Keysight (Agilent) 87421A Service Note

87421A-01-S
SUPERSEDES: None
DATE: 21 February 1992
87421A Power Supply
Serial Numbers: 3116A00101 / 3116A00150
A potential shock hazard is present if the rivets fastening the AC power input module cause the plastic case to fracture.
To Be Performed By: Agilent Service Facility
Situation:
The power input module is secured to the plastic case of the power supply by two metal rivets. If improperly installed, these rivets can cause the plastic case to fracture. Should this occur, the power input module may pull out of the housing, exposing a potential shock hazard.
Solution/Action:
Agilent will replace any unit that has metal rivets. Should the housing be found to be bro­ken, remove the AC (mains) power to the power supply at the AC supply outlet.
1 of 2
AUTHOR: LL
ENTITY: 4500
© 1992 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
SAFETY
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBILE UNTIL: ALWAYS
ADDITIONAL INFORMATION:
ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 0.5 Hours
Continued
Page 2 Service Note 87421A-01-S
DO NOT REMOVE THE POWER CORD FROM THE POWER SUPPLY
WHILE STILL PLUGGED INTO THE MAINS.
Return the power supply to the nearest Agilent sales and service office. The power sup­ply will be exchanged for one with black nylon fasteners securing the AC input module.
All units with serial number 3116A00151 and above use black nylon fasteners to secure the AC input module. This service note does not apply to units with black nylon fasten­ers or with serial number 3116A00151 and above.
WARNING