Keysight (Agilent) 8713B Service Note

8713B-02
SUPERSEDES: None
DATE: June 1996
8713B Network Analyzers
Serial Numbers:
8711B-02 Serial Numbers: GB35100000 / GB35100176 - except S/N 0175 8712B-02 Serial Numbers: GB35100000 / GB35100145 - except S/N 0143 8713B-02 Serial Numbers: GB35100000 / GB35100200 8714B-02 Serial Numbers: GB35100000 / GB35100178
To Be Performed By: Agilent-Qualified Personnel
Parts Required: Part No. Qty.
8120-2587 3 8120-2586 1
1 of 2
AUTHOR: JV
ENTITY: 5320
© 1996 AGILENT TECHNOLOGIES PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION CATEGORY:
LOCATION CATEGORY:
CUSTOMER INSTALLABLE ON-SITE SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: June 1998
ADDITIONAL INFORMATION:
IMMEDIATELY ON SPECIFIED FAILURE AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN SCRAP SEE TEXT
SERVICE INVENTORY:
RETURN SCRAP SEE TEXT
USED PARTS:
STANDARDS:
Labor 1.0 Hour
Continued
Page 2 Service Note 8713B-02
Situation:
QMD has discovered that the PHONO cables used in the A3 Frac-N board assembly were improperly made and may be intermittent. All 871xB instruments shipped prior to mid-January, 1996 with a “GB” serial prefix within the above range may be affected. These units are generally shipped to European destinations.
These cables are the four RCA phono cables. Examination of these parts reveals insuffi­cient solder at the tips of the connector. Since there are several cables per board, the failure symptoms may be varied. The RF output of analyzer may not change frequency, the trace may be noisy, or excess spurs may exist.
Solution/Action:
For any analyzer in the above range, it is recommended that all four cables on the A3 Frac-N assembly be replaced if any are found to be a problem. There are two different lengths involved; three of one length and one of another. Make sure that each end of each cable is fully seated on the connector.