8711A-03
SUPERSEDES: None
Continued
DATE: 01 April 1993
8711A Network Analyzer
Serial Numbers: 0000A00000 / 9999A99999
Incorrect part in service kit/incorrect procedure in manual
Parts Required: Part No. 08711-60017
Situation:
It was recently discovered that a cable supplied in the 8711 service kit was improperly
designed. While it may work in some cases, it could produce a failure in service test
adjustment #103, External Detector Correction Adjustment.
In addtion, the procedure in the service manual for this adjustment is also incorrect.
Service manual 08711-90003 edition 2 (and before) are incorrect; edition 3 is correct.
Solution/Action:
1. If you already have the service kit (p/n 08711-60010), either re-order the external
detector adapter cable (p/n 08711-60017) or modifiy it as follows.
Add another wire (green) from pin K of the detector connector to an alligator clip. This
will be your ground connector. Also note that some previous versions of this cable may be
miswired. Verify the red wire goes to pin D and the black wire to pin C of the connector.
1 of 2
AUTHOR: JV
ENTITY: 5300
© 1993 AGILENT TECHNOLOGIES
PRINTED IN U.S.A.
SERVICE NOTE
ADMINISTRATIVE INFORMATION
MODIFICATION RECOMMENDED
SERVICE NOTE CLASSIFICATION:
ACTION
CATEGORY:
LOCATION
CATEGORY:
CUSTOMER INSTALLABLE
ON-SITE
SERVICE CENTER
AGILENT RESPONSIBLE UNTIL: 01 April 1994
ADDITIONAL INFORMATION:
IMMEDIATELY
ON SPECIFIED FAILURE
AGREEABLE TIME
AVAILABILITY:
PRODUCT’S SUPPORT LIFE
RETURN
SCRAP
SEE TEXT
SERVICE
INVENTORY:
RETURN
SCRAP
SEE TEXT
USED
PARTS:
STANDARDS:
Labor 0.0 hrs

Page 2 Service Note 8711A-03
2. Change the manual as follows.
Under External Detector Correction Adjustment in section 3 of the manual, change the
CAUTION statement referring to a differential voltage “isolated from earth ground” to
“with respect to earth ground.”
Add to step 5; ...and ground (green wire) to pin K.