Keysight (Agilent) 8719A Service Note

8719A-02
SUPERSEDES: NONE
DATE: 26 MARCH 1992
8719A Vector Network Analyzers
Serial Numbers: 0000A00000 / 9999A00000
Change to the ALC OFF adjustment in the 8720A/B 8719A Service Manual
Parts Required: None
Some cases of intermittant phase lock failure have been traced to an insufficient level of RF power during retrace. This level is adjusted by the ALCOFF adjustment in the service manual. In a few instruments we have found that the adjustment procedure in the service manual will not result in sufficient power to reliably phase lock under all circumstances.
Solution/Action:
The following procedure replaces step 1 of the AUTOMATIC LEVELING CONTROL ADJUST­MENT in the Adjustments section of the 8719A 8720A/B service manual.
1. Press [PRESET]{SYSTEM}{SERVICE MENU}{ALC OFF}{INPUT PORTS} {R} [MKR]. Slowly adjust the ALC OFF down until a phase lock error occurs. Note the marker value. Adjust ALC OFF up to 6 dB above the marker reading previously noted. Press [MENU]{CW FREQ} and continue with step 2 of the AUTOMATIC LEVELING CONTROL ADJUSTMENT in the adjust­ments section of the 8719A 8720A/B service manual.
1 of 1
AUTHOR:
JK
ENTITY:
4500
© 1992 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
INFORMATION ONLY
SERVICE NOTE CLASSIFICATION:
ADDITIONAL INFORMATION: