANTY TERMS CONTAINED IN THIS STATEMENT, EXCEPT TO
THE EXTENT LAWFULLY PERMITTED, DO NOT EXCLUDE RESTRICT OR
MODIFY AND ARE IN ADDITION TO THE MANDATORY STATUTORY
RIGHTS APPLICABLE TO THE SALE OF THIS PRODUCT TO YOU.
ASSISTANCEProduct maintenance agreements and other customer assistance agreements are
available for Agilent Technologies products. For any assistance, contact your
nearest Agilent Technologies Sales and Service Office.
7
DECLARA TION OF CONFORMITY
according to ISO/IEC Guide 22 and EN 45014
Manufacturer’s Name:
Agilent Technologies
Manufacturer’s Address:
24001 E. Mission Avenue
Liberty Lake, Washington 99019-9599
USA
declares that the product
Product Name:
Model Number:
Product Options:
CDMA Mobile Station Test Set
Agilent Technologies 8920
This declaration covers all options of the above
product.
conforms to the following Product specifications:
Safety:IEC 1010-1:1990+A1+A2/EN 6101 0-1:1 993
EMC:CISPR 11:1990 / EN 55011:1991 Group 1, Class A
EN 500 8 2 - 1 : 199 2
IEC 801-2:1991 - 4 kV CD, 8 kV AD
IEC 801-3:1984 - 3V/m
IEC 801-4:1988 - 0.5 kV Sig. Lines, 1 kV Power Lines
Supplementary Information:
This is a class A product. In a domestic environment this product may cause radio interference in
which case the user may be required to take adequate measures.
This product herewith complies with the requirements of the Low Voltage Directive
73/23/EEC and the EMC Directive 89/336/EEC and carries the CD-marking accordingly
Spokane, Washington USANovember 20, 1995 Vince Roland/Quality Manager
8
S:\agilent\8920\ALR\MANUAL\front.fb
.
9
Table 1Regional Sales Offices
United States of America:
Agilent Technologies
Test and Measurement Call Center
P.O. Box 4026
Englewood, CO 80155-4026
(tel) 1 800 452 4844
Japan:
Agilent T echnologies Japan Ltd.
Measurement Assistance Center
9-1 Takakura-Cho, Hachioji-Shi,
Tokyo 192-8510, Japan
(tel) (81) 456-56-7832
(fax) (81) 426-56-7840
Asia Pacific:
Agilent Technologies
24/F, Cityplaza One,
111 Kings Road,
Taikoo Shing, Hong Kong
Canada:
Agilent Technologies Canada Inc.
5150 Spectrum Way
Mississauga, Ontario
L4W 5G1
(tel) 1 877 894 4414
Latin America:
Agilent Technologies
Latin America Region
Headquarters
5200 Blue Lagoon Drive,
Suite #950
Miami, Florida 33126
U.S. A.
(tel) (305) 267 4245
(fax) (305) 267 4286
Europe:
Agilent Technologies
European Marketing Organization
P.O. Box 999
1180 AZ Amstelveen
The Netherlands
(tel) (3120) 547 9999
Australia/New Zealand:
Agilent Technologies
Australia Pty Ltd.
347 Burwood Highway
Forest Hill, Victoria 3131
Any adjustment, maintenance, or repair of this product must be performed by
qualified personnel. Contact your customer engineer through your local Agilent
Technologies Service Center. Y ou can find a list of local service representatives on
the Web at:
http://www.agilent-tech.com/services/English/index.html
If you do not have access to the Internet, one of these cent ers can direct y ou to your
nearest representative:
United States Test and Measurement Call Center
(To ll free in US)
Europe
Canada
Japan Measurement Assistance Center
Latin America
Australia/New Zealand
Asia-Pacific
(800) 452-4844
(31 20) 547 9 900
(905) 206-4725
(81) 426 56 7832
|(81) 426 56 7840 (FAX)
(305) 267 4288 (FAX)
1 800 629 485 (Australia)
0800 738 378 (New Zealand)
(852) 2599 7777
(852) 2506 9285 (FAX)
11
Manufacturer’s Declaration
This statement is pr ovided t o comply wi th the req uiremen ts of the Ger man Sound
Emission Directive, from 18 January 1991.
This product has a sound pressure emission (at the operator position) < 70 dB(A).
•Sound Pressure Lp < 70 dB(A).
•At Operator Position.
•Normal Operation.
•According to ISO 7779:1988/EN 27779:1991 (Type Test).
Herstellerbescheinigung
Diese Information steht im Zusammenhang mit den Anforderungen der
Maschinenlärminformationsverordnung vom 18 Januar 1991.
•Schalldruckpegel Lp < 70 dB(A).
•Am Arbeitsplatz.
•Normaler Betrieb.
•Nach ISO 7779:1988/EN 27779:1991 (Typprüfung).
12
S:\agilent\8920\ALR\MANUAL\front.fb
Power Cables
Power Cables
Plug Type
Used in the following locations
Peru
Plug Type
Used in the following locations
Switzerland
Plug Descriptions
male/female
Straight/Straight8120-069890 inches, black
Plug Descriptions
male/female
Straight/Straight
Straight/90°
Part #
(cable & plug)
Part #
(cable & p lug)
8120-2104
8120-2296
Cable Descriptions
Cable Descriptions
79 inches, gray
79 inches, gray
13
Power Cables
Power Cables
Plug Type
Used in the following locations
Afghanistan, Albania, Algeria, Angola, Armenia,Austria, Azerbaijan, Azores
Bangladesh, Belgium, Benin, Bolivia,Boznia-Herzegovina, Bulgaria, Burkina Faso, Burma, Burundi,Bye lar us
Cameroon, Canary Islands, Central AfricanRepublic, Chad, Chile, Comoros, Congo, Croatia, Czech
Republic,Czechoslovakia
Denmark, Djibouti
East Germany, Egypt, Estonia, Ethiopia
Finland, France, French Guiana, French IndianOcean Areas
Gabon, Gaza Strip, Georgia, Germany, Gozo, Greece
Hungary
Iceland, Indonesia, Iran, Iraq, Israel, Italy,Ivory Coast
Jordan
Kazakhstan, Korea, Kyrgystan
Latvia, Lebanon, Libya, Lithuania, Luxembourg
Macedonia, Madeira Islands, Malagasy Republic,Mali, Malta, Mauritania, Miquelon, Moldova, Mongolia,
Morocco,Mozambique
Nepal, Netherlands, Netherlands Antilles, Niger,Norway
Oman
Pakistan, Pa ra guay , Po l a nd, Portugal
Rep. South Africa, Romania, Russia, Rwanda
Saudi Arabia (220V), Senegal, Slovak Republic,Slovenia, Somalia, Spain, Spanish Africa, Sri Lanka, St.
Pierre Islands
Sweden, Syria
Tajikistan, Thailand, Togo, Tunisia, Turkey,Turkmenistan
USSR, Ukraine, Uzbekistan
Western Africa, Western Sahara
Yugoslavia
Zaire
Plug Descriptions
male/female
Straight/Straight
Straight/90°
Part #
(cable & plug)
8120-1689
8120-1692
Cable Descriptions
79 inches, mint gray
79 inches, mint gray
14
S:\agilent\8920\ALR\MANUAL\front.fb
Power Cables
Plug Type
Used in the following locations
Peru
Plug Type
Used in the following locations
Switzerland
Plug Descriptions
male/female
Straight/Straight8120-069890 inches, black
Plug Descriptions
male/female
Straight/Straight
Straight/90°
Part #
(cable & plug)
Part #
(cable & p lug)
8120-2104
8120-2296
Cable Descriptions
Cable Descriptions
79 inches, gray
79 inches, gray
15
Power Cables
125V
Plug Type
Plug Descriptions
male/female
Straight/Straight
Straight/90°
Straight/Straight
Part #
(cable & plug)
8120-1378
8120-1521
8120-1751
Used in the following locations
American Samoa
Bahamas, Barbados, Belize, Bermuda, Brazil,
Caicos, Cambodia, Canada, Cayman Islands,Columbia, Costa Rica, Cuba
Dominican Republic
Ecuador, El Salvador
French West Indies
Guam, Guatemala, Guyana
Haiti, Honduras
Jamaica
Korea
Laos, Leeward and Windward Is., Liberia
Mexico, Midway Islands
Nicaragua
Other Pacific Islands
Panama, Philippines, Puerto Rico
Saudi Arabia (115V,127V), Suriname
Taiwan, Tobago, Trinidad, Trust Territories ofPacific Islands
Turks Island
United States
Venezuela, Vietnam, Virgin Islands of the US
Wake Island
Bahrain, British Indian Ocean Terr., Brunei
Canton, Cyprus
Enderbury Island, Equatorial Guinea
Falkland Islands, French Pacific Islands
Gambia, Ghana, Gibraltar, Guinea
Hong Kong
Ireland
Kenya, Kuwait
Macao, Malaysia, Mauritius
Nigeria
Qatar
Seychelles, Sierra Leone, Singapore, SouthernAsia, Southern Pacific Islands, St. Helena, Sudan
Tanzania
Uganda, United Arab Emirates, United Kingdom
Yeman (Aden & Sana)
Descriptions
male/female
90°/Straight
90°/90°
Part # (cable&
plug)
8120-1351
8120-1703
Cable Descriptions
90 inches, mint gray
90 inches, mint gray
Plug Type
Used in the following locations
Argentina, Australia
China (People’s Republic)
New Zealand
Papua New Guinea
Uruguay
Western Samoa
19
Plug Descriptions
male/female
Straight/Straight
Straight/90°
Part #
(cable & p lug)
8120-1369
8120-0696
Cable Descriptions
79 inches, gray
80 inches, gray
Power Cables
This instrument was constructed in as ESD (electro-static discharge) protected environment. This is
because most of the semi-conductor devices used in this instrument are susceptible to damage by
static discharge.
ATTENTION
Static Sensitive
Devices
Depending on the magnitude of the charge, device substrates can punctured or destroyed by contact
or mere proximity of a static charge. The result can caused degradation of device performance, ea rl y
failure, or immediate destruction.
These charges are generated in numerous ways such as simple contact, separation of materials, and
normal motions of persons working with static sensitive devices.
When handling or servicing equip ment containin g static sen sitive devices, adequate pr ecauti ons must
be taken to prevent device damage or destruction.
Only those who are thoroughly familiar with industry accepted techniques for handling static
sensitive devices should attempt to service circuitry with these devices.
In all instances, measures must be taken to prevent static charge build-up on work surfaces and
persons handling the devices.
20
S:\agilent\8920\ALR\MANUAL\front.fb
Contents
1. Introduction
Conventions Used 34
Test Set Description 36
Troubleshooting 38
Repair Process 39
Calibration 40
Service Tools, Equipment and Documentation 41
Self-Support Information 42
Support Contacts 43
Hardware and Firmware Enhancements 44
2. Troubleshooting
Four Basic Steps In Troubleshooting the Test Set 47
More About Step 2 - Run the Self-Test Diagnostic 53
More About Step 3 - Run the Functional Diagnostics 55
3. Repair
Introduction 68
Downloading Calibration Data 69
Troubleshooting Aids for Assemblies 70
Disassembly and Replacement Procedures 74
A23 RF Input Assembly 77
A1 Front Panel (includes keyboard) 79
A11 Receiver Mixer 81
A10 Power Supply Regulator 83
A9 Power Supply 85
A25 Motherboard 87
Fan 89
High-Power Attenuator 91
A20 Display (CRT) 93
A7 Controller 94
A8 Memory Assembly 95
4. P reventative Maintenance
Hardware Maintenance 98
Firmware and Memory Assembly Maintenance 102
5. Calibration
Introduction 104
Performance Tests 116
RF Gen AM Distortion
Performance Test 1 121
21
Contents
RF Gen AM Accuracy
Performance Test 2 123
RF Gen AM Flatness
Performance Test 3 125
RF Gen FM Distortion
Performance Test 4 127
RF Gen FM Accuracy
Performance Test 5 129
RF Gen FM Flatness
Performance Test 6 131
RF Gen Residual FM
Performance Test 7 133
RF Gen Duplex Output High Level Accuracy
Performance Test 8 136
RF Gen Duplex Output Low Level Accuracy
Performance Test 9 138
RF Gen RF IN/OUT Level Accuracy
Performance Test 10 141
RF Gen Harmonics Spectral Purity
Performance Test 11 143
RF Gen Spurious Spectral Purity
Performance Test 12 145
AF Gen AC Level Accuracy
Performance Test 13 147
AF Gen DC Level Accuracy
Performance Test 14 149
AF Gen Residual Distortion
Performance Test 15 151
AF Gen Frequency Accuracy
Performance Test 16 153
AF Analyzer AC Voltage Accuracy
Performance Test 17 155
AF Analyzer Residual Noise
Performance Test 18 157
AF Analyzer Distortion and SINAD Accuracy
Performance Test 19 159
AF Analyzer DC Level Accuracy
Performance Test 20 161
AF Analyzer Frequency Accuracy to 100 kHz
Performance Test 21 163
AF Analyzer Frequency Accuracy at 400 kHz
Performance Test 22 165
22
Contents
Oscilloscope
Performance Test 23 167
RF Analyzer Level Accuracy
Performance Test 24 169
RF Analyzer AM Accuracy
Performance Test 25 174
RF Analyzer AM Distortion
Performance Test 26 177
RF Analyzer Residual AM
Performance Test 27 181
RF Analyzer FM Accuracy
Performance Test 28 183
RF Analyzer FM Distortion
Performance Test 29 186
RF Analyzer FM Bandwidth
Performance Test 30 190
RF Analyzer Residual FM
Performance Test 31 194
RF Analyzer SSB Demodulation
Performance Test 32 196
Spectrum Analyzer Image Rejection
Performance Test 33 201
6. Performance Test Records
8920A Performance Test Record 206
8920A
RF Gen AM Distortion
Performance Test 1 207
8920A
RF Gen AM Accuracy
Performance Test 2 209
8920A
RF Gen AM Flatness
Performance Test 3 211
8920A
RF Gen FM Distortion
Performance Test 4 212
8920A
RF Gen FM Accuracy
Performance Test 5 214
8920A
RF Gen FM Flatness
23
Contents
Performance Test 6 216
8920A
RF Gen Residual FM
Performance Test 7 217
8920A
RF Gen Duplex Output High Level Accuracy
Performance Test 8 219
8920A
RF Gen Duplex Output Low Level Accuracy
Performance Test 9 221
8920A
RF Gen RF IN/OUT Level Accuracy
Performance Test 10 227
8920A
RF Gen Harmonics Spectral Purity
Performance Test 11 230
8920A
RF Gen Spurious Spectral Purity
Performance Test 12 234
8920A
AF Gen AC Level Accuracy
Performance Test 13 236
8920A
AF Gen DC Level Accuracy
Performance Test 14 238
8920A
AF Gen Residual Distortion
Performance Test 15 239
8920A
AF Gen Frequency Accuracy
Performance Test 16 241
8920A
AF Analyzer AC Voltage Accuracy
Performance Test 17 242
8920A
AF Analyzer Residual Noise
Performance Test 18 243
8920A
AF Analyzer Distortion and SINAD Accuracy
Performance Test 19 244
8920A
AF Analyzer DC Level Accuracy
24
Contents
Performance Test 20 245
8920A
AF Analyzer Frequency Accuracy to 100 kHz
Performance Test 21 246
8920A
AF Analyzer Frequency Accuracy at 400 kHz
Performance Test 22 247
8920A
Oscilloscope
Performance Test 23 248
8920A
RF Analyzer Level Accuracy
Performance Test 24 249
8920A
RF Analyzer AM Accuracy
Performance Test 25 250
8920A
RF Analyzer AM Distortion
Performance Test 26 252
8920A
RF Analyzer Residual AM
Performance Test 27 253
8920A
RF Analyzer FM Accuracy
Performance Test 28 254
8920A
RF Analyzer FM Distortion
Performance Test 29 255
8920A
RF Analyzer FM Bandwidth
Performance Test 30 256
8920A
RF Analyzer Residual FM
Performance Test 31 257
8920A
RF Analyzer SSB Demodulation
Performance Test 32 258
8920A
Spectrum Analyzer Image Rejection
Performance Test 33 259
8920B Performance Test Record 261
8920B
25
Contents
RF Gen AM Distortion
Performance Test 1 262
8920B
RF Gen AM Accuracy
Performance Test 2 264
8920B
RF Gen AM Flatness
Performance Test 3 266
8920B
RF Gen FM Distortion
Performance Test 4 267
8920B
RF Gen FM Accuracy
Performance Test 5 269
8920B
RF Gen FM Flatness
Performance Test 6 271
8920B
RF Gen Residual FM
Performance Test 7 272
8920B
RF Gen Duplex Output High Level Accuracy
Performance Test 8 274
8920B
RF Gen Duplex Output Low Level Accuracy
Performance Test 9 276
8920B
RF Gen RF IN/OUT Level Accuracy
Performance Test 10 281
8920B
RF Gen Harmonics Spectral Purity
Performance Test 11 283
8920B
RF Gen Spurious Spectral Purity
Performance Test 12 286
8920B
AF Gen AC Level Accuracy
Performance Test 13 288
8920B
AF Gen DC Level Accuracy
Performance Test 14 290
8920B
26
Contents
AF Gen Residual Distortion
Performance Test 15 291
8920B
AF Gen Frequency Accuracy
Performance Test 16 293
8920B
AF Analyzer AC Voltage Accuracy
Performance Test 17 294
8920B
AF Analyzer Residual Noise
Performance Test 18 295
8920B
AF Analyzer Distortion and SINAD Accuracy
Performance Test 19 296
8920B
AF Analyzer DC Level Accuracy
Performance Test 20 297
8920B
AF Analyzer Frequency Accuracy to 100 kHz
Performance Test 21 298
8920B
AF Analyzer Frequency Accuracy at 400 kHz
Performance Test 22 299
8920B
Oscilloscope
Performance Test 23 300
8920B
RF Analyzer Level Accuracy
Performance Test 24 301
8920B
RF Analyzer AM Accuracy
Performance Test 25 302
8920B
RF Analyzer AM Distortion
Performance Test 26 304
8920B
RF Analyzer Residual AM
Performance Test 27 305
8920B
RF Analyzer FM Accuracy
Performance Test 28 306
8920B
27
Contents
RF Analyzer FM Distortion
Performance Test 29 307
8920B
RF Analyzer FM Bandwidth
Performance Test 30 308
8920B
RF Analyzer Residual FM
Performance Test 31 309
8920B
RF Analyzer SSB Demodulation
Performance Test 32 310
8920B
Spectrum Analyzer Image Rejection
Performance Test 33 311
8921A Performance Test Record 312
8921A
RF Gen FM Distortion
Performance Test 4 313
8921A
RF Gen FM Accuracy
Performance Test 5 315
8921A
RF Gen FM Flatness
Performance Test 6 317
8921A
RF Gen Residual FM
Performance Test 7 318
8921A
RF Gen Duplex Output High Level Accuracy
Performance Test 8 320
8921A
RF Gen Duplex Output Low Level Accuracy
Performance Test 9 322
8921A
RF Gen RF IN/OUT Level Accuracy
Performance Test 10 328
8921A
RF Gen Harmonics Spectral Purity
Performance Test 11 331
8921A
RF Gen Spurious Spectral Purity
Performance Test 12 335
28
Contents
8921A
AF Gen AC Level Accuracy
Performance Test 13 337
8921A
AF Gen DC Level Accuracy
Performance Test 14 339
8921A
AF Gen Residual Distortion
Performance Test 15 340
8921A
AF Gen Frequency Accuracy
Performance Test 16 342
8921A
AF Analyzer AC Voltage Accuracy
Performance Test 17 343
8921A
AF Analyzer Residual Noise
Performance Test 18 344
8921A
AF Analyzer Distortion and SINAD Accuracy
Performance Test 19 345
8921A
AF Analyzer DC Level Accuracy
Performance Test 20 346
8921A
AF Analyzer Frequency Accuracy to 100 kHz
Performance Test 21 347
8921A
AF Analyzer Frequency Accuracy at 400 kHz
Performance Test 22 348
8921A
Oscilloscope
Performance Test 23 349
8921A
RF Analyzer Level Accuracy
Performance Test 24 350
8921A
RF Analyzer FM Accuracy
Performance Test 28 351
8921A
RF Analyzer FM Distortion
Performance Test 29 352
29
Contents
8921A
RF Analyzer FM Bandwidth
Performance Test 30 353
8921A
RF Analyzer Residual FM
Performance Test 31 354
8921A
RF Analyzer SSB Demodulation
Performance Test 32 355
8921A
Spectrum Analyzer Image Rejection
Performance Test 33 356
7. Specifications
8920A 361
Signal Generator Specifications 362
Audio Source Specifications 367
RF Analyzer Specifications 368
AF Analyzer Specifications 372
Oscilloscope Specifications 375
Spectrum Analyzer Specifications (Option 102) 376
Signaling (Option 004) 379
DC Current Meter (Option 103) 380
Remote Programming (Option 103) 381
Reference Oscillator Specifications 382
Save/Recall Registers 383
General Specifications 384
8920B 385
8920B Specifications 386
8921A 427
Signal Generator Specifications 428
Audio Source Specifications 431
RF Analyzer Specifications 432
AF Analyzer Specifications 434
Oscilloscope Specifications 436
Spectrum Analyzer Specifications 437
Signaling 440
DC Current Meter 441
Remote Programming 442
Save/Recall Registers 443
General Specifications 444