Keysight (Agilent) 8751A Service Note

8751A-23
SUPERSEDES: NONE
Continued
DATE: 03 November 1993
8751A Network Analyzer
Firmware Revision: Rev 5.00 only
Individual Internal Tests Verify the Self Diagnosis Failure
To Be Performed By: Customer
Parts Required: None
Situation:
When {SELF DIAGNOSE} softkey is pressed to perform a series of internal tests, the fol­lowing tests may fail in spite of the normal operation:
25: RF PWR LVL ALC (LF) 26: RF PWR LVL ALC (HF) 27: RF AMP FREQ FLATNESS
Solution/Action:
If the above test failures are found in the “SELF DIAGNOSE” execution, perform each individual test to verify the test results by the following procedure:
1. Press [SYSTEM] hardkey and {SERVICE MENU} {TESTS} softkeys.
2. Enter the test number by pressing [2] [5] [x1] hardkeys.
3. Execute the test by pressing {EXECUTE TEST} softkey.
1 of 2
AUTHOR:
KI
ENTITY:
3355
© 1993 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
INFORMATION ONLY
SERVICE NOTE CLASSIFICATION:
ADDITIONAL INFORMATION:
Page 2 Service Note 8751A-23
4. Confirm the test result. If the test result “PASSED” is displayed in the individual internal test, the 8751A actually passes the test. If the test result “FAILED” is displayed in the individual internal test, the 8751A actually failed the test.
5. Execute the other individual tests, “26: RF PWR LVL ALC (HF)” and “28: RF
AMP FREQ FLATNESS”, and confirm the test results.